0% found this document useful (0 votes)
10 views

Radiacode RC-10x_Device_Manual

Uploaded by

Cristiano Fidani
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
10 views

Radiacode RC-10x_Device_Manual

Uploaded by

Cristiano Fidani
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 43

I Personal radiation detector Radiacode-10X series. User manual.

Contents
Chapter I Introduction 1
1 Appointment of the device
................................................................................................................................... 2
2 Device...................................................................................................................................
design 3
3 Technical specifications
................................................................................................................................... 4
4 Developer contacts
................................................................................................................................... 5

Chapter II Safety precautions and warnings 6

Chapter III Stand-alone use of the device 9


1 Before use
................................................................................................................................... 10
2 Buttons
................................................................................................................................... 11
3 Turning on and off
................................................................................................................................... 12
4 Display
................................................................................................................................... 12
5 Icons................................................................................................................................... 13
6 Indication: sound, vibration, light
................................................................................................................................... 15
7 Battery charging
................................................................................................................................... 15
8 Firmware update
................................................................................................................................... 16
9 Menu...................................................................................................................................
usage 16
10 Device settings
................................................................................................................................... 18
Selecting units
..........................................................................................................................................................
of m easurem ent 18
Count rate - alarm
..........................................................................................................................................................
levels 19
Dose rate - alarm
..........................................................................................................................................................
levels 20
Dose - alarm ..........................................................................................................................................................
levels 21
Display settings
.......................................................................................................................................................... 22
Signals - sound,
..........................................................................................................................................................
vibration, light 24
Bluetooth settings
.......................................................................................................................................................... 27
Language selection
.......................................................................................................................................................... 27
Tim e settings
.......................................................................................................................................................... 28
Reset to factory
..........................................................................................................................................................
settings 28
11 Display Modes
................................................................................................................................... 29
Spectrum .......................................................................................................................................................... 30
Monitor .......................................................................................................................................................... 30
Dose .......................................................................................................................................................... 32
Search .......................................................................................................................................................... 33
Device info .......................................................................................................................................................... 34
12 Alarms
................................................................................................................................... 34

Chapter IV Working with a smartphone 37

Chapter V Work with a PC 39

Index 41

© 2024 Radiacode Ltd.


Chapter

I
2 Introduction

1 Introduction
This section contains the general information about dosimeter Radiacode-10X series. Later in the
document, the dosimeter Radiacode-10X series is also referred to as "device" and "dosimeter".

1.1 Appointment of the device

Radiacode-10X series

Portable dosimeter personal radiation detector Radiacode-10X series is designed to assess the radiation
situation in real time. In addition, the instrument automatically keeps a log of measurement results and
events. This makes it possible to analyze them, including with the help of standard software on a
smartphone or personal computer. The dosimeter is suitable for work on the ground, with samples of
construction materials, with specially selected and prepared food or soil samples.

The instrument has high sensitivity to gamma and X-ray radiation.


It is suitable for assessing dose rate, accumulated dose, absorbed radiation energy spectrum and
radiation hazard warning.

Personal radiation detector Radiacode-10X series is a universal tool for radiation safety assessment.
With it, it is always available for:
site reconnaissance;
· inspection of objects and materials;
· continuous monitoring of the radiation situation;
· automatic data storage in the instrument's memory;
· accumulated dose monitoring;
· quick search for radiation sources;
· two-channel alarm signaling of radiation danger and changes in the radiation situation;
· time-based database;
· possibility of autonomous operation, or in pair with a smartphone;
· wireless data transmission;
· wide possibilities of data processing in Windows, Android and IOS applications.

Instrument features
· Fast response to changes in the radiation environment, thanks to the CsI(Tl) scintillator detector
working together with a silicon photomultiplier tube and adaptive processing of incoming data.
· Continuous assessment of the radiation situation as long as the instrument is switched on. Results at
any moment can be presented in the form of dose rate, count rate, accumulated dose, absorbed
radiation energy spectrum. Assessment and accumulation of data do not depend on the change of the
display mode.
· Splash-proof, dust-proof - IP64 class.
· Long autonomous operation from the built-in battery.
· Independent alarm thresholds for count rate exceedance, dose rate exceedance and accumulated
dose exceedance. Possibility to set 2 thresholds for each value.
· Built-in non-volatile memory for data accumulation, the volume of about 1000 hours of autonomous
observations. Possibility to transfer data to a computer via USB and to a smartphone via Bluetooth.
Storage of results in an external database.

© 2024 Radiacode Ltd.


Introduction 3

· Binding of results to geographic coordinates and display on Google and OSM maps when used in
conjunction with a smartphone.
· Availability of a program for personal computer and applications for Android and IOS.
· Continuous development of the project. Possibility of software upgrades.
· Economical graphic display with automatic backlighting in the dark. Rotation of the image on the
screen when the device is rotated.
· Sound, light and vibration alarms on the device, and the ability to duplicate them on a smartphone.

Radiacode-10X series is a highly sophisticated and sensitive instrument. Please read the
instructions to obtain reliable results.

1.2 Device design


Design elements, controls and indications
controls and indications personal radiation detector Radiacode-10X series

1. Location of the radiation sensor, digital thermometer, and instrument orientation sensor.
2. Ambient light sensor to turn on the screen backlight in the dark.
3. Status of the built-in battery: charge level, charging, end of charging (powered from external
source).
4. Control button lock indication.
5. External connection: USB, Bluetooth.
6. Audio indication status.
7. Presence, source and level of alarm.
8. Temperature of the radiation sensor.
9. Current time.
10. Swing button "up", when the display image is reversed, "down".
11. USB type C connector for charging the device and data exchange.
12. Power and confirmation button, round.

© 2024 Radiacode Ltd.


4 Introduction

13. Swing button "down", when the display image is turned upside down.
14. Graphical representation of dose rate.
15. Numerical representation of dose rate.
16. Source of audible indication.
17. Units of numerical representation of dose rate.
18. Light indication:
- charge (blue color);
- gamma-quantum registration (green in normal, red when the alarm threshold is exceeded).
19. Random error of dose rate estimation.

