VWR Basic Incubator
VWR Basic Incubator
Trademarks
All trademarks mentioned in the operating instructions are the exclusive property of the respective manufacturers.
VWR International, LLC provides this document to its customers with a product purchase to use in the product
operation. This document is copyright protected This document is copyright protected and any reproduction of the
whole or any part of this document is strictly prohibited, except with the written authorization of VWR International,
LLC.
The content of this document are subject to change without notice.
All technical information in this document is for reference purposes only. System configurations and specifications in
this document supersede all previous information received by the purchaser.
This document is not part of any sales contract between VWR International, LLC and a purchaser. This document shall
in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all
conflicting information between the two documents.
Contents
Chapter 1 Safety Notes ..................................................................................................................... 1-1
Basic Operating Precautions ............................................................................................................... 1-1
Operational Safety Rules .................................................................................................................... 1-2
Warranty ............................................................................................................................................ 1-2
Explanation of Safety Information and Symbols ................................................................................. 1-3
Intended Purpose of the Incubator ..................................................................................................... 1-6
Standards and Directives .................................................................................................................... 1-6
List of Figures
Figure 3-1 Table-top incubators, dimensions and required clearances .................................................................... 3-3
Figure 3-2 Floor stand incubators, dimensions and required clearances.................................................................. 3-4
Figure 3-3 Lift Points............................................................................................................................................. 3-5
Figure 4-1 89511-418, 89511-420 and 89511-422 Front View............................................................................. 4-2
Figure 4-2 89511-418, 89511-420 and 89511-422 Rear View .............................................................................. 4-3
Figure 4-3 89511-424, 89511-426 and 89511-428 Front View............................................................................. 4-5
Figure 4-4 9511-424, 89511-426 and 89511-428 Rear View ................................................................................ 4-6
Figure 4-5 89511-430 Front View ......................................................................................................................... 4-7
Figure 4-6 89511-430 Rear View........................................................................................................................... 4-8
Figure 4-7 Sensor System (for table-top incubators)............................................................................................... 4-9
Figure 4-8 Sensor System (for floor stand incubators) ............................................................................................ 4-9
Figure 4-9 Signal Interfaces and Power Socket ....................................................................................................... 4-10
Figure 4-10 Gravity Convection Shelf System........................................................................................................ 4-12
Figure 4-11 Forced Air Shelf System...................................................................................................................... 4-13
Figure 5-1 Sliding the Retaining Spring into the Support Rail ............................................................................... 5-1
Figure 5-2 Installing the Shelving .......................................................................................................................... 5-2
Figure 5-3 Support Rail Installation....................................................................................................................... 5-3
Figure 5-4 Shelf Support Installation ..................................................................................................................... 5-4
Figure 5-5 Installing the Perforated Shelves ........................................................................................................... 5-4
Figure 5-6 Removing the bottom plate .................................................................................................................. 5-5
Figure 5-7 Removing the left and right support profiles......................................................................................... 5-5
Figure 5-8 Removing the rear air baffle 85911-430................................................................................................ 5-5
Figure 5-9 AC Power Supply Socket ...................................................................................................................... 5-7
Figure 7-1 Control Panel for VWR Incubators ...................................................................................................... 7-1
Figure 10-1 Door Seal Replacement ...................................................................................................................... 10-4
Safety Notes
VWR incubators have been manufactured to the latest state of the art and have been tested
thoroughly for flawless functioning prior to shipping. However, the incubator may present
potential hazards, particularly if it is operated by inadequately trained personnel or if it is not
used in accordance with the intended purpose. Therefore, the following must be observed for
the sake of accident prevention:
• Never step into the unit.
• VWR incubators must be operated by adequately trained and authorized professional
personnel.
• VWR incubators must not be operated unless these operating instructions have been fully
read and understood.
