0% found this document useful (0 votes)
70 views

Metrel MI 3210 User Manual

Uploaded by

Rotaru Bogdan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
70 views

Metrel MI 3210 User Manual

Uploaded by

Rotaru Bogdan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 60

TeraOhmXA 10 kV

MI 3210
Instruction manual
Version 1.3; Code No. 20 752 185
Distributor:

Manufacturer:

METREL d.d.
Ljubljanska cesta 77
1354 Horjul
Slovenia
web site: http://www.metrel.si
e-mail: [email protected]

Mark on your equipment certifies that this equipment meets the requirements of the EU
(European Union) concerning safety and electromagnetic compatibility regulations

© 2013 METREL

No part of this publication may be reproduced or utilized in any form or by any means
without permission in writing from METREL.

2
MI 3210 TeraOhm XA 10kV Table of contents

Table of contents
1 General Description ............................................................................................ 5
1.1 Features........................................................................................................ 5
2 Safety and operational considerations ............................................................. 6
2.1 Warnings and notes ...................................................................................... 6
2.2 Battery and charging of Lead – Acid battery ................................................. 8
2.2.1 Precharge .............................................................................................. 9
2.3 Standards applied ....................................................................................... 11
3 Instrument description ..................................................................................... 12
3.1 Instrument casing ....................................................................................... 12
3.2 Operator’s panel ......................................................................................... 12
3.3 Accessories ................................................................................................ 14
3.3.1 Test leads ............................................................................................ 14
3.4 Display organization.................................................................................... 16
3.4.1 Measurement result window ................................................................ 16
3.4.2 Measurement control window .............................................................. 17
3.4.3 Message window ................................................................................. 18
3.4.4 Battery, time and communication indication ........................................ 19
3.4.5 Measurement result line ...................................................................... 20
3.4.6 Graphical presentation of measured data ............................................ 20
3.4.7 Backlight operation .............................................................................. 20
4 Main menu ......................................................................................................... 21
4.1 Instrument Main menu ................................................................................ 21
4.2 Custom tests ............................................................................................... 22
4.2.1 Creating custom test ............................................................................ 22
4.3 Memory menu ............................................................................................. 23
4.3.1 Saving results ...................................................................................... 23
4.3.2 Recalling results .................................................................................. 24
4.3.3 Deleting results .................................................................................... 25
4.3.4 Deleting custom test ............................................................................ 25
4.3.5 Clearing complete memory content ..................................................... 26
4.4 Settings menu ............................................................................................. 26
4.4.1 Language selection.............................................................................. 27
4.4.2 Initial Settings selection ....................................................................... 27
4.4.3 Time selection...................................................................................... 27
4.4.4 Date selection ...................................................................................... 27
4.4.5 Transfer mode ..................................................................................... 28
4.4.6 Contrast selection ................................................................................ 28
4.4.7 Warning sound..................................................................................... 28
4.4.8 Graph selection.................................................................................... 28
4.4.9 Break Down selection .......................................................................... 29
4.4.10 Instrument info ..................................................................................... 29
4.5 Help Menu .................................................................................................. 29
5 Measurements ................................................................................................... 30
5.1 Generally information about high voltage testing ........................................ 30
5.1.1 The purpose of insulation tests ............................................................ 30
5.1.2 DC vs. AC testing voltage .................................................................... 30
5.1.3 Typical insulation tests......................................................................... 30

3
MI 3210 TeraOhm XA 10kV Table of contents

5.1.4 Electrical representation of insulating material .................................... 31


5.2 Some application examples ........................................................................ 32
5.2.1 Basic Insulation resistance test ........................................................... 32
5.2.2 Voltage dependence test – Step Voltage Test ..................................... 32
5.2.3 Time dependence test – Diagnostic Test............................................. 32
5.2.4 Withstanding Voltage Test ................................................................... 34
5.3 Guard terminal ............................................................................................ 35
5.4 Averaging options ....................................................................................... 36
5.4.1 The purpose of averaging .................................................................... 36
5.4.2 Example of averaging .......................................................................... 37
5.5 Measurement menu .................................................................................... 38
5.6 Insulation Resistance Measurement ........................................................... 39
5.6.1 Set Limit ............................................................................................... 41
5.7 Diagnostic test ............................................................................................ 42
5.7.1 Dielectric Absorption Ratio (DAR)........................................................ 44
5.7.2 Polarization Index (PI) ......................................................................... 44
5.7.3 Dielectric Discharge Testing (DD)........................................................ 45
5.8 Step Voltage testing .................................................................................... 47
5.9 Withstanding Voltage Test .......................................................................... 49
5.10 True RMS Voltmeter ................................................................................... 51
6 Communication ................................................................................................. 52
7 Maintenance ...................................................................................................... 53
7.1 Batteries insertion and replacement ........................................................... 53
7.2 Cleaning...................................................................................................... 55
7.3 Periodic calibration...................................................................................... 55
7.4 Service ........................................................................................................ 55
8 Technical specifications ................................................................................... 56
8.1 True RMS voltmeter .................................................................................... 58
8.2 General data ............................................................................................... 59
Appendix A – Remote Control ............................................................................... 60

4
MI 3210 TeraOhm XA 10kV General Description

1 General Description
1.1 Features
TeraOhm XA 10kV (MI 3210) is a portable battery or mains powered test instrument
with excellent IP protection (IP65), intended for diagnosing of Insulation Resistance by
using high test voltages of up to 10 kV. It is designed and produced with the extensive
knowledge and experience acquired through many years of working in this field.

Available functions and features offered by the TeraOhm XA 10kV meter:


 Wide measuring range (5 kΩ ... 20 TΩ);
 Insulation Measurement;
 Step Voltage Test;
 Withstanding Voltage Test (DC) up to 10 kV;
 Voltage and frequency measurement up to 550 V TRMS
 Polarization Index (PI);
 Dielectric Absorption ratio (DAR);
 Dielectric Discharge ratio (DD);
 Graph R(t);
 Adjustable test voltage (50 V...10 kV) 50 V and 100 V step;
 Programmable timer;
 Automatic discharge of test object after completion of measurement;
 Capacitance measurement;
 Input AC current noise rejection (1 mA@600 V);
 High voltage breakdown detection;
 Limit status;
 Additional averaging of the result (5, 10, 30, 60)
 USB and RS232 communication;
 High overvoltage category CAT IV / 600 V.

A 320x240 dot matrix LCD offers easy-to-read results and all associated parameters.
The operation is straightforward and clear to enable the user to operate the instrument
without the need for special training (except reading and understanding this Instruction
Manual).

Test results can be stored on the instrument. PC software HVLink PRO that is supplied
as a part of standard set enables transfer of measured results to PC where can be
analyzed or printed.

5
MI 3210 TeraOhm XA 10kV Safety and operational considerations

2 Safety and operational considerations


2.1 Warnings and notes
In order to maintain the highest level of operator safety while carrying out various tests
and measurements Metrel recommends keeping your TeraOhm XA 10kV instruments in
good condition and undamaged. When using the instrument, consider the following
general warnings:

 The symbol on the instrument means »Read the Instruction manual with
special care for safe operation«. The symbol requires an action!

 The symbol on the instrument means “Hazardous voltage may be


present at the test terminals!”.
 If the test equipment is used in a manner not specified in this user manual,
the protection provided by the equipment could be impaired!
 Read this Instruction manual carefully, otherwise the use of the instrument
may be dangerous for the operator, the instrument or for the equipment
under test!
 Do not use the instrument or any of the accessories if any damage is
noticed!
 Consider all generally known precautions in order to avoid risk of electric
shock while dealing with hazardous voltages!
 Do not use the instrument in supply systems with voltages higher than
600 V!
 Service intervention or adjustment is only allowed to be carried out by
competent authorized personnel!
 Use only standard or optional test accessories supplied by your distributor!
 Hazardous voltages exist inside the instrument. Disconnect all test leads,
remove the power supply cable and switch off the instrument before
opening the battery compartment.
 All normal safety precautions must be taken in order to avoid risk of electric
shock while working on electrical installations!

6
MI 3210 TeraOhm XA 10kV Safety and operational considerations

Warnings related to measurement functions:

Working with the instrument


 Use only standard or optional test accessories supplied by your distributor!
 Equipment under test must be switched off (i.e. de-energized) before test leads
are connected to the equipment.
 Test tips should only be used for TRMS voltage measurement (CAT IV 600 V).
 Do not use high voltage test lead with tip for TRMS voltage measurement in CAT
III or CAT IV environment. Risk of bridging two high-energy conductors with tip
resulting in arc flash and short circuit.
 Always connect accessories to the instrument and to the test object before
starting high voltage measurement. Do not touch test leads or crocodile clips
during measurement. Only hand-held part of high voltage test lead with tip is
allowed to be touched (held) during measurement.
 Do not touch any conductive parts of equipment under test during the test, risk of
electric shock!
 Make sure that the tested object is disconnected (mains voltage disconnected)
before starting the Insulation Resistance measurement!
 In case of a capacitive test object (long tested cable etc.), automatic discharge of
the object may not be done immediately after finishing the measurement –
“Please wait, discharging” message will be displayed.
 Do not connect the instrument to a mains voltage different from the one defined
on the label adjacent to the mains connector, otherwise the instrument may be
damaged.
 Do not connect test terminals to an external voltage higher than 600 V DC or AC
(CAT IV environment) to prevent any damage to the test instrument!