1.3 Technical specifications


General technical characteristics

Energy range of registered photon radiation (X-ray and 0.02…3 MeV


gamma)
Dose rate assessment:
§ range (by Cs-137) 0.1…1000 uSv/h
§ error ±15%
§ time (display update period) 0.5 s
Accumulated dose assessment:
§ range 0…10 Sv
Non-volatile memory:
§ registration volume of Standalone mode last 1000 h
Continuous operation time (for fully charged battery) up to 300 h
Built-in battery:
§ type and capacity Li-Pol 3.7 V, 1000 mAh
§ charging voltage, typical 5V
§ charging current, max 500 mA
§ full charge time, typical 2h
Operating temperature range -10…+45°C
Relative humidity, max 85 %
Wire connection:
§ type USB 2.0 and higher
§ cable type C
§ current consumption, max 500 mA
Wireless connection BLE (Bluetooth 4.0 and higher)
Display Monochrome graphic LCD, 128x48
pixels, 34x13 mm, FSTN,
Transflective, Positive
Dimensions, W*H*D 123x34x18 mm
Weight 65 g
Dust and moisture protection, code IP64

Requirements for
The device works with personal computers with minimum system requirements:
- Windows XP/7/8/10 operating system;

© 2024 Radiacode Ltd.


Introduction 5

- RAM (RAM) not less than 2 GB;


- video card with resolution not worse than 1024x768;
- free space on the hard disk not less than 1 GB;
- USB port 1 pc.

Technical requirements to the mobile device


The device works with tablets or smartphones with minimum system requirements:
Android operating system version 6.0 and higher.

The device works with iPhone and iPad devices with minimum system requirements:
iOS operating system 17.0 or higher, or macOS 14.0 or newer and MAC with Apple chips.

1.4 Developer contacts


Radiacode Ltd.
3 Thalia Street, Office 222, Limassol 3011, Cyprus
Тел. +357 96 207695
E-mail: [email protected]
Web:https://radiacode.com

© 2024 Radiacode Ltd.


Chapter

II
Safety precautions and warnings 7

2 Safety precautions and warnings


Read this manual carefully before using the device.

· Keep the device out of the reach of children and pets!


· Do not disassemble the device! Repair of the device is allowed only in certified service centers.
· Do not use defective chargers! Use only dedicated chargers or USB port of computer to charge
the device! A device damaged as a result of improper charging is not eligible for warranty repair.
· Protect your device from extreme temperatures (below -20 ° C or above +45 ° C). Temperatures that
are too high or too low will reduce the capacity of the battery and shorten its life.
· Avoid getting the device wet. Moisture can cause serious damage to the device. Moisture penetration
into the device will void the manufacturer's warranty.
· If you brought the package with the device from frost to a warm room, leave it to warm up for two hours
without opening. Otherwise, moisture may condense inside the device and cause the device to
malfunction.
· Do not use or store the device in dirty areas.
· Avoid exposing the device to strong electromagnetic fields.The action of the electromagnetic field can
damage the device.
· Mobile phones, electronic and household appliances can interfere with the operation of the device.
· External shocks and rough handling can seriously damage the electronics of the device.
· Do not grip the device strongly in your hand, press the buttons lightly.
· Handle the device with care, the display is made of glass and can be broken if handled roughly.
· Do not disconnect the device from the computer while the firmware is being updated. An attempt to
disrupt the download process may lead to a malfunction of the device, which can only be fixed by a
certified service center.

FCC warning.

Federal Communications Commission (FCC) Statement. This device complies with part 15 of the FCC
Rules. Operation is subject to the following twoconditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired
operation. Note: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide Reasonable protection
against harmful interference in a residential installation. This equipment generates, uses and can radiate
radio frequency energy and, if not installed and used in accordance with the instructions, may cause
harmful interference to radio communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on, the user is encouraged to try to
correct the interference by one ormore of the following measures:
· Reorient or relocate the receiving antenna.
· Increase the separation between the equipment and receiver.
· Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
· Consult the dealer or an experienced radio/TV technician for help.

Warning: Changes or modifications made to this device not expressly approved by Radiacode LTD may
void the FCC authorization to operate this device.Note: The manufacturer is not responsible for any radio
or TV interference caused by unauthorized modifications to this equipment. Such modifications could
void the user’s authority to operate the equipment.

© 2024 Radiacode Ltd.


8 Safety precautions and warnings

RF exposure statement:
This device complies with FCC RF radiation exposure limits set forth for an uncontrolled environment.
The device is installed and operated without restriction.

© 2024 Radiacode Ltd.


Chapter

III
10 Stand-alone use of the device

3 Stand-alone use of the device


Portable dosimeter personal radiation detector Radiacode-10X series is designed to assess the radiation
situation in real time. In addition, the instrument automatically keeps a log of measurement results and
events. This makes it possible to analyze them, including with the help of standard software on a
smartphone or personal computer. The dosimeter is suitable for work on the ground, with samples of
construction materials, with specially selected and prepared food or soil samples.

Radiacode-10X series is designed for long-term autonomous use. Results at any moment can be
presented in the form of dose rate estimation, count rate, accumulated dose, absorbed radiation energy
spectrum. Assessment and accumulation of data do not depend on the change of display mode.
The instrument allows storing the accumulated data in the internal memory up to thousand last hours of
observations.

In addition to autonomous operation personal radiation detector can be connected to a personal


computer via USB cable or to an Android or iOS smartphone via Bluetooth. Connection to an Android
smartphone is also possible via USB OTG. When connected to a computer or smartphone, the
accumulated data are transferred to an external device, and the newly received data are transferred there
immediately without being stored in the built-in memory.

To connect the device to a computer it is necessary to install software Radiacode.


The program is available for download at https://space.radiacode.com.
With the help of the program Radiacode you can transfer data to a computer, view it and manage it
personal radiation detectorом.
For a detailed description of the program interface, see in chapter "Working with the use of a computer".

To work with a smartphone install the app on it Radiacode,


available at https://space.radiacode.com and GooglePlay,
or use the QR code:

3.1 Before use


Note: If you brought the package with the device from frost to a warm room, leave it to warm up
for two hours without opening. Otherwise, moisture may condense inside the device and cause
the device to malfunction. Moisture penetration into the device will void the manufacturer's warranty.

