RX-DC Hyper. Operating Manual
RX-DC Hyper. Operating Manual
REF 97050900
Rev. 03
2018-11
RX DC
EN
2 OPERATOR'S MANUAL EN
ITALIANO
Contents
1. GENERAL WARNINGS ............................................................................................................................................. 5
1.1. SYMBOLS ............................................................................................................................................................ 5
1.2. STANDARDS AND REGULATIONS .................................................................................................................... 6
1.3. INTENDED USE ................................................................................................................................................... 6
1.4. CLASSIFICATION ................................................................................................................................................ 6
1.5. ENVIRONMENTAL CONDITIONS ....................................................................................................................... 7
1.6. WARRANTY ......................................................................................................................................................... 7
1.7. PROTECTION AGAINST RADIATION ................................................................................................................ 8
1.8. ELECTROMAGNETIC SAFETY ........................................................................................................................... 9
2. DESCRIPTION OF THE X-RAY DEVICE ............................................................................................................... 12
2.1. INSTALLATION TYPE ........................................................................................................................................ 12
2.2. TYPE OF X-RAY HEAD ..................................................................................................................................... 15
2.3. HANDHELD ........................................................................................................................................................ 16
3. SWITCHING ON AND OFF THE X-RAY DEVICE .................................................................................................. 17
4. HANDHELD FUNCTIONS ....................................................................................................................................... 18
4.1. HANDHELD DISPLAY FUNCTIONS.................................................................................................................. 19
4.2. USE OF HANDHELD ......................................................................................................................................... 20
4.3. CHECKING THE PARAMETERS....................................................................................................................... 21
4.4. FACTORY SETTINGS ....................................................................................................................................... 22
5. USE OF THE X-RAY DEVICE................................................................................................................................. 23
5.1. PATIENT POSITIONING .................................................................................................................................... 23
5.2. POSITIONING THE X-RAY HEAD ..................................................................................................................... 23
5.3. POSITION OF THE X-RAY PLATE OR SENSOR ............................................................................................. 24
5.4. SETTING THE EXPOSURE MODE AND TIME ................................................................................................. 25
5.5. PROCEDURE TO BE FOLLOWED WHEN TAKING THE X-RAY ..................................................................... 27
6. ADVANCED OPTIONS ........................................................................................................................................... 28
6.1. SETTING THE OPERATING MODE .................................................................................................................. 29
6.2. SETTING TYPE OF MOVABLE COLLIMATOR ................................................................................................. 30
6.3. RESTORING FACTORY SETTINGS ................................................................................................................. 30
7. ERROR MESSAGES .............................................................................................................................................. 31
8. PERIODIC MAINTENANCE .................................................................................................................................... 34
9. CLEANING AND DISINFECTION ........................................................................................................................... 35
10. DISPOSING THE EQUIPMENT WHEN NO LONGER USED ................................................................................ 36
11. TECHNICAL DATA .................................................................................................................................................. 37
11.1. X-RAY TUBES ............................................................................................................................................... 39
11.2. TECHNICAL FACTOR MEASURE ................................................................................................................ 40
12. DIMENSIONAL CHARACTERISTICS ..................................................................................................................... 41
13. IDENTIFICATION PLATES ..................................................................................................................................... 48
14. TIMES/SENSITIVITY CHARTS ............................................................................................................................... 50
15. NOMINAL DOSE EMISSION VALUES ................................................................................................................... 51
16. INSPECTION AND MAINTENANCE ....................................................................................................................... 61
16.1. USER INSPECTION ...................................................................................................................................... 61
16.2. TECHNICAL MAINTENANCE ........................................................................................................................ 62
EN OPERATOR'S MANUAL 3
4 OPERATOR'S MANUAL EN
1. GENERAL WARNINGS
These instructions explain how to correctly use the RX DC x-ray unit. Please carefully read this manual before using
the device.
NOTE: This manual does not specify all the obligations and warnings for possessing a source of ionising
radiation as each country has its own laws. Only the most common ones shall be mentioned and this means
that it is the user’s responsibility to check local standards and observe the relevant laws.
The contents of this publication are valuable trade secrets and must not be given to third parties, stored, copied,
reproduced, disclosed or transferred in any manner (via computer, photocopies, translations or other means) without
the prior written consent of the Manufacturer.
The Manufacturer has a company policy of continual development. Therefore, some of the instructions, specifications
and figures given in this manual may slightly differ from the purchased product. The manufacturer reserves the right to
make changes to this manual without giving prior notice.
The original text is in Italian; this is a translation from the original in Italian.
The Manufacturer’s website contains a list of authorised agents.
1.1. SYMBOLS
Type of protection against direct and indirect contact: Class I. Level of protection against direct and
indirect contact: TYPE B.
WARNING! Failure to observe may result in equipment damage or injury to the user and/or patient.
NOTE: Indicates information that is especially important for the user and/or assistant.
Alternating current.
On.
Off.
Ionising radiations.
Operating instructions. Consult the enclosed documentation before using the device.
EN OPERATOR'S MANUAL 5
It is necessary to refer to the user manual.
Pushing prohibited.
Stepping prohibited.
Manufacturer.
Date of manufacture.
Technical Standards:
IEC 60601-1:2005
IEC 60601-1-2:2014
IEC 60601-1-3:2008
IEC 60601-2-65:2012
IEC 60601-1-6:2010
IEC 62366:2007
The CE marking certifies compliance of the product as described by Medical Device Directive 93/42/EEC and
subsequent amendments.
This x-ray unit is designed for use in the dental surgery to make endo-oral x-rays for diagnostic purposes.
This equipment can be used to produce traditional x-rays developed using chemicals or, alternatively, it can be used
with digital x-ray sensors.
1.4. CLASSIFICATION
6 OPERATOR'S MANUAL EN
1.5. ENVIRONMENTAL CONDITIONS
1.6. WARRANTY
The Manufacturer stands behind its products warranting safety reliability and performance. The warranty is valid only
under the following terms:
- Closely observe the conditions specified in the warranty certificate itself.
- The equipment is only to be used as instructed in this manual.
- Equipment installation, expansion and technical support must be performed exclusively by personnel authorised
by the Manufacturer to carry out these operations.
- Do not open the device covers: installation, repairs and in general all the operations that require opening the device
must be carried out exclusively by technicians authorised by the Manufacturer.
- The equipment is to be installed in rooms that follow the requirements specified in paragraph 1.2.2. “Environmental
conditions”.
- The room where the x-ray unit is installed must comply with official regulations regarding protection against
radiation in the country where the equipment is used.
SAFETY WARNINGS.