• The present operating instructions, applicable safety data sheets, plant hygiene guidelines
and the corresponding technical rules issued by the operator shall be used to create written
procedures targeted at personnel working with the subject matter device, detailing:
• the decontamination measures to be employed for the incubator and the accessories
used with it,
• the safety precautions to be taken when processing specific agents,
• the measures to be taken in case of accidents.
• Repair work on the incubator must be carried out only by trained and authorized expert
personnel.
• The contents of these operating instructions are subject to change at any time without
further notice.
• Concerning translations into foreign languages, the German version of these operating
instructions is binding.
• Keep these operating instructions close to the incubator so that safety instructions and
important information are always accessible.
• Should you encounter problems that are not detailed adequately in these operating
instructions, please contact VWR International, LLC immediately for your own safety.
Warranty
VWR International, LLC warrants the operational safety and functions of the VWR
incubators only under the condition that:
• the incubator is operated and serviced exclusively in accordance with its intended purpose
and as described in these operating instructions,
• the incubator is not modified,
• only original spare parts and accessories that have been approved by VWRare used
(third-party spares without VWR International, LLC approval void the limited warranty),
• inspections and maintenance are performed at the specified intervals,
• an operation verification test is performed after each repair activity.
The warranty is valid from the date of delivery of the incubator to the customer.
Harmful liquids!
Electric shock!
Hot surfaces!
Fire hazard!
Explosion hazard!
Suffocation hazard!
Biological hazard!
Contamination hazard!
Danger of tipping!
Mark of conformity USA/Canada
120 Volts AC power socket
Incorrect Use
To avoid the risk of explosion do not load the incubator with tissue, material, or liquids that:
• are easily flammable or explosive,
• release vapor or dust that forms combustible or explosive mixtures when exposed to air,
• release poisons,
• do not pour any liquids on the bottom of the interior surface or into a collecting basin
inside the unit.
• release dust
• exhibit exothermic reactions
• are pyrotechnical substances
• refrain also from pouring any liquids onto the internal base plate or inserting bowls filled
with liquids into the sample compartment.
Packaging
VWR incubators are delivered in a rugged packaging box. All packaging materials can be
separated and are reusable:
Packaging materials
Acceptance Inspection
After the incubator has been delivered, check the delivery immediately for:
• completeness,
• possible damage.
If components are missing or damage is found on the incubator or the packaging, in particular
damage caused by humidity and/or water, please notify the carrier as well as VWR
International, LLC Technical Support immediately.
Risk of injury
Scope of Supply
Incubators
Gravity Convection Forced Air
Quantity of components supplied (pieces)
Incubators Incubators
Perforated shelves 2 2
Support rail for shelf table-top incubators 4 2
Shelf support 4 4
Power cord 1 1
Clip springs for table-top incubators 4 2
Plug 1 1
Anti-tilt anchor 1 1
Operating manual 1 1
Installation
Ambient Conditions
Location Requirements
Built-in incubators can, heating and drying ovens must be operated with an air exhaust system
and exhaust hose (only original VWR accessory should be used).
For safety reasons, the installation space should be made of non-combustible materials,
according to DIN 4102.
During installation of built-in units, ensure that the escaping air will
be safely discharged out of the installation space.
The incubator must only be operated in a location that meets all of the ambient condition
requirements listed below:
• Installation location indoors in dry areas free from drafts.
• The dust burden may not exceed the contamination category 2 based on EN 61010-1.
Using the incubator in an atmosphere with electrically conductive dust is prohibited.
• The minimal distance to adjacent surfaces must be observed on all sides.(See the
requirements on Table 3-2 on page 3-3.
• The operating room must be equipped with appropriate ventilation.
• Solid, level, fire-proof surface; no flammable materials opposite to the rear panel of the
incubator.
• Vibration-proof substructure (floor stand, lab table) capable of bearing the dead weight of
the incubator and its accessories (particularly if two devices are stacked).
• The electrical circuitry of the incubator has been designed for an operating height of up to
2000 m above sea level.
• Relative humidity up to 80% (maximum; preferably 60-70%), non condensing.
