AAnalyst 800 Detection Limits-nickel Data
AAnalyst 800 Detection Limits-nickel Data
technical note
Mean: 0.0002
SD: 0.00030
Instrument Conditions
3 x SD x mo 3 x 0.0003 x 17.5 pg
Lamp HCL IDL = =
0.0044 0.0044
Wavelength 232.0 nm
= 3.6 pg
Slit 0.2 nm
Mean: 0.0246
SD: 0.0008
RSD: 3.4%
Comments
• Solution detection limit for 50-µL sample volume = 0.07 µg L-1.
• A pyrolysis temperature of up to 1100°C can be used for most samples.
• Full Stabilized Temperature Platform Furnace (STPF) performance achieved
for a refractory element such as Ni. Platform atomization at 2300°C showed
virtually no tailing.