Co2 Ecotech - Manual
Co2 Ecotech - Manual
Analyser
Service Manual
Revision: B
www.ecotech.com
EC9820 CO2 ANALYZER SERVICE MANUAL
Table of Contents
98207601 Rev. B 1
EC9820 CO2 ANALYZER SERVICE MANUAL
4.3.2 Preamp......................................................................................................................................................46
4.3.3 Electric ......................................................................................................................................................46
4.3.4 Use of Diagnostic Modes ..........................................................................................................................46
4.4 TROUBLESHOOTING GUIDE ................................................................................................................................47
INDEX ........................................................................................................................................................................49
2 98207601 Rev B
EC9820 CO2 ANALYZER SERVICE MANUAL
Manual History
The scope of this new manual covers the following analyzers:
This manual should only be used in conjunction with the EC9820 Carbon
Monoxide Analyzer, Operation Manual PN: 98207600 Rev B, April 2007
Revision History
Rev Date Summary Affected Pages
A September New Release for new Microprocessor Board Based All
2005 on original manuals.
B April 2007 Updated specifications, language and links within All
pdf manual created.
98207601 Rev. B 3
EC9820 CO2 ANALYZER SERVICE MANUAL
Safety Requirements
To reduce risk of personal injury caused by electrical shock, follow all
safety notices and warnings in this documentation.
This equipment should always be used with a protective earth installed.
The EC9820 is compliant with the requirements of EN61010-1 A2:1995,
Safety Requirements for Equipment for Measurement, Control, and
Laboratory Use.
If the equipment is used for purposes not specified by the manufacturer,
the protection provided by this equipment may be impaired.
Replacement of any part should only be carried out by qualified personnel,
only using parts specified by the manufacturer. Always disconnect power
source before removing or replacing any components.
Surfaces marked with a “Caution, Hot Surface” (see internationally
recognized symbols on page 4) sticker may get hot and deliver burns.
Measure the temperature on the surface before making any contact with it.
Equipment Rating
100-120/220-240V~ ±10%
50/60 Hz
250 VA max
FUSE: 5/3.15A T 250V
All wiring must be in accordance with local norms and be carried out by
experienced personnel.
Service and Spare Parts
For world wide customer service & spare parts contact ECOTECH:
Web: www.ecotech.com.au
4 98207601 Rev B
CHAPTER 1 INSTALLATION
1.0 Installation
1.1 Initial Check
Verify that the serial number label on the documentation and the serial number(s)
on the analyzer match.
Check to make certain your instrument arrived undamaged. If you find damage,
report it as described in the preface, on the page titled Claims for Damaged
Shipments and Shipping Discrepancies in the Operation manual.
Analyzers are shipped ready to power up. Occasionally, however, rough handling
during shipment causes dislodged PC boards, disconnected cables, or incorrectly
positioned switches. Verify that your instrument is in operating condition by
performing the following procedure.
98207601 Rev. B 5
EC9820 CO2 ANALYZER SERVICE MANUAL
DC Power ON
Pump ON
ServiceIN
The Reset switch is not a toggle switch and is only activated when pressed. It
resets the microprocessor. The pump switch is not applicable to the B-line
instrumentation because the flow is generated through the use of an external
pump.
When in the OUT position, the Service switch sets the OUT OF SERVICE bit in the
50-pin I/O interface and in the status word from the serial port. The OUT position
has no other effect on the operation or validity of the data obtained from the
analyzer. When the Service switch is set from OUT to IN, the instrument returns
to the normal operating conditions.
6 98207601 Rev B
CHAPTER 1 INSTALLATION
98207601 Rev. B 7
EC9820 CO2 ANALYZER SERVICE MANUAL
8 98207601 Rev B
CHAPTER 2 THEORY OF OPERATION
The process further defined consists of radiation from the IR source encountering
a gas filter wheel alternating between CO2, N2, and a mask. The radiation passes
through a multiple pass measurement cell where the sample gas absorption
process occurs, then through a narrow-band-pass filter to limit absorption to
optical wavelengths of interest for CO2 absorption, after which it encounters the
IR detector where the amount of absorption is measured.