- If any person who is not an authorised technician changes the product in any way by replacing parts or
components with other ones not used by the Manufacturer, they shall assume responsibility for the
product.
- Do not forget to turn off the main switch on the equipment before leaving the surgery.
- The equipment is not protected against liquid penetration (risk of electrocution).
- The equipment is not suitable for use in the presence of a mixture of flammable anaesthetic gas with
oxygen or nitrous oxide.
- This equipment must be stored properly so that it is kept in top working order at all times.
- Use in the dental surgery or nearby of electric scalpels or other electric devices that do not comply with
standard I.E.C. 60601-1-2 may cause electromagnetic or other types of interferences resulting in
equipment malfunctions. In these cases shut off the power supply to the equipment beforehand.
- The Manufacturer shall not be held responsible (under civil and criminal law) for misuse, carelessness or
improper use of the equipment.
- The equipment may only be used by authorised and adequately trained staff (dentists and paramedics).
- The user must be present at all times when the equipment is turned on or ready for start-up. In particular,
never leave the equipment unattended in the presence of children/the mentally disabled or other
unauthorised personnel in general.
- If the x-ray equipment is damaged or oil leaks, do not use the equipment and contact customer service
immediately.
EN OPERATOR'S MANUAL 7
1.7. PROTECTION AGAINST RADIATION
8 OPERATOR'S MANUAL EN
1.8. ELECTROMAGNETIC SAFETY
The device is intended for use in environments recognised as professional health facilities, as described in IEC 60601-
1-2:2014. The device belongs to CISPR 11 Class A Group 1 and complies with immunity test levels specified by IEC
60601-1-2:2014 for professional health facilities.
Before using any electronic device in health facilities, always check that it is compatible with the other equipment
present.
Even if the device complies with standard IEC 60601-1-2, it is recommended not to use it near life-support
equipment (e.g.: pacemakers or cardiac stimulators). For further information, see the equipment instructions
for use.
Use of this equipment adjacent to or stacked with other equipment should be avoided, because it could result
in improper operation. If such use is necessary, this equipment and the other equipment should be observed
to verify that they are operating normally.
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the device, including cables specified
by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Do not subject the device to strong electromagnetic disturbances. These disturbances could degrade the
essential performance of the device.
The emission characteristics of this equipment make it suitable for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required)
this equipment might not offer adequate protection to radio-frequency communication services. The user
might need to take mitigation measures, such as relocating or re-orienting the equipment.
EN OPERATOR'S MANUAL 9
Guidance and Manufacturer's declaration - Electromagnetic emissions
RX DC is designed to operate in the specified electromagnetic environment. The customer or the user of RX DC
must ensure its use in an electromagnetic environment with the following features:
Emission test Conformity Electromagnetic Environment
RX DC uses RF energy only for its internal operation.
RF emissions
Group 1 For this, the RF emissions are very low and do not
CISPR 11
interfere with the electronic devices nearby.
RX DC must be used only by adequately trained
personnel (dentists and paramedics). RX DC may
cause radio interferences or disturb the operation of
RF emissions
Class A the nearby equipment. It may be necessary to adopt
CISPR 11
countermeasures, such as re-orienting or moving RX
DC or shielding the installation site.
10 OPERATOR'S MANUAL EN
Guidance and Manufacturer's declaration - Electromagnetic immunity
RX DC is designed to operate in the specified electromagnetic environment. The customer or user of RX DC must
ensure that it is used in such environment.
Immunity test IEC 60601-1-2 Level of Electromagnetic Environment
Test level conformity
The RF communication devices (portable
and mobile) must not be used at a distance
from RX DC and its components, including
cables, lower than the recommended
distance using the corresponding equation
applicable to the transmitter frequency.
Recommended distance.
3 V/m d = 1.2 x √P 80 MHz to 800MHz
Radiated RF IEC 60601-1-2
From 80 MHz to 2.7 d = 2.3 x √P 800 MHz to 2.7GHz
EN 61000-4-3 Test level
GHz
3V d = 1.2 x √P
from 150 kHz to 80
Conducted RF IEC 60601-1-2
MHz
EN 61000-4-6 Test level
6V
ISM frequencies
Where P is the maximum output power of
the transmitter in Watt (W) according to the
transmitter Manufacturer, and d is the
recommended distance in metres (m).
The field intensity of the fixed RF
transmitters, determined based on an
electromagnetic site, could be lower than
the conformity level in each frequency
interval.
Near the equipment with the following
symbol interferences can be caused:
Recommended distance between the RF portable and mobile communication devices and RX DC.
RX DC is intended for use in electromagnetic environment where RF irradiated disturbances are controlled. The
customer or the user of the RX DC can prevent electromagnetic interferences by ensuring a minimum distance
between RF mobile and portable (transmitter) communication devices and RX DC as shown below, according to
maximum power output of the communication devices.
Distance according the transmitter frequency
Transmitter maximum nominal
(m)
output
150KHz to 80MHz 80KHz to 800MHz 800KHz to 2.7MHz
(W)
d = 1.2 x √P d = 1.2 x √P d = 2.3 x √P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters having a maximum nominal output power not listed above, the recommended distance d in metres
(m) can be determined using the corresponding equation applicable to the transmitter frequency where P is the
maximum output power of the transmitter in Watt (W) according to the transmitter Manufacturer.
Note:
(1) At 80MHz and 800MHz it is necessary to apply the distance defined for the highest frequency interval.
(2) These guidelines cannot be applicable to all situations. The electromagnetic propagation is influenced by the
absorption and reflection of structures, objects and people.
EN OPERATOR'S MANUAL 11
2. DESCRIPTION OF THE X-RAY DEVICE
The x-ray unit is available in several versions, which differ in type of installation, x-ray head and handheld.
It is possible to identify the different versions via the REF on the nameplate.
WALL-MOUNTED VERSION
a. X-ray generator.
Constant potential high frequency x-ray generator.
b. Removable collimator (cone).
The generator can work with different types of collimator that are automatically recognised:
- 8” cylindrical COLLIMATOR (incorporated in the generator): minimum skin/focus distance 20cm and 60mm output
beam.
- removable 12” rectangular COLLIMATOR (only for MRPU****S): minimum skin/focus distance 30cm and 45x35
mm output beam (rectangular collimator attached).
- removable 12” round COLLIMATOR (standard for MRXU****S, optional for MRPU****S): minimum source/skin
distance 30cm and diameter of collimator output beam 55mm (with collimator attached).
The following rectangular collimators to be attached to a 12” round collimator are also available as optionals:
- rectangular COLLIMATOR 22x35 mm
- rectangular COLLIMATOR 31x41 mm.
c. Focus spot.
d. Double pantograph arm.
e. Extension arm.
