Session 5 - Topic 3 - Afterloader System Commissioning & QA
Session 5 - Topic 3 - Afterloader System Commissioning & QA
Session 5; Topic 3
Afterloader System Commissioning & QA
date
Educator
Institution
Learning Objectives
2
Document Checklist
❑ Resource Documents
❑ IAEA
❑ ESTRO
❑ Local regulations 3
Delivery and Install (D&I)
D&I: Checking Transport Damage
• Afterloader:
• Visible scratches, dents?
• Frayed cables?
• Shipping Manifesto:
• Agrees with purchase order?
• All items delivered and
accounted for?
5
Acceptance Testing
When does acceptance testing happen?
Acceptance
Delivery and Commissioning:
Testing: Vendor
Install: Vendor Site Physicist
& Site Physicist
• Delivery and Install:
• Machine is property of vendor (or installer).
• Anything that breaks will be fixed by vendor at no cost to customer
• Acceptance testing:
• Performed by vendor and site physicist together (can also mean vendor
logged in remotely)
10
Example acceptance tests
4. Setting Up QA:
• Daily
• Quarterly
• Annual
13
Commissioning HDR Afterloader Stage 2:
• Connectivity tests:
• MRI/CT/Simulator to planning system
• Planning system to afterloader
• Planning system to electronic medical record system (if applicable)
• Planning system and afterloader to printer
• USB transfer of data in case network is down
(also called “sneaker net”)
15
System Administrators
• Administer software (time & date)
• User Administration
• Assigning roles and rights to individuals
18
Commissioning HDR Afterloader Stage 3:
20
Checking Safety Interlocks
21
Components of system tests
22
Source Position Accuracy I
• Several commercial tools available
• Embedded distance markers for
reference (highly accurate
machining of ± 0.1 mm)
• Autoradiograph of source on film
• From most distal source to range
used for treatment
• Tolerance ± 1 mm
23
Source Position Accuracy II
• Verifies source positioning accuracy
• Run for every channel and transfer
tube at commissioning
• Multi-channel tools also available
• With camera mounted above,
position of dummy source relative to
live source can be confirmed
• By disconnecting transfer tube,
channel obstruction can be checked
24
Components of system tests
25
Timer Function
4. Tolerance ± 0.5 s
26
Timer Accuracy (Transit Dose/Source Velocity)
o Recall this from session 4
27
Poll Question #1
28
Poll Question #1
30
Components of system tests
31
Testing Treatment Workflow (End-to-End Test)
• Create a treatment workflow
• Perform end-to-end test using a phantom
• Every step of the workflow should be
performed by the staff doing it in clinic
• Check that the afterloader functions as
intended
• Check that staff is clear about how to use
equipment
• Start with easy patients
32
Example of How-to-Do End to End test
33
Other Commissioning Tasks:
Transfer Tubes
Applicators
Treatment Planning System
Ongoing equipment QA
35
Daily QA
• Checks to be done on the day of a patient treatment (before treatment):
• Date/time in system
• Source strength/decay
• Source position (film or filmless)
• Timer accuracy
• Safety interlocks
– Door
– Transfer tube
– Applicator
• Intercom/CCTV functional
• Radiation warning/lights functional
• Interrupt and emergency stop functional
• Emergency equipment present
36
Source exchange or quarterly QA
• Checks to be done after a source exchange, or quarterly:
• Well chamber constancy check
• Source strength measurement**
• Source positional check (film)**
• Timer linearity & error check
• Transfer tube check (length)
• Daily QA tests**
• Add new source information in treatment unit and planning system
• Generation of new decay tables for the treatment console and
planning area
**THESE CHECKS SHOULD ALSO BE DONE AFTER ANY REPAIRS THAT
REQUIRE REMOVAL OF THE SOURCE FROM THE AFTERLOADER
37
Annual QA
• Checks to be done annually:
• Applicator checks/recommissioning
• Check well chamber calibration date (may need recalibration)
• Check electrometer calibration date (may need recalibration)
• Review of all documentation, checklists and forms
• Emergency procedures refresher and dry-run
38
Poll Question #2
39
Poll Question #2
40
Why do all of these tests?
• Mistakes happen. Things break. We are human.
Examples:
• HDR units damaged in transit
• Source cable separated from the drive unit (remained out of the afterloader)
• New source strength not added to afterloader at source exchange
• Source detached from cable (remained in catheter and inside patient for 4
days – patient died 1 day later)
• Inadequate bunker shielding
41
Resources
42