Handling with capacitive loads


 Note that 40 nF charged to 1 kV or 9 nF charged to 10 kV are hazardous live!
 Never touch the measured object during the testing until it is totally discharged.
 Maximum external voltage between any two leads is 600 V (CAT IV
environment).

Warnings related to Batteries:


 Never dispose of the batteries in a fire as it may cause them to explode or
generate a toxic gas
 Do not attempt to disassemble, crush or puncture the batteries in any way,
as it could cause leakage of sulfuric acid that could cause injury.
 Do not short circuit or reverse polarity the external contacts on a battery.
 Keep the battery away from children.
 Avoid exposing the battery to excessive shock/impacts or vibration.
 Do not use a damaged battery.
 Instrument uses Valve Regulated (sealed) Lead – Acid battery (LC –
R123R4PG). It is designed to automatically charge and maintain battery
capacity according to the use.

7
MI 3210 TeraOhm XA 10kV Safety and operational considerations

2.2 Battery and charging of Lead – Acid battery


The instrument is designed to be powered by rechargeable Lead – Acid battery or with
mains supply. The LCD contains an indication of battery condition and the power source
(upper left section of LCD). In case the battery is too weak the instrument indicates this
as shown in Figure 2.1.

Symbol:
Indication of low battery.

Figure 2.1: Battery test

The battery is charged whenever the power supply is connected to the instrument. The
power supply socket is shown in Figure 2.2. Internal circuit controls (CC, CV) charging
and assures maximum battery lifetime. Nominal operating time is declared for battery
with nominal capacity of 3.4 Ah.

Figure 2.2: Power supply socket (C7)

The instrument automatically recognizes the connected power supply and begins
charging.

Symbol:
Indication of battery charging

Figure 2.3: Charging indication

Battery and charging characteristic Typical


Battery type LC-R123R4PG
Charging mode CC / CV
Nominal voltage 12,0 V
Rated capacity 3,4 Ah
Max charging voltage 14,0 V
Max charging current 1,2 A
Max discharge current 2,5 A
Typical charging time 4 hours

8
MI 3210 TeraOhm XA 10kV Safety and operational considerations

Typical charging profile which is also used in this instrument is shown in Figure 2.4.

Figure 2.4: Typical charging profile

where:
VREG ............................... Battery charging voltage
VLOWV.............................. Precharge threshold voltage
ICH................................... Battery charging current
ICH/8................................. 1/8 of the charging current

2.2.1 Precharge
On power up, if the battery voltage is below the VLOWV threshold, the charger applies 1/8
of the charging current to the battery. The precharge feature is intended to revive deeply
discharged battery. If the VLOWV threshold is not reached within 30 minutes of initiating
precharge, the charger turns off and a FAULT is indicated.

Figure 2.5: No battery indication

Note:
 As a safety backup, the charger also provides an internal 5-hour charge timer for
fast charge.

9
MI 3210 TeraOhm XA 10kV Safety and operational considerations

Typical charging time is 4 hours in the temperature range of 5°C to 60°C.

Current

TWARM
TCOOL

THTF
TLTF

Figure 2.6: Typical charging current vs temperature profile

where:
TLTF................................. Cold temperature threshold (typ. -15°C)
TCOOL .............................. Cool temperature threshold (typ. 0°C)
TWARM ............................. Warm temperature threshold (typ. +60°C)
THTF ................................ Hot temperature threshold (typ. +75°C)

The charger continuously monitors battery temperature. To initiate a charge cycle, the
battery temperature must be within the TLTF to THTF thresholds. If battery temperature is
outside of this range, the controller suspends charge and waits until the battery
temperature is within the TLTF to THTF range.
If the battery temperature is between the TLTF and TCOOL thresholds or between the
TWARM and THTW thresholds, charge is automatically reduced to ICH/8 (1/8 of the charging
current).

10
MI 3210 TeraOhm XA 10kV Safety and operational considerations

2.3 Standards applied


The TeraOhm XA 10kV instrument is manufactured and tested in accordance with the
following regulations:
Electromagnetic compatibility (EMC)
EN 61326 Electrical equipment for measurement, control and laboratory
use – EMC requirements Class A
Safety (LVD)
EN 61010 - 1 Safety requirements for electrical equipment for measurement,
control and laboratory use – Part 1: General requirements
EN 61010 - 2 - 030 Safety requirements for electrical equipment for measurement,
control and laboratory use – Part 2-030: Particular requirements
for testing and measuring circuits
EN 61010 - 2 - 033 Safety requirements for electrical equipment for measurement,
control and laboratory use – Part 2-033: Particular requirements
for hand-held multimeters and other meters, for domestic and
professional use, capable of measuring mains voltage
EN 61010 - 031 Safety requirements for hand-held probe assemblies for electrical
measurement and test.
Some further recommendations
IEEE 43 – 2000 Recommended Practice for Testing Insulation Resistance of Rotating
Machinery:
 1 MΩ + 1 MΩ / 1000 V rating of equipment for insulation systems prior to 1970;
 5 MΩ for random wound motors under 600 Volts;
 100 MΩ for form wound motors over 600 Volts, and armatures;
IEC 60439-1 Low-voltage switchgear and control gear assemblies - Part 1:
Type-tested and partially type-tested assemblies:
 Scope of the insulation resistance test: alternative method for verification of
dielectric properties by insulation resistance measurement;
 Description: a d.c. test voltage (500 V) is applied to the insulation and its
resistance is measured. Insulation is proper, if its resistance is high enough (1000
Ω / V of circuit rated voltage);

IEC 61558 Isolating transformers and safety isolating transformers:


 Test voltage: 500 V, measurement period: 1 min;
 Minimum insulation resistance for basic insulation: 2 MΩ;
 Minimum insulation resistance for supplementary insulation: 5 MΩ;
 Minimum insulation resistance for reinforced insulation: 7 MΩ;

Note:
Immunity to radiated RF fields (Field strength: 10V/m, Modulation: AM, 80%, 1 kHz)
Voltage Range Operational conditions Disturbance < 5 % Disturbance > 5 %
50 V 100 MΩ 300 MHz ÷ 900 MHz 900 MHz ÷ 1 GHz
1000 V 200 GΩ / 300 MHz ÷ 1 GHz
10000 V 200 GΩ 300 MHz ÷ 600 MHz 600 MHz ÷ 800 MHz

Note about EN and IEC standards:


 Text of this manual contains references to European standards. All standards of
EN 6XXXX (e.g. EN 61010) series are equivalent to IEC standards with the same
number (e.g. IEC 61010) and differ only in amended parts required by European
harmonization procedure.

11
MI 3210 TeraOhm XA 10kV Instrument description

3 Instrument description
3.1 Instrument casing
The instrument is housed in a plastic box that maintains the protection class defined in
the general specifications.

3.2 Operator’s panel


The operator’s panel is shown in Figure 3.1 below.

Figure 3.1: The operator’s panel

Keypad Section:
1 START / STOP Start or stop measurement.
Switches the instrument power on or off.
2 ON / OFF The instrument automatically turns off 15 minutes after the
last key was pressed.
3 MEM Store / recall / clear tests in memory of the instrument.
To enter set-up mode for the selected function or to select
4 SELECT
the active parameter to be set.
5,6   Select an option upward, downward.
7,8   Decrease, increase the selected parameter.
9 ESC Exit the selected mode.
Turn the display backlight ON or OFF.
10 ☼ Instrument’s RESET (hold key for 5 s or more).
Display Section:
20 High voltage warning light (red).
Terminal Section:
11 + Rx Positive high voltage output terminal.

12
MI 3210 TeraOhm XA 10kV Instrument description

12,13 GUARD Guard input terminals.


14 - Rx Negative high voltage input terminal.
Power and Communication Section:
19 C7 Input power supply socket (C7)
18 USB USB communication port (standard USB connector - type B)
RS232 communication port (standard RS232 9-pin D
18 RS232
female)

Warnings!
 Maximum allowed voltage between any test terminal and ground is 600 V!
 Maximum allowed voltage between test terminals is 600 V!
 Use original test accessories only!

