RTK-TDS-RE3624 Controller data sheet
RTK-TDS-RE3624 Controller data sheet
CONTROLLER RE3624
Transmitter/controller for conductivity, TDS,
resistance, standard signals and temperature
Brief description
The RE3624 controller is a compact, modular instrument. It is highly flexible (for example 3 slots Controller RE3624
for optional boards) and capable of performing a wide range of tasks. The main input of the type 202552/01... in panel case
RE3624 is used for sensors for measuring electrolytic conductivity, specific resistance, or the
TDS value. Both conductive two-electrode and four-electrode cells can be connected to the
instrument. The second analog input (compensation input) is designed for resistance
thermometers Pt100 and Pt1000, NTC/PTC or standard signals 0(4) to 20 mA or 0 to 10 V. The
two binary inputs can be used either as initiators for actions (e.g. HOLD, keyboard inhibit) or when
connecting pulse generators (for example impeller sensors) for flow-rate measurement. The
high-contrast graphic display allows for several options including display of input signal with
numbers or as bar graph. Parameters are displayed in plain text for easily comprehensible and
reliable operation.
The RE3624 can be used as a two-point or three-point controller, a three-point modulating
controller, or as a continuous controller. All controller outputs can be configured to P, PI, PD or
PID action. The software for the controllers includes parameter set selection, a math module and
more.
A setup program is available for convenient configuration via PC. The instrument can be
integrated into a data network by means of an RS422/485 or PROFIBUS-DP interface. Screw
terminals on the back are used for the electrical connection.
Some applications:
• Industrial and process water. Controller RE3624
• Drinking and well water. type 202552/05... in surface-mounted case
• Pure, ultra-pure and pharmaceutical water (e.g. as per USP, Ph. Eur., WFI).
• Cleaning processes in pharmaceutical applications (four-electrode cells in conjunction with
measuring range selection).
CONTROL
SYSTEMS
3624-7011
board Sensors that return one of the output signals H2SO4 0 to 28 % by wt. 0 to 100 ° C
listed above can be connected to the H2SO4 36 to 85 % by wt. 0 to 115 ° C
Conductivity measurement instrument for the following measurement
H2SO4 92 to 99 % by wt. 0 to 115 ° C
variables, for example:
The measurement can be conducted either • free chlorine, chlorine dioxide, ozone, Hydrochloric acid
with standard two-electrode or with four- hydrogen peroxide and peracetic acid as HCl 0 to 18 % by wt. 0 to 65 ° C
electrodes cells. per data sheet 202630. Hal 22 to 44 % by wt. -20 to 65 ° C
Two-electrode cells can be connected in the • pH value or redox potential as per data
usual grid of cell constants (K = 0.01; 0.1; 1.0; sheet 202701.
• Liquid level measurements. Cust. specs. with table
3.0 and 10.0). The "relative cell constant“ can
be adjusted over wide ranges, which makes it • Flow rate measurements etc. Non-linear correlations between the input and
possible to connect sensors with different cell output variable can be processed in this mode.
The instrument provides the following
constants as well (for example K = 0.2). Typical applications include measuring the
calibration options in this function mode:
level of liquid in horizontal, cylindrical
• Zero point
Values K = 0.5 and 1.0 are predefined for four- containers or simply measuring the
electrode cells. In this case as well, the device • Limit value concentration.
can be adjusted to sensors with different cell • Zero point and limit value The input values are processed in a table
constants (for example K = 0.4). • Cell constant (max. 20 value pairs). Values can only be
• Temperature coefficient entered in the table using the optional setup
The instrument is able to perform an automatic
This allows optimum adaptation of the program.
temperature compensation.
instrument to the sensor. The units used for display and control are:
• µ S/cm
Resistance Linear scaling
Select this mode when the input signal will be • mS/cm
The instrument can be switched to resistance • Cust. specs. (5 characters)
displayed linearly.
measurement for applications in which display • Use the offset parameter to adjust the
One of the following units is used for display or
of the resistance value is preferred over the display.
control:
conductivity value.
