Mechanical Measurements and Instrumentation Lab Manual
Mechanical Measurements and Instrumentation Lab Manual
LABORATORY MANUAL
IV Year B. Tech I- Semester
MECHANICAL ENGINEERING
AY: 2023-24
MECHANICAL MEASUREMENTS AND INSTRUMENTATION LAB
R20A0388
Prepared by:
Dr. D. DAMODARA REDDY
Associate Professor
CONTENTS
AIM:
Calibrate the given Pressure Cell with Pressure Gauge for its performance
APPARATUS REQUIRED:
Pressure cell
THEORY BEHIND:
Pressure is defined as force per unit area and is measured in Newton per square meter (Pascal)
or in terms of an equivalent head of some standard liquid.
Thus, a pressure gauge is connected to the hydraulic line and the gauge itself stands in
atmospheric pressure.
Fluid medium whose parameter has to be measured is allowed to deflect the diaphragm (sensing
member), which is a single block material and forms an integral part of the pressure transducer.
Usually, the Pressure Transducers are made up non-magnetic stainless steel and thus has the
advantage of avoids the yielding effects and leakage problems. The slight deflection of the diaphragms
due to the pressure provides an electrical output.
The material most commonly used for manufacture of diaphragms are steel, phosphor bronze,
nickel silver and beryllium copper. The deflection generally follows a linear variation with the
diaphragm thickness.
PANEL DETIALS:
PRESSURE INDICATOR: To indicate the Pressure in digital format with Zero knob facility.
PREPARATION OF EQUIPMENT:
1. Connect the pressure cell to the pressure indicator with given cable.
2. Connect the instrument to 1ph, 230V AC supply which is having proper earthing.
4. Close the release valve of pressure gauge tester and apply the 10kg dead weight on flange.
PROCEDURE:
1. Slowly rotate the screw road in clockwise direction with the help of handle until flange lift up
(so that pressure is developed up to applied load). Now observed the digital reading. If it is not
showing zero then make it zero by rotating ZERO knob. Now instrument is calibrated.
2. Apply the load up to 10Kgs one by one on the flange and give pressure by rotating the screw rod
such that the dial gauge reads 1 to 10 with respect to load applied.
3. Note down the readings of dial gauge and pressure indicator, simultaneously in every step.
TABULATIONS:
Techniques which can be adopted to minimize the errors in all aspects i.e., from startup to end.
LIMITATIONS
APPLICATIONS:
1.In compressors
2.In boilers
VIVA-VOCE QUESTIONS:
2. CALIBRATION OF LVDT
AIM:
Calibrate Linear Variable Differential Transformer (LVDT) for the performance using Micrometer.
APPARATUS REQUIRED:
LVDT
MICROMETER
THEORY BEHIND:
LVDT is an inductive transducer used to translate the linear motion into electrical signal LVDT
consists of a single primary winding ‘P’ and two secondary windings (S1 & S2) wounds on a cylindrical
armature. An AC source is connected to the primary winding. A movable soft iron core attached with
an arm placed inside the armature.
The primary winding produces an alternating magnetic field which induces alternating voltage
in the secondary windings. Single voltage is obtained by connecting the two secondary windings in
series. Thus, the output voltage of the transducer is the difference of the two voltages.
When the core is at null position, the flux linking with both the secondary windings is equal.
Since both the secondary winding have equal number of turns, M the induced emf is same in them. The
output voltage is the difference of the two emf say E1 & E2. When they are equal, the voltage is zero
at null position.
When the core is moved to the left side from null position more flux links with S1. The output
voltage is V=E2-E1, is greater, the V value is negative (–ve). Means the voltage is read in terms of mm
length on the display board indicates the negative value. When the core is moved to the right side of
the null position, more flux links with S2 induces voltages which is +ve. The display board indicates
the +ve value in mm of length.
The voltage output is linear and is depending on the position of the core. Hence LVDT can be
conveniently used to measure the thickness ranging from fraction of mm to a few cms. Normally LVDT
can give better result up to 5mm.