13
MI 3210 TeraOhm XA 10kV Instrument description

3.3 Accessories
The accessories consist of standard and optional accessories. Optional accessories can
be delivered upon request. See attached list for standard configuration and options or
contact your distributor or see the METREL home page: http://www.metrel.si.

3.3.1 Test leads


The standard length of high voltage shielded test lead with tip is 2 m. The standard
length of high voltage shielded test leads with (red, black) banana connectors is 3 m;
optional lengths are 8 m and 15 m. For more details see attached list for standard
configuration and options or contact your distributor or see the METREL home page:
http://www.metrel.si.
All test leads are made of high voltage shielded cable, because shielded cable provides
higher accuracy and immunity to disturbance of measurements that can occur in
industrial environments.

High voltage shielded test lead with High voltage tip

Application notes:
This test lead is designed for hand held testing
of insulation.
Insulation ratings:
 Hand held part 10 kV d.c. (reinforced
insulation)
 High voltage banana connector (red)
for connection to the instrument: 10 kV
d.c. (basic insulation);
 Guard banana connector (green): 600
V CAT IV (reinforced insulation);
 Cable (yellow): 12 kV (shielded).

Figure 3.2: HV test lead with high voltage tip

14
MI 3210 TeraOhm XA 10kV Instrument description

High voltage shielded test leads with High voltage alligator clips.

Application notes:
This test leads are designed for diagnostic
testing of insulation. They can also be used
for hand held testing with test voltages up to
5 kV d.c..
Insulation ratings:
 High voltage banana connectors (red,
black): 10 kV d.c. (basic insulation),
5 kV d.c. (reinforced insulation);
 Alligator clips (red, black):
10 kV d.c. (basic insulation),
5 kV d.c. (reinforced insulation);
 Guard banana connector (green): 600
V CAT IV (reinforced insulation);
 Cable (yellow): 12 kV (shielded).

Figure 3.3: HV test leads with alligator clips

Guard test lead with alligator clips

Insulation ratings:
 Guard test lead with banana
connectors (green): 600 V CAT IV
(reinforced insulation);
 Alligator clip (green): 600 V CAT IV
(reinforced insulation)
Figure 3.4: Guard lead with alligator clip

Test tips

Application notes:
 Test tips applied on High voltage
shielded test leads are designed for
CAT IV 600 V TRMS voltage
measurement.
Figure 3.5: Test tips

15
MI 3210 TeraOhm XA 10kV Instrument description

3.4 Display organization

Figure 3.6: Typical function display and graph screen

1 Measurement result window


2 Measurement control window
3 Messages window
4 Battery, time and communication indication
5 Measurement result line
6 Graphical presentation of measured data

3.4.1 Measurement result window


Measurement window shows all relevant data during measurement campaign.

Figure 3.7: Measurement window

Measured Insulation Is shown in the center of a screen with the largest font. During
resistance measurement campaign this result is refreshed every few
seconds. When the measurement is completed result is hold on
screen, until new measurement is started.
Bar graph Graphically represents measured insulation resistance in respect
to the measurement range. It also displays limit value if it is
enabled.
U Shows the output voltage. During measurement campaign this
result is refreshed every few seconds. When the measurement is
completed result is hold on screen, until new measurement is
started.

16
MI 3210 TeraOhm XA 10kV Instrument description

I Shows the input current. During measurement campaign this


result is refreshed every few seconds. When the measurement is
completed result is hold on screen, until new measurement is
started.
C Shows the capacitance measured at the output terminals. Value
of capacitance is measured during the final discharge of the test
object.
R1, R2, R3 Show resistances measured at Timer1, Timer2 and Timer3.
When the measurement is completed result is hold on screen,
until new measurement is started (appears only in Diagnostic
Test).
R1, R2, R3,R4, R5 Show resistances measured at steps one to five. When the
measurement is completed result is hold on screen, until new
measurement is started (appears only in Step Voltage Test).
U1, U2, U3, U4, U5 Show voltages measured at steps one to five. When the
measurement is completed result is hold on screen, until new
measurement is started (appears only in Step Voltage Test).
DAR Show the Dielectric Absorption Ratio. When the measurement is
completed result is hold on screen, until new measurement is
started (appears only in Diagnostic Test).
PI Show the Polarization Index. When the measurement is
completed result is hold on screen, until new measurement is
started (appears only in Diagnostic Test).
DD Show the Dielectric Discharge result. When the measurement is
completed result is hold on screen, until new measurement is
started (appears only in Diagnostic Test).
f Shows frequency of the measured voltage (appears only in True
RMS Voltmeter).

t Shows test time (mm:ss).

3.4.2 Measurement control window


Control window permit user to modify control measurement parameters.

Figure 3.8: Control window

Un Allows user to set the desired test voltage.

17
MI 3210 TeraOhm XA 10kV Instrument description

Timer1 Allows user to set the desired measurement duration in


Insulation Resistance test.
Delay for starting DAR measurement in Diagnostic test.
(mm:ss) – step 1 s (max time 99 min).
Timer2 Delay for starting PI measurement (mm:ss) – step 1 s (max time
99 min).

Timer3 Allows user to set the desired measurement duration (mm:ss) –


step 1 s (max time 99 min).

DD Allows user to enable or disable Dielectric Discharge


measurement.

Itrgg Allows user to set the desired trigger level – step 100 μA (max
current 5 mA).

Tstart Allows user to set the start test voltage time (mm:ss) – step 1s
(max time 99 min).

Tend Allows user to set the constant end test voltage time (mm:ss) –
step 1 s (max time 99 min).

Ustart Allows user to set the desired start test voltage.

Uend Allows user to set the desired end test voltage.

Tramp Allows user to set the duration of the test ramp (mm:ss) – step 1
s (max time 99 min).

HI Lim Allows user to set the high limit (value is evaluated at the end of
the measurement).

AVG Allows user to set the additional averaging of the result (OFF, 5,
10, 30, 60).

3.4.3 Message window


In the message field warnings and messages are displayed.

Figure 3.9: Message window

18
MI 3210 TeraOhm XA 10kV Instrument description

High voltage is present on measuring terminals (> 50 V rms).

Test result can be saved.

AC noise is present on the measuring terminals (+ Rx, - Rx).

A Breakdown or Over voltage event has accrued.

Instrument is overheated. Measurement process is disabled.

Graph is enabled.

Graph logging is enabled (internal Flash memory).

Internal Flash memory is full (graph logging is disabled).

Measurement result is within defined limits.

Measurement result is out of defined limits.

3.4.4 Battery, time and communication indication


These symbols indicate the charge condition of a battery, charger connection and
communication status. Time indication is also present.

Battery capacity indication.

Low battery.
Recharge the battery.

Recharging in progress (if mains supply cable is connected).

Time indication (hh:mm).

USB communication is enabled.

Bluetooth communication is enabled.

Note:
 Date and time is attached to each stored result.

19
MI 3210 TeraOhm XA 10kV Instrument description

3.4.5 Measurement result line

R Shows Insulation resistance. During measurement campaign this


result is refreshed every few seconds. When the measurement is
completed it represents Insulation resistance at cursor position.

U Shows the output voltage. During measurement campaign this


result is refreshed every few seconds. When the measurement is
completed it represents output voltage at cursor position.
t Shows test time (mm:ss). When the measurement is completed
it represents time at cursor position.

Note:
 Averiging in measurement result line is disabled during measurement campaign
in all functions regardless of settings.

3.4.6 Graphical presentation of measured data

Figure 3.10: Graph screen

Measured or averaged Insulation resistance values in relation to time of measurement


are represented in this two dimensional R/t diagram. Online plotting, during
measurement, can also be observed. When the measurement is completed cursor is
attached to the graph for detail analysis.

3.4.7 Backlight operation


After turning the instrument ON the LCD backlight is automatically turned ON. It can be
turned OFF and ON by simply clicking the ☼ (LIGHT) key.

Note:
 If you press and hold the Light (☼) key for approximately 5 s the instrument will
RESET!

20
MI 3210 TeraOhm XA 10kV Main menu

4 Main menu
4.1 Instrument Main menu
From the Main menu of the instrument there are five options available: Measurements,
Custom tests, Memory menu, Settings menu and Help menu.

Figure 4.1: Instrument Main menu

Keys:
Select one of the following menu items:
<Measurements> See chapter 4.5;
  <Custom tests> See chapter 4.2;
<Memory menu> Memory management, see chapter 4.3;
<Settings menu> Setup of the instrument, see chapter 4.4;
<Help menu> Help screens, see chapter 4.6;
SELECT Confirms selection.

21
MI 3210 TeraOhm XA 10kV Main menu

4.2 Custom tests


This menu contains a list of custom prepared tests. The commonly used tests are added to
the list by default or user. Up to 30 custom tests can be pre-programmed.

Figure 4.2: Custom Test Menu

Keys:
  Select one of the following menu items:
SELECT Confirms selection.
ESC Returns to the Main menu.

