TUR Is Not Enough 1 1
TUR Is Not Enough 1 1
The question to ask is if they know why they may need a 4:1 TUR and if they may understand the
rationale for requiring a 4:1. The thought here is that a 4:1 ratio is based on specific false accept and
false reject risk, and a 4:1 ratio is a simple way of achieving it if certain conditions can be met.
That thought process alone is dangerous if one does not have enough historical data to use a joint
probability density function associated with many TUR-based methods.
If one does the math, a 4:1 TUR with a coverage probability of k = 2 for the measurement uncertainty
and a 95 % End of Period reliability can equate to less than 1 % false accept and slightly over 1.5 % false
reject (these terms are covered later).
In simplistic terms, End of Period Reliability is defined as the number of calibrations resulting in
acceptance criteria being met divided by the total number of calibrations. The formula to determine the
required sample size from "In-Tolerance" Reliability from historical data is easy to replicate in Excel. The
formula is Sample Size = ln(1-Confidence)/ln(Target Reliability)
If we use the formula for Sample Size above, we would need over 59 (58.4) samples to use a joint
probability distribution associated with many TUR-based methods.
4:1 may sound good on paper, though many laboratories might use the boilerplate language on a
purchase order asking for a 4:1 TUR, likely without the appropriate sample size.
And then, there are different disciplines that, like equipment, cannot easily be grouped into the
calculation based on a global risk approach, equipment that might have different usages, fixturing, wear
patterns, lots with sub-par quality control, different calibration intervals, and more. These different
usages and conditions can lead to statistical independence from the population of like instruments.
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 1
When dealing with physical changes to the instrumentation, like material deformation, as found in many
force and torque measuring equipment, it isn't the same as if we were measuring the voltage of
batteries from a large production lot.
Thus, we must understand the limitations when we analyze requiring a 4:1 TUR as a risk mitigation
strategy to control our probability of false accept risk.
Many labs may use a 4:1 TUR properly and understand what decision rule is best to use to manage their
application's false accept and reject risks.
The paper makes several assumptions about standards, assuming the end-user might request a 4:1 TUR
based on insufficient sample size. These assumptions are based on the author's perception of what
might be happening in the industry.
We will discuss TUR, why the location of the measurement matters, PFA, and some common guard
banding methods used to limit the PFA risk and stay within the lines.
• The requirement is merely a ratio of UUT tolerance limits relative to the expanded
uncertainty of the measurement process. It is, at best, a crude risk control tool, i.e., one that
does not control risks to any specified level. Moreover, in some cases, it may be superfluous.
For instance, what if all UUT attributes of a given manufacturer/model are in-tolerance
prior to test or calibration? In this case, the false accept risk is zero regardless of the TUR.
• The requirement is not applicable when UUT tolerances are single-sided.
• The requirement is only approximately applicable when tolerances are two-sided but
asymmetric and the UUT bias is distributed such that its mode value is zero [2]
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 2
In addition, many fail to realize what is described in "Introduction to Statistics in Metrology,"
section 5.2.1.5 states, "While the 4:1 TUR requirement is commonly used to ensure a
measurement is adequate for making an accept/reject determination, this metric assumes that
the process distribution is centered between the specification limit. If this is not the case, TUR
cannot be reliably used as an indicator of risk" [3]
All of this matters because it is a requirement of ISO/IEC 17025:2017. Section 7.8.6.1 states,
"When a statement of conformity to a specification or standard for test or calibration is
provided, the laboratory shall document the decision rule employed, taking into account the
level of risk (such as false accept and false reject and statistical assumptions) associated with
the decision rule employed and apply the decision rule." [4]
We can think about the measurement risk this way. We have a car, and we need to park it between
two lines. The lines represent the upper and lower specification limit of our device. The width of
our car is the Calibration Process Uncertainty, and parking lines are our tolerance specification
limits.
