2019a Investigation of Collapsed-cone Algorithm Accuracy
2019a Investigation of Collapsed-cone Algorithm Accuracy
Department, Students
Algorithm Accuracy in Small Fields and Research Committee,
Kermanshah University
Heterogeneous Environments of Medical Sciences,
Kermanshah, Iran
2
MD, Radiation oncolo-
gist, Faculty of Medicine,
Hamedan University
Razieh Zaghian1 , Abdolazim SedighiPashaki2, Abbas Hagh- of Medical Sciences,
parast3* , Mohammad Hadi Gholami4, Mahdi Mohammadi1 Hamedan, Iran
3
PhD, Department of
Medical Physics, Faculty
of Medicine, Kerman-
ABSTRACT shah University of Medi-
cal Sciences, Kerman-
Background: The use of small fields has increased by the emergence of advanced shah, Iran
radiotherapy. Dose calculations of these fields are complex and challenging for many 4
PhD, Medical Physicist
reasons such as lack of electrical equilibrium even in homogeneous environments, of Radiotherapy, Mah-
and this complexity will increase in presence of heterogeneity. According to the dieh Radiotherapy and
importance of delivery the accurate prescription dose to the target volume in the pa- Brachytherapy Charity
tient’s body, the dose calculation accuracy of used commercial algorithms in clinical Center, Hamedan, Iran
treatment planning systems (TPS) should be evaluated.
Objective: The present study aims to evaluate the accuracy of Collapsed-cone
dose measurement algorithm in Isogray treatment planning system.
Material and Methods: In this analytical study, the measurements were
made in tissue equivalent solid water phantom with lung and bone heterogeneities
by Pinpoint dosimeter (0.015 cm3 sensitive volume) in several radiation fields (1×1
to 5×5 cm2). The phantoms were irradiated with 6, 10 and 18 MV photon beams and
finally, the results of experimental calculations were compared with treatment plan-
ning outputs.
Results: In all setups, the maximum deviation occurred in the field of 1×1 cm2.
Then, the maximum deviation was observed for 2×2 cm2 field size; however, it was
up to 5% for homogeneous water phantom and lung heterogeneity. In 3×3 cm2 and
larger fields, there was a good agreement between the results of the TPS and experi-
mental dosimetry. The maximum deviation was observed in water-bone heterogene-
ity.
Conclusion: This algorithm was able to pass the standard audit criteria, but it is
better to be used more cautiously in bone heterogeneity, especially in low energies.
Citation: Zaghian R, SedighiPashaki A, Haghparast A, Gholami MH, Mohammadi M. Investigation of Collapsed-cone Algorithm Accuracy in
Small Fields and Heterogeneous Environments. J Biomed Phys Eng. 2021;11(2):143-150. doi: 10.31661/jbpe.v0i0.1121.
Keywords
Radiotherapy; Dosimetry; Small Field; Algorithms; Phantoms, Imaging; Ab-
*Corresponding author:
sorbed Dose
Abbas Haghparast
Department of Medi-
cal Physics, Faculty of
R
Introduction Medicine, Kermanshah
University of Medical
adiotherapy with ionizing radiation is one of the most frequent Sciences, Kermanshah,
techniques for treating cancers [1, 2]. Radiotherapy can lead to Iran
the DNA damage of cancer cells and stop cell dividing or growth, E-mail: abbas.haghpar-
[email protected]
or prevent them from returning. Irradiation can also damage healthy
Received: 22 January 2019
cells, but the normal cells are more radiation-resistant. However, the Accepted: 16 February 2019
Figure 1: Solid water phantom with a) lung and b) bone heterogeneity. The placement of dosim-
eter is shown as a small black circle.
dosimeter can be used in air, water, and solid lent material (solid water slabs) from top and
phantom with a polarity effect less than 2%. 5 cm below (Figure 1a). To evaluate the inho-
The thickness of the sensitive pinpoint wall mogeneity of soft tissue-bone, 5 cm solid wa-
is 0.57 mm of PMMA with a density of 1.19 ter slabs were used in addition to 3 cm PTFE
g/cm3 and 0.90 mm graphite and a density of equivalent to bone and again 5 cm of solid wa-
1.85 g/cm3. The central electrode is made of ter (Figure 1b). After heterogeneity, the Pin-
aluminum with a diameter of 0.3 mm. point dosimeter was placed at 0.5 cm in solid
water slab (i.e.at 8.5 depth for the bone and
Radiation to the heterogeneous 10.5 cm for the lung). The location of dosim-
soft tissue-lung and soft tissue- eter is shown in Figure 2.
bone phantoms Then the phantoms were irradiated with an
In order to investigate the effect of soft tis- Elekta linear accelerator (Synergy) with three
sue-lung heterogeneity, 5cm Cork equivalent photon beams energies of 6, 10 and 18 MV
was sandwiched between 5 cm tissue equiva- and measurements made in radiation fields
Figure 2: The location of pinpoint is shown as a black line at the solid water slab’s groove and
in the center of it.
=%E
( DM − DT ) ×100 gies studied, a significant decrease of deviation
DM was observed with increasing field size to 3×3
cm2, but no significant change was observed
In which DM is calculated dose empirically for larger field sizes. Deviation increased with
and DT is the result of TPS calculations. incremental energy in 1×1 cm2. Equal to more
than 2×2 cm2 fields, there was no significant
Results difference between the relative errors in differ-
Figure 4 illustrates the relationship between ent energies but founded the increasing mag-
TPS relative error percentage and field size for nitude of error from 10-18 MV for 2×2 cm2
three studied energies in all setups. field size. In the homogeneous water environ-
ment, at minimum and maximum deviations
Homogeneous water environment were 0.2% and 14% for energy 18 MV and
In this environment, for each of the 3 ener- in 3×3 and 1×1 cm2 field sizes, respectively.
