Shah 1995
Shah 1995
Abstract
Keywords: Automation; Dissolution testing; Flow-injection analysis; Immediate release; Liquid chromatog-
raphy; Microdialysis sampling; Multivessel; Sustained release
"-
O 0 " I.oo.,
1 2 3 ..o*:o° ! " probes I I
~o I
.c o°n
6 5 4 ]
sion Release T M tablets was performed accord- dissolution system. Similar perfusion rate and
ing to USP XXII specifications, using the pad- injection intervals were used with the auto-
dle method. In an attempt to demonstrate two mated single-vessel dissolution system, but only
different analytical techniques, the ac- one tablet per dissolution run was studied and
etaminophen was determined by on-line LC the dissolution vessel was sampled every
and the phenylpropanolamine was determined 10min. Dissolution profiles of Accutrim '"~
by flow-injection analysis (FIA). tablets measured using various systems were
compared to the previously reported dissolu-
Acetaminophen tablets tion profile [3].
The dissolution of acetaminophen tablets
was studied in 900 ml of water at 37 °C and
50 rpm. A DL-5 microdialysis loop probe was
suspended in each dissolution vessel and per- 3. Results and discussion
fused continuously with water at a rate of
50 ~tl min-~. The dialysate was collected in a
2 ~tl on-line injection loop and injections were 3.1. Principle of operation
made every 96 s, followed by a blank injection
for 36 s. The latter injection was necessary to The STT E6 switching valve has two sets of
flush the system and eliminate carryover. Only inlets and outlets and six pairs of ports for
three tablets could be studied per dissolution sampling up to six dissolution vessels simulta-
run using the STT E6 valve owing to the rapid neously (see Fig. 1). The syringes mounted on
sampling intervals necessary for the adequate the dual syringe pump were filled with the
description of the dissolution profile of ac- perfusion medium and connected to the two
etaminophen which was released rapidly from inlets of the STT E6 valve. A DL-5 microdialy-
the tablets. Each vessel was sampled once every sis probe was suspended in each dissolution
6.5min. The samples were analyzed on-line vessel and both ends of the probe were con-
using a fast LC column with UV detection at nected to a pair of ports on the STT E6 valve.
280 nm. The mobile phase used to elute ac- At any instance, two probes were perfused ---
etaminophen consisted of methanol-phosphate one probe was flushed to waste and the other
buffer (50 mM, pH 6.2) (15:85, v/v) at a flow connected to the injector. An uneven injection
rate of 2 ml rain ~. The dissolution profiles time sequence was built into the detector
measured using the automated multivessel dis- method to sample and analyze the dissolution
solution testing system were compared to man- medium in an automated fashion. At the end
ual and automated single-vessel dissolution of the injection interval, the STT E6 valve,
testing systems. rotated counter-clockwise, and an on-line injec-
tor, which was an integral part of the LC
Accutrim ~ 16h Precision Release T M tablets system, were switched simultaneously through
The dissolution of Accutrim ~ 16 h Precision a relay on the system controller. The on-line
Release T M tablets was studied in 1 1 water at injector made an injection into the LC column,
37 °C and 100 rpm. Dissolution profiles were whereas the STT E6 valve moved to the next
measured by the automated multivessel disso- position. Fig. 1 shows port # 4 in the flush
lution system described here, by the single-ves- mode (dashed line) and port # 3 in the sample
sel dissolution system described in a previous mode (solid line). Once an injection had been
paper [1], and by manual sampling. Six tablets made and the STT E6 valve moved counter-
could be monitored per dissolution run using clockwise, port # 5 would be in the flush mode
the automated multivessel dissolution system. and port # 4 in the sample mode. The sam-
A DL-5 loop probe was suspended in each pling sequence was designed so that the stag-
dissolution vessel and perfused with water at nant probe was flushed prior to the actual
10 ~tl min J. The perfusate was collected in a sampling from that probe. The flush interval
2 ~tl injection loop and analyzed on-line using was designed to allow the previously stagnant
FIA with water at 1 ml min-~ as a carrier and probe to reach a steady-state condition neces-
UV detection at 215 nm. The six dissolution sary for reproducible sampling. Once the injec-
vessels were sampled sequentially every 10 min. tion sequence was initiated, the data and
Therefore, each dissolution vessel was sampled chromatograms were recorded and stored auto-
every hour using the automated multivessel matically in the system controller.
1238 K.P. Shah et al./ J. Pharm. Biomed. Anal. 13 (1995) 1235-1241
20 1 30 Probe-to-probe variations
30 l 21 Multivessel sampling using the STT E6 valve
30 2 l0 required use of one microdialysis probe per
30 3 5.7 vessel. Inter-probe variations were determined
60 1 14 using three DL-5 loop probes. The probes were
75 1 9.0
100 1 <1 immer-sed in an acetaminophen solution
10 10 <1 (350 lag ml -~) at 37 °C and stirred at 50 rpm to
simulate the conditions of the dissolution ex-
K.P. Shah et al./ J. Pharm. Biomed. Anal. 13 (1995) 1235-1241 1239
periments. Each probe was perfused with wa- probe, a slight different curve was obtained
ter at 50 lal min-1 and the dialysate was ana- with each probe. The range of the slopes was
lyzed to determine the drug recovery through approximately 15%, consistent with the inter-
the probe. The range for the recoveries of the probe variability described in the previous
three probes was approximately 15%. There- section. The linear calibration curves sug-
fore, it was necessary to calibrate each probe gested that the microdialysis recovery did not
individually. change with drug concentration in the range
The automated multivessel dissolution sys- studied.
tem was calibrated as follows. Before the dis- Similarly, the calibration curves of phenyl-
solution study, the drug recovery from each propanolamine were linear in the concentra-
microdialysis probe at the condition of the tion range of 10-100 lag m l - l . However,
experiment was determined using a standard below 10 lag ml ~ the response was not linear
solution. Next, dissolution studies were con- with concentration, presumably owing to ad-
ducted on a required number of tablets. The sorption of the drug onto the connecting tub-
experiment was concluded by repeating the ing and/or the dialysis membrane. The
same drug recovery experiment. Average peak average microdialysis recovery of drug from
heights of the standard solution before and aqueous solutions at 37 °C and 10 lal min ~ 1
after the dissolution experiments were used to perfusion rate was about 45% over the linear
quantify the results of the dissolution study range of the assay (10 100lagml l).
using a single-point calibration method. The repeatability of the multivessel auto-
mated system was assessed by injecting each
L#tearity and repeatability standard solution in triplicate during the cali-
The linearity of the integrated sampling
bration study. The RSD was less than 3%
and chromatographic system for ac-
for all the standards measured by manual
etaminophen was determined using three DL-
and microdialysis sampling, except for the
5 microdialysis loop probes and a STT E6
100lalml -~ standard measured by probe 2
multiposition valve. Standard solutions of ac-
where the RSD was 4.3%. In general, the
etaminophen in the concentration range 2 5 -
RSD value measured at each concentration
500 lag ml -~ were placed in the dissolution
by microdialysis sampling was lower than
vessel at 37 °C and stirred at 50 rpm (i.e. the
that by manual sampling.
conditions of the dissolution study). A probe
was immersed in each dissolution vessel and
perfused with water at 50 lal min -1 for on-
3.3. Dissolution of acetaminophen tablets
line sampling. The calibration curves of ac-
The dissolution profiles of acetaminophen
etaminophen were linear in the concentration
range studied, as shown in Fig. 3. Since mi- tablets obtained with an automated multives-
crodialysis recovery varied from probe to sel dissolution testing system compared well
with profiles measured by manual and auto-
mated single-vessel dissolution systems [1], as
4
shown in Fig. 4. The sampling interval with
the automated multivessel dissolution system
was limited by the dead volume of the STT
E6 switching valve. An uneven injection se-
2
quence consisting of a 96 s sampling mode
alternated by a 36 s flush mode was used to
avoid carry-over of the sample from the pre-
a. 1
vious injection. The flush time of 36 s was
determined by the retention time of ac-
0 etaminophen on the LC column. With the
0 100 200 300 400 500 600
present configuration, only three tablets per
Concentration (~g/mL)
dissolution run could be studied and each
Fig. 3. Calibration curves of acetaminophen determined by vessel was sampled about every 6.5min.
three different DL-5 loop probes simultaneously using the
Faster sampling rates could be achieved if a
automated multivessel dissolution system. Study condi-
tions: 37 °C and 50 rpm; perfusion rate, 50 ~tl rain-l; perfu- switching valve with a lower dead volume
sion medium, water. were available.
1240 K.P. Shah et al. / J. Pharm. Biomed. Anal. 13 (1995) 1235-1241