BioTester User Manual
BioTester User Manual
User Manual
version 6.5
Mechanical measurement and analysis of biomedical materials
CellScale provides scientific and medical researchers with turn-key systems for
measuring the mechanical properties of biomaterials.
Our focus is soft tissue – including skin, ligaments, blood vessels, heart valves,
sclera, membranes and scaffolds. We provide user-friendly software, an easy-to-
use patented attachment system, and effective data analysis tools.
Our foundation was laid at one of the world’s leading research institutions – The
University of Waterloo. We understand research and aim to provide effective
solutions at a reasonable price.
Explore our web site www.cellscale.com or contact us to learn more about our
measurement systems.
© 2014 CellScale. All rights reserved. This material may not be reproduced,
displayed, modified or distributed without the express prior written permission of the
copyright holder. For permission, contact CellScale Biomaterials Testing at
[email protected].
1. General Information
The BioTester is a precision test instrument specifically designed for the measurement and
analysis of small biological specimens. The system includes a compact biaxial test station and an
integrated software interface to run and analyze test results. The system is intended for indoor
use only.
System Assembly
The unit requires some assembly. See Appendix A and C.
Power Connections
Connect power supply into properly grounded 100-240VAC power source to ensure safe
operation. Ensure that the power cord is easily accessible at all times. The use of an
Uninterruptible Power Supply (UPS) is recommended to protect against data loss.
The mains supply voltage fluctuations should not exceed 10% of the nominal supply voltage.
Safety Warnings
This equipment must be used in accordance with the procedures outlined in this manual to
prevent injury and/or damage.
System Alert
The plastic covers must remain on this unit at all times to prevent injury and/or damage.
This equipment must not be disassembled by the user or modified in any way.
1
BioTester User Manual
There are two switches on the front of the unit, one for powering the control board and the LED
lighting, and the second for powering the heating element.
The front of the unit also contains a pull rod which is used to raise and lower the specimen stage
and fluid chamber. This rod is spring-loaded so that the chamber can be held at any height. To
raise or lower the stage, lift the rod and then pull or push.
General Maintenance
Clean the system as needed with mild soap and water or alcohol based cleaning solutions.
2
BioTester User Manual
2. Testing Terminology
The BioTester is designed to apply biaxial forces to soft tissue specimens (deforms by more than
1% in vivo) with an in-plane dimension of 3 - 15 mm. This includes biological material such as eye
tissue, heart valves, pericardium, joint capsules, large blood vessels, scaffolds, and polymers.
3
BioTester User Manual
4
BioTester User Manual
Control Modes
There are two control modes which define the basic approach to a given test: displacement
control and force control:
Under displacement control, the displacement of the specimen is predefined. The
BioTester stretches the specimen until the predefined displacement is achieved. The
force required to achieve the displacement is an output of the test.
Under force control, the force applied to the specimen is predefined. The BioTester
stretches the specimen until the predefined force is achieved. The displacement required
to achieve the force is an output of the test.
5
BioTester User Manual
Control Functions
The BioTester makes it possible to test specimens under several control functions:
Under displacement control:
The true strain function applies the displacement at a constant true strain rate, which
accounts for the current specimen length while the specimen is being stretched.
The ramp function applies the displacement at a constant nominal rate. This is
equivalent to a constant engineering strain or constant velocity.
The sine function applies the displacement according to a sinusoid with the desired
displacement magnitude and duration.
The custom load functions allow users to customize the rate and extent of
displacement application by specifying a table of time versus displacement values.
Displacement
Time
Force
Time
Step
The control mode, load function and load magnitude can be set independently for each of the two
loading directions.
6
BioTester User Manual
3. System Overview
System Components
The BioTester brings together several high-precision components, making it the state-of-the-art
instrument for testing biomaterials.
The BioTester
7
BioTester User Manual
Software Overview
The software included with the BioTester is called LabJoy. It is divided into two modules, a data
collection module and a review and analysis module. The data collection module is used to set
test parameters, enable specimen loading and testing, and monitor test progress. The screen
layout for this module is shown below:
Test
Parameters
Image List
Image Playback
Image Tracking
Data Overlay
8
BioTester User Manual
Each test folder will also contain 2 subfolders. The “Logs” subfolder contains a text file of the
content of the text dialogue portion of the main screen. It is useful as reference to troubleshoot
problems with your system, should they occur.