• Should condensation exist, wait until the moisture has evaporated completely before
connecting the incubator to a power source and powering up.
• If a high-voltage test is to be performed on the unit, it must first be heated for around 30
minutes at 75°C.
• The ambient temperature must be within a range of +18°C to +32°C (64.4°F to 89.6°F).
• Avoid direct exposure to sunlight.
• Devices that produce excessive amounts of heat must not be placed near the incubator.
• Place the incubator on a floor stand (option; to be ordered separately), never on the lab
floor. This prevents the penetration of dust or dirt into the device.
• Power line voltage variations must not exceed ±10 % of the nominal voltage.
• Transient surges must lie within the range of levels that normally occur in the power
supply system. The impulse withstand voltage based on surge category II of
IEC 60364-4-443 shall be applied at the nominal voltage level.
• Consider installing one dedicated upstream circuit breaker per incubator to avoid multiple
device failures in case of an electrical fault.
Contamination hazard
Do not place the incubator directly on the lab floor, but mount it
on the floor stand or on a lab work surface (option; to be ordered
separately). Contaminants, such as bacteria, viruses, fungi,
prions, and other biological substances may use the open door
to migrate easily from the floor into the incubator’s work space.
Intermediate Storage
When the incubator is placed in intermediate storage, which is permissible for a maximum of
four weeks, make sure that the ambient temperature is between 20°C to 60°C (68°F to 140°F)
and the maximum relative humidity does not exceed 90%, non-condensing.
Room Ventilation
Heat dissipating from the incubator during continuous operation may cause a change in the
room climate.
• Therefore, the incubator must only be installed in rooms with sufficient ventilation.
• Do not install the incubator in room recesses without ventilation.
• When several devices are to be placed in the same room, additional ventilation may have
to be provided as necessary.
• To avoid any impact of the heat dissipated by the incubator on the ambient climate the
room must be vented by means of a laboratory-grade ventilation system that complies
with applicable local and national health and safety regulations and has sufficient capacity.
• If excessive temperatures tend to occur in the operating room, be sure to provide a thermal
protection means that cuts out the power supply to mitigate the impact of
overtemperature scenarios.
Table-top incubators
Transport
Table-top incubators
For transport, do not lift the incubator using the doors or components attached to the
incubator as lift points.
The floor stand incubators come equipped with four (4) casters. The lever for releasing the
caster is located above the locking lever. After positioning the unit in its installation location
ensure that the locking levers are pressed down on the casters.
To ensure the degree of stability specified by safety requirements the front casters must be
turned so that they are facing forward after the unit has been positioned in its installation
location and the locking levers pressed down on these casters.
Danger of tipping when moving!
Disconnect the unit from the power source before moving it.
Always ensure that the doors are closed when moving the unit.
1. Do not use this position if the door is hinged on this side. Right-hand hinges represent the
standard configuration.
2. Preferred position.
3. Alternative position. Do not use if the door is hinged on this side.
Remove the bracket screws. Use the preferred position, if possible.
Fix the anti-tilt anchor with the bracket side down to the unit.
Position the unit with the anti-tilt anchor to in an angle of approx. 90° +/- 20%.
Take care that the stacking feet of the unit are still in correct place on the lower unit or on the
stacking adapter.
Fix the anti-tilt anchor to a solid part of the building.
You should be aware at all times that stacked devices do not form a
stable unit, even when the stacking pads and frames are correctly
used. The top device may tip over and drop down when being
transported in a stack. To avoid injury to persons and damage to
equipment, do not attempt to move stacked devices as a unit!
Separate and move each device one by one, then restack them.
Product Description
• an on/off timer
• two perforated shelves
• an access port for tubing, sensor leads, etc.
The individual features of forced air incubators are shown in the figures below.
Safety Devices
The incubators are equipped with the following safety features:
• a sample protection feature that safeguards the samples against destruction through
overheating in case of controller failure;
• dual fuses rated at 16 amperes.
