Procedure revised: 08/27/03 Scan Data Analysis Tools (SCAN, Tracking dd, CAL) 1 of 20
GE MEDICAL SYSTEMS
DIRECTION 2349185-800, REVISION 02 LIGHTSPEED 5.X SYSTEM SERVICE INFORMATION
VVC (Views vs. Channels): This is a way to graphically represent the data values from each
channel for each view of data from the MDAS as a shade of grey. The display has the views stacked
vertically and the channels arranged across the display horizontally.
3.1 UPDATE
The UPDATE selection will refresh the List Select display if new scan files have been created since
the Scan Analysis Tool was started.
Procedure revised: 08/27/03 Scan Data Analysis Tools (SCAN, Tracking dd, CAL) 3 of 20
GE MEDICAL SYSTEMS
DIRECTION 2349185-800, REVISION 02 LIGHTSPEED 5.X SYSTEM SERVICE INFORMATION
created. Once created, dd File can be viewed or compared with other files to check for specific
operating characteristics.
Procedure revised: 08/27/03 Scan Data Analysis Tools (SCAN, Tracking dd, CAL) 5 of 20
GE MEDICAL SYSTEMS
DIRECTION 2349185-800, REVISION 02 LIGHTSPEED 5.X SYSTEM SERVICE INFORMATION
Procedure revised: 08/27/03 Scan Data Analysis Tools (SCAN, Tracking dd, CAL) 7 of 20
GE MEDICAL SYSTEMS
DIRECTION 2349185-800, REVISION 02 LIGHTSPEED 5.X SYSTEM SERVICE INFORMATION
4.10 dd Math Operations in ddLS
The dd math operation buttons become insensitive if no files are selected in the dd math operation
panel.
The user may start dd math operation(s) by selecting the file(s) and putting them into the selection field
by clicking the button FILE #1 or FILE #2. If the selected file is not a dd file, the application will not put it
into the dd math operation field. A message window will pop up and ask user to select a dd file.
If only one file is selected and it is of the file type RTS dd file or MSD dd file, both CHANNEL TO
CHANNEL and RATIO OF MEANS VS. STDV become sensitive. If the selected file is not of the
type MSD or RTS, only CHANNEL TO CHANNEL become sensitive. When two dd files are
selected, ADD, SUBTRACT, MULTIPLY, and DIVIDE become sensitive and CHANNEL TO
CHANNEL and RATIO OF MEANS VS. STDV become insensitive.
The user can specify the output file name when the dd file output mode is set. Otherwise the system
offers a default dd file name.
The default output scaling factor is 1.0. The user can set the scaling factor to any real number.
When the dd math operation buttons are sensitive, the user can select the desired button to start
the dd math operation.
5 Z-Axis Tracking
The Z-AXIS TRACKING tool is a new TAB, located within the Analysis Tool. The tool can be
used to plot various tracking functions, using a Scan Data Set. For a scan data set, the analysis
package can plot different data versus views in UN-FILTERED (the default) or FILTERED (20 pt.
Boxcar) formats. Numerical information (“Max”, “Min”, “Mean”, and “Std. Dev.”) is also provided. In
some plots, the numerical information provided can be used for further analysis by comparing it to
a “specification” value, as an indication of a pass/fail condition. Whereas other plots are more
general, and in some cases may be useful, they are typically only used for troubleshooting.
A value is not considered to be out of specification unless the limit is exceeded for a sustained
interval of 100 views or more. In the cases where specifications are not given, consider plots
informational only.
Procedure revised: 08/27/03 Scan Data Analysis Tools (SCAN, Tracking dd, CAL) 9 of 20
GE MEDICAL SYSTEMS
DIRECTION 2349185-800, REVISION 02 LIGHTSPEED 5.X SYSTEM SERVICE INFORMATION
5.2 LOOP ERROR (MBP)
A LOOP ERROR MBP (Mean Beam Position) plot is the same as the loop error plot, except that
the display represents the loop error relative to the mean beam position during FASTCAL. The
FASTCAL beam position is stored in the calibration database.
5.3 Z RATIO
Z-RATIO Plot computes the ratio of outer row Z-channels (Channels 763, 764, & 765 averaged) to
inner row Z-channels. This is done for both the “A” and “B” sides.
5.5 APERTURE
APERTURE plot indicates the width of the Collimator Cam aperture in millimeters. Due to the
distance magnification factor, the width at the collimator is smaller than prescribed acquisition
mode, or width of the beam at the detector window.
Procedure revised: 08/27/03 Scan Data Analysis Tools (SCAN, Tracking dd, CAL) 11 of 20
GE MEDICAL SYSTEMS
DIRECTION 2349185-800, REVISION 02 LIGHTSPEED 5.X SYSTEM SERVICE INFORMATION
5.6 FOCAL SPOT POSITION (A/B)
FOCAL SPOT POSITION plot shows the calculated focal spot position from the centerline of either
the A or B side. The vertical scale (in millimeter) represents that portion of the focal spot length. The
absolute value of the A side plus the B side should equal the focal spot length (Small Spot = 0.7mm,
Large Spot = 1.2mm).
Procedure revised: 08/27/03 Scan Data Analysis Tools (SCAN, Tracking dd, CAL) 13 of 20
GE MEDICAL SYSTEMS
DIRECTION 2349185-800, REVISION 02 LIGHTSPEED 5.X SYSTEM SERVICE INFORMATION
5.10 ROTOR RUN
Figure 0-1 ROTOR RUN provides a plot of high frequency variations that are IN phase, such as
the small periodic movement of the anode at the rotor run frequency. Typical values are less than
0.1 count values.
Procedure revised: 08/27/03 Scan Data Analysis Tools (SCAN, Tracking dd, CAL) 15 of 20
GE MEDICAL SYSTEMS
DIRECTION 2349185-800, REVISION 02 LIGHTSPEED 5.X SYSTEM SERVICE INFORMATION
Channel Cursor
Once again the dip in the KV values reported in the view data corresponds to views around 615.
From the previous examples, it is easy to correlate the views with suspect data from the VVC
Display with the view by view plots for kV, mA, and Channels. Specific information to look for on the
examples:
• The Min, Max, and average values for kV, mA, and channel data. This information provides a
quick way to determine the scale of the information that you are viewing.
• The cursor report information provides a continuous update, depending upon the type of data
that is being displayed: data values, view number, channel number.
Procedure revised: 08/27/03 Scan Data Analysis Tools (SCAN, Tracking dd, CAL) 17 of 20
GE MEDICAL SYSTEMS
DIRECTION 2349185-800, REVISION 02 LIGHTSPEED 5.X SYSTEM SERVICE INFORMATION
18 of 20 7 - Typical Examples of CAL Plots with Scan Analysis Procedure revised: 08/27/03
GE MEDICAL SYSTEMS
DIRECTION 2349185-800, REVISION 02 LIGHTSPEED 5.X SYSTEM SERVICE INFORMATION
Figure 23: Calibration Vector "B1" / Head Filter
Procedure revised: 08/27/03 Scan Data Analysis Tools (SCAN, Tracking dd, CAL) 19 of 20
GE MEDICAL SYSTEMS
DIRECTION 2349185-800, REVISION 02 LIGHTSPEED 5.X SYSTEM SERVICE INFORMATION
Figure 26: Scaled P-Cal - Head Filter
20 of 20 7 - Typical Examples of CAL Plots with Scan Analysis Procedure revised: 08/27/03