500DT_Operator Manual_RevA_02092005
500DT_Operator Manual_RevA_02092005
OPERATORS MANUAL
Welcome
Smiths Detection prides itself on its comprehensive customer service. If any assistance
is needed for scientific, operational or maintenance issues, or if problems arise, please
contact your local Smiths Detection representative, or Smiths Detection directly.
Warranty
The delivery of repair or replacement parts shall not interrupt or prolong the term of
the warranty. Smiths Detection’s warranty ceases to be effective if Purchaser fails to
operate and use the equipment sold hereunder in a safe and reasonable manner and in
accordance with Smiths Detection’s written instructions.
Notwithstanding anything in this warranty to the contrary, Smiths Detection shall not
in any event be liable to Purchaser or any other person for any liability, claim, loss,
damage or expense of any nature whatsoever caused directly or indirectly by the
equipment or any inadequacy thereof for any purpose, or any deficiency or defect
therein, or the use or maintenance thereof, or any delay in providing or failure to
provide servicing or adjustments thereto, or any interruption or loss of service or use
thereof, or any loss of business, or any incidental or consequential damages (including
loss of profit), whatsoever or howsoever caused.
Table of Contents
Chapter 1 Safety
Smiths Detection instruments are designed with user safety in mind. Throughout this
manual, the following warnings are prominently displayed followed by a detailed
description of the hazard.
WARNING!
Radiation Hazard
WARNING!
Burn Hazard
WARNING!
Risk of Electrical Shock
The IONSCAN 500DT meets all the following safety and regulatory mandates:
• IEC 61010-2 • IEC 61326, • 73/23/EEC • FCC Part 15,
• IEC 61000-4-2 Class B • 89/336/EEC Subpart B,
Class B
• IEC 6100-4-3 • ISA S82.01 • 89/392/EEC
• IEC 6100-4-4 • ISO 1677 • CSA 22.2-1010.1 • AECB R-116
• IEC 61000-4-6 • ISO 2919 • ANSI N542 • UL3101-1
The IONSCAN 500DT detection system contains two sealed nickel-63, 63Ni, foil
radioactive sources of 555 MBq (15 mCi) activity. The source is fully enclosed inside
the unit assembly resulting in no direct radiation hazard.
WARNING!
Radiation Hazard - Risk of radiation exposure.
The equipment must be handled in accordance with the licensing requirements of the
applicable nuclear regulatory agency. Refer to “Handling of Radioactive Sources
(USA)” on page B-1.
Should the IONSCAN 500DT require disposal or no longer be required for use, return
the system to Smiths Detection for radioactive source removal.
Chapter 2 Overview
The sample of interest is heated to vapourize the compounds collected, which are then
introduced into the reaction region, through the membrane, by a carrier flow of clean,
dry air. The reaction region has a weak radioactive ionizing source, a Nickel 63 (63Ni)
Beta emitter. Both positive and negative ions are formed by collisions of Beta particles
with the gases in the sample stream. Explosive substances form negative ions, while
most narcotics, such as heroin and cocaine, form positive ions.
Exhaust Drift
Flow Flow
Inlet
Decreasing Potential
Collector
Repelling Focusing
Sample Grid Rings
Desorber Guard Grid
Ionizing
Source Gating Grid
Sample
Flow
The IMS has an electric field applied along its length. The polarity of this field is either
positive or negative depending on the type of substance tested for. The ions of the
correct polarity are allowed to move from the reaction region into the drift region by
the momentary opening of a gating grid. The various ions travel at different, but
characteristic speeds and arrive at the collector electrode at unique drift times. Thus,
three ions (X, Y and Z) of differing masses will produce the drift spectrum, called a
plasmagram. Refer to Figure 2-2, where ion X arrives first having the lowest mass, and
thus the shortest drift time.
The time required for ions of a specific substance to drift down the tube is precisely
known and programmed. A reference substance (the calibrant) of known mobility is
used to ensure the system is operating properly. A microprocessor monitors the
collector electrode signal at the programmed drift times to detect the presence of
target ions using custom-written software for the control and peak identification.
When a substance of interest is detected, the software immediately triggers an alarm
to alert the operator.
Colour LCD
Touch Screen
Sample Inlet
Built-in printer
On/Off Button
USB port
The IONSCAN 500DT sample collection system uses a sampling wand which ensures
that every sample is positioned correctly for desorption, with minimal operator
handling, resulting in a complete and uncontaminated sample analysis.
The sampling wand features a detachable sampling frame which, when containing a
sampling swab, is inserted into the sample inlet in the front of the detector unit and
locked into place. The sampling arm is then detached from the sampling frame and
removed, while the sample remains in the detector for analysis.
After analysis, the operator reattaches the sampling arm to the sampling frame to
remove it, and the IONSCAN 500DT is immediately ready for the next sample.
The graphical user interface has been designed for ease of use. Using touch screen
technology, operators can easily access the menus and screens for operational or
informational purposes. The Smiths button displays the current user and instrument
information. The time and date display at all times. The status bar is colour coded
(green for ready, red for alarm or yellow for waiting) to indicate the current detector
status. The message bar indicates the current progress during waiting and also
displays other important messages during the course of detector operation. Refer to
Figure 2-5 for a detailed view of the touch screen display area.
Two IMS drift tubes are incorporated into a single detector unit. A common inlet
connects the two drift tubes to the analysis of a single swab by both drift tubes, each
being controlled with respect to polarity, electric field gradient, temperature, and flow.
Each IMS drift tube is completely independent, allowing for separate conditions, and
each tube is optimized for the substances being searched for.
Electronics
APS System
Dual IMS
Drift Tubes
Inlet Area
The regenerative air purification system (APS) is a heated regenerative compressed air
dryer, designed to remove hydrocarbons from the purified processed air. The APS
uses two types of molecular sieve. Refer to Figure 2-7.
The regeneration of the molecular sieve uses head and purge flow to remove
accumulated water and contaminants. Under ideal conditions, the regenerative
process can be repeated indefinitely. However, environmental air contaminants and
operation in less than ideal conditions will shorten the service life of the molecular
sieve.
During operation, the APS uses two towers which contain molecular sieves. While one
tower generates dry clean air, the second tower is regenerated, or purged. At fixed
intervals, the towers alternate their functions. The tower that produced dry clean air is
regenerated, while the previously regenerated tower provides dry clean air.
