f701c
f701c
OPERATION MANUAL
01 Apr. 2015
Rev. 1.16
INTRODUCTION
INTRODUCTION
Thank you very much for purchasing our Weighing Controller F701-C.
For good performance, and proper and safe use of the F701-C, be sure to read this instruction
manual and properly understand the contents of it before use. Also, carefully keep this instruction
manual so that it can be referred to at any time.
SAFETY PRECAUTIONS
For safety reasons, please read the following safety precautions thoroughly.
In order to have an F701-C Weighing Controller used safely, notes we would like you to surely
follow divide into " WARNING " and " CAUTION ", and are indicated by the following
documents. Notes indicated here are the serious contents related to safely. Please use F701-C after
understanding the contents well.
WARNING
This sign forewarns the presence of hazards that could result in serious injury or
fatality when incorrectly handled.
CAUTION
This sign forewarns the presence of hazards that could result in personnel injury or
property damage when incorrectly handled.
I
SAFETY PRECAUTIONS
Warning on design
● For the entire system to function safely when the F701-C becomes faulty or malfunctions,
provide a safety circuit outside the F701-C.
● Before using the F701-C as described below, make sure to consult with our sales personnel.
- Use in environments not described in the operation manual.
- Use greatly impacting human lives and assets, such as medical devices, transport devices
entertainment devices, and safety devices.
Warning on installation
● Do not disassemble, repair, or modify the F701-C. Doing so may cause a fire or an electric
shock.
● Do not install in the following environments.
- Places containing corrosive gas or flammable gas.
- Where the product may be splashed with water, oil or chemicals.
Warning on wiring
● Do not connect a commercial power source directly to the signal input/output terminals.
● Be sure to ground the protective ground terminal.
● The attached AC cable is designed for domestic use in Japan, and its rating is 125V AC, 7A.
For use at voltages exceeding the rating and for overseas use, have a separate AC cable
prepared.
● Before performing the following, make sure that no power is applied.
- Attachment/detachment of connectors of options, etc.
- Wiring/connection of cables to the signal input/output terminals.
- Connection of the earth cable.
● For connection to the signal input/output terminals, check the signal names and pin
assignment numbers, and then carry out wiring properly.
● Be sure to use crimp contacts for connection to the terminal blocks, and do not to connect
bare wires as they are.
● After wiring, be sure to mount the attached terminal block cover. Otherwise, it may cause an
electric shock.
● Do not connect anything to unused terminal(s).
● Before applying power, carefully check the wiring, etc.
II
SAFETY PRECAUTIONS
Caution on installation
● Use the F701-C as it is incorporated in a control panel, etc.
● Do not install in the following environments:
- Locations where temperature or humidity exceeds specifications;
- Locations subjected to drastic temperature fluctuations or icing and condensing;
- Outdoors or locations above 2,000m;
- Locations exposed to direct sunlight;
- Locations subject to dust accumulation;
- Locations with poor ventilation;
- Locations with a lot of salt and metal powder;
- Locations where the main unit is subject to direct vibration and shock.
● Take adequate shielding measures when using at the following locations.
- Near a power line.
- Where a strong electric field or magnetic field is formed.
- Where static electricity, relay noise or the like is generated.
● Install the F701-C as far away from devices generating high frequency, high voltage, large
current, surge, etc., as possible. Also, carry out wiring separately from their power lines. Do
not carry out parallel wiring and common wiring.
● Do not use it, broken down.
III
This sign forewarns the presence of hazards
CAUTION that could result in personnel injury or property
damage when incorrectly handled.
Caution on wiring
● Tighten the screws for the signal input/output terminals at the specified torque.
If they are loose, shorts, fire or malfunctions may occur.
Tightening torque: 0.5N • m
● For sensors, external inputs/outputs and options, use shielded cables.
● The temporary overvoltage applied to the power should not exceed 1500V.
● Remove the lithium batteries used in the F701-C and make sure to dispose them according
to classification of waste collection.
IV
CONTENTS
CONTENTS
2.CONNECTION ............................................................................8
2-1.Load cell Connection ................................................................................... 8
2-1-1.6-wire Connection ........................................................................................ 10
2-1-2.4-wire Connection ........................................................................................ 10
2-1-3.Connecting Load cells in Parallel ............................................................... 11
2-1-4.Sensor Cable ................................................................................................. 11
2-2.Connection of the Power Input Terminals ............................................... 12
2-3.Connection of the Guard Ground ............................................................. 12
2-4.SI/F Connection .......................................................................................... 13
2-5.Connection of the Control Connector ...................................................... 14
2-5-1.Control Connector-Pin Assignment ........................................................... 14
2-5-2.How to Assemble the Connector ................................................................ 15
2-5-3.Equivalent Circuit (Input) ............................................................................. 16
2-5-4.Equivalent Circuit (Output) .......................................................................... 16
4.CALIBRATION .........................................................................41
4-1.Span Calibration ......................................................................................... 41
4-2.Span Calibration Procedure ...................................................................... 42
4-3.Secondary Calibration Procedure (Equivalent Calibration) ................... 43
4-4.Preparation for Calibration ........................................................................ 45
4-4-1.LOCK Release ............................................................................................... 45
4-4-2.Unit ................................................................................................................. 46
4-4-3.Decimal Place ............................................................................................... 46
4-4-4.Capacity ......................................................................................................... 47
V
CONTENTS
VI
CONTENTS
VII
CONTENTS
10.INTERFACE .........................................................................124
10-1.2-wire Serial Interface (SI/F) .................................................................. 124
10-1-1.Connection ................................................................................................ 124
10-1-2.Auto Print Command ................................................................................ 124
10-2.RS-232C Interface ................................................................................... 125
10-2-1.Communication Specifications ............................................................... 125
10-2-2.Cable .......................................................................................................... 125
10-2-3.Setting Values for RS-232C ..................................................................... 126
10-2-4.Communication Mode .............................................................................. 126
10-2-5.Transmission Format ............................................................................... 128
10-2-6.Communication Format ........................................................................... 129
10-3.BCD Parallel Data Output Interface ...................................................... 137
10-3-1.Control Connector-Pin Assignment ....................................................... 137
10-3-2.Equivalent Circuit (Input) ......................................................................... 137
10-3-3.Equivalent Circuit (Output) ...................................................................... 138
10-3-4.BCD Data Output ...................................................................................... 138
10-3-5.Polarity Output (MINUS) ........................................................................... 138
10-3-6.Over Status Output (OVER) ..................................................................... 139
10-3-7.Print Command Output (P.C) ................................................................... 139
10-3-8.Data Strobe (STROBE) ............................................................................. 139
10-3-9.BCD Data Update Rate Selection ............................................................ 140
10-3-10.Data Hold Input ....................................................................................... 140
10-3-11.Logic Switching Input ............................................................................ 140
10-3-12.Output Selection Input ........................................................................... 140
10-4.D/A Converter ......................................................................................... 141
10-4-1.Name of Each Part .................................................................................... 141
10-4-2.Specifications ........................................................................................... 143
10-4-3.Method of Adjusting the D/A Zero and Gain .......................................... 143
10-4-4.About D/A Resolution .............................................................................. 145
10-5.RS-485 Communication Interface ......................................................... 146
10-5-1.Communication Specifications ............................................................... 146
10-5-2.RS-485 Connection ................................................................................... 146
10-5-3.Setting Values for RS-485 ........................................................................ 147
10-5-4.Communication Method ........................................................................... 148
10-5-5.Communication Format ........................................................................... 149
VIII
CONTENTS
15.DIMENSIONS .......................................................................168
17.SPECIFICATIONS ................................................................170
17-1.Analog Section ....................................................................................... 170
17-2.Display Section ....................................................................................... 171
17-3.Setting Section ....................................................................................... 171
17-4.External Input/Output ............................................................................. 172
17-5.Interface .................................................................................................. 172
17-6.General Performance ............................................................................. 174
17-7.Attachment .............................................................................................. 175
IX
CONTENTS
M E M O
X
1.APPEARANCE DESCRIPTION
1. APPEARANCE DESCRIPTION
1-1. Front Panel
② Subdisplay
④ Status display
⑥ Function keys
⑤ Setting keys
① Main display
1
1.APPEARANCE DESCRIPTION
② Subdisplay
Weighing data, such as Accumulation Value, and various setting values are displayed by setting.
(6) None
2
1.APPEARANCE DESCRIPTION
Latest Accumulation Data Displays the latest data accumulated by the auto accumulation
command or external input/output signal, or through RS-232C
communication, etc.
Final Displays the “Final” value set in the setting mode 0-9.
Over Displays the “Over” value set in the setting mode 0-7.
Under Displays the “Under” value set in the setting mode 0-8.
Near Zero Displays the “Near zero” value set in the setting mode 0-3.
Upper Limit Displays the “upper limit” value set in the setting mode 0-1.
Lower Limit Displays the “lower limit” value set in the setting mode 0-2.
③ Unit indication
The unit can select from six types: t, kg, g, N, lb and None.
④ Status display
3
1.APPEARANCE DESCRIPTION
Flashes when the voltage of the lithium battery for memory backup has
dropped.Replace with a new battery.
For the method of replacement, refer to "13.REPLACEMENT OF THE BACKUP
BATTERY" on P.166.
* The expression “ ** ” in this instruction manual shows that 「 」on the status display lights,
goes out, or flashes.
⑤ Setting keys
0
0 When the key is pressed, the Tare weight is displayed, and
“ TARE ” flashes.
0
(When the Tare weight display with the key is valid under
Restriction on the Tare Subtraction Function in setting mode 4.)
0
To go back to the weight display, press the key again.
1
UPPER ~ 9
FINAL
Numerical keys to make settings.
⑥ Function keys
4
1.APPEARANCE DESCRIPTION
ZERO
When ZERO → CNG/ENT is pressed, Gross is zeroed
5
1.APPEARANCE DESCRIPTION
② Protective ground
③ Frame ground
② Protective ground
③ Frame ground
Please ground the frame ground terminal to prevent failures due to static electricity.
(The frame and the frame ground terminal are conducted.)
It may be better to remove depending on the environment of the installation location.
④ Control connector
This is a connector to input external signals and output control signals. The Input/Output circuit and
internal circuit are photocoupler-insulated electrically.
The applicable connector is the following (accessory) manufactured by Fujitsu Component or an
equivalent:
Connector : FCN-361J024-AU
Cover : FCN-360C024-B
6
1.APPEARANCE DESCRIPTION
⑤ RS-232C connector
RS-232C connector for receiving and transmitting weight data and status information.
The applicable connector is JAE DE-09SN or its equivalent.
2-wire serial interface is to connect unipulse peripheral equipment such as printer, remote display
or data converter.
⑦ LOCK switch
LOCK switch for avoiding changes of setting value, it prohibits to change setting value while the
switch is ON.
⑨ Option space
7
2.CONNECTION
2. CONNECTION
2-1. Load cell Connection
The voltage application of F701-C is 10V, and the maximum current is 120mA, to which up to four
350Ω load cells can be connected in parallel.
1 +SIG +SIG
2 -SIG -SIG
3 +EXC +EXC
4 +S (Connect 3 to 4)
5 -EXC -EXC
6 -S (Connect 5 to 6)
7 SHIELD SHIELD
8
2.CONNECTION
Method of connection
2) Twist the end to such an extent that it will not become loose.
Hole
Turn clockwise.
Request
UP
When mounting the terminal block to the Insertion side
F701-C body, check its vertical orientation.
(See the illustration on the right-hand side.) DOWN
9
2.CONNECTION
The load cell input of the F701-C is a 6-wire (remote sense) connection. 6-wire shielded load cell
cable should be used and kept separate from AC or other noise generating wire.
Load cell
3 +EXC
4 +S
+IN
6 -S
5 -EXC
-OUT +OUT
1 +SIG
2 -SIG
7 SHIELD
-IN
* Remote sense lines are used to detect and correct variations in excitation voltage over long cable
runs.
Load cell
3 +EXC
4 +S
+IN
6 -S
5 -EXC
-OUT +OUT
1 +SIG
2 -SIG
7 SHIELD
-IN
CAUTION
- The load cell excitation voltage of the F701-C is 10V. Heating or breakage may occur
unless the load cells maximum excitation voltage is 10V or more.
- When using the F701-C with the four-wire load cell connected, be sure to connect +EXC
and +S, and -EXC and -S. Even if +S and -S are not connected, normal operation is
performed apparently, but heating or breakage may occur because excessive voltage is
applied to the load cell.
10
2.CONNECTION
In some industrial weighing apparatus, two or more load cells may be connected in parallel to form
a hopper scale or track scale. The manner of connection is shown below.
Parallel connection can simply be made by using the optionally available B410 (summing box for
4-point multi load cell).
+EXC
R
-SIG
-EXC
R
+SIG
SHIELD
+S
-S
Request
When connecting several load cells in parallel, load cell capacity should be higher than
expected load to compensate for mechanical shock or eccentric loading.
Cable colors of sensors may differ from one manufacturer to another (it may even differ from one
model to another for some products). Refer to the sensor manual (or data sheet) and check signal
names and colors in order to connect the cables correctly.
11
2.CONNECTION
3) Remove the two screws(M3) of the terminal block, align the crimp contacts with the
screw holes, and then tighten the screws.
5) Remove the screws(M4) of the protective ground, align the crimp contacts with the
screw holes, and then tighten the screws.
Frame
ground
Within 6mm
AC power cord
block cover
CAUTION
- Connect with no power applied because it may cause an electric shock.
- The attached AC cable is designed for domestic use in Japan, and its rating is 125V AC,
7A. For use at voltages exceeding the rating and for overseas use, have a separate AC
cable prepared.
12
2.CONNECTION
5 to 6mm
2) Insert the supplied screwdriver into the upper hole and lift upward.
5) Make sure cable is clamped securely and does not come out with a slight tug.
Request
13
2.CONNECTION
A1 * COM B1 * COM
A2 In G/N B2 In Input Selection 1 *1
A3 In D/Z ON B3 In Input Selection 2 *1
A4 In Tare Subtraction ON B4 In Input Selection 3 *1
A5 In Tare Subtraction OFF B5 In Input Selection 4 *1
A6 Out Near Zero B6 Out Lower Limit
A7 Out SP1 B7 Out Upper Limit
A8 Out SP2 B8 Out Stable
A9 Out SP3 B9 Out Output Selection 1 *2
A10 Out Under B10 Out Output Selection 2 *2
A11 Out Over B11 Out Output Selection 3 *2
A12 * COM B12 * COM
* *1 : Selectable by setting.
(For details, please refer to "9-4-11.Input Selection" on P.117)
* *2 : Selectable by setting. (For details, please refer to "6-11.Weight Error / Sequence Error" on
P.101, "6-8.Complete Signal Output Mode / Complete Output Time / Judging Time /
Comparison Inhibit Time / Output Selection 2" on P.93, "9-5-7.Accumulation Error" on
P.119)
14
2.CONNECTION
Nut M2 (four)
Pan-head machine screw Pan-head machine screw
M2×8 (short) (two) M2×10 (long) (two)
Washer (two)
Screw (two)
(1) Set the connector and screws (two) into the grooves of the case (one side).
(2) Cover with the other case, and fit the cases.
15
2.CONNECTION
A signal is inputted to the signal input circuit by short-circuiting or opening the input terminal and
the COM terminal. Short-circuiting is effected by means of a contact (such as a relay or a switch) or
a noncontact (such as a transistor or an open-collector TTL).
Vcc
F701-C Open OFF
+12V Inside
Short ON
Approx.