The work space temperature sensor provides the inputs to the incubator’s built-in controller,
which continuously compares the measured values to the user-specified set value and adjusts
the heaters according to the result.
Do not remove the protection hose from the sensor.
Protect the sensor from mechanical damage.
The unit features a thermal protection function that is factory-preprogrammed and not
adjustable. It protects the cultures in the work space from overheating: Thermal protection
kicks in on a brief violation of the upper limit, based on the defined setpoint temperature, at
between 2 and 3°C (35.6°F and 37.4°F) (37°C (99°F): 2°C (35.6°F), > 50°C (122°F): 3°C
(37.4°F)), automatically reducing the work space temperature to the user-specified set value
and allowing the incubation process to be continued even in case of a controller malfunction.
If the thermal protection is activated, the error message (E111) “Temperature too high”
appears in the display window and an audible alarm is sounded.
When the user acknowledges the error message, the red alarm icon (D4 in figure 7-1 on
page 7-1) is illuminated and the Temperature Set Value icon (see table 7-3 on page 7-3) is
highlighted by a red border to indicate that thermal protection has kicked in.
RS-232 Interface
The RS- 232 interface (item 2 in figure 5 below) may be used to connect VWR incubators to
the serial interface port of a computer to allow for the computer-aided acquisition and
documentation of major operating parameters (temperature, etc.).
AC Power Socket
The incubator is connected to the AC supply mains via the socket (item 4 in figure 5), which
accepts a power cord with an IEC standard plug.
Fuses
Two 16 A slow-blow fuses mounted on the incubator’s main electronic circuit board protect
internal circuitry from the impact of excessive power consumption.
Replacement should only be carried out by skilled and authorized
qualified personnel of electrotechnology/signal engineering!
Fuse replacement
Access Port
A re-sealable, capped access port (can be closed off using the plugs delivered with the unit)
allows cables, hoses or additional sensor leads to be routed into the work space of the
incubator.
The access port [2] has a diameter of 42 mm.
Operating conditions
Shelf System
The incubator is supplied with two perforated shelves. The shelf support rails [1] have an
alternating pattern of oblong and round perforations spaced evenly at 30 mm, allowing the
shelf support brackets [8] to be inserted without any room for error, yet in a very flexible way
to accommodate any required height of sample container. The shelves [2] have an integrated
tilt protection and pull-out stop. For details on using the shelf system, see “Installing the Shelf
System for Table-top Units” on page 5-1.
Start-up
Initial Installation
1. Peel off the protective foil from the support rails.
2. Push the retaining spring [1] into the guide on the support rail [2], making sure that the
locking nub [3] on the retaining spring safely engages with the matching hole in the
support rail.
.
Figure 5-1 Sliding the Retaining Spring into the Support Rail
Disinfection
The embossings at [2] and [5] act as lateral guides for the support rails, while the embossings
at [1] and [6] secure the support rails in place. For the support rails to install correctly the
retaining spring [3] must be facing upwards.
1. Place the support rail [4] on the lower embossing [6] and tilt it upwards against the work
space side wall so that the rail is positioned over the two embossings at [5] and [2].
2. Clamp the retaining spring [3] behind the upper embossing [1].
3. To remove the support rails, pull the retaining spring tab down out of the embossing and
remove the support rail assembly.
2. Make sure that the two vertical elements [2] of the shelf support butt against the support rail.
1. Push the shelf [4] onto the shelf support with the tilt protection devices [2] facing the rear
panel of the incubator.
2. Slightly raise the perforated shelf so that the pull-out stops [1] and [3] can slide over the
shelf support.
3. Make sure that the shelves and both of their tilt protection devices are free to move over
the shelf support.
2. Manually adjust the levelling feet until the shelf is horizontally aligned in all directions.
Perform the adjustment of the levelling feet from left to right and from rear to front.