To purge the molecular sieve, the regenerated tower is supplied with a 5%-10%
portion of dry clean air generated by the purifying tower. In addition, the regenerated
tower is heated to ~200°C to volatilize the adsorbed water and contaminants to be
purged by the purge flow. The purge flow and water trap drain are connected to a
sponge that collects the expelled water, and a fan directed at the sponge aids in
evaporating the water.
APS functions are controlled by a microprocessor. The microprocessor controls the
towers intervals, purged tower temperature, and water trap drain. In case of a power
interruption to the APS, the microprocessor records the last sequence, which will be
continued when power is restored. In addition, the microprocessor visually notifies
the operator of an impending molecular sieve replacement, based on the number of
tower cycles.
Sample testing using the IONSCAN 500DT consists of the following steps:
1. A swab is inserted into the head of the sampling wand, by hand. Each swab will
have a multi-use capability of approximately 15-20 samples.
2. A sample is collected by wiping a suspect surface with the swab.
3. The sampling head is inserted into the front of the detector unit. Once inserted, a
lock is triggered and the sample is locked into place.
4. The sampling arm is detached from the sampling head.
5. The analysis cycle begins with the desorber moving up and sealing the sample
between the desorber heater and the IMS inlet. The screen will display
“Analyzing” to indicate that an analysis is being performed.
As the sample is heated, the sample carrier gas (clean, dry air) sweeps the
desorbed vapours into the IMS detector where they are ionized in the reaction
region, forming ions or ionic clusters of specific mobilities.
6. The gating grid opens, letting ions of the correct polarity enter the drift region. The
ions are then focused and accelerated by the electric field along the drift region of
the IMS tube to arrive at the collector electrode (typically in 10 to 20 milliseconds).
The magnitude of the collector current, as a function of time, is proportional to the
number of ions arriving at the moment. The IMS detector is extremely selective
because the drift times (ion mobilities) are very specific for individual chemical
compounds.
The time required for ions of a specific substance to drift down the tube is
precisely known and pre-programmed. A calibration substance (the calibrant) of
known mobility is used to ensure the system is operating properly.
7. A microprocessor monitors the collector electrode signal at the programmed drift
times to detect the presence of target ions using custom-written software for
control and peak identification.
8. In the event of a pass analysis, the screen will change to a green background and
“No alarm detected” will display.
In the event of an alarm analysis, the screen will change to a red background and
the detailed information of the compound(s) detected will display. Optionally, an
audible alarm will sound.
9. Once sampling is complete, the operator reattaches the sampling arm to the
sampling head to remove it from the detector. The detector is immediately ready
for the next sample.
2.8.3 Software
The IONSCAN 500DT is based on embedded software running in a Linux operating
environment in real time mode and has the following specifications.
2.8.4 Consumables
The following table lists the consumables required for the IONSCAN 500DT.
Chapter 3 Installation
Network
USB Port
USB Port
External
Printer
AC Power
On/Off Switch
A regular power cable is used to plug the IONSCAN 500DT Detector into a grounded
AC outlet. A power cable with approved plugs can be purchased from Smiths
Detection or from a local electrical retailer.
This section guides the operator through the set up options available in the IONSCAN
500DT, as well as provides a basic understanding of IONSCAN 500DT features.
All navigation between the menus and screens is performed by touching the buttons
directly on the LCD touch screen.
For information on IONSCAN 500DT operating functions, refer to Chapter 5.
The Settings menu enables the set up of the screen saver, audio settings, alarm script,
display settings, verific script and preferences options.
Use the following procedure to access the Settings menu.
1. From the Ready screen, press the Menu button to access the Main menu. Refer to
Figure 4-2.
2. From the Main menu, press the Settings button to access the Settings menu. Refer
to Figure 4-3.
Refer to sections 4.1.1 through 4.1.5 to for detailed information on the options
available in the Settings menu.
Use the following procedure to enable and modify the screen saver options.
1. From the Ready screen, press the Menu button to access the Main menu.
2. From the Main menu, press the Settings button to access the Settings menu.
3. From the Settings menu, press the Screen Saver button to proceed.
4. Select the “Enable screen saver” check box to activate the screen saver option.
5. Using the up or down arrows, set the time in minutes that the IONSCAN 500DT
will remain idle prior to activating the screen saver and entering standby mode.
6. Press the Save button to save the screen saver settings and return to the Settings
menu.
Option Description
3. Press the Save button to save the settings and return to the Preferences menu.
Option Description
Option Description
3. Press the Save button to save the changes and return to the Preferences menu.
Option Description
3. Press the Save button to save the settings and return to the Preferences menu.
To set the date and time options, use the following procedure.
1. From the Preferences menu, press the Date/Time Settings button to proceed.
2. Using the up and down arrows, set the current date, then select the preferred date
format.
3. Using the up and down arrows, set the current time, then select the preferred time
format.
4. Press the Save button to save the selections. A pop-up message will display
indicating that changes have been saved. Press the OK button, then press either
Main to return to the Main screen, or Menu to return to the Main menu.
At any time during operation, the Screen Settings button can be selected to modify the
brightness and contrast of the screen display.
Use the following procedure to set the screen settings.
1. Press the Screen Settings button on the right side of the display. The button will
change to display the brightness and contrast controls. Refer to Figure 4-15.
2. Adjust the screen settings, then press the centre of the button to set the changes
and exit screen settings.
This chapter describes the basic operation of the IONSCAN 500DT and includes
information about the following topics:
• User interface
• Starting up
• Preparing the sampling wand
• Running a verification
• Running a blank
• Performing an analysis
• Analysis results
• Viewing saved samples
Advanced features, such as accessing Main menu functions, modifying and viewing
control parameters, and performing administrative tasks are discussed in Chapter 6.
All navigation between the menus and screens is performed by touching the buttons
directly on the LCD touch screen.
The Main screen is the primary screen the operator uses to monitor the status of the
IONSCAN 500DT. Refer to Figure 5-1 for a view of the Main screen, and Table 5-1 for a
detailed description of the menus, messages and status fields.
Feature Description
On/Off
Switch
On/Off
Button
The IONSCAN 500DT will begin a self-test and warm-up. The warm-up procedure
takes approximately 30 minutes during normal use. If the unit has been stored for
any length of time, or is being started for the first time, warm up may take a little
longer. The Main screen will display prompting the operator to log in.
NOTE: During operation, the IONCAN 500DT continuously monitors its
components. In the event of a component error, the LCD will remain
yellow and an error message will be displayed to alert the operator. The
green Ready screen indicates that all system functions are correct and set
points have been reached.