The extension arm is available in three length versions: 40 cm (15.7"), 60 cm (23.6") and 90 cm (35.4").
f. Handheld.
The handheld can be placed either near the control unit or in a remote position. As a result, the doctor can move
conveniently around the room and move out of the area where x-rays are emitted.
g. Handheld holder.
h. Control unit.
I. Main switch.
12 OPERATOR'S MANUAL EN
INSTALLATION INTEGRATED IN DENTAL UNIT
NOTE: This section applies only to models RX DC REF: MRPU****I
(Character * can be any alphanumeric value)
a. X-ray generator.
Constant potential high frequency x-ray generator.
b. Removable collimator (cone).
The generator can work with different types of collimator that are automatically recognised:
- 8” cylindrical COLLIMATOR (incorporated in the generator): minimum skin/focus distance 20cm and 60mm output
beam.
- removable 12” rectangular COLLIMATOR (only for MRPU****I): minimum skin/focus distance 30cm and 45x35
mm output beam (rectangular collimator attached).
- removable 12” round COLLIMATOR (optional for MRPU****I): minimum source/skin distance 30cm and diameter
of collimator output beam 55mm (with collimator attached).
The following rectangular collimators to be attached to a 12” round collimator are also available as optionals:
- Rectangular COLLIMATOR 22x35 mm
- Rectangular COLLIMATOR 31x41 mm.
c. Focus spot.
d. Double pantograph arm.
e. Extension arm.
The only size of the extension arm is 30 cm (11.8”).
f. Handheld.
The handheld can be placed either near the control unit or in a remote position. As a result, the doctor can move
conveniently around the room and move out of the area where x-rays are emitted.
g. Handheld holder.
NOTE: when integrating the x-ray unit in the dental unit, the integration must be suitable and in line with the
provisions of IEC60601-1. For installation instructions and details, refer to the technical manual.
EN OPERATOR'S MANUAL 13
MOBILE STAND INSTALLATION
WARNING!
Never move the mobile stand x-ray unit without first securing the support arm with the special strap.
WARNING!
Moving the mobile stand x-ray unit without using the handles can cause the device to fall, unbalance or tip
over. Be very careful and always use the handles.
WARNING!
During the movement of the mobile stand, pay attention to the presence of steps and / or horizontal
obstacles as they may cause a situation of instability and / or tip over the cart. If you wish to move the x-ray
over a small obstacle, gently tilt the base by pressing with your foot near the rear wheels.
WARNING!
The tray can hold a maximum of 5kg.
WARNING!
Do not step on the mobile stand or parts of it.
To stop the mobile stand in the desired position, lock the wheel brakes by pressing the lever down. Upwards to unlock.
14 OPERATOR'S MANUAL EN
2.2. TYPE OF X-RAY HEAD
The mechanical fitting with which the x-ray head is connected to the pantograph arm is a ball joint. The ball is provided
with a mechanical brake to allow the x-ray head to maintain the position set by the operator.
The generator can freely rotate on the horizontal plane. On the vertical plane, the upward rotation is limited by a
mechanical end stop.
X-RAY GENERATOR INDICATOR LIGHT FOR X-RAY UNIT WITH BALL JOINT:
The x-ray generator is provided with an indicator light (B) that signals the unit status.
Key to colours:
- violet > x-ray unit on (regular condition)
- flashing violet > stand-by (low consumption)
- blue > x-ray unit on – head locked
- yellow > x-ray being emitted
- red > fault
The mechanical fitting with which the x-ray head is connected to the pantograph arm is an arch. The arch allows the x-
ray head to maintain the position set by the operator.
The generator is can freely rotate both on the horizontal and vertical plane. The rotation is limited by suitable
mechanical stops.
EN OPERATOR'S MANUAL 15
2.3. HANDHELD
The handheld is turned on by pressing any key, except for the one for x-ray emission.
WIRELESS HANDHELD
This handheld uses a wireless connection to communicate with the x-ray unit.
The wireless communication complies with specifications IEEE 802.11 b/g/n. The connection is protected through
cryptography and no other wireless product, except for the handheld, can connect to the unit.
Handheld batteries:
- Type: 2 x AA - Alkaline 1.5V.
The wireless communication can be affected by interferences caused by other equipment such as:
- WiFi routers
- surveillance systems
- Audio/Video repeaters
- cordless phones
With wireless communication interferences proceed as indicated in the technical manual (§10.4).
With a too high level of interferences the handheld cannot guarantee a correct operation and it deactivates the wireless
communication. In these latter cases it is preferable to use a wired handheld.
WIRED HANDHELD
This handheld uses a cable connection to communicate with the x-ray unit.
16 OPERATOR'S MANUAL EN
3. SWITCHING ON AND OFF THE X-RAY DEVICE
The control unit is turned on and off from the main switch
(A).
The switch lights up to indicate that the control unit is
powered.
The control unit is turned on and off from the main switch
(A).
The switch lights up to indicate that the control unit is
powered.
NOTE: The technical specifications of the switch are outlined in paragraph 1.5.
Whenever turned on, the equipment performs an operational test that takes a few seconds. A beep is emitted
at the end of the test.
NOTE: The exposure time and the parameters displayed on the handheld when the unit is turned on are the
last ones set before the control unit was turned off.
If the control unit is left untouched for a few minutes, it will go into standby mode. Simply press any key on
the control panel to reactivate it.
EN OPERATOR'S MANUAL 17
4. HANDHELD FUNCTIONS
The handheld is turned on by pressing any key, except for the one for x-ray emission.
A buzzer rings to confirm that the unit has been turned on. The unit will be in the standard configuration and it will start
searching for the base it works with.
If the base is off, the handheld will not indicate the field or the status “ready”. If the base is later turned on, the handheld
will detect it within thirty seconds or by pressing any function key on the push-button panel.
NOTE: To optimise the range of the handheld while it is being used, keep it away from walls and metal
instruments and, above all, do not cover its built-in antenna on top of the screen. In addition, performance
may be reduced if the handheld is moved too quickly while x-rays are being taken. Error E 31 may be displayed
if out of range problems occur.
WIRELESS HANDHELD
NOTE: This section applies only to models RX DC REF: MRPU***W*, MRXU***W*
(Character * can be any alphanumeric value)
Battery fully charged (no symbol appears in the area that shows the battery charge level).
Battery half-charged.
Battery charge level low or almost dead (causing the handheld to automatically shut off).
NOTE: The batteries should be removed from the handheld if it is not going to be used for an extended period.
