codemaster_xl_m1722a
codemaster_xl_m1722a
H
HP Part No. M1722-91908
Printed in USA November 1994
Edition 6
E1194
0 Notice
Hewlett-Packard makes no warranty of any kind on this material, including but not
limited to, the implied warranties of merchantability and fitness for a particular
purpose. Hewlett-Packard shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the furnishing,
performance, or use of this material.
Before using the instrument, read this guide and become thoroughly familiar with
the contents.
the electrical installation of the relevant room complies with the IEC or national
requirements, and
the instrument is used according to the instructions for use presented in this man-
ual.
NOTE As with all electronic equipment, radio frequency interference between the
defibrillator and any existing RF transmitting or receiving equipment at the
installation site, including electrosurgical equipment, should be evaluated carefully
and any limitations noted before the equipment is placed in service. Hewlett-
ii
Packard assumes no liability for failures resulting from RF interference between HP
medical electronics and any radio frequency generating equipment.
This is to certify that this equipment is in accordance with the Radio Interference
Requirements of Directive FTZ 1046/84. The German Bundespost was notified that
this equipment was put into circulation, the right to check the device for compliance
with the requirements was granted.
If Test and Measurement Equipment is operated with unscreened cables and/or used
for measurements in open set-ups, the user has to assure that under these operating
conditions the Radio Interference Limits are still met at the border of his premises.
Printing History
March 1992 Edition 1
June 1992 Edition 2
December 1992 Edition 3
July 1993 Edition 4
October 1994 Edition 5
November 1994 Edition 6
iii
0 Safety Symbols
Off (Standby)
On/Off
Ground
Shock hazard
CAUTION Caution statements describe conditions or actions that can result in damage to the
equipment or loss of data.
iv
NOTE Notes contain additional information on usage.
v
vi
1 Contents
Getting acquainted
Operating Controls and Indications 1-1
Safety considerations 1-8
AC and DC (Battery) Operation 1-10
Battery Life 1-11
Defibrillating
Defibrillating a Patient 2-1
1. Select Energy 2-2
2. Charge 2-2
3. Shock 2-3
After Using the Defibrillator 2-4
Defibrillating with Alternate Paddle Sets 2-4
Monitoring
Using Leads to Monitor 3-1
Preparing the Leads for Monitoring 3-2
Preparing the Patient 3-4
Monitoring Electrodes 3-5
Monitoring 3-6
Heart Rate Alarms 3-7
Printing the Event Summary Record 3-8
Recording 3-9
Automatic Recordings 3-9
Post Shock Data 3-10
Recorder Errors 3-10
External Monitoring 3-11
Contents-1
Performing Synchronized Cardioversion 4-2
Pacing (Optional)
Using the Pacer 5-2
Defibrillation During Pacing 5-4
Troubleshooting
Troubleshooting 7-1
Troubleshooting the Defibrillator 7-2
Troubleshooting the Pacer 7-5
Troubleshooting SpO2 7-6
Performing Diagnostics 7-7
Operational Checks 7-8
Every Shift 7-9
Every Day 7-9
Every Week 7-12
Every Three Months 7-12
Every Six Months 7-12
Contents-2
Maintaining the Defibrillator
Changing the Recorder Paper 8-1
Cleaning the Recorder Printhead 8-2
Maintaining the Battery 8-3
Battery Capacity Check 8-4
Replacing the Battery 8-5
Cleaning Exterior Surfaces 8-6
Cleaning and Sterilizing the Internal Paddles 8-7
Steam Sterilizing the Internal Paddles 8-7
Ethylene Oxide Sterilization 8-8
Supplies 8-8
Contents-3
List of Tables
Contents - 5
List of Tables
Contents - 6
List of figures
Contents - 4
1 Getting acquainted
This User’s Guide provides operational and basic maintenance instructions for safe
use and proper care of the Hewlett-Packard M1722A/B CodeMaster XL+
defibrillator/monitor.
Figure 1-1
The following figures and tables detail the controls and indications on the
CodeMaster XL+ defibrillator/monitor.
1-1
Getting acquainted
Operating Controls and Indications
Figure 1-2
Charge button
Shock buttons
Sync button
Control Description
Energy Select/ Turns the instrument power on and off and selects energy level.
power control
Charg button Charges defibrillator to energy level set on Energy Select control.
Shock but- Administers shock. Labelled .
tons
1-2
Getting acquainted
Operating Controls and Indications
Figure 1-3
Record button
Mark button
Control Description
Mark • When the recorder is on, pressing Mark will annotate the
ECG at that point.
• If the recorder is not on and the unit is set up to do so, pressing
Mark will print an ECG strip.
See Appendix A for information about setting up the CodeMaster
XL+ for recording when you press Mark (Record on Mark).
1-3
Getting acquainted
Operating Controls and Indications
Figure 1-4
ECG Output
Control Description
HR Controls HR Alarms.
Review Prints an Event Summary record. The message "ES" is printed at the
top of the ECG strip when you print the Event Summary record. The
recorder must be off to print an Event Summary with this key.
1-4
Getting acquainted
Operating Controls and Indications
Figure 1-5
Pacer On button
Rate button
Start/Stop button
Output button
Mode button
Control Description
1-5
Getting acquainted
Operating Controls and Indications
Figure 1-6
SpO2 Input
Control Description
1-6
Getting acquainted
Operating Controls and Indications
Figure 1-7
Charge Done
Sync Light
AC Power
Battery Charge
Indicator Lights
Indicator Description
Charge Done Indicates that unit is charged and ready to use. Charge lights on
key panel and on Apex paddle.
Battery Charge Indicates that the unit is plugged into AC power and that the battery
is being charged.
1-7
Getting acquainted
Safety considerations
Indicator Description
Charge Done tone Sounds when instrument is charged and ready to deliver a shock.
Can be disabled in setup.
Auto disarm tone Sounds during the last ten seconds of the Charge Done tone. Beeps
intermittently until disarmed.
CRT alerts Three beeps each time a message appears on the screen. Can be dis-
abled in setup.
HR alarms Sounds if the heart rate is above the higher alarm limit or below the
lower alarm limit.
SpO2 alarm Sounds if the SpO2 level is above the high SpO2 alarm limit or
below the low SpO2 alarm limit.
Shutdown warn- Alternating pitch sounds for 60 seconds when the system is about to
ing turn off. An alert to plug the unit into AC power.
1 Safety considerations
The CodeMaster XL+ stores high voltage energy and is capable of delivering up to
360 joules of DC energy to a 50 ohm impedance.
• To remove power from the instrument, you must turn the Energy Select control to
Off (Standby). Disconnecting the CodeMaster XL+ from an AC outlet will not
remove power because the instrument is battery powered.
1-8
Getting acquainted
Safety considerations
m Place the paddles in their holders and depress both Shock buttons.
or
m Leave the unit charged for 60 seconds and it will automatically disarm.
CAUTION • Do not leave the instrument turned on when it is not in use and it is not plugged
into AC power.
• Do not discharge the defibrillator with the paddles shorted together. To do so can
cause burning and pitting of the metal paddle contacts.
• Disconnect any other medical electronic equipment from the patient during
• Keep the CodeMaster XL+ and the immediate area clean and
dry at all times to avoid creating potentially dangerous electrical
paths.
1-9
Getting acquainted
AC and DC (Battery) Operation
• The battery will charge when the instrument is connected to AC power even if the
Energy Select switch is in the Off (Standby) position.