PANEL DETIALS:
PREPARATION OF EQUIPMENT:
1. Make the Micrometer to Read 10mm on the scale. (ZERO POSITION of LVDT)
2. Connect the instrument to 1ph, 230V AC supply which is having proper earthing.
LIMITATIONS
PROCEDURE:
1. Slowly rotate the screw head of the micrometer either clockwise or anticlockwise to measure
1mm on it.
Note: Clockwise will give readings in –ve direction and Anticlockwise will give in +ve direction.
TABULATIONS:
RESULT: -----------------------
APPLICATIONS:
1.In pressure cells
2.In force cells
3.In accelerometers
VIVA-VOCE QUESTIONS:
4.What are the major considerations which govern the selection of an instrument
transducer?
5.Define displacement?
6.What is the use of wire wound potentiometer for the measurement of linear and rotary
Transducers?
10
11
12
13
Weights
THEORY BEHIND:
A body subjected to external forces is in a condition of both stress and strain. Stress can be
directly measured but its effect i.e., change of shape of the body can be measured. If there is a
relationship between stress and strain, stresses occurring in a body can be computed if sufficient
strain information is available. The constant connecting the stress and strain in elastic material
under the direct stresses is the modulus of elasticity,
i.e., E=σ / є
the principle of the electrical resistance strain gauge was discovered by Lord Kelvin, when he
observed that a stress applied to a metal wire, besides changing resistance strain gauges are
made into two basic forms, bonded wire and bonded foil. Wire gauges are sandwiched between
two sheets thin paper and foil gauges are sandwiched between two thin sheets of epoxy.
The resistance factor ‘R’ of a metal depends on its electrical resistively, , its area, a and the
length l, according to the equation R = l / a.
Thus, to obtain a high resistance gauge occupying a small area, the metal chosen has a high
resistively, a large number of grid loops and a very small cross-sectional area. The most common
material for strain gauge is a copper - -nickel alloy known as Advance.
The strain gauge is connected to the material in which it is required to measure the strain, with
a thin coat of adhesive. Most common adhesive used is Eastman, Deco Cement, etc. as the test
specimens extends or contracts under stress in the direction of windings, the length and cross-
14
GAUGE FACTOR:
The dimension less relationship between the change in gauge resistance and change in length is
called Gauge factor of the strain, which is expressed mathematically,
In this relationship R and I represent, respectively the initial resistance and initial length of the
strain gauge filament, while R & l represents the small change in resistance and length, which
occurs as the gauge is strained along with the surface to which it is bonded. This gauge factor
of a gauge is a measure of the amount of resistance change for a given strain. The higher the
gauge factor greater the electrical output for indication or recording purpose. The gauge factor
is supplied by the manufacturer and may range from 1.7 to 4.
The usual method of measuring the change of resistance in a gauge element is by means of
Wheatstone bridge as shown in figure. It consists of Galvanometer, 4 resistor & a battery.
Resistance R1 is the strain gauge is used for strain measurement, which is mounted on the
specimen. The three resisters R2, R3 and R4 are internal to the device.
Let us assume that the resistance has been adjusted so that the bridge is balanced.
The most common bridge arrangements are single arm, two arm and four arm mode.
15
This bridge arrangement consists of a single active gauge in position, say R 1 and three resistor
are internal to the device. Temperature compensation is possible only if a self temperature
compensating strain gauge is used.
In this mode, two resistors are internal to the device and the remaining two are strain gauges.
One arm of this bridge is commonly labeled as active arm and the other as compensating arm.
The bridge is temperature compensated.
In this bridge arrangement, four active gauges are placed in the bridge with one gauge in
each of the four arms. If the gauges are placed on a beam in bending as shown in fig of the
elastic constant by bending test experiment, the single from each of the four gauges will add.
This bridge arrangement is temperature compensated.
PANEL DETIALS:
PREPARATION OF EQUIPMENT:
1. Connect the instrument to 1ph, 230V AC supply which is having proper earthing.
16
Sl.
