Hyc-55 Opr
Hyc-55 Opr
Applicable model:
HYC-55 / 75
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
MRC.9.21
Content
1
1. INTRODUCTION
2
2. PRECAUTIONS FOR SAFE OPERATION
4
3. PRECAUTIONS FOR USE
4
4. Installation
4
4.1 Installation environment:
4
4.2 Installation site:
5
4.3 Cautions before use
4.4 Standing 5
5. Refrigerator Components 6
11
7. Defrost ●Unused ● Maintenance
7.1 Defrost, Unused and Maintenance 11
7.2 Disposal 12
9. Specifications 15
Please read the page4-page7 carefully as it contains the important safety notices. Only
trained or authorized personnel could operate this medical refrigerator.
Only qualified service personnel or authorized agent could install and maintenance this
device.
Use MRC spare parts as always. If users want to use other accessories, MRC
bio-medical will not be responsible for the adverse consequences. However users could
apply for verifying the eligibility of these accessories from MRC before use them.
Should inspect and maintenance the device in a specified time interval.
Due to the differences between each models and the improvements of products, actual
product may differ from the diagram. Please refer to the final product.
Everyone has an obligation to be responsible for his or her own safety.
Put on dry gloves when you take out refrigerated articles from the refrigerator.
Handing frozen contents or the inside walls with naked hands may cause frostbit.
1
2. PRECAUTIONS FOR SAFE OPERATION
For the first time to use this device, uses must read the meaning of the following warning labels
very carefully. Items and procedures are described so that you can use this unit correctly and safely.
If the precautions advised are followed, this will prevent possible injury to the user and any other
person.
WARNING: CAUTION.
Failure to observe WARNING signs could result in a hazard to Failure to observe WARNING signs could result in injury to
personnel possibly resulting in serious injury or death. personnel and damage to the unit and associated property.
WARNING.
Failure to observe WARNING signs could result in a hazard to personnel possibly resulting in serious injury
or death.
As with any equipment that uses CO2 gas, there is a likelihood of oxygen depletion in the vicinity of the
equipment.It is important that you assess the work site to ensure there is suitable and sufficientventilation.If
restricted ventilation is suspected, then other methods of ensuring asafe environment mustbe considered.
These may include atmosphere monitoring and warning devices.
Do not touch any electrical parts such as the power supply plug or any switches with a wet hand.This may
cause electric shock.
Be sure to install the unit on a sturdy floor. If the floor is not strong enough or the installation site is not
adequate, this may result in injury from the unit falling or tipping over
Carefully with the power cord to avoid short circuit or open circuit.When removing the plug from the power
supply outlet, grip the power supply plug, not the cord.Pulling the cord may result in electric shock or fire by
short circuit. Don't make the power line pack and pressed by furnish or heavy goods.Also please don't close
to the compressor and heat source.
Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply outlet is not
grounded, it will be necessary to install a ground by qualified engineers.
Don't lengthen the line randomly. If you need, To use 2.5mm2 copper line,you should keep 4mm2 line to
connect the electrical outlet. Or may cause fire.
Make sure a dedicated power source is used as indicated on the rating label attached to the unit. Out of the
rate, should install a property transformer and a proper voltage stabilizer for securely operation. Or the
refrigerator may be damaged, and may cause injury.
Never install the unit in a flammable or volatile location. This may cause explosion or fire.
Never install the unit in a humid place or outdoor or a place where it is likely to be basked
straightly.Deterioration of the insulation may result which could cause current leakage or electric shock.
Do not place the device later al tilt, do not impact the device; the device is equipped with
refrigeration systems, roll or shock will easily damage the refrigerator.
2
Be sure to install the device in a dry dust-free environment to avoid overheating, short circuit and other dangers
If there is an unexpected sound, smell, smoke when the power is turned on, unplug the power and contact
the manufacturer or supplier.Continuedabnormal operation may cause electric shock or fire.
Make sure to put the refrigerator in a dry and ventilated environment, to ensure that equipment vents and
wall surface of the instrument or other items have not been blocked; Do not use the device in a poorly
ventilated environment, or the equipment may be damaged by the release of heat.
Never disassemble, repair, or modify the unit yourself.
Any such work carried out by anunauthorized person may result in fire or injury due to a malfunction. Meiing
will be no responsible for such work.
Never store volatile or flammable substances in this unit. This may cause explosion or fire.Never store
corrosive substances in this unit. This may lead to damage to the inner componentsor electric parts.