Connecting Power
Electric shock
The incubator has a class I, protection-earthed enclosure. To minimize the risk of electrical
shock, use the AC power cord supplied to connect the incubator to a correctly installed and
protection-earthed power supply source, with the following features in place for each
incubator:
• T 16 A slow-blow fusing
• B 16 circuit breaker
• FI circuit breaker
Keep the power outlet accessible!
Condensation
When taking the incubator into operation for the first time allow
some time before switching on for stabilization to avoid
condensation forming on live parts.
If a high-voltage test is to be performed on the unit, it must first be
heated for around 30 minutes at 75°C.
The RS-232 data communication interface supports the querying of status information and
temperature data from the incubator by entering basic commands in a standard terminal
window provided by your computer’s operating system. The interconnection requires a
standard RS-232 cable with 9-pin connectors and a straight “1:1” pinout without any crossed
wires, which is not supplied with the incubator.
Users may employ the RS-232 command inventory listed in table 5-1 below for automating
process data logging - for example, by embedding these commands in scripts that run on a
remote computer.
RS-232 interface compatibility
Operation
Device Check
Prior to starting operation, the following incubator components must be checked for their
correct function:
• The door seal in the front frame must not be damaged.
• The glass door must not be damaged.
• The shelving components must be installed safely.
• Disinfecting the Incubator’s Work Space
Disinfect the work space according to the operator-specified hygiene guidelines.
Starting Operation
1. Turn the incubator on using the control panel.
2. Adjust the temperature set value on the control panel.
3. The temperature controller starts adjusting the work space to the user-specified
temperature set value now.
To avoid any risk of explosion or fire
Hot surfaces
The screen of the glass door, the interior panel of the outer door
as well as the surfaces of the shelving and the work space
become hot while the incubator is running through its heating
cycles and need some time to cool down.
Risk of overloading
Proper loading
VWR incubators come with a front panel mounted control unit consisting of a
multifunctional display, four control buttons, and an on/off button. The four control buttons
interact with the display window to let users access all of the user control functions and
adjustments of the incubator, including - for example, the temperature set value, timer,
energizing/de-energizing, as well as a variety of other functions.
Under normal operating conditions the display presents user with the work space temperature.
The display returns to its default mode upon completion of the adjustments or whenever no
entries have been made for a period of 30 seconds.
The graphic below shows the VWR incubators control panel with all of its visualization
elements and controls.
The table below contains brief descriptions of the buttons on the control panel (items K1
through K5 in Figure 7-1).
The table below contains brief descriptions of the display features of the control panel (items
D1 through D3 in figure ; the identifiers K1 through K4 refer to the buttons shown in that
figure).
The table below contains brief descriptions of the menu bar icons (item D2 in figure ).
.
Powering Up
1. Plug the power plug of the incubator into a suitable protection-earthed AC power outlet.
In the display window on the front panel the readiness indicator icon (rightmost icon in
the menu bar at D2 in figure on page 7-1) is illuminated.
An initialization routine will be run after the incubator has been powered up. On
completion of the initialization, the display will light up and the current work space
temperature will appear in the temperature display field (item D1 in figure on page 7-1).
The incubator is ready for use now.
The display window goes out, except for the readiness indicator icon (rightmost icon in
the menu bar at D2 in figure on page 7-1) and a residual heat temperature readout in case
the work space temperature is still higher than 50°C (122°F). The incubator is switched
off now.
2. If required, unplug the AC power plug to power down the incubator completely.
VWR incubators allow for setting the desired work space temperature directly using only a few
button presses. After confirming the new temperature set value, you may trace the resulting
temperature change in the temperature display field (item D1 in figure on page 7-1.
Table 7-4 Adjusting the Temperature Set Value
Timer
The Timer feature from the menu bar enables the user to set a “countdown-type” on or off
timer that switches the incubator on or off after a preset period of time. Instructions for setting
an off timer are given in table 7-5 (see below), while the usage of an on timer is described in
table 7-6 on page 7-7.
