Ima Lab Manual
Ima Lab Manual
(BP705P)
Sr.
Title of experiments Page no.
no.
Syllabus
organic compounds
AIM
compounds.
PRINCIPLE
The solvent dissolves the drug substance and exerts an intense influence on the quality and
shape of the UV-visible spectrum. Hence the absorption spectrum of drug substance changes
mostly as per the change of solvent that has been used to dissolve the drug substance. Here,
THEORY
ultraviolet and the full, adjacent visible regions of the electromagnetic spectrum.
Absorption spectroscopy deals with the spectroscopic techniques that measure the
sample. The sample absorbs energy, i.e., photons, from the radiating field. The intensity of
the absorption varies as a function of frequency, and this variation is the absorption spectrum.
presence of a particular substance in a sample and, in many cases, to quantify the amount of
the material present. Infrared and ultraviolet–visible spectroscopies are particularly common
There are a wide range of experimental approaches for measuring absorption spectra.
The most common arrangement is to direct a generated beam of radiation at a sample and
detect the intensity of the radiation that passes through it. The transmitted energy can be used
to calculate the absorption. The source, sample arrangement and detection technique vary
significantly depending on the frequency range and the purpose of the experiment.
Lambda max refers to the wavelength along the absorption spectrum where a substance has
its strongest photon absorption. Simply, the wavelength at which a substance displays
maximum absorption is called as lambda max (figure 1). Different compounds may have very
examined in dilute solution (absorbance value less than 1), so that significant light energy is
received by the detector, and this requires the use of completely transparent (non-absorbing)
solvents. The most commonly used solvents are water, ethanol, hexane and cyclohexane.
Solvents having double or triple bonds, or heavy atoms (e.g. S, Br & I) are generally avoided.
The Beer–Lambert:
The Beer–Lambert law states that the absorbance of a solution is directly proportional to the
concentration of the absorbing species in the solution and the path length.[3] Thus, for a fixed
path length, UV/Vis spectroscopy can be used to determine the concentration of the absorber
in a solution.
Ultraviolet–visible spectrophotometer
spectrophotometer. It measures the intensity of light after passing through a sample and
Choice of solvents
Every solvent is supposed to exhibit UV-vis absorbance cut-off wavelength. The solvent cut-
off is the wavelength below which the solvent itself absorbs all of the light. So when
choosing a solvent student has to be careful of its absorbance cut-off. If the solvent is
showing cut-off near the absorption maxima of the substance under examination, another
solvent is to be chosen.
Water 180
Ethanol 205
Toluene 285
Acetone 329
Benzene 278
Applications:
electrons within the metal atoms can be excited from one electronic state to another.
The colour of metal ion solutions is strongly affected by the presence of other species,
such as certain anions or ligands. For instance, the colour of a dilute solution of
copper sulfate is a very light blue; adding ammonia intensifies the colour and changes
3. Organic compounds, especially those with a high degree of conjugation, also absorb
reaction.
5. Spectroscopic analysis is commonly carried out in solutions but solids and gases may
also be studied.
REQUIREMENTS
Apparatus: Glass beakers, Measuring flasks, Whatmann filter paper, Measuring cylinder,
etc.
Chemicals: Paracetamol, Distilled water, Ethanol, 0.1 N NaOH, 0.1N HCl etc.
PROCEDURE
solvent (2/3 volume), in 100mL of measuring flask, shake well to dissolve completely
and makeup the volume upto mark to prepare 100 ppm of stock solution.
2. Pipette out 1 mL of solution from stock solution and add to 10 mL measuring flask
and make up the volume with fresh solvent to prepare 10 ppm of solution.
Distilled water
Ethanol
0.1 N NaOH
0.1N HCl
RESULTS
The absorption maxima (λmax) of paracetamol in various solvents like distilled water,
ethanol, 0.1 N NaOH, and 0.1N HCl was observed and compiled.
MCQ’s
1. When absorption intensity of compound is decreased it is called
A. Red shift (Bathochromic shift)
C. Hypochromic shift
D. Hyperchromic shift
C. Hypochromic shift
D. Hyperchromic shift
3. If solvent is polar and electron transition is nπ∗ which type of shift are seen in graph?
A. Hypsochromic shift
B. Red shift
C. Blue shift
D. A and C
4. If solvent is non-polar and electron transition is nπ ∗which type of shift are seen in
graph?
A. Hypsochromic shift
B. Red shift
C. Blue shift
D. A and C
5. Which of the following compounds does not absorb light in the UV/visible spectrum?
A. Aspirin
B. Paracetamol
C. Phenobarbitone
D. Chloral hydrate
Questions
In simple terms, the technique maps the interaction between light and matter and
generates what is known as a spectrum. This allows scientists to measure the rate at which
of electrons from the excited state to the ground state, while absorption measures
i. Solvent
ii. sample PH
v. effect of conjugation
Ans:
functional group of some other structural feature of which gives a color to compound.
For example- Nitro group is a chromophore because its presence in a compound gives
yellow color to the compound. But these days the term chromophore is used in a
much broader sense which may be defined as “any group which exhibit absorption of
electromagnetic radiation in a visible or ultra-visible region “It may or may not impart
any color to the compound. Some of the important chromophores are: ethylene,
1. Chromophores in which the groups have π electrons undergo π-π* transitions. For
two types of transitions. i.e., π-π* and n-π*, for examples: - carbonyls, nitriles, azo
(b) AUXOCHROMES: It is a group which itself does not act as a chromophore but when
–NH2 group is attached to benzene ring. Its absorption change from λ max 225 (ɛmax
the red color in the visible spectrum has a longer wavelength than most other colors, the
the blue color in the visible spectrum has a shorter wavelength than most other colors, this
PRINCIPLE
P-hydroxybenzoic acid hydrazide (PHBAH) reacts with dextrose in alkaline medium, which
gives colour products, these colored products can be analysed by analytical technique of
colorimetry. Dextrose reacts with PHBAH reagent and produces colored product; the
THEORY
light. To use this device, different solutions must be made, and a control (usually a mixture of
distilled water and another solution) is first filled into a cuvette and placed inside a
colorimeter to calibrate the machine. Only after the device has been calibrated you can use it
The technique is based upon Beer-Lambert’s law. Overall the process is accomplished
with the measurement of analyte in a colored solution or development of color that appears in
the solution following reaction with specific reagent. In alkaline medium p-hydroxybenzoic
acid hydrazide (PHBAH) reacts with dextrose that give products which can be assayed by
colorimetry. Dextrose reacts with p-hydroxybenzoic acid hydrazide (PHBAH) reagent to give
Applications of colorimeters
1. Besides being valuable for basic research in chemistry laboratories, colorimeters have
many practical applications. For instance, they are used to test for water quality, by
screening for chemicals such as: chlorine, fluoride, cyanide, dissolved oxygen, iron,
2. They are also used to determine the concentrations of plant nutrients (such as
phosphorus, nitrate and ammonia) in the soil or haemoglobin in the blood and to
3. In addition, they are used by the food industry and by manufacturers of paints and
colors in paints and fabrics, to ensure that every batch comes out looking the same.