Bridge Mode Color No of Wires Connect To
No
ORANGE 2 A&B
2 Half
GREEN 2 B&C
WHITE 2 A&D
ORANGE 2 A&B
3 FULL
GREEN 2 B&C
BLACK 2 C&D
LIMITATIONS
PROCEDURE:
1. For the Bridge selected and connected wires, slowly add the Weights in steps of 1kg.
17
TABULATIONS:
Graph Plotting
Techniques which can be adopted to minimize the errors in all aspects i.e, from startup to end.
APPLICATIONS:
1.In the ropes
2.In in the beams
VIVA-VOCE QUESTIONS:
1.Define strain?
3.What are the various types of strain gauges for different applications?
4.What is the difference between bonded and unbounded type of resistance strain gage?
18
11.Which gauge can be detached from the test specimen and used again?
19
20
21
APPARATUS REQUIRED:
THEORY BEHIND:
Capacitive type transducers are used essentially for displacement or positioning measurements.
But they are more susceptible to environmental factors such as dust or moisture in the atmosphere than
inductive type transducers.
The meshing area between two stator and rotor plates of the capacitor goes on changing as the
shaft capacitor is rotated. The arrangement is used to demonstrate the measurement of angular
displacement.
The transducer is mounted on to the face of a protractor which indicates the angle of displacement
and the readout display the amount of displacement.
22
PANEL DETIALS:
PREPARATION OF EQUIPMENT:
1. Connect the instrument to 1ph, 230V AC supply which is having proper earthing
3. Set ZERO position on the Manual Indicator by pressing the INCREMENTAL Button.
LIMITATIONS
PROCEDURE:
1. After the Preparation of equipment, Now press the INCREMENTAL button so the stepper motor
sets accordingly with the selected mode of previously.
23
TABULATIONS:
Graph Plotting
Techniques which can be adopted to minimize the errors in all aspects i.e., from startup to end.
APPLICATIONS:
1.In the rotating shafts
VIVA-VOCE QUESTIONS:
24
25
26
AIM:
Calibrate PHOTO SPEED SENSOR for the performance using MAGNETIC Speed Sensor.
APPARATUS REQUIRED:
THEORY BEHIND:
The measurement of rotational velocity is more common. For velocity (speed) measurement the most
convenient calibrator scheme uses a combination of toothed wheel, a simple magnetic proximity pickup
a photo couple sensor and an electronic indicator to measure the speed. The angular rotation is provided
by some adjustable speed drive of adequate stability. The toothed wheel mounted with iron rods while
passing under magnetic and photo pickup produces an electric pulse. These pulses are fed to signal
conditioner unit and displays reading visually. The stability of the rotational drive is easily checked by
observing the variation of display reading.
PANEL DETIALS:
27
PREPARATION OF EQUIPMENT:
2. Connect the instrument to 1ph, 230V AC supply which is having proper earthing.
LIMITATIONS
PROCEDURE:
1. Once the equipment is prepared, slowly rotate the Speed controller to set the speed desired in
range.
Note: Clockwise to increase the speed and Anticlockwise to reduce the speed.
TABULATIONS:
28
RESULT: -------------------
APPLICATIONS:
1.To measure the electric motors shaft speed.
2.To measure the turbine shaft speed.
3.To measure the engines shaft speed.
VIVA-VOCE QUESTIONS:
29
30
31
AIM:
Calibrate the Vacuum Cell (McLeod Gauge) in comparison with Vacuum Dial Gauge.
APPARATUS REQUIRED :
CYLINDRICAL CELL
THEORY BEHIND:
In everyday usage, vacuum is a volume of space that is essentially empty of matter, such
that its gaseous pressure is much less than atmospheric pressure. The word comes from the Latin
for “empty”. A perfect vacuum would be one with no particles in it at all, which is impossible
to achieve in practice. Physicists often discuss ideal test results that would occur in a perfect
vacuum, which they simply call “vacuum” or “free space”, and use the term partial vacuum to
refer to real vacuum. The Latin term in vacuum is also used to describe an object as being in
what would otherwise be a vacuum.