• µ S/cm
Calibration
TDS • mS/cm
• % Calibration logbook
Display/control with the unit ppm. The last five successful calibrations can be
• mV
accessed from the calibration logbook. This
The specific TDS factor can also be entered in • pH makes it possible to evaluate the aging of the
this mode. • ppm connected sensor.
• Cust. specs. (5 characters) The logbook can be deleted if necessary
Temperature compensation
Electrolytic conductivity (useful when changing the sensor).
The conductivity or resistance of aqueous µ S/cm or mS/cm are the units used for display If a datalogger has been fitted (optional
solutions often depends greatly on the and control. board), additional information such as the
temperature. The instrument provides the date and time are documented.
Specific resistance (ultra-pure water)
following procedures for temperature Calibration timer
Display/control with the unit k × cm or M
compensation, depending on the display size: The calibration timer indicates (on request) a
× cm.
• Off (e.g. USP) required routine calibration. The calibration
TDS timer is activated by entering the number of
• Linear Display/control with the unit ppm. days that must expire before there is a
The specific TDS factor can also be entered in scheduled re-calibration (specified by the
• ASTM this mode. system or the operator).
• Natural waters (EN 27888/ISO 7888) Concentration
In this mode, the concentration of a liquid can
Analog input for main board be determined from its uncompensated
conductivity.
0(4) to 20 mA; 0 to 10 V and Pt100/Pt1000/ % or "Cust. specs." are the units used for
NTC/PTC (max. 30 k )/cust. specs. display and control.
Typical application: Compensation input for
temperature compensation of the main
measurement variable.
3624-7013
Switching period
50% 50%
50%
tOn tOff
10% Setup interface
90% The setup interface is integrated into the
10% RE3624 by default. You can use it, together
0%
with the setup program (accessory) and a
Proportional band XP Process value X
setup interface (accessory), to configure the
0 X-W 1 instrument.
XP
RS232/RS485 interface
Setpoint W The serial interface is used for communication
with higher-level systems when the Modbus/
If actual value x exceeds setpoint w, the P controller will control in proportion to the control Jbus protocol is used.
deviation. When the proportional range is exceeded, the controller operates with an output level
of 100 % (100 % clock ratio). PROFIBUS-DP
The RE3624 can be integrated into a fieldbus
Pulse frequency controller system according to the PROFIBUS-DP
(output active with x > w and P control structure) standard via the PROFIBUS- DP interface.
This PROFIBUS-DP version is especially
100% designed for communication between
Output level y
Setpoint W
If actual value x exceeds setpoint w, the P controller will control in proportion to the control
deviation. When the proportional range is exceeded, the controller operates with an output level
of 100 % (maximum switching frequency).
3624-7016
Example
To conduct a measurement in the range from 10 µ S/cm to 500 µ S/cm,
select a conductivity cell with a cell constant K = 0.1 1/cm.
Configure the unit µ S/cm on the instrument without places after the decimal.
It may be assumed that the measurement scope of the instrument is always greater than the
recommended or practically usable range of the conductivity cell that is used.
The smaller range (instrument or conductivity cell) determines the maximum usable range.
Example
What measurement scope can the instrument cover with a specified cell constant?