4.2.1 Creating custom test


The user has an option to save any custom made parameter settings.
Simply enter the desired measurement, edit the test parameters end press MEM key.
The insert text screen will appear if no result is left for save.

Figure 4.3: Insert text screen

Keys in Insert text screen:


  Selects a letter.
SELECT Selects the next letter.
MEM Confirm name and return to the selected measurement.
Deletes last letter.
ESC
Returns to selected measurement without changes.

22
MI 3210 TeraOhm XA 10kV Main menu

4.3 Memory menu


Measurement result with all relevant parameters can be stored into the instrument’s
memory. The instrument’s memory space is divided into 3 levels: OBJECT, DUT and
LINE. The OBJECT, DUT and LINE level can contain up to 199 locations.

Figure 4.4: Memory Menu

Keys:
  Select one of the following items.
SELECT Confirms selection.
ESC Returns to the Main menu.

4.3.1 Saving results


After the completion of a test, the results and parameters are ready for storing. By
pressing the MEM key the user can enter storing menu.

Figure 4.5: Storing Menu

Keys:
  Selects OBJECT, DUT and LINE number.
  Jump to a different location.
Saves measurement result to selected location and returns to the
MEM
Measurement result screen.
ESC Returns to the Measurement result screen without save.

The instrument will beep in order to indicate that result is successfully saved into the
memory.

Note:
 Every stored test result also includes date and time stamp (dd:mm:yyyy, hh:mm).

23
MI 3210 TeraOhm XA 10kV Main menu

4.3.2 Recalling results


To enter Recall results menu in Memory menu SELECT key should be pressed.

Figure 4.6: Recall Results menu Figure 4.7: Recalled result screen

Keys in Recall menu:


  Selects one of the following items OBJECT / DUT / LINE.
  Decrees or increase the parameter.
SELECT Recalls result in selected location.
ESC Returns to the Memory menu.

Keys in Recall result screen:


  Toggles between the saved results under selected OBJECT / DUT / LINE.
SELECT Enters Recall result graph screen if available.
ESC Returns to the Recall results menu.

Keys in Recall result graph screen:


  Scroll the cursor along logged data.
SELECT Returns to Recall result screen.
ESC Returns to the Recall results menu.

24
MI 3210 TeraOhm XA 10kV Main menu

4.3.3 Deleting results


To enter Delete results menu in Memory menu SELECT key should be pressed. Single
measurement or all measurements under selected OBJECT, DUT and LINE can be
deleted.

Figure 4.8: Delete all measurements under selected Object

Keys in Delete results menu:


  Selects one of the following items OBJECT / DUT / LINE.
  Decrees or increase the parameter.
SELECT Enters Delete confirmation screen.
MEM Enters measurements field for deleting individual measurements.
ESC Returns to the Memory menu.

Keys in measurements field for deleting individual measurements:

  Select measurement to be deleted.


SELECT Enters Delete confirmation screen.
ESC / MEM Returns to the structure field.

Keys in Delete confirmation screen:


SELECT Deletes result(s) in selected location.
ESC Returns to the Delete results menu without changes.

4.3.4 Deleting custom test


A single custom test can be deleted by simply selecting the test from the list of all tests
and pressing the MEM key.

Figure 4.9: Deleting custom test

Keys in Deleting custom test menu:


  Selects one of the following test.
SELECT Delete selected test.
ESC Returns to the Memory menu.

25
MI 3210 TeraOhm XA 10kV Main menu

4.3.5 Clearing complete memory content


When selecting the Clear all memory function in Memory menu all the memory content
will be deleted.

Keys in Clear all memory confirmation screen:


  Toggles between YES or NO.
SELECT Clears all memory content (if YES is selected).
ESC Returns to the Main menu without changes.

4.4 Settings menu


In the Settings menu different parameters and settings of the instrument can be viewed or
set.

Figure 4.10: Settings menu

Keys:
Select the setting to adjust or view:
<Language> instrument language;
<Initial Settings> factory settings;
<Time> time settings;
<Date> date settings;
  <Transfer Mode> communication mode selection;
<Contrast> LCD contrast settings;
<Warning Sound> enables or disables high voltage warning sound;
<Graph> switches graph on or off;
<Break Down> enables or disables breakdown detection;
<Instrument Info> basic instrument information.
SELECT Confirms selection.
ESC Returns to the Main menu.

26
MI 3210 TeraOhm XA 10kV Main menu

4.4.1 Language selection


The instrument language can be set.
Keys:
  Toggles between different languages (changes are stored automatically).

Note:
 No confirmation is needed to set the desired language.

4.4.2 Initial Settings selection


In this menu the following instrument parameters can be set to their initial values:
 All measurement parameters;
 Language;
 Transfer Mode;
 Contrast settings;
 Custom Measurements

Keys:
  Toggles between YES or NO.
SELECT Confirms selection and restart of the instrument (if YES is selected).
ESC Returns to Main menu without changes.

Note:
 Applying the initial settings will re-boot the instrument.

4.4.3 Time selection


In this selection the time of the instrument can be set.
Keys:
  Decrees or increase the parameter (changes are stored automatically).
SELECT Select the parameter to be changed.
Time is attached to each stored result.

4.4.4 Date selection


In this selection the date of the instrument can be set.
Keys:
  Decrees or increase the parameter (changes are stored automatically).
SELECT Select the parameter to be changed.
Date is attached to each stored result.

Note:
 If the battery is removed the instrument time and date will be lost.

27
MI 3210 TeraOhm XA 10kV Main menu

4.4.5 Transfer mode


The instrument communication mode can be set.
Keys:
  Toggles between RS232, USB and Bluetooth.

Note:
 No confirmation is needed to set the desired transfer mode.

4.4.6 Contrast selection


In this selection the contrast of the display can be set.
Keys:
  Sets contrast value (changes are stored automatically).

Note:
 When using the instrument in a cold environment contrast level should be
increased.

4.4.7 Warning sound


In this selection the Warning sound can be set. When enabled the Warning sound will
beep when high voltage (>=50V rms) is present on the input terminals +Rx to –Rx.

Keys:
  Toggles between YES or NO (changes are stored automatically).

4.4.8 Graph selection


In this selection the graph R(t) logging can be set.
Keys:
  Toggles between YES or NO (changes are stored automatically).

Note:

 When icon is displayed the internal flash memory is full and graph logging is
disabled.

28
MI 3210 TeraOhm XA 10kV Main menu

4.4.9 Break Down selection


In this selection the Break down can be set. When triggered the Break down circuit
automatically stops the measuring procedure.

Keys:
  Toggles between YES or NO (changes are stored automatically).

Notes:
 Break down is not active in Withstanding Voltage test and in the charging state of
the high voltage generator.
 Break down is active if output voltages higher or equal then 1kV are set!

4.4.10 Instrument info


In this menu the following instrument data is shown:

 firmware version;
 serial number;
 calibration date;
 battery type.

4.5 Help Menu


The Help menu contains schematic diagrams for illustrating how to properly connect the
instrument to the various test objects.
Keys in help menu:
  Selects next / previous help screen.
ESC Returns to the Settings menu.

Figure 4.11: Examples of help screens

29
MI 3210 TeraOhm XA 10kV Measurements

5 Measurements
5.1 Generally information about high voltage testing
5.1.1 The purpose of insulation tests
Insulating materials are important parts of almost every electrical product. The material’s
properties depend not only on its compound characteristics but also on temperature,
pollution, moisture, ageing, electrical and mechanical stress, etc. Safety and operational
reliability require the regular maintenance and testing of the insulation material to ensure
it is kept in good operational condition. High voltage tests are used to test insulating
materials.

5.1.2 DC vs. AC testing voltage


Testing with a DC voltage is widely accepted as being useful as testing with AC and / or
pulsed voltages. DC voltages can be used for breakdown tests especially where high
capacitive leakage currents interfere with measurements using AC or pulsed voltages.
DC is mostly used for insulation resistance measurement tests. In this type of test, the
voltage is defined by the appropriate product application group. This voltage is lower
than the voltage used in the withstanding voltage test so the tests can be applied more
frequently without stressing the test material.

5.1.3 Typical insulation tests


In general, insulation resistance tests consist of the following possible procedures:
 Simple insulation resistance measurement also called a spot test;
 Measurement of the relationship between voltage and insulation resistance;
 Measurement of the relationship between time and insulation resistance
 Test of residual charge after the dielectric discharge.
The results of this test can indicate whether the replacement of the insulation system is
required.
Typical examples of where testing insulation resistance and its diagnosis are
recommended are transformer and motor insulation systems, cables and other electrical
equipment.

30
MI 3210 TeraOhm XA 10kV Measurements

5.1.4 Electrical representation of insulating material


The Figure 5.1 represents the equivalent electrical circuit of an insulating material.