Programming a turn-on time causes the incubator to switch off until it is scheduled to restart,
while a turn-off time keeps the device running before it shuts down at the user-specified time.
The timer starts running immediately as soon as the user confirms his or her entries.
Table 7-5 Setting a Countdown-type Off Timer
selection with .
Stopping a Timer
Table 7-7 Stopping an Off Timer Before It Expires
Settings
The Settings menu item opens a submenu populated with various commands for viewing
general status information on the VWR unit and setting for the operation of the incubator or
its display window:
• Read access to error log
• Calibrating the incubator
• Toggling the temperature display unit between °C and °F
• Entering a configuration control code
Error Log
Users calling customer service for support may be asked by the VWR agent to supply
information from the error log of the incubator. It enables the user to browse through the
most recent 22 alarm messages that were caused by hardware or control loop errors. Each error
is displayed with an internal error code.
Error codes and instructions for clearing alarm conditions appear in the section “Error Codes”
on page 12-1.
Table 7-9 Reading the Error Log
Use the button to select the first entry in the error log,
numbered E01 (Error 01).
After a few seconds, the display pane automatically switches
to the internal error code - for example, 109.
E01 shows the latest fault, E22 shows the oldest fault.
To exit from the error log and return to normal display mode
press twice.
The Settings icon in the menu bar will go out.
Calibration
The Settings -> Calibration menu item enables the user to initiate a temperature calibration
process (see “Temperature Calibration Procedure” on page 10-3) for the built-in temperature
sensors and choose whether calibration should be accomplished manually or automatically:
• The Manual option allows for entering an absolute temperature directly, as measured - for
example, using an external reference sensor.
Calibration Prerequisites
Note This setting does not have any impact on data logging via the RS-232 interface. Any
temperature data that is logged to a computer for operational parameter documentation purposes is
handed over in °C.
Table 7-11 Toggling the Temperature Display Unit
Shut-down
Contamination hazard
To rule out any risk for subsequent users, perform a full cleaning,
disinfection and decontamination cycle to the standards set out
in the section “Cleaning and Disinfection” on page 9-1 if you
suspect (or if you are positive) that harmful biological material
has been processed with the incubator.
1. Remove the containers with the cultures, all accessories, and other objects from the work
space.
2. Clean and disinfect the work space, as explained in the section “Cleaning and
Disinfection” on page 9-1.
3. When cleaning and disinfection and/or decontamination are done, turn the incubator off
using the control panel.
4. Unplug the power cord and secure it against accidental reconnection.
5. Until the incubator is shut down, the work space must be continuously ventilated. Leave
the glass door and the outer door open and secure them against accidental closure.
Cleaning
Incompatible cleaners
Moisture-sensitive components
Do not spray cleaning agent onto the control panel and the
interfaces at the rear of the incubator. When wiping the incubator
clean, always make sure that no moisture enters into these
components.
Wipe the display window down with a slightly dampened cloth,
then wipe dry with a rag made of 100% microfiber.
Alcoholic disinfectants!
After the disinfectant has been allowed to react, wipe the cleaned
incubator components thoroughly dry.
Chloride-containing disinfectants!
Chloride-containing disinfectants can corrode stainless steel.
Health hazard
Predisinfection
1. Remove all samples from the work space and store them in a safe place.
2. Spray disinfectant onto the surfaces of the work space and of the accessories or wipe the surfaces
clean using disinfectant.
Risk of injury caused by breaking of
glass panel
The glass panel may only be removed by properly trained and
authorized specialists.
Check to ensure that the air baffles are securely screwed into
place after cleaning and moving the unit.
Cleaning
1. Remove all internals from the specimen chamber.
2. Wipe off the specimen chamber surfaces and the internals that have been removed from the
chamber using lukewarm water mixed with standard rinsing agents. Completely remove any
tenacious impurities using rinsing agent and warm water.
3. Re-rinse the cleansed surfaces 3 to 5 times with autoclaved water to completely remove and
cleaning agent residues.