Use this unit in safe area when treating the poison, harmful or radiate articles. Improper use may cause bad
effect on your health or environment.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such grounding may
cause electric shock in the case of an incomplete circuit.
Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply outlet is not
grounded, it will be necessary to install a ground by qualified engineers
CAUTION:
Failure to observe WARNING signs could result in injury to personnel and damage to the unit and associated
property.
The medical refrigerator is not available to store non-living things, flowers, or other critical articles which
is not suitable for low temperature storage.
The temperature inside the refrigerator is very low during the normal working. Do not touch the interior
surface of the chamber or the object inside without wearing protective gear.
Always disconnect the power plug when the unit is not used for long periods.
Make sure to prepare a safety check sheet when you request any repair or maintenance for the safety of
service personnel. Be sure to check set point of the controller prior to restart the refrigerator.
Always hold the handle when closing the door. This will reduce the likelihood of a trapped finger.
Keep the key properly avoiding the children take it to open the back door which may result in unexpected injury.
Select a level and sturdy floor for installation. This precaution will prevent the unit from tipping.
Improper installation may result in water spillage or injury from the unit tipping over.
Check the filter mentioned in this manual and clean it as necessary. A dusty filter may cause.
Do not tilt the unit more than 45 degrees when moving the unit. All transportation should be carefully.
3
3. PRECAUTIONS FOR USE
◆ Before you put the items to the medical refrigerator, please make sure whether the temperature in the refrigerator has
reached the set temperature and then put the items to the refrigerator in batches. No more than 1/3 of the refrigerator rating
volume per time, in case the temperature rises too much.
◆ Medical refrigerator temperature display value shows the temperature around the temperature sensors. Sometimes, there
is certain difference between the display temperature and the actual temperature in the center of the refrigerator, but it will
gradually close to the real temperature.
◆ Clean the refrigerator with the diluted neutral detergent. Do not use brushes, acid, gasoline, soap powder, polish or hot
water to clean the medical refrigerator, as the above material may damage the painted surface and plastic rubber parts. Be
careful not to use volatile solvents such as gasoline to wipe plastic rubber parts.
◆ The power should be cut off while the medical refrigerator does not use for a long time.
4. Installation
4.1 Installation environment:
Ambient temperature:16℃ ~ 32℃ ,optimal ambient temperature :18℃ ~ 25℃ , Air conditioning
system is required if necessary.
Relative humidity:≤80%RH.
No strong vibration and no corrosive gases around.
Without the presence of a lot of dust .
No shaking or vibrating of the refrigerator.
Altitude of the place where the refrigerator is located :≤ 2000m
Input voltage ≤220+10% (V).
No direct sunlight or any other cooling or heating source, no electromagnetic interference, or the refrigerator
will not run properly.
4
Warning
● Dot through gas pipe, water pipe, line or lightning rod to medical grounding cooler, easy cause electric shock
● After installation, power plug must be within reach, convenient unplug the power cord in time in case of an emergency
● Any articles shall not cover medical air vents of the medical refrigerator
4.3 Cautions before use
a. Remove all package components(include the protection foam inside the package)
Caution:Do not put the packing plastic bag within reach of children as suffocation may result
b. Check the device, accessories and document with the device as per list of packing
c. Clean:make a clean of the device before use it.
d. Before use, make sure the temperature control probe is immersed in the test liquid.
4.4 Standing
Do not turn on the refrigerator at once when the refrigerator is well located, must wait 24 hours for the first start.
5
5.Refrigerator Components
Display panel
Test port
Door
Cabinet
Adjustable
foot RS485
Power switch
Remote alarm
As the product is improved and the model is different, the actual product may be different from the simple diagram.
The above picture is only the example of identifier function piece. Structure and composition: the product consists of a
box, a glass door structure, a refrigeration system and a control system. Scope of application: This product is suitable for
hospitals, pharmacies, epidemic prevention stations, research institutions, biopharmaceutical departments and so on.
6
6. Display panel function description
6 7 8
7
h. Print indicator
When the printer is not working, the indicator lights out; when the printer is working, the indicator light is lit.
i. Serial port indicator
When the refrigerator is not connected to the reserved RS-485 serial port, the serial port light is extinguished; when the
refrigerator is successfully connected to the RS-485 serial port, the serial port light is lit.
j. Fan indicator lamp
When the evaporator is opened, the fan indicator is opened; when the evaporator is closed, the fan indicator is
extinguished.
k. Mute indicator lamp
When the alarm tone is silent, the indicator light is lit; when the alarm sound function is cancelled, the indicator light is
extinguished.
