200_-1-
200_-1-
Remote exposure button Signal EXPSW switch and its wiring. Use
does not operate. Sr 74 IOC.
Panel exposure button Signal PNLEXPSW switch and its wiring.
does not operate. Use Sr 74 IOC.
Unit is used in Test mode. The exposure mode selection in the
control panel. Select A or M instead.
Installation. The I/O board jumper X11. Set X11
jumper to OFF or turn S2 to OFF.
Exhibition mode is set when exposure
lights are on but no buzzer is heard
during the exposure.
NOTE!
Detailed patient positioning instructions can be found from OP200 User
Manual.
Sharp image layer is not correct See OP200 / OC200 User Manuals
for patient positioning details
Overexposed image at the end of Check that Ceph LAT-switch
CEPH LAT program senses LAT position - if not - there
isn’t soft-tissue compensation in
Ceph LAT image.
kV display
WARNING!
If the unit is further used, FAIL failure may cause malfunction.
mA display
The mA-display shows the actual numeric failure code. Each failure code
has a unique number, to differ one malfunction from another:
kV MA
Ch 1 to 15
Sy 20 to 33
Er 40 to 46
s display
kV Time display
Ch failure codes can be reset by correcting the reason for the failure code
(eg. changing collimator position).
Ch and Sy failures can be reset by pushing any key in the control panel
(up-down-right-left-OK) or in the patient positioning panel.
Er failures can not be reset. Switch the unit off and on, to test whether the
failure was only temporary.
S 31 S 32 S 33 COLLIMATOR
COL1SW COL2SW COL3SW POSITION
closed closed open Quality Assurance collimator
closed open open Panoramic collimator
closed closed closed Cephalostat TOMOcollimato
collimator: r:
24 x 30 cm AV
or
10 x 12 in AV
or
*open *closed *open 10 x 8 in AH * TOMO
open closed closed 18 x 24 cm or 18 x 24 cm AV
8 x 10 in AV
or
24 x 30 cm AV
or
8 x 10 in AV or
10 x12 in AV or
10 x 8 in AH
open open closed 18 x 24 cm SV 18 x 24 cm SV
or or
8 x 10 in SV 8 x 10 in SV
open open open No valid collimator
2.3.4 Ch 4 COL
2.3.5 Sy 25 AEC
NOTE!
The unit can be used in Manual
exposure mode.
2.3.6 Sy 28 PoC
D 2 / S 24 D 3 / S 25 CARRIAGEPOSITION
CASLIMSW CASMIDSW (viewed from tubehead towards
In0 LED7 In0 LED6 cassette)
closed open left limit (end)
open open left half
open closed right half
closed closed right limit (end)
2.3.7 Sy 30 PoH
Replace Interface
board with Interface
board OT and change
the PAL version. New
board type has
stand-by current to
hold the motor.
Cassette holder stuck, Release the grid lock:
does not move up nor
down. Gear assembly remove X21 on
in gridlock. Interface board, press
cass up or down key.
S 22 S 21 HOLDER
RACKLIMSW RACKMIDSW POSITION
In0 LED2 In0 LED3
closed closed up
open closed upper half
open open lower half
closed open down
1. No respond from
camera because:
is FO LINK ok
communication ok from TxD led in digital I/O blinking
No cephalometric camera camera to PC and from PC to Camera RxD and TxD blinking
present when cephalometric OP?
exposure initiated Check system info for found
CV started cameras
CEc
Is camera in ceph head and
connected properly Camera powered
S 31 S 32 S 33 S QA COLLIMATOR
COL1SW COL2SW COL3SW COL2SW POSITION
closed open open closed Quality
Assurance
collimator
closed open open open Panoramic
collimator
open closed open open Cephalostat
collimator:
2.4.5 Ch 15 bPL
Wiring fault between terminal and CPU PC not connected or POWER OFF
Fiber optic link NOT OK (check LINK_OK
Missing FILT5 from the CPU (done in the The AECFRQ doesn't reach OPD led) Fault on the fiber optic cable or on the
terminal, should give Er43 or Er44 on the CPU optical connector
POWER UP) Terminal brd doesn't sent the frq
Fault on the PAN AEC terminal brd
Fault on the Digital I/O or Core module brd
Fault on the CEPH head brd
NOTE!
Both frequency lines (paecfrq and caecfrq) drive the same digital I/O
board line.
2.4.7 Sy 28 CCd
NOTE!