REQUIREMENTS
hydrazide (PHBAH): 50 ml
PROCEDURE
Alkaline solution
Put 3-5 pellets of sodium hydroxide in about 50 ml of water, stir to mix well.
Diluent
Add 10.0 ml of p-hydroxybenzoic acid hydrazide (PHBAH) solution with 30.0ml of water
Weigh 5.0 mg of dextrose; add in a test tube containing 15.0 ml of water, 2-3 drops of
sodium hydroxide and 5.0 ml of p-hydroxybenzoic acid hydrazide (PHBAH) solution, and
mix well.
gently to mix. Keep the above sample solution for 30 -35 minutes at normal temperature.
Again shake gently, then add 5.0 ml of p-hydroxybenzoic acid hydrazide (PHBAH) coloured
solution and few drops of sodium hydroxide, mix well and keep the test tube in water bath in
boiling mode. After about 5 minutes withdraw test tube from water bath, allow to cool, then
Instrumentation
Before measuring absorbance by colorimeter clean the glass tube with water, and ensure zero
To finalize the light source, take sample solution in glass tube and keep first in blue light
source and repeat with the available light source simultaneously recording absorbance.
Use the light source with highest absorbance and record the absorbance for the followings.
1. Standard stock
Note the above three different absorbance. Now plot the graph of two known solution (1 is
standard stock, 2 is working standard) i.e. absorbance (Y-axis) against concentration (X-
axis), plot a straight line in between two points and find the concentration of unknown sample
CONCLUSION
The chelate formed from the reaction of sucrose and p-hydroxybenzoic acid hydrazide
MCQs
1. What is the name of an instrument used to measure the absorbance of a coloured
compound in solution?
A. Colorimeter.
B. Colourymeter
C. Coulometer.
D. Calorimeter.
2. The dichromate ion absorbs light of wavelength close to 500 nm. Based on this
3. The wavelength of absorption is 495 nm. In what part of the electromagnetic spectrum
A. Ultraviolet-visible.
B. Microwave.
C. Infrared.
D. Radiowave.
A. Light source.
B. Filter.
C. Sample.
D. Photo detector.
5. The plot of absorption verses wavelength for a particular compound is referred to as the
A. Path length.
B. Emission spectrum.
C. Absorption spectrum.
D. None of these.
Questions
Ans: The working principle of the colorimeter is based on Beer-Lambert's law, which
states that the amount of light absorbed by a color solution is directly proportional to the
concentration of the solution and the length of a light path through the solution.
single detector may be used to measure incident and transmitted light, alternately.
The latter design is both cheaper and analytically better, because it eliminates
variations between detectors. The filter is used here to obtain an appropriate range
Colorimeter, experienced person is not the requirement. The method is fast and
convenient to that of the gravimetric or volumetric processes, and they are easily
Ans: Colorimeters are widely used to monitor the growth of a bacterial or yeast culture.
They provide reliable and highly accurate results when used for the assessment of color in
bird plumage. They are used to measure and monitor the color in various foods and
beverages, including vegetable products and sugar. Certain colorimeters can measure the
colors that are used in copy machines, fax machines, and printers. Besides being used for
such as testing water quality by screening chemicals such as chlorine, fluoride, cyanide,
AIM
PRINCIPLE
THEORY
light. To use this device, different solutions must be made, and a control (usually a mixture of
distilled water and another solution) is first filled into a cuvette and placed inside a
colorimeter to calibrate the machine. Only after the device has been calibrated you can use it
The technique is based upon Beer-Lambert’s law. Overall the process is accomplished
with the measurement of analyte in a colored solution or development of color that appears in
Applications of colorimeters
1. Besides being valuable for basic research in chemistry laboratories, colorimeters have
many practical applications. For instance, they are used to test for water quality, by
screening for chemicals such as: chlorine, fluoride, cyanide, dissolved oxygen, iron,
phosphorus, nitrate and ammonia) in the soil or haemoglobin in the blood and to
3. In addition, they are used by the food industry and by manufacturers of paints and
colors in paints and fabrics, to ensure that every batch comes out looking the same.
Requirements:
Chemicals: Sulphanilamide, 0.1 N NaOH solution, Conc. H2SO4, Ethanol Solution (95%v/)
Distilled water.
calorimetrically after derivatization at 510 nm. Visible spectrophotometry is one of the most
by substance in a solution. The instrument that measures the ratio or the function of the ratio
of the intensity of the two beams of light in the visible region is called as colorimeter.
The single point procedure involves the measurement of the absorbance of a sample solution
The standard and sample solutions are prepared in a similar manner; ideally the concentration
The concentration of the substance in the sample is calculated from the proportional
Procedure:
1. Weigh accurately about 0.15 gm of Sulphanilamide and add 1 ml of conc. sulphuric acid
in 100 mL measuring flask. Now add 2 ml of 0.1 N NaOH solutions and make the volume
2. Pipette out 10 ml of this solution and make the volume up to 100 ml with distilled water.
3. Again, pipette out 10 ml of this solution and add 10 ml of ethanol solution 95% v/v and
Observation Table:
1.
2.
Calculation: 𝐴 = 𝑎 × 𝑏 × 𝑐
A = Absorbance
a = Absorptivity
Result
Conclusion:
The quantitative spectrophotometric analysis and the estimation of impurity level in the
A. X rays.
B. Visible light
C. IR Rays.
D. UV light.
A. Analysis of paint.
B. Analysis of ink.
C. Analysis of cosmetic.
D. Composition detection.
A. True
B. False
A. True
B. False
5. Which of the following is the purpose of balance indicator in double beam photometer or
colorimeter?
states that the amount of light absorbed by a color solution is directly proportional to the
concentration of the solution and the length of a light path through the solution.
Beer law: The intensity of beam of monochromatic light decreases exponentially with
Ans: SOURCE OF LIGHT: The visible spectrum ranges from 400nm to 800 nm.
Hence, any lamp source, which gives adequate intensity of radiation over the entire
wavelength region, can be used. The requirements of a source of light for colorimeter
are:
i. Tungsten lamp: As it satisfies the above criteria, this lamp finds its place in most
vacuum bulb similar to the ones used domestically. However, it offers sufficient
intensity.
ii. Carbon arc lamp: For a source of very high intensity, carbon arc lamp can be
i. Absorption filters: These filters are made up of glass, coated with pigments or
they are made up of dyed gelatin. They absorb the unwanted radiation and
ii. Interference Filters: This filter is known as Fabry – Perot filter. The features
include (1): It has dielectric spacer film made up of CaF2, MgF2 or SiO, between
Ans: Monochromators are better and more efficient than filters in converting a
PRINCPLE:
Estimation of combined dosage form such as ibuprofen and paracetamol tablet can be done
using the Vierodt’s method. Here two absorbing drugs, display absorption at the wavelength
of each other, and the concentrations of both the drugs can be analysed by technique of
THEORY
If a sample contains two absorbing drugs (X and Y) each of which absorb at the Lambda Max
of the other, it may be possible to determine both drugs by the technique of simultaneous
respectively.