1. temperature display window, in normal running state, display the average temperature inside the box, the unit is
centigrade.
8
The default lamp on the refrigerator is closed. The key is to press the light, and the turning light can be turned out. When
the lamp is closed, the lamp is lit and the lamp closes when the door is closed; when the light is on, the light does not change
when the door is closed. This machine has two functions of automatic lighting and manual lighting. In the automatic
lighting condition, the door opened the door immediately lights lit, turn off the lights after 5 seconds off; the user can
manually control the lights turn on the lights, not switch the door has been lit (indicating lamp button above the light), only
once again press the key to open the lamp lighting lamp can turn off manually enter the automatic lighting state after
automatic lighting (indicator button above out).
7. USB data export;
Automatic export: when the U disk is connected to the USB interface, the recorder buzzer calls once, displaying "on",
and the PDF files in the U disk are generated during the month and the last month. After the data transmission is completed,
the buzzer calls once, displaying "End", and then returning to normal display after 6S.
Note: when the data is less, "on" and "End" are not displayed on the digital tube.
Manually export: button keep unlocked, U disk connected, and not in the file generation, press the button 3 seconds,
digital tube display "D01", according to the raised key or down button to adjust the "d00~d12" . press button, U disk
made this file (D00) or push forward the month (1-12) PDF document data.
Note: when the digital tube alarm flickers "LoF", the recorder is not started;press at the same time, the "LoF"
is disappearing for 3 seconds ,the buzzer sounded once and the recorder is started.
(2) functional settings:
A. connect the power supply, open the back power switch of the back of the box, and the machine can enter the working
state.
B. user parameter settings:
To unlock: under normal operating conditions, and at the same time press for 3 seconds, the digital display
parameters
The code "000", enter the password "005" (the input menu user password, enter "099" can be the key to restore the
default lock password "005".) It is unlocked at this time. After unlocking press for 3 seconds, digital display
The display parameter code "PS1", enter the set parameters.
Use to adjust
Display→PS1→ b1→ b2→Set→ H→ L
n→y→ r→ S→ F→Pt→tH1→ P1→ P2
Press to confirm the parameters, the first parameter name of this type of parameters will be displayed;
(1) rolling parameters with or ;
(2) displays the corresponding parameter value;
(3) use or increase or decrease the value;
(4) press to return to the display parameters, value of the temporary storage.
(5)If other parameters are modified, repeat step 1 ~ (4);
(6)press the above 3 seconds, modify the parameters stored and returned to the display parameter category.
C. if there is no key in 60S. Exits the parameter setting program
D. parameter display
9
min
10
(3)Alarm Display
Just to switch on the power supply for medical refrigerator do not immediately add items, make empty operation after a
period of time (about 24 hours), then refrigerated goods into storage box.
To add items, if the item is too much water or excessive drying will influence the change of the humidity, the best goods
sealed; the same medical refrigerator working environment humidity size will also affect the change of the humidity of the
door and the door body ,especially too much not close well.
For each store items can not exceed 1/3 box capacity, achieve the actual temperature and normal operation after 24
hours, then add another 1/3 box storage capacity.
To add items please do not block the outlet and inlet.
The goods can not be placed directly on the bottom of the medical refrigerator should be placed on the bottom shelf,
otherwise the effect of medical refrigerator cooling effect.
The cooling process as far as possible not to open the door, otherwise it will cause temperature rise.
Due refrigeration inertia, the product reaches a set value when stopping the display temperature and the set temperature
may have a certain difference
Different, this is the normal phenomenon.
Note: children are not allowed to play games in a medical refrigerator.
Caution
For personal safety, please cut the power supply before any maintenance!
Don't inhale medications or aerosols around the device while maintaining a medical refrigerator, or it will endanger
your health. The medical refrigerator is direct-cooling and may have the risk of icing. It is recommended to observe the thickness
of the ice and remove it in time (30 days/time at 25 °C).
7.1 Defrost, Unused and Maintenance
The medical refrigerator will be automatically defrosted during the working process.
Unused:If the unit is not used for long time, please disconnect power supply and maintain
as the following procedures.
Maintenance: Once in a while, the medical refrigerator should be cleaned and maintained. (For the sake of safety,
remove the power plug firstly) and wipe the inner and outer surfaces of the refrigerator with a soft cloth.