1. Operators must log in to perform analyses. It is possible to log in while the system
is warming. Press the Login button to display the Login screen. Refer to Figure 5-4.
NOTE: Once a user is logged in, the Login button will change to a Logout button.
2. Enter the user name and password, and then press the OK button to continue. The
Main screen will display and indicate that the user has logged in to the system.
The swabs used in the sampling wand can be used 15-20 times, depending on the
cleanliness and dryness of the surfaces being sampled. Optionally, upon completion of
an analysis, a message may display to prompt the operator to change the swab in the
sampling wand.
Use the following procedure to prepare the sampling wand for analyses.
1. Wearing powder-free latex gloves, press the trigger on the arm of the sampling
wand to lift the shaper.
4. Snap the lid on the sampling frame back into place, then release the shaper.
The IONSCAN 500DT should be verified as soon as the unit becomes Ready at the
beginning of each shift to ensure that it will alarm correctly. Verification can also be
performed at any time the user needs to verify the unit is operating correctly.
Use the following procedure to run a verification.
1. Press the Menu button to display the Main menu.
2. From the Main menu screen, press the Verific button. The Verification screen will
display. Refer to Figure 5-8.
CAUTION!
To avoid damage to the desorber and sampling inlet, always disengage the arm
from the sampling wand immediately after the sampling frame is inserted into the
sampling inlet and the analysis begins.
6. Insert the sampling frame into the sample inlet of the IONSCAN 500DT with the
swab facing upwards. Analysis will begin automatically. Immediately disengage
the arm from the sampling wand as soon as analysis begins.
Sampling
Frame
Magnet
Sampling Arm
CAUTION!
When running a verification, use caution to ensure the IONSCAN 500DT, sampling
wand, work area and the operator do not become contaminated.
10. Wearing clean powder-free latex gloves, lightly apply the verification standard to
the sample area of the clean swab.
11. Roll the sampling wand with your fingertips until the swab is in the upright
position.
12. Squeeze the trigger to release the shaper from the sampling frame.
13. Insert the wand with the verification standard into the sample inlet of the
IONSCAN 500DT with the swab facing upwards. Analysis will begin
automatically. Immediately disengage the arm from the sampling wand as soon as
analysis begins.
14. After the verification analysis, insert the arm of the sampling wand into the
sampling inlet. The magnet on the sampling arm will clasp the sampling frame.
15. Gently remove the sampling frame from the sampling inlet.
16. Refer to “Verification Successful” on page 5-9 and “Verification Unsuccessful” on
page 5-9.
5. Press the Back button, then press the Save button if the settings have been
modified, or press the Cancel button to return to the Control Parameter Sets screen
without saving changes.
CAUTION!
To avoid damage to the desorber and sampling inlet, always disengage the arm
from the sampling wand immediately after the sampling frame is inserted into the
sampling inlet and the analysis begins.
5. Insert the sampling wand into the sample inlet on the IONSCAN 500DT. Analysis
will begin automatically. Immediately disengage the arm from the sampling wand
as soon as analysis begins.
Sampling
Frame
Magnet
Sampling Arm
If the blank procedure was unsuccessful, indicated by a red “alarm” screen, press the
Reset Alarm button. Repeat “Running a Blank” on page 5-10 until two successful
blanks are obtained. In the event of persistent contamination, refer to “Cleaning After
an Alarm” on page 5-16.
Sampling Tips:
• Smiths Detection recommends wearing powder-free latex gloves.
• Use a firm swaying motion with the sampling wand.
• Sample areas where hands and fingertips would touch, such as handles, carrying
straps, zippers, car door handles, steering wheel, etc.
• Swabs can be used 15-20 times, depending on the cleanliness and dryness of the
swab.
3. Roll the sampling wand with your fingertips until the swab is in the upright
position.
4. Squeeze the trigger to release the shaper from the sampling frame.
CAUTION!
To avoid damage to the desorber and sampling inlet, always disengage the arm
from the sampling wand immediately after the sampling frame is inserted into the
sampling inlet and the analysis begins.
5. Insert the sampling frame into the sample inlet of the IONSCAN 500DT with the
swab facing upwards. Analysis will begin automatically. Immediately disengage
the arm from the sampling wand as soon as analysis begins. Refer to Figure 5-14.
Sampling
Frame
Magnet
Sampling Arm
6. After the sample analysis, insert the arm of the sampling wand into the sampling
inlet. The magnet on the sampling arm will clasp the sampling frame.
NOTE: When analysis is complete, a message may display to prompt the operator
to change the swab in the sampling wand. Replace the swab in the
sampling wand and follow the prompts to reset the counter.
7. Gently remove the sampling frame from the sampling inlet.
8. Refer to “Pass Result” on page 5-14 and “Alarm Result” on page 5-14.
In the event of an alarm result, the colour coded status bar will change to red, the
alarm substance(s) and strength will appear on the display, and “Press Alarm Reset.
Run a blank” will display in the message bar.
1. Press the Reset Alarm button to clear the alarm.
2. Press the Show Alarm Script button to display the alarm resolution protocols.
3. Replace the swab in the sampling wand with a new swab.
4. Smiths Detection recommends performing a blank procedure after an alarm result.
Refer to “Running a Blank” on page 5-10.
Detail Description
Detail Description
Selecting the Samples option enables the user to view the last sample saved and, with
appropriate access, view any of the archived samples and relevant data.
Use the following procedure to view saved samples.
1. Press the Menu button to access the Main menu.
2. Select the Samples button to proceed. The Samples screen will appear. Refer to
Figure 5-18.
3. Select the sample of interest by either entering a range in the From, To and Type
fields, or use the vertical scroll bar to navigate through the saved samples.
4. Select the appropriate display option from the buttons in the top of the screen. Use
the left and right scroll arrows to navigate through the display options, if
necessary.
Field Description
Field Description
5.7.1 Logout
Pressing the Logout button from any screen will log the current operator out of the
system. Users may log out of the system at any time, except when:
1. The machine is actively analyzing a sample, including verific, running a blank, or
performing background analysis.
2. Any modal dialog is presented including a bakeout cancel confirmation, or alarm
reset.
3. Certain processes are in progress, such as powering down and during a clean
cycle.
WARNING!
Risk of electrical shock - When performing maintenance duties, use
the following procedure to shut down the system, then switch the
On/Off button at the rear of the detector to the Off position.
Selecting the system shutdown option will automatically log off the current user. All
data will be saved and software components will be shut down in a controlled
manner, ensuring data integrity.