Itest
surface
+
m aterial
R iss1 Itest
C pi
R iso
IPI
C iso
G u ard R pi
R iss2
ICiso
IRiso IRiss
-

Figure 5.1: Insulating material

where:
Riss1 and Riss2 ................. Surface resistivity (position of optional guard connection)
Riso ................................. Actual insulation resistance of material
Ciso ................................. Capacitance of material
Cpi, Rpi ............................ Represents polarization effects
Itest .................................. Overall test current (Itest= IPI+ IRISO+ IRISS)
IPI .................................... Polarization absorption current
IRISO ................................ Actual insulation current
IRISS ................................ Surface leakage current

31
MI 3210 TeraOhm XA 10kV Measurements

5.2 Some application examples


5.2.1 Basic Insulation resistance test
Virtually every standard concerning the safety of electrical equipment and installations
requires the performance of a basic insulation testing. When testing lower values (in the
range of M), the basic insulation resistance (Riso) usually dominates. The results are
adequate and stabilize quickly.
It is important to remember the following:
 The voltage, time and limit are usually given in the appropriate standard or
regulation.
 Measuring time should be set to 60 s or the minimum time required for the
Insulation capacitance Ciso to be charged up.
 Sometimes it is required to take ambient temperature into account and adjust the
result for a standard temperature of 40 C.
 If surface leakage currents interfere with the measurements (see Riss above) use
the guard connection. This becomes critical when measured values are in the G
range.

5.2.2 Voltage dependence test – Step Voltage Test


This test shows if the insulation under test has been electrically or mechanically
stressed. In this instance the quantity and size of insulation anomalies (e.g. cracks, local
breakdowns, conductive parts, etc). is increased and the overall breakdown voltage is
reduced. Excessive humidity and pollution have an important role especially in the case
of mechanical stress.
 The test voltage steps are usually close to those required in the DC withstanding
test.
 Sometimes it is recommended that the maximum voltage for this test should not
be higher than 60 % of the withstanding voltage.
If the results of successive tests show a reduction in the tested insulation resistance the
insulation should be replaced.

5.2.3 Time dependence test – Diagnostic Test

5.2.3.1 Polarization Index – PI


The purpose of this diagnostic test is to evaluate the influence of the polarization part of
insulation (Rpi, Cpi). After applying a high voltage to an insulator the electric dipoles
distributed in the insulator align themselves with the applied electrical field. This
phenomenon is called polarization. As the molecules polarize, a polarization (absorption)
current lowers the overall insulation resistance of the material.
The absorption current (IPI) typically collapses after a few minutes. If the overall
resistance of the material does not increase, this means that other currents (e.g. surface
leakages) dominate the overall insulation resistance.
 PI is defined as the ratio of the measured resistances in two time slots. The most
typical ratio is 10 min value to 1 min value but this is not a rule.

32
MI 3210 TeraOhm XA 10kV Measurements

 The test is typically performed at the same voltage as the insulation resistance
test.
 If the one-minute insulation resistance is greater than 5000 MΩ, then this
measurement may not be valid (new modern types of insulation).
 Oiled paper used in transformers or motors is a typical insulation material that
requires this test.
In general, insulators that are in good condition will show a “high” polarization index
while insulators that are damaged will not. Note that this rule is not always valid. Refer to
Metrel’s handbook Insulation Testing Techniques for more information.

General applicable values:


PI value Tested material status
1 – 1.5 Not acceptable (older types)
2 – 4 (typically 3) Considered as good insulation (older types)
> 4 (very high insulation resistance) Modern type of (good) insulation systems

Example for minimum acceptable values for motor insulation (IEEE 43):
Class A =1.5, Class B = 2.0, Class F =2.0, Class H =2.0.

5.2.3.2 Dielectric Discharge – DD


An additional effect of polarization is the recovered charge (from Cpi) after the regular
discharging of a completed test. This can also be a supplementary measurement for
evaluation of the quality of insulating material. This effect is generally found in insulating
systems with large capacitance Ciso.
The polarization effect (described in “Polarization Index”) causes a capacitance to form
(Cpi). Ideally this charge would dissipate immediately a voltage was removed from the
material. In practice, this is not the case.
In conjunction with the polarization index (PI), Dialectic Discharge (DD) is another way to
check the quality and suitability of a insulation material. A material that discharges
quickly would provide a low value while a material that takes a long time to discharge will
provide a higher value (described in the table below, for more information see section
5.6).
DD value Tested material status
>4 Bad
2–4 Critical
<2 Good

33
MI 3210 TeraOhm XA 10kV Measurements

5.2.4 Withstanding Voltage Test


Some standards allow the use of a DC voltage as an alternative to AC withstanding
voltage testing. For this purpose the test voltage has to be present across the insulation
under test for a specific time. The insulation material only passes if there is no
breakdown or flash over. Standards recommend that the test starts with a low voltage
and reaches the final test voltage with a slope that keeps the charging current under the
limit of the current threshold. The test duration normally takes 1 min.
Withstanding voltage test or dielectric test is usually applied for:
 Type (acceptance) tests when a new product is being prepared for manufacture,
 Routine (production) tests for the verification of safety on each product,
 Maintenance and after service tests for any equipment where the insulation
system can be exposed to degradation.
Some examples for DC withstanding test voltage values:

Standard (only sample values) Voltage


EN/IEC 61010-1 CAT II 300 V (mains circuits) basic insulation 2100 V
EN/IEC 61010-1 CAT II 300 V (mains circuits) reinforced insulation 4200 V
IEC 60439-1 (clearance between live parts…), withstanding 4700 V
impulse voltage 4 kV, 500 m
IEC 60598-1 2120 V

Humidity and insulation resistance measurements

When testing outside the reference ambient conditions, the quality of the insulation
resistance measurements can be affected by humidity. Humidity adds leakage paths
onto the surface of the complete measuring system, (i.e. the insulator under test, the test
leads, the measuring instrument etc). The influence of humidity reduces accuracy
especially when testing very high resistances (i.e. Tera ohms). The worst conditions
arise in environments containing high condensation, which can also reduce safety. In the
case of high humidity, it is recommended to ventilate the test areas before and during
the measurements. In the case of condensed humidity the measuring system must dry
and it can take several hours or even few days to recover.

34
MI 3210 TeraOhm XA 10kV Measurements

5.3 Guard terminal


The purpose of the GUARD terminal is to lead away potential leakage currents (e.g.
surface currents), which are not a result of the measured insulation material itself but are
a result of the surface contamination and moisture. This current interferes with the
measurement i.e. the Insulation Resistance result is influenced by this current. The
GUARD terminal is internally connected to the same potential as the negative test
terminal (black one). The GUARDs test clip should be connected to the test object so as
to collect most of the unwanted leakage current, see the Figure below.

IL

IM IM

IL
IA IA IL
+OUT -OUT +OUT -OUT GUARD

Ut A Ut A

Figure 5.2: Connection of GUARD terminal to measured object

where:
Ut ................................... Test voltage
IL .................................... Leakage current (resulted by surface dirt and moisture)
IM .................................... Material current (resulted by material conditions)
IA .................................... A-meter current

Result without using GUARD terminal:

R Ut  U t
Incorrect result.
ins
I A
( I M  I L)

Result using GUARD terminal:

R ins
 U t  U t Correct result.
I A I M

It is recommended to use the GUARD connection when high insulation resistance


(>10 G) are measured.

Notes:
 The guard terminal is protected by internal impedance (400 k).
 The instrument has two guard terminals to allow easy connection of shielded
measuring leads.

35
MI 3210 TeraOhm XA 10kV Measurements

5.4 Averaging options


Filters and additional averaging are built in to reduce the influence of noise on
measurement results. This option enables more stable results especially when dealing
with high Insulation Resistances.
In Insulation Measurement the status of the averaging option is shown in measurement
control window of the LCD screen. The table below contains a definition of the individual
filter options:

Averaging Settling
Meaning
options time
--- 0s Averaging is disabled
5 results 5s Moving average of 5 results
10 results 10 s Moving average of 10 results
30 results 30 s Moving average of 30 results
60 results 60 s Moving average of 60 results

5.4.1 The purpose of averaging


In simple terms the averaging smooth’s the measured.
There are various sources of disturbance:
 AC currents at the mains frequency and its harmonics, switching transients etc,
cause the results to become unstable. These currents are mostly cross talk
through insulation capacitances close to live systems,
 Other currents induced or coupled in the electromagnetic environment of the
insulation under test.
 Ripple current from internal high voltage source.
dU (t )
i (t )  C 
dt
 Charging effects of high capacitive loads and / or long cables.

Voltage changes are relatively narrow on high resistance insulation, so the most
important point is to filter the measured current.