Push TTL open collector
IC=6mA Toggle switch Relay Transistor
switch (ON when IN is HI)
IN
IN
COM
Vcc +12V
Vext Spark killer
F701-C
Inside
DC Power
Load
Relay
COM Varistor
Spark killer
Vceo=30V(max) AC power
Ic =120mA (max)
Load
- Transistor status
Output data Tr - Use external power source (up to DC30V) for driving relay
(Vext).
0 OFF
- Do not short-circuit the load, such as a coil of relay, that will
1 ON
break the output transistor.
- Connect a surge absorber or a speark killer to the relay
circuit as shown in the draft so that to reduce noise trouble
and extend the life of relay.
Noise trouble can be reduced, and the relay’s life can be
extended.
16
3.METHODS OF SETTING
3. METHODS OF SETTING
3-1. Setting Procedure
Change settings in the order of “setting mode selection” → “setting item selection” → “setting
value entry”.
3
F → CNG/ENT → NEAR Z. → CNG/ENT
17
3.METHODS OF SETTING
3
3) Select the setting mode number. ( NEAR Z. )
18
3.METHODS OF SETTING
1
UPPER → CNG/ENT → → CNG/ENT
0 5 0 0 0
3) Input the setting value. ( SP2 )
4) After the correct setting value is input, press the CNG/ENT key to enter the setting value.
19
3.METHODS OF SETTING
6
CPS → CNG/ENT → → CNG/ENT
0 0 3 0
3) Input the setting value. ( NEAR Z. → F )
4) After the correct setting value is input, press the CNG/ENT key to enter the setting value.
20
3.METHODS OF SETTING
(Example 3) For setting the 1/4 scale division display to OFF (Setting from choices)
7
OVER → CNG/ENT → ZERO ZERO ZERO → CNG/ENT
3) Move with the ZERO key until the digit you want to set blinks. ( ZERO ZERO ZERO )
Not defined
1/4 scale division display
Decimal place
Display frequency
Subdisplay selection
Setting item number
Setting mode number
21
3.METHODS OF SETTING
5) After the correct choice is input, press the CNG/ENT key to enter the choice.
22
3.METHODS OF SETTING
3 Near Zero Complete Output Time Weighing Function 3 Min. Scale Division
NEAR Z. / P.24 / P.26 / P.29 / P.31
5 Set Point 2 Auto Zero Times Function Key Invalid Gross Over
SP2 / P.24 / P.26 / P.30 / P.31
ZERO
2 D/A Zero Output Weight Output Selection Max. Value Equivalent Calibration
LOWER / P.33 / P.36 / P.38 / P.54
CC-Link I/F
3 D/A Full Scale Number of occupied station & Min. Value
NEAR Z. / P.33 Transmission speed / P.38
(Refer to separate-volume spec.)
* When each setting mode is selected, the 1 ~ 9 keys function as setting item selection keys.
23
3.METHODS OF SETTING
In setting mode 0, setting values for final discharging control are to be set.
- Upper Limit
1
UPPER (0 ~ 99999)
- Lower Limit
2
LOWER (0 ~ 99999)
- Near Zero
3
NEAR Z. (0 ~ 99999)
- Set Point 1
4
SP1 (0 ~ 99999)
- Set Point 2
5
SP2 (0 ~ 99999)
- Compensation
6
CPS (-9999 ~ 9999)
24
3.METHODS OF SETTING
- Over
7
OVER (0 ~ 999)
- Under
8
UNDER (0 ~ 999)
- Final
9
FINAL (0 ~ 99999)
25
3.METHODS OF SETTING
In setting mode 1, output signals for final discharging control and parameters in sequence mode,
etc., are to be set.
1
UPPER . sec. (0.00 ~ 9.99)
- Judging Time
2
LOWER . sec. (0.00 ~ 9.99)
3
NEAR Z. . sec. (0.00 ~ 9.99)
4
SP1 . sec. (0.00 ~ 9.99)
5
SP2 (0 ~ 99)
6
CPS (0 ~ 99)
26
3.METHODS OF SETTING
7
OVER (0 ~ 99999)
- Analog Filter
8
UNDER (0 ~ 3)
- Tare Weight
9
FINAL (0 ~ 99999)
27
3.METHODS OF SETTING
In setting mode 2, the display and internal functions of the F701-C are to be tuned.
- Weighing Function 1
1
UPPER
- Weighing Function 2
2
LOWER
28
3.METHODS OF SETTING
- Weighing Function 3
3
NEAR Z.
2 : AFFC. ON
(Communication OP invalid)
1: 〃 ON
0: 〃 OFF
(For details, please refer to
"6-3" on P.83.)
- Sequence Mode
4
SP1
(Sequence control)
1 : Sequence mode 1 : At start-time, weight
0 : Simple comparison mode value check ON
(For details, please refer to 0 : 〃 OFF
"6-2-3" on P.82.) (For details, please refer to
"6-9" on P.96.)
1 : Adjust feeding ON
(Sequence control)
0: 〃 OFF
1 : At start-time,
(For details, please refer to Near Zero check ON
"6-9" on P.96.) 0 : 〃 OFF
(For details, please refer to
"6-9" on P.96.)
29
3.METHODS OF SETTING
5
SP2
- Digital Filter
6
CPS
Digital Filter
(0 ~ 256)
(For details, please refer to "5-3" on P.56.)
- Motion Detection
7
OVER . -
8
UNDER .
Period (sec.)
(0.0 ~ 9.9)
(For details, please refer to "5-7" on P.60.)
9
FINAL
Range
(0 ~ 9999)
(For details, please refer to "5-7" on P.60.)
30
3.METHODS OF SETTING
1
UPPER (0 ~ 99999)
- Capacity
2
LOWER (0 ~ 99999)
3
NEAR Z. (1 ~ 100)
- Net Over
4
SP1 (0 ~ 99999)
- Gross Over
5
SP2 (0 ~ 99999)
- DZ Regulation Value
6
CPS (0 ~ 9999)
31
3.METHODS OF SETTING
- Function Selection
7
OVER
Request
When using the F701-C for any balance weight to undergo type approval by the
Measurement Law, set the 1/4 scale division display to OFF. If it is set to OFF,
the “zero point” lights only at a true zero point (0±1/4 scale division).
8
UNDER (0 ~ 16)
9
(9.700 ~ 9.999)
FINAL .
(For details, please refer to "4-4-7" on P.48.)
32
3.METHODS OF SETTING
1
UPPER
2
LOWER (0 ~ 99999)
3
NEAR Z. (0 ~ 99999)
- RS-485 I/F
4
SP1
Parity bit
Character length
2 : Even
1 : 8bit
1 : Odd
0 : 7bit
0 : None
33
3.METHODS OF SETTING
- ID Number
5
SP2 (0 ~ 9999)
- RS-232C I/F
6
CPS
Stop bit
1 : 2bit
0 : 1bit
Parity bit
2 : Even
1 : Odd
0 : None
34
3.METHODS OF SETTING
7
OVER
8
UNDER
LOCK1 LOCK2
1 : ON 1 : ON
0 : OFF 0 : OFF
9
FINAL
35
3.METHODS OF SETTING
- Input Selection
1
UPPER
- Output Selection
2
LOWER
36
3.METHODS OF SETTING
37
3.METHODS OF SETTING
- Average Weight
1
UPPER (0 ~ 99999)
- Max. Value
2
LOWER (0 ~ 99999)
- Min. Value
3
NEAR Z. (0 ~ 99999)
4
SP1 (0 ~ 99999)
5
SP2 (0 ~ 99999)
6
CPS (0 ~ 10000)
7
OVER (0 ~ 99999)
38
3.METHODS OF SETTING
8
UNDER (0 ~ 99999)
- Optional Board
9
FINAL
● Calculation Formula
Σ(χ
2
i - χ) Use all the data of the finite population and
σn = i=1 find the standard deviation of the population.
n
Σ(χ
2
i - χ) Use the sample data among the population and
σn-1 = i=1 the standard deviation of the population.
n-1
39
3.METHODS OF SETTING
【Example】
Accumulation clear
2 40.090 20.040 20.045 20.050 20.040 0.010 0.005 0.007
3 60.160 20.070 20.053 20.070 20.040 0.030 0.012 0.015
4 80.240 20.080 20.060 20.080 20.040 0.040 0.016 0.018
5 100.260 20.020 20.052 20.080 20.020 0.060 0.021 0.024
6 120.260 20.000 20.043 20.080 20.000 0.080 0.027 0.030
7 140.270 20.010 20.039 20.080 20.000 0.080 0.028 0.030
8 160.250 19.980 20.031 20.080 19.980 0.100 0.033 0.035
9 180.360 20.110 20.040 20.110 19.980 0.130 0.039 0.042
10 200.370 20.010 20.037 20.110 19.980 0.130 0.038 0.041
40
4.CALIBRATION
4. CALIBRATION
4-1. Span Calibration
Calibration is performed for matching the F701-C to a load cell. For example, it is work to adjust so
that the F701-C accurately displays 100.00kg when an actual load (or weight) of 100kg is applied to
the load cell (balance section) of the weighing apparatus to which the F701-C is connected. This
operation is called Span Calibration.
After calibration.....
100kg
41
4.CALIBRATION
Release Calibration LOCK Release Setting Value LOCK which inhibits the calibration.
Min. Scale Division Register the minimum unit (scale division) of the scale.
Balance Weight Value Register the value of load (balance weight) for load cell.
Place the load (balance weight) on the load cell and register the
Span Calibration
span value.
42
4.CALIBRATION
Calibration procedure performed by entering the rated output value (mV/V) and rated capacity
value of load cell without using real load.
Secondary calibration function is provided for provisional calibration when the F701-C develops
trouble or the calibration value is mistakenly changed.
The secondary calibration is only a provisional method. Calibration with actual load must be done
as soon as possible.
Release Calibration LOCK Release Setting Value LOCK which inhibits the calibration.
Min. Scale Division Register the minimum unit (scale division) of the scale.
Equivalent Calibration *1 Register the rated output value (mV/V) of load cell.
43
4.CALIBRATION
Request
44
4.CALIBRATION
F701-C features a LOCK function for disabling changes in calibration and setting values.
The software LOCK is performed with the operation on the display; the hardware LOCK is located
on rear panel. Release both of locks when the calibration is performed.
Operation
LOCK
OFF
4
F → CNG/ENT → SP1 → CNG/ENT
8
UNDER → CNG/ENT → ZERO → CNG/ENT
LOCK2
1 : ON
0 : OFF
LOCK is released through above procedure. After the calibration is finished, LOCK ON to protect
the calibration value.
45
4.CALIBRATION
4-4-2. Unit
Register the unit of the scale. The unit can select from t, kg, g, N, lb or None.
Operation
1) Select setting mode 3.
3
F → CNG/ENT → NEAR Z. → CNG/ENT
7
OVER → CNG/ENT →
Unit
5:N 4 : lb
3 : kg 2:g
1:t 0 : None
Set the decimal place for weight-related displays, setting items, etc. Decimal place should be
selected from 0, 0.0, 0.00 or 0.000. This selection results in a common decimal place.
Operation
1) Select setting mode 3.
3
F → CNG/ENT → NEAR Z. → CNG/ENT
7
OVER → CNG/ENT → ZERO ZERO → CNG/ENT
Decimal place
3:
2:
1:
0:
In the F701-C, the decimal place is all fixed except for weight-related
descriptions.
* It cannot be changed.
46
4.CALIBRATION
4-4-4. Capacity
Register the maximum capacity of the scale. If the registered value exceeds by 9 scale divisions,
Operation
1) Select setting mode 3.
3
F → CNG/ENT → NEAR Z. → CNG/ENT
2
LOWER → CNG/ENT → → CNG/ENT
Capacity
(0 ~ 99999)
Register the minimum unit (scale division) of the scale. (Input range / 1 ~ 100)
Operation
3
F → CNG/ENT → NEAR Z. → CNG/ENT
3
NEAR Z. → CNG/ENT → → CNG/ENT
47
4.CALIBRATION
Register the value of load (balance weight) before the Span Calibration.
(Input range / 0 ~ 99999)
Operation
3
F → CNG/ENT → NEAR Z. → CNG/ENT
1
UPPER → CNG/ENT → → CNG/ENT
Operation
3
F → CNG/ENT → NEAR Z. → CNG/ENT
8
UNDER → CNG/ENT → → CNG/ENT
Area number
(0 ~ 16)
48
4.CALIBRATION
9
FINAL → CNG/ENT → → CNG/ENT
Acceleration
(9.700 ~ 9.999)
49
4.CALIBRATION
It divides the minimum scale division into four (4) parts. The “ ZERO ” (center zero) lamp turns on
when the weight is between +1/4 division and -1/4 division.
1/4 scale division selects ON/OFF.
Operation
3
F → CNG/ENT → NEAR Z. → CNG/ENT
7
OVER → CNG/ENT → ZERO ZERO ZERO → CNG/ENT
n n+1
1/4 scale
“ ZERO ” ON “ ZERO ” ON
When the 1/4 scale division setting is OFF,“ ZERO “ lamp only works at the zero
point.
0
-1 1/4 scale 1
“ ZERO ” ON
50
4.CALIBRATION
- Verify there are no excess loads applied to load cell (or scale).
- Check that “ STAB ” is ON. (Correct calibration can not be completed if signal is unstable.)
Operation
9
F → CNG/ENT → FINAL → CNG/ENT
ZERO → CNG/ENT
, kg In-zero-calibration display.
kg
Weight value
Setting mode number
If a calibration error is displayed, redo Zero Calibration corresponding to the following error
messages. (Please refer to "11.OVER SCALE & ERROR" on P.157.)
-“ ”
Initial dead load is above zero adjustment range.
Remove any excess load from load cell or scale. If is still displayed, connect a
resistor between +EXC and -SIG load cell connections. This should shift the zero point.
51
4.CALIBRATION
Resistance Strain
Calculated value Approx.value μ-STRAIN mV/V
875 kΩ 866 kΩ 200 0.1
437 kΩ 442 kΩ 400 0.2
291 kΩ 294 kΩ 600 0.3
219 kΩ 221 kΩ 800 0.4
175 kΩ 174 kΩ 1000 0.5
146 kΩ 147 kΩ 1200 0.6
125 kΩ 124 kΩ 1400 0.7
109 kΩ 110 kΩ 1600 0.8
97 kΩ 97.6 kΩ 1800 0.9
87.3 kΩ 86.6 kΩ 2000 1.0
79.4 kΩ 78.7 kΩ 2200 1.1
72.7 kΩ 73.2 kΩ 2400 1.2
67.1 kΩ 66.5 kΩ 2600 1.3
62.3 kΩ 61.9 kΩ 2800 1.4
58.2 kΩ 57.6 kΩ 3000 1.5
54.5 kΩ 54.9 kΩ 3200 1.6
51.3 kΩ 51.1 kΩ 3400 1.7
48.4 kΩ 48.7 kΩ 3600 1.8
45.9 kΩ 46.4 kΩ 3800 1.9
43.6 kΩ 43.2 kΩ 4000 2.0
41.5 kΩ 41.2 kΩ 4200 2.1
39.6 kΩ 39.2 kΩ 4400 2.2
37.9 kΩ 38.3 kΩ 4600 2.3
36.3 kΩ 36.5 kΩ 4800 2.4
34.8 kΩ 34.8 kΩ 5000 2.5
-“ ”
Inditial dead load is negative.
Check that load cell is mounted in the correct direction; check that load is being applied to
the load cell in the correct direction; check that the +SIG and -SIG lines are propely
connected.