To shut down the IONSCAN 500DT at the end of a shift or to perform maintenance
duties, use the following procedure.
1. Press the Menu button to access the Main menu.
2. From the Main menu, press the System Shutdown button. The System Shutdown
screen will display. Refer to Figure 5-19.
3. Press the OK button to proceed with shutting down the system. Switch the On/Off
button at the rear of the detector to the Off position if the system was shut down to
perform maintenance duties.
This chapter describes the advanced features of the IONSCAN 500DT and includes
information about the following topics:
• Main menu options
• Administrative menu options
• Modifying and viewing control parameters
• Modifying and adding user profiles
• Data management
• Generating Reports
The Main Menu enables operators with appropriate access to perform preventive
maintenance procedures, view saved samples, adjust the settings, adjust the control
parameters, perform administrative tasks, and shut down the system.
6.1.1 Verific
For information on running a verification, refer to “Running a Verification” on
page 5-7.
6.1.2 Samples
For information on viewing saved samples, refer to “Viewing Saved Samples” on
page 5-17.
6.1.3 Maintenance
For information on maintenance options, refer to “Maintenance Menu” on page 7-3.
6.1.4 Settings
For information on settings options, refer to “Settings Menu” on page 4-1.
CAUTION!
Smiths Detection recommends proceeding with caution. Changing operating
parameters could impact the validity of detection results. Contact a Smiths
Detection Service Engineer for guidance prior to performing any modifications.
Use the following procedure to access the control parameters for a specific parameter
set.
1. From the Main Menu screen, press the Control Parameters button to proceed. The
Control Parameter Sets screen will display.
2. Identify the control parameter set to edit from the list, then press the Select button.
The Control Parameter Set menu will display.
NOTE: The Save and Save As buttons will only be active when changes have been
made within the control parameter options.
Button/Field Description
Button/Field Description
6.1.5.2 Description
Selecting the Description option enables a user with appropriate access to change the
description of the selected control parameter set.
Use the following procedure to change the description of the currently selected control
parameter set.
1. From the Control Parameter Set Menu, press the Description button to proceed.
An on-screen keyboard will display. Refer to Figure 6-4.
2. Using the keyboard, edit the control parameter set description and then press
Enter to accept the change.
3. Press the Back button to return to the Control Parameters Set menu. The
Description button text will display in blue, indicating that a parameter change
within the option has not yet been saved.
4. Press Save to update the control parameter set, or press the Save As button to save
the amended control parameter set as a new set, so as not to change the original
parameter set.
6.1.6 Administration
The Administration Menu options control the settings for access levels, user profiles,
configure devices, data management, report management, and e-mail settings.
To access the Administration menu, use the following procedure.
1. From the Main menu, press the Administration button to proceed. The
Administration menu will display. Refer to Figure 6-7.
2. Select the function from the list. Use the slider, or up and down scroll bar arrows to
view the entire list of functions.
3. Select the option within the selected function that requires the access level change
and then use the Increase or Decrease buttons to alter the access level for that
option.
NOTE: All access levels higher than that level also have access to that item. For
example, if an item is given an access level of 3, then operators with access
levels 3, 4, or 5 have access to that item.
4. Repeat Step 2 and Step 3 for each option that requires an access level modification.
5. Press the Save button to save the changes, acknowledge the pop-up confirmation
message, and then press the Administration button to return to the Administration
menu.
3. Using the on-screen keyboard, select each field to type in the user name, login ID,
password, confirmation of password, and access level for the new user.
If the new user profile is a peer administrator, select the Peer Administrator
option.
4. Press the Save button to return to the User Profiles screen. The new profile will be
displayed in the list.
6.1.6.4.1 Backup
To backup the saved data in the IONSCAN 500DT, use the following procedure.
1. From the Administration menu, press the Data Management button to proceed.
The Data Management screen will display.
2. Press the Backup button. The screen will display all saved data.
3. Press the Begin Backup button to proceed. The message “Processing, please wait”
will appear, and the backup location and file name will display. When the backup
procedure is complete, the message “Backup complete” will display.
To export the saved data in the IONSCAN 500DT, use the following procedure.
1. From the Administration menu, press the Data Management button to proceed.
The Data Management screen will display.
2. Press the Select for Export button. The screen will display all saved samples and
parameters.
3. Either choose the Select All button to select all the saved samples, or scroll through
the list and select the samples individually.
4. Press the Select Params button. Either choose the Select All button to select all the
saved parameters, or scroll through the list and select the parameters individually.
5. Press the Back button to return to the Data Management screen, then select the
Export option.
6. Using the on-screen keyboard, select each field to type in the description, export
type, password, and confirmation of password.
If the data is to be deleted from the system after it has been exported, select the
“Delete exported samples” option.
7. Press the Export button. Insert a USB memory stick into the USB port on the front
of the IONSCAN 500DT at the prompt, then press the OK button to proceed.
Selecting the Report Management option enables a user with appropriate access to
produce a historical report.
A historical report enables specific sample result parameters for the report data to be
set.
2. Press the Historical button. The Historical Report options screen will display. Refer
to Figure 6-17.
3. Select the information to be included in the report, and the date range for the
report.
4. Press the Print button. A print progress pop-up will appear displaying the number
of pages being generated.
3. Using the on-screen keyboard, enter the new e-mail address information in each
field.
If the new email address is to be the default, select the Default option.
4. Press the Save button to save the new address and return to the E-Mail Settings
screen. The new e-mail address details will display in the list.
3. Select the field and modify the e-mail information. Press the Save button to save
the changes and return to the E-Mail Settings screen.
6.1.7 Audit
Selecting the Audit option enables an operator with the appropriate access level to
display, and then print or export a range of detector events in report format, by date,
and also identify the user performing the event and the user’s access levels.
To produce an audit report, use the following procedure.
1. From the Main menu, press the Audit button. The Audit screen will display a list
of all detector events. Refer to Figure 6-21.
NOTE: Review all event records by using the left and right navigation buttons at the
bottom of the screen. Jump to the first or last event record using the First
Record button or the Last Record button.
2. To set the date range for the audit report, press the down arrow to the right of the
Date From field. A calender will display where the start date for the audit report
can be selected, by month, day, and year. To clear any selected date, press the Clear
button.
3. Press the down arrow to the right of the Date To field. A calendar will display
where the end date for the audit report can be selected, by month, day, and year.
To clear any selected date, press the Clear button.
