SENSOR & INSTRUMENTATION UNIT-1 NOTES
SENSOR & INSTRUMENTATION UNIT-1 NOTES
The desirable features of sensors are Range /span Errors, accuracy, Non
linearity, Hysteresis, Dead band and Saturation, Output impedance,
Repeatability, Reliability, Sensitivity, Resolution, Frequency Response,
Response time calibration.
Sensor
A sensor is a device that responds to any change in physical phenomena or
environmental variables like heat, pressure, humidity, movement etc. This change
affects the physical, chemical or electromagnetic properties of the sensors which is
further processed to a more usable and readable form. Sensor is the heart of a
measurement system. It is the first element that comes in contact with
environmental variables to generate an output.
Note that a transducer and a sensor are not the same. In the above given example of
thermocouple. The thermocouple acts as a transducer but the additional circuits or
components needed like the voltmeter, a display etc together from a temperature
sensor.
Hence the transducer will just convert the energy from one form to another and all
the remaining work is done by the additional circuits connected. This whole device
forms a sensor. Sensors and transducers are closely related to each other.
Characteristics of Sensors
A good sensor should have the following characteristics
1. High Sensitivity: Sensitivity indicates how much the output of the device
changes with unit change in input (quantity to be measured). For example
the voltage of a temperature sensor changes by 1mV for every 1 oC change in
temperature than the sensitivity of the sensor is said to be 1mV/oC.
2. Linearity: The output should change linearly with the input.
Types of Sensors
Sensor classification
(3) Analog and digital sensor: An analog sensor converts the physical quantity
being measured to analog form (continuous in time). Thermocouple, RTD, Strain
gauge are called analog sensors. A digital sensor produces output in the form of
pulse. Encoders are example of digital sensors.
(4) Inverse sensors: There are some sensors which are capable of sensing a
physical quantity to convert it to other form and also sense the output signal form
to get back the quantity in original form. For example a piezoelectric crystal when
subjected to vibration generates voltage. At the same time when a piezo crystal is
subjected to varying voltage they begin to vibrate. This property make them
suitable to use in microphone and speakers.
Transducer
Transducer is a device which converts one form of energy into another form i.e,.
the given non-electrical energy is converted into an electrical energy.
Types of Transducers
Active transducer is a device which converts the given non-electrical energy into
electrical energy by itself. Thermocouple, Photovoltaic cell and more are the best
examples of the transducers
Passive transducers
Passive transducer is a device which converts the given non-electrical energy into
electrical energy by external force. Resistance strain gauge, Differential
Transformer are the examples for the Passive transducers.
Classification of Transducers
Transducers
Basically, Transducer converts one form of energy into another form of energy.
The transducer, which converts non-electrical form of energy into electrical form
of energy is known as electrical transducer. The block diagram of electrical
transducer is shown in below figure.
As shown in the figure, electrical transducer will produce an output, which has
electrical energy. The output of electrical transducer is equivalent to the input,
which has non-electrical energy.
Types of Electrical Transducers
Mainly, the electrical transducers can be classified into the following two types.
Active Transducers
Passive Transducers
Active Transducers
The transducer, which can produce one of the electrical quantities such as voltage
and current is known as active transducer. It is also called self-generating
transducer, since it doesn‟t require any external power supply.
The block diagram of active transducer is shown in below figure.
Thermistor Transducer:
The resistor, which depends on temperature is called thermal resistor. In short, it
is called Thermistor. The temperature coefficient of thermistor is negative. That
means, as temperature increases, the resistance of thermistor decreases.
Mathematically, the relation between resistance of thermistor and temperature
can be represented as
Where,
R1: is the resistance of thermistor at temperature T10K
R2: is the resistance of thermistor at temperature T20K
β: is the temperature constant
The advantage of Thermistor transducer is that it will produce a fast and stable
response.
Thermocouple Transducer:
Thermocouple transducer produces an output voltage for a corresponding change
of temperature at the input. If two wires of different metals are joined together in
order to create two junctions, then that entire configuration is
called Thermocouple. The circuit diagram of basic thermocouple is shown below
Where,
e : is the emf generated by thermocouple
The above thermocouple circuit can be represented as shown in below figure for
practical applications.
The part of the circuit, which lies between hot & cold junctions including those
two junctions is an equivalent model of basic thermocouple. A PMMC
galvanometer is connected across the cold junction and it deflects according to the
emf generated across cold junction. Thermocouple transducer is the most
commonly used thermoelectric transducer.
As shown in the figure, passive transducer will produce variation in the passive
element in accordance with the variation in the non-electrical input quantity (or
signal).