If still displayed, connect a resistor between -EXC and -SIG load cell
connections. This should shift the zero point. Do Zero Calibration again.
52
4.CALIBRATION
Span Calibration means putting a load (test weight) on the load cell (or scale) and calibrating so the
F701-C indicates correct weight.
- Put the Balance Weight Value which is set the Balance Weight Value, on the load cell (or scale).
(Calibrating with more than 50% of the Capacity is favorable in respect of linearity.)
- Verify there are no excess loads applied to load cell (or scale).
- Check that “ STAB ” is on. (Correct calibration can not be completed if signal is unstable.)
Operation
9
F → CNG/ENT → FINAL → CNG/ENT
1
UPPER → CNG/ENT → → CNG/ENT
, kg In-span-calibration display.
3) When the weight value display becomes equal to the Balance Weight Value, Span Calibration is
completed.
kg
Weight value
Setting mode number
* If any error message appears, refer to "11.OVER SCALE & ERROR" on P.157.
53
4.CALIBRATION
By key-inputting the weight value corresponding to the output value (mV/V) of the load cell,
calibration is performed so that the input of the entered output value results in the entered weight
value display.
Operation
3
F → CNG/ENT → NEAR Z. → CNG/ENT
2) Input the Balance Weight Value corresponding to the output value to be entered in 4).
(1-Balance Weight Value)
1
UPPER → CNG/ENT → → CNG/ENT
9
F → CNG/ENT → FINAL → CNG/ENT
2
LOWER → CNG/ENT → → CNG/ENT
* If any error message appears, refer to "11.OVER SCALE & ERROR" on P.157.
* If “ ” (Zero Calibration again) appears, redo Zero Calibration, and then perform
Equivalent Calibration again.
54
5.FUNCTION SETTINGS
5. FUNCTION SETTINGS
5-1. Display Frequency
Select the F701-C display frequency. The numbers of updates on the display per second is only
selected here.
Internal A/D conversion speed and CPU processing speed are not changed. The available display
frequencies are: 25, 13, 6 or 3 times/sec.. 25 times/sec. is recommended for normal operation.
If the display flickers, select a lower frequency.
Operation
1) Select setting mode 3.
3
F → CNG/ENT → NEAR Z. → CNG/ENT
7
OVER → CNG/ENT → ZERO → CNG/ENT
Display frequency
3 : 25 times/sec.
2 : 13 times/sec.
1 : 6 times/sec.
0 : 3 times/sec.
55
5.FUNCTION SETTINGS
Operation
1) Select setting mode 3.
3
F → CNG/ENT → NEAR Z. → CNG/ENT
7
OVER → CNG/ENT → → CNG/ENT
Subdisplay selection
5 : Near Zero / Upper Limit / Lower Limit
4 : Final / Over / Under
3 : Accumulation Count / Latest Accumulation Data /
Final
2 : Latest Accumulation Data / Accumulation Value
1 : Accumulation Count / Accumulation Value
0 : None
Operation
1) Select setting mode 2.
2
F → CNG/ENT → LOWER → CNG/ENT
6
CPS → CNG/ENT → → CNG/ENT
56
5.FUNCTION SETTINGS
Operation
1) Select setting mode 1.
1
F → CNG/ENT → UPPER → CNG/ENT
8
UNDER → CNG/ENT → → CNG/ENT
Analog Filter
3 : 8Hz
2 : 6Hz
1 : 4Hz
0 : 2Hz
Operation
1) Select setting mode 4.
4
F → CNG/ENT → SP1 → CNG/ENT
7
OVER → CNG/ENT → ZERO → CNG/ENT
57
5.FUNCTION SETTINGS
When the range of weight changes is within the set range and the status lasts for set period of time,
the Stab. turns ON considering to be the weight stable.
There are stable mode and checker mode in Motion Detection.
Stable mode
D1 ~ D5 shown below are compared with a set range at each A/D conversion.
* D1 represents the difference between current weight value and the value 1 second previous. If
one of them exceeds the range, the Stab. turns OFF at once.
〈e.g.1〉 D5
D4
D3
D1 D2
0.3
0.6
0.8
0.95
Weight value 1.00 sec.
〈e.g.2〉
D1 D2 D3
D4 D5
Weight value
Checker mode
D1 ~ D3 shown below are compared with a set range at each A/D conversion.
If one of them exceeds the range, the Stab. turns OFF at once.
* D1 represents the difference between current weight value and the value 0.09 second previous.
D1 D2 D3
0.03
0.06
0.09 sec.
58
5.FUNCTION SETTINGS
When the Stab. is ON, it enables to insert the Digital Filter to restrain the instability of weight
value. (Please refer to "5-5.Filter in Stable Condition" on P.57.)
Set under Analog Filter in setting mode 1, Set ON/OFF under External Function
and Digital Filter in setting mode 2. Selection in setting mode 4.
MD mode Select the stable condition from stable mode or checker mode.
Operation
4
F → CNG/ENT → SP1 → CNG/ENT
7
OVER → CNG/ENT → ZERO ZERO → CNG/ENT
MD range Comparing the value (setting value×minimum scale division) with the range of
weight change.
(Input range / 0 ~ 99)
Operation
1) Select setting mode 2.
2
F → CNG/ENT → LOWER → CNG/ENT
7
OVER → CNG/ENT → → CNG/ENT
59
5.FUNCTION SETTINGS
- Zero Tracking adjusts the zero point every set period when the shifting of the
zero point is within the set point.
- The Tracking Period must be set between 0.0 and 9.9 seconds. The Tracking
Range (digit) must be set from 0 ~ 9999 using a 1/4 scale division as a unit.
(0002=0.5 divisions, 0012=3 divisions)
Zero Tracking does not work when the period is set 0.0 or the range is set to
00.
From the point when it returned within the range,
Boundary of zero track counting will be resumed.
0
Tracking range
-
Weighing value
Tracking period Tracking period
Operation
a)
1) Select setting mode 2.
2
F → CNG/ENT → LOWER → CNG/ENT
8
UNDER → CNG/ENT → → CNG/ENT
Period (sec.)
(0.0 ~ 9.9)
b)
1) Select setting mode 2.
2
F → CNG/ENT → LOWER → CNG/ENT
9
FINAL → CNG/ENT → → CNG/ENT
Range
(0 ~ 9999)
60
5.FUNCTION SETTINGS
The Gross weight is forcibly zeroed by this function. The Net weight changes according to the
following expression: (Net weight) = (Gross weight) - (Tare weight).
If the digital zero operation is performed where Gross weight exceeds the DZ Regulation Value
(Please refer to "5-10.Zero Regulation Value" on P.62.), “ ZALM ” flashes to give a warning of the
problem.
In this case DZ Regulation Value is subtracted from Gross weight.
Operation
2) Gross is zeroed.
By short-circuiting A3pin (D/Z ON) and COM on the control connector, Gross weight is
immediately zeroed.
Net weight changes according to the following expression:
(Net weight) = (Gross weight) - (Tare weight).
OFF
ON
* In a case where 0 is not indicated when performing digital zero operation or when “ ZALM ”
flashes, the following causes are considered.
Cause Measure
Digital zero operation was performed - Change the set value of the DZ Regulation
when the DZ Regulation Value Value and perform digital zero operation
(default value 200) was exceeded. again. (However, since this is a temporary
measure, perform Zero Calibration at an
early time.)
- Remove lightweight residue adhered to the
tank.
- Check that there is no mechanical contact.
61
5.FUNCTION SETTINGS
If this operation is performed when “ ZALM ” flashes, the digital zero is cleared, and “ ZALM ” goes
out.
Operation
9
F → CNG/ENT → FINAL → CNG/ENT
Set a range of zero point adjustment (a gap from the registered Zero Calibration value) by digital
zero or zero tracking.
If the digital zero operation is performed or zero tracking operates where the DZ Regulation Value
Operation
3
F → CNG/ENT → NEAR Z. → CNG/ENT
6
CPS → CNG/ENT → → CNG/ENT
DZ Regulation Value
(0 ~ 9999)
62
5.FUNCTION SETTINGS
However, the operation is performed only when “ STAB ” is on depending on the setting of
Restriction on the Tare Subtraction Function in setting mode 4. The range of Tare subtraction can be
selected from all range or 0 < Tare ≦ Capacity.
Operation
1) Press TARE .
2) “ TARE ” lights.
By short-circuiting A4pin (Tare subtraction ON) and COM on the control connector, Net weight is
OFF
ON
* If Net weight is not zeroed by performing the operation of Tare subtraction, the following causes
are presumable.
Cause Measure
The indicated value is Gross weight. GROSS
Press the /NET
ESC key to change the
indicated value into Net weight.
63
5.FUNCTION SETTINGS
The subtracted Tare can be restored and Net weight becomes equal to Gross weight.
Operation
TARE
1) Press RESET .
By short-circuiting A5pin (Tare subtraction OFF) and COM on the control connector, Tare
subtraction is immediately reset, and Net weight and Gross weight are equalized.
OFF
ON
* If Net weight and Gross weight are not equalized by resetting Tare subtraction, the following
cause is presumable.
Cause Measure
The digital tare subtraction Set the value of Tare Weight in setting mode
function is valid. 1 to 0, or set digital tare subtraction
ON/OFF in setting mode 2 to OFF.
64
5.FUNCTION SETTINGS
Set Tare Weight by pressing TARE key then set digital tare subtraction ON.
Operation
a)
1) Select setting mode 2.
2
F → CNG/ENT → LOWER → CNG/ENT
3
NEAR Z. → CNG/ENT → ZERO ZERO ZERO ZERO → CNG/ENT
b)
1) Select setting mode 1.
1
F → CNG/ENT → UPPER → CNG/ENT
9
FINAL → CNG/ENT → → CNG/ENT
Tare Weight
(0 ~ 99999)
65
5.FUNCTION SETTINGS
It can be restricted the function of digital tare subtraction or one touch tare subtraction.
Operation
4
F → CNG/ENT → SP1 → CNG/ENT
9
FINAL → CNG/ENT → → CNG/ENT
66
5.FUNCTION SETTINGS
be switched over ON/OFF by using the signal from control connector to avoid false operation on
the display.
- External input mode ......... Switching with the signal (revel input) from the control connector.
Operation
4
F → CNG/ENT → SP1 → CNG/ENT
7
OVER → CNG/ENT → ZERO ZERO ZERO → CNG/ENT
OFF OFF
ON ON
Net weight display results at ON, and Gross weight display results at OFF.
OFF
Gross weight display Net weight display
ON
67
5.FUNCTION SETTINGS
Operation
1) Select setting mode 4.
4
F → CNG/ENT → SP1 → CNG/ENT
7
OVER → CNG/ENT →
Operation
a), b), c), d)
1) Select setting mode 2.
2
F → CNG/ENT → LOWER → CNG/ENT
5
SP2 → CNG/ENT → → CNG/ENT
68
6.DISCHARGING CONTROL MODE
In this example system, raw materials are to be fed from the material tank to the weighing tank. At
first, the feeding valves are to be fully opened to feed, and closed in the order of Set Point 1 → Set
Point 2, and Set Point 2 → Set Point 3 at the time of Final - Set Point 1, and Final - Set Point 2,
respectively. The feeding valves are to be completely closed at the time of Final - Compensation.
The weighed raw materials are to be discharged to a container by opening the discharge valves.
F701-C
69
6.DISCHARGING CONTROL MODE
(1) By setting external input Tare subtraction ON (or with the TARE key on the panel), Net
(2) The feeding valves are fully opened to start feeding. When the weighed value reaches (Final -
Set Point 1), Set Point 1 output signal turns ON. (If the comparison inhibit timer is set, the timer
starts.) Set Point 1 valve of the material tank is closed.
(3) When the weighed value reaches (Final - Set Point 2), Set Point 2 output signal turns ON. (If the
comparison inhibit timer is set, the timer starts.) Set Point 2 valve of the material tank is
closed”.
(4) When the weighed value reaches (Final - Compensation), Set Point 3 output signal turns ON,
and if the judging timer is set, the timer starts. The feeding valves are completely closed.
(5) After the judging timer expires, Over/Under judgment is made. If the weighed value exceeds the
range of the Over/Under setting value, the over signal or under signal turns ON.
(6) Raw materials are discharged from the weighing tank to a container. The valves of the weighing
tank should be opened. Completion of discharge is confirmed with the Near Zero signal. In
weighing from the second time onward, (1) ~ (5) are repeated.
The feeding valves and discharge valves should be opened and closed through
the sequencer or relay sequence according to the control signals from the
F701-C.
70
6.DISCHARGING CONTROL MODE
NET
Final Over
Final – CPS Under
Final – SP2
Final – SP1
Near Zero
0 TIME
Tare subtraction
SP1
SP2
SP3
Comparison
Inhibit Timer
Judging Timer
Near Zero
Judgment
(Go, Over or Under)
71
6.DISCHARGING CONTROL MODE
First supply raw material into a metering tank (hopper), then controls a weight while discharging
raw material from the metering tank.
In discharging weighing, the amount of discharge can be weighed by adding negative weighed
values. In this example system, raw materials are to be fed from the material tank to the weighing
tank, and a fixed quantity of raw materials are to be discharged from the weighing tank to a
container. At first, the discharge valves of the weighing tank are to be fully opened to discharge raw
materials, and the discharge valves are to be closed in the order of Set Point 1 → Set Point 2, and
Set Point 2 → Set Point 3 at the time of Final - Set Point 1, and Final - Set Point 2, respectively. The
discharge valves are to be completely closed at the time of Final - Compensation, when one
measurement is completed. When the weighing tank runs short, the feeding valves are to be opened
to replenish the weighing tank with raw materials from the material tank to weigh.
F701-C
72
6.DISCHARGING CONTROL MODE
(1) The valves of the material tank are opened by the lower limit signal, and raw materials are
fed into the weighing tank.
(2) It is detected by the upper limit signal that the weighing tank is full, and the valves of the mate-
rial tank are closed.
(3) By setting external input Tare subtraction ON (or with the TARE key on the panel), Net
(4) The discharge valves are fully opened to start discharging. When the weighed value reaches
(Final - Set Point 1), Set Point 1 output signal turns ON. (If the comparison inhibit timer is set,
the timer starts.) The discharge valve (Set Point 1) is closed.
(5) When the weighed value reaches (Final - Set Point 2), Set Point 2 output signal turns ON. (If the
comparison inhibit timer is set, the timer starts.) The discharge valve (Set Point 2) is closed.
(6) When the weighed value reaches (Final - Compensation), Set Point 3 output signal turns ON,
and if the judging timer is set, the timer starts. The discharge valves are completely closed.
(7) After the judging timer expires, Over/Under judgment is made. If the weighed value exceeds the
range of the Over/Under setting value, the over signal or under signal is output. In weighing
from the second time onward, (3) ~ (6) are repeated in a like manner.
(8) When the weighing tank runs short of raw materials, the valves of the material tank are opened
by the output of the lower limit signal to feed raw materials into the weighing tank.
The feeding valves and discharge valves should be opened and closed through
the sequencer or relay sequence according to the control signals from the
F701-C.
73
6.DISCHARGING CONTROL MODE
0 TIME
Final – SP1
Final – SP2
Final – CPS
NET
Tare subtraction
SP1
SP2
SP3
Comparison
Inhibit Timer
Judging Timer
Lower Limit
Judgment
(Go, Over or Under)
The Near Zero signal is used for discharge completion confirmation as in the
case of feeding weighing.
74
6.DISCHARGING CONTROL MODE
Set the feeding or discharging. Selectable methods are feeding control, discharging control and
external selection (Changing the feed/Discharge by the signal from control connector).