Notes:
 Any of the selected averaging options increases the settling time.
 It is necessary to pay close attention to the selection of time intervals when using
the averaging.
 The recommended minimum measuring times when using averaging are the
settling times of the selected averaging option.

36
MI 3210 TeraOhm XA 10kV Measurements

5.4.2 Example of averaging

Capacitive test object 200 nF


Insulation Resistance Measurement
Test parameters:
Un = 5.00 kV
Timer1 = 5:00 min

Figure 5.3: Insulation Meas. (AVG _ _ _ ) Figure 5.4: Insulation Meas. (AVG 5s )

Figure 5.5: Insulation Meas. (AVG 30s ) Figure 5.6: Insulation Meas. (AVG 60s )

37
MI 3210 TeraOhm XA 10kV Measurements

5.5 Measurement menu


In the Measurement menu 5 different measurements and tests can be selected.

Figure 5.7: Measurement menu

Keys:
  Selects measurement or test.
SELECT Enters selected measurement function window.
ESC Returns to the Main menu.

38
MI 3210 TeraOhm XA 10kV Measurements

5.6 Insulation Resistance Measurement


Test can be started from the Insulation Resistance Measurement window. Before
carrying out a test the parameters output voltage, timer, high limit and additional
averaging can be edited.

Figure 5.8: Insulation Resistance Measurement menu

Test parameters for Insulation resistance Measurement


Un Set test voltage – step 50 V (50 V – 1 kV) and 100 V (1 kV – 10 kV).
Timer1 Duration of measurement (mm:ss) – step 1 s (max time 99 min).
Hi Lim Limit value selection (OFF, 0.50 MΩ – 1.0 TΩ).
AVG Additional averaging of the result value (OFF, 5, 10, 30, 60).

Keys:
  Selects the field to be changed.
  Modifies selected field.
SELECT Toggles between graph view and results view. (Graph must be
enabled in the Settings menu).
MEM Enters Save custom test menu.
Saves results (if present).
START/STOP Starts or Stops the Insulation Resistance Measurement.
ESC Exits to Measurements menu.

Keys in graph screen – measurement completed:


  Scroll the cursor along displayed graph data.
  Cursors on / off

39
MI 3210 TeraOhm XA 10kV Measurements

Insulation Measurement procedure:


 Select the Insulation Resistance Measurement function.
 Set the test parameters (voltage, timer, high limit, averaging).
 Connect the test leads to the instrument and to the test object.
 Press the START/STOP key to start the measurement.
 Press SELECT key to toggle between graph view and result view (optional).
 Wait until the test result has stabilized then press the START/STOP key again
to stop the measurement or wait for the set timer to run out.
 Wait for the object under test to discharge.
 Store the result by pressing MEM key (optional).

Figure 5.9: Example of Insulation Figure 5.10: Example of Insulation


resistance measurement result Resistance measurement graph view

Warnings:
 Refer to Warnings chapter for safety precautions!
 Do not touch the test object during the measurement or before it is fully
discharged! Risk of electric shock!

Notes:
 Consider displayed warnings when starting the measurement!
 A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a potentially dangerous test voltage.
 Value of capacitance is measured during the final discharge of the test object.

40
MI 3210 TeraOhm XA 10kV Measurements

5.6.1 Set Limit


With the high limit the user is allowed to set the limit resistance value. Measured
resistance is compared against the limit. Result is validated only if it is within the given
limit. Limit indication is shown by the bar Bar – graph marker (see Figure 5.10).

Figure 5.11:Limits marker

Message window:

Measurement result is within defined limit.

Measurement result is out of defined limit.

Note:
 Pass / Fail indication is only displayed if limit is set and if no Break down, Over
voltage or Noise is detected during measurement.

41
MI 3210 TeraOhm XA 10kV Measurements

5.7 Diagnostic test


Diagnostic test is a long duration test for evaluating the quality of the insulation material
under test. The results of this test enable the decision to be made on the preventive
replacement of the insulation material. Test can be started from the Diagnostic Test
window. Before carrying out a test the parameters can be edited.

Figure 5.12: Diagnostic test menu

Test parameters for Diagnostic test


Un Set test voltage – step 50 V (50 V – 1 kV) and 100 V (1 kV – 10 kV).
Timer1 Delay for starting DAR measurement (mm:ss) – step 1 s (max time 99 min).
Timer2 Delay for starting PI measurement (mm:ss) – step 1 s (max time 99 min).
Timer3 Duration of measurement (mm:ss) – step 1 s (max time 99 min).
DD Enable or disable dielectric discharge testing.
AVG Additional averaging of the result value (OFF, 5, 10, 30, 60).

Keys:
  Selects the field to be changed.
  Modifies selected field.
SELECT Toggles between graph view and results view. (Graph must be
enabled in the Settings menu).
MEM Enters Save custom test menu.
Saves results (if present).
START/STOP Starts or Stops the Diagnostic Test.
ESC Exits to Measurements menu.

Keys in graph screen – measurement completed:


  Scroll the cursor along displayed graph data.
  Cursors on / off

42
MI 3210 TeraOhm XA 10kV Measurements

Diagnostic Test procedure:


 Select the Diagnostic test function.
 Set the test parameters (voltage, timer1 …).
 Connect the test leads to the instrument and to the test object.
 Press the START/STOP key to start the measurement.
 Press SELECT key to toggle between graph view and result view (optional).
 Wait for the set timers to run out.
 Wait for the object under test to discharge.
 Store the result by pressing MEM key (optional).

Figure 5.13: Example of Diagnostic test Figure 5.14: Example of Diagnostic test
result graph view

Timer1, Timer2 and Timer3 are timers with the same starting point. The value of each
presents the duration from the start of the measurement. The maximum time is limited to
99 min. The following Figure shows the timer relationships.

30 s 1min 10min
DAR PI DD

Time1 ≤ Time2
Time2 ≤ Time3

Figure 5.15 Timer relations

43
MI 3210 TeraOhm XA 10kV Measurements

5.7.1 Dielectric Absorption Ratio (DAR)


DAR is ratio of Insulation Resistance values measured after 30 s and after 1 minute. The
DC test voltage is present during the whole period of the test (also an Insulation
Resistance measurement is continually running). At the end, the DAR ratio is displayed:

DAR  R Timer 2 _(1 min) 


iso

R Timer1 _(30 s) 
iso
Some applicable values for DAR (Timer1 = 30 s and Timer2 = 1 min):
DAR value Tested material status
<1 Bad insulation
1 ≤ DAR ≤1.25 Acceptable insulation
> 1.4 Very good insulation

Note:
 When determining Riso (30 s) pay close attention to the capacitance of the test
objects. It has to be charged-up in the first time section (30 s). Approximate
maximum capacitance using:

s  3

C F   t 10
max
U V
Where:
t ...................................... period of first time unit (e.g. 30 s).
U .................................... Test voltage.

5.7.2 Polarization Index (PI)


PI is the ratio of Insulation Resistance values measured after 1 minute and after 10
minutes. The DC test voltage is present during the whole period of the measurement (an
Insulation Resistance measurement is also running). On completion of the test the PI
ratio is displayed:

Timer 3 _(10 min) 


PI  Riso
Riso Timer 2 _(1 min) 
Some applicable values for PI (Timer2 = 1 min and Timer3 = 10 min):
PI value Tested material status
1 – 1.5 Not acceptable (older types)
2–4 Considered as good insulation (older types)
4 Modern type of good insulation systems

44
MI 3210 TeraOhm XA 10kV Measurements

Note:
 When determining Riso (1 min) pay close attention to the capacitance of the test
objects. It has to be charged-up in the first time section (1 min). Approximate
maximum capacitance using:

t s 10 3

C F  
max
U V 
where:
t ...................................... period of first time unit (e.g. 1 min).
U .................................... Test voltage.

Analyzing the change in the measured insulation resistance over time and calculating
the DAR and PI are very useful maintenance tests of an insulating material.

5.7.3 Dielectric Discharge Testing (DD)


DD is the diagnostic insulation test carried out after the completion of the Insulation
Resistance measurement. Typically the insulation material is left connected to the test
voltage for 1  30 min and then discharged before the DD test is carried out. After 1
minute a discharge current is measured to detect the charge re-absorption of the
insulation material. A high re-absorption current indicates contaminated insulation
(mainly based on moisture):
Idis1 min nA
DD 
U V  C F 
where:
Idis 1 min ........................ Discharging current measured 1 min after regular discharge.
C .................................... Capacitance of test object.
U .................................... Test voltage.

A high re-absorption current shows that the insulation has been contaminated, usually
by moisture. Typical values of dielectric discharge are shown in table.