The probability of us getting a ding or denting another vehicle is our PFA, depending on how
centered we are within the parking lines. If we try to park too close to one side, we may risk not
being able to open the door, or if we misjudge entirely,, we may run right into the car in the other
lane and cause substantial damage. If we park centered on the line, 50 % of our car will be in the
next lane no matter what size our car is.
Many examples cited are assumed to be based on discrete measurements at the bench level
(Specific Risk).
Specific risk is that we are testing an instrument when we do not have a high enough sample size or
information other than where the result is located in relation to the tolerance requested and
calculate our uncertainty correctly to calculate the false accept risk. Figure 1 below shows this
concept.
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 3
Nominal Value 10000
6
Lower specification Limit 9995
Upper Specification Limit 10005
Measured Value 10005 5
Measurement Error 5
Std. Uncert. (k=1) 0.08 4
TUR
TUR or Test Uncertainty Ratio, is defined in Section 3.11 of ANSI/NCSL Z540.3 as, "The ratio of the
span of the tolerance of a measurement quantity subject to calibration, to twice the 95% expanded
uncertainty of the measurement process used for calibration." The TUR tells us how much space
between the lines we must be "in-tolerance." [5]
TUR is commonly used as a simplified approach of evaluating global risk. When we know the
tolerance we are working to, we have a high enough sample size to know the shape and the
distribution of the calibration results and our end-of-period reliability. We can calculate the
appropriate uncertainty that corresponds to the maximum amount of false accept risk we are okay
with.
Per the ANSI/NCSL Z540.3 Handbook, "For the denominator, the 95 % expanded uncertainty of the
measurement process used for calibration following the calibration procedure is to be used to
calculate TUR. The value of this uncertainty estimate should reflect the results that are reasonably
expected from the use of the approved procedure to calibrate the M&TE. Therefore, the estimate
includes all components of error that influence the calibration measurement results, which would
also include the influences of the item being calibrated except for the bias of the M&TE. The
calibration process error, therefore, includes temporary and non-correctable influences incurred
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 4
during the calibration such as repeatability, resolution, error in the measurement source, operator
error, error in correction factors, environmental influences, etc." [5]
PFA Risk
Figure 4: Guard band USL showing a 2 % PFA when Measured Value is at the GB USL
All measurements have a percentage of likelihood of calling something good when it is bad, and
something bad when it is good. You might be familiar with the terms consumer's risk and
producer's risk. Consumer's risk refers to the possibility of a problem occurring in a consumer-
oriented product; occasionally, a product not meeting quality standards passes undetected through
a manufacturer's quality control system and enters the consumer market. The Probability of False
Accept (PFA) is similar to consumer's risk. It is the likelihood of calling a measurement "good" or
stating something is "In Tolerance" when there is a percentage that the measurement is "bad" or
"Out of Tolerance".
ANSI/NCSLI sub-clause 5.3 is the tolerance-type test requirement that "the probability that
incorrect acceptance decisions (false accept) will result from calibration tests shall not exceed 2 %."
With the preponderance of calibrations being of this type, the resources and conditions described
by the calibration procedure will require careful evaluation and determination to achieve the
measurement uncertainty needed for the calibration process to achieve this allowable probability
of false accept." The measurement uncertainty must be accounted for, and the acceptance limits
must be calculated to ensure the likelihood of the measurement being "Out of Tolerance" does not
exceed 2 %.
The entire purpose of analyzing the PFA is to ensure your measurements are "In Tolerance" with
risk that does not exceed 2 %. And why just knowing you have a 4:1 TUR without analyzing the PFA
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 5
regarding the location of the measurement is not enough to minimize measurement risk as shown
in Figure 3. Figure 3 shows the upper and lower guard banded limits to ensure a PFA of 2 % or less.
If the measured value is not within the guard band limits, the PFA will be higher than 2 %.