Two equations are constructed based upon the fact that at lambda 1 and lambda 2 the
1. At lambda 1
2. At lambda 2
Absorptivity: E 1%, 1 cm
A = abc------------------------------- (5).
A - Absorbance
b = path length
c = concentration
a = A/bc------------------------------- (6).
b=1cm
c=1% (1gm/100mL)
Procedure:
After studying solubility of both the drugs in various solvents, select a common solvent in
prepare (400 and 325 µg/mL). Withdraw 0.1mL and dilute upto 10 mL with fresh solvent to
Above two solutions were scanned separately between 400-200 nm. The overlain spectra of
both drugs were recorded. λmax of both the drugs were recorded (λ1 and λ2).
Linearity of Ibu and Para was recorded as function of absorbance (absorbance should be less
than 1). Calibration curve was constructed by plotting absorbance versus concentration.
E (1%, 1cm) values of these drugs were calculated using following formula;
𝐴𝑏𝑠𝑜𝑟𝑏𝑎𝑛𝑐𝑒
E (1%, 1cm) =
𝐶𝑜𝑛𝑐𝑒𝑛𝑡𝑟𝑎𝑡𝑖𝑜𝑛 (𝑔/100𝑚𝐿)
3.
4.
5.
6.
Formula:
𝐀𝟐 𝐚𝐲𝟏 − 𝐀𝟏𝐚𝐲𝟐
𝐂𝐱 (𝐈𝐛𝐮) =
𝐚𝐱𝟐 𝐚𝐲𝟏 − 𝐚𝐱𝟏 𝐚𝐲𝟐
𝐀𝟏 𝐚𝐱𝟐 − 𝐀𝟐𝐚𝐱𝟏
𝐂𝐲 (𝐏𝐚𝐫𝐚) =
𝐚𝐱𝟐 𝐚𝐲𝟏 − 𝐚𝐱𝟏𝐚𝐲𝟐
Where,
ax1 and ax2, are E(1%, 1cm) of Ibu at λ1and λ2, respectively;
Result:
respectively.
MCQs
A. Wavelength selector
B. Detector
C. Signal processors
D. Cuvette
A. 400-800 nm
B. 200-800 nm
C. 25 μm-2.5 μm
D. 2.5 μm – 1mm
A. Spectrometer
B. Photometer
C. both of these
D. none of these
A. Light source
B. Colour
C. Sound
D. none of these
Questions:
1. What is simultaneous equation method of analysis?
Ans: If a sample contains two absorbing drugs (X and Y) each of which absorb at the
Lambda Max of the other, it may be possible to determine both drugs by the technique
applied to estimate drug combinations that contain two drugs or more than two drugs in
Ans: photomultiplier tube: The photomultiplier tube is the most popular detector used
several dynodes. Photons entering the tube strike the cathode resulting in emission of
electrons.
Ans: A colorimeter is generally any tool that characterizes color samples to provide
can measure intensity as a function of the color, or more specifically, the wavelength
of light. There are many kinds of spectrophotometers. Among the most important
distinctions used to classify them are the wavelengths they work with, the
measurement techniques they use, how they acquire a spectrum, and the sources of
spectrophotometers include the spectral bandwidth and linear range. The most
PRINCIPLE
THEORY
ultraviolet and the full, adjacent visible regions of the electromagnetic spectrum.
Absorption spectroscopy deals with the spectroscopic techniques that measure the
sample. The sample absorbs energy, i.e., photons, from the radiating field. The intensity of
the absorption varies as a function of frequency, and this variation is the absorption spectrum.
presence of a particular substance in a sample and, in many cases, to quantify the amount of
the material present. Infrared and ultraviolet–visible spectroscopies are particularly common
REQUIREMENTS
PROCEDURE:
Assay: Weigh and powder 20 tablets. Weigh accurately a quantity of the powder containing
about 0.15 g of Paracetamol, add 50 ml of 0.1 M sodium hydroxide, dilute with 100 ml of
water, shake for 15 minutes and add sufficient water to produce 200.0 ml. Mix, filter and
dilute 10.0 ml of the filtrate to 100.0 ml with water. To 10.0 ml of the resulting solution add
10 ml of 0.1 M sodium hydroxide, dilute to 100.0 ml with water and mix. Measure the
absorbance of the resulting solution at the maximum at about 257 nm (2.4.7). Calculate the
Observations
1. 11. (gm)
2. 12.
3. 13.
4. 14.
5. 15.
6. 16.
7. 17.
8. 18.
9. 19.
10. 20.
0.15 × (𝐴)
x= = _________ gm --------- (B)
0.500
Wavelength Absorbance
257 nm
(𝐵) 10 10
% Dilution factor= × × × 100 = _________%w/v----------------(C)
200 100 100
Formula
Result: The Percent purity of the paracetamol tablet was found to be_____% w/w.
MCQ’s
1. Beer Lambert’s law gives the relation between which of the following?
3. Beer’s law states that the intensity of light decreases with respect to __________
A. Distance
B. Concentration
C. Volume
D. Composition
4. Beer’s law states that the intensity of light decreases with respect to __________
A. Distance
B. Concentration
C. Volume
D. Composition
5. What is the unit of absorbance, which can be derived from Beer Lambert’s law?
A. L mol-1 cm-1
B. L gm-1 cm-1
C. cm
D. No unit
Questions
techniques that need to be conducted while characterizing an analyte. All the major
chromophores. Upon absorbing UV/Vis light, these chromophores get excited from
ground state to a higher energy level, thus giving out characteristic spectra, aiding in
Ans: Deuterium lamps are always used with a Tungsten halogen lamp to allow
quartz Iodine lamps, this measure most effectively in the visible region from 320 -
1100 nm.
Ans: There are four basic components to a simple single beam UV/Vis
Ans: A commonly taught acronym that helps people remember the colors in the
visible spectrum is (ROY G BIV) which stands for Red, Orange, Yellow, Green,
Ans: Yes, The absorbance is directly proportional to the concentration (c) of the
solution of the sample used in the experiment. The absorbance is directly proportional
to the length of the light path (l), which is equal to the width of the cuvette.
Experiment No. 6
Aim:
Principle:
Fluorescence is the phenomenon of emission of radiation when there is transition from singlet
excited state to singlet ground state. The wavelength of absorbed radiation is called as
instrument used to measure fluorescence with a wavelength varying between 240 nm to 800
nm.