Operation
2
F → CNG/ENT → LOWER → CNG/ENT
1
UPPER → CNG/ENT → → CNG/ENT
Switch with the pin set at “Feed/Discharge” out of B2 to B5 pin on the control connector.
OFF → Feeding control, ON → Discharging control.
75
6.DISCHARGING CONTROL MODE
NET
Final Over
Final – CPS Under
Final – SP2
Final – SP1
Near Zero
0 TIME
OFF
Near Zero ON
SP1 ON
SP2 ON
SP3 ON
Comparison
Inhibit Timer
t1 t1
Judging Timer
t2
Stable ON
Complete ON
Complete Timer
t3
Over
Go ON
OFF
Under
76
6.DISCHARGING CONTROL MODE
● The Over/Under comparison timing depends on the Over/Under comparison mode setting
(Weighing Function 2 in setting mode 2).
(It is set at “Regularly” in the illustration.)
● The completion signal output timing depends on the complete signal output mode setting
(Weighing Function 2 in setting mode 2).
● Conditional expression
- Near Zero turns ON when Weight value ≦ Set value of Near Zero
- SP1 output turns ON when Weight value ≧ Final - SP1
- SP2 output turns ON when Weight value ≧ Final - SP2
- SP3 output turns ON when Weight value ≧ Final - CPS
- Under turns ON when Weight value < Final - Under
- Over turns ON when Weight value > Final + Over
- Go turns ON when Final + Over ≧ Weight value ≧ Final - Under
● Select the weight value for Near Zero comparison from Gross weight/Net weight (Weighing
Function 1 in setting mode 2).
● Select the comparison weight value for Set Point 1, Set Point 2 and Set Point 3 output signals,
and Over, Go and Under judgment signals from Gross weight/Net weight (Weighing Function 1
in setting mode 2).
This is a control method which compares the weight value with discharging value after starting
signal of weighing ON.
When the starting signal ON, output of each signal of Set Point 1, Set Point 2, and Final turns ON
then the signal turns OFF after weight value satisfied the each condition.
Sequential control can be conducted without connecting external sequencer in sequence control.
Sequence controls are classified into following categories.
77
6.DISCHARGING CONTROL MODE
NET
Final Over
Under
Final – CPS
Final – SP2
Final – SP1
Near Zero
0 TIME
Tare Subtraction
ON
Start ON
OFF
Stop
SP1 ON
SP2 ON
SP3 ON
Complete ON
Comparison
Inhibiti Timer
t1 t1 t2 t3
Judging Timer
Complete Timer
HOLD
Near Zero
● The completion signal output timing depends on the complete signal output mode setting
(Weighing Function 2 in setting mode 2).
● The Over/Under comparison timing is such that the Over/Under comparison mode setting
(Weighing Function 2 in setting mode 2) is ignored, but comparisons are made when the
completion output is ON, and the weight value is held.
● Upper/Lower limit comparison is always made in defiance of the setting of Upper/Lower limit
comparison mode (Weighing Function 2 in setting mode 2).
78
6.DISCHARGING CONTROL MODE
● Conditional expression
- Near Zero turns ON when Weight value ≦ Set value of Near Zero
* Set Point 1 / Set Point 2 / Set Point 3 feeding gate signal is turned ON at the ON edge (OFF
→ ON) of the start signal.
● Select the weight value for Near Zero comparison from Gross weight/Net weight (Weighing
Function 1 in setting mode 2).
● Select the comparison weight value for Set Point 1, Set Point 2 and Set Point 3 output signals,
and Over, Go and Under judgment signals from Gross weight/Net weight (Weighing Function 1
in setting mode 2).
START ON
OFF
STOP
SP1 ON
SP2 ON
SP3 ON
Complete ON
Completion Timer
● When the setting of Judging Times in setting mode 1 is 00, Over/Under judgment is not made.
79
6.DISCHARGING CONTROL MODE
● The completion signal output timing is such that the complete signal output mode setting
(Weighing Function 2 in setting mode 2) is ignored, but it is output at the OFF edge (ON →
OFF) of the final signal (Set Point 3 signal).
START
ON
OFF
STOP
SP1 ON
SP2 ON
SP3 ON
Complete ON
Judging Timer
Completion Timer
HOLD
● The completion signal output timing depends on the complete signal output mode setting
(Weighing Function 2 in setting mode 2).
● The Over/Under comparison timing is such that the Over/Under comparison mode setting
(Weighing Function 2 in setting mode 2) is ignored, but comparisons are made when the
completion output is ON, and the weight value is held.
● Upper/Lower limit comparison is always made in defiance of the setting of Upper/Lower limit
comparison mode (Weighing Function 2 in setting mode 2).
80
6.DISCHARGING CONTROL MODE
START ON
OFF
STOP ON
SP1 ON
SP2 ON
SP3 ON
Sequence error
● When the stop signal turns ON during weight cycle, a sequence error ( ) will result and
all the signals of Set Point 1 output, Set Point 2 output and Set Point 3 output are turned OFF.
START ON
STOP ON ON
OFF
SP1
OFF
SP2
OFF
SP3
Sequence error
● If the start signal is turned ON when the stop signal is ON, it will become a sequence error
81
6.DISCHARGING CONTROL MODE
OFF
START ON
STOP ON
OFF
SP1
OFF
SP2
OFF
SP3
Sequence error
*
● For resetting the sequence error ( * ), input the ON edge (OFF → ON) of the stop signal.
The start signal is not received at the time of a sequence error.
Operation
2
F → CNG/ENT → LOWER → CNG/ENT
4
SP1 → CNG/ENT → → CNG/ENT
1 : Sequence mode
0 : Simple comparison mode
82
6.DISCHARGING CONTROL MODE
Auto free fall compensation adjusts the amount of suspended ingredient automatically to reduce
weight error. Register auto free fall compensation parameter.
Sampling the weight value when the complete signal turns ON after Set Point 3 finished.
Record “n” times(A) the difference(D) between set Final value and actual weight value, average the
difference and multiplied by compensation coefficient, then add the result of Compensation value.
Request
83
6.DISCHARGING CONTROL MODE
84
6.DISCHARGING CONTROL MODE
c) Average count of AFFC. Number of weighing value for calculating the average of
compensation.
(Input range / 1 ~ 9)
1
F → CNG/ENT → UPPER → CNG/ENT
7
OVER → CNG/ENT → → CNG/ENT
2
F → CNG/ENT → LOWER → CNG/ENT
3
NEAR Z. → CNG/ENT → ZERO → CNG/ENT
85
6.DISCHARGING CONTROL MODE
WEIGHT
Final Over
Final – CPS Under
Final – SP2
Final – SP1
Near Zero
0 TIME
OFF
Near Zero ON
SP1 ON
SP2 ON
SP3
ON
Over
Go ON
Under OFF
Outputting the control signal to external part from the control connector by the value set here.
86
6.DISCHARGING CONTROL MODE
* Not for using Set Point 1 and Set Point 2, make the settings of Set Point 1 and Set Point 2 equal
to the Final.
Operation
a)
1) Select setting mode 0.
0
F → CNG/ENT → → CNG/ENT
9
FINAL → CNG/ENT → → CNG/ENT
Final
(0 ~ 99999)
b)
1) Select setting mode 0.
0
F → CNG/ENT → → CNG/ENT
5
SP2 → CNG/ENT → → CNG/ENT
Set Point 2
(0 ~ 99999)
87
6.DISCHARGING CONTROL MODE
c)
1) Select setting mode 0.
0
F → CNG/ENT → → CNG/ENT
4
SP1 → CNG/ENT → → CNG/ENT
Set Point 1
(0 ~ 99999)
d)
1) Select setting mode 0.
0
F → CNG/ENT → → CNG/ENT
6
CPS → CNG/ENT → → CNG/ENT
Compensation
(-9999 ~ 9999)
e)
1) Select setting mode 0.
0
F → CNG/ENT → → CNG/ENT
7
OVER → CNG/ENT → → CNG/ENT
Over
(0 ~ 999)
f)
1) Select setting mode 0.
0
F → CNG/ENT → → CNG/ENT
8
UNDER → CNG/ENT → → CNG/ENT
Under
(0 ~ 999)
88
6.DISCHARGING CONTROL MODE
- Inhibit ....................Not including the negative value of the input range of compensation.
(Input range is 0 ~ 9999.)
Operation
2
F → CNG/ENT → LOWER → CNG/ENT
3
NEAR Z. → CNG/ENT → → CNG/ENT
Negative Compensation
1 : Effect
0 : Inhibit
89
6.DISCHARGING CONTROL MODE
〈 Conditional formula 〉
Operation
a)
1) Select setting mode 0.
0
F → CNG/ENT → → CNG/ENT
3
NEAR Z. → CNG/ENT → → CNG/ENT
Near Zero
(0 ~ 99999)
b)
1) Select setting mode 0.
0
F → CNG/ENT → → CNG/ENT
1
UPPER → CNG/ENT → → CNG/ENT
Upper Limit
(0 ~ 99999)
c)
1) Select setting mode 0.
0
F → CNG/ENT → → CNG/ENT
2
LOWER → CNG/ENT → → CNG/ENT
Lower Limit
(0 ~ 99999)
90
6.DISCHARGING CONTROL MODE
It enables to select the weight (Gross/Net) to be compared and the timing of comparison at the each
comparison point of Upper/Lower limit, Near Zero or Over/Under.
Operation
a), c), d)
1) Select setting mode 2.
2
F → CNG/ENT → LOWER → CNG/ENT
2) Set Upper/Lower limit comparison, Near Zero comparison and Final and Over/Under compari-
son. (1-Weighing Function 1)
1
UPPER → CNG/ENT → ZERO → CNG/ENT
91
6.DISCHARGING CONTROL MODE
b), e)
1) Select setting mode 2.
2
F → CNG/ENT → LOWER → CNG/ENT
2
LOWER → CNG/ENT → ZERO ZERO → CNG/ENT
OFF
ON
Judge
Undecided interval
(within 50ms)
92
6.DISCHARGING CONTROL MODE
The function which inhibits the comparison for certain period of time to eliminate false control or
judgement with the vibration caused by opening or closing of valve.
Final – CPS
Final –SP2
Judging Time
Final –SP1
Setting the period of complete signal (Control connector B10pin) output when the weighing is
completed.
NET
Final Over
Final – CPS Under
Final –SP2
Final –SP1
Near Zero
0 TIME
Stab ON
Complete ON
Complete time
93
6.DISCHARGING CONTROL MODE
a) Complete signal output mode (Select from Judging Time / Judge & Stable / Judging / Stable)
b) Complete Output Time (Input range / 0.00 ~ 9.99)
Operation
a)
1) Select setting mode 2.
2
F → CNG/ENT → LOWER → CNG/ENT
2
LOWER → CNG/ENT → ZERO ZERO ZERO ZERO → CNG/ENT
b)
1) Select setting mode1.
1
F → CNG/ENT → UPPER → CNG/ENT
3
NEAR Z. → CNG/ENT → → CNG/ENT
94
6.DISCHARGING CONTROL MODE
c)
1) Select setting mode 1.
1
F → CNG/ENT → UPPER → CNG/ENT
2
LOWER → CNG/ENT → → CNG/ENT
d)
1) Select setting mode 1.
1
F → CNG/ENT → UPPER → CNG/ENT
1
UPPER → CNG/ENT → → CNG/ENT
e)
1) Select setting mode 5.
5
F → CNG/ENT → SP2 → CNG/ENT
2
LOWER → CNG/ENT → ZERO → CNG/ENT
- Unless the weight value once drops down to 25% or less of the Final after the
complete output, the complete output will not turn on after the next weighing.
95
6.DISCHARGING CONTROL MODE
a) Judging Times
The Selection whether conduct the judgment on Over, Under or Go when complete signal is
displayed after completing the weighing.
Set the number from 00 to 99.
00 : Without judging
01 : Conducting judging every time
02 : Conducting an judging for every twice
03 : Conducting an judging for every three times
Complete
AFFC.
Request
96
6.DISCHARGING CONTROL MODE
b) AZ times
00 : Without auto 0
01 : Conduct an auto zero every time.
02 : Conduct an auto zero for every twice.
03 : Conduct an auto zero for every three times.
Start
AZ times 3 times
Auto zero
- Tare subtraction and digital zero can be operated via the front keys or the
control connector even if you set the auto zero to (00) without auto zero.
Setting for confirming whether Near Zero signal ON at the start of weighing.
(Select from ON or OFF).
Weighing will normally start if the Near Zero is ON. “ ” is displayed if the Near Zero is
OFF.
Refer to "6-6.Near Zero / Upper Limit / Lower Limit" on P.90 concerning Near Zero setting.
97
6.DISCHARGING CONTROL MODE
Confirm whether the weighing value has reached SP1 at the start of weighing.
Operation
a)
1) Select setting mode 1.
1
F → CNG/ENT → UPPER → CNG/ENT
6
CPS → CNG/ENT → → CNG/ENT
Judging Times
(0 ~ 99)
b)
1) Select setting mode 1.
1
F → CNG/ENT → UPPER → CNG/ENT
5
SP2 → CNG/ENT → → CNG/ENT
98
6.DISCHARGING CONTROL MODE
c), d), e)
1) Select setting mode 2.
2
F → CNG/ENT → LOWER → CNG/ENT
2) Set Near Zero check at start-time, weight value check at start-time and adjust feeding. (4-
Sequence Mode)
4
SP1 → CNG/ENT → ZERO → CNG/ENT
(Sequence control)
1 : At start-time, weight value
1 : Adjust feeding ON
check ON
0: 〃 OFF
0 : 〃 OFF
(Sequence control)
1 : At start-time, Near Zero
check ON
0 : 〃 OFF
f)
1) Select setting mode 1.
1
F → CNG/ENT → UPPER → CNG/ENT
4
SP1 → CNG/ENT → → CNG/ENT
99
6.DISCHARGING CONTROL MODE
This function warns you that the Net or Gross weight exceeds the entered control value.
Operation
a)
1) Select setting mode 3.
3
F → CNG/ENT → NEAR Z. → CNG/ENT
4
SP1 → CNG/ENT → → CNG/ENT
Net Over
(0 ~ 99999)
b)
1) Select setting mode 3.
3
F → CNG/ENT → NEAR Z. → CNG/ENT
5
SP2 → CNG/ENT → → CNG/ENT
Gross Over
(0 ~ 99999)
100
6.DISCHARGING CONTROL MODE
Weight error
When the display is LOAD, OFL or ZALM, the output turns ON.
Sequence error
Weight error and sequence error output signals are selectable by setting.
Operation
5
F → CNG/ENT → SP2 → CNG/ENT
2
LOWER → CNG/ENT → ZERO ZERO → CNG/ENT
101
7.STATISTICAL AND ACCUMULATION FUNCTION SETTING AND OPERATION
Weight values are accumulated by this function. Based on the accumulated weight data, statistical
data, such as the maximum, minimum, and average, are calculated.
Weight values can be accumulated by the auto accumulation command or external input/output
signal, through RS-232C communication, RS-485 communication (option), CC-Link (option).
Up to 10000 accumulations can be made. For 10001 or more accumulations, the operation
“Accumulation clear” must be performed to clear the accumulated and statistical data.
While the number of accumulations is 10000, an accumulation error results. The accumulation error
can be outputted from B11 pin out of the external input/output signals (by setting).
Even while the accumulation error is given, other weighing operations and the print command (SI/
F), etc., are not affected.
Operations other than those relating to statistics and accumulations are performed normally.