Resistive Transducer
Inductive Transducer
Capacitive Transducer
Resistive Transducer:
A passive transducer is said to be a resistive transducer, when it produces the
variation (change) in resistance value. the following formula for resistance, R of a
metal conductor.
Where,
ρ: is the resistivity of conductor
l: is the length of conductor
A: is the cross sectional area of the conductor
……………………….(1)
Where,
N: is the number of turns of coil
S: is the number of turns of coil
the following formula for reluctance, S of coil.
……………………(2)
Where,
l: is the length of magnetic circuit
μ: is the permeability of core
A: is the area of magnetic circuit through which flux flows
Substitute, equation 2 in equation 1:-
……………………………(3)
From Equation 1 & Equation 3, we can conclude that the inductance value
depends on the three parameters N, S & μ. So, we can make the inductive
Where,
ε: is the permittivity or the dielectric constant
A: is the effective area of two plates
D: is the effective area of two plates
The capacitance value depends on the three parameters ε, A & d. So, we can make
the capacitive transducers based on the variation in one of the three parameters
ε, A & d. Because, the variation in any one of those three parameters changes the
capacitance value.
Measurement of Displacement
The physical quantities such as displacement, velocity, force, temperature & etc.
are all non-electrical quantities. active transducer converts the physical quantity
into an electrical signal. Whereas, passive transducer converts the physical
quantity into the variation in passive element.
So, based on the requirement we can choose either active transducer or passive
transducer. Let us discuss how to measure displacement by using a passive
transducer. If a body that moves from one point to another point in a straight line,
then the length between those two points is called displacement.
We have the following three passive transducers
Resistive Transducer
Inductive Transducer
Capacitive Transducer
Now, let us discuss about the measurement of displacement with these three
passive transducers one by one.
1. Measurement of Displacement using Resistive Transducer
The circuit diagram of resistive transducer, which is used to measure
displacement is shown in below figure.
The above circuit consists of a potentiometer and a voltage source, VS. We can
say that these two are connected in parallel with respect to the points A & B.
The transformer present in above circuit has a primary winding and two secondary
windings. Here, the ending points of two secondary windings are joined together.
So, we can say that these two secondary windings are connected in series
opposition.
The voltage, VP is applied across the primary winding of transformer. Let, the
voltage developed across each secondary winding is 𝑉𝑆1 and 𝑉𝑆2. The output
voltage, V0 is taken across the starting points of two secondary windings.
Mathematically, the output voltage, 𝑉0 can be written as
V0=VS1−VS2
The capacitor, which is present in above circuit has two parallel plates. Among
which, one plate is fixed and the other plate is a movable one. Due to this, the
spacing between these two plates will also vary. the value of capacitance changes
as the spacing between two plates of capacitor changes.
Therefore, we should connect the body whose displacement is to be measured to
the movable plate of a capacitor. So, whenever the body moves in a straight line,
the spacing between the two plates of capacitor varies. Due to this, the capacitance
value changes.
The potentiometer is also called as pots and it one of the most commonly used
devise for measuring the displacement of the body. The potentiometer is the
electrical type of transducer or sensor and it is of resistive type because it works on
the principle of change of resistance of the wire with its length. The resistance of
the wire is directly proportional to the length of the wire, thus as the length of the
wire changes the resistance of the wire also changes.
What is Potentiometer?
As per the resistance law of the conductor, the resistance of the conductor AC
changes as the length of the wire AC changes. Accordingly the output voltage Vo
Vo/Vs = AC/AB
• If the core is displaced from the central position more in secondary coil 1
than in coil 2, then more emf is generated in one coil i.e. coil 1 than the
other, and there is a resultant voltage from the coils. If the magnetic core is
further displaced, then the value of resultant voltage increases in proportion
with the displacement With the help of signal processing devices such as
low pass filters and demodulators, precise displacement can be measured by
using LVDT sensors.
Optical Encoder
The change in resistance normally has very small value, and to sense that small
change, strain gauge has a long thin metallic strip arrange in a zigzag pattern on a
non-conducting material called the carrier, as shown below, so that it can enlarge
When an object shows physical deformation, its electrical resistance gets change
and that change is then measured by gage.
In this circuit, R1 and R3are the ratio arms equal to each other, and R2 is the
rheostat arm has a value equal to the strain gage resistance. When the gauge is
unstrained, the bridge is balanced, and voltmeter shows zero value. As there is a
change in resistance of strain gauge, the bridge gets unbalanced and producing an
indication at the voltmeter. The output voltage from the bridge can be amplified
further by a differential amplifier.
Most of the strain gauges are made of constantan alloy which cancel out the effect
of temperature on the resistance. But some strain gauges are not of an isoelastic
alloy. In such cases, dummy gauge is used in the place of R2 in the quarter bridge
strain gauge circuit which acts as a temperature compensation device.