4. Press the down arrow to the right of the Users, Events, and Level fields to display
drop-down lists where the audit criteria can be selected.
5. Press the Print button to print the audit report, or press the Export button to export
the audit data to a USB memory stick.
6.1.8 Shutdown
For information on shutting down the IONSCAN 500DT, refer to “System Shutdown”
on page 5-18.
Chapter 7 Maintenance
This chapter describes the daily checks and preventive maintenance required to keep
the IONSCAN 500DT in optimum running condition.
Smiths Detection offers a comprehensive preventive maintenance plan which includes
quarterly inspection visits. For information about the plan, emergency service issues,
and troubleshooting and repair information please contact a Smiths Detection Service
Engineer.
CAUTION!
Routine preventive maintenance procedures must be performed regularly to keep
the IONSCAN 500DT in good working condition. Failure to perform routine
maintenance procedures may cause severe malfunctions and repairs, which may not
be covered by the warranty.
The following table outlines Smiths Detections recommended daily checks and
preventive maintenance tasks and schedule.
A maintenance kit is included with the IONSCAN 500DT and contains the following
items:
Re-Order
Item Description Quantity
Part Number
19055 Tweezers 1
WARNING!
Radiation Hazard - The IONSCAN Detector contains two sealed
nickel-60 (63Ni) foil radioactive sources of 555MBq (15 mCi) activity,
as explained in “Radiation Hazard” on page 1-2. UNDER NO
CIRCUMSTANCES should an attempt be made to open or clean the
IMS detector or drift Tube.
In the event of fire, explosion, malfunction, damage or suspected
damage, contact a Smiths Detection Service Engineer immediately.
WARNING!
Burn Hazard - The IMS drift tube operates at high temperatures.
NEVER service these areas until the IMS drift tube has cooled to room
temperature. Turn the unit off and it to cool down before performing
service or maintenance duties.
WARNING!
Risk of Electrical Shock - Observe the same precautions as for any
electrically-powered appliance. Check for frayed cables or broken
connectors. NEVER operate the IONSCAN 500DT on wet surfaces.
NEVER attempt service of any components until the main power
supply has been disconnected.
To shut down the system to perform certain maintenance functions, refer to “System
Shutdown” on page 5-18.
Selecting the Maintenance button from the Main Menu opens the Maintenance Menu,
containing Clean Cycle, Bakeout, Status, Anvil Cycle, Auto Calibration, Run
Background, Cycle Count, Diagnostics and Maintenance History. Refer to Figure 7-1.
2. Press the Start button to begin the clean cycle procedure. The Clean Cycle screen
will be open throughout the procedure.
NOTE: Selecting the Cancel button stops the clean cycle procedure; selecting the
Cancel button again, or before selecting Start exits the bakeout screen and
returns the operator to the Maintenance Menu screen.
7.3.2 Bakeout
During bakeout the drift tubes, inlet temperatures, and standby drift flow reach pre-
programmed bakeout values for a set period of time, and afterwards automatically
return to their normal standby values.
Bakeout is used to vapourize any contaminants that have accumulated on the IMS
tubes over time. The requirement for bakeout depends on the number, frequency and
cleanliness of samples analyzed.
The values for temperatures, flow, duration of the bakeout and the calibrant
monitoring parameters after the bakeout are determined in the control parameters
options for each tube. Refer to “Common Parameters” on page 6-6.
To perform a bakeout, use the following procedure.
1. From the Maintenance menu, press the Bakeout button to proceed. The Bakeout
screen displays. Refer Figure 7-3.
NOTE: On the Bakeout screen, a notification indicates the length of time required
to perform the bakeout. All operator input and menu selection is
unavailable for the duration of the bakeout, and sampling cannot be
performed during this time.
2. On the Bakeout screen, select the Start button to begin the bakeout process.
NOTE: Selecting the Cancel button stops the bakeout procedure; selecting the
Cancel button again, or before selecting Start exits the bakeout screen and
returns the operator to the Maintenance screen.
7.3.3 Status
Selecting this option enables the user to view the overall detector status, and the status
of each drift tube individually.
2. Select the Cycle Continuously option to have the anvil cycle continuously, select
the Go To Top option to have the anvil cycle once to the top, select the Go To
Bottom option to have the anvil cycle once to the bottom, or select the Go To Idle
option to idle the anvil.
NOTE: If the Cycle Continuously option is selected, the button will change to the
Stop Cycling option.
Autocalibration should rarely be necessary, and should only be performed when the
delta values of detected substances are very high or if, following a previous
autocalibration, the display shows ACAL: YES/REDO.
Autocalibration automatically applies corrections for minor changes in control
parameters. These corrections remain valid until the machine is either turned off or
placed into standby mode. Autocalibration can only be performed when the 500DT is
in a ready state.
To perform an autocalibration, use the following procedure.
1. From the Maintenance Menu, select the Auto Calibrate option to proceed. The
Autocalibration screen displays.
2. Analyze a clean swab to ensure the IONSCAN 500DT is free of contamination. A
“Pass” message must displayed after analysis prior to proceeding with this
instruction.
CAUTION!
When running a verification standard, use caution to ensure the IONSCAN 500DT,
sampling wand, work area and the operator do not become contaminated.
3. Wearing clean powder-free latex gloves, lightly apply the verification standard to
the sample area of the clean swab.
4. Squeeze the trigger to release the shaper from the sampling frame.
5. Insert the sampling frame into the sample inlet of the IONSCAN 500DT with the
swab facing upwards. Analysis will begin automatically. Immediately disengage
the arm from the sampling wand as soon as analysis begins.
Sampling
Frame
Magnet
Sampling Arm
6. The Autocalibration screen displays the verific results of the verification analysis.
The result will be either a pass or fail. Refer to Figure 7-9.
It is highly recommended that a clean chemical- and dust-free cloth be used to wipe
down the outside of the IONSCAN 500DT to remove excess dust, dirt and moisture
that may have accumulated. Never use commercial cleansers this highly sensitive
detection system.
Smiths Detection recommends the use of clean, dry Kimwipes® to remove excess dust,
dirt and moisture.
Condenser tubes contain charcoal and a foam sponge. When the foam sponge in the
centre of the tube begins to discolor, the condenser tube needs replacement. Both
condenser tubes can be viewed through the two windows located on the top of the
detector.
WARNING!
Risk of electrical shock - This procedure must be performed with the
power off. Refer to “System Shutdown” on page 5-18 to turn off all
power to the IONSCAN 500DT.