Sy 25 AEC is generated if signal line to Core Module is not functioning.
1.1.2) Fault on the Check that the fiber Fiber cables C41, C50
fiber optic cable or optic cable and C67.
bad connection connectors are
properly connected.
Do the link test
procedure to the fiber
cables.
1.2) No or interfered Monitor the image
image data from the capture window
camera caused by: whether the image
appears there during
exposure?
1.2.1) Missing Check that all the PAN AEC or CEPH
PPOWER (PAN) or supply voltage LEDs Terminal Board,
CPOWER (CEPH) on are lit on the Camera Camera Supply
Terminal Board Supply Board (PAN) or Board, CEPH Head
on the CEPH Head Board, Cables C63,
Board (CEPH) when C67, C50, C52
the exposure button
is pressed.
1.2.2) Missing PIMAGE Check that the PAN AEC or CEPH
(PAN) or CIMAGE PIMAGE LED or Terminal Board,
(CEPH) on Terminal CIMAGE LED is lit on CEPH Head Board,
Board the corresponding Cables C67, C50, C52
Terminal Board.
1.2.3) Missing Measure the PAN AEC or CEPH
PDETCLK (PAN) or frequency between Terminal Board
CDETCLK (CEPH) on the Terminal Board
Terminal Board test points TP5 and
TP6 with a
multimeter: If a
frequency greater
than 0 is found
during the exposure,
the line is working.
1.2.4) Missing supply Check A2a. Check Cables C47, C51, C63,
voltage on the cables C51, C63. C68. PAN AEC or
camera Check cable C47 CEPH Terminal Board
(PAN) or C68 (CEPH).
Replace Terminal
Board. Replace
Camera.
1.2.5) Missing image Check A2d. Replace PAN AEC or CEPH
data control signals Terminal Board. Terminal Board,
(SS, VV, H/L and DS) Check cable C47 Camera Supply
(PAN) or C68 (CEPH). Board, CEPH Head
Replace camera. Board, cable C47
(PAN) or C68 (CEPH).
PAN or CEPH Camera.
1.2.6) Missing IMAGE Check A2b and A2c. PAN AEC or CEPH
or TDI+/- signals on Replace Terminal Terminal Board,
the PAN camera Board. Replace cable Camera Supply
C47 (PAN) or C68 Board, CEPH Head
(CEPH). Replace Board, cable C47
camera (PAN) or C68 (CEPH).
PAN or CEPH Camera.
1) Fault on the Replace Terminal PAN AEC or CEPH
Terminal Board Board Terminal Board
2.1) Wiring fault Check PAECFRQ (PAN) Cables C67, C52 and
between the Terminal or CAECFRQ and C50)
Board and the Core CEPH_AEC (CEPH)
Module lines from the
corresponding
Terminal Board to the
Core Module.
2.2) Missing FILT5 Check by turning the
from the Core Module OP power OFF and
back ON whether
Er43 and Er44
appear: If not, FILT5 is
OK. See Er43 and 44
errors.
2.3) Fault on the Core Replace the board. Core Module.
Module
3) The AECFRQ Check with Sr 90 CIN
frequency is out of that during the
range (0.8 - 655kHz) exposure is within the
limits (0.8 - 655kHz).
3.1)Misadjusted AEC Do the PAN AEC PAN AEC Terminal
frequency on the adjustment Board.
PAN AEC Terminal procedure (see
Board Installation manual).
3.2) Fault on the Check the image PAN AEC or CEPH
Terminal Board quality - if the image Terminal Board.
is OK (no artifacts)
replace the board.
Otherwise check also
A2a-f.
3.3) High exposure Check that you have Pan Sensor and PAN
values with no object object on the X-ray AEC Terminal Board.
on the X-ray field beam and / or
filtration on the
primary beam
2.4.8 Sy 30 PoC
Sr 74IOC in 0:
Led 2 Led 3
LIMSW MIDSW
2.4.9 Sy 32 PoA
Sr 74 IOC in 4
1 Enter Pr 91 Ins and select nch using the controlpanel left, right and
OK buttons.
2 Exit Pr 91 Ins and select cephalometric lateral (P9) or PA (P10)
projection imaging program.
3 start exposure and monitor the beam alignment signals.
NOTE!
(CEPHLOK, CEPHCOK; CEPHROK) with the 3 center most LEDs of the AEC
density scale on the control panel.