WIRED HANDHELD
NOTE: This section applies only to models RX DC REF: MRPU***C*, MRXU***C*
(Character * can be any alphanumeric value)
18 OPERATOR'S MANUAL EN
4.1. HANDHELD DISPLAY FUNCTIONS
EN OPERATOR'S MANUAL 19
4.2. USE OF HANDHELD
As illustrated in the figure below, the handheld has four function keys and a single x-ray emission key.
Increases the exposure times in steps Increases the scroll speed of the values in
according to the set scale. increasing order.
Decreases the exposure times in steps Increases the scroll speed of the values in
according to the set scale. decreasing order.
NO EFFECTS ARE OBTAINED IF THE KEY Starts x-ray exposure (the button has to be held
IS PRESSED FOR LESS THAN A SECOND. down throughout the x-rays emission, “dead man”
function).
NOTE: “Dead man” function: the system that starts x-ray exposure with the dedicated key on the wireless
handheld allows x-rays to be emitted only when the user presses and holds down the exposure key. X-ray
emission will stop if the key is released ahead of time.
20 OPERATOR'S MANUAL EN
NOTE: The function related to pressing the key briefly is performed by pressing the key which will activate
the function assigned to it. On the other hand, to perform the function carried out when the key is held down
longer, press the key until the relative function is started. The buzzer will beep shortly to signal that the function
has been activated.
NOTE: Warm-up: When the equipment has not been used for a prolonged period (more than 3 months) or
when turned on for the first time, it is advisable to perform a series of emissions with short times (0.01-0.02
sec.) and then, progressively, some pictures with 0.1 sec. intervals to better stabilise the operation of the x-
ray tube before using it.
Before actually taking an exposure, make sure the exposure parameters for the examination in progress are correctly
set:
- Controlling the type of collimator selected.
Make sure the cone icon shown on the handheld matches the desired cone:
ICON ON: indicates that the x-ray unit is using the built in round cone (diameter 60mm, focal spot-skin distance
20cm).
ICON OFF: indicates that the external removable rectangular or round cone is attached to the x-ray device
(source-skin distance 30cm).
Refer to paragraph 5.3 to set the type of movable collimator used.
NOTE: After the rectangular collimator (12”) has been attached or detached, the icon and set exposure times
will automatically be modified within a few seconds.
NOTE: After the change has been made, the preset exposure times will automatically be modified.
EN OPERATOR'S MANUAL 21
- Checking the selected type of intraoral exam.
Upper incisors
Lower molars
or front ”bitewing”
22 OPERATOR'S MANUAL EN
5. USE OF THE X-RAY DEVICE
A positioner or alignment device specific for the selected image receiver should always be used to assure the x-rays
are correctly aligned regardless of the position the patient’s head is in.
Position the x-ray head so that the cone is aligned with the image receiver.
In the versions equipped with ball joint, the x-ray head can freely rotate on both
its horizontal and vertical axis.
An electromechanical brake initially locks the x-ray head. To release the head
and let it rotate on the positioning ball, work on the touch sensitive unlocking
areas located on it (see points A in the figure on the side).
Touching the unlocking areas allows positioning the x-ray head at the desired
angle to perform the exposure. To lock it again, simply release the unlocking
areas.
NOTE: Firmly hold the x-ray head with both hands when putting it in place.
It is possible to set a safety unlocking mode that allows the head to be turned only by pressing both unlocking buttons.
This prevents the head from unlocking unexpectedly after one of the two unlocking buttons has been accidentally
pressed. To activate this mode, refer to section 5 “Advanced options”.
ARCH TECHNOLOGY
In the versions equipped with arch technology, the x-ray head can freely rotate
on both its horizontal and vertical axis.
Simply moving the x-ray head allows positioning it at the desired angle to perform
the exposure.
EN OPERATOR'S MANUAL 23
5.3. POSITION OF THE X-RAY PLATE OR SENSOR
The parallel technique, where applicable, provides more accurate images in terms of size compared to the bisecting
technique. A rectangular collimator, with 30 cm (12"), focus-skin distance, is always preferable to obtain better quality
pictures. To avoid exposing the image receiver only partly (whether it is a sensor or photostimulable phosphorus plate
system) an alignment device that gives rectangular collimators guidelines should be used. These lines are usually
given on the alignment ring.
- Parallel technique.
1 Incisors
2 Canines
3 Premolars
4 Molars
5 Upper arch
6 Lower arch
24 OPERATOR'S MANUAL EN
- The x ray emission axis is perpendicular to the image receiver (for example a sensor or photostimulated phosphor
plate) which in turn is parallel with the tooth’s long axis.
- As a result, the picture of the tooth will only be deformed by the divergence of the x rays in relation to the focus
spot.
- Radiographic enlargement may reach up to 15%.
- For some “special” pictures, for example occluded ones, it may be necessary to remove the rectangular collimator
and use the round one if a positioner is not present.
The exposure parameters are set by following the directions given below:
1) Select the tooth to be examined
2) Select the patient size
NOTE: Each tooth and patient size selected is displayed for approximately 1 second according to the
operating mode used (En60, En63 o En65).
The suggested exposure time can be changed with keys and . Exposure times ranging from 0.02s to 1.00s
belonging to the R'20 scale can be set. Random exposure times different from the ones provided in the R’20 scale
cannot be set.
When the exposure time displayed differs from the default setting, icon comes on.
To save the new setting, make sure that icon is on and then press and hold down key for approximately 2
seconds. The handheld will beep shortly to confirm that the setting has been saved. At this point, make sure that icon
is off.
NOTE: If the exposure time is not saved, the change made will be lost after a new entry or as soon as the
handheld changes over to stand-by.
EN OPERATOR'S MANUAL 25
WARNING:
After customised settings have been made, the “Original exposure values charts” are no longer valid.
If icon is displayed while the exposure time is changed, it means that the set time cannot be saved for the selected
tooth-patient size combination. In any case, the x-rays can be taken with the set time.
WARNING:
When the suggested exposure time is changed, the sensitivity factor is also modified (by default set to F=19).
Once this change has been saved, it is applied to all the teeth and both patient sizes.
The exposure time can also be modified by changing the sensitivity factor. To do so, press keys and at the
Use keys and to change the value from 3 to 25. If the displayed value differs from the previous setting, icon
comes on. To quit this mode, press key or to confirm. The change made to the sensitivity factor is
26 OPERATOR'S MANUAL EN
5.5. PROCEDURE TO BE FOLLOWED WHEN TAKING THE X-RAY
- Pick up the handheld and move to a safe distance (at least 2 meters) maintaining visual contact with the patient
and x-ray unit during the exposure. Make sure “ready” is indicated.
NOTE: If the “EMIT X-RAY” key is released at any time, exposure will be interrupted and error code E01 will
appear on the display.