11
Caution:DO NOT sprinkle water on the faces of cabinets, which may decrease insulating property of electric parts
and rust metal parts.DO NOT use hot water, corrosive cleanser and organic solvent!
No violent vibration or collision during transportation. No rain drench.
Suitable conditions: temperature: - 40℃ ~ + 55℃, relative humidity: 10% ~ 90%.
7.2 Disposal
Warning:
If the equipment is stored in unsupervised areas for a long period of time and leaveunused, ensure that the child is not
close to the medical refrigerator and the door cannot be closed completely.The disposal of refrigerator shall be carried out
by corresponding personnel to prevent the occurrence of such accidents as suffocation.
8. After-sale Ser vice
Any product has the possibility of failure. Please observe the operation of the medical refrigerator in the process of use.
If there is any abnormality, please check and compare the errors with the following table.If you can't fix the issue, Please
inform our service center in time. We will serve you wholeheartedly to avoid any losses.
Term of use: 10 years
12
③ The slight flowing noise of running water in the refrigerant pipe.
④ Before call the service engineer,Please clean and disinfect the refrigerator.
Condition:Cannot shake heavily,strike,prevent to drenching.
Storing environment temperature:-40℃~+55℃, Relative humidity: 10%~
90%.
An ne x 1: Rechargeable Battery Maintenance,Installation,Replacement
and Disposal
Long term power failure or in the process of transportation, the main power switch must be turned off. Otherwise, long
time discharge will cause battery loss or permanent damage, and it will be abnormal after re-energized.
Maintenance of rechargeable batteries: In order to prolong the battery life, please avoid idlingthe product. It is better to
run the product more than 24 hours monthly to recharge.
Battery Maintenance
A. If the refrigerator does not run in a long time, it should be connected to the poweron a regular basis (monthly), turn
on the power switch to charge the refrigerator for aperiod of time, and the charging time is not less than 24 hours.
B. When the power supply is interrupted, the power lock switch should be turnedoff in time, otherwise the battery will
lose power, which may cause permanentdamage to the battery.
C. The battery is expendable and the battery life is about 2 to 3 years. If thebattery is not properly used, such as the loss
of electricity or reach the batterylife, it will lead to low battery alarm.(It does not affect the usage of the refrigerator but
there is alarm failure and influence on printing function.It is suggested that users should contact company after-sales
service staff to replace.
1. Battery Installation Position: Top inside of electrical box
2. Battery replacement
a. Turn off the power switch and pull the plug from the socket (Pay attention to the electrical components in
the electrical cabinet. Power supply must be turned off and also unplug the power cord and turn off the power switch of
the refrigerator before opening. The electrical cabinet must be opened by qualified engineer or maintenance personnel).
b. Remove the battery connection plug.(Before unplugging the cord, pay attention to the sequence of the battery's
positive and negative levels and the connecting line, does not upside down the positive and negative levels to prevent the
control system damage from the installation of new batteries.The red line is usually connected to the positive pole, and the
black line is connected to the negative pole)
c. Remove two fixed screws from the battery plate with a screwdriver and remove the battery.
d. New replaced battery model: BT-12M4.0AC(12V4.0AH);
e. The replacement battery is recyclable, please contact the local battery recycling agency for processing.
Note: In order to effectively ensure that the replacement tank battery meets therequirements of the control system and to
avoid the impact of improperoperation on the system during the replacement, it is recommended to
contactthe MRC after-sale service staff to replace or guide.
13
Anex3: Remote alarm terminal and RS485 interface
The remote alarm terminal is installed in the back of refrigerator and the alarm signal is output by the terminal.The
terminal bearing capacity is DC 30V, 2A.
Normal open
N.O.
5 4 3 2 1
COM
9 8 7 6
RS-485 N.C.
1:+5v 5:GND 7:D+ 8:D-
Max DC30V 2A
Normal close
Common
14
9.Specifications
HYC-55 1 2
HYC-75 1 2
Circuit diagram
Battery
ROO ROO
M1 M2
IN1 IN2 IN3 IN4 IN5 IN6 IN7 IN8 +-
+-
SMPS
Control PCB
Evaporator M
motor
LED Light
Tran
M M -sformer
NC COM NO
If the product is improved, the technical data and circuit diagram shall be subject to the final product
nameplate and cabinet circuit label.
15