DD value Tested material status


>4 Bad
2–4 Critical
<2 God

45
MI 3210 TeraOhm XA 10kV Measurements

Itot
Ba d
tion
In sula

Figure 5.16: The current/time diagram of a good and bad insulation tested with
dielectric discharge method

The dielectric discharge test is very useful for testing a multi-layer insulation. This test
can identify excess discharge currents that occur when one layer of a multi-layer
insulation is damaged or contaminated. This condition will not be detected by both the
spot test and the polarization index test. Discharge current will be higher for known
voltage and capacitance if an internal layer is damaged. The time constant of this
individual layer will differ from other layers, causing a higher current than that of a sound
insulation.

Warnings:
 Refer to Warnings chapter for safety precautions!
 Do not touch the test object during the measurement or before it is fully
discharged! Risk of electric shock!

Notes:
 Consider displayed warnings when starting the measurement!
 A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a potentially dangerous test voltage.
 Value of capacitance is measured during the final discharge of the test object.
 If enabled, the instrument measures Dielectric Discharge (DD) when the
capacitance is within the range of 20 nF to 50 F.
 The time duration of Graph R(t) is equal to the value of Timer3.
 The Timer value could be very long (up to 99 minutes), so the Special automatic
decimation algorithm is used to write the Graph R(t) to the LCD.
 If additional averaging (AVG) of the result value is turned on PI and DAR will not
be calculated (---).

46
MI 3210 TeraOhm XA 10kV Measurements

5.8 Step Voltage testing


In this test, the insulation is measured in five equal time periods with test voltages from
one fifth of the final test voltage up to full scale (Figure 5.17). This function illustrates the
relationship of a materials Insulation resistance and its applied voltage.

Figure 5.17: Step Voltage Test menu

Test parameters for Step Voltage Test


Un Set test voltage – step 50 V (50 V – 1 kV) and 100 V (1 kV – 10 kV).
Timer1 Duration of measurement (mm:ss) – step 1 s (max time 99 min).
AVG Additional averaging of the result value (OFF, 5, 10, 30, 60).

Keys:
  Selects the field to be changed.
  Modifies selected field.
SELECT Toggles between graph view and results view. (Graph must be
enabled in the Settings menu).
MEM Enters Save custom test menu.
Saves results (if present).
START/STOP Start or Stop the Step Voltage test.
ESC Exits to Measurements menu.

Keys in graph screen – measurement completed:


  Scroll the cursor along displayed graph data.
  Cursors on / off.

Step Voltage test procedure:


 Select the Step Voltage Test function.
 Set the test parameters (voltage, timer …).
 Connect the test leads to the instrument and to the test object.
 Press the START/STOP key to start the measurement.
 Press SELECT key to toggle between graph view and result view (optional).
 Wait for the set timers to run out.
 Wait for the object under test to discharge.
 Store the result by pressing MEM key (optional).

47
MI 3210 TeraOhm XA 10kV Measurements

Figure 5.18: Example of Step Voltage test Figure 5.19: Example of Step Voltage
result test graph view

Un

0.8Un

0.6Un

0.4Un

0.2Un

0.2T 0.4Ts 0.6Ts 0.8Ts T t

Figure 5.20 Step-up voltage test

Warnings:
 Refer to Warnings chapter for safety precautions!
 Do not touch the test object during the measurement or before it is fully
discharged! Risk of electric shock!

Notes:
 Consider displayed warnings when starting the measurement!
 A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a potentially dangerous test voltage.
 Value of capacitance is measured during the final discharge of the test object.
 Timer information shows the complete measurement period after the completion
of the measurement

48
MI 3210 TeraOhm XA 10kV Measurements

5.9 Withstanding Voltage Test


This function offers Withstanding Voltage Test of insulation material. It covers two types
of tests:
 Breakdown voltage testing of high voltage device, e.g. transient suppressors.
 DC withstanding voltage test for insulation coordination purposes.
Both functions require breakdown current detection. In the function, the test voltage is
increased from the starting voltage to the stop voltage over a predefined time (set by the
parameters). The Stop voltage is then maintained for a predefined test time.

Figure 5.21: Withstanding Voltage Test menu

Test parameters for Withstanding Voltage Test


Itrigg Set trigger leakage current step – 100 A (max 1.00 mA).
Ustart Stop test voltage – step 50 V (50 V – 1 kV) and 100 V (1 kV – 10 kV).
Uend Start test voltage – step 50 V (50 V – 1 kV) and 100 V (1 kV – 10 kV).
Tramp Duration of the test ramp voltage (mm:ss) – step 1 s (max time 99 min).
Tstart Duration of the start test voltage (mm:ss) – step 1 s (max time 99 min).
Tend Duration of constant test voltage after reaching stop value (mm:ss) – step 1
s (max time 99 min).

Keys:
  Selects the field to be changed.
  Modifies selected field.
MEM Enters Save custom test menu.
Saves results (if present).
START/STOP Start or Stop the Withstanding Voltage test.
ESC Exits to Measurements menu.

Withstanding Voltage test procedure:


 Select the Withstanding Voltage test function.
 Set the test parameters (voltage, timer …).
 Connect the test leads to the instrument and to the test object.
 Press the START/STOP key to start the measurement.
 Wait until the set timer runs out or until breakdown occurs, (the result will be
displayed).
 Wait for the object under test to discharge.
 Store the result by pressing MEM key (optional).

49
MI 3210 TeraOhm XA 10kV Measurements

Figure 5.22: Example of Withstanding Voltage test result

Figure 5.23 Test voltage presentation without breakdown

USTART............................ Starting test voltage.


UEND ............................... End test voltage.
TRAMP ............................. Duration of test ramp.
TSTART ............................ Duration of starting test voltage.
TEND ............................... Duration of test voltage after reaching UEND value.

Notes:
 Breakdown is detected when the measured current reaches or exceeds the set
current level Itrigg.
 A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a potentially dangerous test voltage.

50
MI 3210 TeraOhm XA 10kV Measurements

5.10 True RMS Voltmeter


It is a simple function that continuously measures the voltage and frequency across
+Rx and -Rx connectors. Measured voltage and frequency in function True RMS
Voltmeter can be stored.

Test circuit for voltage measurement

Figure 5.24: True RMS Voltmeter display

Measurement procedure:
 Select the True RMS Voltmeter function.
 Insert the test leads in to the instrument.
 Connect test leads with probes or crocodile clips to the measuring points.
 Store the result by pressing MEM key (optional).

Warnings:
 Refer to Warnings chapter for safety precautions!
 Do not touch the test object during the measurement or before it is fully
discharged! Risk of electric shock!

51
MI 3210 TeraOhm XA 10kV Maintenance

6 Communication
The instrument can communicate with the HVLink PRO PC software. The following
action is supported:
 Saved results can be downloaded and stored to a PC.
A special communication program on the PC automatically identifies the instrument and
enables data transfer between the instrument and the PC.
There are three communication interfaces available on the instrument:
USB, RS 232 or Bluetooth.

How to transfer stored data:


 RS 232 communication: connect a PC COM port to the instrument RS 232
terminal using the RS 232 serial communication cable.
 USB communication: connect a PC USB port to the instrument USB connector
using the USB interface cable.
 Switch on the PC and the instrument.
 Set the desired communication port RS 232, USB or Bluetooth.
 Run the HVLink PRO PC software.
 The instrument is prepared to download data to the PC.

Note:
 USB drivers should be installed on PC before using the USB interface. Refer to
USB installation instructions available on installation CD.
 Bluetooth communication device pairing code NNNN.

52
MI 3210 TeraOhm XA 10kV Maintenance

7 Maintenance
Unauthorized persons are not allowed to open the TeraOhm XA 10kV instrument. There
are no user replaceable components inside the instrument, except the battery.

7.1 Batteries insertion and replacement


Battery is stored in the battery compartment of the instrument casing under the battery
cover (see Figure 7.1). In case of defective battery please note the following:
Step 1
Turn off the power, disconnect any measurement accessories and mains supply
cable connected to the instrument before opening the battery cover to avoid
electric shock.

Step 2
Remove the battery cover (see Figure 7.1) !

Figure 7.1: Battey cover screws

53
MI 3210 TeraOhm XA 10kV Maintenance

Step 3
Replace the battery of the same type.

Step 4
Insert the battery correctly and check polarity (see Figure 7.2) !

Figure 7.2: Correctly inserted battery

Step 5
Fix the battery cover back in place.

Ensure batteries are used and disposed of in accordance with Manufacturers guidelines
and in accordance with Local and National Authority guidelines.

Note:
 Operator does not need to disconnect the instrument from mains supply after the
full recharging period. The instrument can be connected permanently.

Warnings:
 Disconnect all measuring accessory, mains supply and switch off the
instrument before opening battery cover!
 Use only LC-R123R4PG rechargeable battery!

54
MI 3210 TeraOhm XA 10kV Maintenance

7.2 Cleaning
No special maintenance is required for the housing. To clean the surface of the
instrument use a soft cloth slightly moistened with soapy water or alcohol. Then leave
the instrument to dry totally before use.