Figure 5: Graph Showing 10,004 as the measured value with a 31.23:1 TUR which is achieved by using a lab with low
uncertainties
Calling an instrument “In Tolerance” is all about location, location, location. It's also about the
uncertainty of the measurement, but a bad location will raise the Probability of False Accept (PFA)
significantly.
The probability of false acceptance is the likelihood of a lab calling a measurement “In Tolerance”
when it is not. The location we are referring to is how close the measurement is to the nominal
value. If the nominal value is 10,000 lbf and the instrument reads 10,004 lbf, the instrument bias is
4 lbf, as shown in Figure 4.
The larger the bias, the worse the location of the measurement. If we go back to our parking
scenario, the worse the bias from nominal, the more likely one side of our automobile will be
damaged, or maybe we are still “in tolerance” but have to exit the vehicle from the other side.
Higher TURs help control PFA. If the End of Period Reliability (EOPR) is fixed, the TUR will decrease
as the measurement process uncertainty increases.
Figure 5 shows this concept as risk level increases as we have switched calibration providers, and
the new provider has a higher CMC uncertainty component of 0.025% than shown in Figure 4
where the calibration provider had a 0.0016% CMC uncertainty component; everything else has
remained the same.
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 6
Nominal Value 10000
0.35
Lower specification Limit 9995
Upper Specification Limit 10005
0.3
Measured Value 10004
Measurement Error 4
0.25
Std. Uncert. (k=1) 1.25
Figure 6: Graph Showing 10,004 as the measured value with a 1.99:1 TUR as the labs Expanded uncertainty is higher than
in figure 4
Why do we care about the location of the measurement if the device is within tolerance? If a device has
a specification of 0.1 % of full scale and the calibrating laboratory reports a value within 0.1%, the device
is “In Tolerance,” right?
The answer is and always will be it depends on the measurement is uncertainty and if the lab
performing the calibration has adequately calculated their Calibration Process Uncertainty correctly and
followed the proper guidelines in determining the uncertainty of measurement when making the
statement of compliance.
If the uncertainty of the measurement is significant, the lab performing the calibration will have to be
very concerned with the location of the measurement.
If their uncertainty of measurement is too high, they may not even be able to perform the calibration at
all, and if the measured value falls right on the specified tolerance line, the PFA can be 50 % or higher.
There are several methods to ensure a 2 % PFA requirement can be met. These TUR-based decision rules
typically set acceptance limits to ensure the PFA is less than 2 %.
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 7
9994 9995 9996 9997 9998 9999 10000 10001 10002 10003 10004 10005 10006
MV LSL Nominal Value USL Uncert. Dist
UGB LGB UM6 LM6
Figure 7: Graph Showing Specification Limits and Acceptance Limits for Both Method 5 and Method 6
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 8
Rules and Statements of Conformity has several examples of decision rules and the corresponding
risk associated with those rules.
Guard Band Method 5, Based on the Expanded Calibration Process Uncertainty. This method is
simple as one subtracts the 95 % expanded process uncertainty from the tolerance limits. The
above graphs in figures 1 through 4 use Method 5. It is very similar to the ILAC G8: 2009 rule.
ILAC-G8:03/2009 states if the specification limit is not breached by the measurement result plus
the expanded uncertainty with a 95 % coverage probability, then compliance with the specification
can be stated. The ILAC rule allows for a PFA of < 2.5 %.
Simply put, if one subtracts the expanded calibration process uncertainty from the upper limit of
the specified tolerance, then the new acceptance limits will assure a PFA of less than 2.275 %. It is
an interesting point as the ANSI/Z540.3 Handbook mentions 2 %, though the calculation gives a
PFA of 2.275 %. One must assume some rounding took place. The only information needed to use
Method 5 is the tolerance and the calibration process uncertainty formula in the figure below.
Note: See ILAC-P14:09/2020 section 5.4 for requirements for calculating CMC and reporting
measurement uncertainty. The requirements align very closely with the formula in Figure 7.