2. Check the main power supply to the IONSCAN 500DT is in the off position.
3. Remove the top and front right covers of the IONSCAN 500DT.
NOTE: Gently lift the four corners of the top cover starting from the back, then
remove the right front cover.
4. Unscrew the rear fastener of the condenser tube, then pull the condenser tube up
and out of the retaining bracket. Refer to Figure 7-13.
Gaskets
Condenser
Tube
5. Unscrew the brass fitting from the front of the condenser tube.
6. Replace the gasket on the retaining bracket and brass fitting with new gaskets.
7. Discard the used condenser tube and hand-tighten a new one to the brass fitting.
8. Reinstall the condenser tube in the retaining bracket and reassemble the unit.
Dust Filter
WARNING!
Risk of electrical shock - This procedure must be performed with the
power off. Refer to “System Shutdown” on page 5-18 to turn off all
power to the IONSCAN 500DT.
2. Check the main power supply to the IONSCAN 500DT is in the off position.
3. Remove the dust filter from the unit.
4. Remove the foam filter from the outer frame.
5. Rinse the dust filter in clean water. Dry thoroughly using a clean lint-free cloth.
6. Reassemble and replace the dust filter.
WARNING!
Risk of electrical shock - This procedure must be performed with the
power off. Refer to “System Shutdown” on page 5-18 to turn off all
power to the IONSCAN 500DT.
Burn Hazard - The IMS Detector uses very high temperatures during
operation. Allow the IONSCAN 500DT to return to room temperature
prior to performing this procedure. Never operate the IONSCAN
500DT without the inlet cover securely in place.
2. Check the main power supply to the IONSCAN 500DT is in the off position, and
the IONSCAN 500DT has cooled to room temperature.
3. Remove the top and front right covers of the IONSCAN 500DT.
NOTE: Gently lift the four corners of the top cover starting from the back, then
remove the right front cover.
4. Wearing powder-free latex gloves, dampen a swab with methanol or iso propanol
and wipe the inlet area (the bottom part of the inlet assembly where the sample
enters the system). Refer to Figure 7-15.
Use the following procedure to clean the IMS detector inlet liner junction.
1. Assemble the following replacement parts and tools:
• Powder-Free Latex Gloves
• Tweezers
• Clean, Warm Water
• Unscented Detergent
• Methanol or Iso Propanol
• Clean Swabs
WARNING!
Risk of electrical shock - This procedure must be performed with the
power off. Refer to “System Shutdown” on page 5-18 to turn off all
power to the IONSCAN 500DT.
Burn Hazard - The IMS Detector uses very high temperatures during
operation. Allow the IONSCAN 500DT to return to room temperature
prior to performing this procedure. Never operate the IONSCAN
500DT without the inlet cover securely in place.
2. Check the main power supply to the IONSCAN 500DT is in the off position, and
the IONSCAN 500DT has cooled to room temperature.
3. Remove the top and front right covers of the IONSCAN 500DT.
NOTE: Gently lift the four corners of the top cover starting from the back, then
remove the right front cover.
4. Wearing powder-free latex gloves, loosen the captive screw on the inlet door and
swing the door to the right to open. Refer to Figure 7-16.
Captive Screw
Inlet Door
5. Using tweezers, carefully remove the inlet liner junction. Refer to Figure 7-17.
Revision A, February 2005 7–17 Cleaning the IMS Junction Inlet Liner
IONSCAN 500DT
Inlet
Liner
Junction
CAUTION!
Under no circumstances should any laboratory solvents, particularly halogenated
solvents such as carbon tetrachloride, chloroform, or methylene chloride, be
introduced into the cell as they will damage the detector.
6. Check the condition of the inlet liner junction. If the inlet liner junction is dirty,
clean using warm water and unscented detergent. Rinse thoroughly and set aside
to dry.
7. Insert a clean inlet liner junction in the inlet area. Close the inlet door and hand-
tighten the captive screw.
NOTE: Do not over-tighten the captive screw.
Radiation leak testing must be carried out at regular intervals in accordance with the
rules proscribed by the local nuclear regulatory agency. The laboratory that performs
this service will provide a kit with instructions and cotton swabs for taking samples.
Samples are obtained from two locations; the IMS inlet and the IONSCAN 500DT
exhaust, as outlined in “IMS Inlet Radiation Leak Test” on page 7-19 and “IONSCAN
500DT Exhaust Radiation Leak Test” on page 7-21. Use a separate swab for each
sample location.
WARNING!
Radiation Hazard - The IONSCAN detector, which includes the IMS
inlet, radiation source and IMS drift tube, must never be opened by
untrained personnel.
Use the following procedure to perform a radiation leak test on the IMS inlet.
WARNING!
Risk of electrical shock - This procedure must be performed with the
power off. Refer to “System Shutdown” on page 5-18 to turn off all
power to the IONSCAN 500DT.
Burn Hazard - The IMS Detector uses very high temperatures during
operation. Allow the IONSCAN 500DT to return to room temperature
prior to performing this procedure. Never operate the IONSCAN
500DT without the inlet cover securely in place.
1. Check the main power supply to the IONSCAN 500DT is in the off position, and
the IONSCAN 500DT has cooled to room temperature.
2. Remove the top and front right covers of the IONSCAN 500DT.
NOTE: Gently lift the four corners of the top cover starting from the back, then
remove the right front cover.
3. Thoroughly swab the underside of the round inlet area. Refer to Figure 7-18.
WARNING!
Radiation Hazard - The IONSCAN detector, which includes the IMS
inlet, radiation source and IMS drift tube, must never be opened by
untrained personnel.
Use the following procedure to perform a radiation leak test on the IONSCAN 500DT
Exhaust.
WARNING!
Risk of electrical shock - This procedure must be performed with the
power off. Refer to “System Shutdown” on page 5-18 to turn off all
power to the IONSCAN 500DT.
Burn Hazard - The IMS Detector uses very high temperatures during
operation. Allow the IONSCAN 500DT to return to room temperature
prior to performing this procedure. Never operate the IONSCAN
500DT without the inlet cover securely in place.
1. Check the main power supply to the IONSCAN 500DT is in the off position, and
the IONSCAN 500DT has cooled to room temperature.
2. Remove the top cover of the IONSCAN 500DT.
NOTE: Gently lift the four corners of the top cover starting from the back.
