USBe
USBe
consisting of
Control Unit (a) 732 689USB 01
Drive and Brake Unit (b) 732 689USB 02
[D&B Unit]
732 689USB 01
The control unit 732 689USB-01 is used in conjunction with the • Power supply 230 V, 47…63 Hz, 2,5 kW, supplied via
D & B Unit 732 689USB -02 as a testing system in the electrical connection cable with earthing-pin connector
machine laboratory.
• Automatic digital speed control:
The D & B Unit is a cradle-type three-phase asynchronous
machine with integrated torque pick-up for connection to the (-5000 rpm ...0...+5000 rpm)
control unit. This special machine is equipped with sufficient
power and torque reserves to brake to standstill or drive any of • Automatic digital torque control (±30.0 Nm)
the machines contained in the electrical machine system
1.0 kW. Torque pick-up is performed with strain-gauges • Reference variable and limiting value can be set via
mounted to a durable, highly elastic stainless steel bending bar incremental control knob
and includes integrated electronics. • Automatic run-up and load characteristics
Overload protection is guaranteed using thermal circuit-breaker 2
protection. Connection is carried out using a round 7-pin • Load simulations: Load start-up (M ~ n ), Heavy load
connector plug and a 25-pin cable contained within the scope start-up (M ~ 1/n), arbitrary load characteristic Mi ~ ni,
of supply. Flywheel
The control unit is a microcontroller-controlled device with inte-
grated frequency converter for the power supply and control of • External control: -10 V...+10 V
the three-phase pendulum machine and display of the speed • Switchover between two / four quadrant operation
and torque measured values. Different operating modes permit
manual and automatic recording of characteristics. Further- • 26 mm seven-segment display; speed 4 digit, torque
more, the machines can approach unstable operating points, 3 digit
like when the squirrel-cage machines are set to lower speeds
1
than the pull-out speeds, which are then maintained exactly by • Temperature monitoring: device under test (DUT ),
digital control loops. The recording of the characteristics, even D & B Unit, control unit
over several quadrants, is performed point-by-point, via the XY
recorder or the PC. • Four-quadrant display for load type of DUT
• Adjustable torque limiting (overload protection) and
stop speed (for automatic recording of characteristics)
• Direct measurement of DUT’s voltage and current
without the need of isolation amplifier
• USB interface for measured value transmission and
remote control
1
DUT = Device under test – machine to be tested
Seite 2/7
• Restart inhibit can be reset after fault induced switch- 2 Safety Instructions
off
• Analog outputs for connection of an XY recorder or
display meters
The device is designed according to protection
• Highest safety standard: leakage current < 5 mA class 1 and corresponds to the safety stipulations
laid out in VDE 0411. When the device is used
according to the stipulations the safety of both the
Furthermore the device is equipped with a error recognition operator as well as the device is guaranteed.
using six codes as well as protective functions. The following However, safety cannot be guaranteed if the device
parameters are continuously monitored: is used improperly or handled carelessly.
For that reason it is imperative that these
• Generator operation (ERR 1) instructions be read thoroughly and carefully before
• Temperature of the DUT (ERR 2) putting the control unit for the D & B Unit into
• Overvoltage in the link operation and that all of the points listed below be
circuit of the control unit (ERR 3) adhered to.
This device may not be operated by persons who
• Temperature of the D & B Unit (ERR 4) are unable to recognize contact hazards and
• Temperature of the control unit (ERR 5) incapable of meeting safety measures.
• DUT is switched off via the In general only work with safety connecting leads so
output DUT ENABLE by removing that contact hazards with dangerous voltages can be
the coupling guard or the excluded.
shaft end guard (ERR 6) Modifications to the experiment set up may only be
performed when the system is switched off (zero
current / zero voltage). For this turn off the three-
On switching on the control unit using the main switch (a1) and phase mains and the control unit for the D & B Unit
pressing the nSTOP (a28) button, the Software version is shown with switch a1.
on the displays (a29) and (a30). Before switch off operate the D & B Unit at no load
by actuating pushbutton M = 0.
Always use coupling and shaftend guards to protect
against contact to rotating parts.
Normally it is not necessary to open the device's
housing. However, should this prove necessary, this
may only be performed by a technician and only
under the condition that the mains plug and all
connecting leads have been disconnected.
Only apply safety extra-low voltages to the
CONTROL IN sockets.
Never connect external voltage sources to the
outputs.
Warning: This operating equipment defined class A,
group 1 as set forth in the EN55011 standard and
may cause radio interference in residential areas. In
this case it may be required of the person operating
the equipment to take necessary preventative
measures and to bear the expenses. However, when
the equipment is operated in the appropriate
technical setting or lab of a general education or
vocational institution or other training center it is
assumed that no radio interference will arise in
residential areas as long as a safety zone of at least
30 m exists and the operating equipment is operated
only intermittently.