The “Tracking” subfolder is initially empty. If tracking is performed on any of the images in the
test folder using the “Analyze and Review” software module, there will be data files that contain
the tracking information stored in this directory.
While using the “Analyze and Review” software module, you may create additional data files such
as text files (*.csv) containing tracking results, images with force data or tracking results overlaid
(*.wmf), or movie files (*.avi). These additional files can be stored in the test directory or
elsewhere on your computer’s hard drive without interfering with the software application (once
created, they cannot be opened by the LabJoy application).
When working with the data, you should be aware of a few details regarding the data output:
1. The output specimen sizes are based on the spacing between the BioRake tines and do
not account for the specimen material outside of the test region.
2. The software cannot calculate stresses since the thickness of the specimen is not known.
To calculate stress, you will have to manually measure the thickness of the material
before or after the test is performed.
3. Strains can be calculated using the output displacement values (which are based on the
BioRake tine motions). The specimen may actually be subjected to less strain than the
calculated values due to attachment site effects and tissue tearing. The image tracking
module is useful for determining the actual strain values and variations within the
specimen.
9
BioTester User Manual
1. Select a template that matches the type of test you wish to perform. See the BioTester
Tip on how to select and use a template. You can modify the template parameters in
Step 4, below.
2. Name your test. The dialog will have a default test name. You can use the default or
rename the test. Each time you start a new test, the default name will continue to
increment the number at the end of the name.
3. If desired, you can change the location of the output data. The location of your output
data and images is determined by specifying a Test Name and Data Directory. The
template and data directories are user specific (computer login name). Each system user
can store their files to a different default location. The system uses the last location
selected as the default.
10
BioTester User Manual
4.
To reset the actuators, select Reset Actuator from the tools menu, or press on the toolbar.
System Alert
You should not reset the actuators if the specimen is already loaded. Doing so will
damage your specimen as well as possibly damage the BioRakes or Load Cells.
11
BioTester User Manual
To move the actuators to the specified position, select Move Actuators to Specified Size from the
Tools menu, or press on the toolbar.
To zero the load cells, select Zero Load Cells from the tools menu or press on the toolbar.
System Alert
You should not zero the load cells if the specimen is already loaded, or the load cell
zeroing process will not work properly and an offset introduced into your force data.
12
BioTester User Manual
13
BioTester User Manual
14
BioTester User Manual
press on the tool bar, or from the Tools menu select Move Actuators to
Specified Size
use the Actuator Control jog button arrows.
15
BioTester User Manual
5. Additional Settings
Configuring Output Data Files
From the Settings menu, select Data Output to display the dialogue shown on the left. Click the
Configure button to select which of the following columns to output.
16
BioTester User Manual
Kx and Ky (Default = 100, integer values) are proportional gain tuning parameters (one for
each of the coordinate axes). They control the sensitivity or the amount of corrective action
that is applied to the difference between the current and target force values. At low gain
values the system may be safe and stable but may be sluggish in response to changing
conditions. If the proportional gain is increased the system becomes more responsive and
loads are achieved more quickly. If Kx and Ky are set too high, an unstable oscillating
system may result and forces may never be achieved. The optimal value for these settings
depends on the properties of the material being testing. For stiff tissues values may range
between 1 and 10. For soft tissues values may range between 50 and 200.Kx and Ky usually
are set equal to each other unless testing materials are strongly anisotropic.
Tolerance (mN) (Default = 3) controls how close to the specified force the actual load values
need to be. For example, a specified force of 100 mN with a Tolerance of 3 mN will be
considered achieved when the loads are between 97 and 103 mN. Keep in mind the
tolerance of the load cell being used when choosing the value for this parameter. The load
cells typically fluctuate by about 0.1% of their full scale during the course of a test. For a
1000mN load cell this is about 1mN but for a 23N load cell, this is about 23mN. This value
should usually be 0.2-0.5% of full scale to ensure stable operation.