First steps:
· take the device out of the package;
· make sure it is intact;
· read the instructions;
· check the completeness according to the product passport;
· check the availability and correctness of filling in the passport/warranty card;
· fully charge the battery.

© 2024 Radiacode Ltd.


Stand-alone use of the device 11

After the battery is fully charged, turn on the device and set the current time. After it, you may
additionally:
· select menu language;
· set display parameters;
· set up signals;
· update firmware of the device.

3.2 Buttons
Personal radiation detector Radiacode-10X series has an intuitive menu system and is controlled by
three buttons.
Sound and vibration indication are also helpful in operating the instrument.

The device is in the left hand in the right hand


1. round button for turning on and confirming;
2. swing button "down";
3. swing button "up".

There are two types of pressing the buttons: short and long (at least two seconds).
A long press on the round button [1] is used to
· turn on the instrument;
· unlock the swing buttons;
· call up the menu;
· confirm viewing of the presented list of maximum alarm levels;
· in some setup screens.
A short press of the round button is used to:
· to lock the swing buttons;
· return from the main menu to the display mode;
· confirm selection of the highlighted menu item;
· toggle the selection options.
· to turn off the alarms.

The swing buttons are automatically locked after 5, 10, 15, 30 seconds of inactivity. At the same time,
the backlight of the screen is switched off, if it was on. In addition, there are 2 additional time intervals for
the display backlighting - 2 and 5 minutes, the swing buttons are locked after 30 seconds of inactivity.
The time until the backlight auto-locks and turns off depends on the settings.

You can unlock these buttons by long pressing the round button [1]. This will also turn on the screen
backlight, if it is enabled in the instrument settings.
If there is sufficient light in the "Auto" mode, the backlight will not be switched on. The buttons can be
switched back to the locked state by briefly pressing the round button [1].

A short press on the up/down swing buttons [2, 3] is used to:


· move between menu items;
· setting numerical values.
· additional options in different modes.

© 2024 Radiacode Ltd.


12 Stand-alone use of the device

Long press on the up/down swing buttons[2, 3] is used for different purposes in different cases.
In menus, a long press on the up/down swing buttons[2, 3] leads to a quick selection of items.
In scene display modes, a long press on the "up" swing button turns the sound indication on and off.

Which of the swing buttons will be "up" and which will be "down" depends on the screen orientation
settings.
When moving through the menu at any level, the selected item will be presented in negative - light on
dark.

Detailed description of using the buttons in different modes can be found in the sections "Display Modes"
и "Menu Usage".

Pressing any button is accompanied by a short sound and vibration, depending on the settings.

*The sound and vibration accompanying button presses can be turned off.

3.3 Turning on and off


To include Radiacode-10X series click on round button and hold it down for three seconds.
The device can be switched off via the menu.

· To turn on the instrument, press and hold the round button until the
instrument turns on (at least 3 seconds).
o When the instrument is turned on, it will be in Monitor mode.
o With the appropriate settings, the power on. sound will be played at this
moment.
o With the appropriate settings, the vibration indication. will be played at
this moment.
o At appropriate settings, the display backlight will be turned on at this
point.

· To turn off, press and hold the round button of the instrument until the
menu icons appear (at least 3 seconds).

o Icon display indicates that the swing buttons are locked.


o To unlock, press and hold the round button (at least 3 seconds).

· Short presses on the up or down swing buttons move the cursor to the

shutdown icon .
· A short press on the round button will turn the appliance off.
o With the appropriate settings, a shutdown sound will be played at this
point.
o The instrument screen will be turned off.

3.4 Display
The instrument is equipped with a graphical monochrome
display.

© 2024 Radiacode Ltd.


Stand-alone use of the device 13

· Monochrome graphic LCD;


· 128x48 dots;
· 34x13 mm;
· FSTN, transflective, positive;
· built-in backlight.
*Handle with care, the display may be brok en or scratched if handled carelessly.

The screen is equipped with a backlight for operation in the dark.


The device automatically detects the light level when the setting is set to "Auto". In this case, the
backlight is turned on only when the button is pressed. In bright light, for example in the sun, the
backlight will not turn on.
The backlight brightness has ten gradations.

You can set the orientation of the image on the screen for left or right handed operation, or select the
automatic rotation mode.

Power on conditions, backlight time, backlight brightness and image orientation are set via the menu,
section Settings.

The display is used to show:


· instrument status;
· graphical dose rate level;
· control menu;
· settings menu;
· spectrum mode, for estimation of absorbed radiation energies;
· monitor mode, for estimation of dose rate;
· Dose mode, for estimation of accumulated dose;
· search mode, for visual estimation of count rate.

The status panel is placed in the upper part of the display. The content of the status panel depends on
the selection of the display mode.
The type and meaning of the icons used are described in more detail in the section Icons.

3.5 Icons
In all display modes except settings, the display shows a logarithmic indicator:

Each horizontal bar shows a graphical representation of a value (dose rate, dose, count rate) on a
logarithmic scale.
The ranges of the bars differ by a factor of 10. For example, for dose rate (µSv/h) the ranges will look as
follows:

© 2024 Radiacode Ltd.


14 Stand-alone use of the device

Various icons are used to label fields on the instrument status panel.

At least 60% charge.

At least 30% charge.

At least 10% charge.

Less than 10% charge, the battery needs to be recharged.

The battery is being charged.

Charge completed, external power supply.

The "swing" buttons are locked.

USB communication has been established.

Bluetooth communication has been established.

All sounds are off.

Sounds included.

Level 1 alarm on Count Rate.

Level 2 alarm on Count Rate.

Count Rate over-range alarm.

Level 1 Dose Rate alarm.

Level 2 Dose Rate alarm.

Dose Rate over-range alarm.

Level 1 alarm on Dose.

Level 2 alarm on Dose.

Dose off-scale alarm.

Linear scale of the amplitude of the energy spectrum of photon radiation.

Logarithmic scale of the amplitude of the energy spectrum of photon radiation.

In the menu (enter by long pressing the round button), the sections are indicated by icons and text.

