NEONATAL INTENSIVE CARE INCUBATOR INC 100

ROBUST • RELIABLE • USER-FRIENDLY

Total Maternal and Infant Care


STURDY construction NEONATAL INTENSIVE CARE INCUBATOR INC 100

STABLE environment
The INC 100. A must for
every well run hospital.

The INC 100 has been 3


specially designed to provide the safest and 1
2
most stable environment. And, with its wide
range of accessories it can be tailored to meet
4
any situation in the care of the critical neonate.
5

Its excellent nursing facilities, safe, reliable


operation, easy cleaning and sturdy
6
construction makes it a must for every well
run, busy hospital.

FEATURES:
1 Elbow operated ports

Allows immediate access and eliminates


cross infection

2Large front
Provides direct access for major procedures
and easy removal of baby Self supporting tube and lead apertures
(made from autoclavable silicone)
Externally operated 5 trendle and fowler Easy and safe to insert tube and cords
positioning
(Maximum 8º on both sides) Microprocessor based 6 control panel
with unique blower and heater
Easy to pull out 4 baby tray Easy-to-read and touch operated
With X-Ray cassette compartment

Canopy with unique hinges


3
The R&D department at would be
Easy to clean inside the incubator happy to provide you a solution if your practice has any
special requirement.

S ERVICE SU
CK P
PO
O
D THE CL

RT :
UN

Your nearest distributor:


O

Phoenix Medical Systems Private Limited


An ISO 9001:2000 Company (with Design)
No.32/4, Jawaharlal Nehru Salai, Ekkattuthangal, P.O. Box 3205. Chennai - 600 097, India.
Tel.:91-44-2225 1177, 2225 1178 Fax : 91-44-2225 0194.
E-mail : [email protected]
Website:www.phoenixmedicalsystems.com
NEONATAL INTENSIVE CARE INCUBATOR
Technical specifications
INC 100 AIR
MODE

INCUBATOR
CONTROL

TEMPERATURE

OVERRIDE
TEMPERATURE

PHOENIX
PATIENT

HEATER
OUTPUT
TEMPERATURE

100%
75%
50%
ALARM

25%
& MESSAGE

System
Failure

Power Failure

Care
Intensive
Neonatal Incubator

Model No. INC100


DIMENSIONS:
Width-70cm
Length-94.5cm
Height with Castors-136.5cm
Castors-4” diameter, 2locking and 2 non-locking

CANOPY/HOOD DIMENSIONS: PERFORMANCE SPECIFICATIONS:


Width-50.6cm Temperature rise time: 20-30 minimum per 10ºC
Length-80.7cm Temperature variability: 0.2ºC
Height-45cm Temperature overshoot: 0.5ºC
Air velocity over mattress: < 10cm sec
MATTRESS DIMENSIONS: Noise level within unit: < 60 dB
Width-40cm Bed Tilting: ± 8ºC maximum continuously variable on both
Length-60cm sides
Height-5cm
OPERATIONAL ENVIRONMENT:
Electrical Power Requirements: Operating temperature range-20ºC to 30ºC
Voltage rating - 230V Storage temperature range-25ºC to 60ºC
Power rating - 350w Operating humidity range-0 to 95%
Current rating - 3.1A
frequency - 50Hz OXYGEN: Oxygen inpu-Average concentration range
Leakage current -0.5mA 5 lettres minimum-25% to 45%
Power factor - 1.000 10 lettres minimum-35% to 60%
Electronics: Microprocessor based servo control system 15 lettres minimum-45% to 70%
(Self tests are performed periodically)
HUMIDITY: Water Capacity-1000ml
Humidification upto 50%-+10%
Mobility: 3 Swivel castors fixed on sleek - T-shaped base
Humidity control range-40 to 90 + 5%
(2 wheels with brakes)
Humidity Display resolution-+ 5%

TEMPERATURE MODULE : ALARMS: Patient Temperature (more than 0.5ºC difference)


Heater Element: Encapsulated heater with Aluminum fins Air Temperature (more than 0.5ºC difference)
Heater Output: 0% to 100% , 25% increment Temperature probe failure/Disconnect
Probe: Thermistor based interchangeable probe System failure, Heater failure, Air flow failure, Too much
Probe Calibration: Not Required weight, Power Fail
Patient Temperature Range: 32º C to 38º C
Air Temperature Range: 20º C to 38º C COATING:
Accuracy: ± 0.2º C Epoxy powder coated body for scratch protection and Rust
Resolution: 0.1º C prevention and polyurethane painted baby bed and air
Power: 220V to 250V at 50Hz 500W circulation path for easy cleaning and baby safety.
Temperature Display: Bright 1” numerical LED display
Over Temperature Protection: Automatic cutoff of heater at 39ºC

S ERVICE SU
CK P
PO
O
D THE CL

RT :

Your nearest distributor:


Phoenix Medical Systems Private Limited
UN
O

R
An ISO 9001:2000 Company (with Design)
No.32/4, Jawaharlal Nehru Salai, Ekkattuthangal, P.O. Box 3205. Chennai - 600 97, India.
Tel.:91-44-2225 1177, 2225 1178 Fax : 91-44-2225 0194.
E-mail : [email protected]
Website:www.phoenixmedicalsystems.com