- Once exposure has been completed, it is possible to proceed with the next exposure unless the x-ray unit has
reached the maximum allowable temperature. The percentage the cone exceeds the maximum allowable
temperature is always shown on the screen (see icon below).
- Once the temperature has been reached, wait the pause time for cooling signaled by symbol .
- At this point the exposure function will be disabled until the screen shows “ready” again
- As soon as “READY” appears on the handheld, another exposure can be taken.
EN OPERATOR'S MANUAL 27
6. ADVANCED OPTIONS
The handheld allows the user to view, edit and set some operating parameters by simply combining the keys provided.
Follow the steps outlined below to access:
KEY COMBINATION DESCRIPTION OF COMMAND
Press these two keys to adjust the sensitivity levels (determined based on the table
given below and type of sensor/receiver used), modifying the current value from
the minimum to the maximum allowable one (on a scale from 3 to 25), with keys
“+” and “-”. Press key “size” to confirm the desired level and go back to the main
+ screen.
Hold down these two keys to go to the set-up menu (from P 01 to P 07).
Press key “size” to make the selection. Once within the individual configurations,
they can be scrolled with keys “+” and “-” and selected by pressing key “size” again.
Key "tooth" quits the configuration without saving the setting.
The configurations are outlined in detail below:
- P 01: Sets the stand-by time (from a minimum of 5 to a maximum of 30
minutes).
- P 02: Assigns an identification tag to the x-ray unit base (from 1 to 5 or none).
- P 03: Shows the list of software versions.
+ - P 04: Handheld unique code display.
- P 05: Activates/deactivates the safety unlock mode (only with ball joint).
- P 06: Operating mode.
- P 07: Sets the type of removable collimator used.
- P 08: Set current value (only for 70kV x-ray heads).
X-ray unit has a safety unlocking mode for the ball joint.
The default setting allows the ball joint to be disengaged by simply touching one of the buttons present on the front of
the head. To prevent accidental contact with the buttons unexpectedly disengaging the ball joint (and therefore causing
undesired movement of the head), the safety unlocking mode can be activated. In this mode, the ball joint is disengaged
only if both buttons are activated at the same time.
To set the safety unlocking mode, press keys and to go to the set-up menu.
Scroll the parameters up to parameter P05 and press key . Scroll the options to select “ON” and then press key
28 OPERATOR'S MANUAL EN
6.1. SETTING THE OPERATING MODE
For both operating modes, the exposure time is in the range 0.02s - 1s. The permitted anode voltage and current
combinations are shown in the following table:
NOTE: This section applies only to models RX DC REF: MRPU*6***, MRXU*6***
(Character * can be any alphanumeric value)
Table 1
60-7 60 kV 7 mA
63-6 63 kV 6 mA
65-6 65 kV 6 mA
Table 2
60-8 60 kV 8 mA
65-8 65 kV 8 mA
70-8 70 kV 8 mA
60-4 60 kV 4 mA
65-4 65 kV 4 mA
70-4 70 kV 4 mA
To set the operative mode, press keys and to go to the setup menu.
Scroll the parameters up to parameter P06 and press key . Scroll the options to find the desired operating mode
EN OPERATOR'S MANUAL 29
6.2. SETTING TYPE OF MOVABLE COLLIMATOR
NOTE: For an ideal use of the x-ray unit, set the collimator depending on the type used.
To set the type of collimator, press keys and to go to the set-up menu.
Scroll the parameters up to parameter P07 and press key . Scroll the options to find the type of collimator used
To restore the factory settings (see paragraph 4.4) press keys and to go to the set-up menu.
Press keys and simultaneously. “rESS” will briefly appear and the handheld will be rebooted.
30 OPERATOR'S MANUAL EN
7. ERROR MESSAGES
ERROR
ERROR TYPE DESCRIPTION SOLUTION
CODE
The warning signals that the x-ray unit does not contain
GENERAL Configuration data reading configuration data. Perform the handheld pairing
W0005
WARNING error procedure and if the warning persists after restarting
the machine contact the service department.
X-RAY UNIT Mains voltage lower than Check the power supplied to the machine. The
E104
ERROR 260 V. shooting cycle cannot start until it is restored.
E105 X-RAY UNIT Try to repeat exposure. If the problem persists replace
Software error
E107 ERROR the x-ray unit control board.
EN OPERATOR'S MANUAL 31
ERROR
ERROR TYPE DESCRIPTION SOLUTION
CODE
EXTERNAL X-
E206 External x-ray key released Hold down the key until the image has been captured.
RAY KEY
E211 in advance If the issue persists, replace the external shooting key.
ERROR
32 OPERATOR'S MANUAL EN
ERROR
ERROR TYPE DESCRIPTION SOLUTION
CODE
NOTE: As regards the other error codes, CONTACT the technical service department.
EN OPERATOR'S MANUAL 33
8. PERIODIC MAINTENANCE
WARNING:
Any technical maintenance work required must be carried out by qualified personnel or by a specialised
technician authorised by the Manufacturer. It is the user’s responsibility to check that an authorised technician
carries out routine maintenance at least every year. The maintenance methods are specified in the Technical
Service Manual possessed by the Authorised Technicians
For safety reasons and for the health of the patient, operator and third-parts inspections and maintenance must be
carried out at regular intervals.
Quality control by means of a dental phantom for image acquiring systems, according to IEC 61223-3-4:2000:
The quality control consists in performing a radiological investigation by means of the tested device and a suitable
acquisition system. The sensitivity of the x-ray unit must be set according to what is stated in paragraph 14.
Alternatively, you can verify that the measured load factors (kV, mA, ms) fall within the accuracy limits stated in Section
11.
Periodic monitoring ensures the proper functioning of the device and the conformity of the results obtained.
34 OPERATOR'S MANUAL EN
9. CLEANING AND DISINFECTION
- Cleaning is the first necessary step for any disinfection process. Rubbing with detergents and surfactants
and rinsing with water removes allows removing a considerable number of microorganisms. If a surface
is not cleaned first, the disinfection process cannot be successful.
- The recommended products have been tested: they are technically compatible with the device materials.
- Damages to surfaces and materials due to the use of different products cannot be excluded even if they
are not included in the exceptions mentioned above.
EN OPERATOR'S MANUAL 35
10. DISPOSING THE EQUIPMENT WHEN NO LONGER USED
WARNING!
Never remove the device covers.
The device does not contain parts that can be repaired directly by the user. In the event of malfunctioning, do not
attempt to carry out any type of maintenance operation. If you find or suspect any kind of system malfunctioning, do
not attempt to carry out any type of maintenance operation and do not use the system on a patient, but directly contact
your local distributor.