Settings Instrument settings menu.

© 2024 Radiacode Ltd.


Stand-alone use of the device 15

Spectrum Display of the energy spectrum of radiation.

Monitor Display the current dose rate or count rate.

Dose Display of accumulated dose of photon radiation (gamma and X-ray).

Search Search mode - display the count rate as a graph.

Power off Turning off the device.

About device Information about the dosimeter serial number and software version.

3.6 Indication: sound, vibration, light


In personal radiation detectorе Radiacode-10X series sound and vibration sources are built in to
acknowledge button presses and to indicate events and alarms. A light indication is provided for remote
monitoring

Sound can be accompanied by:


· switching the device on and off;
· pressing a button;
· establishment of communication via Bluetooth;
· registration of radiation quantum;
· exceeding alarm thresholds;
· low battery warning;
· response to the "Search for device" request.

Vibration can be accompanied by:


· switching the device on and off;
· pressing the button;
· exceeding alarm thresholds;
· response to the "Search for the device" request.

Light indication:
· battery charging - blue color, continuous glow;
· registration of radiation quantum in normal conditions - green color, flash;
· registration of radiation quantum in case of alarm - red color, flash;
· exceeding the alarm threshold - switching on the screen backlight.

It is possible to turn signals on or off the alarms by separate groups:


· all sounds;
· button pressing sound;
· sound of radiation quantum registration;
· all types of vibration;
· light indication.

3.7 Battery charging


To charge the battery:
· plug the USB cable to the connector located at the end of the device ([14] in the general view);
· plug the USB cable to the connector of a computer or a special charger;
· leave the device connected to the charging source until it is fully charged.
The indicator of the activity of the charging process is the glowing of the blue indicator ([18] in the general
view). If the device is turned on, then the screen displays the icon ([3] in the general view). At the end of
the charging procedure, the blue indicator will turn off and the icon will change to the sign of power
supply from an external source.

© 2024 Radiacode Ltd.


16 Stand-alone use of the device

Pay attention to the charge indicator icon while using the device. If the charge indicator is empty ,
then connect the device to the charger.

Attention! Use only dedicated chargers or a computer USB port to charge the device. Connecting
the device to the wrong charging source may damage it. A device damaged as a result of
improper charging is not eligible for warranty repair.
If the battery is completely discharged, the instrument may not respond to attempts to turn on. To
start the dosimeter, the battery must be recharged. Then disconnect the cable and reconnect it.
The device will be ready for operation.

3.8 Firmware update


The device is constantly being improved and the firmware (built-in software) is regularly updated by the
developer company. The current version of the device firmware is available through the Device info menu
item.
To update the firmware take the following steps:
· download the latest version of Radiacode software at https://space.radiacode.com;
· install Radiacode on your computer and run it;
· connect the device directly to your computer's USB port using the USB cable type-C (supplied with the
device);
· turn on the device;
· if there is a fresh version of the firmware, you will be prompted to update it;
· wait for the update process to finish.

Update the firmware using your smartphone and the Android app:
· install the latest version of the Android application in the application store: Radiacode or download from
the link: https://space.radiacode.com;

If the device bootloader is version 4.0 or higher - it is possible to update the firmware via Bluetooth:
launch the application;
· call the menu by pressing the button with three horizontal dashes on the top left;
· select "Devices";
· select Bluetooth communication interface and confirm by pressing the [OK] button;
· after connecting the device, confirm the firmware update of the device;
· wait for the download and update process to complete. The red LED will be on during the update
process.

If the device bootloader is lower than version 4.0 you will need a USB OTG adapter for your
smartphone:
· connect the instrument with the cable and adapter to your smartphone;
· Allow the RadiaCode app to be used with the connected device;
· launch the app;
· access the menu by pressing the button with three horizontal dashes on the top left;
· select "Devices";
· select the USB communication interface and confirm by pressing the [OK] button;
· confirm the firmware update of the instrument;
· wait for the update process to complete.

Attention! Do not turn off the instrument and do not disconnect the USB cable until the
process is complete!
Violation of the firmware update process may lead to instrument failure.
Please contact the service department to restore operation.

3.9 Menu usage


Let's take the state "The device is off" as the initial state.

© 2024 Radiacode Ltd.


Stand-alone use of the device 17

· Press and hold the round button until the instrument


turns on (at least 3 seconds);
o When turned on, the instrument will be in the Monitor
and display the dose rate;
o With the appropriate settings, the sound power on will
be played at this time.
o At appropriate settings at this moment the backlight
of the display will be turned on.

· Press and hold the round button to move to the menu;


o The name of the current menu item will be displayed
at the top. (In this case it is "Monitor");
o The icons of the main menu items will be placed
vertically;
o The active item will be highlighted with a negative
icon.

There are seven menu items available on the main menu level.

Settings Device settings - go to the submenu;

Spectrum Photon radiation energy spectrum (gamma and X-ray) - display mode;

Monitor Dose rate, count rate - display mode;

Dose Accumulated dose - display mode;

Search Count rate search graph - display mode

Power off Turning off the appliance is an action;

About device Serial number and software version information.

Use a short press on the up/down swing buttons to select the desired item. The selected item will be
highlighted with a negative icon. To confirm the selection, use a short press on the round button.
Only three selections are available on the instrument screen at a time, the others will become available
when the up/down swing buttons are pressed again.

Of the seven items, four [ Spectrum, Monitor, Dose, Search] give access to the
selection of the radiation situation display mode.
Item [ Power off] – is an action point, if it is confirmed with the round button, the instrument will be
switched off.
Item [ Settings] is used to go to the submenu of the instrument settings.
Item [ About device] is used to obtain information about the instrument number and firmware and
bootloader versions.

At all levels, the menu is arranged in a cyclical fashion. For example, on the main level, pressing the
down swing button moves the focus from item to item in the following order: --> --> --> -->

--> --> --> ...

© 2024 Radiacode Ltd.


18 Stand-alone use of the device

3.10 Device settings

The instrument settings can be accessed from the corresponding instrument submenu –
"Settings".

· Press and hold the round button until you reach the
menu;
· The name of the current menu item will appear at the
top. (In this case it is "Monitor");
o The icons of the main menu items will be placed
vertically;
o The active item will be highlighted with a negative
icon.

