Computer Controlled Distillation Column
Computer Controlled Distillation Column
COLLEGE OF ENGINEERING
GROUP 2B
DICKSON PRECIOUS
LIST OF TABLES............................................................................................................2
CHAPTER 1......................................................................................................................3
1.0 INTRODUCTION...................................................................................................3
CHAPTER 2....................................................................................................................10
2.1 PROCEDURE........................................................................................................11
CHAPTER 3....................................................................................................................12
3.3 DISCUSSION........................................................................................................18
CHAPTER 4....................................................................................................................19
4.0 CONCLUSION......................................................................................................19
CHAPTER 5....................................................................................................................21
5.2 REFERENCES......................................................................................................25
LIST OF FIGURES
LIST OF TABLES
1.0 INTRODUCTION
Distillation is a method of separating the components of a solution which depends upon
the distribution of the substances between a liquid and a gas phase, applied to cases
where all components are present in both phases (Treybal, 1980). Instead of introducing
a new substance into the mixture in order to provide the second phase, as is done in gas
absorption or desorption, the new phase is created from the original solution by
vaporization or condensation. During most of the century, distillation was by far the
most widely used method for separating liquid mixtures of chemical components.
(Taylor and Krishna, 1993). Despite the high energy requirements, distillation is
typically favoured over other separation processes since it needs less upfront investment
and produces high purity products. Despite the recent advent of various novel separation
xylene, distilled water, alcohol, paraffin, kerosene, and a number of other liquids. It is
also used to cleanse and separate liquid mixes of one or more liquids. Distillation is
boiling points of the constituent components. Furthermore, the boiling point of the
liquid will vary based on the amounts of the components present. As a result, distillation
boiling points into the gas phase. The gas is then condensed and collected in liquid
form. The overhead distillate and a bottom product are among the products. Double
distillation is the technique of repeating the procedure on the collected liquid to increase
example, nitrogen, oxygen, and argon may all be distilled from air.
Steam distillation, simple distillation, fractional distillation, and vacuum distillation are
all forms of distillation. Heat-sensitive components are separated via steam distillation.
When steam is introduced to the mixture, some of it vaporizes. This vapor is condensed
into two liquid parts after cooling. The fractions are sometimes collected separately or
given distinct density values so that they can separate on their own. Steam distillation of
flowers, for example, produces volatile oil and a water-based distillate. When the
boiling points of two liquids are significantly different, simple distillation can be
convert the most volatile component from a liquid to vapor in simple distillation. The
vapor rises to the top and condenses in a condenser. Typically, the condenser is cooled
(e.g., by circulating cold water around it) to promote vapor condensation and collection.
Vacuum distillation is employed to separate components that have high boiling points.
Lowering the pressure of the equipment also lowers boiling points. Otherwise, the
method is comparable to other types of distillation and it is especially useful when the
When the boiling points of the components of a mixture are close to one other, as
mixture is heated so that vapor rises and enters the fractionating column in fractionation.
As the vapor cools, it condenses on the column's packing. This liquid vaporizes again as
a result of the rising vapor's temperature, pushing it up the column and finally giving a
can be adapted (plate columns and Raschig rings column). An eight-plate column, a
reflux system, a tank for distillation, a vacuum pump and a feeding pump (for
continuous feeding) are included. The steam that goes to the head of the column is sent
to a total condenser. The cooling water flow that crosses the condenser is regulated and
indicated in a flow sensor. Distillation can be carried out at low pressures with the help
of an adjustable vacuum pump. The pressure loss in the column can be measured with a
pressure sensor. It allows continuous operation and batch operation. For proceeding to
continuous feeding, a pump that can inject the substance directly into the boiler or in
any of the plates is available. The temperatures of the system are measured by sensors
placed in strategic positions. This Computer Controlled Unit is supplied with the
EDIBON Computer Control System (SCADA), and includes: The unit itself, a control
interface box, a data acquisition board, a computer control, data acquisition and data
management software packages, for controlling the process and all parameters involved
in the process. The use of the computer-controlled distillation unit helps to analyze real-
time data for the distillation of alcohol and water in this lab work. A brix refractometer
is then used to measure the %brix of the distillate and reboiler at various time intervals
ii) To analyze the distillation operating line using the McCabe Thiele diagram.
1.2 RAOULT’S LAW
The value of the mole fraction of alcohol in water at a given temperature is obtained
using Raoult's Law, allowing the bottom curve of the phase diagram to be drawn. It is
mathematically defined as;
¿
¿ ¿ P−P B (T )
P= X A P (T )+(1− X A )P (T )=¿ X A = ¿
A B ¿
P A (T )−P B (T )
The relationship between the mole fractions of the elements in the liquid and vapor
phases must be employed to depict the upper curve of the diagram.
