FB03102

BILIMETER 3D
Bilihandy®

Specifications
Range: 0,00 – 99,99 mg/dl (Total Bilirubin)
0 - 1690 µmol (Total Bilirubin)
Light Source: LED
Light Receptacle: Photodiode
Measurement Volume: ~ 4 µl
Measurement time: ~ 1 second
Display: 2,4“ TFT – 320x240 RGB
Filters: ~445nm / ~560nm
Instrument Power: 5V --- ; 5A
5W, Standby: <1W
Printer: Thermoprinter
Dimensions: 245 x 130 x 105 (mm)
Weight: ~850 g

Exclusive Distributor UK

Unit 20 Howard Way


PFAFF MEDICAL GMBH
Cromwell Business Centre - Interchange Park
DR.-KARL-LEXER-WEG 299
Newport Pagnell – Buckinghamshire MK16 9QS
86633 NEUBURG DONAU - GERMANY
PHONE 01908 612598
TEL. +49 (0) 8431 - 41669
FAX 01908 217974
[email protected]
[email protected]
WWW.BILIMETER.DE
www.djblabcare.co.uk
FB03102

O p e r a t i n g i n s t r u c t i o n s

Turn on Insert empty and clean sample holder into the


power switch
instrument. The notch on the yellow handle
must face upwards.

Please wait until display shows “0.0” mg/dl (or


“0” µmol). Instrument is now ready for use.

Quality control: Please follow the instructions provided by your government and
/ or your hospital’s Clinical Chemistry department.

Insert Sample Remove sample holder from the instrument.


Insert patient sample into the holder. The serum
part must cover the measuring slit completely.

Insert Sample Insert the sample holder into the instrument.


Holder
The indicated value is flashing during the
measurement. The measurement is finished
when the result is shown permanently. The
result is printable. To continue with another
measurement proceed as described.

RE-
RE-ZERO Please perform a Re-Zero when the instrument
(Standby)
has not been switched off for a longer period.
Be sure the sample holder is inserted without
any capillary tube. Push the RE-ZERO button. You
can continue measuring Bilirubin when the
display shows “0.0” mg/dl or “0” µmol.

Print Print out the result if needed (if your device is


equipped with an optional printer).

Correctness in mg/dl Linearity in mg/dl Precision in mg/dl Hemoglobincompensation


in mg/dl
25 35
A B C Standard Value Readings
20
30
1 19.1 11.4 0.8 1 20.5 20.7
2 19.1 11.4 0.8 2 10.4 10.5
Reference Value

25

15
20
3 19.1 11.4 0.8 3 5.5 5.6
RefG

4 19.1 11.4 0.8 4 2.8 2.9


10 15
5 19.1 11.4 0.8
10 250mg/dl Hemoglobin
5
MEAN 19.1 11.4 0.8
5
SD 0.0 0.0 0.0
0
0 5 10 15 20 25 0
0 5 10 15 20 25 30 35
CV (%) 0.0 0.0 0.0
Bilimeter Bilimeter Readings IVD