Vacuum is useful in a variety of processes and devices. Its first widespread use was in the
incandescent light bulb to protect the filament from chemical degradation. The chemical
inertness produced by a vacuum is also useful for electron beam welding, cold welding, vacuum
packing and vacuum frying. Ultra-high vacuum is used in the study of atomically clean
substrates, as only a very good vacuum preserves atomic-scale clean surfaces for a reasonably
long time (on the order of minutes to days). High to ultra-high vacuum removes the obstruction
of air, allowing particle beams to deposit or remove materials without contamination. This is
the principle behind chemical vapor deposition, physical vapor deposition, and dry etching
which are essential to the fabrication of semi conductors and optical coatings, and to surface
science. The reduction of convection provides the thermal insulation of thermos bottles. Deep
vacuum lowers the boiling point of liquids and promotes low temperature out gassing which is
32
used in freeze drying, adhesive preparation, distillation, metallurgy, and process purging. The
electrical properties of vacuum make electron microscopes and vacuum tubes possible,
including cathode ray tubes. The elimination of air friction is useful for flywheel energy storage
and ultracentrifuges.
PANEL DETIALS:
PREPARATION OF EQUIPMENT:
2. Connect the instrument to 1ph, 230V AC supply which is having proper earthing.
LIMITATIONS
33
PROCEDURE:
5. Now slowly open the Valve on the cylinder and set to the required Vacuum.
TABULATIONS:
RESULT: --------------------------
APPLICATIONS:
1.To clear halls
34
VIVA-VOCE QUESTIONS:
35
36
37
AIM:
Calibrate TEMPERATURE SENSORS for the performance using STANDARD water bath.
APPARATUS REQUIRED:
THEORY BEHIND:
There are many categories like carbon resistors, film and wire wound types are the most widely used.
• Carbon resistors are widely available and are very inexpensive. They have very reproducible
results at low temperatures. They are the most reliable from at extremely low temperatures.
They generally do not suffer from significant hysteresis or strain gauge effects.
• Film thermometer have a layer of platinum on a substrate, the layer may be extremely thin,
perhaps one micrometer. Advantages of this type are relatively low cost and fast response. Such
devices have improved performance although the different expansion rates of the substrate and
platinum give “strain gauge” effects and stability problems.
• Wire – wound thermometers can have greater accuracy, especially for wide temperature ranges.
• Coil element has largely replaced wire wound elements in industry. This design has a wire coil
which can expand freely over temperature, held in place by some mechanical support which lets
the coil keep its shape.
38
THERMOCOUPLES
39
PANEL DETIALS:
PREPARATION OF EQUIPMENT:
1. Fill the water into the Water Bath to the required level.
3. Connect the instrument to 1ph, 230V AC supply which is having proper earthing.
LIMITATIONS
PROCEDURE:
1. Switch on the Console to activate the Panel and then switch on the water bath.
3. Select the Sensor under study {however other sensors will also be working condition, this is only
done to concentrate on particular sensor type}and Note the readings of the selected Sensor
indicator for every two degree rise of water bath temperature.
40
TABULATIONS:
RESULT: --------------------
APPLICATIONS:
1.To control of gas flow
2.In electric iron boxes
3.In domestic ovens
VIVA-VOCE QUESTIONS:
Below-400C?
41
42
43
44
8.ROTAMETER SETUP
i. AIM:
ii. PROCEDURE:
45
iii. OBSERVATIONS:
Time for
‘R’ cm Height
Rotameter
rise in measured
Sl. Reading, ‘Q’
water on Log
No Lpm
scale, cm
‘T’ sec
iv. CALCULATIONS:
Where,
2. Actual Discharge, QA
𝐴𝑥𝑅
𝑄𝐴 = 𝑚3 /𝑠
𝑡 𝑥 100
Where,
46
3. Co – efficient of discharge, CD
𝑄𝐴
𝐶𝑑 =
𝑄𝑇𝐻
Where,
QA = Actual Discharge.
A. For Rotameter:
47
B. Graphs
vi. RESULTS:
For Rotameter:
4. PRECAUTIONS
48
49
50