Settling Frames (Default = 5) controls how many successive load samples (sampled at 100
Hz) need to be within the Tolerance criteria for the force to be considered achieved. For
example, a specified force of 100 mN with a Tolerance of 3 mN and Settling Frames = 5 will
be considered achieved when the loads are between 97 and 103 mN for 5 successive data
samples.
Velocity (1-10 integer values) matches actuator jog speed. Typically for harder tissues, this
value should be between 4 and 6 and for soft tissues, this value should be between 6 and 10.
17
BioTester User Manual
Acceleration (1-10 integer values). Typically this should be set to the default value of 10
(maximum acceleration). Changes made to the acceleration will affect tests under both force
and displacement control. Reducing this value may help smooth force controlled tests under
low loads or displacement controlled tests at high speeds. Tests run with lower accelerations
will have softer transitions at direction or speed changes but may take longer than specified
to achieve testing cycles.
Soft Limits
To avoid rake collisions or sample destruction, it is useful to assign reasonable travel limits to
the actuators. This is especially important during force control or preloading, where if
specified loads cannot be reached, the actuators will keep moving until they reach the soft
limits.
Min and Max Displacement, in µm, set the minimum and maximum positions to which the
rakes are allowed to move.
Min and Max Force, in mN; this feature is currently not functional. Appropriate limits should
be set using Min and Max Displacement.
18
BioTester User Manual
19
BioTester User Manual
Graph Data
Force versus Time
This graph shows how the measured X and Y forces
are changing with time. Peak loads per cycle,
differences in load between the X and Y axes, and
force relaxation are all easily seen in this type of graph.
Force is proportional to nominal (engineering) stress.
20
BioTester User Manual
Image List
Image Playback
Image Tracking
Data Overlay
21
BioTester User Manual
Selecting Images
The Images are shown in a tree structure on the top left panel.
The tree structure organizes the images by:
Set (line on the test parameter window)
Cycle (iteration of a given Set)
Phase (Preload/Stretch/Hold/Recovery/Rest of each Cycle)
Image (the one or more individual images from each Phase)
You can select which images to include in a playback set in
one of several ways:
hold down the shift key and click on individual images
select entire sets, cycles, or phases (will playback all
contained images)
select images which have data associated with a
particular tracking set
Next/Prev Display next or previous image (as dictated by “Play Every” parameter).
Play/Stop Start and stop the playback.
Prev/Next Cycle Jumps to a corresponding image and phase in the next or previous cycle. Useful when reviewing
sets that have multiple cycles.
Loop Sequences can be played continuously in a loop or only once.
Playback Rate +/- changes the speed of the playback. A playback rate can also be manually entered in the
display box.
Play Every Allows the users to skip some frames to expedite playback.
22
BioTester User Manual
23
BioTester User Manual
24
BioTester User Manual
Figure 1: Initial tracking grid displayed as Figure 2: Tracking to 20% strain displayed as
points. points.
25
BioTester User Manual
26
BioTester User Manual
Figure 4: Tracking to 20% strain displayed as Figure 5: Tracking to 20% strain with text
strain ellipses. showing x% strain.
27
BioTester User Manual
Data Overlay
Data associated with each image is displayed below
the tracking controls on the left side of the main
window. Checking the Show Graph box displays this
data as a graph overlaid on the images
28
BioTester User Manual
29
BioTester User Manual
30
BioTester User Manual
Camera
The camera uses a ½” monochrome CCD sensor to
obtain a resolution of 1280 pixels by 960 pixels. The
camera shutter can be controlled during the setup
phase of the test.
The camera captures images at 15 frames per second
during every test and writes these images to a buffer.
The test controller extracts images from this buffer at
the frequency specified in the Set Parameter Editor
(15Hz, 5Hz, 1Hz, 0.1Hz, 0.01Hz, or 0.001Hz). These
images are written to the output directory. The file
name is the ideal capture time. The actual capture time
may differ from the ideal capture time by as much as
1/30 of a second since the images are extracted from a
buffer.
The test software creates an image capture log file in the output directory should this difference
be significant. This log stores the actual capture time and the idealized capture time.
The camera default shutter and gain are specific to each test template and can be set by
selecting Hardware from the Settings menu. Typically the gain is best left at 20 and the shutter
changed to alter the brightness of the image.