¿
PB = the partial pressure of component B
1.3 AZEOTROPIC MIXTURES
A minimal boiling azeotrope is formed at a certain composition by a solution that
instance. After distillation, azeotropic mixtures have the same chemical ingredient ratio;
at a given pressure, boiling creates a vapor with the same chemical constituent ratio as
the liquid. An ideal mixture follows Raoult’s law. The point at which the vapor and the
liquid compositions are the same is called an azeotropic point. At the Azeotropic point,
equal to the heat (energy) required to vaporize or condense a mole of water, and is
approximately equal to the heat (energy) required to vaporize or condense any mixture
this link, the tower may be explored using visual techniques that employ straight lines.
The 45° line depicts the probable areas where the vapor concentration equals the liquid
such vapor-liquid mix would need an endlessly tall distillation tower if the equilibrium
curve passes the 45°line. In addition, if the equilibrium curve crosses the 45° line, the
combination has generated an azeotrope. Simple rectification would not be able to distill
Water
Computer Control Distillation Unit with SCADA and PID control (UDCC)
Test tubes
Beaker
Volumetric flask
Alcohol meter
Measuring cylinder
Funnel
Stop watch
2.1 PROCEDURE
The alcohol meter's calibration curve was created by measuring the percent V of water,
pure ethanol, and a few combinations of water and ethanol whose compositions were
known.
The percent volume was then shown as a function of the mole fraction of ethanol (x).
To make the operation easier, the test solution was poured into the boiler through the
proper intake using a funnel. The refrigeration system was turned on. The SCADA
The heating mantle was turned on, and the set-point was set to a temperature that was
greater than the boiling temperature of the most volatile component and lower than the
boiling temperature of the least volatile component.The reflux valve was turned to the
The distillate flow was stabilized, the solenoid valve AVS-1 was close, volume of
distillate was measured. The distillate tank (using V-2 &3 valve) was drained, at the
same time a sample in the boiler was taken into another beaker (using V-1 valve).
The time that indicates the table of values from the SCADA program was written down,
to be able to find the temperature (and pressure) when the sample was taken. The
sample was allowed to cool down quickly and the vessel that contain them was kept
closed. Every 10 minutes, steps 7 and 8 were repeated, and the time provided by the
software was typed down once again. The samples were labeled, and the time and
location of their collection were specified. The alcohol meter was used to evaluate the
samples. Using tables, the relevant values of the mole fraction of each ingredient in the
mixture were calculated using percent volume and the density value. The mole fractions
XA VA VB Brix (%)
A and B represent Ethanol and Water respectively. Also, V A and VB represent the
volume of ethanol and water as well. XA represent the mole fraction of ethanol.
sample
1 5 16 15.5 92 19.1
X Y
0.0190 0.1700
0.0721 0.3891
0.0966 0.4375
0.1238 0.4704
0.1661 0.5089
0.2337 0.5445
0.2608 0.5580
0.3273 0.5826
0.3965 0.6122
0.5079 0.6564
0.5198 0.6599
0.5732 0.6841
0.6763 0.7385
0.7472 0.7815
0.8943 0.8943
CALIBRATION CURVE
25
20
15
Brix(%)
10
0
0 0.2 0.4 0.6 0.8 1 1.2
EQUILIBRIUM CURVE
1
0.9
0.8
Vapour Mole Fraction, Y
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1
brix
18.2−14.3 18.2−15.6
= x = 0.270
0.4−0.2 0.4−x
14.3−12.8 0.2−x
=
14.3−5 0.2−0.09
x=0.182
19.3−18.7 0.6−x
=
19.3−18.2 0.6−0.4
x=0.491
R 1
y= x+ x ………………………….. (1)
R+1 R+1 D
At the y-intercept of a line, x=0. Hence, equation (1) can be written as:
R 1
y= (0)+ x
R+1 R+ 1 D
1
y= x ……………………………........... (2)
R+1 D
1
y∗¿ x …………………………….... (3)
Rmin +1 D
Minimum reflux ratio can be found by making Rmin the subject of equation (3).
Assumptions: Feed condition is saturated liquid feed, R=1.5 R min, Let composition of
EQUILIBRIUM CURVE
1
0.9
Vapour mole fraction, y
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1
1
0.35= 0.85
R min +1
Rmin =1.4286
Now the actual operating line of rectification can be found by using the equation below.