CHECKLIST
• LINK is OK – from the PCI board
– if NOT – check that red light goes to dark fibre / grey receiver chip
in the brd
– check terminal LINK_OK leds
– make ”linktest” to the whole imaging chain / individual brd
• Start CV, select patient – but don’t initiate exposure - yet
• Start DebugView and check communication in steady state:
– Camera found (HWID) and corresponding gainfiles found
• enable image capturing
– ”CASxxyy”- message is OK
– xx defines found new cameras and yy cameras at a system overall
(see also camera LEDs material)
• if selected collimator & program corresponds to the found camera ->
OP goes ready
• Make exposure (with or without x-rays)
– Program label (exp. values) appear on the image capturing screen
title
– short after this image starts to run into window
– after x-rays stopped there should be end values in the Debugview.
After this DICC set timer to 1,5s since last DMA transfer. Based on
this time delay is image end interpreted.
• Exposure button should turn on the powers in the Cam Supply brd
(observe also Camera Power LEDs)
• Image LED is turned on when image layer movements start
• Test pattern values tell if all databits are there (Cam TDI (=clk) LED).
See the chapter TEST PATTERN IMAGE PROCEDURE below.
• Go through all brds and leds that should be on during exposure
If values differ all databits aren't present. Check camera connection and
cabling from the camera to terminal.
Testpattern example image from ceph sensor (panoramic image has only two lower sections):
NOTE!
All signalling can be enabled (except X-rays) in exhibition mode (option
switch). Test mode from ctrl panel only makes movement, but it does not
for instance power the camera.
2.6.2 Ch 6 POS
How is it PAN: When the panoramic mode has been selected the
detected collimator has to be in PAN position. If these conditions are
? not true, the error is generated and the exposure is
prevented.
CEPH: When exposure is activated
NOTE!
If cassette holder vertical movement limit, Pr 56 HLI is
selected “on” and the cassette holder is driven down below
the midpoint by this feature, the unit enters a special state
where the error code is not generated although
RACKMIDSW is inactive. If any of the cassette holder
movement buttons is pressed, this special state is cleared.
2.6.3 Ch 7 ***
2.6.4 Ch 8 PSE
NOTE!
This feature can be disabled when
Pr 59 PSE is set to “OFF”. This error
code has no effect to the unit’s
normal operation.
2.6.5 Ch 9 rEo
2.6.6 Ch 11 PAr
2.6.7 Ch 12 dCC
2.6.8 Ch 16 StP
NOTE!
This can be done in either one or
both heads.
2.6.9 Sy 20 ***
OP200 is not ready for the next exposure. Wait until the unit is
ready or use lower values
if possible. Elapsing
waiting time (***) in
seconds in s-display.
NOTE!
Manual mode might be possible to perform because AEC need more heat
capacity due to unknown exposure values.
2.6.10 Sy 21 HHo
2.6.11 Sy 22 Arc
2.6.12 Sy 23 Inu
2.6.13 Sy 24 FIL
2.6.14 Sy 26 EEP
2.6.15 Sy 27 Por
Unit rotates, but Use Sr 74 IOC. Test the S11 - S14, C11,
microswitches S 11 to S signals, rotate by hand. X9, Core Module
14 may not operate
properly. Signals
ROT1SW to ROT4SW.
Problem with motor Check the control from Interface Board,
control. Rotation Core Module to X16, Core
stepping motor or Interface Board. Check Module, X17,
motor driving circuitry the motor control red M3, X112
in Interface Board may LED’s: all should lit when
not operate properly. unit is rotating. Check
X16and X17
connections.
Problem with motor Check F2 on Power Power Supply
power. Power Supply F2 Supply Board. Board, F2
blown.
Unit rotates, but not Clean the friction
enough friction surface with alcohol.
between the drive Use Sr 80 ro-. Adjust the
wheel and friction spring tension.
surface
Fig 2.2. Rotation angle reference points with Sr 74 IOC LED states
S 14 S 13 S 12 S 11 TUBEHEAD
ROT4SW(I ROT3SW(I ROT2SW ROT1SW POSITION
n5 LED8) n5 LED7) (In5 LED6) (In5 LED5)
2.6.16 Sy 29 PoL
Linear movement, Use Sr 74 IOC. Test the S15 - S17, C11, X6,
but microswitches S signals, rotate fly wheel Core Module
15 to S 17 may not by hand or press switch
operate properly. actuator arms.
Signals LINMIDSW
and LINLIMSW.
