Replacements of Parts
AXIOM Iconos R200
© Siemens
Siemens,, 2010
2010
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.
Copyright
“© Siemens, 2010“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany,
Germany, Siemens Shenzhen Magnetic Resonance
Resona nce Ltd. - China, Siemens
Shanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA,
Siemens Healthcare Diagnostics Inc. - USA and/or Siemens Healthcare Diagnostics Prod-
ucts GmbH - Germany.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout.
p rintout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.
http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel
p ersonnel who are specially
trained for such installation and/or service.
Table of Contents 3
0TableofContents
1 ____
_______
___ Removing
Removing the Luminos
Luminos dRF Unit Support
Support ____
________
_________
_________
________
_________
______
_ 4
Preparatory Measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Cover Panels
Panels on the Lifting Base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Cover Panels on the Extender . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Cover Panels
Panels on the Unit Support
Suppor t . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Unit Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Installing the Transport
Transport Carriage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2 ____
_______
___ Removing
Removing the Iconos
Iconos R200
R200 Unit Support
Support ____
_________
_________
________
________
________
______
__ 15
Cover Panels
Panels on the Lifting Base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Cover Panels on the Extender . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Cover Panels
Panels on the Unit Support
Suppor t . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Unit Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Installing the Transport
Transport Carriage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3 ____
_______
___ Removing
Removing the
the Lifting
Lifting Base ____
________
________
_________
_________
_________
_________
________
_________
_____ 25
4 ____
_______
___ Installatio
Installation
n and Adjustment
Adjustment ____
________
_________
_________
________
________
________
________
_________
_____ 27
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Adjusting the Safety Limit Switches: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Adjusting the Potentiometers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Adjustments with the Luminos dRF with Bucky Wall Stand . . . . . . . . . . . . . . . . . . . 28
Installing the Cover Panels
Panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5 ____
_______
___ ____
________
_________
_________
_________
__________
_________
_________
_________
_________
_________
________
_________
_________
____ 30
6 ____
_______
___ List of Hazard
Hazard IDs ____
________
________
________
________
_________
_________
________
________
________
_________
_____ 31
4 Removing the Luminos dRF Unit Support
1-
Preparatory Measures
1Removingthe Luminos dRFUnitSupport
If the Luminos dRF system has a Bucky wall stand, the position of the lifting base must be
marked as exactly as possible on the floor.
The new lifting base must then be installed again in exactly (millimeter accuracy) the same
position so that center of the tube <-> detector center exactly match.
It is absolutely necessary to work very accurately here, because correction of a difference
takes a lot of time.
Removing the Luminos dRF Unit Support 5
1. Remove the footrest and the other parts on the patient table.
2. Remove the side cover panels on the unit support(1-3/Fig. 4 / p. 7).
3. Remove the fan cover panel at the head end on the unit support(1/Fig. 5 / p. 7).
Fig. 4: Cover panels at foot end Fig. 5: Cover panels at head end
20 Removing the Iconos R200 Unit Support
• Unplug the connections between the lifting base and the M1 assembly (line power
assembly).
Fig. 26:
• Remove the contact protector from the F1 trip breaker (1/Fig. 27 / p. 20).
• Label the motor cable to the Micromasters (1/Fig. 28 / p. 21) and disconnect it.
Fig. 28:
• Remove the transformer (2/Fig. 28 / p. 21). It will be used again.
• Pull all cables to the right side of the base in the direction of the transformer out of the
base. Be careful of the connectors (Fig. 26 / p. 20).
• Remove the support plate (3/Fig. 28 / p. 21) with the Micromasters. Take note of the
cabling! The support plate and the Micromaster will be used again.
22 Removing the Iconos R200 Unit Support
1. Install the plates at the head and foot ends of the unit support and tighten all screws.
2. Install the transport carriages at the head and foot ends (thread the screws in only
loosely) and make sure that the safety screws to prevent tilting of the unit support are
inserted:
3. Install the transverse strut ((1/Fig. 33 / p. 23) and (1/Fig. 34 / p. 23)) and angled struts (
(2/Fig. 33 / p. 23) and (2/Fig. 34 / p. 23)) (thread in the screws only loosely). Prior to
installation, the transverse struts must be connected together (tighten the screws).
Fig. 33: Transverse and angled strut, head Fig. 34: Transverse and angled strut, foot end
end
4. First tighten all screws on the transport carriage and then the screws for the transverse
strut/struts and angled struts (if installed).
5. Install the transport castors at the head and foot ends.
6. Turn down the transport castors with the spindles so that the transport castors are
under light pressure.
7. Remove the screws between the unit support and the extender.
8. Evenly turn down the transport castors with the spindles so that the unit support and
extender come apart.
9. Move the unit support so that the base is accessible from the front and side.Pay care-