The user may not carry out maintenance on any mechanical or electronic part of the x-ray system.
Opening the cases to access the internal circuits may cause device breakage and failure of the electrical safety devices
and will lead to forfeiture of the warranty.
Any maintenance, repairs and modifications of the device must be carried out only by personnel directly authorised by
the Manufacturer or by third parties expressly authorised by the Manufacturer and must be carried out according to the
laws in force and the generally accepted technical standards.
All the system components must be checked and replaced, if necessary, by qualified personnel.
For any maintenance operation, please contact the Manufacturer via the website indicated on the cover of this manual
by filling in the Information Request form.
Should you for any reason need to return the device or its parts to the Manufacturer or a Technical Service centre,
disinfect all the external parts of the device using a specific product (see the paragraph “Cleaning and disinfection”)
and preferably return it in its original packaging.
At the end of its lifetime, dispose of the device in accordance with the regulations in force. It is also advisable to disinfect
all the external parts of the device before disposal and to separate the materials for differentiated waste collection.
In compliance with Directives 2011/65/EU and 2012/19/EU regarding restriction of the use of certain hazardous
substances in electrical and electronic equipment along with waste electrical and electronic equipment it is forbidden
to dispose of this equipment in the municipal waste stream as unsorted municipal waste.
When purchasing a new device of an equivalent type, one for one, the device that has come to the end of its lifetime
should be returned to the dealer for disposal.
As regards reuse, recycling and other forms of recovery of the above mentioned waste, the Manufacturer carries out
the functions defined in the individual national legislations.
Appropriate differentiated waste collection for subsequent recycling treatment and environmentally friendly disposal
contributes to preventing possible negative effects on the environment and health and encourages recycling of the
materials of which the device is made up.
The symbol indicating separate collection for electrical and electronic equipment consists of the crossed out bin marked
on the equipment.
WARNING!
Under local legislation, fines can be imposed if the equipment is disposed in an illegal manner.
36 OPERATOR'S MANUAL EN
11. TECHNICAL DATA
WEIGHTS
- Weight of the unit with packaging: 38Kg (84lb) max.
- Weight of the x-ray unit: 25Kg (55 lb).
- Weight of the handheld: 0.3Kg (0.7 lb).
- Weight of the x-ray mobile stand: 20Kg (44 lb).
EN OPERATOR'S MANUAL 37
NOTE: This section applies only to models RX DC REF: MRPU*6***, MRXU*6***
(Character * can be any alphanumeric value)
WEIGHTS
- Weight of the unit with packaging: 38Kg (84lb) max.
- Weight of the x-ray unit: 25Kg (55 lb).
- Weight of the handheld: 0.3Kg (0.7 lb).
- Weight of the x-ray mobile stand: 20Kg (44 lb).
38 OPERATOR'S MANUAL EN
11.1. X-RAY TUBES
EN OPERATOR'S MANUAL 39
X-RAY TUBE CANON / TOSHIBA D-0711, Kailong KL21-
0.7.
- Anode inclination: 16.0°
- Anode thermal load: 7.0 KJ
- Maximum continuous heat dissipation: 170 W.
- Operating cycle: 1:60 (1 second exposure - 60 seconds
pause time).
40 OPERATOR'S MANUAL EN
12. DIMENSIONAL CHARACTERISTICS
EN OPERATOR'S MANUAL 41
42 OPERATOR'S MANUAL EN
WALL-MOUNTED VERSION WITH BALL JOINT
EN OPERATOR'S MANUAL 43
44 OPERATOR'S MANUAL EN
INTEGRATED VERSION
EN OPERATOR'S MANUAL 45
MOBILE STAND VERSION
46 OPERATOR'S MANUAL EN
EN OPERATOR'S MANUAL 47
13. IDENTIFICATION PLATES
WARNING:
Never remove the identification nameplates provided on the generator, control unit and collimator cone.
48 OPERATOR'S MANUAL EN
X-ray head.
The nameplate is located on the lower cover at the back of
the x-ray head.
Data given on plate:
- Name of the Manufacturer.
- Name of equipment.
- Technical specifications.
- Model and x-ray tube serial number.
- Equipment serial number.
- Month and year of manufacture.
- Symbol required by standards.
Collimator.
The nameplate for the rectangular collimator is found
outside it.
Data given on plates:
- Name and address of the Manufacturer.
- Equipment code.
- Technical data.
- DHHS compliance.
- Serial number.
- Month and year of manufacture.
Wireless handheld.
The nameplate for the handheld is found in the battery
compartment.
Data given on plate:
- Name of the Manufacturer.
- Name of equipment.
- Rated voltage.
- Number and type of batteries.
- Serial number.
Wired handheld.
The handheld control plate is located at the centre on the
back.
Plate images are purely illustrative; refer to the plate placed on the device.
EN OPERATOR'S MANUAL 49
14. TIMES/SENSITIVITY CHARTS
When using the x-ray unit, in conjunction with the intraoral sensor of the Manufacturer, it is recommended to use a
default value of sensitivity equal to F15.
If using phosphorous plates of the Manufacturer, it is recommended to set a sensitivity value equal to F16.
For the analogue films the optimum sensitivity is F19.
For sensors other than the previous ones, you need to find the appropriate exposure time in relation to the image that
you want to obtain. Once identified this time for a specific combination of tooth and patient size, you can set the
sensitivity parameter in accordance with the exposure time found (procedure described in paragraph 5.4).