4. After this, wipe the cleansed surfaces and internals dry with a soft, sterile cloth.
Final Disinfection
1. Spray the specimen chamber surfaces and the internals removed from this chamber again with
disinfectant, or wipe them down.
2. Let the disinfectant work on the surfaces/internals as detailed in the manufacturer’s instructions.
3. Re-install the internals in the specimen chamber.
Recommendations for
decontamination
Bio-hazard
Maintenance
Maintenance and inspection at regular intervals of the features and components listed below
are mission-critical to maintain the product in a fully operative and safe condition and avoid
malfunctions due to aging and wear. Failure to perform maintenance on a regular basis may
result in:
• deviations in heating performance
• damage to samples
• loss of control over temperature distribution throughout the work space
Regular Checks
• Check the incubator for overall cleanliness and remove any residues from previous
processes.
• Check the locking screw for the glass panel
Semi-annual Inspection
• Check integrity and proper seating of the seal.
• Perform functional check of the control panel and of the incubator’s built-in controller.
• Perform electrical safety check in accordance with the relevant national regulations.
Functional Check
Service Intervals
During running operation, the following service works must be performed:
3-monthly service
• Perform the comparative temperature measurement outlined in the following section.
Annual Service
• Have the incubator inspected and services by a VWR authorized Technical Service agent.
Service contract
Use a calibrated measuring instrument with an accuracy of < ± 0.1°C (0.18°F) for this test.
To minimize temperature variations during the measurement, put the measuring sensor in an
isothermal container (such as a bowl filled with glycerol) before placing it in the work space.
Use the center of the work space as the reference location for the comparison measurement.
Isothermal container
Excessive work space temperatures
Excessive work space temperature
Contamination hazard
The incubator may have been used for treating and processing
infectious substances, which may have caused contamination of
the incubator and its components.
Prior to return shipment, it is therefore mandatory that all
incubator components be properly decontaminated.
• Clean the incubator components thoroughly, then disinfect or
decontaminate them (depending on application).
• Fill in and attach a safety declaration with details on
decontamination activities performed to the items that are to
be repaired.
Disposal
Contamination hazard
The incubator may have been used for treating and processing
infectious substances, which may have caused contamination of
the incubator and its components.
Prior to disposal, it is therefore mandatory that all incubator
components be properly decontaminated.
Component Material
Interior containers, installed compo- Stainless steel 1.4016 + 1.4301
nents and shelves
Door frame seal Silicone
Glass screen Sodium silicate glass
Cables Plastic-sheathed stranded copper wire
Packaging Corrugated board, polyethylene film, and
styrofoam, chemically untreated wood
Error Codes
The table 12-1 below lists the error messages that may appear in the control panel display
window (see “Error Log” on page 7-9) and provides instructions for clearing such alarms.
Table 12-1 VWR Incubator Error Codes
Error Message & Code Root Cause Alarm Response Alarm Clearing Instructions*
Display Error (E002) Display communication Audible alarm activated, Power cycle the device by
error. The built-in message shown on unplugging, then
controller was unable to display. Reset after 30 s. reconnecting the power
restore communication with cord. If this doesn't solve
the control panel. the problem, call service.
Mirrored Parameter The controller was unable Fallback to mirrored Check the latest settings,
Loaded (E003) to read the user-specific parameter storage. Device for example the set value.
settings and had to resort continues to run without
to an emergency loss of functionality,
parameter set held in mir- including user-specific
rored storage. settings.
Factory Parameter The controller was unable Fallback to factory-preset Acknowledge by pressing
Loaded (E004) to read the mirrored parameters. Audible alarm . Re-enter
parameter set and had to activated, message shown
customer-specific settings.
resort to factory-preset on display. User-specific
parameters. settings may be lost - for
example, the temperature
display unit preference, or
user programs.
Default Parameter The controller was unable Fallback to default Call service.
Loaded (E005) to read the factory-preset parameters. Audible alarm
parameters.and had to activated, message shown
resort to default settings on display. The device is
completely inoperative.