31
BioTester User Manual
Load Cells
The load cells are semiconductor strain gage-based
with mechanical overload protection. They have
accuracy equal to 0.2% of the rated full scale load. The
load cells are available in 0.5N, 1N, 2.5N, 5N, 10N and
23N sizes. Each load cell has a unique calibration
factor which is stored in a chip attached to the load cell
connector.
While the load cells should not have significant
hysteresis or thermal drift, temperature changes or
prolonged loading can introduce small offsets. It is
recommended that the zero load state be redefined
periodically. This can be done by selecting Zero Load
Cells from the Tools menu or pressing .
The load cells are sampled at 100Hz, with hardware sample averaging to reduce noise in the
data. Further sample averaging can be implemented by selecting Advanced from the Settings
menu.
Each load cell is calibrated and has a specific calibration factor which is stored on a chip in the
load cell connector. The range and calibration factor can be viewed by selecting Hardware from
the Settings menu, but are not user controlled. In addition, the load cells need to be calibrated
when the system is initially set up and whenever load cells are changed or re-mounted. See
Section 8 for this procedure.
32
BioTester User Manual
Temperature
To activate the fluid chamber heater, both power
switches on the front of the BioTester need to be in the
On position (the computer and software do not need to
be on yet). At this point, the heater will warm the fluid to
the last used temperature set point. The temperature
set point can be changed by selecting Hardware from
the Settings menu.
Typically, the heater will take 20 to 40 minutes to warm
a full fluid chamber from room temperature to 37°C.
When Show Temperature Warnings is enabled and the
Current Temperature is not at the Temperature Set
Point plus or minus the specified tolerance -- a red
indicator button will be displayed next to the Current
Temperature live output and a warning dialogue will
appear when you execute the test.
System Alert
In the event of a dislodged or faulty temperature sensor, or if there is a lack of fluid in
the fluid chamber, the system is designed to reach a maximum temperature of 80°C.
Caution: Avoid touching the specimen stage when the heater is on, it can become
very hot!
33
BioTester User Manual
34
BioTester User Manual
If the load cells are changed, moved, or impacted, it is good practice to perform the load cell
calibration routine. This calibration can be performed in the following steps:
1. Launch the calibration utility by selecting Tools, Advanced, then Load Cell Calibration
2. Input the K value (spring constant) for the calibration spring provided for the load cell
being used (shown on the spring case label, provided upon system delivery)
3. The load cell capacity should be automatically read and displayed, and the appropriate
calibration preload calculated
4. Lower the fluid chamber and install the calibration spring brackets.
5. Zero the load cell using the Zero Load Cells button
6. Using the Jog+ and Jog– buttons, drive the actuators until the posts are correctly spaced
so that the calibration spring can be installed without applying any significant load. Set
the calibration spring on the posts (see above figure)
7. Using the Jog+ and Jog– buttons, move the actuators until the spring is carrying a
tension that is out of the non-linear “dead zone” and reaches the appropriate preload
displayed in the Calibration Preload dialogue. For most setups, the spring should be
35
BioTester User Manual
driven to 30% of the load cell rating ie. for a 2.5N load cell, a preload of 750mN is
specified. If it is suspected that the load cell is more than 15% out of calibration, contact
the manufacturer for another recommendation for calibration pre-load.
8. Visually verify that the spring is loaded in a stable, horizontal position.
9. Click the Run button, wait for calibration to complete.
10. The number (A) that is displayed in the main dialogue box
at the end of the calibration procedure is the ratio of the
current calibration value to the previous calibration value.
If there has been no change to the load cell, A should be
between 0.99 and 1.01.
11. If the A value is greater than 1.05 or less than 0.95, the
load cells were significantly out of calibration. If this is the
case, the calibration procedure above should be repeated
to ensure that the preload was correct and an accurate
calibration achieved.
36
BioTester User Manual
The camera position can be adjusted using the knobs at the back and side of the camera
housing. If larger position adjustments are needed, the bolts that attach the camera to the mast
can be loosened to allow rotation of the camera housing.