The accumulation value and statistical data can be referred to in the subdisplay area or in the setting
mode 8.
(For details, please refer to "1-1.Front Panel"②Subdisplay on P.1, "3-2-7.Setting Mode 8" on P.38.)
102
7.STATISTICAL AND ACCUMULATION FUNCTION SETTING AND OPERATION
Operation
8
F → CNG/ENT → UNDER → CNG/ENT
1
UPPER → CNG/ENT → → CNG/ENT
2
LOWER
1-Average Weight (0 ~ 99999)
3
NEAR Z. 2-Max. Value (0 ~ 99999)
4
SP1
3-Min. Value (0 ~ 99999)
5
4-General Standard Deviation (0 ~ 99999)
SP2
5-Sample Standard Deviation (0 ~ 99999)
6
CPS 6-Accumulation Count (0 ~ 10000)
7 7-Latest Accumulation Data (0 ~ 99999)
OVER
8-Max. - Min. (0 ~ 99999)
8
UNDER
Weight values can be accumulated by the external input/output signal or through the RS-232C, RS-
485 (option), CC-Link (option) when desired.
Set one of the B2 to 5 pins out of the external input/output signals to “accumulation command”.
An immediate accumulation is executed at the ON edge of an input to the set pin.
OFF
ON
Pulse width 50msec or more
103
7.STATISTICAL AND ACCUMULATION FUNCTION SETTING AND OPERATION
Operation
1) Select setting mode 5.
5
F → CNG/ENT → SP2 → CNG/ENT
2) One of input selections 1~4 is set to the Accumulation Command. (1-Input Selection)
1
UPPER → CNG/ENT → → CNG/ENT
An immediate accumulation is executed when the accumulation command “CI” is received from
the host.
Host C I CR
104
7.STATISTICAL AND ACCUMULATION FUNCTION SETTING AND OPERATION
Operation
9
F → CNG/ENT → FINAL → CNG/ENT
9 1 2 3 5
FINAL → CNG/ENT → UPPER LOWER NEAR Z. SP2 → CNG/ENT
Accumulations can be cleared by the password input or external input/output signal or through the
RS-232C, RS-485 (option), CC-Link (option).
Set one of the B2 to 5 pins out of the external input/output signals to “accumulation clear”.
Immediate accumulation clearing is executed at the ON edge of an input to the set pin.
OFF
ON
Pulse width 50msec or more
105
7.STATISTICAL AND ACCUMULATION FUNCTION SETTING AND OPERATION
Operation
5
F → CNG/ENT → SP2 → CNG/ENT
2) One of input selections 1~4 is set to the Accumulation Clear. (1-Input Selection)
1
UPPER → CNG/ENT → → CNG/ENT
Immediate accumulation clearing is executed when the accumulation command “CJ” is received
from the host.
Host C J CR
106
7.STATISTICAL AND ACCUMULATION FUNCTION SETTING AND OPERATION
Select whether or not to automatically accumulate weight values in the F701-C. If the auto accumu-
lation command is on, an accumulation is made when the complete signal of the F701-C turns on.
At this time, the count, accumulation value and statistical data are updated, and also an auto print
command is sent onto the SI/F format.
However, in the case where the result of weighing is negative or in the case of “over scale”, no
accumulation is made if no judgment is set.
Weighing Value
Final over
Final - CPS under
Final - Setpoint 2
Final - Setpoint 1
Near Zero
0 Time
OFF
Near Zero ON
SP1 ON
SP2 ON
SP3 ON
Judging Time
t2
Stable
ON
Complete
ON
Accumulation
ON
t2 : Judging time
107
7.STATISTICAL AND ACCUMULATION FUNCTION SETTING AND OPERATION
2) Sequence control
Weighing Value
Final over
Final - CPS under
Final - Setpoint 2
Final - Setpoint 1
0 Time
Tare
ON
Start
ON OFF
SP1 ON
SP2 ON
SP3 ON
Complete
ON
Judging Time
Accumulation ON
Operation
2
F → CNG/ENT → LOWER → CNG/ENT
2
LOWER → CNG/ENT → ZERO → CNG/ENT
108
8.SYSTEM MODE
8. SYSTEM MODE
8-1. LOCK (soft)
Lock for protecting from false operation, refer to "18.THE LIST OF INITIAL SETTING VALUE"
on P.176 regarding effective setting value for LOCK (soft).
Select from ON/OFF.
Operation
4
F → CNG/ENT → SP1 → CNG/ENT
8
UNDER → CNG/ENT → → CNG/ENT
LOCK1 LOCK2
1 : ON 1 : ON
0 : OFF 0 : OFF
8-2. Password
Operation
9
F → CNG/ENT → FINAL → CNG/ENT
9 1 2 3 5
FINAL → CNG/ENT → UPPER LOWER NEAR Z. SP2 → CNG/ENT
109
8.SYSTEM MODE
This equipment is provided with the self-check function by which the memory is automatically
checked to detect problems, and the visual check function by which the display can be checked
visually.
Description Type
1 Software version Display
2 All lighting on the display Display
3 RAM Read/Write Check Auto ← is displayed when error occurs.
4 ROM check sum Check Auto ← is displayed when error occurs.
5 Status sequential lighting Visual
6 7-segment lighting on the display Visual
7 NOV RAM Read/Write Check Auto ← is displayed when error occurs.
PASS display and check end Display
* The software version display may vary according to the time of purchase.
* When the LOCK switch on the rear panel is ON, the NOV RAM is not checked.
* If checking in progress stops or the display is not correct, it is faulty. Ask our agency or us for
repair.
● Memory clear
(setting values and work areas) is cleared, and the self-check program is executed.
* However, the settings and calibrated values stored in the NOV RAM (nonvolatile RAM) remain
unchanged. (Refer to "18.THE LIST OF INITIAL SETTING VALUE" on P.176.)
110
9.EXTERNAL INPUT/OUTPUT SIGNALS (CONTROL CONNECTOR)
A1 * COM B1 * COM
A2 In G/N B2 In Input Selection 1 *1
A3 In D/Z ON B3 In Input Selection 2 *1
A4 In Tare ON B4 In Input Selection 3 *1
A5 In Tare OFF B5 In Input Selection 4 *1
A6 Out Near Zero B6 Out Lower Limit
A7 Out SP1 B7 Out Upper Limit
A8 Out SP2 B8 Out Stable
A9 Out SP3 B9 Out Output Selection 1 *2
A10 Out Under B10 Out Output Selection 2 *2
A11 Out Over B11 Out Output Selection 3 *2
A12 * COM B12 * COM
* *1 : Selectable by setting.
(For details, please refer to "9-4-11.Input Selection" on P.117)
* *2 : Selectable by setting. (For details, please refer to "6-11.Weight Error / Sequence Error" on
P.101, "6-8.Complete Signal Output Mode / Complete Output Time / Judging Time /
Comparison Inhibit Time / Output Selection 2" on P.93, "9-5-7.Accumulation Error" on
P.119)
111
9.EXTERNAL INPUT/OUTPUT SIGNALS (CONTROL CONNECTOR)
A signal is inputted to the signal input circuit by short-circuiting or opening the input terminal and
the COM terminal. Short-circuiting is effected by means of a contact (such as a relay or a switch) or
a noncontact (such as a transistor or an open-collector TTL).
Vcc
approx.
Push TTL Open Collector
IC=6mA Toggle Switch Relay Transistor
Switch (ON when IN is HI)
IN
IN
COM
Vcc +12V
Vext Spark killer
F701-C
Inside
DC Power
Load
Relay
COM Varistor
Spark killer
Vceo=30V(max) AC power
Ic =120mA (max)
Load
- Transistor status
Output data Tr - Use external power source (up to DC30V) for driving relay
(Vext).
0 OFF
- Do not short-circuit the load, such as a coil of relay, that will
1 ON
break the output transistor.
- Connect a surge absorber or a speark killer to the relay
circuit as shown in the draft so that to reduce noise trouble
and extend the life of relay.
Noise trouble can be reduced, and the relay’s life can be
extended.
112
9.EXTERNAL INPUT/OUTPUT SIGNALS (CONTROL CONNECTOR)
Edge input or level input is selectable by the setting of Gross weight/Net weight display (External
Function Selection in setting mode 4).
* For the method of setting, refer to "5-15.Switching Gross Weight/Net Weight Display" on P.67.
When external input is ON edge (OFF → ON), display switches to Net weight.
When external input is OFF edge (ON → OFF), display switches to Gross weight.
OFF OFF
ON ON
Net weight display results at ON, and Gross weight display results at OFF.
OFF
Gross weight display Net weight display
ON
When external input is ON edge (OFF → ON), it brings the Gross weight to zero.
This function works within the range of DZ Regulation Value.
If out of this range, zero will not result, but “ ZALM ” flashes.
The ZERO → CNG/ENT key operation also results in the same movement (when the ZERO key is
valid under the Function Key Invalid setting in setting mode 2).
OFF
ON
113
9.EXTERNAL INPUT/OUTPUT SIGNALS (CONTROL CONNECTOR)
Subtract the Tare at once by external input ON edge (OFF → ON), it brings the Net weight to zero.
The TARE key operation also results in the same movement (when [TARE] key is valid under the
However, the operation is performed only when “ STAB ” is on depending on the setting of
Restriction on the Tare Subtraction Function in setting mode 4. The range of Tare subtraction can be
selected from all range or 0 < Tare ≦ Capacity.
OFF
ON
Pulse width 50msec or more
TARE
The RESET key operation also results in the same movement (when [TARE RESET]
key is valid under the Function Key Invalid setting in setting mode 2)
OFF
ON
Pulse width 50msec or more
OFF
ON
HOLD
undefined section
(within 50mSec.)
114
9.EXTERNAL INPUT/OUTPUT SIGNALS (CONTROL CONNECTOR)
* For the method of setting, refer to "6-7.Upper/Lower Limit Comparison / Upper/Lower Limit
Comparison Mode / Near Zero Comparison / Final and Comparison / Over/Under Comparison
Mode" on P.91.
* Judge function is ineffective in Sequence Mode.
OFF
ON
Judge ON
undefined section
(within 50mSec.)
Feeding control is activated when it is OFF. Discharging control is activated when it is ON.
115
9.EXTERNAL INPUT/OUTPUT SIGNALS (CONTROL CONNECTOR)
An immediate accumulation is executed at the ON edge (OFF → ON) of the external input.
The weight value (gross or net) set in the over/under comparison mode is accumulated.
OFF
ON
Pulse width 50msec or more
Immediate clearing of accumulated data and statistical data is executed at the ON edge (OFF →
ON) of the external input.
OFF
ON
Pulse width 50msec or more
116
9.EXTERNAL INPUT/OUTPUT SIGNALS (CONTROL CONNECTOR)
On each pin of input selection 1 (B2) to 4 (B5) of the external input/output signals, the input signal
can be selected by setting.
Settable items are accumulation clear, accumulation command, stop, start, feed/discharge, judge,
and hold (common to input selection 1 to 4).
However, except the hold signal, the same signal cannot be selected on two or more pins. If the
same signal is selected on two or more pins, forced setting of “HOLD” will result.
When “HOLD” is selected on two or more pins, the weight value is held if at least one signal is ON.
Operation
5
F → CNG/ENT → SP2 → CNG/ENT
1
UPPER → CNG/ENT → → CNG/ENT
117
9.EXTERNAL INPUT/OUTPUT SIGNALS (CONTROL CONNECTOR)
* For the method of setting, refer to "6-6.Near Zero / Upper Limit / Lower Limit" on P.90, "6-
7.Upper/Lower Limit Comparison / Upper/Lower Limit Comparison Mode / Near Zero
Comparison / Final and Comparison / Over/Under Comparison Mode" on P.91.
Conditional expression
The lower limit output turns ON when Weight value < Lower Limit
The upper limit output turns ON when Weight value > Upper Limit
Upper/Lower Limit
in setting mode 0
Upper/Lower limit comparison under Weighing Function 1
in setting mode 2
* For the method of setting, refer to "6-6.Near Zero / Upper Limit / Lower Limit" on P.90, "6-
7.Upper/Lower Limit Comparison / Upper/Lower Limit Comparison Mode / Near Zero
Comparison / Final and Comparison / Over/Under Comparison Mode" on P.91.
9-5-3. Stable
This output turns ON when the weight value is stable.
* For the method of setting, refer to "6-11.Weight Error / Sequence Error" on P.101.
* For the error display, refer to "11.OVER SCALE & ERROR" on P.157.
118
9.EXTERNAL INPUT/OUTPUT SIGNALS (CONTROL CONNECTOR)
* For the method of setting, refer to "6-11.Weight Error / Sequence Error" on P.101.
* For the error display, refer to "11.OVER SCALE & ERROR" on P.157.
9-5-6. RUN
Operation
5
F → CNG/ENT → SP2 → CNG/ENT
2
LOWER → CNG/ENT → → CNG/ENT
119
9.EXTERNAL INPUT/OUTPUT SIGNALS (CONTROL CONNECTOR)
* For the method of setting, refer to "6-4.Final / Set Point 2 / Set Point 1 / Compensation / Over /
Under" on P.86, "6-7.Upper/Lower Limit Comparison / Upper/Lower Limit Comparison Mode /
Near Zero Comparison / Final and Comparison / Over/Under Comparison Mode" on P.91.
◆ In sequence mode
In sequence mode, the weighing sequence is started at the ON edge of the start signal (OFF → ON),
Set Point 1, Set Point 2 and Set Point 3 turn ON.
* For the method of setting, refer to "6-4.Final / Set Point 2 / Set Point 1 / Compensation / Over /
Under" on P.86, "6-7.Upper/Lower Limit Comparison / Upper/Lower Limit Comparison Mode /
Near Zero Comparison / Final and Comparison / Over/Under Comparison Mode" on P.91, "9-4-
11.Input Selection" on P.117.
120
9.EXTERNAL INPUT/OUTPUT SIGNALS (CONTROL CONNECTOR)
Select comparison mode in Over/Under comparison mode (Weighing Function 2 in setting mode 2).
The go signal turns ON when Final + Over ≧ Weight value ≧ Final - Under
* For the method of setting, refer to "6-4.Final / Set Point 2 / Set Point 1 / Compensation / Over /
Under" on P.86, "6-7.Upper/Lower Limit Comparison / Upper/Lower Limit Comparison Mode /
Near Zero Comparison / Final and Comparison / Over/Under Comparison Mode" on P.91, "6-
8.Complete Signal Output Mode / Complete Output Time / Judging Time / Comparison Inhibit
Time / Output Selection 2" on P.93.
◆ In sequence mode
Compares the weight value when complete signal ON and hold it, while ignoring the setting of
Over/Under comparison mode (Weighing Function 2 in setting mode 2).
(only when judge ON)
The go signal turns ON when Final + Over ≧ Weight value ≧ Final - Under
* For the method of setting, refer to "6-4.Final / Set Point 2 / Set Point 1 / Compensation / Over /
Under" on P.86, "6-7.Upper/Lower Limit Comparison / Upper/Lower Limit Comparison Mode /
Near Zero Comparison / Final and Comparison / Over/Under Comparison Mode" on P.91, "6-
8.Complete Signal Output Mode / Complete Output Time / Judging Time / Comparison Inhibit
Time / Output Selection 2" on P.93.
121
9.EXTERNAL INPUT/OUTPUT SIGNALS (CONTROL CONNECTOR)
9-5-10. Complete
Select the complete signal output timing in complete signal output mode (Weighing Function 2 in
setting mode 2).