Power Down Error Power has been cut off Audible alarm activated, Check the power supply.
(E007) (power outage) while the message shown on Power up then device, then
device was running. display. acknowledge the alarm by
pressing .
Config Error (E012) General device Audible alarm activated, Call service.
configuration error. message shown on
display.
OTP error (E013) Klixon contact not closed. Overtemperature Restart the device. If this
Protection fault. Audible doesn't solve the problem,
alarm activated, message call service.
shown on display. Bridging
across Klixon contact has
failed.
Calibration Value Too The calibration reference Fall back to previous Check the external
low (E105) value calculated on the calibration reference. reference sensor for proper
basis of the user input Audible alarm activated, function and replace, if
exceeds the lower limit for message shown on necessary. If this does not
calibration references. display. solve the problem, call
service.
Constant Sensor Sig- None of the decimal places Control transferred to Call customer service for
nal (E106) of the A/D-converter reference sensor, audible support. Shut down any
output for the process alarm activated, message device’s operation. Any
sensor has changed over a shown on display. If both further operation of the
specific time period. sensors are defective, all device is at your own risk
control circuits will be and may result in damage
disabled. to the device and the
samples contained in the
device.
Technical Data
The technical data are valid only for an empty device equipped with three shelves, a spray-painted outer enclosure
and a power line voltage 120 V/60 Hz. Options may have an impact on the specified performance.
Table 13-1 Technical Data - Gravity Convection Incubators
Parameter Unit 89511-418 89511-420 89511-422
Process
Work space atmosphere Ambient Ambient Ambient
Min. °C/°F temperature temperature temperature
plus plus plus
Max. °C/°F 5°C/9°F 5°C/9°F 5°C/9°F
75 °C/167 °F 75 °C/167 °F
75 °C/167 °F
Temperature deviation from set value at 37°C K ± 0.7/±0.6 ± 0.7/±0.6 ± 0.7/±0.6
(98.6°F), spatial. Max. value /typical value
Temperature deviation from set value at K ±0.2 ±0.2 ±0.2
37°C (98.6°F), over time
Heat-up time (work space unoccupied, from min 52 42 47
25°C (77°F) to 98% of set temperature of
37°C/98.6°F)
Recovery time (work space unoccupied, min 5/3 5/4 9/8
door open for 30 s, to set temperature).
Max. value/typical value
Heat dissipation to environment (at set W 21 ±10% 26 ±10% 31 ±10%
temperature of 37°C (98.6°F) and room
temperature of 25°C/77°F)
Overall dimensions
Height mm/in 720/ 820/ 920/
28.3 32.3 36.2
mm/in 530/ 640/ 640/
Width
20.8 25.2 25.2
Depth mm/in 565/ 565/ 738/
25.2 25.2 29.1
Overall weight kg/lbs 40/88 51/112 65/143
Loading capacity
Loading capacity per shelf kg/lbs 25/55
Max. overall loading capacity per device kg/lbs 50/110 50/110 75/165
Electrical data
Power consumption W 300 540 720
Maximum current A 2.5 4.5 6.0
Overall dimensions
Height mm/in 720/ 820/ 920/ 1655/
28.3 32.3 36.2 65.2
Width mm/in 530/ 640/ 640/ 755/
20.8 25.2 25.2 29.7
Depth mm/in 565/ 565/ 738/ 770/
25.2 25.2 29.1 30.3
Overall weight kg/lbs 45/99 56/123 70/154 145/320
Loading capacity per shelf kg/lbs 25/55 30/66
Max. overall loading kg/lbs 50/110 50/110 75/165 75/165
capacity per device
Electrical data
Power consumption W 600 840 1020 1380
Maximum current A 5.0 7.0 8.5 11.5
Earthing system (e. g. 1/N/PE 1/N/PE 1/N/PE 1/N/PE
1/N/PE)
Power line frequency Hz 60 60
Device Log