The N2 filter in the gas filter correlation wheel is transparent to IR and yields a
measure beam that can be absorbed by CO2 in the measurement cell. The CO2
filter in the wheel yields a beam that cannot be further attenuated by the CO2 in
the measurement cell, thus it is a reference beam. The mask creates a signal used
to determine the strength of the other two signals.
The CO2 concentration in the measurement cell absorbs the measure beam and
does not absorb the reference beam, modulating the IR radiation, or detector input
signal between the gas filters. Other gases absorb the reference and measure
beams equally and thus do not cause modulation of the detector signal. Using this
method, the system responds specifically to CO2.
The EC9820 analyzer uses the advanced digital Kalman filter. This filter provides
the best possible compromise between response time and noise reduction for the
type of signal and noise present in ambient air analyzers and their application.
The system continuously analyzes the signal and uses the appropriate filtering
time.
98207601 Rev. B 9
EC9820 CO2 ANALYZER SERVICE MANUAL
2.1.1.3 Microprocessor
The microprocessor board contains a battery backed clock/calendar and an
onboard 16-bit microprocessor (MC68HC12) operating at 16 MHz. The
microprocessor board is the control center for input and output apparatus such as
the 2 inch by 4 inch liquid crystal display (LCD), keyboard switches, the serial
ports, and the 50-pin I/O connector on the rear panel. The 50-pin I/O connector
input accepts control lines from the rear panel and sends status and failure signals
to solid state relay drivers. Support circuitry for the liquid crystal display includes
a -20 V power supply and digitally adjusted potentiometers for contrast level.
All analog voltages from the sensor assembly are digitized by the analog-to-
digital (A/D) converter for microprocessor use. Digital-to-analog (D/A)
conversion of three channels is used to send 0 to 20 mA analog signals to the
50-pin I/O connector.
10 98207601 Rev B
CHAPTER 2 THEORY OF OPERATION
The microprocessor has electrically erasable ROMs which store the operating
program and internally logged data. Program upgrades can be easily made
through the serial port. The Service and Reset switches are located on the front of
the board and are accessible when the top is removed or when the front panel is
opened. The microprocessor also has provisions for USB and optional TCIP
connections through the rare panel.
98207601 Rev. B 11
EC9820 CO2 ANALYZER SERVICE MANUAL
2.1.2.1 Pneumatics
The pneumatic system continuously supplies sample air to the measurement cell
at a rate that allows the sample to be measured before exiting the analyzer. The
pneumatic system flow is illustrated in Figure 6 The pump causes sample air to
be drawn into the sample inlet and through the 5 micron particulate filter.
During the startup and auto-zero cycles, the sample port is disconnected and air is
pulled into the system from then into the measurement cell. This provides the
analyzer with a stable reference of zero air to automatically correct for changes in
the measurement path.
12 98207601 Rev B
CHAPTER 2 THEORY OF OPERATION
98207601 Rev. B 13
EC9820 CO2 ANALYZER SERVICE MANUAL
2.1.2.2 Optics
The optics of the EC9820 analyzer include an IR source, an optical detector, a gas
filter correlation wheel, measurement cell, narrow-band-pass filter, IR detector,
and preamp board as described in the following paragraphs. Figure 7 illustrates
the optical path.
Gas Filter Correlation Wheel. The gas filter correlation wheel includes
two gas filled sapphire chambers and mask areas. As the wheel rotates, the
IR beam passes through either 1) the reference chamber which is filled
with CO2, removing the CO2-sensitive wavelengths which are centered at
4.5 microns and passing the CO2-insensitive wavelengths; 2) the measure
chamber which is filled with nitrogen and passes the CO2-sensitive
wavelengths; or 3) a mask which blocks the beam. When the mask blocks
the transmission of IR, the detector generates the mask signal used as a
zero reference to determine the strengths of the other two signals. The
differences between signals are used to compute the concentration of CO2.
Figure 8 shows the gas filter correlation wheel. The wheel is mounted on a
shaft driven by a brushless DC motor.
Opto-detector. A reflective optical detector reacts to timing marks on the
periphery of the wheel and generates a signal indicating when the center of
each sector is the target of the IR source beam.
14 98207601 Rev B
CHAPTER 2 THEORY OF OPERATION
98207601 Rev. B 15
EC9820 CO2 ANALYZER SERVICE MANUAL
16 98207601 Rev B
CHAPTER 2 THEORY OF OPERATION
2.2.1.5 Background
Background allows the analyzer to sample zero air and measure the level of the
concentration voltage. This reading is taken as the zero signal level and this is
subtracted from any subsequent readings. This is the analyzer’s fine zero
measurement. The background is re-run nightly at midnight (unless background is
disabled), when manually selected through the CALIBRATION MENU, when a
temperature change of greater than 4°C is sensed, or when another automatic
startup routine is performed.
Note
To perform a differential background:
1. Ensure that zero air is not connected to zero port otherwise
a standard “absolute background” will be performed.
2. Connect the zero port to ambient air (outside of sampling
area) and run background readings.
3. Once this background has been performed then the analyzer
can be placed in sampling environment and proceed
sampling.
4. Ensure background air is NEVER taken from ambient air in
sampling environment; this will result in erroneous sample
readings.
98207601 Rev. B 17
EC9820 CO2 ANALYZER SERVICE MANUAL
18 98207601 Rev B
CHAPTER 2 THEORY OF OPERATION
2.2.3.1 Background
The auto-zero function (background) is performed nightly at midnight unless
BACKGROUND is DISABLED. This compensates for drift in the measurement
baseline of the analyzer. The nightly background routine is identical to the
background described in the startup routine.
98207601 Rev. B 19
EC9820 CO2 ANALYZER SERVICE MANUAL
3.0 Maintenance
3.1 Maintenance Schedule
The following outlines a periodic maintenance schedule for the EC9820 analyzer.
This schedule is based on experience under normal operating conditions, and may
need to be modified to suit specific operating conditions and regulations. It is
recommended that this schedule be followed in order to maintain reliable, long-
term operation of the analyzer.
1
Suggested intervals for normal operation and actual intervals will vary depending upon application. The user can
refer to this table as a guideline, but should develop a maintenance schedule to suit their specific requirements.
20 98207601 Rev B
CHAPTER 3 MAINTENANCE
Level 1: General maintenance supplies and expendables such as filters, O-rings, lamps, etc.
Level 2: Critical items that are known from experience to have a higher failure rate, such as
pumps, heaters, converters, valves, and circuit boards.
Level 3: Other miscellaneous items not included in Level 1 or 2. This level includes other spare
parts that are not expected to fail over a given time frame.
98207601 Rev. B 21
EC9820 CO2 ANALYZER SERVICE MANUAL
Components marked with shading are essential components which need to be kept on hand at all
times.
22 98207601 Rev B
CHAPTER 3 MAINTENANCE
Toolbox
Oscilloscope
Digital multimeter (DMM)
Computer or remote data terminal and connection cable for RS232
communication
Pressure transducer (absolute) and connection tubing, calibrated in torr
Flowmeter (2 slpm nominal)
Wire strippers
Soldering iron
Minifit extraction tool
Orifice removal tool
Assortment of 1/4” and 1/8” tubing and fittings
Test zero air source (CO2-free)
Test span gas source
Leak tester.
98207601 Rev. B 23
EC9820 CO2 ANALYZER SERVICE MANUAL
Several factors affect the filter replacement schedule. In the springtime, for
example, pollens and dust might accumulate in the filter. Man-made
environmental changes such as construction dust might indicate more frequent
change, or a climate where dry, dusty conditions are normal might dictate more
frequent filter replacement than climates with few natural pollutants.
Determining the schedule for changing the filter is best developed by monitoring
the filter at weekly intervals for the first few months, then adapting the schedule
to fit the specific site.
