Chapter 1
Chapter 1
CONTROL
Chapter 1 – Introduction
This chapter will introduce you to instrumentation, the various measurement units used, and the
reason why process control relies extensively on instrumentation. It will help you become
familiar with instrument terminology and standards.
This ensures that the industrial machines run smoothly and safely in
factories and efficiently use energy to transform raw materials into
high-quality finished products with reliable consistency while
reducing energy waste and economic costs, something which could
not be achieved purely by human manual control.
Process control is the automatic control of an output variable by
sensing the amplitude of the output parameter from the process and
comparing it to the desired or set level and feeding an error signal
back to control an input variable.
Block Diagram of a process control loop
Definitions of the Elements in a Control Loop
Transducers are devices that can change one form of energy to another, e.g., a
resistance thermometer converts temperature into electrical resistance, or a
thermocouple converts temperature into voltage. Both of these devices give an
output that is proportional to the temperature.
Converters are devices that are used to change the format of a signal without
changing the energy form, i.e., a change from a voltage to a current signal.
Actuators are devices that are used to control an input variable in response to a
signal from a controller. Atypical actuator will be a flow-control valve that can
control the rate of flow of a fluid in proportion to the amplitude of an electrical
signal from the controller. Other types of actuators are magnetic relays that
turn electrical power on and off. Examples are actuators that control power to
the fans and compressor in an air-conditioning system in response to signals
from the room temperature sensors.
Controllers are devices that monitor signals from transducers and take the
necessary action to keep the process within specified limits according to a
predefined program by activating and controlling the necessary actuators.
Programmable logic controllers (PLC) are used in process-control
applications, and are microprocessor-based systems.
An Error signal is the difference between the set point and the
amplitude of the measured variable.
An electrical supply is required for all control systems and must meet all
standards in force at the plant. Power failure can mean plant shutdown and the
loss of complete production runs.
Water supply is required in many cleaning and cooling operations, and for
steam generation. Domestic water supplies contain large quantities of
particulates and impurities, and may be satisfactory for cooling, but are not
suitable for most cleaning operations.
The span of an instrument is its range from the minimum to maximum scale
value, i.e., a thermometer whose scale goes from −40°C to 100°C has a span of
140°C. When the accuracy is expressed as the percentage of span, it is the
deviation from true expressed as a percentage of the span.
Reading accuracy is the deviation from true at the point the reading is being
taken and is expressed as a percentage, i.e., if a deviation of ±4.35 psi in
Example 1.6 was measured at 28.5 psi, the reading accuracy would be
(4.35/28.5) × 100 = ±15.26% of reading.
The absolute accuracy of an instrument is the deviation from true as a number not as a
percentage, i.e., if a voltmeter has an absolute accuracy of ±3 V in the100-volt range, the
deviation is ±3 V at all the scale readings, e.g., 10 ± 3 V, 70 ± 3 V and so on.
Precision refers to the limits within which a signal can be read and may be somewhat
subjective. In the analog instrument shown in Fig. 1.6a, the scale is graduated in divisions of
0.2 psi, the position of the needle could be estimated to within 0.02 psi, and hence, the
precision of the instrument is 0.02 psi. With a digital scale the last digit may change in steps of
0.01 psi so that the precision is 0.01 psi.
Reproducibility is the ability of an instrument to repeatedly read the same signal over time,
and give the same output under the same conditions. An instrument may not be accurate but
can have good reproducibility, i.e., an instrument could read 20 psi as having a range from17.5
to 17.6 psi over 20 readings.
Sensitivity is a measure of the change in the output of an instrument for a change in the
measured variable, and is known as the transfer function, i.e., when the output of a pressure
transducer changes by 3.2 mV for a change in pressure of 1 psi, the sensitivity is 3.2 mV/psi.
High sensitivity in an instrument
is preferred as this gives higher output amplitudes, but this may have to be weighted against
linearity, range, and accuracy.
Linearity is a measure of the proportionality between the actual value of a variable being
measured and the output of the instrument over its operating range.