Sensitivity 19*
Collimator (focus-skin distance) 20 cm (8") 30 cm (12")
Mode En60 En63 En65 En60 En63 En65
Patient size A B A B A B A B A B A B
Lower incisors 0.125 0.080 0.110 0.071 0.100 0.063 0.250 0.160 0.220 0.140 0.200 0.125
Lower canines/bicuspids 0.160 0.100 0.140 0.090 0.125 0.080 0.320 0.200 0.280 0.180 0.250 0.160
Upper incisors/front “bitewing” 0.160 0.100 0.140 0.090 0.125 0.080 0.320 0.200 0.280 0.180 0.250 0.160
Lower molars 0.200 0.125 0.180 0.110 0.160 0.100 0.400 0.250 0.360 0.220 0.320 0.200
Upper canines/bicuspids/rear 0.200 0.125 0.180 0.110 0.160 0.100 0.400 0.250 0.360 0.220 0.320 0.200
“bitewing”
Upper molars 0.250 0.160 0.220 0.140 0.200 0.125 0.500 0.320 0.450 0.280 0.400 0.250
Sensitivity 19*
Collimator (focus-skin distance) 20 cm (8") 30 cm (12")
Mode En60 En65 En70 En60 En65 En70
Patient size A B A B A B A B A B A B
Lower incisors 0.110 0.071 0.080 0.050 0.063 0.040 0.220 0.140 0.160 0.100 0.140 0.080
Lower canines/bicuspids 0.140 0.090 0.100 0.063 0.080 0.056 0.280 0.180 0.200 0.125 0.160 0.110
Upper incisors/front “bitewing” 0.140 0.090 0.100 0.063 0.080 0.056 0.280 0.180 0.200 0.125 0.160 0.110
Lower molars 0.180 0.110 0.125 0.080 0.110 0.063 0.360 0.220 0.250 0.160 0.220 0.140
Upper canines/bicuspids/rear 0.180 0.110 0.125 0.080 0.110 0.063 0.360 0.220 0.250 0.160 0.220 0.140
“bitewing”
Upper molars 0.220 0.140 0.160 0.100 0.140 0.080 0.450 0.280 0.320 0.200 0.280 0.160
* = Default settings
A = Adult
B= Child
50 OPERATOR'S MANUAL EN
15. NOMINAL DOSE EMISSION VALUES
EN OPERATOR'S MANUAL 51
NOTE: This section applies only to models RX DC REF: *****7***
(Character * can be any alphanumeric value)
52 OPERATOR'S MANUAL EN
NOTE: This section applies only to models RX DC REF: *****6***
(Character * can be any alphanumeric value)
EN OPERATOR'S MANUAL 53
NOTE: This section applies only to models RX DC REF: *****7***
(Character * can be any alphanumeric value)
54 OPERATOR'S MANUAL EN
NOTE: This section applies only to models RX DC REF: *****6***
(Character * can be any alphanumeric value)
Collimator Round ø 55 mm
SSD 30 cm
A (cm2) 23.75
KV - mA 60kV 7mA 63kV 6mA 65kV 6mA
t (s) DOSE DAP DOSE DAP DOSE DAP
mGy mGy · cm2 mGy mGy · cm2 mGy mGy · cm2
0.020 0.09 2.0 0.08 1.9 0.08 2.0
0.022 0.09 2.2 0.09 2.1 0.09 2.2
0.025 0.11 2.5 0.10 2.4 0.11 2.5
0.028 0.12 2.8 0.11 2.7 0.12 2.8
0.032 0.14 3.2 0.13 3.0 0.14 3.2
0.036 0.15 3.7 0.14 3.4 0.15 3.6
0.040 0.17 4.1 0.16 3.8 0.17 4.0
0.045 0.19 4.6 0.18 4.3 0.19 4.5
0.050 0.21 5.1 0.20 4.7 0.21 5.0
0.056 0.24 5.7 0.22 5.3 0.24 5.6
0.063 0.27 6.4 0.25 6.0 0.27 6.3
0.071 0.30 7.2 0.28 6.7 0.30 7.1
0.080 0.34 8.1 0.32 7.6 0.34 8.0
0.090 0.38 9.1 0.36 8.5 0.38 9.0
0.100 0.43 10.1 0.40 9.5 0.42 10.0
0.110 0.47 11.2 0.44 10.4 0.47 11.1
0.125 0.53 12.7 0.50 11.8 0.53 12.6
0.140 0.60 14.2 0.56 13.3 0.59 14.1
0.160 0.68 16.2 0.64 15.2 0.68 16.1
0.180 0.77 18.3 0.72 17.1 0.76 18.1
0.200 0.85 20.3 0.80 19.0 0.85 20.1
0.220 0.94 22.3 0.88 20.8 0.93 22.1
0.250 1.07 25.4 1.00 23.7 1.06 25.1
0.280 1.20 28.4 1.12 26.5 1.18 28.1
0.320 1.37 32.5 1.28 30.3 1.35 32.1
0.360 1.54 36.5 1.44 34.1 1.52 36.2
0.400 1.71 40.6 1.60 37.9 1.69 40.2
0.450 1.92 45.6 1.80 42.6 1.90 45.2
0.500 2.14 50.7 2.00 47.4 2.12 50.2
0.560 2.39 56.8 2.23 53.1 2.37 56.3
0.630 2.69 63.9 2.51 59.7 2.66 63.3
0.710 3.03 72.0 2.83 67.3 3.00 71.3
0.800 3.42 81.1 3.19 75.8 3.38 80.4
0.900 3.84 91.3 3.59 85.3 3.81 90.4
1.000 4.27 101.4 3.99 94.8 4.23 100.5
EN OPERATOR'S MANUAL 55
NOTE: This section applies only to models RX DC REF: *****7***
(Character * can be any alphanumeric value)
Collimator Round ø 55 mm
SSD 30 cm
A (cm2) 23.75
KV - mA 60kV 8mA 65kV 8mA 70kV 8mA
t (s) DOSE DAP DOSE DAP DOSE DAP
mGy mGy · cm2 mGy mGy · cm2 mGy mGy · cm2
0.020 0.10 2.3 0.11 2.7 0.13 3.0
0.022 0.11 2.5 0.12 2.9 0.14 3.3
0.025 0.12 2.9 0.14 3.3 0.16 3.8
0.028 0.14 3.2 0.16 3.8 0.18 4.2
0.032 0.16 3.7 0.18 4.3 0.20 4.8
0.036 0.18 4.2 0.20 4.8 0.23 5.4
0.040 0.20 4.6 0.23 5.4 0.25 6.0
0.045 0.22 5.2 0.25 6.0 0.29 6.8
0.050 0.24 5.8 0.28 6.7 0.32 7.6
0.056 0.27 6.5 0.32 7.5 0.36 8.5
0.063 0.31 7.3 0.36 8.4 0.40 9.5
0.071 0.35 8.2 0.40 9.5 0.45 10.7
0.080 0.39 9.3 0.45 10.7 0.51 12.1
0.090 0.44 10.4 0.51 12.1 0.57 13.6
0.100 0.49 11.6 0.56 13.4 0.64 15.1
0.110 0.54 12.7 0.62 14.7 0.70 16.6
0.125 0.61 14.5 0.71 16.7 0.80 18.9
0.140 0.68 16.2 0.79 18.8 0.89 21.1
0.160 0.78 18.5 0.90 21.4 1.02 24.2
0.180 0.88 20.9 1.02 24.1 1.14 27.2
0.200 0.98 23.2 1.13 26.8 1.27 30.2
0.220 1.07 25.5 1.24 29.5 1.40 33.2
0.250 1.22 29.0 1.41 33.5 1.59 37.8
0.280 1.37 32.5 1.58 37.5 1.78 42.3
0.320 1.56 37.1 1.80 42.9 2.04 48.3
0.360 1.76 41.7 2.03 48.2 2.29 54.4
0.400 1.95 46.4 2.26 53.6 2.54 60.4
0.450 2.20 52.2 2.54 60.3 2.86 68.0
0.500 2.44 58.0 2.82 67.0 3.18 75.5
0.560 2.73 64.9 3.16 75.0 3.56 84.6
0.630 3.07 73.0 3.55 84.4 4.01 95.2
0.710 3.46 82.3 4.00 95.1 4.52 107.2
0.800 3.90 92.7 4.51 107.2 5.09 120.8
0.900 4.39 104.3 5.08 120.6 5.72 135.9
1.000 4.88 115.9 5.64 134.0 6.36 151.1
56 OPERATOR'S MANUAL EN
NOTE: This section applies only to models RX DC REF: *****6***
(Character * can be any alphanumeric value)
EN OPERATOR'S MANUAL 57
NOTE: This section applies only to models RX DC REF: *****7***
(Character * can be any alphanumeric value)
58 OPERATOR'S MANUAL EN
NOTE: This section applies only to models RX DC REF: *****6***
(Character * can be any alphanumeric value)
EN OPERATOR'S MANUAL 59
NOTE: This section applies only to models RX DC REF: *****7***
(Character * can be any alphanumeric value)
An indirect measure of the DAP can be obtained by a measuring instrument of the absorbed dose.