Requirement:
Instrument: Fluorometer
3) Take 10 mL of stock solution and dilute to 100 mL with 0.1 N H 2SO4 (100 μg/mL)
4) Again add 1 mL of above solution and dilute to 100 mL with 0.1 N H2SO4 (10 μg/mL)
5) Prepare concentration of 0.5, 1, 1.5, 2, and 2.5 μg/mL from above solution and dilute to
1) Pipette out 1 mL of given sample solution and make up the volume to 10 mL with with
0.1 N H2SO4
2) Switch on the instrument, set the excitation and emission filters at the wavelength 360 to
460 nm respectively
3) Set the Fluorescent Intensity (FI) to 0 % by using 0.1 N H 2SO4 as blank and 100 % by
4) Repeat the same at least for two or more time to avoid instrumental error
6) Plot the graph between the concentration vs FI and determine the concentration of
Observation:
1 0
2 0.5
3 1
4 1.5
5 2
6 2.5
7 unknown
90
80
70
60
50
% FI 40
30
20
10
0
0 0.5 1 1.5 2 2.5
CONC (μg/mL)
Result:
The concentration of quinine sulphate in the given sample was found to be ……. mg/mL
MCQ`s
1) The process of raising electrons in the atoms or molecules to higher energy level is
called?
a) Partition
b) Transition
c) Excitation
d) Derivatization
2) ……….. is the general term applied to the absorption and emission of radiant energy
a) Fluorescence
b) Phosphorescence
c) Luminescence
d) Affinity Chromatography
a) Phosphorescence
b) Fluorescence
c) Adsorption
d) Absorption
4) An increase in the intensity of the light incident on the sample produces proportional
a) Decrease
b) Increase
c) Alternation
d) Conjugation
5) Electron donating groups like -NH2, -OH, etc. often ………. Fluorescence.
a) Increase
b) Decrease
c) Nullify
d) Withdraw
Questions
Ans: Fluorescence is the emission of visible light by a substance that has absorbed
Ans:
i. Photoluminescence
a. Fluorescence
b. Phosphorescence
ii. Chemiluminescence
iii. Bioluminescence
iv. Triboluminescnce
v. Cathodluminescence
vi. Thermoluminescence
3. Define Phosphorescence?
Ans:
i. Temperature
ii. pH
iv. Solvent
v. Adsorption
vi. Conjugation
5. Define Photoluminescence
Ans: The absorption of light is due to the absorption of photon (light) this
Aim:
Requirements:
Instrument: fluorometric
Principle:
certain ions, constituents of the solution itself. These effects may be due to various factors
like Concentration, pH, presence of specific chemical substances, temperature, viscosity, etc.
Types of Quenching
Self-quenching
Chemical quenching
Static quenching
Collisional quenching
Procedure:
Take 10 mL of above solution and dilute to 100 mL with 0.1 N H2SO4 (100 μg/mL).
For final conc. Again add 1 mL of above solution and dilute to 100 mL with 0.1 N H2SO4 (10
μg/mL)
4. For final conc. Again add 1 mL of above solution and dilute to 100 mL with 0.1 N
6. Add 1,2,3,4 and 5 mL of KI solution in each flask and make up the volume to 10 mL.
7. Set the FI to 0% by using 0.1 N H2SO4 as blank and 100 % by using standard solution
(without KI)
1 0
2 1
3 2
4 3
5 4
6 5
Result:
A decrease in fluorescence intensity shall be observed for quinine solution with the addition
MCQ`s
a) Quenching
b) Filtering
c) Excitation
d) Derivatization
a) Excitation
b) Emission
c) A and B
d) None of above
b) Decreases
c) Remain unaffected
d) None of above
a) Colloidal effect
b) Collisional effect
c) Complex formation
d) Chemical changes
5) When conc. increases, fluorescence intensity does not increase proportionally. This
a) Self-Quenching
b) Chemical Quenching
c) Static Quenching
d) Collisional Quenching
Questions
1. Define Quenching.
Ans: In materials science, quenching is the rapid cooling of a work piece in water,
oil or air to obtain certain material properties. A type of heat treating, quenching
occurring.
Ans:
Ans:
i. Photovoltaic Cell
ii. Phototube
4. What is the effect of viscosity in Fluorescent intensity?
Ans:
i. Filter monochromator
ii. Spectrofluorometer
Experiment No. 8
Aim:
Requirement:
Principle:
When a solution of metallic salt is sprayed on the flame, fine droplets are formed. Due to the
thermal energy of the flame, the solvent in the droplets evaporate, leaving behind fine
residue, which are converted to neutral atoms. These neutral atoms are converted to excited
state atoms by the thermal energy of the flame. As the excited state is not stable, these excited
atoms return to ground state, with the emission of radiation of specific wavelength. The
wavelength of the radiation emitted is characteristic of the element and is used to identify the
element. The intensity of the radiation emitted depends upon the conc. of the element
analysed.
Neutral
atoms
3) Prepare a series of standard solutions of 10, 20, 30, 40 and 50 μg/mL concentrations.
kg/cm3
6) Atomize the Flame intensity to 0 % using distilled water with the knob.
7) Atomize the Flame intensity to 100 % using 50 μg/mL standard solutions (highest conc.).
8) Measure the percent flame intensity of all the standard solutions (10, 20, 30, 40 and 50
μg/mL concentrations)
10) From the graph, mark the % flame intensity of the unknown sample and by extrapolating,
Dilutions:
Observation Table:
1 0
2 10
3 20
4 30
5 40
6 50
7 unknown
Result:
The conc. of sodium ion in the given sample of NaCl solution was found to be ……. μg/mL
by Flame photometry
MCQ`s
1) To get higher temperature in the burner, in flame photometer, the following combination
is used …….
Hydrogen/Air
Hydrogen/Oxygen
Acetylene/Air
Acetylene/Oxygen
Lithium
Calcium
Sodium
Beryllium
3) Which of the following is not the requirement of a good flame in flame photometer?
Radiation is absorbed by non-excited atoms in vapour state and are excited to higher
states
To assay of drug
Questions
Ans:
Ans:
v. Optical System
Ans:
Ans:
i. Flame atomization
Aim:
Requirement:
Principle:
When a solution of metallic salt is sprayed on the flame, fine droplets are formed. Due to the
thermal energy of the flame, the solvent in the droplets evaporate, leaving behind fine
residue, which are converted to neutral atoms. These neutral atoms are converted to excited
state atoms by the thermal energy of the flame. As the excited state is not stable, these excited
atoms return to ground state, with the emission of radiation of specific wavelength. The
wavelength of the radiation emitted is characteristic of the element and is used to identify the
element. The intensity of the radiation emitted depends upon the conc. of the element
analysed.
Neutral
atoms
3) Prepare a series of standard solutions of 10, 20, 30, 40 and 50 μg/mL concentrations.
kg/cm3
6) Atomize the Flame intensity to 0 % using distilled water with the knob.
7) Atomize the Flame intensity to 100 % using 50 μg/mL standard solution (highest conc.).
8) Measure the percent flame intensity of all the standard solutions (10, 20, 30, 40 and 50
μg/mL concentrations)
10) From the graph, mark the % flame intensity of the unknown sample and by extrapolating,
Dilutions:
Observation Table:
1 0
2 10
3 20
4 30
5 40
6 50
7 unknown
Result:
The conc. of potassium ion in the given sample of KCl solution was found to be ……. μg/mL
by Flame photometry
MCQ`s
Radiation is absorbed by non-excited atoms in vapour state and are excited to higher
states
Assay of metformin.