Whenever temperature changes, the resistance will change in the same proportion
in the both arms of the rheostat, and the bridge remains in the state of balance.
Effect of temperature get nullifies. It is good to keep voltage low so that the self-
heating of strain gauge could be evaded. Self-heating of gauge depends upon its
mechanical behavior.
Measurement
of pressure using LVDT based diaphragm
The pressure measurement using bourdon tube and LVDT is shown in above
figure. In this the, the bourdon tube act as primary transducer and LVDT which
follows the output of bourdon tube act as a secondary transducer. The bourdon
tube senses the pressure and converts it into a displacement. The free end of
bourdon tube shows this displacement. A cord is used to connect the free end of
bourdon tube to the core of LVDT as shown in figure. When the free end shows
the displacement, the core of LVDT also moves. This movement of core is
proportional to the displacement of free end, which is proportional to the applied
pressure. The LVDT gives analogues output which is a conversion of displacement
into respective emf. This set up is used for measurement of pressure which is
converted into electrical signal by LVDT.
Piezoelectric pressure sensors exploit this effect by measuring the voltage across a
piezoelectric element generated by the applied pressure. They are very robust and
are used in a wide range of industrial applications.
Working principle
A given static force results in a corresponding charge across the sensor. However,
this will leak away over time due to imperfect insulation, the internal sensor
resistance, the attached electronics, etc.
As a result, piezoelectric sensors are not normally suitable for measuring static
pressure. The output signal will gradually drop to zero, even in the presence of
constant pressure. They are, however, sensitive to dynamic changes in pressure
across a wide range of frequencies and pressures.
Function
An internal amplifier makes the sensor simpler to use. For example, it makes it
possible to use long signal cables to connect to the sensor. The amplifier can also
include signal-conditioning circuitry to filter the output, adjust for temperature and
compensate for the changing sensitivity of the sensing element.
The presence of the electronic components does, however, limit the operating
temperature to not much more than 120ºC.
The low frequency response of the sensor is determined by the discharge time of
the amplifier.
Construction
The piezoelectric effect requires materials with a specific asymmetry in the crystal
structure. This includes some natural crystals, such as quartz or tourmaline.
Because the piezoelectric materials are rigid, only a very small deflection of the
material is required to get a usable output signal. This makes the sensors very
robust and tolerant of over-pressure conditions. It also means they respond rapidly
to changes in pressure.
The pressure sensor can be affected by any external force on the piezoelectric
element, for example, by forces caused by acceleration or noise.
Microsensors can be constructed using thin films. Zinc oxide was one of the first
materials used. This has largely been replaced by ceramics made from materials
such as lead zirconate titanate (PZT) because of their larger piezoelectric effect.
Piezoelectric materials are also used in some other types of MEMS sensors. For
example, the inverse piezoelectric effect is used to generate surface acoustic waves
through a diaphragm. The distortion of the surface under pressure can then be
detected by the changes it causes in the waves that are received by another
piezoelectric element.
Design
Piezoelectric pressure sensors are often constructed in a threaded tube (as shown in
the diagram below) to make it easy to mount them in equipment where pressure is
to be monitored. Care is needed when installing these because over-tightening can
affect the output sensitivity.
The effects of thermal shock can be minimised by the design of the enclosure and
mounting the sensor to provide isolation.
Sensitivity
The output is linear over a wide range, typically 0.7 KPa to 70 MPa (0.1 to 10000
psi) with an accuracy of about 1%.
Ceramic sensors are subject to a loss of sensitivity over time. But this is usually
quite small; typically less than 1% per year.
There may also be a small loss in sensitivity when first exposed to high pressure
and temperature. The effects of this can be avoided by cycling the sensor through
the maximum expected pressure and temperature before deploying them.
Applications
The robustness, high frequency and rapid response time of piezoelectric pressure
sensors means they can be used in a wide range of industrial and aerospace
applications where they‟ll be exposed to high temperatures and pressures.
They are often used for measuring dynamic pressure, for example in turbulence,
blast, and engine combustion. These all require fast response, ruggedness and a
wide range of operation.
Their sensitivity and low power consumption also makes them useful for some
medical applications. For example, a thin-film plastic sensor can be attached to the
skin and used for real-time monitoring of the arterial pulse.
Apart from the associated electronics, piezoelectric sensors can be used at high
temperatures. Some materials will work at up to 1,000ºC. The sensitivity may
change with temperature but this can be minimised by appropriate choice of
materials.
Piezoelectric sensors can be easily made using inexpensive materials (for example
quartz or tourmaline), so they can provide a low cost solution for industrial
pressure measurement.