R 1
y= x+ x
R+1 R+1 D
At the y intercept, x=0. The above equation is further reduced to the equation below.
1
y= x
R+1 D
1
y= 0.85=0.2545
2.1429+1
Figure 3. 4: Determining the number of stages using Maccabee and Thiele Method
3.3 DISCUSSION
Distillation preferentially separates a more volatile component from a less volatile one
vapour produced is richer in the more volatile component. In order to have a larger
degree of separation, multistage contact between the vapour and the liquid phases is
arranged in the distillation column. Some part of the condensate is withdrawn as the top
product and the rest is fed back into the column as reflux which flows down the trays.
The more the number of trays and higher reflux ratio, the better the degree of
separation.
Observations from the calibration curve, shows the mole fraction of ethanol increased
with increasing %Brix to a peak value of 19.3. Ethanol is known to evaporate first from
the mixture and condenses as distillate because of its lower boiling point. As a result,
the mole fraction and volumes of ethanol in the reboiler and distillate showed that
The minimum reflux ratio corresponds to the minimum vapour flow in the tower and
hence the minimum reboiler and condenser sizes. From the results obtained, the
minimum reflux ratios decreased substantially as the mole fraction of ethanol in the
From the McCabe and Thiele graphical method, 15 number of stages were obtained. A
high number of stages signifies that there is a higher degree of separation of ethanol
from the mixture. Therefore, the degree of separation of ethanol from the mixture was
4.0 CONCLUSION
The graph of T as a function of vapour and liquid composition shows that water-ethanol
mixture is not an ideal binary mixture and therefore does not obey Raoult’s law; it
exhibits an azeotropic behavior. The number of trays in a distillation column affects the
quality of products at the top and bottoms product of the distillation column and
flask were rinsed well with the appropriate reagent before use.
Careful transfer of test solution into the boiler to prevent spillage and damage to
distillation unit.
help accuracy.
Bottom liquids were allowed to cool before the percentage volumes were
measured.
The ebullition points of Ethanol and distilled water may vary from the standard
In continuous distillation, the mixture is fed continuously into the column and separated
gravity acts on the liquid to run down the column whilst the vapour runs up the column.
leaving the top stage is totally condensed to give a liquid distillate product and liquid
reflux that is returned to the top stage; a partial reboiler in which liquid from the bottom
stage is partially vaporized to give a liquid bottoms product and vapour boil-up that is
The feed, which contains a more volatile component-the light key, LK and a less
volatile component-the heavy key, HK-enters the column at a feed stage, f. At the feed-
stage pressure, the feed may be liquid, vapour, or a mixture of liquid and vapour, with
its overall mole fraction composition with respect to the light component. Vapour rising
in the section above the feed (called the enriching or rectifying section) is washed with
liquid to remove or absorb the heavy key. In the section below the feed (stripping or
exhausting section), the liquid is stripped of the light key by the rising vapour. Inside
the tower, the liquids and vapours are always at their bubble points and dew points,
respectively, so that the highest temperatures are at the bottom, the lowest at the top.
In Batch Distillation, the mixture is added to the distillation unit at the start of the
process and distillates are taken sequentially in time during the process and the bottoms
fraction is removed at the end. A bottom receiver Reboiler which is charged with the
feed to be processed and which provides the heat transfer surface. A rectifying column
(either a tray or packed column) superimposed on the reboiler, coupled with either a
connected to the product streams to collect the main and or the intermediate distillate
fractions. Operation of such a column involves carrying out the fractionation until a
desired amount has been distilled off. The overhead composition varies during the
operation and usually a number of cuts are made. Some of the cuts are desired products
(main-cuts) while others are intermediate fractions (off-cuts) that can be recycled to
subsequent batches to obtain further separation. A residual bottom fraction may or may
Each of the fraction streams is taken Only one exit point for distillate is
simultaneously throughout operation and needed and for different chemicals, the
hence needs separate exit for each distillate can be just switched to a
heights
Optimal reflux ratio is easily obtained by Difficult to establish optimal reflux ratio
mixture change
Can only process one mixture at a time. Gives production flexibility where
experienced
Separated fractions are removed Some of the cuts are desired products
to the other.
Schematic Diagram for Batch Rectifier
5. Edibon. (2014). Computer Controlled Biogas Process Unit , with SCADA and PID
6. Katzen, R., Madson, P. W., & Moon, G. D. (1999). Ethanol distillation : the
fundamentals. In The Alcohol Textbook: A reference for the beverage, fuel and
industrial alcohol industries (1st ed., pp. 269–288). Ohio: KATZEN International Inc.