Photronic cell
Photovoltaic cell
Photoemissive tube
Chromatogram
5) If Hydrogen and air are used in burner how many °C temp. is produced?
1900°C
2200°C
2300°C
2100°C
Questions
Ans:
Ans:
ii. Quick
iii. Convenient
Ans:
i. Propane
ii. Hydrogen
iii. Acetylene
4. Define Interferences.
Ans:
Differences between the sample & the standard solutions can lead to systemic errors
Ans:
i. Spectral Interferences
v. Anion Interferences
Aim:
Requirement:
Chloride standard solution, 0.1N Silver Nitrate, Sulphate Standard solution, Dilute nitric acid,
Principle:
suspended particles. The intensity of scattered light is proportional to the concentration. The
intensity of scattered light is normally measured at 90º. It can also be measured at any
suspended particles. The intensity of transmitted light is measured at 180º. The intensity of
transmitted light as a function of concentration i.e. when concentration is more, it is less and
Procedure:
Determination of Chloride:
2) Add 10 mL of Dilute HNO3 and make up the volume to 50 mL with distilled water.
3) Add 1 mL of 0.1 N AgNO3 solution and stir immediately to produce the turbidity.
2) Add 10 mL of dilute HNO3 and make up the volume to 50 mL with distilled water
3) Add 1 mL of 0.1 N AgNO3 and stir immediately.
4) Allow the resulting solution to stand for 5 minutes protected from light.
4) Discard the water and take standard opalescence and adjust the reading to 100 %.
Determination of Sulphate:
1) Pipette out 1.5 mL of ethanolic standard sulphate solution into Nessler cylinder.
4) To this add, 15 mL of (10 ppm) standard sulfate solution and 0.15 mL of 5 M acetic acid.
5) Stir immediately with glass rod and allow to stand for 5 minutes.
1) Pipette out 1.5 mL of ethanolic standard sulphate solution into Nessler cylinder.
7) Stir immediately with glass rod and allow to stand for 5 minutes.
4) Discard the water and take standard opalescence and adjust the reading to 100 %.
Observation Table:
For Chloride:
No. Reading
1. Standard 1000
2. Standard
For Sulphates:
No. Reading
1. Standard 1000
2. Standard
Result:
As the NTU range x meter reading of the sample found less than that of the standard solution,
therefore the given sample passes/fails the limit test for sulphate. Similarly, as the NTU range
x meter reading of the sample found less than that of the standard solution, therefore the
MCQ`s
1) Smaller particles undergo scattering to give rise a symmetrical pattern of secondary rays
2) 180
3) 60
4) 45
1) Nessler reagent
3) Barium sulphate
4) Silver chloride
1) Vitamin B12
2) Vitamin A
3) Vitamin B1
4) Vitamin B6
1) 50 times
2) 100 times
3) 80 times
4) 90 times
1) Number of molecules
2) Concentration
3) Molecular weight
4) Polarity of Molecules
Questions
1. Define Quenching.
Ans: In materials science, quenching is the rapid cooling of a work piece in water,
oil or air to obtain certain material properties. A type of heat treating, quenching
occurring.
Ans:
ii. Collisional
iii. Static
iv. Chemical
Ans:
i. Photovoltaic Cell
ii. Phototube
Ans:
i. Filter monochromator
ii. Spectrofluorometer
Experiment No. 11
AIM- To perform the paper chromatography for the separation of amino acids present in the
given sample.
THEORY
The chromatographic techniques used to separate out mixtures of different substances into
their discrete components. All forms of chromatographic techniques work on the same
principle. They all have basic requirements of stationary phase (a solid or a liquid supported
on a solid) and a mobile phase (a liquid or a gas). The mobile phase tides through the
stationary phase and carries the components of the mixture with it. Different components
travel at different rates based on their affinities toward stationary phase and mobile phase. In
paper chromatography, the stationary phase is a very uniform adsorbent paper and the mobile
Retention factor value (Rf Value) is defined the ratio of the distance travelled by the solute to
the distance travelled by solvent. The distance travelled relative to the solvent is called the Rf
value.
REQUIREMENTS
Chemicals: n-butanol, Glacial acetic acid, Distilled water (4:1:5), Amino acids (Tryptophan
PROCEDURE
Solvents system and methods preparation – For solvents system and methods preparation
n-butanol and water are taken in ratios of 4:5 in a flask and allow it to saturate for 24 hours,
by using the separating funnel separate out the n - butanol and water. The saturated n -
butanol and glacial acetic acid both are taken in the 4:1 ratio which can be used as a solvent
other methods of chromatography. The chromatography paper is cut into rectangular strips
and marks a line on the paper with pencil at about 2 cm from the bottom. With the help of
capillary tube, the samples are applied at different points on the starting line and place the
chromatography paper in the developing chamber, which contains the mobile phase. While
placing the paper, it is important that the solvent level should not reach the starting line or the
sample spots and paper shouldn’t touch the walls of the developing chamber. After sometime
the solvent rises up the paper or the stationary phase by capillary action and dissolves the
sample. The components of the sample move along with the solvent in upward direction.
Checkered if the solvent has reached near the top level of chromatography paper, then the
paper is removed when it reaches the top and marked the level with pencil. This level (or)
height is called the “solvent front”. By using UV light, Ninhydrin or iodine vapors examined
The distance moved by tryptophan and threonine is …… cm and ….. cm respectively, and
Rf value of tryptophan is ……
MCQ’s
a. HPLC
b. Ascending
c. Descending
d. Two dimensional
3. When there are too complex mixture are separated which development technique of paper
Chromatography is used?
a. Ascending
b. Two dimensional
c. Descending
d. Radiation
a. 24 hour
b. 6 hour
c. 12 hour
d. 10 hour
5. Which force is control for the separation of the components in descending paper
chromatography?
a. Partition
b. Adsorption
c. Gravity
Questions
1. What do you mean by paper chromatography?
Ans. Paper chromatography is applicable as a separation technique where separation
of a mixture of compound is mainly achieved by a flow of solvent on a
chromatographic paper and stationary phase has bound or absorbed water present with
the cellulose of the chromatographic paper and the mobile phase is an organic solvent
which is immiscible with stationary phase.
2. What is a principle of paper chromatography?
Ans. Paper chromatography is mainly work on principal of the adsorption &
partition. The mechanism of separation in paper chromatography is mainly partition
in Type. So, when a drop of the solute is treating with the solvent on the paper, the
more strong components back while less strong components are move forward.
3. Define Rf value with their formula.
Ans. The relative rate of the movement of solvent and solute is expressed by a term
Rf. It is defined as the ratio of the distance travelled by the compound at
its maximum.