The BioTester comes equipped with a 75mm focal length lens and a set of lens spacers. Using
these products it is possible to achieve a field of view ranging from 9mm to 30mm. The details of
the appropriate number of spacers and the required standoff are shown in the table below. The
image magnification can be changed by adding or removing spacer tubes located between the
camera body and the lens. Increasing the total spacer tube length will result in higher
magnification while reducing the tube length will result in lower magnification. Changing the tube
length will also change the working distance of the lens. To adjust the vertical position of the lens
and camera assembly, loosen the bolts that clamp the camera housing to the mast and raise or
lower the camera housing until the image is in focus. It is helpful to make sure the lens focus ring
is in the middle of its rotation so that fine tuning the image focus can be done once the camera
housing is re-clamped.
INCREASING THE TOTAL LENGTH OF THE SPACER TUBES WILL INCREASE
MAGNIFICATION (DECREASE FIELD OF VIEW) AND DECREASE WORKING DISTANCE
(STAND-OFF).
If the camera position or magnification is changed, it is important to capture a single image scale
shot. A glass calibration target or a ruler is acceptable for this purpose. The scale shot will be
useful later to characterize the size of test specimens or test specimen features.
As the table and graph also show, it is also possible to swap in a lens with a different focal length
to achieve a different range of magnifications. This data was collected using the family of fixed
focal length lenses available from Edmund Optics (the part numbers are shown in the table). The
camera has a C-mount connection, making it possible to mount a wide variety of optical
components to best suit any application. Currently, only the standard camera is supported.
25mm (NT56-529) 35mm (NT56-530) 50mm (NT56-531) 75mm (NT56-532) 100mm (NT56-675)
Stand Stand Stand Stand Stand
Spacer off pix/c FOV off pix/c FOV off pix/c FOV off pix/c FOV off pix/c FOV
(mm) (mm) m (mm) (mm) m (mm) (mm) m (mm) (mm) m (mm) (mm) m (mm)
0 685 78.6 162.8 1070 71 179.3 930 121 105.9
5 100 520 24.6 215 373 34.3 370 329 38.9 820 206 62.2 1540 147 86.8
10 53 935 13.7 125 682 18.8 235 560 22.9 450 404 31.7 950 258 49.6
15 90 993 12.9 185 779 16.4 350 534 24.0 675 366 34.9
20 73 1273 10.1 160 978 13.1 285 711 18.0 535 473 27.0
25 135 1186 10.8 245 859 14.9 455 578 22.1
30 125 1410 9.1 220 959 13.3 400 687 18.6
35 115 1630 7.9 200 1111 11.5 355 797 16.1
40 105 1842 6.9 185 1260 10.2 320 897 14.3
45 100 2068 6.2 170 1392 9.2 295 1003 12.8
37
BioTester User Manual
600
500
400
Standoff (mm)
25mm
300 35mm
50mm
200 75mm
100mm
100
0
0 20 40 60 80 100 120 140
Field of View (mm)
NOTE: The physical camera standoff will not go below 100mm or exceed 500mm, the preceding
table and chart provides guidelines only
38
BioTester User Manual
Update Firmware
1. With the BioTester connected to the PC and turned on, launch the firmware update
software located in the Windows start menu under LabJoy>Utilities.
2. Load the firmware file using the “Load Update File” button.
3. Click the “Connect” button.
4. Execute the update using the “Update” button and wait for update to complete. This
can take many minutes.
39
BioTester User Manual
9. Troubleshooting
40
BioTester User Manual
The BioTester has been packed in a special container to protect it during shipping. In addition,
system accessories have been packed in several cartons contained in a separate box. The
system components listed below may be unpacked and laid out before system setup. A space
24” x 24” with 40” overhead clearance is required for setup.
Tools
1 BioTester
16 6-32 X 1/2” Socket Head Cap Screws
2 6-32 Hex Nuts
12 5mm Magnets
12 7.5mm Magnets
20 BioRakes
1 Calibration Spring
2 Calibration Spring Brackets
1 Set of Camera Lens Spacer Tubes
1 Specimen Mounting Bridge
2 Load Cell Goosenecks
2 No-Load Cell Goosenecks
6 6-32 Hex Nuts (2 extra)
1 Standard Specimen Cutting Jig
4 Razor Blades for Cutting Jig
1 Fluid Bath Chamber
1 Power Cord and Supply
1 6’ (1.8m) USB Cable
1 Set of Spacer Plates
1 Sheet of Rubber Backing Material
2 Load Cells
41
BioTester User Manual
Unpack System
Remove the BioTester from the shipping container. Carefully remove the boxes and protective
shipping material from the container such that the BioTester can be lifted out. This will require 2
people. Take special care with the camera mast as it is connected to the main unit by cables. In
addition, protective foam will have to be removed from the main unit and camera mast.
