Chemistry Unit-5 Material
Chemistry Unit-5 Material
Radio waves: A radio basically captures radio waves that are transmitted by radio stations. Radio
waves can also be emitted by gases and stars in space. Radio waves are mainly used for TV/mobile
communication.
Microwave: This type of radiation is found in microwaves and helps in cooking at home/office. It
is also used by astronomers to determine and understand the structure of nearby galaxies and stars.
Infrared: It is used widely in night vision goggles. These devices can read and capture the infrared
light emitted by our skin and objects with heat. In space, infrared light helps to map interstellar
dust.
X-ray: X-rays can be used in many instances. For example, a doctor can use an X-ray machine to
take an image of our bones or teeth. Airport security personnel use it to see through and check
bags. X-rays are also given out by hot gases in the universe.
Gamma-ray: It has a wide application in the medical field. Gamma-ray imaging is used to see
inside our bodies. Interestingly, the universe is the biggest gamma-ray generator of all.
Ultraviolet: The Sun is the main source of ultraviolet radiation. It causes skin tanning and burns.
Hot materials that are in space also emit UV radiation.
Visible: Visible light can be detected by our eyes. Light bulbs, stars, etc., emit visible light.
Absorption of radiation:
When ground state electrons or lower energy state electrons (E1) absorbs sufficient energy from
photons, they jump into the next higher energy level or higher energy state (E 2). In other words,
when the ground state electrons absorb energy which is equal to the energy difference between the
two energy states (E2 – E1), the electrons jumps from ground state (E1) to the excited state or
higher energy level (E2). The electrons in the higher energy level are called excited electrons.
The light or photons energy applied to excite the electrons can be mathematically written as
hv = E2 – E1
V = Frequency of photon
It can also be defined as the intensity of transmitted light decreases exponentially as the
concentration of the substance and distance travelled through the substance increases.
Applications of Beer-Lambert law:
UV-VIS Spectroscopy:
When the interaction between incident radiation and the electron cloud in a chromophore results in
an electronic transition involving the promotion of one or more of the outer shell or the bonding
electrons from a ground state into a higher energy state, ultraviolet-visible (UV-Vis) spectra are
derived.
Generally, the UV and visible spectral bands of substances are large and may not exhibit a high
degree of compound recognition accuracy. Nonetheless, they are sufficient for quantitative assays
and are useful as an alternate means of detection for several substances.
Electronic Transitions:
Electronic transition is that when the energy from UV or visible light (higher energy radiation in
the UV (200-400 nm) and visible (400-700 nm)) is absorbed by a molecule, one of its electrons
jumps from a lower energy to a higher energy molecular orbital.
σ → σ* Transition:
Example: Molecular hydrogen, H2, the molecular orbital picture consists of one
bonding σ MO (HOMO), and a higher energy antibonding σ* MO (LUMO).
π -π* Transition:
Where UV-vis spectroscopy becomes useful to most organic and biological chemists is in
the study of molecules with conjugated π systems.
In these groups, the energy gap for π -π* transitions is smaller than for isolated double
bonds, and thus the wavelength absorbed is longer.
Molecules or parts of molecules that absorb light strongly in the UV-vis region are
called chromophores.
Example: MO diagram for 1,3-butadiene, the simplest conjugated system contains two
bonding orbitals, and two antibonding orbitals.
n - π* Transition:
The conjugated pi system in 4-methyl-3-penten-2-one gives rise to a strong UV absorbance at 236
nm due to a π - π* transition. However, this molecule also absorbs at 314 nm. This second
absorbance is due to the transition of a non-bonding (lone pair) electron on the oxygen up to a π*
antibonding MO
This is referred to as an n - π* transition. The nonbonding (n) MO’s are higher in energy than the
highest bonding p orbitals, so the energy gap for an n→π∗ transition is smaller than that of a π - π*
transition – and thus the n - π* peak is at a longer wavelength. In general, n - π* transitions are
weaker (less light absorbed) than those due to π - π* transitions.
Light Source: UV-Visible spectrophotometers have a light source that emits a broad spectrum of
light covering the ultraviolet and visible regions. Tungsten filament lamps are commonly used for
the visible region (350-800 nm), while deuterium lamps or xenon arc lamps are used for the
ultraviolet region (190-350 nm).
Sample Compartment: The sample compartment holds the sample that is being analyzed. It
typically consists of a cuvette transparent to the wavelength or a sample holder through which light
passes.
Detector: The detector measures the intensity of light that passes through the sample.
Photomultiplier tubes (PMTs) or charge-coupled devices (CCDs) are commonly used detectors in
UV-Visible spectrophotometers. The detector converts the light intensity into an electrical signal,
which is then used to calculate the absorbance of the sample.