24 98207601 Rev B
CHAPTER 3 MAINTENANCE
1. Open the front panel to access the service switches and particulate filter.
Position the Pump switch in the secondary front panel to OFF.
3. Pull the filter plunger out of the body, carefully resting it in a secure place.
The O-ring and filter membrane are now exposed inside the filter body.
4. Remove, inspect, and wipe down the O-ring. Replace the O-ring only if
damaged.
6. Reinstall the O-ring over the new membrane, reinstall the plunger, and hand-
thread the cap back into place by turning it clockwise. Do not use tools.
98207601 Rev. B 25
EC9820 CO2 ANALYZER SERVICE MANUAL
Caution
If the detector signal is adjusted, the analyzer
will require recalibration.
Required equipment:
Oscilloscope
3. Verify that the two signals on the detector signal differ in amplitude by
approximately 15% (refer to Figure 11). If they differ by less than 15%, the
correlation wheel may require replacement.
26 98207601 Rev B
CHAPTER 3 MAINTENANCE
4. Disconnect the oscilloscope and reinstall the cover. If the signal was
readjusted, the analyzer should be reset and recalibrated after its startup
sequence is complete.
2. Disconnect the 1/4” tubing from the measurement cell inlet and outlet and all
electrical connections from the bench to the sensor tub. Remove the four
mounting screws from the bench and remove the bench from the sensor tub.
3. Remove the four screws securing the source/timing assembly to the optical
bench and remove the block from the bench.
4. Remove the correlation wheel by loosening the set screw in the side of the
wheel body (0.5 mm hex screw). Slide the correlation wheel off of the motor
shaft.
5. Remove the four screws securing the motor to the mounting plate and remove
the motor.
6. Remove the two screws that fasten the IR source into the source block.
Remove the IR source from the block.
98207601 Rev. B 27
EC9820 CO2 ANALYZER SERVICE MANUAL
7. Using a minifit extraction tool, remove the two wires from the source to the
electrical connection. Remove the IR source and install the new source,
rewiring the new source in place of the old one.
10. Reinstall the correlation wheel. When securing the correlation wheel to the
motor shaft, ensure the set screw secures to the flat face of the motor shaft.
11. Reassemble the source/timing assembly to the optical bench and reinstall the
bench in the sensor tub.
12. Perform the detector adjustment from earlier in this section. Be sure to
recalibrate the analyzer.
2. Disconnect the 1/4” tubing from the measurement cell inlet and outlet and all
electrical connections from the bench to the sensor tub. Remove the four
mounting screws from the bench and remove the bench from the sensor tub.
28 98207601 Rev B
CHAPTER 3 MAINTENANCE
3. Remove the four screws securing the detector cover to the detector assembly.
4. Remove the four screws securing the detector assembly to the bench.
5. Remove the detector assembly. Install the new detector in reverse order of the
above steps.
1. Enter the TEST MENU and select OUTPUT TEST MENU, then VALVE TEST MENU;
from these items pick VALVE SEQUENCING and set to OFF.
3. Press <Pg Up> to return to the OUTPUT TEST MENU and choose FLOW CONTROL
POTS.
4. Turn the Pump Switch on the secondary panel to OFF and verify FLOW: 0.00
SLPM. If it does not read 0.00 the Flow Control PCA requires recalibration.
Return the Pump Switch to ON.
6. Verify GAS FLOW in the FLOW CONTROL POTS screen reads 0.00. If the reading
is not 0.00, there is a leak somewhere between the valves and the flow block.
7. Set VALVE SEQUENCING to ON and then press <Exit>. Remove the plug from
the INLET port.
8. Verify that the flow returns to the original setting by observing the
INSTRUMENT STATUS menu.
The actual flow through the analyzer should be checked by turning on the pump
and connecting a flow meter to the measure inlet (ensure that the analyzer is in
SAMPLE MEASURE). Actual flow through the analyzer should be approximately
1 slpm. If flow is too low, perform the flow calibration in section 3.5 below.