42
BioTester User Manual
Thread the mast into the receiver. The knurled collar must be rotated to fasten the camera mast.
It will take many turns but eventually the mast will lock into place. Try to position the camera
approximately as shown, although the final position can be adjusted later in the procedure. Be
sure to reduce the slack on the cables before threading the post, so as not to pinch the cables
while tightening the collar.
43
BioTester User Manual
Check the orientation by attempting to place a BioRake in the gooseneck. If the magnets are
correct, then the BioRake will snap in to place. If the BioRake is repelled by the magnet, then
they have been installed backwards.
44
BioTester User Manual
Align the temperature probe mounting hole with temperature probe cable.
Remove protective caps on the lens and camera body and then screw the lens into the camera
body. Spacer tubes are placed between the lens and the camera body to change the image
scale. This also affects the working distance of the lens and therefore the camera housing may
need to be moved up or down on the camera mast to accommodate changes in the number of
spacer tubes.
45
BioTester User Manual
A Windows based PC with at least 2 USB 2.0 ports is required to operate the BioTester. Included
in your shipment is a USB key which contains the LabJoy software. These files contain the
software installer for LabJoy, which is the user interface and control software for the BioTester.
To install the software, simply insert the USB key, open the flash drive, and click on “setup.exe”.
Details of typical dialogue boxes are shown below.
46
BioTester User Manual
47
BioTester User Manual
48
BioTester User Manual
49
BioTester User Manual
50
BioTester User Manual
51
BioTester User Manual
Connect USB cables to the PC USB 2.0 ports and then proceed
with the following steps.
Ensure the correct orientation of the load cells. Tighten the nuts gently to avoid damaging the
load cells (just a little bit tighter than finger tight). Keep in mind that the load cells are sensitive to
excessive torque, excessive overload, transverse loading, and bending forces. Ensure the load
cell cable angle is approximately as shown (see following 4 images).
52
BioTester User Manual
53
BioTester User Manual
See the following 4 images. The nut on the backside of the load cell is not as sensitive as the
front side nut. Make sure that the gooseneck is rotated such that it is level when fastening the
backside nut.
54
BioTester User Manual
55
BioTester User Manual
After connecting the load cells, tuck the cables inside the plastic cover, taking care to handle the
connectors by the edges. A circuit board assembly may be attached to the connector and must
be handled carefully. Position the cables inside the cover so that they do not interfere with the
actuator carriage travel, particularly during actuator reset where the carriages are fully retracted.
56
BioTester User Manual
Launch the LabJoy software. The default screen should appear as:
The details of the software are outlined in this user’s manual. For assembly purposes, it is only
required that we initiate the default test. Select OK.
57
BioTester User Manual
Next place a BioRake in each of the 4 goosenecks so that the BioTester looks like:
58
BioTester User Manual
On the LabJoy control screen, ensure that the specimen size is set to 5000µm in X and Y and
then click the “Move to Size” button.
59
BioTester User Manual
The camera position has both coarse and fine position adjustments and coarse and fine focus
adjustments. To start, make sure the lens fine focus knob (lowest ring on lens) is approximately
in the middle of its rotation. Next, loosen the bolts that clamp the camera housing to the mast.
The camera assembly can then be rotated until the camera is positioned approximately over the
centre of the 4 BioRakes. As well as rotating, the camera assembly can slide up and down to
bring the BioRakes into approximate focus. Be careful to move the camera housing gently in
small increments so as not to crash it into the goosenecks or BioRakes. Attempt to achieve the
setup approximately as shown below:
Next, fine tune the camera position using the knobs at the back and side of the camera housing
and fine tune the image focus using the lowest ring on the lens.