In the mentioned environment, deviation was
negligible and up to 2% with the exception of
1×1 and 2×2 cm2.
Water/Lung heterogeneity
The results obtained from this environment
were similar to the homogeneous water media;
thus, the deviation decreased with the increase
of field size. But it increased in 18 MV with
increasing the field size from 3×3 to 4×4 cm2.
This increasing was still with an acceptable
level of error. From another view, the energy
dependence of the heterogeneity was similar
to the homogeneous water phantom. In this
environment, the minimum error occurred in
4×4 and 5×5 cm2 field size with 6 MV photon
energy, and the maximum error occurred in
1×1 cm2 field size with 18 MV photon energy
and their values were 0.03% and 22.2%, re-
spectively. In this environment, deviation was
up to 2% with the exception of 1×1 and 2×2
cm2.
Water/bone heterogeneity
In this heterogeneity, which was compared
to the previous two environments, there was
a greater difference between the calculations
Figure 4: Curves of relative error percent- of TPS and experimental measurements. The
age of calculated Collapsed-cone algorithm lowest and the highest deviation occurred in
for every field size in three studied media a) 3×3 cm2 and 1×1 cm2 fields for energy 18 MV
solid water phantom with lung heterogene- similar to the homogeneous water environ-
ity b) solid water phantom with bone hetero- ment and deviations were 1.7% and 17.5%,
geneity and c) homogenous water phantom. respectively. Although there was no signifi-
cant difference in this energy for the magni- measuring the absolute dose. The signal quan-
tude of error in 3×3 cm2 and larger fields, the tum noise of the dosimeters depends on the
minimum deviation occurred in the energy of detector`s material, in addition to the sensitive
6 MV (2.6%) in 1×1 cm2 field size. Regard- volume of the detector. This noise is very low
less of this data, deviation was reduced with in pinpoint dosimeter, even rather than diode
increasing the field size, but there was no dif- type E with a sensitive volume of 0.03 mm3.
ference between 4×4 cm2 and 5×5 cm2 field As a result, the precision and speed of the Pin-
sizes for 6 MV photons. For energy of 10 and point dosimeters are very high in dose mea-
18 MV, it decreased by increasing the size to surement [17]. Furthermore, in a similar study
3×3 cm2, and a slight increase was observed in carried out by Alagar [4], a scintillator plastic
larger field sizes than 3×3 cm2. In all fields, a dosimeter was used, which has a high level
deviation reduction occurred with increasing of noise compared to chambers; the highest
energy. quantum noise can be expected from these de-
tectors and dosimetry measurement is a time-
Discussion consuming performance.
In this study, the accuracy of Collapsed-cone This study was carried out in 1×1 to 5×5 cm2
algorithm dose calculating from Isogray TPS field sizes. For all setups, the greatest devia-
was compared with the experimental results tion was observed in the 1×1 cm2 field size,
obtained using the pinpoint dosimeter in the except for bone heterogeneity at 6 MV ener-
homogeneous water phantom, heterogeneous gy, which was consonant with the results of
water-lung and water-bone phantoms exposed Alagar’s study [4]. The maximum deviations
to photon beam radiation with 6, 10 and 18 of the 1×1 cm2 field size for water-lung and
MV in 1×1 to 5×5 cm2 field sizes. In most sim- water-bone heterogeneity were 22% and 17%,
ilar studies, the Monte Carlo simulation has respectively, and for the homogeneous water
been used to verify the accuracy of dose cal- environment was 14%. The most important
culation for clinical measurement algorithms reason for this significant error could be in-
[5, 14, 15], although the Monte Carlo simula- appropriateness of the pinpoint dosimeter in
tion can’t be considered as a golden standard field sizes smaller than 2×2 cm2. The effect of
without experimental confirmation [16]. charged particle disequilibrium in this small
Due to the absence of a single protocol for field should not be ignored.
absolute dose measurement in small fields, the Although the magnitude of error decreased
complexity of direct measurement and com- significantly in 2×2 cm2 field size, the final
paring with calculations of the TPS, most of value was still significant. This behavior can
the studies have investigated the accuracy of be described based on the electrical disequi-
algorithms with measuring the percentage of librium effect. Deviation in this field for 6 and
depth doses. The use of absolute dosimetry 10 MV of photon energy in the lung and water
instead of the relative dosimetry for compar- heterogeneity was relatively lower and about
ing the TPS results and also Pinpoint ionizing 3.5%. While for 18 MV energy, it was 9.1%
chamber (an ideal chamber for small field do- and 5.9% in the lungs and water, respectively.
simetry) led into difference between this study Due to the enhancement of the photon energy,
from other studies. the range of the secondary scatter electron has
Due to the challenges of small-field dosime- increased and the effect of electrical disequi-
try, selecting a suitable detector for dose mea- librium, which is usually not considered in the
surement in these fields is very necessary. Ac- treatment planning algorithms, leads to more
cording to study carried out by J. U. Wuerfel,et deviation.
al., ionizing chambers are the best choice for Generally, for field sizes larger than 2×2