The specified cell constant is K = 0.4
The measurement scope of the instrument = 0.1 µ S/cm × 0.4 1/cm to 1250 mS/cm × 0.4 1/cm
➝ 0.04 µ S/cm to 500 mS/cm
3624-7017
Technical data
Inputs (main board)
Main input Measuring range/control range Accuracy Effect of temperature
µ S/cm 0.000 to 9.999 0.6 % of range + 0.3 µ S × 0.2 %/10 K
00.00 to 99.99 cell constant (K)
000.0 to 999.9
0000 to 9999
mS/cm 0.000 to 9.999 0.6 % of range + 0.3 µ S × 0.2 %/10 K
00.00 to 99.99 cell constant (K)
000.0 to 999.9
0000 to 9999a
k × cm 0.000 to 9.999 0.6 % of range + 0.3 µ S × 0.2 %/10 K
00.00 to 99.99 cell constant (K)
000.0 to 999.9
0000 to 9999
M × cm 0.000 to 9.999 0.6 % of range + 0.3 µ S × 0.2 %/10 K
00.00 to 99.99 cell constant (K)
000.0 to 999.9
0000 to 9999
Secondary input
Temperature Pt100/1000 -50 to +250 ° Cb 0.25 % of range 0.2 %/10 K
Temperature NTC/PTC 0.1 to 30 k 1.5 % of range 0.2 %/10 K
Entry via table with 20 value pairs
Standard signal 0(4) to 20 mA or 0 to 10 V 0.25% of range 0.2 %/10 K
Resistance transmitter Minimum: 100 ±5 0.1 %/10 K
Maximum: 3 k
a
In the range between 1 to 10 S the accuracy is 1 % of the measuring range.
b Selectable in ° F
Temperature compensation
Type of compensation Rangea
Linear 0 to 8 %/K -10 to +160 ° C
ASTM D1125 - 95 (ultra-pure water) 0 to 100 ° C
Natural waters (ISO 7888) 0 to 36 ° C
Reference temperature
Adjustable from 15 to 30 ° C;
preset to 25 ° C (default)
a
Note the sensor operating temperature range!
Two-electrode systems
Cell constant Setting range Resulting usable range
[1/cm] of the relative cell constant [1/cm]
0.01 0.002 to 0.05
0.1 0.02 to 0.5 1.0
20 to 500 % 0.2 to 5
3.0 0.6 to 15 10.0
2.0 to 50
Four-electrode systems
Cell constant Setting range Resulting usable range
[1/cm] of the relative cell constant [1/cm]
0.5 0.1 to 0.75
20 to 150 %
1.0 0.2 to 1.5
Binary input
Activation Floating contact is open: function is not active
Floating contact is closed: function is active
Function Key lock, manual mode, HOLD, HOLD inverse, alarm suppression, freeze measured value,
level lock, reset day counter, reset total counter, parameter set changeover, flow-rate
measurement
Pulse input for flow measurement Binary input 1: approx. 3 to 2000 Hz, resolution 2 Hz
Binary input 2: approx. 4 to 300 Hz, resolution 0,5 Hz
At the device only one binary input for flow measurement can be used.
Controller
Controller type Limit comparators, limit controllers, pulse length controllers, pulse frequency
controllers, modulating controllers, continuous controllers
Controller structure P/PI/PD/PID
3624-7019
Outputs
Relay (changeover) PSU board
Contact rating 5 A at AC 240 V resistive load
Contact service life 350,000 operations at nominal load/750,000 operations at 1 A
Voltage supply for PSU board Electrically isolated, non-controlled
2-wire transmitter DC 17 V at 20 mA, open-circuit voltage approx. DC 25 V
Voltage supply for Optional board DC 12 V; 10 mA
inductive proximity switch
Relay (changeover) Optional board
Contact rating 8 A at AC 240 V resistive load
Contact service life 100,000 operations at nominal load/350,000 operations at 3 A
Relay SPST (normally open) Optional board
Contact rating 3 A at AC 240 V resistive load
Contact service life 350,000 operations at nominal load/900,000 operations at 1 A
Semiconductor relay Optional board
Contact rating 1 A at 240 V
Protective circuit Varistor
PhotoMOS® relay Optional board U AC/DC 50 V
I 200 mA
Voltage Optional board
Output signals 0 to 10 V or 2 to 10 V
Load resistance Rload 500
Accuracy 0.5 %
Electrical current Optional board
Output signals 0 to 20 mA or 4 to 20 mA
Load resistance Rload 500
Accuracy 0.5 %
Display
Type LC graphic display, blue with background lighting, 122 × 32 pixels
Electrical data
Voltage supply (switch-mode PSU) AC 110 to 240 V +10/-15 %; 48 to 63 Hz or
AC/DC 20to 30 V; 48 to 63 Hz
Electrical safety To DIN EN 61010, Part 1
overvoltage category II, pollution degree 2
Power consumption Max. 14 VA (20 A fuse max.)
