QC Lecture 2
QC Lecture 2
(Lecture 2)
• Quality Control - QC refers to the measures that must be included during each assay run to verify that
• Quality Assurance - QA is defined as the overall program that ensures that the final results reported by
the laboratory are correct. (the systems or procedures in place to avoid errors occurring)
• “The aim of quality control is simply to ensure that the results generated by the test are correct. However,
quality assurance is concerned with much more: that the right test is carried out on the right specimen,
and that the right result and right interpretation is delivered to the right person at the right time”.
QC , overview
means to determine the quality of the results generated by the laboratory. Quality
• Accepted true value - the value approximating the true value, the difference
between the two values is negligible.
• Error - the discrepancy between the result of a measurement and the true (or
accepted true value).
QC , overview
Unreliable Performance ?
• Potential consequences include:-
• patient misdiagnosis
• delays in treatment
• increased costs
• avoidable retests cost US 200million USD per year
• Even a small calibration bias can effect treatment rates:
• 1% +ve bias in cholesterol result
5% increase in patients exceeding the treatment cut-off
• 3% +ve bias
15% increase in patient treatment.
QC , overview
Random Error
• An error which varies in an unpredictable manner, in magnitude and sign, when a large
number of measurements of the same quantity are made under effectively identical
conditions.
• Random errors create a characteristic spread of results for any test method and cannot be
accounted for by applying corrections. Random errors are difficult to eliminate but
repetition reduces the influences of random errors.
• Examples of random errors include errors in pipetting and changes in incubation period.
Random errors can be minimized by training, supervision and adherence to standard
operating procedures.
QC , overview
Random Errors
x
x x
x x
True x x x x
Value x x x
x x x
x
x
x
QC , overview
Systematic Error
• An error which, in the course of a number of measurements of the same value of a
given quantity, remains constant when measurements are made under the same
conditions, or varies according to a definite law when conditions change.
• Systematic errors create a characteristic bias in the test results and can be accounted for
by applying a correction.
• Systematic errors may be induced by factors such as variations in incubation
temperature, blockage of plate washer, change in the reagent batch or modifications in
testing method.
QC , overview
Systematic Errors
x
x x x x x x x
True x
Value
QC , overview
Error Classification..
• Pre-analytical:-
– errors before the sample reaches the laboratory
• Analytical:-
– errors during the analysis of the sample
• Post-analytical:-
– errors occurring after the analysis
QC , overview
30
Post- Analytical Causes of Error
• Incorrect reference values
• Incorrect data entry of lab result
32
Random Analytical Error (RE)
Factors contributing to random analytical error (RE) are those that
affect the reproducibility of measurement. These include:
• instability of the instrument,
• variations in the temperature,
• variations in the reagents and calibrators (and calibration-curve
stability),
• variability in handling techniques such as pipetting, mixing, and
timing, and
• variability in operators.
These factors superimpose their effects on each other at different
times. Some cause rapid fluctuations, and others occur over a longer
time. Thus RE has different components of variation that are related
to the actual laboratory setting.
Systematic error
• Systematic change in the test system resulting in a displacement of
the mean from the original value
• Systematic error of an analytic system is predictable and causes shifts
or trends on control charts that are consistently low or high
36
Causes of Systematic Error
• Change in reagent or calibrator lot numbers
• Wrong calibrator values
• Improperly prepared reagents
• Deterioration of reagents or calibrators
• Inappropriate storage of reagents or calibrators
• Variation in sample or reagent volumes due to pipettor misalignments
• Variation in temperature or reaction chambers
• Deterioration of photometric light source
• Variation in procedure between technologists
38
Shifts and Trends
• Shift
• QC data results are distributed on one side of the mean for 6-7 consecutive
days
Trend
• Consistent increase or decrease of QC data points over a period of 6-7 days
40
Random Error
• Imprecision of the test system causing a scatter or spread of control
values around the mean
42
Causes of Random Error
• Air bubbles in reagent
• Improperly mixed reagents
• Reagent lines, sampling, or reagent syringes
• Improperly fitting pipette tips
• Clogged or imprecise pipetter
• Fluctuations in power supply
44
Random Analytical Error (RE) Components
1. sampling
2. pipetting precision
is caused by:
1. instability of calibration curve
2. differences in recalibration that occur throughout the day,
3. longer term variations in the instrument,
4. small changes in the condition of the calibrator and reagents,
5. changes in the condition of the laboratory during the day, and
6. fatigue of the laboratory staff.
Between-day component of variation (bd)
is caused by:
RE = t