· Short presses on the upper swing button (or the lower

one, the menu is circular) focus on the item


"Settings";
· Press the round button briefly to confirm access to
the "Settings" submenu.

Next, short presses on the up/down buttons of the swings can be used to move the focus to the desired
item and make the required settings.
This and deeper menu levels can be exited via [< Back] and [<< Menu quit].

Selecting units of measurement


Count rate - alarm levels
Dose rate - alarm levels
Dose - alarm levels
Display settings
Signals - sound, vibration, light
Bluetooth settings
Language selection
Time settings
Device info
Reset to factory settings

3.10.1 Selecting units of measurement

· Briefly press the up/down rocker buttons to move


the focus to [Measurement units];
· Short press the round button to move to
[Measurement units].

© 2024 Radiacode Ltd.


Stand-alone use of the device 19

· Short presses on the round button to select dose


rate measurement units:
· Sv - Sievert or R - Roentgen;
· Short presses on the up/down swing buttons
move the focus to the item [Count Rate].

· Short presses on the round button to select the


units of count rate estimation:
· CPS - Count Per Second;
· CPM - Count Per Minute.

· To return to the previous level or exit the menu,


briefly press the up/down swing buttons to move
the focus to [< Back] or [<< Menu quit];
· Briefly press the round button to confirm your
selection;

3.10.2 Count rate - alarm levels

· Briefly press the up/down rocker buttons to move


the focus to [Count rate];
· Short press the round button to move
to [Count rate].];

· Short press the round button to move to


[Alarm 1];
· Short press the round button to select the
desired digit of the threshold value;
· Short presses on the up/down swing buttons to
change the value of the selected digit;
· Set all digits;

· Short presses on the round button to select


[Enter] or [Esc];
· Long press the round button to confirm your
selection;

© 2024 Radiacode Ltd.


20 Stand-alone use of the device

This and deeper menu levels can be exited via [< Back] and [<< Menu quit].
or go to [Alarm 2].

· Short press the round button to move to


[Alarm 2];
· Short press the round button to select the
desired digit of the threshold value;
· Short presses on the up/down swing buttons to
change the value of the selected digit;
· Set all digits;

· Short presses on the round button to select


[Enter] or [Esc];
· Long press the round button to confirm your
selection;

· To return to the previous level or exit the menu,


briefly press the up/down swing buttons to move
the focus to [< Back] or [<< Menu quit];
· Briefly press the round button to confirm your
selection;

3.10.3 Dose rate - alarm levels

· Briefly press the up/down rocker buttons to move


the focus to [Dose rate];
· Short press the round button to move
to [Dose rate].

· Short press the round button to move to


[Alarm 1];
· Short press the round button to select the
desired digit of the threshold value;
· Short presses on the up/down swing buttons to
change the value of the selected digit;
· Set all digits;

© 2024 Radiacode Ltd.


Stand-alone use of the device 21

· Short presses on the round button to select


[Enter] or [Esc];
· Long press the round button to confirm your
selection;

This and deeper menu levels can be exited via [< Back] and [<< Menu quit].
or go to [Alarm 2].

· Short press the round button to move to


[Alarm 2];
· Short press the round button to select the
desired digit of the threshold value;
· Short presses on the up/down swing buttons to
change the value of the selected digit;
· Set all digits;

· Short presses on the round button to select


[Enter] or [Esc];
· Long press the round button to confirm your
selection;

· To return to the previous level or exit the menu,


briefly press the up/down swing buttons to move
the focus to [< Back] or [<< Menu quit];
· Briefly press the round button to confirm your
selection;

3.10.4 Dose - alarm levels

· Briefly press the up/down rocker buttons to move


the focus to [Dose];
· Short press the round button to move
to [Dose].

© 2024 Radiacode Ltd.


22 Stand-alone use of the device

· Short press the round button to move to


[Alarm 1];
· Short press the round button to select the
desired digit of the threshold value;
· Short presses on the up/down swing buttons to
change the value of the selected digit;
· Set all digits.

· Short presses on the round button to select


[Enter] or [Esc];
· Long press the round button to confirm your
selection.

This and deeper menu levels can be exited via [< Back] and [<< Menu quit].
or go to [Alarm 2].

· Short press the round button to move to


[Alarm 2];
· Short press the round button to select the
desired digit of the threshold value;
· Short presses on the up/down swing buttons to
change the value of the selected digit;
· Set all digits.

· Short presses on the round button to select


[Enter] or [Esc];
· Long press the round button to confirm your
selection.

· To return to the previous level or exit the menu,


briefly press the up/down swing buttons to move
the focus to [< Back] or [<< Menu quit];
· Briefly press the round button to confirm your
selection.

3.10.5 Display settings


You can customize the screen properties to suit your needs.

You can select one of ten screen backlight brightness levels (0 to 9). The default setting is 5. Lower
levels use up battery power more slowly. The screen backlight will turn off automatically after a period of
time when no buttons are pressed. You can select a shutdown delay of 5, 10, 15, 30 seconds and 2, 5
minutes. The screen backlight can be disabled completely, allowed to be unconditionally turned on by
pressing the buttons or by pressing the button in case of insufficient light level.

© 2024 Radiacode Ltd.


Stand-alone use of the device 23

By default the device is set in the mode of automatic screen rotation when working with it with right or left
hand. You can disable the automatic screen rotation option and select its orientation in the right or left
position.

· Briefly press the up/down rocker buttons to move


the focus to [Screen];
· Short press the round button to move
to [Screen].

· Short press the round button to move to


[Backlight];
· Briefly press the up/down swing buttons to move
the focus to [Turning on];
· Short presses on the round button to select the
backlight on mode:
[Auto], [Never], or [By Button];

· Briefly press the up/down swing buttons to move


the focus to [Duration];
· Short press the round button to select the delay
until the screen backlight turns off automatically:
5, 10, 15, 30 sec. and 2, 5 min;

· Short press the up/down swing buttons to move


the focus to [Brightness];
· Short press the round button to select the
backlight brightness level from 0 to 9;

· To return to the previous level, briefly press the


up/down swing buttons to move the focus to
[< Back];
· Briefly press the round button to confirm your
selection;

© 2024 Radiacode Ltd.