Rf = Distance travelled by solute / Distance travelled by solvents
In many cases it has been observed that the solvent front os run off the end of the
chromatogram. Rx value is the ratio of the distance travelled by a substance to the
distance travelled by a reference standard.
- Ascending
- Descending
- Radial
- Two dimensions
Experiment No. 12
AIM: To perform the separation and analysis of some sugars by using thin layer
chromatography (TLC)
THEORY
The inordinate consideration has been given just too some comparatively rapid techniques of
laboratory tool for routine analytical work. Its low cost effective, ease, and rapidity along
with its capacity for separating and identifying small quantities of compound mixtures make
the technique important tool for research. The components with less affinity towards the
stationary phases are travels rapidly. The major objective of this investigation was to adapt a
processing liquors, thick juice from storage, and beet storage samples.
REQUIREMENTS
Apparatus: Glass beakers, measuring cylinder, developing chamber glass plates etc.
Chemicals: Chloroform, acetic acid, and water, 85% Orthophospheric acid, diphenylamine,
PROCEDURE
For the preformation of TLC practical be prepared following things firstly as Glass plates are
precoated with 0.25 mm dry silica gel. Solvent system which consists of a mixture of
chloroform, acetic acid, and water (3:3.5:0.5) by volume, respectively and spraying agent is
made from 1 gram diphenylamine and 1 ml of aniline in 100 ml acetone. This mixture is
further mixed with 85% orthophosphoric acid prior to use (10: 1 v/v, respectively).
Sample is applied and then the plate is dried in air for approximately 30 minutes. The solvent
system irrigated within the ascending direction in a tight container. The solvent is allowed to
move upward about 12.5 cm, this usually requires 90 minutes. The plate is removed from the
tank. Then it is kept for drying in air for about 30 minutes. The plate is placed back in the
same developing solvent and let the solvent to move in the same direction to the same
distance of 12.5 cm. It takes 45minutes and plates should then be dried for approximately 30
minutes.
RESULTS
The distance moved by glucose .… and …… is …… cm and ….. cm respectively and the
solvent is ……cm.
Rf value of glucose is ……
MCQ’s
a. Travels slowly
b. Travels rapidly
c. A and B
a. Partition coefficient
b. Temperature
c. Capillary action
c. It is very costly.
a. Absorption
b. Adsorption
c. Partition
d. A and B
a. Temperature.
d. Chemical nature
Questions
1. What is a principle of thin layer Chromatography (TLC)
Ans. The principle of TLC is separation is adsorption. When the mixture contains one
or more components is spotted on a TLC of absorbent coated on a chromatographic
plate and introduce into mobile phase, which are continuing distributing. The mobile
phase is flow by capillary action. The components with more affinity travel slow and
that of lesser affinity travel faster. TLC is method of analysis in which the stationary
phase is spread as a thin layer on a rigid supporting plate. The mobile phase a liquid is
allowed to migrate across the stationary of plate. TLC is a solid- liquid form of
Chromatography, where the stationary phase is polar and mobile phase is non polar.
2. Give characteristics of thin layer Chromatography (TLC)
Ans. - It’s a simple and rapid technique
- Low cost technique and takes less time
- Applied for all types of mixture samples
- Detection is easy and not tedious
- More efficiency of separation
- Corrosive spray reagents can be used without any damage to plate
THEORY
The plants leaves contain a different colored pigments generally falling into two categories,
compounds called terpenes and they are tetraterpenes (eight isoprene units). Chlorophylls A
and B are the pigments that make plants look green. Lycopene, the compound responsible for
the red coloring of watermelon, tomatoes and β -carotene, the compound that causes carrots
and apricots to be orange, are examples of carotenoids. Spinach leaves contain chlorophyll a
and b and β -carotene as major pigments as well as smaller amounts of other pigments.
Chlorophyll A and B are same in structure and may not be able to be resolved in this process.
REQUIREMENTS
Apparatus: Round bottom flask, chromatography column pipettes, glass beakers, measuring
cylinder etc.
PROCEDURE
Extraction of the pigments – For Extraction of the pigments takes about 5 grams of leaves
and dried and placed it in a mortar and then pigments are extracted by grinding the leaves
with a pestle with about 5-10 ml of in the ratio 80:20 mixture (v/v), petroleum ether (hexane)
and acetone after it liquid decanted into a 50 ml round bottom flask. A quick filtration is must
require.
Preparation of the column – For Preparation of the column wet pack method is used. The
chromatography column made with plastic tip with frit, the one-way stopcock, and the plastic
funnel. The column is filled with enough alumina to get the required height. The dry alumina
is poured into a beaker and hexane (pet ether) is added. The mixture is swirled and then
poured into the column. The column is tapped gently, so air is not trapped as the alumina
settles. Then it is added with a small amount of sand after the alumina has been settled. The
column should not contain air bubbles and should be homogeneous. Then the solvent level is
Running the column - Using a long pipette, some of the pigment mixture is added directly
onto the sand. Then it is added enough to fill the sand layer with color. Then the stopcock is
opened and let the liquid level falls to the top of the alumina. Gently add petroleum ether to
fill the sand layer. Then the stopcock is opened and let the liquid level fall to the top of the
alumina. These steps are repeated at least three times or until all the colored compounds are
Kindly do not ever let the column run dry, then the open stopcock to allow a drip rate of
around 1 drop per second. First the yellow-orange β-carotene is eluted. As the yellow-orange
colored product is eluted, it was collected in a test tube. When the β-carotene has been
eluted, the elutions of the chlorophylls are eluted by using a more polar solvent. Let the
solvent level fall to the top of the alumina. Gently the column was filled with either pure
acetone combination, it might be able to separate chlorophylls A and B and then the
RESULTS
MCQ’s
1. . What do you mean eluent?
b. is a liquid solution
stationary phase
b. Polar compound
c. Both A and B
3. Chromatogram is a _______________
c. Both A and B
a. Hydrogen bonding
b. London force
a. Reversed chromatography
b. Ion exchange
c. Liquid Chromatography
Questions
1. What is principle of column chromatography?
Ans. The principle behind column chromatography is adsorption, in which a
mixture of components dissolved in the mobile phase is introduced in to
the column and the components move depending on their relative affinities. The
choice of the solvent depends on the solubility characteristics of the mixture.
2. Enlist steps involved in column chromatography?
Ans. - Preparation / Packing of an adsorbent column
- Selection of Solvent system / Mobile phase
- Application of samples
- Sample elution
- Collection and analysis of fraction
3. What are the kinds of column chromatography?
Ans. Five distinct chromatographic methods that are use columns are
gas chromatography (GC), liquid chromatography (LC), Ion exchange
chromatography (SEC) and Chiral Chromatography.
4. Why is Rf useful?
Ans. Rf values can be used to identify unknown chemicals if they can be compared to
a range of reference substances. The Rf value for a particular substance is always the
same if the same solvent and stationary phase are used.