Problem with motor Check the control from Interface Board,
control. Linear Core Module to Interface X16, Core Module,
stepping motor or Board. Check the motor X17, M2, X111
motor driving control red LED’s: all
circuitry in Interface should lit when unit is
Board may not rotating. Check X16 and
operate properly. X17 connections.
LINLIMSW
LINMIDSW
S 15 is for the mid position of the linear movement, also indicating front
and rear segments of the movement. It is used to set OP200 rotating unit
for patient positioning (Programs 1 to 4 and 6 to 9) and it serves as a
reference point for TMJ pointer movements in TMJ imaging (Programs 6 to
9) and linear tomography (Programs 11 and 12).
2.6.17 Sy 31 PoU
Detent pieces. Check the top detent Short & long detent
Microswitches trigger pieces: longer one piece
in wrong order. higher than short
one.
No motor power. Check the Power X16, Interface Board,
12VDC missing. Supply & Interface X15, Power Supply
Problem with Board LED’s. Board
Interface Board.
Motor operates all Motor wiring. Motor M1, C3, X16,
the time. Motor or Interface Board relay. Interface Board
diving circuitry If problem with panel
shorted. Positioning key, it gives first Er 45
panel key problem. INP.
Clutch. Motor Check the motor Clutch at column top
operates, but slow or clutch tension. Adjust
no carriage with 14mm wrench.
movement
Mechanical. Counter Check the amount of Trim weights
weight trims are too trims: OP200 4 smalls
many/few or they are & 7 big trims, OC200
touching the column 4 small & 23big ones.
interior. sliding
carriage roller
movements
Mechanical. Problem S/N 70xxx: adjust Sliding carriage.
with sliding carriage with shim plates.
rollers. From S/N 71xxx:
adjust the roller slack.
There are two position indicators for the vertical carriage movement.
These microswitches are located inside the rear support assembly, at the
rear of the column unit. The height of the vertical carriage is detected by
two short detent pieces located in the groove of the column. These pieces
are adjusted at the factory so that they will stop the carriage 10 - 20 mm
before the mechanical limits.
The height of the cassette holder can be limited by lowering the cassette
holder to the half way up position. When this feature is activated, cassette
holder is lowered when S4 is actuated by the long detent piece located in
the right groove of the column (looking from behind the column). This
piece is adjusted at the factory so that the cassette holder always stays
below the height of the column. This feature is activated by user program
“Pr 56 HLI” -> “on”. Note that the longer detent piece is also used to
distinguish the difference between upper and lower limits.
ZLIMSW
ZMIDSW
2.6.21 Er 43 ***
NOTE!
When the unit is connected to 230V line with 110V settings, fuses normally
blow before this error is displayed.
NOTE!
If +34VDC is missing also +5VDC from PAN AEC terminal is missing
resulting Er 44 Fil. This is possible by pressing emergency stop switch.
2.6.22 Er 44 FIL
Cables (check
wiring diagram)
Problem with wiring Power - Core Module - C67, X4, X38
Filament - PAN AEC
Problem with Check Filament control
operating voltages. Board, or PAN AEC
Signal FILT5. Terminal Board +5V signal.
Broken Core Module. Replace Core Module.
Optoisolator. Wire.
Power Supply Board Power off. Check the switch Power Supply
switches SW1-SW4 in positions. Check the main Board
incorrect position. fuses.
Error message
sometimes barely
seen.
2.6.23 Er 45 InP
2.6.24 Er 46 PAy
All OP200 operations have been See Service program Sr 71 PAy for
halted by software. details. Increase the limit or
disable this feature.
All LED indicators can be identified from the schematics and printed
circuit boards by the name of the signal and the component number of
the LED: eg. LED for + 5 V voltage in the Core Module is marked H1 +
5 V.
All supply voltages are indicated by green LED’s, and the most important
or critical signals are indicated by red LED’s in the printed circuit boards.
When the LED is lit, it means that the supply voltage is available or that
the signal is active.
All test points can be identified from the schematics and printed circuit
boards by the name of the signal and the TP number: eg. the test point for
PROJLIT signal in the Interface Board is marked TP1 PROJLIT.
Instrumentarium Dental
P.O.Box 20, FI-04301 Tuusula, Finland
Tel. +358 45 7882 2000
Fax +358 45 7882 2506
Americas:
Instrumentarium Dental Inc.
Milwaukee, Wisconsin, U.S.A.
Tel. 800 558 6120
Fax 414 481 8665