Protocol
Protocol
To determine the Ethylene oxide sterilization process consistently performs as intended by running
the system and recording all relevant information. The Data and Test results mist demonstrate that
the process meets pre-determined specifications and under normal conditions as well as worst case
conditions.
SCOPE:
The scope of validation protocol is to provide sterilization validation strategies for ethylene oxide
sterilization of medical device. This document will shoe two approaches for reducing or eliminating
bioburden on medical device. This will also show test method, sampling method and acceptance
criteria used in validating ethylene oxide sterilization of medical devices.
RESPONSIBILITY:
Person Responsibility
Validation team Preparation of protocol
Organizing validation activities
Collecting the samples and sending to QC
Review and interpretation of final results
Preparation of reports
Quality control Review of proto9col and report
Analysing the test samples
Reporting and interpretation of results
Production Review of protocol and report
Conducting the validation activity as per the protocol
Quality assurance manager Review and approve the validation protocol
LIST OF DOCUMENTS:
PRE-REQUISITE:
The following requisite must be fulfilled before the ethylene oxide sterilization process validation.
3. Duration of Exposure
BI or PCD placement in the product load: - after the product load challenge has been
identified, the BI or PCD positioning and placement can be determined. BIs and PCDs should
be distributed throughout the product load and, as much as possible, in the same orientation
(e.g., vertical).
Following minimum number of Bis/PCDs to be included in each validation cycle (as per ISO
11135): -
Load Temperature: - ISO 11135:2014 requires that the minimum temperature of product
permitted to enter preconditioning be determined. Load temperature prior to entrance to
preconditioning should be determined at the lowest temperature zone. Direct placement of
cold loads into preconditioning can result in excessive water condensation and load wetting,
which can cause product damage and reduced lethality. In process validation it requires to
simulate the worst condition to which the load will ever be exposed. The anticipated load
temperature extremes can be simulated during validation by following techniques described
in the appendices 1.
Identification of the worst case location or cold spot: - temperature is the easiest variable to
measure and monitor, therefore temperature is used as an indicator of the worst case
location in the sterilization load. During the preconditioning or conditioning phase
temperature profile against time of the sterilization load measured and based on that data
the worst case location or cold spot is identified.
Place Biological Indicator (BI) with 106 spores of Bacillus atropheus in the PCDs and product
samples in each pallet with humidity and temperature sensor. (place PCDs to cold spot of the
product load).
If product sterility samples show survival, repeat the run on new load with elevated gas exposure
time until no survival in product and growth in Bis (evaluation of BI appropriateness)
Sterility test should show no growth for both product and BIs samples
Perform one full cycle (routine process exposure time) on the aerated sterilization load.
Remove the product sample and PCDs and perform following test on product sample
b) Bioburden /BI method: - this method required that bioburden level be demonstrated to be
relatively consistent over time and the resistance of the bioburden be shown to be equal to,
or less resistant then, the resistance of the biological indicator. During the bioburden
assessment, if the characterization of bioburden performed and level of bioburden is less
than or equal to 100 CFU which indicating a lesser challenge than the BI than the following
steps can be followed.
Establish worst case location in the product load based on temperature distribution and
humidity of load
Place BI within known amount of microorganism in the PCDs, temperature and humidity sensor
Perform 5 fractional cycles with graded exposure time (e.g. 0, 4, 8, 12, 16 minutes)
Calculate rate of inactivation (D value) using survival curve or fractional negative method
Place product samples and PCDs in each pallet with temperature and humidity sensor
Run 3 half cycle according to prescribed cycle parameter
Aeration
Remove product samples and PCDs and perform sterility test on product sample and BIs. Additionally
perform fungi statis/Bacteriostasis test.
Perform one full cycle (routine process exposure time) on the aerated sterilization load. Place
addition product sample for the additional test.