3. Pull the condenser tube up and out of the retaining bracket.
Condenser
Tube
Condenser
Tube
4. Unscrew the brass fitting from the front of the condenser tube.
5. Using a clean swab, wipe the fitting.
Explosive Substances
SUBSTANCE DESCRIPTION
DNT n/a
SUBSTANCE DESCRIPTION
Narcotic Substances
SUBSTANCE DESCRIPTION
Barbiturates n/a
PART 31
Authority: Secs. 81, 161, 183, 68 Stat. 935,948,954, as amended (42 U.S.C.
2111,2201,2233); secs. 201, as amended, 202,88 Stat. 1242, as amended, 1244 (42 U.S.C.
5841,5842). Section 31.6 also issued under sec. 274,73 Stat. 688 (42 U.S.C. 2021).
disposed of.
(5) Shall immediately suspend operation of the device if there is a failure of, or
damage to, or any indication of a possible failure of or damage to, the shielding of the
radioactive material or the on-off mechanism or indicator, or upon the detection of 185
bequerel (0.005 microcurie) or more removable radioactive material. The device may
not be operated until it has been repaired by the manufacturer or other person holding
a specific license to repair such devices that was issued under parts 30 and 32 of this
chapter or by an Agreement State. The device and any radioactive material from the
device may only be disposed of by transfer to a person authorized by a specific license
to receive the by-product material in the device or as otherwise approved by the
Commission. A report containing a brief description of the event and the remedial
action taken; and, in the case of detection of 0.005 microcurie or more removable
radioactive material or failure of or damage to a source likely to result in
contamination of the premises or the environs, a plan for ensuring that the premises
and environs are acceptable for unrestricted use, must be furnished to the Director of
Nuclear Material Safety and Safeguards, ATTN: GLTS, U.S. Nuclear Regulatory
Commission, Washington, DC 20555-0001 within 30 days. Under these circumstances,
the criteria set out in Sec. 20.1402, ʺRadiological criteria for unrestricted use,ʺ may be
applicable, as determined by the Commission on a case-by-case basis;
(6) Shall not abandon the device containing by-product material;
(7) Shall not export the device containing by-product material except in accordance
with part 110 of this chapter;
(8)(i) Shall transfer or dispose of the device containing by-product material only by
export as provided by paragraph (c)(7) of this section, by transfer to another general
licensee as authorized in paragraph (c)(9) of this section, or to a person authorized to
receive the device by a specific license issued under parts 30 and 32 of this chapter, or
part 30 of this chapter that authorizes waste collection, or equivalent regulations of an
Agreement State, or as otherwise approved under paragraph (c)(8)(iii) of this section.
(ii) Shall furnish a report to the Director of Nuclear Material Safety and Safeguards,
ATTN: GLTS, U.S. Nuclear Regulatory Commission, Washington, DC 20555-0001
within 30 days after the transfer of a device to a specific licensee or export. The report
must contain--
(A) The identification of the device by manufacturer’s (or initial transferor’s) name,
model number, and serial number;
(B) The name, address, and license number of the person receiving the device (license
number not applicable if exported); and
(C) The date of the transfer.
(iii) Shall obtain written NRC approval before transferring the device to any other
specific licensee not specifically identified in paragraph (c)(8)(i) of this section.
(9) Shall transfer the device to another general licensee only if--
(i) The device remains in use at a particular location. In this case, the transferor shall
give the transferee a copy of this section, a copy of Secs. 31.2, 30.51, 20.2201, and
20.2202 of this chapter, and any safety documents identified in the label of the device.
Within 30 days of the transfer, the transferor shall report to the Director of Nuclear
Material Safety and Safeguards, ATTN: GLTS, U.S. Nuclear Regulatory Commission,
Washington, DC 20555-0001--
(A) The manufacturer's (or initial transferor's) name;
(B) The model number and the serial number of the device transferred;
(C) The transferee's name and mailing address for the location of use; and
(D) The name, title, and phone number of the responsible individual identified by the transferee
in accordance with paragraph (c)(12) of this section to have knowledge of and authority to take
actions to ensure compliance with the appropriate regulations and requirements; or
(ii) The device is held in storage by an intermediate person in the original shipping container at
its intended location of use prior to initial use by a general licensee.
(10) Shall comply with the provisions of §§20.2201, and 20.2202 of this chapter for reporting
radiation incidents, theft or loss of licensed material, but shall be exempt from the other
requirements of parts 19,20, and 21, of this chapter.
(11) Shall respond to written requests from the Nuclear Regulatory Commission to provide
information relating to the general license within 30 calendar days of the date of the request, or
other time specified in the request. If the general licensee cannot provide the requested
information within the allotted time, it shall, within that same time period, request a longer
period to supply the information by submitting a letter to the Director, Office of Nuclear
Material Safety and Safeguards, U.S. Nuclear Regulatory Commission, Washington, DC
20555-0001 and provide written justification as to why it cannot comply.
(12) Shall appoint an individual responsible for having knowledge of the appropriate
regulations and requirements and the authority for taking required actions to comply with
appropriate regulations and requirements. The general licensee, through this individual, shall
ensure the day-to-day compliance with appropriate regulations and requirements.This
appointment does not relieve the general licensee of any of its responsibility in this regard.
(13)(i) Shall register, in accordance with paragraphs (c)(13)(ii) and (iii) of this section, devices
containing at least 370 MBq (10 mCi) of cesium-137, 3.7 MBq (0.1 mCi) of strontium-90, 37
MBq (1 mCi) of cobalt-60, or 37 MBq (1 mCi) of americium-241 or any other transuranic (i.e.,
element with atomic number greater than uranium (92), based on the activity indicated on the
label. Each address for a location of use, as described under paragraph (c)(13)(iii)(D) of this
section, represents a separate general licensee and requires a separate registration and fee.
(ii) If in possession of a device meeting the criteria of paragraph (c)(13)(i) of this section, shall
register these devices annually with the Commission and shall pay the fee required by Sec.
170.31 of this chapter. Registration must be done by verifying, correcting, and/or adding to the
information provided in a request for registration received from the Commission. The
registration information must be submitted to the NRC within 30 days of the date of the request
for registration or as otherwise indicated in the request. In addition, a general licensee holding
devices meeting the criteria of paragraph (c)(13)(i) of this section is subject to the bankruptcy
notification requirement in Sec. 30.34(h) of this chapter.
(iii) In registering devices, the general licensee shall furnish the following information and any
other information specifically requested by the Commission--
(A) Name and mailing address of the general licensee.
(B) Information about each device: the manufacturer (or initial transferor), model number,
serial number, the radioisotope and activity (as indicated on the label).