• A fully depleted battery will recharge to 90% of full capacity in two hours and
100% capacity in 18 hours. To preserve battery integrity, the battery must be fully
recharged each time the battery is depleted.
• A new battery or one that has been stored for an extended period requires 24 hours
of charging before use.
• When the unit is not in use, connect it to AC power with the monitor turned off.
This is to maintain a full battery charge and to prolong battery life.
• To operate on internal battery power only, disconnect the power cord from the AC
outlet.
• A fully charged battery will nominally provide fifty 360 joule charge-shock cycles
or 2.5 hours of continuous monitoring (15C-40C).
1-10
Getting acquainted
AC and DC (Battery) Operation
NOTE Continuous recording will reduce monitoring time available when you are using the
unit on battery power.
CAUTION When the LOW BATTERY message is displayed on the monitor, plug the unit into
AC power.
From the time the LOW BATTERY message is first displayed to when the battery
capacity is fully depleted (instrument shutdown), there is typically enough reserve
battery capacity to provide either 30 minutes of monitoring, or five 360 joule
charge-shock cycles.
A unique audible alarm will sound continuously when there are 60 seconds of
battery capacity remaining. The instrument will automatically shut off after 60
seconds.
If the battery has been fully depleted, plugging the instrument into AC will
immediately restore full operation.
Frequent battery discharges to the low battery level will degrade battery life.
Battery Life
The sealed lead-acid battery used in the CodeMaster XL+ will provide optimum life
when the unit is continually connected to AC power when not in use. The battery
operates best when it is fully charged after each use. To fully charge a depleted
battery requires 18 hours of continuous charge time. Because it is not always
practical to allow a full charge cycle between uses, the CodeMaster XL+ can charge
a depleted battery to 90% of its capacity in approximately two hours. However,
battery capacity and battery life will be reduced if the battery is not allowed to fully
charge after each use. For improved battery life, consider ways to reduce the number
of instrument uses between full charge cycles.
When the instrument is not plugged into AC power, some current is drawn from the
battery to maintain memory and startup logic. Remove the battery if the instrument
1-11
Getting acquainted
AC and DC (Battery) Operation
is to be stored for extended periods (more than one month) without AC power. Note
on the instrument that the battery has been removed. After an extended storage
period, test the battery according to the battery capacity check as described in
Maintaining the Battery on page 8-3.
This battery was selected because it provides optimum performance and battery life
over a wide range of operating conditions. The life expectancy of this battery is
dependent on many variables, including temperature and usage. Periodically check
the battery capacity to determine whether to replace it. The battery capacity check is
described in Maintaining the Battery on page 8-3.
NOTE When plugged into AC power, the CodeMaster XL+ will function normally with no
battery installed, however the time required to charge the defibrillator will increase.
1-12
2 Defibrillating
This chapter contains information about defibrillating a patient and using different
paddle sets for defibrillating.
Figure 2-1
Charge button
Shock buttons
Sync button
2 Defibrillating a Patient
1 Select Energy
2 Charge
3 Shock
2-1
Defibrillating
Defibrillating a Patient
1. Select Energy
1 Turn the Energy Select control to the desired energy level. The defibrillator is
now on.
4 Rub the paddles lightly against the skin to increase contact between the patient
skin and the paddles. Then keep the paddles still to reduce motion artifact on the
monitor.
WARNING Do not spread paste between the paddle electrodes on the chest.
The patient can be burned if the paste forms a path between the
electrodes.
2. Charge
Press the Charge button on either the right (Apex) paddle or on the instrument front
panel.
2-2
Defibrillating
Defibrillating a Patient
Wait for the Charge Done indicators. When the unit is armed, the monitor Delivered
Energy display shows the available energy in joules.
To increase or decrease the selected energy level after pressing the Charge button,
perform the following steps.
3. Shock
To shock the patient, perform the following steps.
1 Briefly adjust paddle pressure and placement to optimize patient contact, as reg-
istered on the paddle contact indicator (if supplied).
2 Verify that no one is in contact with the patient, the monitoring cable or leads,
the bed rails, or another potential current pathway.
3 Call out "Clear!" to alert personnel to stand away from the patient.
WARNING Keep hands clear of the paddle electrode edges. Use your thumbs
to depress the Shock buttons on the paddle handles.
4 Press and briefly hold both Shock buttons (one on each paddle) simultaneously,
to deliver energy to the patient.
If the defibrillator does not shock, refer to Chapter 7, Troubleshooting.
NOTE If you must disarm the charged defibrillator (if countershock is not needed), turn the
Energy Select control to Monitor On. Any stored energy will be discharged
internally and the available energy on the display will return to 0.
2-3
Defibrillating
Defibrillating a Patient
2 Return the instrument to its storage location, and plug the power cord into an AC
power outlet. Verify that the BATT and [ AC lights are on.
3 Clean all paddles, controls, and cables. Refer to Chapter 8, Maintaining the
Defibrillator for cleaning instructions.
4 Check that sufficient recorder paper and electrolyte paste or defibrillator pads
are available for the next use of the defibrillator.
• Internal Paddles
The CodeMaster XL+ paddle set comes with pediatric paddles. To use the pediatric
paddle set, depress the release latch at the front of the standard external paddle set
while pulling forward on the adult paddle surface. This action will remove the adult
paddle contact surface and expose the smaller pediatric contact surface.
2-4
Defibrillating
Defibrillating a Patient
WARNING The clinician must select an appropriate energy level for the
pediatric patient. There is no energy limit lockout for the
pediatric paddle set.
The CodeMaster XL+ has an external pads adaptor that is optional. This adaptor
allows defibrillation through external adhesive pads.
• You can perform synchronized cardioversion without using an ECG lead set, while
monitoring through the pads.
• If the optional pacer is installed, you can switch between the pacing and defibrilla-
tion modes of operation quickly.
1 Attach the pads adaptor cable (HP M1750A/B) to the paddle connector on the
front of the defibrillator.
2 Slide the paddle connector lock towards the front of the defibrillator to secure
the cable.
2-5
Defibrillating
Defibrillating a Patient
4 Connect the pads to the pads adaptor cable. The pads are correctly connected
when the locking ring is twisted, locking the ears of the connector to the adaptor
cable.
WARNING Failure to correctly connect the pads to the adaptor cable can
result in a failure to deliver energy to the patient.
If the PADS OFF monitor message is displayed, check all patient connections.
5 Select pads as the ECG source by pressing Lead until PADS appears on
the display under the heart rate.
7 Press Charge .
9 Press both Shock buttons at once to defibrillate. The Shock buttons for
external pads are on the cable connector housing.
You can perform internal defibrillation using one of the optional internal
defibrillation paddle sets.
The Switchless paddle sets attach to the internal paddles adaptor cable (M1740A/B).
The Shock buttons are on the connector housing of the internal paddle adaptor
cable.
The Switched paddle set does not require the internal paddles adaptor cable. This
paddle set has a single Shock button on the right-hand paddle handle.
2-6
Defibrillating
Defibrillating a Patient
1 Attach the internal paddles adaptor cable (HP M1740A/B) to the paddle connec-
tor on the front of the defibrillator.
2 Slide the paddle connector lock towards the front of the defibrillator to secure
the cable.
3 Select correct paddle set size from Table 2-1, “Switchless Internal Paddles
Selection,” on page 2-8.
4 Attach the internal paddles set to the internal paddles adaptor cable. The pads
are correctly connected when the locking ring is twisted, locking the ears of the
connector to the adaptor cable.