The output ON time depends on the Complete Output Time in setting mode 1.
* For the method of setting, refer to "6-8.Complete Signal Output Mode / Complete Output Time /
Judging Time / Comparison Inhibit Time / Output Selection 2" on P.93.
- Unless the weight value once drops down to 25% or less of the Final after the
complete output, the complete output will not turn ON after the next weighing.
◆ In sequence mode
Select the complete signal output timing in complete signal output mode (Weighing
Function 2 in setting mode 2).
The output ON time depends on the Complete Output Time in setting mode 1.
The complete output turns ON in defiance of the setting of complete signal output mode
(Weighing Function 2 in setting mode 2) on the OFF edge (ON → OFF) of the set point 3
signal.
The output ON time depends on the Complete Output Time in setting mode 1.
* The complete output signal is used for go output by setting (output selection 2 under Output
Selection in setting mode 5).
(Either go or complete is selectable.)
* For the method of setting, refer to "6-8.Complete Signal Output Mode / Complete Output Time /
Judging Time / Comparison Inhibit Time / Output Selection 2" on P.93.
122
9.EXTERNAL INPUT/OUTPUT SIGNALS (CONTROL CONNECTOR)
Each signal of the weight error, sequence error, complete, go, RUN, and accumulation error outputs
is selective by setting.
The weight error and sequence error are outputted from output selection 1 of B9 pin.
Complete and Go
The complete and go are outputted from output selection 2 of B10 pin.
The RUN and accumulation error are outputted from output selection 3 of B11 pin.
Operation
5
F → CNG/ENT → SP2 → CNG/ENT
2
LOWER → CNG/ENT → → CNG/ENT
123
10.INTERFACE
10. INTERFACE
10-1. 2-wire Serial Interface (SI/F)
This 2-wire serial interface is for connecting the F701-C to peripheral equipment such as a printer
or remote display. A two-core parallel cable or a cabtyre cable (Wire with covering thickened for
construction) may be used for connection. When a two-core parallel cable or a cabtyre cable is
used, the transmitting distance is approximately 30m. When a two-core shielded twisted pair cable
is used, the transmitting distance is approximately 300m.
10-1-1. Connection
Up to three nonpolarized external devices can be connected. It should not run along side AC or
other high-voltage wiring. It may cause of malfunction.
Example 1 : SIF
LD514 Remote display
Tare display
Example 2 : SIF
LD557A
Remote display
LD514 M252B
Remote display Printer
* These indicated above are examples for the connection of remote displays (LD557A, LD514,
M252B). Each display can select the data (Gross weight, Net weight, Tare weight) individually.
124
10.INTERFACE
10-2-2. Cable
F701-C personal computer (9-pin)
1 Cross type 1 CD
2 RXD 2 RXD
3 TXD 3 TXD
4 4 DTR
6 6 DSR
7 RTS 7 RTS
8 CTS 8 CTS
5 GND 5 GND
9 9 RI
The above diagram is for connecting a personal computer as a DTE (Data Terminal Equipment)
device. If it is a DCE (Data Circuit-terminating Equipment) device, connect pin to pin (DTR to
DTR, DSR to DSR etc.).
Cables should be prepared after checking connector type and pin assignments of the connected
device.
125
10.INTERFACE
Operation
4
F → CNG/ENT → SP1 → CNG/ENT
6
CPS → CNG/ENT → → CNG/ENT
Stop bit
1 : 2bit
0 : 1bit
Parity bit
2 : Even
1 : Odd
0 : None
2. The initial set for RS-232C port of connecting personal computer and sequencer
should be the same setting of F701-C.
- Communication mode 0
F701-C communicates with the command from the host computer.
The terminator can be selected “CR” or “CR+LF”.
- Communication mode 1
F701-C sends Gross weight continuously.
Command is invalid at this mode.
126
10.INTERFACE
- Communication mode 2
F701-C sends Net weight continuously.
Command is invalid at this mode.
- Communication mode 3
F701-C sends Gross and Net weight continuously.
Command is invalid at this mode.
- Communication mode 4
F701-C sends Gross weight at the timing of auto print command.
Command is invalid at this mode.
- Communication mode 5
F701-C sends Net weight at the timing of auto print command.
Command is invalid at this mode.
- Communication mode 6
F701-C sends Gross and Net weight at the timing of auto print command.
Command is invalid at this mode.
Request
127
10.INTERFACE
- Transmission format 1
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
G S , *1 , *2 , *3 , *4 , ± CR LF
- Transmission format 2
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
N T , *1 , *2 , *3 , *4 , ± CR LF
- Transmission format 3
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
SOH W T STX *1 *2 *3 *4 *5 ± ± ETX BCC
*3 H : Upper limit ON
L : Lower limit ON
G : Upper limit and Lower limit OFF
N : Upper limit and Lower limit ON
F : Comparison OFF
Priority N > (H or L)
128
10.INTERFACE
- Reading out the Gross weight (sign, 5-digt weight value, decimal point)
Host R A CR
F701-C R A + 1 0 0 . 0 0 CR LF
- Reading out the Net weight (sign, 5-digit weight value, decimal point)
Host R B CR
F701-C R B + 1 0 0 . 0 0 CR LF
- Reading out the Tare weight (sign, 5-digit weight value, decimal point)
Host R C CR
F701-C R C + 1 0 0 . 0 0 CR LF
Host R D CR
F701-C R D 0 0 0 0 0 0 0 CR LF
Weight value 1 : ON
hold 0 : OFF * Not defined
Weight value 1 : ON
Zero alarm 0 : OFF
1 : ON Weight 1 : Net
Tare subtracted 0 : OFF display 0 : Gross
129
10.INTERFACE
Host R E CR
F701-C R E 0 0 0 0 0 0 0 CR LF
Complte 1 : ON
SP1 1 : ON output signal 0 : OFF
output signal 0 : OFF
Over 1 : ON
output signal 0 : OFF
SP2 1 : ON
output signal 0 : OFF Go 1 : ON
output signal 0 : OFF
SP3 1 : ON
output signal 0 : OFF Under 1 : ON
output signal 0 : OFF
Host R F CR
F701-C R F 0 0 0 0 0 0 0 CR LF
* Not defined
Near zero 1 : ON
output signal 0 : OFF
Lower limit 1 : ON
output signal 0 : OFF
Upper limit 1 : ON
output signal 0 : OFF
130
10.INTERFACE
Host R G CR
F701-C R G 0 0 0 0 0 0 0 CR LF
Weight error 1 : ON
output signal 0 : OFF
* Not defined
Error 1 : ON
output signal 0 : OFF
Sequence error
0 or 1 ~ 5
3 : Span calibrating
Operation 2 : Zero calibrating * Same as the error
mode 1 : Weighing display of F701-C.
0:
Host R H CR
F701-C R H 9 9 9 9 9 CR LF
Host R I CR
F701-C R I 9 9 9 9 9 9 9 . 9 9 CR LF
Host R J CR
F701-C R J 0 0 + 1 0 0 . 0 0 CR LF
Host R J CR
F701-C R J CR LF
131
10.INTERFACE
Host W CR LF
F701-C W CR LF
Setting value No.
Setting value
* For setting value No., refer to "Setting value communication format" on P.134.
- Zero Calibration
Host C A CR
F701-C N A 0 CR LF
Calibrating
0: Normally completed, 1 ~ 9 : Error
(same as the error display of F701-C)
- Span Calibration
Host C B CR
F701-C N B 0 CR LF
Calibrating
0: Normally completed, 1 ~ 9 : Error
(same as the error display of F701-C)
Request
Before sending this command, set the Capacity, Min. Scale division and
Balance Weight Value.
132
10.INTERFACE
Host C C CR
Host C D CR
- Tare subtraction
Host C E CR
Host C F CR
- Digital zero
Host C G CR
Host C H CR
- Accumulation Command
Host C I CR
- Accumulation Clear
Host C J CR
133
10.INTERFACE
Use the following for reading and writing set values. For the communication formats, see p.132.
㻿㼑㼠㻌㻼㼛㼕㼚㼠㻌㻝
䠳 㻝 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻿㼑㼠㻌㻼㼛㼕㼚㼠㻌㻞
䠳 㻝 㻝 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻲㼕㼚㼍㼘
䠳 㻝 㻞 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻻㼢㼑㼞
䠳 㻝 㻟 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼁㼚㼐㼑㼞
䠳 㻝 㻠 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻯㼛㼙㼜㼑㼚㼟㼍㼠㼕㼛㼚
䠳 㻝 㻡 䠉 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻖㻌㻿㼕㼓㼚㻛㻌㻜㻦㻌㻼㼘㼡㼟㻌㻌䠉㻦㻌㻹㼕㼚㼡㼟
㻭㼡㼠㼛㻌㻲㼞㼑㼑㻌㻲㼍㼘㼘㻌㻯㼛㼙㼜㼑㼚㼟㼍㼠㼕㼛㼚㻌㻾㼑㼓㼡㼘㼍㼠㼕㼛㼚
䠳 㻝 㻢 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻭㼐㼖㼡㼟㼠㻌㻲㼑㼑㼐㼕㼚㼓㻌㼀㼕㼙㼑
䠳 㻝 㻣 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻶㼡㼐㼓㼕㼚㼓㻌㼀㼕㼙㼑
䠳 㻞 㻜 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻯㼛㼙㼜㼍㼞㼕㼟㼛㼚㻌㻵㼚㼔㼕㼎㼕㼠㻌㼀㼕㼙㼑
䠳 㻞 㻝 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼁㼜㼜㼑㼞㻌㻸㼕㼙㼕㼠
䠳 㻞 㻞 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻸㼛㼣㼑㼞㻌㻸㼕㼙㼕㼠
䠳 㻞 㻟 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻺㼑㼍㼞㻌㼆㼑㼞㼛
䠳 㻞 㻠 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼀㼍㼞㼑㻌㼃㼑㼕㼓㼔㼠
䠳 㻞 㻡 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻭㼡㼠㼛㻌㼆㼑㼞㼛㻌㼀㼕㼙㼑㼟
䠳 㻞 㻢 㻜 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻶㼡㼐㼓㼕㼚㼓㻌㼀㼕㼙㼑㼟
䠳 㻞 㻣 㻜 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻯㼛㼙㼜㼘㼑㼠㼑㻌㻻㼡㼠㼜㼡㼠㻌㼀㼕㼙㼑
䠳 㻞 㻤 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
134
10.INTERFACE
㻿㼑㼝㼡㼑㼚㼏㼑㻌㻹㼛㼐㼑
䠳 㻟 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼃㼑㼕㼓㼔㼕㼚㼓㻌㻲㼡㼚㼏㼠㼕㼛㼚㻌㻝
䠳 㻟 㻝 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼃㼑㼕㼓㼔㼕㼚㼓㻌㻲㼡㼚㼏㼠㼕㼛㼚㻌㻞
䠳 㻟 㻞 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼃㼑㼕㼓㼔㼕㼚㼓㻌㻲㼡㼚㼏㼠㼕㼛㼚㻌㻟
䠳 㻟 㻟 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻲㼡㼚㼏㼠㼕㼛㼚㻌㻷㼑㼥㻌㻵㼚㼢㼍㼘㼕㼐
䠳 㻟 㻠 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻭㼚㼍㼘㼛㼓㻌㻲㼕㼘㼠㼑㼞
䠳 㻟 㻡 㻜 㻜 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻰㼕㼓㼕㼠㼍㼘㻌㻲㼕㼘㼠㼑㼞
䠳 㻟 㻢 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻹㼛㼠㼕㼛㼚㻌㻰㼑㼠㼑㼏㼠㼕㼛㼚
䠳 㻟 㻣 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼆㼑㼞㼛㻌㼀㼞㼍㼏㼗㼕㼚㼓㻌㻼㼑㼞㼕㼛㼐
䠳 㻟 㻤 㻜 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼆㼑㼞㼛㻌㼀㼞㼍㼏㼗㼕㼚㼓㻌㻾㼍㼚㼓㼑
䠳 㻟 㻥 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻿㼑㼠㼠㼕㼚㼓㻌㼂㼍㼘㼡㼑㻌㻸㼛㼏㼗
䠳 㻟 㻭 㻜 㻜 㻜 䠟䠮 䠨䠢
㻮㼍㼘㼍㼚㼏㼑㻌㼃㼑㼕㼓㼔㼠㻌㼂㼍㼘㼡㼑
䠳 㻠 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻯㼍㼜㼍㼏㼕㼠㼥
䠳 㻠 㻝 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻹㼕㼚㻚㻌㻿㼏㼍㼘㼑㻌㻰㼕㼢㼕㼟㼕㼛㼚
䠳 㻠 㻞 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻺㼑㼠㻌㻻㼢㼑㼞
䠳 㻠 㻟 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻳㼞㼛㼟㼟㻌㻻㼢㼑㼞
䠳 㻠 㻠 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻲㼡㼚㼏㼠㼕㼛㼚㻌㻿㼑㼘㼑㼏㼠㼕㼛㼚
䠳 㻠 㻡 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻳㼞㼍㼢㼕㼠㼍㼠㼕㼛㼚㼍㼘㻌㻭㼏㼏㼑㼘㼑㼞㼍㼠㼕㼛㼚㻌㻔㼍㼞㼑㼍㻌㼚㼡㼙㼎㼑㼞㻌㼕㼚㼜㼡㼠㻕
䠳 㻠 㻢 㻜 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻰㼆㻌㻾㼑㼓㼡㼘㼍㼠㼕㼛㼚㻌㼂㼍㼘㼡㼑
䠳 㻠 㻣 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻳㼞㼍㼢㼕㼠㼍㼠㼕㼛㼚㼍㼘㻌㻭㼏㼏㼑㼘㼑㼞㼍㼠㼕㼛㼚㻌㻔㼍㼏㼏㼑㼘㼑㼞㼍㼠㼕㼛㼚㻌㼕㼚㼜㼡㼠㻕
䠳 㻠 㻤 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
135
10.INTERFACE
㪜㫏㫋㪼㫉㫅㪸㫃㩷㪝㫌㫅㪺㫋㫀㫆㫅㩷㪪㪼㫃㪼㪺㫋㫀㫆㫅
䌗 㪌 㪇 䌃䌒 䌌䌆 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩
㪩㪼㫊㫋㫉㫀㪺㫋㫀㫆㫅㩷㫆㫅㩷㫋㪿㪼㩷㪫㪸㫉㪼㩷㪪㫌㪹㫋㫉㪸㪺㫋㫀㫆㫅㩷㪝㫌㫅㪺㫋㫀㫆㫅
䌗 㪌 㪈 㪇 䌃䌒 䌌䌆 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩
㪛㪆㪘㩷㪦㫌㫋㫇㫌㫋㩷㪤㫆㪻㪼
䌗 㪌 㪉 㪇 㪇 㪇 䌃䌒 䌌䌆 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩
㪛㪆㪘㩷㪱㪼㫉㫆㩷㪦㫌㫋㫇㫌㫋㩷㪮㪼㫀㪾㪿㫋
䌗 㪌 㪊 䌃䌒 䌌䌆 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩
㪛㪆㪘㩷㪝㫌㫃㫃㩷㪪㪺㪸㫃㪼
䌗 㪌 㪋 䌃䌒 䌌䌆 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩
㪠㫅㫇㫌㫋㩷㪪㪼㫃㪼㪺㫋㫀㫆㫅
䌗 㪌 㪌 㪇 䌃䌒 䌌䌆 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩
㪦㫌㫋㫇㫌㫋㩷㪪㪼㫃㪼㪺㫋㫀㫆㫅
䌗 㪌 㪍 㪇 㪇 䌃䌒 䌌䌆 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩
㪘㫍㪼㫉㪸㪾㪼㩷㪮㪼㫀㪾㪿㫋
䌗 㪏 㪇 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
㪤㪸㫏㪅㩷㪭㪸㫃㫌㪼
䌗 㪏 㪈 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
㪤㫀㫅㪅㩷㪭㪸㫃㫌㪼
䌗 㪏 㪉 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
㪞㪼㫅㪼㫉㪸㫃㩷㪪㫋㪸㫅㪻㪸㫉㪻㩷㪛㪼㫍㫀㪸㫋㫀㫆㫅
䌗 㪏 㪊 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
㪪㪸㫄㫇㫃㪼㩷㪪㫋㪸㫅㪻㪸㫉㪻㩷㪛㪼㫍㫀㪸㫋㫀㫆㫅
䌗 㪏 㪋 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
㪤㪸㫏㪅㩷㪄㩷㪤㫀㫅㪅
䌗 㪏 㪌 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
㪘㪺㪺㫌㫄㫌㫃㪸㫋㫀㫆㫅㩷㪚㫆㫌㫅㫋
䌗 㪏 㪍 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
㪣㪸㫋㪼㫊㫋㩷㪘㪺㪺㫌㫄㫌㫃㪸㫋㫀㫆㫅㩷㪛㪸㫋㪸
䌗 㪏 㪎 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
136
10.INTERFACE
Approx.
Push TTL open collector
IC=6mA Toggle switch Relay Transistor
switch (ON when IN is HI)
IN
IN
COM
137
10.INTERFACE
Vcc +12V
F701-C Vext
Inside Outside
Vceo=30V(max)
Ic =120mA (max)
COM
0 OFF ON 0 H L
1 ON OFF 1 L H
digit data 8 4 2 1
0 0 0 0 0
1 0 0 0 1
2 0 0 1 0
3 0 0 1 1
4 0 1 0 0
5 0 1 0 1
6 0 1 1 0
7 0 1 1 1
8 1 0 0 0
9 1 0 0 1
138
10.INTERFACE
When weight values output as BCD data are on the following conditions, 1 (OVER) is output.
BCD data
BCD data is updated at each time of A/D conversion, and strobe pulses are output in
synchronization with the BCD data.
For reading data, use the pulse end edge.
BCD data
The data update rate is normally 200 times/sec., which is the same as the operating speed of the
main unit, but data may not be read with some connected external devices because the speed is too
fast. In such a case, set the data update rate longer (reduce the number of times).
139
10.INTERFACE
Operation
4
F → CNG/ENT → SP1 → CNG/ENT
7
OVER → CNG/ENT → → CNG/ENT
When this input terminal is short-circuited with the COM terminal, the BCD data is held.
(No strobe pulses are output.)
Open : negative
Short : positive
140
10.INTERFACE
The D/A converter is an interface to output measured weight values as electrical signals. Current in
proportion to the weight value (4mA ~ 20mA) can be output.
The over range is ±5% of the full scale.
Request
CAUTION
- Pick up the plug then handle the lever with
thumb.
- Do not handle the lever when the plug is attached
to the base otherwise it may cause some trouble.
141
10.INTERFACE
Connect an external equipment (load resistance 350Ω or less) between CUR.(+) ~ G(-) of F701-C.
F701-C
← Inside Outside →
SIG
CUR. (+) +
External equipment
G (-) -
Load resistance:
GND 350Ω or less
CAUTION
Do not apply voltage or current to the current output terminal externally by
mistake.
Doing so will cause breakage.
142
10.INTERFACE
10-4-2. Specifications
Resolution 1/10000
The D/A converter of the F701-C obtains analog output by setting the weight value to output 4mA
of current and the weight value to output 20mA of current, respectively. Input each setting value
and select D/A Output Mode in setting mode 4.
Current
(mA)
20
Resolution 1/10000
4
Weight value
143
10.INTERFACE
Operation
a)
1) Select setting mode 4.
4
F → CNG/ENT → SP1 → CNG/ENT
1
UPPER → CNG/ENT → → CNG/ENT
b)
1) Select setting mode 4.
4
F → CNG/ENT → SP1 → CNG/ENT
2) Input the D/A Zero Output Weight. (2-D/A Zero Output Weight)
2
LOWER → CNG/ENT → → CNG/ENT
c)
1) Select setting mode 4.
4
F → CNG/ENT → SP1 → CNG/ENT
3
NEAR Z. → CNG/ENT → → CNG/ENT
144
10.INTERFACE
Example
Current
(mA)
20
Resolution 1/10000
4 Net
500 2000 2500
480 3.84
Zero 500 4.00
1000 8.00
1500 12.00
Full scale 2500 20.00
2520 20.16
145
10.INTERFACE
Sequencer,
host computer, etc. F701-C
* Logic
Rt=terminal resistor
A A - Mark status (OFF)
SD RD VA - VB < -0.2V
B B
A A* - Space status (ON)
B B VA - VB > 0.2V
A A
RD SD
B B VA : RD A terminal voltage
Rt=terminal resistor
VB : RD B terminal voltage
(Master) (Slave)
- Use a twisted pair cable for connection. (The noise margin will increase.) However, a twin cable
is sufficient for short-distance connection.
- Couple a terminal resistor of 100 ~ 200Ω to the receiving end.
- The F701-C side has internal terminal resistor. By
turning ON the Rt switch, it is not necessary to
install external resistance.
- The terminal SG is a grand terminal used on the
circuit for protecting the circuit.When the main Turn ON the Rt switch.
body of F701-C and the device connected to F701-C
are grounded by D type ground, there is usually no need to use the terminal SG.
However, confirm the specifications of the devise connected before connecting the terminal SG,
when it is necessary to connect it according to the situation of the site.
Request
146
10.INTERFACE
SD RD SD RD SD RD SD RD
A B A B A B A B A B A B A B A B
Rt Rt Rt Rt
OFF OFF OFF ON
Rt
A 120ΩRt can be
SD
B installed or removed
according to the switch.
A
RD
B
Rt Twisted-para-cable
The switch is available on the option
sheet metal section at the back
of the main unit.
Operation
4
F → CNG/ENT → SP1 → CNG/ENT
4
SP1 → CNG/ENT → → CNG/ENT
Parity bit
Character length
2 : Even
1 : 8bit
1 : Odd
0 : 7bit
0 : None
2. The initial set for RS-485 port of connecting personal computer and
sequencer should be the same setting of F701-C.
147
10.INTERFACE
1. For connecting more than one F701-C, assign an ID number to each F701-C.
2. When a start command including an ID number is sent from the host, only one F701-C is brought
into transmittable condition, in which reading of weight data, reading and changing of setting
values, commands, etc., become valid.
* Since Tri-state control is performed according to the start command and end command, if only
the start command is sent to two or more F701-Cs, their outputs come into collision, so that
normal communications cannot be carried out.
4. ID number setting.
Operation
1) Select setting mode 4.
4
F → CNG/ENT → SP1 → CNG/ENT
5
SP2 → CNG/ENT → → CNG/ENT
ID Number
(0 ~ 9999)
* When the ID number is 0000, transmittable and receivable conditions are brought about at
power-on, therefore set ID numbers other than 0000 for connecting two or more F701-Cs.
* If the ID number setting is other than 0000, other formats (R・・, W・・, C・・, etc.) are invalid
until the following start command is accepted after power-on.
5. Start command.
Host ID number
S 1 2 3 4 CR
F701-C I D 1 2 3 4 CR LF
ID number
* Return only when corresponding
to the key-set ID number.
6. End command.
Host ID number
E 1 2 3 4 CR
F701-C I D 1 2 3 4 CR LF
ID number
* Return only when corresponding
to the key-set ID number.
148
10.INTERFACE
- Reading out the Gross weight (sign, 5-digt weight value, decimal point)
Host R A CR
F701-C R A + 1 0 0 . 0 0 CR LF
- Reading out the Net weight (sign, 5-digit weight value, decimal point)
Host R B CR
F701-C R B + 1 0 0 . 0 0 CR LF
- Reading out the Tare weight (sign, 5-digit weight value, decimal point)
Host R C CR
F701-C R C + 1 0 0 . 0 0 CR LF
Host R D CR
F701-C R D 0 0 0 0 0 0 0 CR LF
Weight value 1 : ON
hold 0 : OFF * Not defined
Weight value 1 : ON
Zero alarm 0 : OFF
1 : ON Weight 1 : Net
Tare subtracted 0 : OFF display 0 : Gross
149
10.INTERFACE
Host R E CR
F701-C R E 0 0 0 0 0 0 0 CR LF
Complte 1 : ON
SP1 1 : ON output signal 0 : OFF
output signal 0 : OFF
Over 1 : ON
output signal 0 : OFF
SP2 1 : ON
output signal 0 : OFF Go 1 : ON
output signal 0 : OFF
SP3 1 : ON
output signal 0 : OFF Under 1 : ON
output signal 0 : OFF
Host R F CR
F701-C R F 0 0 0 0 0 0 0 CR LF
* Not defined
Near zero 1 : ON
output signal 0 : OFF
Lower limit 1 : ON
output signal 0 : OFF
Upper limit 1 : ON
output signal 0 : OFF
150
10.INTERFACE
Host R G CR
F701-C R G 0 0 0 0 0 0 0 CR LF
Weight error 1 : ON
output signal 0 : OFF
* Not defined
Error 1 : ON
output signal 0 : OFF
Sequence error
0 or 1 ~ 5
3 : Span calibrating
Operation 2 : Zero calibrating * Same as the error
mode 1 : Weighing display of F701-C.
0:
Host R H CR
F701-C R H 9 9 9 9 9 CR LF
Host R I CR
F701-C R I 9 9 9 9 9 9 9 . 9 9 CR LF
Host R J CR
F701-C R J 0 0 + 1 0 0 . 0 0 CR LF
Host R J CR
F701-C R J CR LF
151
10.INTERFACE
Host W CR LF
F701-C W CR LF
Setting value No.
Setting value
* For setting value No., refer to "Setting value communication format" on P.154.
- Zero Calibration
Host C A CR
F701-C N A 0 CR LF
Calibrating
0: Normally completed, 1 ~ 9 : Error
(same as the error display of F701-C)
- Span Calibration
Host C B CR
F701-C N B 0 CR LF
Calibrating
0: Normally completed, 1 ~ 9 : Error
(same as the error display of F701-C)
Request
Before sending this command, set the Capacity, Min. Scale division and
Balance Weight Value.
152
10.INTERFACE
Host C C CR
Host C D CR
- Tare subtraction
Host C E CR
Host C F CR
- Digital zero
Host C G CR
Host C H CR
- Accumulation Command
Host C I CR
- Accumulation Clear
Host C J CR
153
10.INTERFACE
Use the following for reading and writing set values. For the communication formats, see p.152.
㻿㼑㼠㻌㻼㼛㼕㼚㼠㻌㻝
䠳 㻝 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻿㼑㼠㻌㻼㼛㼕㼚㼠㻌㻞
䠳 㻝 㻝 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻲㼕㼚㼍㼘
䠳 㻝 㻞 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻻㼢㼑㼞
䠳 㻝 㻟 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼁㼚㼐㼑㼞
䠳 㻝 㻠 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻯㼛㼙㼜㼑㼚㼟㼍㼠㼕㼛㼚
䠳 㻝 㻡 䠉 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻖㻌㻿㼕㼓㼚㻛㻌㻜㻦㻌㻼㼘㼡㼟㻌㻌䠉㻦㻌㻹㼕㼚㼡㼟
㻭㼡㼠㼛㻌㻲㼞㼑㼑㻌㻲㼍㼘㼘㻌㻯㼛㼙㼜㼑㼚㼟㼍㼠㼕㼛㼚㻌㻾㼑㼓㼡㼘㼍㼠㼕㼛㼚
䠳 㻝 㻢 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻭㼐㼖㼡㼟㼠㻌㻲㼑㼑㼐㼕㼚㼓㻌㼀㼕㼙㼑
䠳 㻝 㻣 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻶㼡㼐㼓㼕㼚㼓㻌㼀㼕㼙㼑
䠳 㻞 㻜 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻯㼛㼙㼜㼍㼞㼕㼟㼛㼚㻌㻵㼚㼔㼕㼎㼕㼠㻌㼀㼕㼙㼑
䠳 㻞 㻝 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼁㼜㼜㼑㼞㻌㻸㼕㼙㼕㼠
䠳 㻞 㻞 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻸㼛㼣㼑㼞㻌㻸㼕㼙㼕㼠
䠳 㻞 㻟 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻺㼑㼍㼞㻌㼆㼑㼞㼛
䠳 㻞 㻠 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼀㼍㼞㼑㻌㼃㼑㼕㼓㼔㼠
䠳 㻞 㻡 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻭㼡㼠㼛㻌㼆㼑㼞㼛㻌㼀㼕㼙㼑㼟
䠳 㻞 㻢 㻜 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻶㼡㼐㼓㼕㼚㼓㻌㼀㼕㼙㼑㼟
䠳 㻞 㻣 㻜 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻯㼛㼙㼜㼘㼑㼠㼑㻌㻻㼡㼠㼜㼡㼠㻌㼀㼕㼙㼑
䠳 㻞 㻤 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
154
10.INTERFACE
㻿㼑㼝㼡㼑㼚㼏㼑㻌㻹㼛㼐㼑
䠳 㻟 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼃㼑㼕㼓㼔㼕㼚㼓㻌㻲㼡㼚㼏㼠㼕㼛㼚㻌㻝
䠳 㻟 㻝 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼃㼑㼕㼓㼔㼕㼚㼓㻌㻲㼡㼚㼏㼠㼕㼛㼚㻌㻞
䠳 㻟 㻞 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼃㼑㼕㼓㼔㼕㼚㼓㻌㻲㼡㼚㼏㼠㼕㼛㼚㻌㻟
䠳 㻟 㻟 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻲㼡㼚㼏㼠㼕㼛㼚㻌㻷㼑㼥㻌㻵㼚㼢㼍㼘㼕㼐
䠳 㻟 㻠 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻭㼚㼍㼘㼛㼓㻌㻲㼕㼘㼠㼑㼞
䠳 㻟 㻡 㻜 㻜 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻰㼕㼓㼕㼠㼍㼘㻌㻲㼕㼘㼠㼑㼞
䠳 㻟 㻢 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻹㼛㼠㼕㼛㼚㻌㻰㼑㼠㼑㼏㼠㼕㼛㼚
䠳 㻟 㻣 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼆㼑㼞㼛㻌㼀㼞㼍㼏㼗㼕㼚㼓㻌㻼㼑㼞㼕㼛㼐
䠳 㻟 㻤 㻜 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㼆㼑㼞㼛㻌㼀㼞㼍㼏㼗㼕㼚㼓㻌㻾㼍㼚㼓㼑
䠳 㻟 㻥 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻿㼑㼠㼠㼕㼚㼓㻌㼂㼍㼘㼡㼑㻌㻸㼛㼏㼗
䠳 㻟 㻭 㻜 㻜 㻜 䠟䠮 䠨䠢
㻮㼍㼘㼍㼚㼏㼑㻌㼃㼑㼕㼓㼔㼠㻌㼂㼍㼘㼡㼑
䠳 㻠 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻯㼍㼜㼍㼏㼕㼠㼥
䠳 㻠 㻝 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻹㼕㼚㻚㻌㻿㼏㼍㼘㼑㻌㻰㼕㼢㼕㼟㼕㼛㼚
䠳 㻠 㻞 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻺㼑㼠㻌㻻㼢㼑㼞
䠳 㻠 㻟 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻳㼞㼛㼟㼟㻌㻻㼢㼑㼞
䠳 㻠 㻠 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻲㼡㼚㼏㼠㼕㼛㼚㻌㻿㼑㼘㼑㼏㼠㼕㼛㼚
䠳 㻠 㻡 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻳㼞㼍㼢㼕㼠㼍㼠㼕㼛㼚㼍㼘㻌㻭㼏㼏㼑㼘㼑㼞㼍㼠㼕㼛㼚㻌㻔㼍㼞㼑㼍㻌㼚㼡㼙㼎㼑㼞㻌㼕㼚㼜㼡㼠㻕
䠳 㻠 㻢 㻜 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻰㼆㻌㻾㼑㼓㼡㼘㼍㼠㼕㼛㼚㻌㼂㼍㼘㼡㼑
䠳 㻠 㻣 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
㻳㼞㼍㼢㼕㼠㼍㼠㼕㼛㼚㼍㼘㻌㻭㼏㼏㼑㼘㼑㼞㼍㼠㼕㼛㼚㻌㻔㼍㼏㼏㼑㼘㼑㼞㼍㼠㼕㼛㼚㻌㼕㼚㼜㼡㼠㻕
䠳 㻠 㻤 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅
155
10.INTERFACE
㪜㫏㫋㪼㫉㫅㪸㫃㩷㪝㫌㫅㪺㫋㫀㫆㫅㩷㪪㪼㫃㪼㪺㫋㫀㫆㫅
䌗 㪌 㪇 䌃䌒 䌌䌆 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩
㪩㪼㫊㫋㫉㫀㪺㫋㫀㫆㫅㩷㫆㫅㩷㫋㪿㪼㩷㪫㪸㫉㪼㩷㪪㫌㪹㫋㫉㪸㪺㫋㫀㫆㫅㩷㪝㫌㫅㪺㫋㫀㫆㫅
䌗 㪌 㪈 㪇 䌃䌒 䌌䌆 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩
㪛㪆㪘㩷㪦㫌㫋㫇㫌㫋㩷㪤㫆㪻㪼
䌗 㪌 㪉 㪇 㪇 㪇 䌃䌒 䌌䌆 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩
㪛㪆㪘㩷㪱㪼㫉㫆㩷㪦㫌㫋㫇㫌㫋㩷㪮㪼㫀㪾㪿㫋
䌗 㪌 㪊 䌃䌒 䌌䌆 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩
㪛㪆㪘㩷㪝㫌㫃㫃㩷㪪㪺㪸㫃㪼
䌗 㪌 㪋 䌃䌒 䌌䌆 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩
㪠㫅㫇㫌㫋㩷㪪㪼㫃㪼㪺㫋㫀㫆㫅
䌗 㪌 㪌 㪇 䌃䌒 䌌䌆 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩
㪦㫌㫋㫇㫌㫋㩷㪪㪼㫃㪼㪺㫋㫀㫆㫅
䌗 㪌 㪍 㪇 㪇 䌃䌒 䌌䌆 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩
㪘㫍㪼㫉㪸㪾㪼㩷㪮㪼㫀㪾㪿㫋
䌗 㪏 㪇 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
㪤㪸㫏㪅㩷㪭㪸㫃㫌㪼
䌗 㪏 㪈 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
㪤㫀㫅㪅㩷㪭㪸㫃㫌㪼
䌗 㪏 㪉 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
㪞㪼㫅㪼㫉㪸㫃㩷㪪㫋㪸㫅㪻㪸㫉㪻㩷㪛㪼㫍㫀㪸㫋㫀㫆㫅
䌗 㪏 㪊 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
㪪㪸㫄㫇㫃㪼㩷㪪㫋㪸㫅㪻㪸㫉㪻㩷㪛㪼㫍㫀㪸㫋㫀㫆㫅
䌗 㪏 㪋 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
㪤㪸㫏㪅㩷㪄㩷㪤㫀㫅㪅
䌗 㪏 㪌 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
㪘㪺㪺㫌㫄㫌㫃㪸㫋㫀㫆㫅㩷㪚㫆㫌㫅㫋
䌗 㪏 㪍 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
㪣㪸㫋㪼㫊㫋㩷㪘㪺㪺㫌㫄㫌㫃㪸㫋㫀㫆㫅㩷㪛㪸㫋㪸
䌗 㪏 㪎 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀
156
11.OVER SCALE & ERROR
157
11.OVER SCALE & ERROR
The load cell output dose not reach the span adjustment
range.
The load cell output is miunus.
158
12.TROUBLE SHOOTING
An input signal from the load cell exceeds F701-C span adjustment range.
Check whether the output of the load cell is within span adjutment range;
Check there are breakages in the cable which is connected to the F701-C and the load cell;
And whether the load cell connector on the rear panel is open or not, when the load cell input termi-
nal block is open (not connected with the load cell) will display “ ” also.
An input signal from the load cell is lower than F701-C span adjustment range.
Check whether the output of the load cell for falling short of the span adjustment range;
Check there are breakages in the cable which is connected to the F701-C and the load cell;
And whether the load cell connector on the rear panel is open or not, when the load cell input termi-
nal block is open (not connected with the load cell) will display “ ” also.
Request
Scale a Capacity is primary data for the F701-C. You must re-calibrate if
Capacity is changed .
Do not change Capacity to reset over scale message ( ).
159
12.TROUBLE SHOOTING
● Error display
(Sequence error)
(Sequence error)
(Sequence error)
It is displayed if a ZALM condition is brought about when the auto zero is in operation by sequence
control.
Remove the causes of zero out of adjustment (adherents, etc.), and reset the digital zero to resolve
the zero alarm.
(Sequence error)
When the start of weighing, Near Zero output signal turns OFF ("At start, Near Zero confirmation
ON" is selected for sequence control).
First, check the Near Zero setting value and the Near Zero comparison target. Next, check the
following for occurrence:
“A start was made before complete discharge.”
“The start ON timing was too early.”
“Discharge is in a jam, etc.”
Set the stop signal from OFF to ON to resolve the sequence error.
(Sequence error)
It is displayed if the Set Point 1 output (SP1) signal is ON when weighing starts (when it is set to
check the weight value when weighing starts).
First, check the Set Point 1 setting value and Final setting value. Next, check the following for
occurrence:
“A start was made before complete discharge.”
“The start ON timing was too early.”
“Discharge is in a jam, etc.”
Set the stop signal from OFF to ON to resolve the sequence error.
Or, it is displayed when it is (Final - Compensation) < 0 at the start of weighing.
160
12.TROUBLE SHOOTING
(Calibration error)
(Calibration error)
Resistor Strain
Ideal Approx. μ-STRAIN mV/V
875 KΩ 866 KΩ 200 0.1
437 KΩ 442 KΩ 400 0.2
291 KΩ 294 KΩ 600 0.3
219 KΩ 221 KΩ 800 0.4
175 KΩ 174 KΩ 1000 0.5
146 KΩ 147 KΩ 1200 0.6
125 KΩ 124 KΩ 1400 0.7
109 KΩ 110 KΩ 1600 0.8
97 KΩ 97.6 KΩ 1800 0.9
87.3 KΩ 86.6 KΩ 2000 1.0
79.4 KΩ 78.7 KΩ 2200 1.1
72.7 KΩ 73.2 KΩ 2400 1.2
67.1 KΩ 66.5 KΩ 2600 1.3
62.3 KΩ 61.9 KΩ 2800 1.4
58.2 KΩ 57.6 KΩ 3000 1.5
54.5 KΩ 54.9 KΩ 3200 1.6
51.3 KΩ 51.1 KΩ 3400 1.7
48.4 KΩ 48.7 KΩ 3600 1.8
45.9 KΩ 46.4 KΩ 3800 1.9
43.6 KΩ 43.2 KΩ 4000 2.0
41.5 KΩ 41.2 KΩ 4200 2.1
39.6 KΩ 39.2 KΩ 4400 2.2
37.9 KΩ 38.3 KΩ 4600 2.3
36.3 KΩ 36.5 KΩ 4800 2.4
34.8 KΩ 34.8 KΩ 5000 2.5
161
12.TROUBLE SHOOTING
(Calibration error)
Inditial dead load is negative.
Check that load cell is mounted in the correct direction; check that load is being applied to the load
cell in the correct direction ; check that the +SIG and -SIG lines are propely connected.
If still displayed, connect a resistor between -EXC and -SIG load cell connections. This
should shift the zero point. Do Zero Calibration again.
+EXC
Connect the resistor
between- EXC and -SIG.
+SIG
-EXC
-SIG
(Calibration error)
The Balance Weight Value is larger than the Capacity.
Re-enter the Balance Weight Value equal to or less than the Capacity.
Do Span Calibration again.
Load
+9 scale division
Capacity
Balance Weight Value
50%
Balance Weight Value must be between 50% to 100% of Capacity in order to do Span Calibration
correctly.
(Calibration error)
The set Balance Weight Value is “00000”. Set adequate value to Balance Weight Value.
(Calibration error)
The load cell output dose not reach the span adjustment range of the F701-C.
Check how load is applied to load cell; check load cell has sufficient outpul (mV/V) to reach span
range. Do Span Calibration again.
Or, when an equivalent calibration is done by resolution that exceeds 1/10000, it is assumed the one
not reaching within the range of the span adjustment.
162
12.TROUBLE SHOOTING
(Calibration error)
(Calibration error)
(Calibration error)
Adjust the stable period and stable range of Motion Detection so “ STAB ” display lights during
Calibration. Do Calibration again.
Load
Capacity
163
12.TROUBLE SHOOTING
(EXC Error)
The voltage of the lithium battery for memory backup has dropped.
Replace with a new battery.
For the method of replacement, refer to "13.REPLACEMENT OF THE BACKUP BATTERY" on
P.166.
There is a possibility that writing to the NOV RAM is interrupted, so that the setting value is not
changed.
If any setting value stored in the NOV RAM (refer to "18.THE LIST OF INITIAL SETTING
VALUE" on P.176) is changed, the “・” (position shown below) lights (may not light in some cases).
While “・” is lit, do not turn off the power of the F701-C.
● Setting values have been changed when the power is turned on.
The voltage of the lithium battery for memory backup may have dropped.
The setting values stored in the SRAM (refer to "18.THE LIST OF INITIAL SETTING VALUE"
on P.176) are backed up by the lithium battery.
If the power of the F701-C is turned off with the voltage of the lithium battery having dropped, the
setting values are lost.
Replace the lithium battery, and change the setting values again.
164
12.TROUBLE SHOOTING
(Checksum error)
165
13.REPLACEMENT OF THE BACKUP BATTERY
2. Remove the two M3 screws at the top of the rear panel, and pull out the upper cover by sliding it.
4. Remove the screw fixing the nylon clamp, and replace the lithium battery with a new one.
Nylon clamp
Connector
Screw * The connector of the battery can
be removed with a strong pull.
Lithium battery
5. After connection of the connector of the new lithium battery, make a cold start.
Request
The setting values in the “SRAM” (refer to "18.THE LIST OF INITIAL SETTING
VALUE" on P.176 are lost when the battery is removed. Record the setting
values beforehand.
166
LOCK
SW
BUZZER
RESET
CPU CPU
16bit WD Monitoring
function
-SIG
SHIELD RS-232C
FRAME
24 OUT
BCD
DISPLAY
DATA OUTPUT
8 IN
For fluorescent
VCC character display tube ISOLATION
+9V -9V +5V AC4V +13V +5V +12V D/A
KEY INPUT 4 ~ 20mA
CONVERTER
CC-Link
FRAME
167
15.DIMENSIONS
15. DIMENSIONS
Unit:mm
168
16.MOUNTING ON A PANEL
138 Unit : mm
Panel cutout size
138W×68H (mm)
(2) Remove the mounting rails on both sides of the controller, and insert the controller into the
panel.
(3) Insert the mounting rails into both sides from the back of the controller.
(4) Securely fix the mounting hardware on both sides with the attached M4 screws.
CAUTION
For transportation after panel-mounting, consideration should be given so as not to give
excessive shocks or vibration.
169
17.SPECIFICATIONS
17. SPECIFICATIONS
17-1. Analog Section
170
17.SPECIFICATIONS
Center zero A true zero point or the center of each value is displayed.
Status display SP1, SP2, SP3, NEAR Z, HOLD, ZALM, STAB, TARE, NET,
ZERO,
Protection of set values Setting operation is inhibited so that initial setting values and
calibration will not be changed by wrong operation.
Setting item - Upper Limit / Lower Limit / Near Zero / Set Point 1 / Set Point 2 /
Compensation / Over / Under / Final
- Comparison Inhibit Time / Judging Time / Complete Output Time /
Adjust Feeding Time / Auto Zero Times / Judging Times /
Auto Free Fall Compensation Regulation / Analog Filter / Tare Weight
171
17.SPECIFICATIONS
Input signals G・N / Digital Zero / Tare ON / Tare OFF / Hold / Judge / Feed・Discharge /
(8 points) Start / Stop / Accumulation Command / Accumulation Clear
ON when short-circuited with the COM terminal by contact
(relay, switch, etc.) or non-contact (transistor, open collector output, etc.)
Output signals Near Zero / SP1 / SP2 / SP3 / Under / Over / Upper Limit / Lower Limit /
(12 points) Stab / Weight Error・Error / Go・Complete / RUN・Accumulation Error
Transistor open collector output
(Emitter = COM terminal)
Output LO with transistor ON.
17-5. Interface
< Standard equipment >
172
17.SPECIFICATIONS
- D/A converter
Current output : 4 ~ 20mA
D/A conversion speed : 200 times/sec.
Resolution : 1/10000
Over range : ±5% of the full scale
Current 3.2mA ~ 20.8mA
Zero drift : Within 0.3μV/ ℃
Gain drift : Within 30ppm/ ℃
Non-linearity : Within 0.02%/FS
* Not including the drift of the analog input section
173
17.SPECIFICATIONS
+2 +1
Panel cutout size 138W( -0 )×68H
( -0 )[mm]
174
17.SPECIFICATIONS
17-7. Attachment
AC input cord (3m)........................................................................1
Minus screw driver (For WAGO terminal)....................................1
Control signal Input/Output connector ..........................................1
Connector for BCD output (with BCD option) .............................1
Connector for D/A converter (with D/A converter option) ...........1
Connector for RS-485 (with the RS-485 option)...........................1
Connector for CC-Link (with the CC-Link option).......................1
Operation manual...........................................................................1
Jumper line.....................................................................................2
- The power cable attached to this product as standard equipment can be used in the
AC100V power supply in Japan. (Official ratings voltage AC125V)
Please use the power cable authorized in the country when you use this product
outside Japan.
- Our company sells following resistance pressure cable AC250V (European standard
product) separately.
Please purchase it from us when you need after confirming its plug shape/voltage.
175
18.THE LIST OF INITIAL SETTING VALUE
0
F → CNG/ENT → → CNG/ENT
1
F → CNG/ENT → UPPER → CNG/ENT
176
18.THE LIST OF INITIAL SETTING VALUE
2
F → CNG/ENT → LOWER → CNG/ENT
3
F → CNG/ENT → NEAR Z. → CNG/ENT
4
F → CNG/ENT → SP1 → CNG/ENT
*D/A Zero Output Weight and D/A Full Scale cannot be set larger value than the value of Capacity.
The value of Capacity automatically reaches a set value.
177
18.THE LIST OF INITIAL SETTING VALUE
5
F → CNG/ENT → SP2 → CNG/ENT
8
F → CNG/ENT → UNDER → CNG/ENT
9
F → CNG/ENT → FINAL → CNG/ENT
178