The downside of Method 5 is that the test limit is based on the worst-case PFA, which means they
may be too aggressive, resulting in more false rejects from the reference laboratory. Being overly
aggressive and needing to adjust more equipment lends one to look for an alternative method.
Guard Band Method 6, Based on Test Uncertainty Ratio:
This method is also simple as it depends only on the measurement uncertainty when compared
with the specification limits of what is being calibrated. Per ANSI/NCSLI Z540.3 Handbook, "It
makes use of an observation that for a given Test Uncertainty Ratio (TUR), there is a maximum PFA
value for all values of the M&TE test point in-tolerance probability. Applying a guard band based on
this maximum PFA value, and the corresponding TUR ensures that the PFA is 2 % or less regardless
of the in-tolerance probability." It also results in guard bands with acceptance limits much larger
than that of method 5.
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 9
The downside of Method 6 is it only works with TUR ratios of 0.76:1 through 4.6:1. Any ratio higher
or lower will cause errors with not calculating the acceptance limits properly.
Comparing Method 5 versus Method 6
Below is a table using the same 10,000 lbf device, using the same variables as shown in figures 2-4,
which are a 0.01 resolution and a CMC uncertainty component of 0.0016 % from Morehouse, who
was used as the reference laboratory resulting in a 0.08 lbf calibration process at the 10,000 lbf pt.
DIFFERENCE IN ACCEPTANCE LIMITS METHOD 5 VS METHOD 6
Force or Torque Applied Avg Instrument Reading Method 6 AL PASS/FAIL Method 5 AL PASS/FAIL % Diff in AL
1000 1000.00 5.00 PASS 4.98 PASS 0.33%
2000 2000.00 5.00 PASS 4.97 PASS 0.63%
3000 3000.00 5.00 PASS 4.95 PASS 0.94%
4000 4000.00 5.00 PASS 4.93 PASS 1.25%
5000 5000.00 5.00 PASS 4.92 PASS 1.57%
6000 6000.00 4.99 PASS 4.90 PASS 1.88%
7000 7000.00 4.99 PASS 4.88 PASS 2.19%
8000 8000.00 4.99 PASS 4.87 PASS 2.50%
9000 9000.00 4.99 PASS 4.85 PASS 2.81%
10000 10000.00 4.99 PASS 4.84 PASS 3.13%
Figure 9: Difference in Acceptance Limits Method 5 versus Method 6 using a Reference Standard with an Expanded
Uncertainty of 0.0016 %
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 10
DIFFERENCE IN ACCEPTANCE LIMITS METHOD 5 VS METHOD 6
Force or Torque Applied Avg Instrument Reading Method 6 AL PASS/FAIL Method 5 AL PASS/FAIL % Diff in AL
1000 1000.00 4.99 PASS 4.74 PASS 4.89%
2000 2000.00 4.97 PASS 4.49 PASS 9.80%
3000 3000.00 4.96 PASS 4.23 PASS 14.74%
4000 4000.00 4.95 PASS 3.97 PASS 19.70%
5000 5000.00 4.93 PASS 3.72 PASS 24.69%
6000 6000.00 4.84 PASS 3.46 PASS 28.45%
7000 7000.00 4.72 PASS 3.20 PASS 32.10%
8000 8000.00 4.59 PASS 2.95 PASS 35.84%
9000 9000.00 4.46 PASS 2.69 PASS 39.71%
10000 10000.00 4.32 PASS 2.43 PASS 43.74%
Figure 10: Difference in Acceptance Limits Method 5 versus Method 6 using a Reference Standard with an Expanded
Uncertainty of 0.025 %
Force or Torque Applied Avg Instrument Reading Method 6 AL PASS/FAIL Method 5 AL PASS/FAIL % Diff in AL
1000 1000.00 4.97 PASS 4.49 PASS 9.80%
2000 2000.00 4.95 PASS 3.97 PASS 19.70%
3000 3000.00 4.84 PASS 3.46 PASS 28.45%
4000 4000.00 4.59 PASS 2.95 PASS 35.84%
5000 5000.00 4.32 PASS 2.43 PASS 43.74%
6000 6000.00 4.04 PASS 1.92 PASS 52.48%
7000 7000.00 3.74 PASS 1.41 PASS 62.40%
8000 8000.00 3.43 PASS 0.89 PASS 73.97%
9000 9000.00 3.11 PASS 0.38 PASS 87.81%
10000 10000.00 2.78 PASS -0.13 FAIL 104.83%
Figure 11: Difference in Acceptance Limits Method 5 versus Method 6 using a Reference Standard with an Expanded
Uncertainty of 0.05 %
When we analyze the data in Figures 8 through 10, it becomes apparent that the differences
between Method 5 and Method 6 start to become quite drastic as the calibration process
uncertainty increases. The CMC uncertainty component of the reference laboratory impacts the
calibration process uncertainty, the resolution of the Test Instrument, and possibly the
repeatability of the Test Instrument, which may or may not have been included in the calibration
process uncertainty.