24 Stand-alone use of the device

· Briefly press the up/down swing buttons to move


the focus to [Rotate];
· Short press the round button to select the screen
orientation:
[Auto], [Right] or [Left];

· To return to the previous level, briefly press the


up/down swing buttons to move the focus to
[< Back];
· Briefly press the round button to confirm your
selection;

· To return to the previous level or exit the menu,


briefly press the up/down swing buttons to move
the focus to [< Back] or [<< Menu quit];
· Briefly press the round button to confirm your
selection.

3.10.6 Signals - sound, vibration, light


Personal radiation detector Radiacode-10X series is equipped with a combined interface for user
interaction:
· display;
· buttons;
· sound emitter;
· LEDs;
· vibration signaling device.

Sound, light, vibration complement the information presented on the display and provide feedback when
pressing the buttons.

Event/state Display Sound Vibration Led/Backlight

Switching on – + + –
Switching off – + – –
Establishing a connection , + – –
Loss of connection – + – –
Low battery charge + + –
Battery charge process – – blue - continuous

Device search – + + –

© 2024 Radiacode Ltd.


Stand-alone use of the device 25

Pressing the button – + + –


Quantum registration – + – green, red - flashes

Alarm 1 , , + + screen backlight

Alarm 1 , , + + screen backlight

Out of scale , , + + screen backlight

In the menu you can select the operation mode of sound, light and vibration alarms.
Separate permissions to play alarm sounds, buttons, clicks are available.

· Briefly press the up/down rocker buttons to move


the focus to [Signals];
· Short press the round button to move the focus to
[Signals].

· Briefly press the up/down swing buttons to move


the focus to the item [Sound];
· Short presses on the round button to select the
sound signaling mode:
[On] - all allowed sounds;
[Off] - only the sounds of searching for the
instrument and going out of the measuring scale
remain, if it is allowed by the settings.

o Vibration and light modes can be set in the same


way:
· Move the focus to [Vibro] or [Light] by short
presses on the up/down swing buttons;
· Short presses on the round button to select the
[On] or [Off] mode.

· The presence of gamma and X-ray photon


registration clicks can be controlled separately.
· By short presses on the up/down swing buttons
move the focus to the item [Clicks];
· Short presses on the round button to select the
[On] or [Off] mode.

· Briefly press the up/down swing buttons to move


the focus to the [Buttons];
· Short press the up/down swing buttons to move
the focus to [Buttons] or [Vibro];
· Short press the round button to select the
desired response type.

© 2024 Radiacode Ltd.


26 Stand-alone use of the device

o Separately, you can control the presence of the


device connection and disconnection signal.
· Briefly press the up/down swing buttons to move
the focus to [Connection];
· Short press the round button to select
[On] or [Off].

o The presence of the on/off signal can be controlled


separately.
· Short press the up/down swing buttons to move
the focus to [Power];
· Short press the round button to select
[On] or [Off].

· To set the alarms, move the focus to [Alarms];


· Press the round button briefly to confirm your
selection.

· Short presses on the round button to select the


alarm indication mode
[One-shot] or [Continuous].

· To select an alarm channel, move the focus to the


desired item, e.g. [Dose rate];
· Press the round button briefly to confirm your
selection.

· Briefly press the up/down swing buttons to move


the focus to the desired alarm level
[Alarm 1] or [Alarm 2];
· Press the round button briefly to confirm your
selection.

© 2024 Radiacode Ltd.


Stand-alone use of the device 27

· Short press the up/down swing buttons to move


the focus to [Sound] or [Vibration];
· Short press the round button to select
[On] or [Off];

· To return to the previous level or exit the menu,


briefly press the up/down swing buttons to move
the focus to [< Back] or [<< Menu quit];
· Briefly press the round button to confirm your
selection.

3.10.7 Bluetooth settings

· Short press the up/down swing buttons to move


the focus to [Bluetooth];
· Short press the round button to select the
Bluetooth wireless communication mode:
[On] or [Off];

· To return to the previous level or exit the menu,


briefly press the up/down swing buttons to move
the focus to [< Back] or [<< Menu quit];
· Briefly press the round button to confirm your
selection.

3.10.8 Language selection


The device can support multiple languages for menus and messages.

· Briefly press the up/down swing buttons to move


the focus to the item [Language];
· Short press the round button to select the
instrument interface language: [English] or other
supported languages.

© 2024 Radiacode Ltd.


28 Stand-alone use of the device

· To return to the previous level or exit the menu,


briefly press the up/down swing buttons to move
the focus to [< Back] or [<< Menu quit];
· Briefly press the round button to confirm your
selection.

3.10.9 Time settings


When connecting Radiacode-10X series The time on the device will be set automatically when you
connect it to your smartphone or computer.
Time setting is also available from the local menu of the device:

· Briefly press the up/down rocker buttons to move


the focus to [Time];
· Short press the round button to confirm your
selection.

· Short presses on the round button to select the


desired time entry field;
· Short press the up/down swing button to change
the value of the selected field;
· To quickly change values - hold the up or down
swing button.

· Short press the round button to select


[Enter] or [Esc];
· Long press the round button to confirm your
selection;
o This will return to the previous menu level.

· To return to the previous level or exit the menu,


briefly press the up/down swing buttons to move
the focus to [< Back] or [<< Menu quit];
· Briefly press the round button to confirm your
selection.

3.10.10 Reset to factory settings


The menu provides an item for resetting the instrument to its original factory settings.

© 2024 Radiacode Ltd.


Stand-alone use of the device 29

· Briefly press the up/down swing buttons to move


the focus to the item
· [Factory settings];
· Press the round button briefly to confirm your
selection;

· Short presses on the round button to select the


[Yes] or [No] field;

· Long press the round button to confirm your


selection;
o Confirming [Yes] will reset the instrument to its
original settings and return to the previous menu
level.

· To return to the previous level or exit the menu,


briefly press the up/down swing buttons to move
the focus to [< Back] or [<< Menu quit];
· Briefly press the round button to confirm your
selection.