Remove the product sample and PCDs and perform following test on product sample
Maintenance of Validation: -
Requalification: -
Initial evaluation: -
Humidification: -
Heat
Acceptance Criteria met (yes / no)
Comments
Duration of exposure
Vacuum rate
Aeration duration
Anticipated load temperature extremes during transport, handling, and storage can be simulated by
following techniques: -
a) Use of cold storage – For example, the product load is stored under refrigeration or
deliberately induced cold temperature. The temperature of the storage area should be less
than or equal to the lowest temperature the product is expected to be exposed to
throughout the year. Product temperature data are recorded during the storage time. The
time in storage should be equal to or greater than the maximum time any product load is
expected to exist under such conditions, or the maximum product temperature while in
storage becomes the minimum acceptable product temperature for a load to be admitted to
the preconditioning phase. This applies to preconditioning or to conditioning when
preconditioning is not used. The use of refrigeration can result in unrealistically low humidity
levels, with the resultant desiccation of the load. A desiccated load can be much more
difficult to sterilize than a non-desiccated load.
c) Seasonal Validation – preconditioning validation can target those seasons that present the
most extreme temperature and humidity conditions. For example, a validation performed
during the coldest part of the year (which would also present the lowest ambient humidity
level) could yield minimum preconditioning parameters valid for routine production cycles
during the entire year. However, in cases of validations performed during the summer, it is
advisable to validate the precondition process with simulated cold storage, mathematical
modeling or to repeat the preconditioning validation at least one during the winter to
confirm the validity of the parameters.
Volume Fluid
Appendix 3 (GLOSSARY)
Aeration
Part of the sterilization process during which ethylene oxide and/ or its reaction products desorb
(outgas) from the medical device until predetermined levels are reached. This may be performed
within the sterilizer and/or in a separate chamber or room.
Aeration area
Either a chamber or a room in which aeration occurs. The temperature in the aeration area is
elevated and controlled to assure that the aeration process is accelerated and repeatable. Air is
recirculated in the room to maintain food heat distribution with a portion of the air stripped off and
cleansed with the pollution control equipment. The stripping of the air reduces the amount of free
ethylene oxide resident in the rooms, thus prevents recontamination of product load.
Bacteriostasis/Fungi stasis Test
Test performed to evaluate the presence of microbial growth inhibiting properties of a healthcare
product. This test assures that any reaction of the growth medium with the different materials used
in the manufacture of the device will not mask or prevent growth from any viable organism
remaining on the device after being subjected to a sterilization process.
Bioburden
Population of viable microorganisms on a raw material, component, finished product, and/or
package. Unlike the gamma irradiation process, bioburden data is not utilized in ethylene oxide
sterilization to determine the sterilization process. It is collected to quantify the amount of product
contamination, which is then compared to the population of the biological indicator used in the EO
sterilization process.
Biological Indicator
Inoculated carrier contained within its primary pack ready for use providing a defined resistance to
the specified sterilization process.
Calibration
Comparison of a measurement system or device of unknown accuracy to a measurement system or
device of unknown accuracy (traceable to national standards) to detect, correlate, report, or
eliminate by adjustment, any variation from the required performance limits of the unverified
measurement system or device. All critical measuring devices utilized on ethylene oxide sterilizers
which may impact the quality of the process are calibrated to traceable national or international
standards.
Chamber
Enclosed area which only accommodates sufficient product to fill the sterilizer. EO chambers are
constructed of steel or stainless steel and are designed to withstand the extreme pressures and
elevated temperatures utilized in the EO sterilization process.
Commissioning
Obtaining and documenting evidence that equipment has been provided and installed in accordance
with its specification and that it functions within predetermined limits when operated in accordance
with operational instructions. All EO sterilization equipment which complies with ANSI/AAMI/ISO
11135 is commissioned.
Conditioning
Treatment of product within the sterilization cycle, but prior to sterilant admission, to attain a
predetermined temperature and relative humidity. This part of the sterilization process may be
carried out either at atmospheric pressure or under vacuum (see also preconditioning).
Critical parameters
Parameters identified as being essential to the sterilization process and requiring monitoring.
D-Value
Time (expressed in minutes) required to achieve inactivation of 90 percent of a population of a test
organism under stated exposure conditions. Also referenced as the D10 Value or decimal reduction
value.
Exposure Time
Time for which the sterilizer chamber is maintained within the specified range for temperature,
sterilant concentration, pressure, and relative humidity. Also, may be referred to as the time for
which a medical device (load) is exposed at the specified sterilization conditions.
Failure
Event in which a component does not perform one or more of its required functions within the
specified limits under specified conditions.
Flushing
Procedure by which sterilant is removed from the load and chamber by either multiple alternate
admissions of filtered air or inert gas and evacuations of the chamber or continuous passage of
filtered air or inert gas through the load and chamber.
Inactivation
Loss of the ability of microorganisms to grow and/or multiply under specified culture conditions.