THEORY
infections. Few of them infection symptoms includes as intestinal infections, middle ear
infections, throat infection, pneumonia, traveler’s diarrhea, and overall for respiratory tract
infections. The molecular weight of azithromycin is 748.996 g/mol and molecular formula is
REQUIREMENTS
Apparatus: Volumetric flask, mortar pestle, weighing balance, filter paper, HPLC
instrument
PROCEDURE
Buffer solution - Dissolve 1.16 gm of dipotassium hydrogen phosphate in water and dilute to
about 50 ml of diluent, mix well and make up the volume up to 100 ml by diluent.
Sample solution - Triturate azithromycin tablets in a mortar pestle. Weight 20 mg of
strongly to dissolve complete for about 10 -15 minutes, make up the volume up to 100 ml by
Column size - 18 column (15 mc x 4.6 mm, 5 μm), Column temperature - 70°C, Mobile
phase flow rate - 1 ml/min, Detection wavelength - 215 nm, Injection volume - 100 μl, Run
the HPLC only with mobile phase at least for 10 minutes for conditioning of column.
At the present inject standard solution six times by taking solution from volumetric flask each
time and record the response/area. Similarly inject filtered sample solution six times by
taking solution from volumetric flask each time and record the response/area.
Observation table -
Avg.
Calculation –
RESULTS
The content (mg/tab) of market brand tablet of Azithromycin was found to be ____.
MCQ’s
c. Identification of polysaccharide
d. Ligand-exchange chromatography
a. Affinity chromatography
c. Exclusion chromatography
a. Reverse Osmosis
b. Vacuum Pumping
c. Distillation
d. Sparging
Questions
1. What is principle of HPLC?
Ans. High performance liquid chromatography (HPLC) is an analytical technique
which useful for separation, identification or quantification each component in a
mixture. The mixture is separated using the basic principle of column chromatography
and then identified and quantified.
2. Which types detectors is used in HPLC?
Ans. Following common detectors used in HPLC are as UV-Vis Detectors, Refractive
Index Detector, Fluorescence Detectors, Evaporative Light Scattering Detector, and
Conductivity Detector.
3. Which typescolumn employed in HPLC?
Ans. Normal Phase HPLC Columns
Reverse Phase HPLC Columns
Ion Exchange HPLC Columns
Size Exclusion HPLC Columns
4. What are troubleshooting errors in HPLC?
Ans. Pumping system is common problem occurs and some of the more common
indications are erratic retention times, noisy baselines, or spikes in the
chromatogram. Leaks at pump fittings or seals will result in poor chromatogram.
5. Gives process for preparation of standard and sample solution in HPLC
Ans. Standard solution - Weigh 20 mg sample in 100 ml volumetric flask containing
about 50 ml of diluent, mix well and make up the volume up to 100 ml by diluent.
Sample solution - Triturate the sample in a mortar pestle and weight 20 mg of
powdered sample in a 100 ml volumetric flask containing about 50 ml of diluent,
shake strongly to dissolve complete for about 10 -15 minutes, make up the volume up to 100
ml by diluent. Filter sample solution by 0.45 μm filter paper.
Experiment No. 15
THEORY
This technique is generally used to separate the gases in gaseous solutions. The more mutual
technique is gas - liquid chromatography in which the stationary phase is a porous solid
covered with an absorbing the liquid. Gas-liquid chromatography (GLC) is commonly used
for separation of wide variety of organic compounds. The basic requirements for GLC are
that the sample may be volatile and that it not decomposes in all vaporization process.
A separating column consisting with metal tubing packed with a solid material coated
A flow control equipment for the maintenance a constant flow of carrier gas
throughout column.
A hot air ovens and heaters for the column, detector and injector temperature control.
An integrator and strip chart recorder combination to provide the maintain record of
these analysis.
The separation of different mixtures into its components depends upon the solubility
differences of sample vapour in a liquid (stationary phase), and generally stationary phase is
coated in a thin layer on the solid particles of large surface area and then packed uniformly
into the column. A constant flow of an inert carrier gas badges through the column and
transports solute molecules in the gas phase. The column is enclosed by an oven for precise
temperature control.
A sample of the analyte is introduced by syringe injection into the heated injector tube, where
it is vaporized and mixed with the carrier gas. As the sample vapor/carrier gas mixture flows
onto the column, the analyte partitions between the gas and liquid phases according to the
analyte component's solubility in the liquid at the column operating temperature. This
equilibrium partitioning continues as the sample is moved through the column by the carrier
gas. The rate at which the sample travels through the column is determined by the sample
solubility in the stationary phase, the carrier gas flow rate, and the temperature gradient
(temperature program) applied. Each component travels at a characteristic rate, and if the
column has sufficient length and resolving power, the sample will be completely separated by
the time it reaches the detector and identified by comparing the "retention time".
INSTRUMENTATIONS
Figure – Instrumentation of Gas Chromatography
1. Carrier gas in a high-pressure cylinder with attendant pressure regulators and flow
meters - In Gas Chromatography Helium, N2, H, and Argon are used as carrier gases in
its high thermal conductivity relative to that of most organic vapors. N2 is preferable
when a large consumption of carrier gas is employed. Carrier gas from the tank passes
through a toggle valve, a flow meter, (1-1000 ml/min), capillary restrictors, and a
pressure gauge (1-4 atm). Flow rate is adjusted by means of a needle valve mounted on
the base of the flow meter and controlled by capillary restrictors. The operating
gas flow.
2. Sample injection system – In sample injection system liquid samples are injected by a
micro syringe with a needle inserted through a self-scaling, silicon-rubber septum into a
heated metal block by a resistance heater. Gaseous samples are injected by a gas-tight
syringe or through a by-pass loop and valves. Typical sample volumes range from 0.1 to
0.2 ml.
3. The column separation - The heart of the gas chromatography is the column which is
made of metals bent in U shape or coiled into an open spiral or a flat pancake shape and
copper is very useful up to 2500 . Swege lock fittings make column insertion easy.
4. Liquid phases - An infinite variety of liquid phases are available limited only by their
volatility, thermal stability and ability to wet the support. No single phase will serve for
points.
on a long non-polar skeleton which can dissolve both polar and non-polar
solutes. For example. Diethyl hexyl phthalate is used for the separation of high
boiling alcohols.
- Polar – Carbo waxes – Liquid phases with a large proportion of polar groups.
- Hydrogen bonding – Polar liquid phases with high hydrogen bonding e.g.
Glycol.
5. Supports - The structure and surface characteristics of the support materials they are
very important parameters, which determines the efficiency of the support and the
immobilizing a large volume of liquid phase as a thin film over its surface. The surface
area should be large to ensure the rapid attainment of equilibrium between stationary and
mobile phases. Support should be strong enough to resist breakdown in handling and be
capable of packed into a uniform bed. Diatomaceous earth, Kieselguhr treated with Na
2CO 3 for 9000 C causes the particle fusion into coarser aggregates. Glass beads with a
low surface area and low porosity can be used to coat up to 3% stationary phases. Porous
close to the column exit and the correct temperature to prevent decomposition.