98207601 Rev. B 29
EC9820 CO2 ANALYZER SERVICE MANUAL
To access the hidden menu, simultaneously press three keys on the front panel
from the primary screen as follows:
HIDDEN MENU
SERIES A
ANALYZER TYPE 9820
SUB TYPE STANDARD
FLOW BLOCK TYPE STANDARD
GAS NAME ___
PRESSURE CALIBRATION MENU
FLOW CALIBRATION MENU
FACTORY DIAGNOSTICS MENU
SERIES
Allows the user to choose the correct series of analyzer. The options are A, B &
S. The selection must reflect the Hardware to which it is being installed. i.e. for
an EC9820 analyzer, it must be set to A, and for an EC9820B analyzer set to B, in
order to operate correctly. The S option is reserved for special system software.
ANALYSER TYPE
Allows the user to choose the correct analyzer type. The options are 9810, 9811,
9812, 9820, 9830, 9841, 9842 & 9850. The selection must reflect the
Hardware to which it is being installed. i.e. for an EC9820 analyzer, it must be set
to 9820 in order to operate correctly. The analyzer type will default to ???? if the
device type has not been set or cannot be read from the preprocessor board.
SUB TYPE
Allows the user to set the correct Hardware options for this analyzer. The options
for the EC9820 are: STANDARD, TRACE & HIGH LVL.
GAS NAME
Allows the user to set a 3 character name for the primary gas name. The name
entered here will be used to replace the primary gas name in all screens. If the
first character is a space then CO will be used as a gas name.
30 98207601 Rev B
CHAPTER 3 MAINTENANCE
DESIRED FLOW
Desired flow rate that the standard flow block will maintain a constant flow at.
Note
If the flow rate is changed the instrument must
be reset in order for the change to take effect.
FLOW SPAN POINT
During the flow calibration of the standard flow block, this is where the externally
measured flow rate is entered.
FLOW TRANSDUCER DF
Selecting set, loads the factory default calibration curve into the earom on the
standard flow control board.
PRESSURE 1 HIGH
Pressure entered during ambient calibration.
PRESSURE 1 LOW
Pressure entered during low pressure calibration.
VALVE SEQUENCING
Turn the valve sequencing on or off. Same as in the Valve Test Menu. When
turned off, the appropriate valve sequence will set ready for pressure calibration.
CONTROL LOOP
When ENABLED, the microprocessor controls the pump speed (via the flow control
pots) to give the desired flow rate. When DISSABLED, the user can manually adjust
the pump speed using the flow control pots. For the standard flow block only.
98207601 Rev. B 31
EC9820 CO2 ANALYZER SERVICE MANUAL
AMBIENT PRESSURE
Current ambient pressure measured from the ISO flow controller.
GAS PRESSURE
Current gas pressure measured from the flow controller.
GAS FLOW
Current gas flow measured from the flow controller.
32 98207601 Rev B
CHAPTER 3 MAINTENANCE
From the HIDDEN MENU, set the FLOW BLOCK TYPE TO STANDARD, and press
Enter> (↵). Press Reset on the analyzer secondary panel.
VALVE SEQUENCING ON
CONTROL LOOP ENABLED
PUMP SPEED COARSE 28
2. Verify the pump is turned off by setting the pump switch to OFF on the
secondary panel.
Note
To convert from millibar to TORR, multiply the
pressure by 0.75.
4. Set the CONTROL LOOP TO DISABLED, and press Enter> (↵).This will disable
the flow control loop.
5. Connect the calibrated pressure transducer to the sample inlet on the rear
panel of the analyzer.
98207601 Rev. B 33
EC9820 CO2 ANALYZER SERVICE MANUAL
10. Disconnect the pressure transducer from the inlet and turn the pump OFF.
11. To verify that the pressure transducers on the standard flow control board are
calibrated, view the GAS PRESSURE reading in the PRESSURE CALIBRATION
MENU. it should be at ambient pressure +/- 2 TORR.
