Updated Rationale For Exemption of Biocompatibility Testing
Updated Rationale For Exemption of Biocompatibility Testing
1. Introduction
This document provides a justification for the exemption of biocompatibility testing for
orthopedic instruments manufactured and sold by Griportho Surgicals Private Limited. The
orthopedic instruments in question are designed for temporary use during surgical
procedures and are classified as non-implantable devices. The rationale is based on the
nature of their contact with the body, the use of sterilization processes, and their
mechanical functions, which do not involve prolonged interaction with biological tissues or
fluids.
Example: Surgical tools like bone drills or retractors manufactured by Griportho Surgicals
are used for a limited period during the surgery and do not remain in the body post-surgery.
3. Non-Implantable Devices
The orthopedic instruments sold by Griportho Surgicals Private Limited are strictly non-
implantable. They are not intended to remain inside the patient’s body after surgery. As
non-implantable devices, these instruments do not require prolonged or permanent
interaction with bodily tissues or fluids, thereby eliminating the need for biocompatibility
testing typically required for implantable medical devices.
Example: A surgical saw or clamp used to assist in the positioning and stabilization of bones
during surgery is removed once the procedure is over and, thus, does not require
biocompatibility testing.
Example: A bone chisel or alignment tool touches the bone or soft tissue briefly and is then
removed.
Example: A depth gauge used to measure the depth of a hole drilled in bone performs an
external, mechanical function and does not interact with tissues or fluids.
7. Conclusion
Orthopedic instruments manufactured by Griportho Surgicals Private Limited are designed
for transient use during surgery, have no prolonged contact with bodily tissues or fluids,
and are sterilized by healthcare professionals before each use. Their non-implantable
nature and mechanical functions further support the exemption of biocompatibility testing.
Therefore, biocompatibility testing is not required for these instruments, as their safe use is
ensured through proper sterilization protocols and their temporary application in clinical
settings.