7. Recorder - The recorder should be generally 10 mv (full scale) fitted with a fast
response pen (1 sec or less). The recorder should be connected with a series of good
quality resistances connected across the input to attenuate the large signals. An integrator
APPLICATIONS
example as promising the quality excellence of products in the industry and for
3. Gas chromatography is very precise if used correctly and for measurement pico-
4. Gas Chromatography also used widely in the forensic science as disciplines as diverse
as solid drug dose qualitative and quantitative analysis, paint chip analysis, and
MCQ’s
1. Which method is useful for separation of thermally stable and volatile substance?
a. Gas Chromatography.
c. Mass Spectroscopy
a. Kieselguhr
b. Diatomaceous earth
c. A and B
a. Alumina
b. Carbon
c. Granular silica
5. Which of the following is not a feature of carrier gas used in gas chromatography?
d. It should be cheap
Questions
1. What is principle of Gas Chromatography?
Ans. The principle of gas chromatography is the sample solution injected into the
instrument enters a gas (stream) which transport the sample into a separation
tube is known as the "column", for that Helium or nitrogen is used as the so-called
carrier gas. The various components and mixture of components are separated
inside the column.
2. Give applications of Gas Chromatography.
Ans. Gas Chromatography is useful to compute the content of a chemical product;
Gas chromatography is also valuable for analysis of: Air borne contaminants, Recital
performance improving drugs in urine samples, oil spills and essential oils in
perfumes. Gas chromatography also used correctly and for measurement pico-
moles of a substance Gas Chromatography widely applicable in the forensic science
as disciplines as diverse as solid drug dose qualitative and quantitative analysis, paint
chip analysis, and toxicology cases, employ gas chromatography to crime-scene
evidence.
Molecu
Polar Boiling p Melting p solubility
lar density
Solvent ity oint oint in H2O1
formul (g/mL)
index (o C) (o C) (g/100g)
a
cyclohexane C6H12 0.006 80.7 6.6 0.779 0.005
pentane C5H12 0.009 36.1 -129.7 0.626 0.0039
hexane C6H14 0.009 69 -95 0.655 0.0014
heptane C7H16 0.012 98 -90.6 0.684 0.0003
carbon tetrachlorid
CCl4 0.052 76.7 -22.4 1.594 0.08
e
carbon disulfide CS2 0.065 46.3 -111.6 1.263 0.2
p-xylene C8H10 0.074 138.3 13.3 0.861 0.02
toluene C7H8 0.099 110.6 -93 0.867 0.05
benzene C6H6 0.111 80.1 5.5 0.879 0.18
ether C4H10O 0.117 34.6 -116.3 0.713 7.5
methyl t-
butyl ether C5H12O 0.124 55.2 -109 0.741 4.8
(MTBE)
diethylamine C4H11N 0.145 56.3 -48 0.706 M
dioxane C4H8O2 0.164 101.1 11.8 1.033 M
N,N-
C8H11N 0.179 194.2 2.4 0.956 0.14
dimethylaniline
chlorobenzene C6H5Cl 0.188 132 -45.6 1.106 0.05
anisole C7H8O 0.198 153.7 -37.5 0.996 0.10
tetrahydrofuran (T
C4H8O 0.207 66 -108.4 0.886 30
HF)
ethyl acetate C4H8O2 0.228 77 -83.6 0.894 8.7
C9H10O
ethyl benzoate 0.228 213 -34.6 1.047 0.07
2
dimethoxyethane C4H10O
0.231 85 -58 0.868 M
(glyme) 2
C6H14O
diglyme 0.244 162 -64 0.945 M
3
methyl acetate C3H6O2 0.253 56.9 -98.1 0.933 24.4
chloroform CHCl3 0.259 61.2 -63.5 1.498 0.8
3-pentanone C5H12O 0.265 101.7 -39.8 0.814 3.4
C2H4Cl
1,1-dichloroethane 0.269 57.3 -97.0 1.176 0.5
2
di-n-butyl C16H22
0.272 340 -35 1.049 0.0011
phthalate O4
cyclohexanone C6H10O 0.281 155.6 -16.4 0.948 2.3
pyridine C5H5N 0.302 115.5 -42 0.982 M
C10H10
dimethylphthalate 0.309 283.8 1 1.190 0.43
O4
methylene chloride CH2Cl2 0.309 39.8 -96.7 1.326 1.32
2-pentanone C5H10O 0.321 102.3 -76.9 0.809 4.3
2-butanone C4H8O 0.327 79.6 -86.3 0.805 25.6
C2H4Cl
1,2-dichloroethane 0.327 83.5 -35.4 1.235 0.87
2
benzonitrile C7H5N 0.333 205 -13 0.996 0.2
acetone C3H6O 0.355 56.2 -94.3 0.786 M
dimethylformamid C3H7N
0.386 153 -61 0.944 M
e (DMF) O
t-butyl alcohol C4H10O 0.389 82.2 25.5 0.786 M
aniline C6H7N 0.420 184.4 -6.0 1.022 3.4
dimethylsulfoxide C2H6O
0.444 189 18.4 1.092 M
(DMSO) S
acetonitrile C2H3N 0.460 81.6 -46 0.786 M
3-pentanol C5H12O 0.463 115.3 -8 0.821 5.1
2-pentanol C5H12O 0.488 119.0 -50 0.810 4.5
2-butanol C4H10O 0.506 99.5 -114.7 0.808 18.1
cyclohexanol C6H12O 0.509 161.1 25.2 0.962 4.2
1-octanol C8H18O 0.537 194.4 -15 0.827 0.096
2-propanol C3H8O 0.546 82.4 -88.5 0.785 M
1-heptanol C7H16O 0.549 176.4 -35 0.819 0.17
i-butanol C4H10O 0.552 107.9 -108.2 0.803 8.5
1-hexanol C6H14O 0.559 158 -46.7 0.814 0.59
1-pentanol C5H12O 0.568 138.0 -78.2 0.814 2.2
acetyl acetone C5H8O2 0.571 140.4 -23 0.975 16
C6H10O
ethyl acetoacetate 0.577 180.4 -80 1.028 2.9
3
1-butanol C4H10O 0.586 117.6 -89.5 0.81 7.7
benzyl alcohol C7H8O 0.608 205.4 -15.3 1.042 3.5
1-propanol C3H8O 0.617 97 -126 0.803 M
acetic acid C2H4O2 0.648 118 16.6 1.049 M
C2H7N
2-aminoethanol 0.651 170.9 10.5 1.018 M
O
ethanol C2H6O 0.654 78.5 -114.1 0.789 M
C4H10O
diethylene glycol 0.713 245 -10 1.118 M
3
methanol CH4O 0.762 64.6 -98 0.791 M
ethylene glycol C2H6O2 0.790 197 -13 1.115 M
glycerin C3H8O3 0.812 290 17.8 1.261 M
water, heavy D2O 0.991 101.3 4 1.107 M
water H2O 1.000 100.00 0.00 0.998 M