2. Verify the pump is turned OFF .and the CONTROL LOOP is DISABLED in the
FLOW CALIBRATION MENU.
3. Load the linearization table into the flow control board EAROM by setting
FLOW TRANSDUCER DF to SET and pressing Enter> (↵). The linearization
table contains the following parameters:
34 98207601 Rev B
CHAPTER 3 MAINTENANCE
4. Connect the calibrated flow meter to the sample inlet on the rear panel of the
analyzer. With the pump turned off, verify that there is no flow.
5. Connect a Digital Volt Meter (DVM) to the FLOW test point on the Flow
Control PCA. TP1 is the 0V connection, and pin 5 of the J2 connector is the
FLOW test point. The Digital Volt Meter should be set to the 2V DC range.
6. Adjust the FLOW CONTROL ZERO pot until the DVM reads slightly positive (0
to 0.5 VDC), then press <Enter>.
7. Set the FLOW ZERO POINT to SET and press Enter> (↵). The GAS FLOW
should now read 0.000 SLPM.
9. Turn the analyzer pump ON and verify the CONTROL LOOP is still
DISABLED.
10. From FLOW CALIBRATION MENU use the PUMP SPEED COARSE and PUMP
SPEED FINE pots to adjust the measured flow (external flowmeter) to as close
as possible to 1.00 SLPM.
Note
If the analyzer attempts to automatically change the
pot position during adjustment then repeat step 9.
11. Record the average measured flow from the flowmeter and enter it into the
FLOW SPAN POINT and press Enter> (↵).
12. Set the DESIRED FLOW to 1.00 SLPM and press Enter> (↵).
13. Press Reset on the analyzer and allow the flow to stabilize. Verify the
flowmeter reading is now 1.0 ±0.05 SLPM.
Note
After the instrument startup is complete, verify the
gas pressure is slightly below actual ambient
pressure (approximately 10 - 20 torr). The gas
pressure and gas flow readings may not bet updated
during the startup routine.
98207601 Rev. B 35
EC9820 CO2 ANALYZER SERVICE MANUAL
1. From the HIDDN MENU set the SERIES to A then press Enter> (↵).
3. Press Reset on the analyzer secondary panel. The display should now display
9820 CO ANALYZER.
4. You may need to erase memory after this procedure to avoid any problems.
Next, download the required software version for your 9800 analyzer from those
listed at the bottom of this page (e.g. V1.00.0002), by right clicking on the link,
and choosing 'Save Target As', and saving the .sx file on your computer. You will
need to select save as type 'All Files' in the download window.
2. Connect the 9800 analyzer to the computer using a standard serial cable (you
can connect to either the Multidrop port on the back of the analyzer, or the
service port on the front of the analyzer).
3. Select 'Serial Port' and the COM Port on the computer from those listed on
the Firmware Update screen.
36 98207601 Rev B
CHAPTER 3 MAINTENANCE
4. Click Next
5. Enter the full path and file name of the firmware (.sx) file you downloaded.
e.g. - C:\TEMP\V1.03.0001.SX assuming the file was saved to C:\TEMP.
6. Click Next
98207601 Rev. B 37
EC9820 CO2 ANALYZER SERVICE MANUAL
8. Ensure the analyzer is switched off using the switch under the front panel.
9. Click 'Start'.
11. The Firmware Updater window will show each step as the firmware is
uploaded. DO NOT turn the analyzer off until the 'Close' button is enabled as
shown below, and the Analyzer is operating as usual again.
38 98207601 Rev B
CHAPTER 3 MAINTENANCE
4.0 Troubleshooting
4.1 DC Power Supply Voltages
Before consulting the troubleshooting section, verify that the DC power supply
voltages are present and within the specifications given for each printed circuit
board listed in the following table. Circuit board illustrations indicating the test
points and other component locations immediately follow the Troubleshooting
Guide.
* The +12V GOOD test point is a 5 volt status output from the power supply to
indicate that the Mains supply is within the correct operating range.