Indicator
Combination of indicator agent and its substrate in the form in which it is intended to be used.
Inoculated Carrier
Carrier on which a defined number of test organisms have been deposited.
Installation Qualification
Obtaining and documenting evidence that equipment has been provided and installed in accordance
with its specifications and that it functions within predetermined limits when operated in accordance
with the operational instructions.
Manufacturer
Natural or legal person packaging or sterilizing a medical device.
Medical Device
Any instrument, apparatus, appliance, material or other article whether used alone in combination,
including the software necessary for its proper application intended by the manufacturer to be used
for human beings for the purpose of:
Diagnosis, prevention, monitoring, treatment, or alleviation of disease;
Diagnosis, monitoring, treatment, alleviation of, or compensation for injury or handicap;
Investigation, replacement, or modification of the anatomy or of a physiological process;
Control of conception; and which does not achieve its principal intended action in or on the
human body by pharmacological, immunological, or metabolic means, but which may be
assisted in its function by such means.
Microbial Barrier
Ability of the packaging system to prevent the ingress of microorganisms under specified conditions.
Microbiological Challenge
Biological indicators, biological-indicator test packs, or inoculated product that contain known
populations of microorganisms and can be used in testing sterilization cycles.
Overkill Sterilization Process
Process which is sufficient to provide at least a 12-logarthimic reduction or 12 D inactivation of an
appropriately resistant biological indicator with an established D value.
Package Integrity
Unimpaired physical condition of a final package.
Parametric Release
Declaring product as sterile based on physical and/or chemical process data rather than on the basis
of sample testing or biological indicator results.
Performance Qualification
Obtaining and documenting evidence that the equipment, as commissioned, will produce acceptable
product when operated according to the processing specifications.
Positive Sterility Test
Sterility test samples which exhibit detectable microbial growth after incubation.
Preconditioning
Treatment of product prior to the sterilization cycle in a room or chamber to attain specified limits
for temperature and relative humidity.
Preconditioning Area
Either a chamber or a room in which preconditioning occurs.
Process lethality
Capability of the sterilization process to destroy microorganisms.
Process Qualification
Obtaining and documenting evidence that the sterilization process will produce acceptable health
care products.
Product
Generic term used to describe raw materials, intermediate products, subassemblies, and finished
medical devices.
Product Qualification
Obtaining and documenting evidence that the health care product will be acceptable for its intended
use after exposure to the sterilization process.
Qualification
Documented evidence that all prescribed design and performance requirements are met.
Requalification
Repetition of part of all of the validation test requirements for the purpose of reconfirming process
reliability.
Sterilant
Microbicidal agent in the physical form in which it is active.
Sterile
Free from viable microorganism.in practice, no such absolute statement regarding the absence of
microorganisms can be proven (see sterilization).
Sterility
State of being free from viable microorganisms. In practice, no such absolute statement regarding
the absence of microorganisms can be proven (see sterilization).
Sterility Assurance Level (SAL)
Probability of a viable microorganisms being present on a product unit after sterilization. SAL is
normally expressed at 10-n.
Sterility Test
Test performed to determine if viable microorganisms are present.
Sterilization
Validated process used to render a product free from viable microorganisms. In a sterilization
process, the nature of microbial death is described by an exponential function. Therefore, the
presence of viable microorganisms on any individual item can be expressed in terms of probability.
While this probability may be reduced to a very low number, it can never be reduced to zero. The
probability can be expressed as a sterility assurance level (SAL).
Sterilization Cycle
Defined sequence of operational steps designed to achieve sterilization that is carried out in a sealed
chamber. Specifically, for EO sterilization, the treatment in a sealed chamber comprising air removal,
conditioning, injection of sterilant, exposure to ethylene oxide, and removal of ethylene oxide.
Sterilization Load
Goods that are to be or have been sterilized simultaneously in the same sterilization chamber.
Sterilization Process
All treatments which are required to accomplish sterilization, including preconditioning, the
sterilization cycle, and aeration.
Validation
A documented procedure for obtaining, recording, and interpreting the results needed to show that
a process will consistently yield a product complying with predetermined specifications. Validation is
considered as a total process that includes written protocol, evidence that the equipment as installed
meets design criteria and specifications (equipment qualification), use of calibrated instruments to
collect data, and evidence that the equipment can deliver the process within the specified tolerances
under established operating conditions and is reproducible as demonstrated by replicate runs and
process challenges (performance qualification).