In this case, the DAP is obtained by multiplying the dose measurement for the irradiated area.
The data obtained by this measurement has a tolerance of not less than 20%.
60 OPERATOR'S MANUAL EN
16. INSPECTION AND MAINTENANCE
16.1. USER INSPECTION
These instructions describe the maintenance procedures for the x-ray unit.
These instructions apply to all versions of said equipment, as well as all the accessories that may have been provided, therefore the description of some parts may not correspond to your equipment.
Inspection and preventive maintenance must be performed at scheduled intervals to protect the health and safety of patients, users and other persons in accordance with national regulations regarding the use and maintenance of dental x-ray
units that are in force in the country where the device is installed.
In order to ensure the operational safety and functional reliability of your product, the system owner should check the equipment at regular intervals (at least once a year) or commission an authorised technician to do so.
If one or more checks to be performed are not satisfactory, please contact your dealer for technical support.
Answer questions with yes (√) or not (−)
Inspection DATE
Step Description Reference in User Manual __/__/20__ __/__/20__ __/__/20__ __/__/20__ __/__/20__
Check that all labels located
- on the wall-mounted cover,
1 Section Identification nameplates
- on the x-ray tube
- inside the collimator/s, are intact, correctly applied and readable.
Check there are no external damages to the equipment, which may reduce
2 Section Description of the x-ray unit
protection against radiation.
3 Check the battery level of the wireless remote control. Section Operation of control handheld
4 Check the remote control functionality: buttons must respond to interaction Section Use of control handheld
Check the power switch verifying that the switch is working properly and the main
5 Section Turning the x-ray unit on and off
switch green light switches on when the switch is in the ON position.
Check the electromechanical brake that lock/unlock the movement of the
6 Section Ball joint technology
generator – ball joint model units only
Check proper functioning of x-ray generator indicator light – ball joint model units
7 Section Type of x-ray unit
only
8 Check the exposure buzzer during a trial x-rays emission Section Performing the exposure
9 Verify that exposure is immediately interrupted when x-ray button is released Section Performing the exposure
10 Check the scissors arm balance Section Description of the x-ray unit
11 Verify that exposure is immediately interrupted when x-ray button is released Section Performing the exposure
Check the x-ray generator functionality performing a complete trial exposure.
Select any exposure time and hold down the emission button throughout the
12 Section Using the x-ray unit
entire exam procedure. Absence of error messages assures proper generator
functionality.
Operator Name
Signature
The undersigned confirms that the equipment was checked for the above criteria and that, in case of any malfunction, an authorised technician of the local dealer was informed.
All inspection and maintenance work performed by the system owner and/or service engineer must be recorded in this document and kept near the unit!
EN OPERATOR'S MANUAL 61
16.2. TECHNICAL MAINTENANCE
These instructions describe the maintenance procedures for the x-ray unit. These instructions apply to all versions of said equipment.
In order to ensure the operational safety and functional reliability of the equipment installed, at least once a year an authorised service technician must perform a full inspection of the device.
When taking measurements that require a multimeter, always use a calibrated digital multimeter.
All the following tests will be carried out. Customer should be notified prior to replacing any parts.
Answer questions with yes (√) or not (−)
Inspection DATE
Step Description Reference in User Manual __/__/20__ __/__/20__ __/__/20__ __/__/20__ __/__/20__
Check that all labels located
- on the wall-mounted cover,
1 User Manual, Section Identification nameplates
- on the x-ray tube
- inside the collimator/s, are intact, correctly applied and readable.
Check there are no external damages to the equipment, which may reduce
2 User Manual, Section Description of the x-ray unit
protection against radiation.
Pull out the collimator and panel stop ring, take off the screw cover caps and
Technical manual, section Replacement of x-ray unit control base
3 loosen the screws that secure the lower cover. Check there is no oil leakage on
board
the tube-head.
Check the electromechanical brake that locks/unlocks the movement of the
4 Technical Manual, section Actuator unit adjustment
generator and adjust it if necessary – ball joint model units only
Power off the unit and remove the wall mounting cover. Disconnect the unit from
the main power supply and check the condition of the main power supply cable.
5 Technical Manual, section Wall-mounted plate wiring connection
Replace it in case of damage. Connect it back making sure the safety ground is
securely connected. Install the wall-mounted cover back again.
6 Check the battery level of the wireless remote control. User manual, section Operation of control handheld
7 Check the remote control functionality: buttons must respond to interaction User manual, section Use of control handheld
Check the power switch verifying that the switch is working properly and the main
8 User Manual, section Switching the x-ray system on and off
switch green light switches on when the switch is in the ON position.
Check proper functioning of x-ray generator indicator light – ball joint model units
9 User Manual, section Type of x-ray unit
only
10 Check the exposure buzzer during a trial x-rays emission User Manual, Section Performing the exposure
11 Verify that exposure is immediately interrupted when x-ray button is released User Manual, Section Performing the exposure
12 Check the scissors arm balance and adjust it if necessary Technical Manual, section Fitting the double pantograph arm
13 Perform a calibration of the x-ray tube housing Technical manual, section User and technician configuration menu
Operator Name
Signature
The undersigned confirms that the equipment was checked for the above criteria and that it was provided in optimal operating conditions.
All inspection and maintenance work performed by the system owner and/or service engineer must be recorded in this document and kept near the unit!
62 OPERATOR'S MANUAL EN