3.11 Display Modes


The Radiacode-10X series dosimeter has a single mode of operation - continuous data collection,
accumulation and analysis.
At any moment one of several ways of displaying the radiation situation assessment is available:

Spectrum – energy spectrum of photon radiation;

Monitor – estimation of dose rate and count rate;

.. Dose – dose assessment;

Search – operational search.

During operation, the instrument also detects the exceeding of all Alarms levels. Informs the user of
radiation hazards or changes in the environment.

© 2024 Radiacode Ltd.


30 Stand-alone use of the device

3.11.1 Spectrum
The mode is intended for representation of the general view of the photon (gamma and X-ray) radiation
energy spectrum.

In the Spectrum display mode the screen shows:


state panel - from top;
energy spectrum of photon (gamma and X-ray) radiation.

The upper part of the screen, from left to right, contains basic information about the instrument status:

o scale of the amplitude scale of the photon radiation energy spectrum - linear or logarithmic;
o time since the beginning of spectrum accumulation: hours, minutes, seconds;
o sign of exceeding thresholds by dose rate, dose rate, count rate;
o audio indication status;
o Bluetooth or USB connection;
o swing buttons lock sign;
o battery status.

At the bottom of the screen from left to right:

o graphical representation of dose rate estimation.


o histogram of the photon radiation energy spectrum;

Labels are placed under the energy scale of the histogram:


o dots - every 100keV;
o dashes every 500keV;
o arrows - every 1000keV.

Using the buttons


Quick control of the spectrum view is available via the buttons:
o a short press on the upper swing button switches the spectrum amplitude scale between linear and
logarithmic;
o a short press on the lower button of the swing cycles the scale of the energy scale 1MeV -> 2MeV ->
3MeV -> 1MeV.... ;
o long press on the lower button of the swing clears the screen and starts a new session of spectrum
accumulation;
o long pressing the upper button of the swing turns on and off the sound indication.

3.11.2 Monitor
This mode is intended to represent dose rate or count rate. When the instrument is turned on, it starts
working in this particular display mode. All incoming data are accumulated and analyzed. If a change in
the radiation situation is detected, the instrument starts a new interval of data accumulation. If there are
no signs of changes in the radiation situation, the averaging continues in order to increase the reliability
of the estimation. Statistical error in the Monitor mode is maintained at the level not more than 15% for
the confidence interval 2δ at sufficient radiation intensity.

© 2024 Radiacode Ltd.


Stand-alone use of the device 31

In the Monitor display mode, the screen shows:


· status panel - from the top;
· parameters of dose rate or count rate estimation - at the bottom.

The upper part of the screen, from left to right, contains basic information about the status of the
instrument:

o current time: hours, minutes;


o detector temperature;
o sign of exceeding the thresholds for dose rate, dose rate, count rate;
o audio indication status;
o presence of Bluetooth or USB connection;
o swing button lock sign;
o battery status.

At the bottom of the screen from left to right:


o graphical representation of the dose rate or count rate estimate;
o value of the dose rate or count rate estimate;
o value of random error in %;
o dose rate measurement units - Sv/h, R/h, or count rate - CPS, CPM.

With the help of the menu the choice between the measurement units is available:
o Sv/h or R/h - when displaying dose rate;
o CPS or CPM - when displaying the count rate.

The type of radiation intensity estimation: Dose rate or count rate can be selected in the menu, or can be
switched operatively with the instrument buttons.

Using the buttons


Quick control in this mode is available via the buttons:
· a short press on the top button of the swing to switch between the types of photon radiation intensity
display:
o dose rate or count rate;
o long pressing on the upper button of the swing turns on and off the sound indication;
o short pressing on the lower button of the swing switches on the sound indication of registered quanta
(clicks).

Setting alarm thresholds


Two alarm thresholds for the dose rate level can be set using the menu.
When the dose rate alarm threshold level is exceeded, a pulsating sign will be displayed in the status
bar:

© 2024 Radiacode Ltd.


32 Stand-alone use of the device

o "Alarm 1" – ;

o "Alarm 2" – .

o "Alarm 3 (off the scale)" – .

Two alarm thresholds for the count rate alarm level can be set using the menu.
When the count rate alarm threshold level is exceeded, a pulsating sign will be displayed in the status
bar:

o "Alarm 1" – ;

o "Alarm 2" – .

o "Alarm 3 (off the scale)" – .

3.11.3 Dose
The mode is intended for presentation of accumulated dose.

In the Dose display mode the screen shows:


· status panel - from the top;
· parameters of accumulated dose estimation.

The upper part of the screen, from left to right, contains basic information about the status of the
instrument:

o time since the beginning of dose accumulation: days, hours, minutes;


o detector temperature;
o sign of exceeding the thresholds for dose rate, dose rate, count rate;
o audio indication status;
o presence of Bluetooth or USB connection;
o swing button lock sign;
o battery status.

At the bottom of the screen from left to right:


o graphical representation of the cumulative dose estimate.
o cumulative dose assessment value;
o dose measurement units - Sv, R;

Using the menu you can choose between the measurement units:
o Sv or P - when displaying dose rate.

Using the buttons

© 2024 Radiacode Ltd.


Stand-alone use of the device 33

Quick operation in this mode is available via the buttons:


o long pressing the lower button resets the value of the accumulated dose;
o long pressing on the upper button of the swing turns on and off the sound indication.

Setting alarm thresholds


Two alarm thresholds for the accumulated dose level can be set using the menu.

When the alarm threshold level is exceeded, a pulsating sign will be displayed in the status bar:

o "Alarm 1" – ;

o "Alarm 2" – ;

o "Alarm 3 (off the scale)" – .

3.11.4 Search
The mode is designed to represent the count rate in the form of a search graph. It is applicable for
operative search of the source or zone with increased radiation level.

In the Search display mode the screen shows:


· status bar - on top;
· search graph of count rate - at the bottom.

The upper part of the screen, from left to right, contains basic information about the status of the
instrument:

o averaging time, price of one bar on the chart in seconds - 1/2, 1, 2, 4 seconds;
o current count rate value;
o count rate units - CPS or CPM;
o sign of exceeding thresholds on dose rate, dose rate, count rate level;
o audio indication status;
o Bluetooth or USB connection;
o swing buttons lock sign;
o battery status.