WARNING If the energy switch is set to a level greater than 50 joules, 50J MAXIMUM will be
displayed. If Charge is pressed, the unit will only charge to 50 joules.
WARNING For safety and sterility when using the paddles, do not touch the
paddle beyond the fingerguard on the handle.
7 Press Charge .
9 Press both Shock buttons at once to defibrillate. The Shock buttons are on
the cable connector housing.
2-7
Defibrillating
Defibrillating a Patient
HP Part
Description
Number
To defibrillate a patient internally, using the switched paddles, perform the following
steps.
1 Select correct paddle set size from Table 2-2, “Switched Internal Paddles Selec-
tion,” on page 2-9.
2 Attach the switched internal paddles cable to the paddle connector on the front
of the defibrillator.
3 Slide the paddle connector lock towards the front of the defibrillator to secure
the cable.
NOTE If the energy switch is set to a level greater than 50 joules, 50J MAXIMUM will be
displayed. If Charge is pressed, the unit will only charge to 50 joules.
WARNING For safety and sterility when using the paddles, do not touch the
paddle beyond the fingerguard on the handle.
2-8
Defibrillating
Defibrillating a Patient
6 Press Charge .
HP Part
Description
Number
2-9
Defibrillating
Defibrillating a Patient
2-10
3 Monitoring
This chapter contains information about monitoring a patient with the CodeMaster
XL+ defibrillator/monitor. This chapter also contains details of patient preparation
that apply to the synchronized cardioversion and pacing procedures described later
in this manual.
The CodeMaster XL+ can be used for either short term or long-term cardiac
monitoring. A fully charged battery pack provides a minimum of two and a half
hours of continuous monitoring. The power cord can be connected to AC power for
unlimited monitoring periods.
The CodeMaster XL+'s monitoring functions can be used for cardiac monitoring,
elective cardioversion, and pacing (optional). Table 3-1, “Cardiac Monitoring
Configurations,” on page 3-1 details the different ECG sources that can be used for
cardiac monitoring and monitoring applications for which each is suited.
3-1
Monitoring
Using Leads to Monitor
Table 3-2, “3-Wire Electrode Placement,” on page 3-2 describes typical lead-wire
placement using the 3-wire patient cable. Table 3-3, “Lead Formation,” on page 3-3
shows how the individual leads are formed using the individual leadwires.
Electrode Placement
LA/Black Near the left midclavicular line, directly below the clavicle.
LL/Red Below the left pectoral muscle on the left midclavicular line.
3-2
Monitoring
Using Leads to Monitor
Lead + — ref
I LA RA LL
II LL RA LA
III LL LA RA
Table 3-4, “5-Wire Electrode Placement,” on page 3-3 describes a typical lead-wire
placement using the 5-wire patient cable. Table 3-5, “Lead Configurations,” on page
3-3 shows how the individual leads are formed using the individual leadwires.
Electrode Placement
LL/Red Below the left pectoral muscle on the left midclavicular line.
RL/Green Below the right pectoral muscle on the right midclavicular line.
I LA - RA
II LL - RA
III LL - LA
3-3
Monitoring
Preparing the Patient
1 If necessary, shave hair from the site to ensure good electrode to skin contact.
2 Clean the skin with soap and water or with alcohol, then wipe it dry.
NOTE You can safely monitor a patient during defibrillation. However, monitoring
electrodes can become polarized during defibrillation shock, causing the ECG
waveform to briefly disappear from the display. You can reduce this effect by using
silver- silver chloride electrodes.
NOTE Be careful to correctly align the cable plug when connecting the patient ECG leads
cable to the defibrillator/monitor. Correctly orient the cable plug key with the
3-4
Monitoring
Preparing the Patient
defibrillator connector slot. If the ECG leads cable falls off or is incorrectly
connected, the message LEADS OFF appears on the display.
Monitoring Electrodes
Using Pads
Standard pads (HP1749A) allow you to monitor through the pads for defibrillation
and synchronized cardioversion. If you wish to use the pads for pacing however, you
must attach separate electrodes for monitoring. To use standard pads, perform the
following steps.
Using Paddles
For an emergency evaluation you can monitor a patient's ECG through the paddles
electrodes when leads are not attached to the patient.
3-5
Monitoring
Monitoring
3 Monitoring
Figure 3-1
ECG Output
To monitor a patient's ECG with the CodeMaster XL+, perform the following steps.
3 Press Lead to select the ECG source. The selected source appears in the
upper right corner of the display. For example, PADS appears on the display
when it has been selected.
m If the message LEADS OFF or PADS OFF appears on the display, inspect the
electrodes, patient cable, leadwires, and associated connections. If the selected
ECG source is not connected, a dashed line will replace the normal ECG trace.
3-6
Monitoring
Heart Rate Alarms
4 Ensure that the ECG size has been automatically adjusted for optimal size. If
you wish to reduce the ECG size, press ECG Size . The "gain bar" along the
left side of the display represents 1 mV of signal amplitude.
NOTE Autogain allows an initial quick setup when the instrument is turned on. To remove
the instrument from Autogain, press ECG Size . You then must adjust the ECG
size manually.
The CodeMaster XL+ provides three configurable pairs of upper and lower heart
rate alarm limits. Each pair of heart rate alarm limits can be defined in setup menu 1
as described in Appendix A. While monitoring, you can select and enable any pair
of pre- defined limits using HR .
When HR alarms are inactive, the monitor will display a bell symbol with a "\"
through it.
To select a pair of HR alarm limits, press HR until the pair of limits you
wish to use are displayed. If you do not press the key again, the displayed HR alarm
limits become active and the limits are replaced by the bell symbol.
If the HR alarm limits are violated, the HR alarm limits replace the bell symbol, and
the violated limit is highlighted. Pressing HR at this point will turn off the
HR alarms.
If the HR Alarms are active and you wish to review the limits, press HR .
The currently active pair of HR alarm limits are displayed momentarily.
If the HR alarms are active and you wish to select another pair of limits, press
HR until the pair of limits you wish to use are displayed. Pressing
HR repeatedly cycles through the three pairs of HR alarm limits and the HR
alarms inactive choice.
3-7
Monitoring
Printing the Event Summary Record
NOTE HR alarms are automatically turned off when you press Charge .
Mark Mark Marker symbol ( ) annotates strip at point Mark was pressed.
• To print the Event Summary on the recorder, press Review . The recorder must
not be printing to print an Event Summary with this key. After printing an event,
you must wait 10 seconds before printing another event.
• To review the Event Summary later, turn the unit on and press Review .
3-8
Monitoring
Recording
NOTE The Event Summary record is cleared each time the defibrillator is turned on and a
new event occurs. This allows you to turn off the defibrillator and return later to
review event information such as code statistics.
Turn the defibrillator off between uses to ensure that Event Summary records are
patient specific.
3 Recording
To print a record of the current ECG and of the monitor status, press Record .
• The upper line of the ECG strip contains a periodic report of monitor parameters
(Date, Time, Heart Rate, ECG Source, ECG Size, and Recorder mode).
• The lower line of the ECG strip records asynchronous events such as Shock deliv-
ery or Heart Rate Alarm violations.
• Several graphic symbols are used to annotate events such as Shock, HR Alarms,
Mark, or Sync.
A 1 mV, 200 ms calibration pulse can be printed on the ECG strip by pressing both
arrows on the ECG Size key simultaneously.
Automatic Recordings
In setup menu 2, you can enable or disable any of the following automatic
recordings:
• Record on Mark
• Record on Charge
3-9
Monitoring
Recording
• Record on Shock
• Record on Alarms
The automatic recordings for both delayed and non-delayed recorder modes of
operation are defined in Table 3-7, “Automatic Recordings,” on page 3-10.
Delayed mode
Delayed mode Non- Delayed mode
Event Post-event
Pre-event time Post-event time
time
• If Post Shock Data is enabled, the defibrillator will record the shock delivery sta-
tistics (Actual Delivered Energy, Patient Impedance, Peak Current).
• If Post Shock Data is disabled, the defibrillator will record the energy to which it
was charged as the delivered energy. For example, if the unit was charged to 200J,
the delivered energy annotation on the ECG strip would be DEL 200J.
3-10
Monitoring
External Monitoring
Recorder Errors
The message CHECK RECORDER appears if an error occurs while recording. If this
message appears, check the recorder paper supply. This message may also appear if
the recorder door is open.
3 External Monitoring
The ECG output provides an analog 1V/mV ECG signal for connection to an
external monitor. Compatible external monitoring divider cables are listed in Table
3-8, “External Monitoring Cables,” on page 3-11.
1000:1 voltage
Divider Cable
HP Part No.
Connector
Type
The ECG output is also compatible with the interface to the HP central station ECG
input.
NOTE Do not use the ECG output to synchronize another defibrillator. (The ECG-In to
ECG-Out delay is 35 milliseconds.)
CAUTION The connection of external equipment may increase leakage currents. Always
request that local safety personnel verify that multiple connected equipment comply
with local regulatory standards before putting such equipment into service.
3-11
Monitoring
External Monitoring
3-12
4 Performing Synchronized Cardioversion
4 Performing Cardioversion
4-1
Performing Synchronized Cardioversion
Performing Cardioversion
2 Plug the input end of the cable from the monitoring equipment into the ECG
input plug on the CodeMaster XL+.
If the PADS OFF monitor message is displayed, check all patient connections.
3 Press Sync once to place the instrument in Sync mode. The message SYNC
will appear on the display.
NOTE If the paddles are selected as the ECG source, the message USE LEADS will appear
on the display. Although the instrument will allow synchronized shock in paddles
4-2
Performing Synchronized Cardioversion
Performing Cardioversion
ECG mode, leads mode is recommended. Artifact induced by moving the paddles
may resemble an R-wave and trigger defibrillator shock.
4 Select the desired energy level with the Energy Select control.
1 Ensure that the paddles connector is attached and the paddles are in their hold-
ers.
2 Ensure that the patient cable is connected to the ECG Input jack in the front of
the defibrillator and the leads are attached to the patient.
WARNING Keep hands clear of the paddle electrode edges. Use your thumbs
to depress the Shock buttons on the paddle handles.
4 Press Charge . Wait for the Charge Done indicators, and for the Delivered
Energy display to read 100 joules.
5 With the paddles pressed firmly into their holders, press and hold both Shock
buttons at once. The defibrillator will shock with the next detected R-wave.
4-3
Performing Synchronized Cardioversion
Performing Cardioversion
After the shock, the ECG strip will indicate whether the 100 joule test shock
passed or failed.
If the defibrillator does not charge, refer to Chapter 7, Troubleshooting.
c. Gently rub the electrode surfaces together to evenly distribute the applied
paste.
3 Rub the paddles slightly against the skin to increase the paddle-to-patient con-
tact.
5 Press the Charge button on either the right (Apex) paddle or on the instrument
front panel.
4-4
Performing Synchronized Cardioversion
Performing Cardioversion
NOTE If you must disarm the charged defibrillator (if countershock is not needed), turn the
Energy Select control to Monitor On. Any stored energy will be discharged
internally and the available energy on the display will return to 0.
To increase or decrease the selected energy level after the Charge button has been
pressed, move the Energy Select control to the new energy level, and wait for the
Charge Done indicators.
1 Verify again that the ECG waveform is stable, and that a marker dot appears
only with each R-wave of the cardiac cycle.
3 Depress both Shock buttons (one on each paddle) until shock occurs. The
defibrillator will shock with the next detected R-wave.
NOTE Depending on how the unit has been configured, it will either remain in the
synchronized shock mode or it will return to defibrillator mode following a
synchronized shock. Refer to Appendix A for instructions on configuring the
defibrillator for operation after synchronized cardioversion.
After using the defibrillator, perform the following steps to ensure that the
defibrillator is ready for use.
4-5
Performing Synchronized Cardioversion
Performing Cardioversion
3 Plug the power cord into an AC power outlet. Verify that the BATT and
AC lights are on.
4 Clean all paddles, controls, and cables as necessary. Refer to Chapter 8, Main-
taining the Defibrillator for information on cleaning the defibrillator.
5 Check that there are adequate recorder paper, electrolyte paste, and defibrillator
pads for the next use.
4-6
5 Pacing (Optional)
This chapter contains information about pacing with the CodeMaster XL+
defibrillator/monitor.
Figure 5-1
Pacer On button
Rate button
Start/Stop button
Output button
Mode button
The CodeMaster XL+ with the optional pacer can perform external transcutaneous
pacing. The pacing option provides demand (synchronous) and fixed
(asynchronous) pacing modes. The patient is connected to the pacer by external
adhesive pads. The patient can be paced and defibrillated through the same set of
pads.
WARNING While pacing, avoid touching the gelled area of the pacing pads or
the patient to prevent electrical shock.
Do not use pads for more than eight hours of continuous pacing.
5-1
Pacing (Optional)
Using the Pacer
The CodeMaster XL+ will pace on battery power alone. However, whenever
possible, plug the CodeMaster XL+ into AC power while pacing.
3 Attach the patient cable to the CodeMaster XL+'s ECG Input connector.
5 Attach the pads adaptor cable (M1750A/B) to the defibrillator output connector.
Pull the latch connector toward the front of the defibrillator to lock the connec-
tor in place.
6 Attach the pads to the pads adaptor cable and turn the twist lock.
8 Press Pacer to turn the pacer on. Pacer parameters will now be displayed at
the bottom of the display (PACER STOP, DEMAND MODE, 70 PPM 30
MA). The rate (ppm) and output (mA) settings for when the pacer is turned on
can be selected in setup menu 1. The original rate and output settings from the
factory are 70 ppm and 30 mA. The pacer is always in Demand mode when it is
turned on.
5-2
Pacing (Optional)
Using the Pacer
NOTE At this point, no pacer pulses are being delivered to the patient. The pacer must be
started before the pacer pulses are delivered at the selected rate (ppm) and output
(mA).
9 Press Lead to select the best Lead for monitoring while pacing. You can
only select Leads as the ECG source when the pacer is on.
m If the message LEADS OFF is displayed, check all patient cable connections.
10 Press Rate to adjust the rate. The selected rate (PPM) is displayed on the
monitor.
11 Press Mode to select the pacing mode (Demand Mode/Fixed Mode). The
selected mode is displayed on the monitor.
m When in Demand mode, the pacer will only deliver pacer pulses when the
patients heart rate is lower than the selected pacer rate.
m When in Fixed mode, the pacer will deliver pacer pulses at the selected pacer
rate.
WARNING Use Demand pacing mode whenever possible. Use Fixed pacing
mode (Asynchronous) for cases when reliably monitoring the
patient is impractical. For example, use Fixed mode when there is
motion artifact or other ECG noise that makes R-wave detection
unreliable.
12 Press Start/ to start pacing. The monitor will now display the message
PACING as well as the selected mode, rate and output.
The pacer will not start pacing if there is a problem with either the pacing pad
connections or the monitoring electrode connections.
If there is a problem with the pacing pad connection, the message ATTACH
PADS will be displayed briefly when you press Start/ .
13 Verify that the pacer pulses are well positioned in the diastole.
5-3
Pacing (Optional)
Using the Pacer
14 Increase output (mA) by pressing Output until the beat is captured. Selecting
an alternate lead can help you to determine capture.
15 To set the lowest possible output level to capture, decrease the current by decre-
ments of 5 mA by pressing Output .
NOTE If the monitoring ECG lead falls off while pacing in Demand mode, the pacer will
stop delivering pulses and the messages PACER STOP and LEADS OFF will
appear. To resume pacing, reattach the lead and press Start/ . Pacing in Fixed
mode does not require leads to be attached for the pacer to deliver pulses.
If a pacing pad comes off during pacing, the pacer will stop delivering pulses and the
messages PACER STOP and PADS OFF will appear. To resume pacing, reattach
the pad and press Start/ .
WARNING HR meters and HR alarms function during pacing, but they can
be unreliable. The HR meter attempts to count QRS activity in
both Demand and Fixed pacing modes. Observe the patient
closely while pacing. Do not rely on HR alarms or the indicated
Heart Rate as a measure of the patient's health.
2 Press Charge . The defibrillator will automatically turn off the pacer and start
charging. The pacer status messages will be cleared and replaced with the
defibrillator Delivered Energy Display.
4 Call out "Clear!" to alert personnel to stand away from the patient.
5 Press and briefly hold both Shock buttons. The shock will be delivered through
the multifunction pads.
5-4
Pacing (Optional)
Using the Pacer
After the shock, the pacer remains off. Resume pacing if it is required.
5-5
Pacing (Optional)
Using the Pacer
5-6
7 Troubleshooting
7 Troubleshooting
This section provides information about messages that appear on the display.
7-1
Troubleshooting
Troubleshooting
50J MAXI- Internal paddles are attached and CodeMaster XL+ will charge
MUM the Energy Select control has been to 50 joules.
turned past 50 joules.
7-2
Troubleshooting
Troubleshooting
3 If the instrument remains unable to charge, turn the Energy Select control to Off
(Standby) and use a backup defibrillator.
4 Press Revie to print an event summary and keep any ECG strips from the
defibrillator for later evaluation.
If the defibrillator does not deliver a shock, perform the following steps.
7-3
Troubleshooting
Troubleshooting
1 Make sure the unit is not in synchronized shock mode. The SYN light is on
when the unit is in synchronized shock mode.
2 If the unit discharges internally (that is, the energy display decrements slowly
then beeps three times and displays the screen message DEFIB DISARMED)
verify proper connections from the patient to the defibrillator. This includes con-
nections to the pads/paddles adaptor cable if one is being used. Also check for
worn or broken areas along the cables.
5 If the unit remains unable to shock, turn the Energy Select control to the Moni-
tor On or Off (Standby) position and use a back-up defibrillator.
6 Press Revie to print an event summary and keep any ECG strips from the
defibrillator for later evaluation.
7-4
Troubleshooting
Troubleshooting
PACER The unit detected a delivered current 1 Change pads and make
FAILURE error. sure pads adaptor cable is
properly connected.
2 Turn the Energy Select
switch to Off (Standby),
then turn back to Monitor
On.
3 Restart the pacer.
If PACER FAILURE hap-
pens again, use a back-up
pacer and call for service.
PACER OUT- The pacer cannot deliver the 1 Stop the pacer.
PUT LOW required current. 2 Change the pads and
Pacer current check the pads adaptor
highlighted on cable for proper connec-
the display. tion.
3 Restart the pacer.
STOP PACER An attempt was made to change pac- Stop the pacer before you
ing mode while pacing. change the pacing mode.
NO PADS Pacer is on and one of the following Attach pads for pacing.
occurs:
• No pads adaptor attached.
• Paddles attached.
• Internal paddles attached.
PADS OFF Pads adaptor cable is attached and a Attach pad for pacing. If pads
pad is off. are already applied to patient,
change to a new set of pads.
7-5
Troubleshooting
Troubleshooting
Troubleshooting SpO2
Table 7-4 SpO2 Messages
SPO2 SENSOR The sensor or adaptor cable is Replace the sensor or adaptor
FAIL broken. cable.
Dashes appear on dis- Can’t derive measurement Check the patient for a pulse.
play instead of SpO2 because: Reapply the sensor, and make
reading. • The sensor is not on the sure it is correctly positioned.
patient. If it doesn’t work, replace the
• No pulse is detected. sensor.
• The sensor is incorrectly
positioned.
• The sensor is defective.
SPO2 LOW SIG- Bad connection to patient, or Check the patient for poor
NAL patient has poor perfusion. perfusion. Reapply disposable
and semi-disposable sensors.
Readjust reusable sensors.
Consider using a different
sensor site.
7-6
Troubleshooting
Performing Diagnostics
7 Performing Diagnostics
1 Plug the power cord into an AC outlet. Verify that the BATT and
AC lights are on.
2 Turn the Energy Select control to the Monitor On position. The monitor trace
will appear within ten seconds.
3 Press Lead until Lead I is displayed and verify that the message LEADS
OFF appears, indicating that one or more leadwires are not connected.
5 With the paper installed, press Recor once to turn on the recorder.
a. Allow the recorder to run for approximately 20 seconds and check that
Date, Time, HR (heart rate), PADDLES (ECG source), and AUTOGAIN
(ECG gain mode) are noted on the ECG strip.
b. Press Mark and verify that the mark (t) symbol is printed. It will be
delayed 6 seconds if the recorder is in the delayed mode.
c. Press Recor to stop the recorder.
6 Press HR . Verify that the configured alarm limits appear briefly, then
are replaced by the bell symbol bell in the upper right of the display. With no
ECG signal, an audible alarm tone should sound within four seconds.
7-7
Troubleshooting
Operational Checks
WARNING Keep hands clear of the paddle electrode edges. Use your thumbs
to depress the Shock buttons on the paddle handles.
11 Grasp the paddle handles, and without removing the paddles from their holders,
press both Shock buttons simultaneously. A brief automatic recorder run prints
the delivered energy test report.
13 Verify that the messages SYNC and USE LEADS appear on the display.
14 Press Lead once to select Lead I, and verify that the message USE
LEADS no longer appears. The message LEADS OFF should now appear, indi-
cating that one or more leads are not connected.
7 Operational Checks
These checks are intended to briefly verify the proper operation of the CodeMaster
XL+ defibrillator/monitor. Regularly perform a test routine incorporating the
following checks along with visual inspection of all cables, paddles, and controls.
Every Shift
Perform the following checks every shift.
7-8
Troubleshooting
Operational Checks
o Check for ECG leads, electrodes, and adequate REDUX® electrolyte paste or
defibrillator pads.
Every Day
o Visually check AC power cord for wear.
o Visually check the patient cables, paddles, cables, and pads adaptor cables for
wear, insulation nicks, and other damage.
o Ensure that the BATT and AC lights are on. If the unit is
plugged in and the AC light is not on, the power cord may have a
broken wire.
o Check that the BATT and AC lights go off when the unit is
unplugged.
o Perform the Delivered Energy and Shock Button Functional Test, which
follows.
o If your instrument has the external pacing option, perform the Quick Pacer
Functionality Test as described in Quick Pacer Functionality Test on page 7-
11.
To check the instrument with the paddles, perform the following steps.
2 Push the paddles completely into their holders (Apex paddle in right pocket,
Sternum in left) and press either Charge button. Wait for the Charge Done indi-
cators.
7-9
Troubleshooting
Operational Checks
WARNING Keep hands clear of the paddle electrode edges. Use your thumbs
to depress the Shock buttons on the paddle handles.
3 With the paddles in their holders, grasp the paddle handles and press the Apex
paddle Shock button. Verify that the defibrillator does not discharge.
4 Release the Apex paddle Shock button, then press the Sternum paddle Shock
button. Verify that the defibrillator does not discharge.
6 Press and hold both Shock buttons. Verify that the defibrillator does not dis-
charge.
8 With the paddles in their holders, press and briefly hold both Shock buttons at
once.
To check the instrument with the pads adaptor cable, perform the following steps.
1 Connect the Test Load (M1781A) to the Pads Adaptor Cable (M1750A/B).
4 Press each Shock button independently. Verify that the defibrillator does not dis-
charge.
5 While depressing Charge, press and briefly hold both Shock buttons.
7-10
Troubleshooting
Operational Checks
NOTE Notify Service Personnel if the ECG strip prints TEST 100J FAILED or if any of
the Shock button tests fail.
7 Disconnect the Test Load (M1781A) from the Pads Adaptor Cable
(M1750A/B).
1 Connect the pads adaptor cable (M1750A/B) to the defibrillator and the test load
(M1781A) to the adaptor cable. Turn the unit on by turning the Energy Select
switch to Monitor On.
4 Adjust the pacer current to 30mA by pressing Output . Adjust the pacer rate
to 60ppm by pressing Rate .
5 Start the pacer by pressing Start/Sto , and start the recorder by pressing
Recor .
6 Verify that the pacer pulses are shown on the recorder strip approximately every
five large boxes (if the recorder is in delay mode it will take several seconds
before the pacer pulses appear). Allow the pacer to run for 10-12 seconds.
7 Turn the pacer off by pressing Pacer On and stop the recorder by pressing
Recor . Disconnect the test load (M1781A) from the adaptor cable.
• The unit beeps three times and displays PACER OUTPUT LOW.
• The pacer pulses are not shown on the recorder strip as described in the test.
7-11
Troubleshooting
Operational Checks
Every Week
Perform the following checks on the internal paddle set every week.
o Check for excessive residue from sterilization on the paddle set and clean as
needed. Oxidation can be an indication the paddle set is old and must be
replaced.
o Ensure that the cable, connector, and electrodes have no cracks in the
insulation.
7-12
8 Maintaining the Defibrillator
This chapter contains information about maintaining and cleaning the CodeMaster
XL+ defibrillator/monitor.
Figure 8-1
1 Slide the recorder door to the right of the defibrillator as shown in Figure 8-1, on
page 8-1. The paper platen will tilt up.
2 Pull up on the plastic removal tag to remove the empty or low paper roll.
8-1
Maintaining the Defibrillator
Cleaning the Recorder Printhead
3 Place a new roll of thermal paper (HP 40457C/D) in the recorder so that the
paper unrolls from the top of the roll and the grid faces down as it comes out of
the recorder.
5 While holding the recorder door open, (to the right) press the platen down over
the paper.
If the ECG strip has light or varying density printing, clean the printhead to remove
any possible buildup of paper residue.
1 Slide the recorder door to the right of the defibrillator. The paper platen will tilt
up.
3 Clean the printhead surface (above the brush) with rubbing alcohol and a cotton
swab.
5 While holding the recorder door open, (to the right) press the platen down over
the paper.
8-2
Maintaining the Defibrillator
Maintaining the Battery
The sealed lead-acid battery used in the CodeMaster XL+ will provide optimum life
when the unit is continuously connected to AC power and fully charged after each
use. To fully charge a depleted battery requires 18 hours of continuous charge time.
Because it is not always possible to allow a full charge cycle between uses, the
CodeMaster XL+ was designed to charge a depleted battery to 90% of its capacity in
approximately two hours. However, battery capacity and battery life will be reduced
if the battery is not allowed to fully charge after each use. For improved battery life,
applications where the CodeMaster XL+ is used frequently between full charge
cycles, consider ways to decrease the number of instrument uses between full charge
cycles.
When the instrument is not plugged into AC power, some current is drawn from the
battery to maintain memory and startup logic. If the battery is to be stored for
extended periods (more than one month) without AC power, observe the caution in
Appendix A, Installing and Charging the Battery.
This battery was selected because it provides optimum performance and battery life
over a wide range of operating conditions. The life expectancy of this battery is
dependent on many variables, including temperature and usage. Periodically check
the battery capacity to determine whether to replace it. The battery capacity check is
described in the next section.
NOTE When plugged into AC power the CodeMaster XL+ will function normally with no
battery installed, however the time required to charge the defibrillator will increase.
8-3
Maintaining the Defibrillator
Maintaining the Battery
A new battery will provide a minimum of 2.5 hours of monitoring time. Hewlett-
Packard recommends that you replace the battery when it fails to provide a
minimum of 2.5 hours of monitoring time or 10 minutes of Low Battery warning
time when starting from a fully charged battery.
To test the battery capacity, perform the following steps. This test will require up to
3.5 hours to perform (not including battery charge times). Allow 30 hours to test the
battery and have it ready for use again.
1 Charge the battery by plugging the instrument into AC power for eight hours.
Verify that the AC and BATT lights are on.
3 Depress Sync HR while turning the Energy Select control from Off
(Standby) to Monitor On. While pressing Sync , turn the Energy Select con-
trol from the Off position to the Monitor On position. The Setup/Diagnostic Ser-
vice menu will appear in a moment.
8-4
Maintaining the Defibrillator
Cleaning Exterior Surfaces
8 Fully recharge the battery by plugging the instrument into AC power for 18
hours. Verify that the AC and BATT lights are on.
CAUTION The battery can be permanently damaged if left uncharged for a prolonged period.
7 Replace the battery door and secure it by turning the two retaining screws a
quarter-turn clockwise.
8 Perform the Battery Capacity Check above before placing instrument in service.
The CodeMaster XL+ and its accessories are chemically resistant to common
hospital cleaning solutions and non-caustic detergents. The following list includes
some approved cleaning solutions.
8-5
Maintaining the Defibrillator
Cleaning and Sterilizing the Internal Paddles
o Ammonia-based cleaners
o Keep the outside of the instrument clean and free of dust and dirt. Clean the
paddles thoroughly to prevent build-up of dried electrolyte paste.
o Do Not allow any fluids to penetrate the instrument case. Avoid pouring fluid
on the unit while cleaning.
o Do not steam sterilize the monitoring leads, submerge them for prolonged
periods, or heat them above 50ºC. If metallic surfaces become oxidized, clean
them with a very light abrasive (toothpaste). Do not use highly abrasive
cleaners such as steel wool or silver polish.
o Clean the electrode surface and handle with standard hospital solution.
o Use a small, soft brush with cleaning solution to clean any contamination from
the electrode surface and edges.
8-6
Maintaining the Defibrillator
Cleaning and Sterilizing the Internal Paddles
Internal paddle sets can be sterilized using any of four different methods. Three of
these methods use steam sterilization, the fourth method uses Ethylene Oxide (EtO).
Although the internal paddle sets are constructed using the highest quality materials,
the severe conditions of steam sterilization will limit their useful life.
Prevacuum Sterilization
Flash Sterilization
Preparation Unwrapped
Exposure Temperature 132–135ºC (270–275ºF)
Exposure Time 3 minutes
Gravity Sterilization
8-7
Maintaining the Defibrillator
Supplies
8 Supplies
Battery
Patient Cables
8-8
Maintaining the Defibrillator
Supplies
Power Cords
Paddles
*
Does not comply with IEC 601-2-4 standard for contact area.
8-9
Maintaining the Defibrillator
Supplies
Pads
Electrolyte (Redux®)
651-1008-050 Redux®
Paper
Cases
8-10
A Installation and Setup
1 Installation
The CodeMaster XL+ is ready for operation when the following tasks have been
properly performed:
o Install battery.
o Install paper.
o Make sure that the paddle set connector is seated and locked.
WARNING Use only three wire power cords with three-pronged grounded
plugs. Make sure that the outlet accepts the three-pronged plugs
and is grounded. Never adapt a three- pronged plug to fit a two-
pronged outlet.
A-1
Installation and Setup
Installation
NOTE To ensure full battery capacity, charge the battery for 24 hours following its
installation in the defibrillator.
CAUTION If the defibrillator will be stored for longer than one month without AC power, first
charge the battery for 48 hours, then remove the battery from the unit. Note on the
instrument that the battery has been removed. Store the battery in a cool, dry
location. Recharge a stored battery for at least 24 hours every six months. This will
ensure that the battery does not completely discharge while in storage. The battery's
shelf life is longer with cooler temperatures, but the battery must not be stored
below freezing. After an extended storage period, the battery should be tested using
the “Battery Capacity Check” (page 8-4).
A-2
Installation and Setup
Installation
Figure 1-1
Battery
Battery compartment
Battery plug
Battery
Paddles Connector
A-3
Installation and Setup
Installation
1 Slide the paddle connector lock on the paddles plug to the unlock position. To
do this, push the lock towards the top of the connector.
2 Insert the paddles/pads adaptor cable plug into the paddles connector on the
defibrillator, as shown in Figure 7-1 on page 7-1.
3 Slide the paddle connector lock to the lock position, to latch the plug in place.
A-4
Installation and Setup
Installation
Figure 1-2
Connector lock
Push the lock in the direction
indicated by the arrow to move the
lock to the lock position (shown).
The ECG Input connector on the defibrillator is a 6-, 8-, or 12-pin connector,
depending on the option purchased with the instrument. For each connector option,
several different patient cables can be used for various ECG sources and
applications.
Refer to Table 3-1, “Cardiac Monitoring Configurations,” on page 3-1 for a list of
available patient cables and lead sets, and their part numbers.
A-5
Installation and Setup
Installation
Connecting a Patient Cable The 3-wire or 5-wire patient cable connects to the ECG
Input connector located on the front of the defibrillator, behind the carrying handle.
The patient cable plug has 6- , 8-, or 12-pins. Before installing the patient cable,
make sure that the pin count of the patient cable plug matches the pin count of the
ECG Input connector. To install the patient cable:
1 Align the keyed cable plug with the slot in the ECG Input connector. See Figure
1-3 on page A-7.
2 Push the cable plug firmly into the ECG Input connector.
NOTE When a lead is selected for monitoring, the message LEADS OFF appears on the
display if the patient cable falls off or is incorrectly connected. Also, a dashed line
appears on the display in place of an ECG trace.
A-6
Installation and Setup
Setup
Figure 1-3
1 Setup
To configure the CodeMaster XL+ you must use the setup menus. Table 1-1, “Setup
Menu 1 Settings,” on page A-9 and Table 1-2, “Setup Menu 2 Settings,” on page A-
10 list the choices you can make on the setup menus. Perform the following steps to
configure the CodeMaster XL+.
A-7
Installation and Setup
Setup
1 Depress Sync HR while turning the Energy Select control from Off
(Standby) to Monitor On. A menu will appear with the following choices.
CALIBRATE DEFIB
SETUP MENU 1
SETUP MENU 2
RESTORE FACTORY SETTINGS
PRINT LOG
TEST DEFIB
TEST ECG
TEST CRT
TEST RECORDER
TEST CONTROLS
TEST INDICATORS
TEST BATTERY
TEST PACER
2 Select SETUP MENU 1 by pressing ECG Size until the highlight appears
on SETUP MENU 1 .
3 Press Lead . SETUP MENU 1 will appear with the current setup values
displayed.
4 Press ECG Size or ECG Size until the highlight appears on the value
you wish to change.
5 Press Lead .
6 Press ECG Size or ECG Size to scroll through the available choices for
this parameter.
7 When the choice you want is displayed, press Lead to set your choice.
8 Press ECG Size or ECG Size until the highlight appears on the next
value you wish to change.
9 Repeat steps 5 through 8 until you are finished configuring the settings in setup
menu 1.
10 Depress both sides of the ECG Size key at once to return to the setup/diag-
nostic menu.
A-8
Installation and Setup
Setup
NOTE You must depress both sides of the ECG Size key at the same time to return to the
setup/diagnostic menu.
11 To change settings in setup menu 2, select SETUP MENU 2 from the setup/
diagnostic menu and repeat the above steps.
12 Turn the Energy Select control to Off (Standby) to leave setup/diagnostic mode.
You can use the factory setting for most values by selecting
RESTORE FACTORY SETTINGS from the setup/diagnostic menu.
WARNING The setting values can have critical impact on how your
defibrillator operates.
Table 1-1, “Setup Menu 1 Settings,” on page A-9 and Table 1-2, “Setup Menu 2
Settings,” on page A-10 show configurable parameters on the instrument.
Language English, Dutch, Swedish, French, German, Printed and displayed text language
Italian, Spanish, Finnish, Danish, Norwegian
Upper Alarm 120, 140, 160, 20 to 280 in increments of 5 Upper heart rate limits LAL to 280
Limits (UAL)
Lower Alarm 40, 60, 90, 20 to 280 in increments of 5 Lower heart rate limits 20 to UAL
Limits (LAL)
A-9
Installation and Setup
Setup
Pacer Rate (pacer 70 (ppm), 40 to 180 in increments of 10 Sets the pacer rate and the initial power-on
option only) default rate.
Pacer Output 30 (mA), 30 to 200 in increments of 10 Sets the pacer current and the initial power-on
(pacer option default pacer current.
only)
Post Shock Data ON, OFF ON prints the delivered energy statistics.
OFF prints Energy Select control setting.
Power On Lead PADDLES, LEAD I, LEAD II, LEAD III Sets the ECG monitoring source that
appears when you turn on the instrument.
A-10
Installation and Setup
Specifications
1 Specifications
Defibrillator
Waveform: Damped sinusoidal (Lown).
Output Energy (Delivered): 2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, and
360 joules.
Charge Time: Less than 5 seconds to 360 joules with battery present. Less than 15
seconds to 360 joules on AC only.
Armed Indicators: Charge done tone, charge done lamp on apex paddle, and
available energy indicated on display.
Paddle Contact Indicator: 3-color LED bar graph array on STERNUM paddle
indicates quality of defibrillator paddle contact before discharge.
A-11
Installation and Setup
Specifications
Monitor
Inputs: ECG may be viewed through paddles or patient cable. Lead I, II, III, or
PADDLES selectable. Additional leads (avR, avF, AvL, V Leads) and PADS are
available. Monitor and recorder indicate selected ECG source.
Lead Fault: LEADS OFF message and dashed baseline appear on monitor if a lead
becomes disconnected.
Common Mode Rejection: Greater than 100 dB measured as per AAMI standards
for cardiac monitors (EC13).
Display Size and Type: 5 inch (12.7 cm) diagonal CRT for 4 seconds of ECG data
on screen; non-fade, fixed trace. Scrolling trace is selectable.
Heart Rate Alarms: Three configurable pairs of high and low heart rate alarm limits
from 20 to 280 BPM.
Annotates: Time, date, HR, event marker, ECG mode, defibrillator mode, selected
energy, actual delivered energy, peak current, and patient impedance.
Speed: 25 mm/sec.
A-12
Installation and Setup
Specifications
Weight: 24 lbs (10.9 kg). Includes external paddles, battery, and recorder paper.
Battery
Type: Rechargeable sealed lead-acid battery. 4 Ah, 12 V nominal.
A-13
Installation and Setup
Specifications
SpO2 (Optional)
Measurement Range: 0 to 100%.
Averaging: 8 beats.
SpO2 alarm delay: ten seconds after value drops below the low alarm setting.
A-14
Installation and Setup
Calling for Service
United States
Canada
Western Region
Tel: 1-800-661-5626
A-15
Installation and Setup
Calling for Service
Netherlands Spain
Hewlett-Packard Nederland Hewlett-Packard Espanolas
B.V. S.A.
Startbaan 16 Crta. de la Coruna
1187 XR Amstelveen km 16, 400
P.O. Box 667 Las Rozas
NL-11 8O AR Amstelveen E-Madrid
Tel: 20/547 69 11 Tel: 1/637 00 11
Norway Sweden
Hewlett-Packard Norge A/S Hewlett-Packard Sverige AB
P.O. Box 34 Skalholtsgatan 9, Kista
Osterndalen 16-18 Box 19
N-1345 Osteras S-163 93 Spanga
Tel: 2/24 60 90 Tel: 8/750 20 00
Finland Italy
Hewlett-Packard OY Hewlett-Packard Italiana
Piispankalliontie 17 S.p.A.
02200 Espoo Via G. di Vittorio, 9
Tel: 0/887 21 20063 Cernusco S/N (MI)
Tel: 2/92 36 91
A-16
6 SpO2 Monitoring (Optional)
SpO2 monitoring gives information on both cardiac and respiratory systems, and
details of oxygen transportation in the body. It is widely used because it is non-
invasive, continuous, easily applied and painless.
You can use the SpO2 monitor with sensors made by other manufacturers as well as
with HP sensors. For a list of approved sensors, see the Sensor Guide (HP part
number M1722-93970).
SaO2 is the term used to indicate the oxygen saturation of arterial blood.
SpO2 is the term used to indicate the oxygen saturation of arterial blood as
measured by pulse oximetry.
6 Application Notes
The pulse oximetry method used for measuring SpO2 uses LEDs (light emitting
diodes) to transmit red and infrared light through suitable peripheral areas of the
body, typically the foot in neonates or the finger in adults. The oxygen saturation is
gauged by measuring the "redness" of the blood in the arterial pulse.
6-1
SpO2 Monitoring (Optional)
Using SpO2 to Monitor a Patient
Figure 6-1
o All transmitted light must pass through the extremity to the detector.
o The light emitter and the photodetector must be opposite each other.
6-2
SpO2 Monitoring (Optional)
Using SpO2 to Monitor a Patient
1 Estimate the patient’s weight, and determine the best site for the sensor.
2 Use the Sensor Guide to select the correct type and size of sensor for the identi-
fied location.
5 Connect the sensor to the monitor. To connect sensors from other manufacturers
you need the HP M1900B Connector Cable.
6-3
SpO2 Monitoring (Optional)
Using SpO2 to Monitor a Patient
Figure 6-2
Push the sensor over the fingertip so the cable lies on the back of the hand, and
secure the cable to the wrist with the wrist-strap supplied. Make sure the finger is
not pinched in the end of the sensor. This ensures that the light sources in the sensor
lie over the base of the fingernail, giving the best measurement results. If the sensor
is not in the correct position, inaccurate readings result. In extreme cases, the
instrument displays dashes instead of an SpO2 reading. When correctly positioned,
the end of the finger just touches the end of the sensor.
CAUTION When non-HP SpO2 sensors are used, application must be consistent with the
manufacturer's own guidelines.
6-4
SpO2 Monitoring (Optional)
Using SpO2 to Monitor a Patient
Patient Movement
Make sure that the application site chosen does not move excessively, which may
adversely affect the performance of the sensor. You may have to replace the sensor
to ensure good adhesion, or you may have to choose another application site.
Inspect the SpO2 sensor site at least once every 2 hours to ensure adhesion, skin
integrity, and correct alignment of the light emitter and photodetector. Should
alterations of skin integrity occur, remove the sensor and reapply at another
recommended site. Avoid application of the sensor to edematous or fragile tissue.
Check circulation distal to the sensor site routinely.
Wrapping the tape too tightly, or using supplemental tape, can cause venous
pulsations that could potentially lead to inaccurate saturation measurements.
Therefore, do not wrap the adhesive too tightly and do not use additional tape to
secure the sensor. High positive intrathoracic airway pressures, valsalva maneuvers,
or other consequences of impaired venous return may also cause venous pulsations.
6-5
SpO2 Monitoring (Optional)
Using SpO2 to Monitor a Patient
Avoid placing the SpO2 sensor on any extremity with an arterial catheter, blood
pressure cuff, or intravascular venous infusion line.
Figure 6-3
SpO2 input
CAUTION Do not force the SpO2 connector into the ECG input socket. Doing so may damage
the pins on the cable connector.
6-6
SpO2 Monitoring (Optional)
Using SpO2 to Monitor a Patient
Start Monitoring
Turn the defibrillator on, if necessary, by turning the Energy Select control to
Monitor On. Press the SpO2 button to display the SpO2 reading in the upper
On/Off
right corner of the display.
SpO2 Readings
Figure 6-4
HR
LEAD I I
SpO2 reading SPO2
PR 72
Pulse rate
The pulse amplitude indicator shows the quality of the SpO 2 signal. Since it is
derived from the patient’s plethysmograph signal, it varies with the pulse of the
patient. If the patient has a very low signal the pulse amplitude indicator does not
vary through its full range. If the signal is noisy, the pulse amplitude indicator does
not vary rhythmically with the pulse.
6-7
SpO2 Monitoring (Optional)
SpO2 Alarms
The pulse rate is derived from the pulse oximeter. It should correlate closely with
the patient’s heart rate.
6 SpO2 Alarms
the SpO2 button repeatedly to cycle through the alarm options and the no-
Alarm
alarm option. Stop when you see the alarm you would like to choose and after three
seconds that alarm will take effect. A symbol replaces the limits to show that
the alarm is active. To review the limit set, press the SpO2 button.
Alarm
If the SpO2 level falls below the low alarm limit, an alarm sounds and the violated
limit is highlighted.
NOTE SpO2 alarms are automatically turned off when you press Charge .
Press the SpO2 button. The symbol to the right of the SpO 2 display
Alarm
6-8
SpO2 Monitoring (Optional)
SpO2 Alarms
Recorder Output
After an alarm event, the recorder prints a strip. The bottom of the strip shows the
alarm violation, and the top of the strip shows the SpO2 reading.
You may print an event summary record, which contains SpO2 information, as
described in “Printing the Event Summary Record” (page 3-8).
6-9
SpO2 Monitoring (Optional)
SpO2 Alarms
6-10