Adjustments to the image scale can be made by changing the overall height of the spacer tubes
between the lens and the camera. See section on “Adjusting the Camera Position and Image
Magnification” (page 37).
60
BioTester User Manual
Because the goosenecks attached to the load cells are connected by only 1 threaded stud, the
vertical and transverse alignment of the end of the gooseneck can be somewhat different each
time the goosenecks are removed and reattached. To accommodate the vertical variation, shims
between the gooseneck brackets and the actuators may need to be added or removed. Thick
and thin shims are provided (3 thins shims = 1 thick shim). At this stage, do not be concerned
with the transverse or axial positioning of the goosenecks as these will be aligned in subsequent
steps.
61
BioTester User Manual
By looking at the BioRakes from the side, it is possible to observe their vertical alignment. Please
add or remove shims to vertically align the BioRakes. Here are pictures of poor vertical alignment
and good vertical alignment:
62
BioTester User Manual
This step requires 2 different adjustment mechanisms. The goal is to make the BioRakes look
like this:
First, make sure the BioRakes are positioned in the transverse centre of the gooseneck. The
magnetic mounts that hold the BioRakes to the gooseneck fix them in the axial direction, but
allow a small amount of adjustment in the transverse direction. This is useful for alignment during
regular testing, but for the purposes of alignment they should be in the neutral position. The first
image below shows a transversely offset BioRake and the second shows a transversely centered
BioRake.
63
BioTester User Manual
To transversely adjust the goosenecks, loosen the bolts that attach the goosenecks to the
actuators, slightly rotate the gooseneck, and retighten the bolts. The on-screen image should be
helpful as a guide.
To axially adjust the goosenecks, use the actuator jog icons in the lower left of the screen. Make
sure the “independent” radio button is selected and not the “mirror matching” button. The jog
speed should be about 7.
Adjust all 4 goosenecks until you are satisfied with their transverse and axial alignment. It may
also be necessary to make small adjustments to the camera position.
It is now necessary to perform a centre position calibration. First, it will be necessary to
temporarily define the current position as the centre position. This calibration can be performed
in the following steps:
1. Lower the fluid chamber and remove the BioRakes.
2. From the Tools menu select Advanced and then select Centre Position Calibration. This
should bring up this dialogue box:
3. Press YES.
6. Press YES. Once the actuators have stopped moving, insert 2 opposing BioRakes on
the X axis.
7. Using the actuator jog control arrows at the lower left of the screen, move the actuators
until all 10 BioRake tips are collinear. Be sure to jog the BioRakes in MIRROR
MATCHING MODE ONLY!!! Note that it is the tips of the BioRakes that should be
collinear and not the outside edge. Because the tips are underneath the BioRakes and
cannot be seen on screen, it is necessary to estimate their position. Please use the
image below as a reference:
64
BioTester User Manual
Example 3: BioRakes do not have any overlap so tips are not collinear.
65
BioTester User Manual
12. Select the reset actuators button (or use >Tools>Reset Actuators).
To check the centre position calibration, use the >Tools>Advanced>Move to Centre command
while having only 2 opposing BioRakes mounted. Compare the position of the rakes to the image
in step 4 above.
Each actuator cover has a number written on the inside. These numbers match up with numbers
written on the inside of the main cover to ensure the best possible fit between the actuator covers
and the main cover.
The best way to install the covers is to place the back 2 pegs into the main cover and then lower
the front 2 pegs while squeezing them together. Be careful not to let the cover slip and
accidentally apply a transverse load to the gooseneck.
Finally, perform load cell calibration (see Section 8 under Load Cell
Calibration).
66
BioTester User Manual
There are many ways to attach specimens to the suture brackets using a variety of suture
designs. This document outlines one way to use double-armed sutures to accomplish specimen
attachment.
1. Assemble equipment: small vice, magnifying goggles, 2 needle nose pliers, small forceps.
67
BioTester User Manual
68
BioTester User Manual
3. Tie off sutures. First hook each bent hook around a secured post held in the vice. Then
tie a standard knot in the suture and use forceps to guide the location of the knot so
that it tightens at the correct location.
69
BioTester User Manual
70
BioTester User Manual
71
BioTester User Manual
72
BioTester User Manual
The resulting mounted specimen should ideally look something like this:
73