At the bottom of the screen from left to right:


o graphical representation of the current count rate.
o search graph of the count rate;

Using the buttons


Quick control in this mode is available via buttons:
o a short press on the upper swing button increases the averaging time;

© 2024 Radiacode Ltd.


34 Stand-alone use of the device

o a short press on the lower swing button decreases the averaging time;
o long press on the lower swing button clears the screen and starts a new search session;
o long pressing the upper swing button turns the sound indication on and off.

3.11.5 Device info


The mode is intended for presenting brief information about the device:
· Instrument model and serial number;
· Firmware version;
· Bootloader version;
o FCC ID. *Not specified for all devices

3.12 Alarms
The sources of alarms in the Radiacode-10X series device are:
current value of count rate value;
current value of dose rate value;
value of accumulated dose;

There are three alarm thresholds for each of these sources in the Radiacode-10X series instrument:
If the current count rate, dose rate, or dose value crosses any of the thresholds from bottom to top, an
Alarm event occurs:
· first threshold - Alarm 1;
· second threshold - Alarm 2;
· upper limit of the scale - over scale;

The values of alarm thresholds of the first and second level, separately for each of the sources, are set in
the device settings. They can be changed at your discretion.
The value of the third threshold is the upper limit of the scale of each of the measured values. These
thresholds cannot be changed through the menu.
The instrument constantly checks the current values of count rate, dose rate, accumulated dose and
compares them with the threshold values.

Exceeding any of the thresholds puts the instrument into the Alarm state.

Each Alarm Source defines one of 4 states:


· Normal - value below the first threshold (no alarm);
· Alarm 1 - value higher than the first threshold but not higher than the second;
· Alarm 2 - value higher than the second threshold, but not higher than the third;
· Over scale - value above the third threshold, going beyond the scale.

© 2024 Radiacode Ltd.


Stand-alone use of the device 35

Exceeding the upper threshold by count rate leads to unreliability of dose rate estimation. In this case
the upper threshold for dose rate is also considered to be exceeded.

If any of the alarm thresholds for dose rate, count rate or accumulated dose is exceeded, the
corresponding alarm will be activated. To stop the alarm for any level, you need to confirm by briefly
pressing the round button that you have accepted the alarm. The instrument will continue to display
exceeding the set level with a flashing icon of the appropriate type. To terminate an alarm based on the
accumulated dose level, you must reset it to zero, or change the corresponding alarm threshold.

Each Alarm Event has corresponding Alarms associated with it:


· audible alarm;
· vibration signal;

Audio and vibration Alarms can be enabled/disabled in the Radiacode-10X series settings for each level
of any of the alarm sources independently.

On an Alarm Event, Radiacode-10X series will play the Alarms allowed in the settings.
You can stop playing the alarm by briefly pressing the round button.
The next Alarm Event will be signaled again.

The Alarm Status is indicated by:


· red LED flashes;
· alarm icons on the status panel of the device display;
· unacknowledged alarms status line.

Alarm 1 on Count Rate.


Alarm 2 on Count Rate.
Alarm 3 on Count Rate over scale.
Alarm 1 on Dose Rate.
Alarm 2 on Dose Rate.
Alarm 3 on Dose Rate over scale.
Alarm 1 on Dose.
Alarm 2 on Dose.
Alarm 3 on Dose over scale

Light signals can be turned on/off in the settings Radiacode-10X series

If the device has been locked, if the alarm indication settings (sound and/or vibro) are enabled, a line of
maximum recorded levels of unconfirmed alarms will be shown on the device display after the thresholds
have been exceeded:

The following rules are adopted for indication of states and events of alarms on the display:

© 2024 Radiacode Ltd.


36 Stand-alone use of the device

· If the swing buttons are locked, the status panel displays the message "Alarms:" and icons of
maximum alarm levels reached from the moment the buttons are locked to the current moment of time
- one icon for each source.
· By long pressing the round button, the user can confirm viewing of the presented list of maximum
alarm levels, after which the swing buttons will be unlocked and the normal instrument status panel will
be displayed.
· If Alarm Events occurred while displaying the menu, upon exiting the menu, the display will show the
list of maximum alarm levels mentioned above. The swing buttons will then be locked.
· If the swing buttons are unlocked and the display shows one of the measurement display modes, the
status bar will show the current alarm level icons for each of the sources, alternating between them.

© 2024 Radiacode Ltd.


Chapter

IV
38 Working with a smartphone

4 Working with a smartphone


To pair with your smartphone, install the app on your smartphone Radiacode,
available at link: https://space.radiacode.com and GooglePlay,
or use the QR-code:

The IOS app is available by clicking here: https://apps.apple.com/ru/app/radiacode/id6449358533?l=en-


GB.
A detailed description of the application interface is included with the application and is available after
installation.

© 2024 Radiacode Ltd.


Chapter

V
40 Work with a PC

5 Work with a PC
To connect the device to a computer, the software must be installed Radiacode.
The program is available for downloading at: https://space.radiacode.com.
With the help of the program Radiacode you can transfer the data to a computer, view it,
and manage personal radiation detector. A detailed description of the program interface is included with
the program and is available after installation.

© 2024 Radiacode Ltd.


Index 41

menu usage 16

Index menu use 16

off 12
Alarm 34 on 12

Battery charging 15 PC 40
Before use 10 pressing 11, 16
Bluetooth 27
button 11, 16

Reset 28
resolution 12
contrast 12
control 11
Count rate alarm 19
screen 22
screen adjustment 22
screen contrast 22
date 28 screen orientation 22
Device settings 18 screen rotation 22
display 12 screen setting 22
Dose alarm 21 short press 11, 16
Dose rate alarm 20 smartphone 38
sound 15
Spectrum energy 30
Stand-alone use of the device 10

Firmware 16
Firmware update 16

Time 28
Time settings 28

icons 13
info 34

vibration 15

Language 27
Language selection 27
light 15
long press 11, 16

menu 16
© 2024 Radiacode Ltd.

You might also like