The laboratory with the low CMC uncertainty component in Figure 8, shows the least amount of %
difference from using Method 5. However, the formulas are based on the measurement process
uncertainty, which includes the UUT's resolution and repeatability.
If the UUT's resolution and repeatability were to increase, the % difference would increase. Method
5 is the most affected as we subtract the measurement process uncertainty from the upper
specification limits and add it to the lower specification limit to obtain our acceptance limits.
Figure 10 shows that the calibration laboratory would not make a conformity assessment using
Method 5 at the last calibrated test point. However, using Method 6 allows that same laboratory to
make a statement of conformity, assuming the measured value falls within the specified tolerance
limits.
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 11
Conclusion
Any method used for calculating PFA will have both positive and negatives associated with
implementation. The ISO/IEC 17025:2017 standard better addresses measurement risk by
requiring the laboratory to report which specifications are not met and the decision rule applied.
The decision rule applied needs to consider the location of the measurement for reporting False
Accept Risk (PFA).
Throughout this paper, the author has demonstrated that TUR only shows the ratio of the specified
tolerance compared to the calibration process uncertainty. If the ratio is too large, a laboratory may
not be able to make a statement of conformance with complying with ISO/IEC 17025:2017.
Furthermore, the author shows why a 4:1 or better TUR might not be enough to control risk
without several other conditions being met.
It is important to analyze the measurement's location to ensure the measured value falls within the
acceptance limits calculated by the accepted guard banding method used.
The best chance of continually meeting tolerance requirements is to use a reference lab (Calibration
vendor) with the lowest CMC uncertainty component that replicates how the instrument is used.
Also, the end-user must purchase the right equipment capable of continually achieving the desired
result or adjust the tolerance appropriately.
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 12
Annex (Sample Calculation of TUR)
Example: A customer sends a 10,000 lbf load cell for calibration with an accuracy specification of ±
0.05 % of full scale. The calibration provider uses a Universal Calibrating Machine to perform the
calibration. When 10,000 lbf is applied, the unit reads 10,001 lbf. The display resolution is 1 lbf.
The device is a 10,000 lbf load cell with an accuracy specification of ± 0.05 %
10 𝑙𝑏𝑓
𝑇𝑈𝑅 =
2
𝐶𝑀𝐶 2 𝑅𝑒𝑠𝑜𝑙𝑢𝑡𝑖𝑜𝑛𝑈𝑈𝑇 2 𝑅𝑒𝑝𝑒𝑎𝑡𝑎𝑏𝑖𝑙𝑖𝑡𝑦𝑈𝑈𝑇 2
2 𝑥 𝑘95% ( √( ) + ( ) + ( 1 ) + ⋯ (𝑢𝑂𝑡ℎ𝑒𝑟 )2 )
𝑘𝐶𝑀𝐶 2
√12
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 13
CMC is the uncertainty at the calibrated force. The Universal Calibrating Machine has an uncertainty
of 0.02 % at 10,000 lbf.