Amplifier and Signal Processor: The electrical signal generated by the detector is often weak,
and an amplifier is used to strengthen it.
Wavelength Selector/Controller: The wavelength selector or controller allows the user to choose
the desired wavelength for the analysis. It controls the monochromator to select specific
wavelengths of light.
Applications of UV-VIS Spectroscopy:
Identification of Functional Groups: UV-Visible spectroscopy can provide information about the
electronic transitions within a molecule, allowing for the identification of functional groups.
Kinetic Studies and Reaction Monitoring: is employed in kinetic studies to monitor the progress
of chemical reactions over time, by measuring the changes in absorbance during a reaction.
Quality Control in Pharmaceuticals: widely used in the pharmaceutical industry for quality
control purposes. It is employed to determine the concentration of active pharmaceutical
ingredients (APIs) in formulations.
Environmental Analysis: applied in environmental science for the analysis of water, air, and soil
samples to detect and quantify pollutants, such as heavy metals, organic compounds, and other
environmental contaminants.
IR Spectroscopy:
IR radiations lie in the wavelength range of 0.7 to 400 µm. IR Spectroscopy is based upon the
selective absorption of IR radiations by the molecule which induces vibration of the molecules of
the compound
Principle of Infrared Spectroscopy:
Infrared (IR) radiation lies in the electromagnetic spectrum between visible light and
microwave radiation.
The technique is based on the interaction of infrared radiation with chemical bonds in
molecules, causing them to vibrate.
Different bonds absorb specific frequencies of IR radiation, resulting in characteristic
absorption patterns that can be used for identification and analysis.
Fundamental modes:
There are two types of fundamental molecular vibrations viz., stretching (change in bond length)
and bending (change in bond angle).
Stretching vibrations: It involves a rhythmic movement along a bond axis with a subsequent
increase and decrease in bond length.
Bending vibrations: A change in the angle occurring between two bonds is known as a bending
vibration. Four bending vibrations exist namely, wagging, twisting, rocking and scissoring.
Selection rules:
A molecule will absorb Infrared radiation if the change in vibrational states is associated with a
change in dipole moment of the molecule
Vibrations which do not change the dipole moment are Infrared inactive.
Homonuclear diatomic molecules are IR inactive. Examples: O2, N2, H2, Cl2
Infrared spectrometers require a suitable source of infrared radiation, which is typically provided
by a heated filament or a globar, a silicon carbide rod. These sources emit a broad range of infrared
wavelengths, covering the mid-infrared (MIR) and near-infrared (NIR) regions.
Sample Compartment:
It may consist of a sample holder, such as a transparent window or an attenuated total reflection
(ATR) accessory, depending on the sample type and measurement technique.
Interferometer:
The interferometer splits the radiation into two beams that travel different paths before
recombining.
Optical Components:
Various mirrors and beam splitters are used within the spectrometer to manipulate and direct the
infrared radiation.
Beam splitters divide the radiation into reference and sample beams, while mirrors ensure proper
alignment and optical efficiency.
Detector:
Detectors include thermal detectors (such as pyroelectric or thermopile detectors) and photo
detectors (such as photodiodes or photoconductive detectors).
Different detectors offer varying levels of sensitivity, speed, and spectral range.
Spectral analysis software is used to analyze and interpret the obtained spectrum, enabling
identification of functional groups and compounds.
Principles of Chromatography:
Gas Chromatography (GC): In GC, the mobile phase is a gas, and separation is based on the
different volatilities and partition coefficients of components.
Liquid Chromatography (LC): In LC, the mobile phase is a liquid, allowing separation based on
factors such as polarity, size, or ion exchange properties.
Thin-Layer Chromatography (TLC): TLC employs a thin layer of adsorbent coated on a solid
support, enabling rapid and convenient separations.
High-performance liquid chromatography (HPLC) is a powerful analytical technique used for the
separation, identification, and quantification of components in a liquid mixture.
Principle:
HPLC Pump: Provides a continuous flow of the mobile phase at a constant pressure.
Injector: Introduces the sample into the mobile phase stream, typically using an injection
valve.
Chromatographic Column: Contains the stationary phase, typically packed with small
particles for efficient separation.
Detector: Measures the concentration of analytes exiting the column and generates a signal.
Eg. UV-Visible (UV-Vis) Detector, Electrochemical Detector:
Data System: Collects and processes the detector signal, displaying and analyzing
chromatographic data.
Modes of Operation:
Normal Phase HPLC: The stationary phase is polar, and the mobile phase is nonpolar. Eg. Silica
hydride, Diol: Diol-based stationary phases contain hydroxyl (-OH) functional groups, Silica
gel.
Reverse Phase HPLC: The stationary phase is non polar, and the mobile phase is polar. Eg:
Phenyl Phases, Bi phenyl phase.
Applications of HPLC: