0% found this document useful (0 votes)
11 views

PBS Lab Practical Study Guide

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
11 views

PBS Lab Practical Study Guide

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 7

PBS Lab Practical Study Guide

Activity 1.1.1 A Sketchy Scene- Crime Scene Search Methods.

Activity 1.1.2 Reliable Witnesses- Polygraph analysis

Activity 1.1.3 Traces We Leave Behind- Fingerprint patterns &


Minutiae, Microscope parts and use

Activity 1.1.4 Blood Evidence- Blood Spatter Analysis, Blood type


AB,A,B,O

Activity 1.1.5 DNA Evidence- Micropipetting, PCR & Restriction


Enzymes

Activity 1.1.6 DNA Analysis- Label parts of DNA, Reading


Electrophoresis Gels

Activity 1.2.4 Histology- Tissue examples, Total magnification


Microscope

Activity 1.2.5 Gross Heart Anatomy- Anatomy of the heart


CRIME SCENE SEARCH METHODS
Control
Question

2
Parts of a Micropipette

Micropipettes are extremely helpful laboratory tools that are easy and fun to use with a little bit of instruction and practice.

Setting the Volume and Measuring Liquids

1. Choose the correct micropipette for the volume you are


measuring. Make sure that the volume to be measured does not
exceed the upper or lower volume setting of the micropipette.
2. To determine the units measured by the micropipette, look at the
volume setting. The setting shows in the window on the side of
the micropipette.
3. To set the volume, twist the top of the pipetting button, or
plunger. Note: In general, twisting the plunger clockwise reduces
the volume, and twisting the plunger counterclockwise increases
the volume.
4. Place a clean tip on the micropipette.
5. Press the plunger down to the first stop. Hold the plunger down
while placing the tip beneath the surface of the liquid.
6. Slowly release the plunger to draw the sample into the pipette tip.
7. Deliver the sample by slowly pressing the plunger to the first
stop. Depress the plunger to the second stop to expel any
remaining sample. Do not release the plunger until the tip is out
of the sample container.
8. To discard the tip, press the eject button. Use a new tip for the
next sample.

How to use a Microscope

Step 1: Connect the light microscope to a power source. If your microscope uses a mirror instead of an illuminator, you can skip

this step. Instead, find a place where natural light is easily accessible

Step 2: Turn the revolving nosepiece so the lowest objective lens is in position.

Step 3: Mount your specimen onto the stage. But before doing so, see to it that your specimen is adequately protected by placing

a coverslip on top of it (if needed).

Step 4: Use the metal clips to keep your slide in place. Make sure the specimen is positioned in the center, right under the lowest

objective lens.

Step 5: Look into the eyepiece and slowly rotate the coarse adjustment knob to bring your specimen to focus. See to it that the

slide does not touch the lens.

Step 6: Adjust the condenser for the maximum amount of light. Since you're on the low power objective, you may have to

decrease the illumination. Use the diaphragm under the stage to adjust.

Step 7: Now slowly rotate the fine adjustment knob until you obtain a clearer image of your specimen.

Step 8: Examine your specimen.

Step 9: After you're done viewing with the lowest power objective, switch to the medium power objective and re-adjust the focus

with the fine adjustment knob (if you need to use course knob return to low power objective lens).

Step 10: Proceed to the high power objective once you have it focused.

You might also like