98207601 Rev. B 39
EC9820 CO2 ANALYZER SERVICE MANUAL
PREPROCESSOR POTS
VALVE TEST MENU
EVENT LOG
INSTRUMENT STATUS
SYSTEM TEMPERATURES
SYSTEM FAULTS.
These menus provide information that may indicate a failure or an operational
problem. If instrument performance appears to have changed dramatically, the
component that is causing the problem can possibly be determined, thereby
speeding up the corrective process. It may assist the operator to periodically check
and record these parameters to establish an operational history of the analyzer. In
addition, information from this section may be requested by the Ecotech Service
Support personnel when assistance is required.
40 98207601 Rev B
CHAPTER 4 TROUBLESHOOTING
The VALVE TEST MENU (see Figure 23) displays the current status of each valve
in the instrument. This menu can be particularly useful in correcting flow
problems in the machine. The valves can be opened and closed from this menu,
thus allowing the operator to determine whether valves are operating correctly.
Valve sequencing must be ON in order for correct gas measurement to be
accomplished.
98207601 Rev. B 41
EC9820 CO2 ANALYZER SERVICE MANUAL
Should the EVENT LOG indicate an error, it will also provide information as to the
portion or component of the instrument which is at fault.
42 98207601 Rev B
CHAPTER 4 TROUBLESHOOTING
98207601 Rev. B 43
EC9820 CO2 ANALYZER SERVICE MANUAL
EXIT
44 98207601 Rev B
CHAPTER 4 TROUBLESHOOTING
The SYSTEM FAULTS display provides pass/fail indications for various parameters
that are continually monitored. These parameters must be within acceptable
operating ranges in order to display PASS. If FAIL is indicated, this indicates a
major failure in that area. If the instrument is in startup mode, START will be
displayed.
Note
The SYSTEM FAULTS screen only indicates PASS or
FAIL for the various analyzer parameters, and is
meant to indicate major failures. Desired operating
ranges are indicated in the INSTRUMENT STATUS
and SYSTEM TEMPERATURE ranges section. If
analyzer readings are not within these ranges, it
could indicate deterioration of certain assemblies
within the analyzer, or minor failures.
The following table lists the possible system fault messages that are displayed on
the primary screen if a major failure occurs. If a fault message is displayed, use
the troubleshooting guide to find the possible cause of the fault.
CHOPPER WHEEL Indicates that the chopper (correlation) wheel is not rotating.
FAILURE
COOLER FAILURE Indicates that the cooler temperature or voltage is not within the
acceptable limits. A fault is indicated if the cooler voltage is above
1.5 volts.
REFERENCE Indicates that the reference voltage is not within the acceptable limits.
VOLTAGE OUT OF A fault is indicated if the reference voltage is below 3 volts or above
RANGE 4.5 volts.
12 VOLT SUPPLY Indicates that the 12 volt supply voltage is not within the acceptable
FAILURE limits. A fault is indicated if the 12 volt supply voltage is below
11.1 volts or above 14.3 volts.
98207601 Rev. B 45
EC9820 CO2 ANALYZER SERVICE MANUAL
4.3.1 Optic
Not supported.
4.3.2 Preamp
The preamp test function generates an electronic test signal which is applied to
the input of the IR detector preamp. This simulates an input from the detector and
is then processed as if it were an actual signal. This test is used to verify the
operation of the detector.
4.3.3 Electric
The electric test function generates an electronic test signal which is applied to the
input of the preprocessor. This simulates a reference and measure input to the
preprocessor and is then processed as if it were an actual signal. This test is used
to verify the operation of the preprocessor board reference and measure channels.
46 98207601 Rev B
CHAPTER 4 TROUBLESHOOTING
If you cannot identify the problem, contact Ecotech at the locations given in
the front of this manual.
98207601 Rev. B 47
EC9820 CO2 ANALYZER SERVICE MANUAL
48 98207601 Rev B
CHAPTER 4 TROUBLESHOOTING
INDEX
98207601 Rev. B 49
EC9830 CO ANALYZER SERVICE MANUAL
2 98207601 Rev B