LD DIDACTIC GMBH . Leyboldstrasse 1 . D-50354 Huerth . Phone +49 2233 604-0 . Fax +49 2233 604-222 . e-mail: [email protected]
3 Equipment Description machine-generator set sheds its load and the stop speed
flashes for approx. 5 seconds.
3(a) Control unit a4 The pushbutton M = 0 serves to interrupt any given
operating state. At the same time the torque display shows
OFF. The D & B Unit no longer develops any torque, comes to
a standstill or is set to its no-load speed by the machine under
test. Consequently, the activation of the pushbutton M = 0
results at all times to the load shedding of the machine-
generator set. Reactivation of the pushbutton M = 0 results in
the reactivation of the control unit and there with the frequency
converter.
a5 Operating mode switch (MODE)
Switchover to the next operating mode occurs each time this
pushbutton is pressed. The current operating mode is
displayed by an LED (a6).
a6 Modes of operation:
a6.1 Load characteristic
The rotation speed, specified by the incremental control knob
or externally, is kept constant. The load of the DUT is the
product of its speed / torque characteristic. The recording of the
characteristic is started by pressing the START button.
a6.2 Run-up characteristic
The speed, which is preset using the incremental control knob
or externally, is kept constant. The load of the DUT is the prod-
uct of its speed / torque characteristic. The recording of the
/
+ characteristic is started by pressing the START button.
a6.3 Automatic torque control (TORQUE CONTROL)
The torque, which is set by means of the incremental control
knob or externally, is kept constant. The DUT operates under
constant load.
2
a6.4 Load simulation M ~ k n
732 689USB 01 2
The D & B Unit simulates the load response of a fan (M ~ k * n
with k = 1).
a6.5 Load simulation M ~ k / n
The D & B Unit simulates the load response of a winding drive
(M ~ k / n with k = 1).
a1 Main switch a6.6 Load simulation Mi ~ ni
The D & B Unit simulates the load response with arbitrary
ON / OFF switch for the control electronics and supply voltages
characteristic edited by the user.
for the pendulum machine.
a6.7 Load simulation flywheel
a2 2- or 4-quadrant operation (ONLY MOTOR OPERATION)
The D & B Unit simulates the load response of a flywheel.
If this LED lights up the control unit automatically switches off
when the DUT is operating in the 2nd or 4th quadrants a7 Error message LED (ERROR)
(generator operation). This is a function used to protect power A multicolored LED is situated with the following significance:
sources which are not energy recovery-proof, if the DUT is off: normal operation
being supplied by such a source. green: An error has been eliminated and has to be
acknowledged
After the device is switched on at switch S ,the LED lights up
(only 2-quadrant operation possible). If the device is switched red: An error appears and is displayed as ERR n
on by pressing the START button (directly below the LED), the (n stands for 1 to 6, see table below
display (a29))
LED does not light up (4-quadrant operation possible).
Acknowledge by pressing RESET button (a8).
a3 Pushbutton to start automatic recording of the
characteristics. a8 RESET
If the operating mode load characteristic or run-up If an error message is displayed or switch off has been carried
characteristic is actice, automatic recording of characteristics out because the torque limit has been exceeded, the
can be started by pressing the pushbutton. control unit can be switched on using this button. However, this
In the case of the load characteristic the speed range, starting is only possible if the corresponding error has been eliminated.
In the case of switch-off due to overheating this entails a
from the momentary speed value and automatically proceeding cooling off of the corresponding components (LED lights
up to the stop speed. For the run-up characteristic the green).
machine-generator set is rapidly braked to the stop speed in or-
der to run up again to the speed applicable for start or a9 Four-quadrant display
recording. Before starting the characteristic recording the Display of the instantaneous load type of the DUT.
momentary speed can be changed using the incremental Quadrant Load type Rotation direction
control knob . At the end of the characteristic recording the 1st quadrant Motor operation clockwise
2nd quadrant Generator operation couterclockwise
Seite 4/7
732 689USB
a19 USB interface (USB)
This interface is electrically isolated and outputs the speed and
torque of the D & B Unit. If the sockets in the DUT sockets field
are connected, also the current consumption and the voltage
supply of the DUT is transmitted. Control of the control unit can
also be carried out using the software CBM10.
LD DIDACTIC GMBH . Leyboldstrasse 1 . D-50354 Huerth . Phone +49 2233 604-0 . Fax +49 2233 604-222 . e-mail: [email protected]
b3 7 pole connector
to connect the power cord to the control unit (a)
4 Experiment setup
Seite 6/7
LD DIDACTIC GMBH . Leyboldstrasse 1 . D-50354 Huerth . Phone +49 2233 604-0 . Fax +49 2233 604-222 . e-mail: [email protected]