Dividing the CMC by 2, the standard uncertainty is reported at one standard deviation. In most
cases, the CMC uncertainty component is reported at approximately 95 %, and a coverage factor of
k = 2 is used.
10 𝑙𝑏𝑓
𝑇𝑈𝑅 =
2
2 𝑙𝑏𝑓 2 𝑅𝑒𝑠𝑜𝑙𝑢𝑡𝑖𝑜𝑛𝑈𝑈𝑇 2 𝑅𝑒𝑝𝑒𝑎𝑡𝑎𝑏𝑖𝑙𝑖𝑡𝑦𝑈𝑈𝑇 2
2 𝑥 𝑘95% ( √( 2 ) + ( 2 ) + ( 1 ) + ⋯ (𝑢𝑂𝑡ℎ𝑒𝑟 )2 )
√12
UUT Resolution
ResolutionUUT for force instrument is calculated by dividing the force applied by the output at
applied force and then multiplying this by the instrument's readability.
To convert 1 lbf resolution to standard uncertainty, it is either divided by the square root of 12, or
the square root of 3 depending on the Type of resolution.
10 𝑙𝑏𝑓
𝑇𝑈𝑅 =
2
2 𝑙𝑏𝑓 2 1 𝑙𝑏𝑓 2 𝑅𝑒𝑝𝑒𝑎𝑡𝑎𝑏𝑖𝑙𝑖𝑡𝑦𝑈𝑈𝑇 2
2 𝑥 𝑘95% ( √( 2 ) + ( 2 ) + ( 1 ) + ⋯ (𝑢𝑂𝑡ℎ𝑒𝑟 )2 )
√12
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 14
Repeatability
Repeatability is obtained by applying a force of 10,000 lbf to the Unit Under Test (UUT) five times,
and the sample standard deviation of five replicated measurements is calculated.
Repeatability of sample size five: (10,000, 10,001, 10,000, 10,001, 10,001) = 0.54772
Since the repeatability is already expressed as one standard deviation, the divisor is 1.
10 𝑙𝑏𝑓
𝑇𝑈𝑅 =
2
2 𝑙𝑏𝑓 2 1 𝑙𝑏𝑓 2 0.54772 2
2 𝑥 𝑘95% ( √( ) + (2 ) + ( ) + ⋯ (𝑢𝑂𝑡ℎ𝑒𝑟 )2 )
2 √12 1
Other error sources attributed to the CPU can be considered for the UUT. Some examples are
environmental influences, error in correction factors, etc. For this example, other error sources are
inherent in repeatability and CMC.
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 15
10 𝑙𝑏𝑓
𝑇𝑈𝑅 =
2
2 𝑙𝑏𝑓 2 1 𝑙𝑏𝑓 2 0.54772 2
2 𝑥 𝑘95% ( √( 2 ) + ( 2 ) + ( 1 ) )
√12
10 𝑙𝑏𝑓
TUR =
2 𝑥 𝑘95% (1.1762)
The specification of 10 lbf is divided by: 2 * k at 95 % Calibration Process Uncertainty (k = 2 for this
example)
TUR = 2.1256
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 16
References
[1] Requirements for the Calibration of Measuring and Test Equipment, 2006, ANSI/NCSL Z540.3-
2006
[2] NCSLI RP – 18 Estimation and Evaluation of Measurement Decision Risk
[3]Crowder, Stephen; Delker, Collin; Forrest, Eric; Martin, Nevin. 2020. Introduction to Statistics in
Metrology. Springer Nature Switzerland AG.
[4] ISO/IEC 17025:2017 General requirements for the competence of testing and calibration
laboratories
[5] Handbook for the ANSI/NCSL Z540.3-2006, 2009, ANSI/NCSL International
Why a 4:1 TUR is not Enough: The Importance of Analyzing the Probability of False Accept Risk
Author: Henry Zumbrun
12/2021 Page 17