Warnings:
 Do not use liquids based on petrol or hydrocarbons!
 Do not spill cleaning liquid over the instrument!

7.3 Periodic calibration


It is essential that the test instrument is regularly calibrated in order that the technical
specification listed in this manual is guaranteed. We recommend an annual calibration.
Only an authorized technical person can do the calibration. Please contact your dealer
for further information.

7.4 Service
For repairs under warranty, or at any other time, please contact your distributor.

55
MI 3210 TeraOhm XA 10kV Technical specifications

8 Technical specifications
Nominal test voltage range ........................50 V – 10 kV
Voltage step ..............................................50 V (50 V – 1 kV) and 100 V (1 kV – 10 kV)
Voltage output accuracy .............................-0%, +10%  10 V

Current capability of test generator ................ > 1 mA


Short circuit / charge current .......................... 5 mA  10 %
Charging rate for capacitive load.................... < 3 s / µF at 10 kV

Automatic discharge....................................... yes


Discharging rate for capacitive load ............... < 0,25 s / µF at 10 kV
Discharging resistance ................................... 41 k  10 %

Bar graph range ............................................. 0  1 T (logarithmic scale)


Guard resistance ............................................ 400 k  5 %
Input AC current noise rejection ..................... 1 mA @ 600 V (4 mA max)
Resistance
Measuring range () Resolution () Accuracy
0.01  9.99 M 10 k
10.0  99.9 M 100 k
100  999 M 1M
(5 % of reading + 3 digits)
1.00  9.99 G 10 M
10.0  99.9 G 100 M
100  999 G 1G
1.0  9.9 T 100 G
(15 % of reading + 1 digits)
10  20 T 1T
Table 8.1: Resistance measurement ranges and accuracy ( 10 kV)

Notes:
 Full-scale value of resistance (RFS) depends on nominal test voltage (UN) and is
defined according to the following equation:

R FS  2 *10  / V  U N V 
9

 Accuracy versus full-scale resistance is defined in table below:

Measuring range () Accuracy


RFS
R (5 % of reading + 3 digits)
20
RFS
 R  RFS (15 % of reading + 1 digits)
20

56
MI 3210 TeraOhm XA 10kV Technical specifications

Current
Measuring range (A) Resolution (A) Accuracy
1.00  5.00 m 10 
100 999  1
10.0  99.9  100 n
(5 % of reading + 3 digits)
1.00  9.99  10 n
100  999 n 1n
10.0  99.9 n 100 p
0.00  9.99 n 10 p (10 % of reading + 0.15 nA)
Table 8.2: Current measurement ranges and accuracy

Voltage
Measuring range (V) Resolution (V) Accuracy
0  999 1
1.00  9.99 k 10 (5 % of reading + 3 digits)
10.0  14.0 k 100

Table 8.3: Voltage measurement ranges and accuracy

Capacitance
Measuring range (F) Resolution (F) Accuracy
20  999 n 1n
1.00  9.99  10 n (5 % of reading + 3 digits)
10.0  50.0  100 n
Table 8.4: Capacitance measurement ranges and accuracy

Nominal voltage range ................................... 500 V  10 kV

Dielectric absorption ratio DAR


Display range DAR Resolution Accuracy
0.01  9.99 0.01
(5 % of reading + 3 digits)
10.0  100.0 0.1

Table 8.5: DAR display ranges and accuracy

Polarization index PI
Display range PI Resolution Accuracy
0.01  9.99 0.01
(5 % of reading + 2 digits)
10.0  100.0 0.1

Table 8.6: PI display ranges and accuracy

57
MI 3210 TeraOhm XA 10kV Technical specifications

Dielectric discharge test DD


Display range DD Resolution Accuracy
0.01  9.99 0.01
(5 % of reading + 2 digits)
10.0  100.0 0.1

Table 8.7: DD display ranges and accuracy

Notes:
 All data regarding accuracy is given for nominal (reference) environment
condition measurements.
 The error in operating conditions could be at most the error for reference
conditions (specified in the manual for each function) ±5% of measured value + 3
digits, unless otherwise specified in the manual for particular function.
 Capacitance range for DD test: 20 nF to 50 F.

12,0

10,0

10 kV
8,0
5 kV
2,5 kV
U [kV]

6,0

4,0

2,0

0,0
0 1 10 100
Rx [MΩ]
Figure 8.8: Generator capability vs resistance

8.1 True RMS voltmeter


Voltage
Measuring range (V) Resolution (V) Accuracy
5.0  99.9 0.1
(2 % of reading + 2 digits)
100  550 1
Result type ..................................................... True r.m.s. (trms)
Nominal frequency range ............................... DC, 45 Hz  65 Hz
Input resistance .............................................. 40 k  10 %
Frequency
Measuring range (Hz) Resolution (Hz) Accuracy
10  500 0.1 (0.2 % of reading + 1 digit)

Nominal voltage range ................................... 5 V  550 V

58
MI 3210 TeraOhm XA 10kV Technical specifications

8.2 General data


Battery power supply......................... 12 V DC (3.4 Ah Lead – Acid)
Battery charging time ........................ typical 4 h (deep discharge)
Battery operation time:
Idle state ........................................... > 24 h
Measurements .................................. > 3 h continues testing 100 MΩ load @ 10 kV
.......................................................... > 5 h continues testing 100 MΩ load @ 5 kV
Auto - off timer................................... 15 min (idle state)
Mains power supply .......................... 90-260 VAC, 45-65 Hz, 100 VA (300 V CAT II)
Protection classification ..................... reinforced insulation
Over-voltage category ....................... 600 V CAT IV
Pollution degree ................................ 2
Degree of protection.......................... IP 65 (case closed)
Dimensions (w  h  d) ..................... 36 x 16 x 33 cm
Weight .............................................. 6.5 kg, (with battery and accessories)
Sound / Visual warning .................... yes
Display .............................................. 320 x 240 dots matrix display with backlight
Reference conditions:
Reference temperature range ........... 25 C ± 5 C
Reference humidity range ................. 40 %RH  60 %RH
Operation conditions:
Working temperature range .............. -10 °C  50 °C
Maximum relative humidity ................ 90 %RH (0 °C  40 °C), non-condensing
Working nominal altitude ................... up to 2000 m
Storage conditions:
Temperature range ........................... -10 °C  +70 °C
Maximum relative humidity ................ 90 %RH (-10 C  +40 C)
80 %RH (40 C  60 C)

RS 232 serial communication............ galvanic separated


Baud rate: ......................................... 9600 baud rate, 1 stop bit, no parity
Connector: ........................................ standard RS232 9-pin D female
USB slave communication ................ galvanic separated
Baud rate .......................................... 9600 baud rate
Connector ......................................... standard USB connector - type B
Bluetooth communication
Device pairing code: .......................... NNNN
Baud rate: ......................................... 115200 baud rate
Memory ............................................. 1000 storage locations (4 MB Flash memory)
Real time clock error .........................  50 ppm

59
MI 3210 TeraOhm XA 10kV Appendix A – Remote Control

Appendix A – Remote Control


Remote Control function is intended for distance control of the instrument MI 3210
TeraOhmXA 10kV via RS232, USB or Bluetooth (optional) communication.

Keypad Driving
All functionalities of the keypad can be driven via Remote Control communication.
Table A.1 shows the appropriate syntax to control the instrument keys. When the
command is received by the instrument it is send back as acknowledge.
Keypad Syntax Meaning
START / STOP ~KEY;START Start or stop measurement.
ON / OFF ~KEY;OFF Switches the instrument power off.
MEM ~KEY;MEM Store / recall / clear tests in memory
of the instrument.
SELECT ~KEY;SELECT To enter set-up mode for the
selected function or to select the
active parameter to be set.
 ~KEY;UP Select an option upward, downward.
 ~KEY;DOWN Select an option upward, downward.
 ~KEY;LEFT Decrease, increase the selected
parameter.
 ~KEY;RIGHT Decrease, increase the selected
parameter.
ESC ~KEY;ESC Exit the selected mode.
☼ ~KEY;LIGHT Turn the display backlight ON or
OFF.
Table A.1: Keypad Driving commands

Test Driving
All stored Custom tests can be run by Remote Control communication. Table A.2 shows
the appropriate syntax to start a Custom test. When the command is received by the
instrument it is send back as acknowledge.
Custom tests Syntax Meaning
CABEL 1kV ~RUN;CUSTOM;CABEL 1kV Run selected custom test.
500V TEST ~RUN;CUSTOM;500V TEST Run selected custom test.
~REZIM;PRINT_RESULTS;ON Enables printing of results (U,I,t).
~REZIM;PRINT_RESULTS;OFF Disables printing of results (U,I,t).
Table A.2: Examples of Custom tests driving commands

60

You might also like