Data backup EEPROM
Electrical connection On the back via screw terminals,
conductor cross-section up to max. 2.5 mm2
Electromagnetic Compatibility (EMC) DIN EN 61326-1
Interference emission Class A
Interference immunity To industrial requirements
3624-7020
Case
Enclosure type Plastic case for panel mounting to DIN IEC 61554 (indoor use)
Depth behind panel 90 mm
Ambient temperature -5 to +55 ° C
Storage temperature -30 to +70 ° C
Climatic rating Rel. humidity 90 % annual mean, no condensation
Site altitude Up to 2000 m above sea level
Operating position Horizontal
Enclosure protection To DIN EN 60529
In panel case Front IP65, rear IP20
In surface-mounted case IP65
Weight (fully fitted) Approx. 380 g
Interface
Modbus
Interface type RS422/RS485
Protocol Modbus, Modbus Integer
Baud rate 9600, 19200, 38400 Device
address 0 to 255
Max. number of nodes 32
PROFIBUS-DP
Device address 0 to 255
Approvals/approval marks
Mark of Testing laboratory Certificates/certification Test basis valid for
conformity numbers
c UL us Underwriters Laboratories E 201387 UL 61010-1 Type 202552/01...
CAN/CSA-C22.2 No. 61010-1
3624-7021
Electrical connection
(c) (b) (a) Mounting information for conductor cross-sections and ferrules
Ferrule Conductor cross- Minimum
section ferrule length
12
11
10
(1)
9
8
7
6
5
4
3
2
1
Minimum Maximum or stripping
Without ferrule 0.34 mm2 2.5 mm2 10 mm (stripping)
Without collar 0.25 mm2 2.5 mm2 10 mm
(2)
15
14
13
12
11
9
8
7
6
4
3
2
1
2 2
With collar up to 0.25 mm 1.5 mm 10 mm
1.5 mm2
Twin, with collar 0.25 mm2 1.5 mm2 12 mm
N(L-)
L1(L+)
(3)
17
16
15
13
12
11
9
8
6
5
4
Resistance transmitter 2 6 10
E 3 7 11
4 8 12
S
A
Electrical current + 3 7 11
4 8 12
-
Voltage + 1 5 9
0(2) to 10 V 2 6 10
-
Voltage + 2 6 10
0 to 1 V 3 7 11
-
Continuous output
Current or voltage + 2 6 10
3 7 11
-
Modbus interface
RS422 RxD+ - - 9
10
RxD- 11
12
TxD+
TxD-
RS485 RxD/TxD+ - - 11
12
RxD/TxD-
3624-7022
DGND
Datalogger interface
RS485 RxD/TxD+ - - 10
11
RxD/TxD-
S
Relay (2× NO, common
K3 1 - K5 9
2 10
K6 3 K8 11
Triac (1 A)
K3 2 K4 6 K5 10
3 7 11
K6 3 K7 7 K8 11
4 8 12
Resistance transmitter 4
E
3
S 2
A
3624-7023
Shield connection
Conductivity cell 10 GND
Binary inputs
Binary input 1 12+
14
n.c. 4
5
6
Relay 1
Switching output K1 O 11
(floating) 12
P 13
S
Relay 2
Switching output K2 O 15
(floating) 16
P 17
S
3624-7024
Dimensions
Panel case
Close mounting
Minimum spacing of Horizontal Vertical
panel cutouts
(1 Without setup 30 mm 11 mm
connector
With setup connector 65 mm 11 mm
(see arrow)
(1
(2
Surface-mounted case
(2)
(1)