(C) Name, title, and telephone number of the responsible person designated as a representative
of the general licensee under paragraph (c)(12) of this section.
(D) Address or location at which the device(s) are used and/or stored. For portable devices, the
address of the primary place of storage.
(E) Certification by the responsible representative of the general licensee that the information
concerning the device(s) has been verified through a physical inventory and checking of label
information.
(F) Certification by the responsible representative of the general licensee that they are aware of
the requirements of the general license.
(iv) Persons generally licensed by an Agreement State with respect to devices meeting the
criteria in paragraph (c)(13)(i) of this section are not subject to registration requirements if the
devices are used in areas subject to NRC jurisdiction for a period less than 180 days in any
calendar year. The Commission will not request registration information from such licensees.
(14) Shall report changes to the mailing address for the location of use (including change in
name of general licensee) to the Director of Nuclear Material Safety and Safeguards, ATTN:
GLTS, U.S. Nuclear Regulatory Commission, Washington, DC 20555-0001 within 30 days of
the effective date of the change. For a portable device, a report of address change is only
required for a change in the device's primary place of storage.
(15) May not hold devices that are not in use for longer than 2 years. If devices with shutters
are not being used, the shutter must be locked in the closed position. The testing required by
paragraph (c)(2) of this section need not be performed during the period of storage only.
However, when devices are put back into service or transferred to another person, and have not
been tested within the required test interval, they must be tested for leakage before use or
transfer and the shutter tested before use. Devices kept in standby for future use are excluded
from the two-year time limit if the general licensee performs quarterly physical inventories of
these devices while they are in standby.
(d) The general license in paragraph (a) of this section does not authorize the manufacture or
import of devices containing by-product material.
2Persons possessing by-product material in devices under a general license in Sec. 31.5
before January 15, 1975, may continue to possess, use, or transfer that material in
accordance with the labeling requirements of Sec. 31.5 in effect on January 14,1975.
[30 FR 8189, June 26, 1965, as amended at 33 FR 6463, Apr. 27, 1968; 38 FR 22220, Aug.
17, 1973; 42 FR 28896, June 6, 1977; 43 FR 6922, Feb. 17, 1978; 56 FR 23471, May 21,
1991; 56 FR 61352, Dec. 3, 1991; 58 FR 67659, Dec. 22,1993]
1Sources generally licensed under this section prior to January 19, 1975 may bear labels
authorized by the regulations in effect on January 1, 1975.
31.11 General license for use of by-product material for certain in vitro clini-
cal or laboratory testing.
(a) A general license is hereby issued to any physician, veterinarian in the practice of veterinary
medicine, clinical laboratory or hospital to receive, acquire, possess, transfer, or use, for any of
the following stated tests, in accordance with the provisions of paragraphs (b), (c), (d), (e), and
(f) of this section, the following by-product materials in prepackaged units:
(1) Iodine-125, in units not exceeding 10 microcuries each for use in in vitro clinical or
laboratory tests not involving internal or external administration of by-product material, or the
radiation therefrom, to human beings or animals.
(2) Iodine-131, in units not exceeding 10 microcuries each for use in in vitro clinical or
laboratory tests not involving internal or external administration of by-product material, or the
radiation therefrom, to human beings or animals.
(3) Carbon-14, in units not exceeding 10 microcuries each for use in in vitro clinical or
laboratory tests not involving internal or external administration of by-product material, or the
radiation therefrom, to human beings or animals.
(4) Hydrogen-3 (tritium), in units not exceeding 50 microcuries each for use in in vitro clinical
(Name of Manufacturer)
(e) The registrant possessing or using by-product materials under the general license of
paragraph (a) of this section shall report in writing to the Director of Nuclear Material Safety
and Safeguards, any changes in the information furnished by him in the "Registration
Certificate --In Vitro Testing With By-product Material Under General License". Form NRC -
483. The report shall be furnished within 30 days after the effective date of such change.
(f) Any person using by-product material pursuant to the general license of paragraph (a) of this
section is exempt from the requirements of parts 19, 20, and 21, of this chapter with respect to
by-product materials covered by that general license, except that such persons using the Mock
Iodine-125 described in paragraph (a)(7) of this section shall comply with the provisions of
§§20.2001, 20.2201, and 20.2202.
[33 FR 16553, Nov. 14, 1968, as amended at 38 FR 1271, Jan. II, 1973; 38 FR 34110, Dec. II,
1973; 39 FR 26147, July 17, 1974; 40 FR 8785, Mar. 3, 1975; 41 FR 16446, Apr. 19, 1976; 42
FR 21604, Apr. 28, 1977; 42 FR 26987, May 26, 1977; 42 FR 28896, June 6, 1977; 44 FR
50325, Aug. 28, 1979; 51 FR 36967, act. 16, 1986; 56 FR 23471, May 21, 1991; 56 FR 61352,
Dec. 3, 1991; 58 FR 67659, Dec. 22, 1993]
1
Labels authorized by the regulations in effect on September 26, 1979, may be used until one
year from September 27, 1979.
31.13 Violations.
(a) The Commission may obtain an injunction or other court order to prevent a violation of the
provisions of --
(1) The Atomic Energy Act of 1954, as amended;
(2) Title II of the Energy Reorganization Act of 1974, as amended; or
(3) A regulation or order issued pursuant to those Acts.
(b) The Commission may obtain a court order for the payment of a civil penalty imposed under
section 234 of the Atomic Energy Act:
(1) For violations of --
(i) Sections 53, 57, 62, 63, 81, 82, 101, 103, 104, 107, or 109 of the Atomic Energy Act of
1954, as amended;
(ii) Section 206 of the Energy Reorganization Act;
(iii) Any rule, regulation, or order issued pursuant to the sections specified in paragraph
(b)(1)(i) of this section;
(iv) Any term, condition, or limitation of any license issued under the sections specified in
paragraph (b)(1)(i) of this section.
(2) For any violation for which a license may be revoked under section 186 of the Atomic
Energy Act of 1954, as amended.
[57 FR 55072, Nov. 24, 1992]
Smiths Detection
Technical Documentation Department
1730 Aimco Blvd.
Mississauga, Ontario
Canada, L4W 1V1
Fax: 905-238-3018
Customer
To help us continually improve the quality of this manual, we appreciate your input.
Please answer the following questions, and fax or mail the completed questionnaire
back to the address above.
4. Would you like to see instructions for any additional procedures? Please list them
in the space below.
Please use the space on the back of this page to add any additional comments you may
have about this manual.