Service CS-2000 En
Service CS-2000 En
Translation
© Eltra GmbH, 42781 Haan, Retsch-Allee 1-5, Germany | 08.08.2016 Version 0001
Copyright
© Copyright by
Eltra GmbH
Retsch-Allee 1-5
D-42781 Haan
Germany
2
Table of Contents
1 Notes on service instructions 7
1.1 Explanations of the Safety Instructions 7
2 Discharging the oscillator 8
3 Faults 12
3.1 Check base lines 12
3.2 Fluctuating gas flow 13
3.3 No combustion 18
3.3.1 Induction furnace operation 18
3.3.2 Resistance furnace operation 27
3.4 Bad combustion 27
3.4.1 Induction furnace operation 27
3.4.2 Resistance furnace operation 28
3.5 Furnace does not close 30
3.6 Analysis takes too long 32
3.6.1 RF-interference on the IR-cell, or on the electronic units 33
3.6.2 Additional reasons for excessive analysis duration 34
3.6.3 One of the IR-sources is unstable 34
3.6.4 The power voltage sinks below 180 VAC 34
3.7 Development of dust when closing the furnace 35
3.8 Erratic results 35
3.9 The results are always 0.0000% or 00.0ppm: 37
3.10 No oxygen flow 38
3.10.1 Induction furnace operation 38
3.10.2 Other faults 39
3.10.3 Resistance furnace operation 39
3.11 IR-cell temperature control is out of range 39
3.12 Breaker turns off at the beginning of combustion 39
3.13 Breaker turns off at the end of the analysis 43
3.14 Dustcloud when opening the furnace 44
3.15 No or low oxygen pressure 45
3.16 Waiting for stability message 47
3.17 Combustion tube breaks frequently 48
3.18 Oxygen pressure falls when the furnace opens 49
3.19 Combustion without starting analysis 50
3.20 Crucible or boat cracks or melts 52
3.20.1 Induction furnace 52
3.20.2 Resitance furnace 52
3.21 Oxygen pressure goes up very slowly 52
3.22 Pump voltage too high 54
3.23 Resistance furnace takes too long to warm up 58
3.24 Negative drift after start 59
3.25 Resistance furnace completly cold or too hot 60
3.26 Fumes in the resistance furnace 61
3.27 Resistance furnace doesn`t reach full temperature 62
3.28 Unstable resistance furnace temperatur 62
3.29 Dust or soot from the resistance furnace 63
3
3.30 Vibrations in induction furnace mode 64
3.31 Analysis cycle stops too early 64
3.32 Chopper motor doesn‘t rotate 65
3.33 Sample burns but no peak on the screen 65
3.34 IR signal does not come down to the base line at the end of analysis 66
4 Adjustments 70
4.1 Gas flow controller-adjustment and jumper settings 70
4.1.1 Flow controller for induction furnace operation 70
4.1.2 Flow controller for resistance furnace operation 73
4.2 Infrared base line adjustments 75
4.2.1 Infrared cell modification - general information 75
4.2.2 General test points 78
4.2.3 Test points IR-ranges 79
4.2.4 Base line adjustments. 80
4.2.5 Infrared cell board IRC1.x sections assignment 80
4.2.6 Voltage of a base line is too high (higher than 9.5V) 81
4.3 IR source voltage setting 83
4.4 Infrared cell temperature regulation 84
4.5 Pneumatics 86
4.6 Pressure regulation 90
4.7 Closing cone adjustment 92
4.8 Linearity correction 93
4.9 Induction generator control 94
4.10 Resistance furnace 96
4.10.1 Setting the furnace regulation card TH 43 / TH 44 96
4.10.2 Furnace regulation board TH 43 / TH 44 97
4.11 Thermocouple testing 98
5 Service 99
5.1 IR-paths, cleaning and replacing 99
5.2 Infrared electronics 100
5.2.1 Removing the IRC-board 100
5.2.2 Changing the IR-source 101
5.3 Chopper 102
5.4 Leak checking 104
5.5 Solenoid valves cleaning 108
5.6 Flow sensor replacement 110
5.7 Gas pump replacement 111
5.8 Thermocouple replacement 112
5.9 Gas saving mode 114
5.10 Gas saving also for resistance furnace 115
5.11 Furnace pneumatics - additional safety features (optional) 116
5.12 Halogen trap - installing 117
6 Miscellaneous 118
6.1 Ordering numbers 118
6.1.1 Front side 118
6.1.2 Left hand side 120
6.1.3 Right hand side 122
4
6.1.4 Induction furnace 124
6.1.5 Furnace cleaning 126
6.1.6 Oscillating circuit 127
6.1.7 Infrared cell 129
6.1.8 Chopper 131
6.1.9 Resistance furnace 132
6.1.10 Oxygen purifying furnace 134
6.1.11 Dust trap 135
6.2 Wiring diagrams 137
6.2.1 Right hand side power wiring 137
6.2.2 Left hand side power wiring 138
6.2.3 Valve wiring 139
6.2.4 Control signal wiring 140
6.2.5 IR-current supply, old version 141
6.2.6 IR-cell internal signals 142
6.2.7 Amplitude circle 143
6.3 Gas flow diagrams 144
6.3.1 Gas flow for induction furnace when furnace is closed 144
6.3.2 Gas flow for induction furnace when furnace is open 145
6.3.3 Gas flow for resistance furnace 146
6.3.4 Valve contolling 147
6.4 Spare parts kits 148
6.4.1 Common spare parts kti for all analysers 148
6.4.2 Spare Parts kit for CS2000 (including spare parts kits CS800 and HTF540) 150
6.5 Maintenance 152
7 Approved methodologies to which Eltra instruments conform 153
7.1 Inorganic materials (Metals) 153
7.2 Organic materials (Oil, Coal, foodstuffs) 154
8 Disposal 155
9 Index 156
5
Notes on service instructions
6
Notes on service instructions
Pos : 1.1 /0025 Ü bersc hriften/1. Ü bers chriften/1 Hi nweis e z ur Ser vic e Anl eitung @ 7\mod_1379492578032_9.doc x @ 48845 @ 1 @ 1
Mortal injury may result from failing to heed these safety warnings. We give you the following
warnings and corresponding content.
DANGER
We also use the following signal word box in the text or in the instructions on action to be taken:
DANGER
Serious injury may result from failing to heed these safety warnings. We give you the following
warnings and corresponding content.
WARNING
Type of danger / personal injury
Source of danger
– Possible consequences if the dangers are not observed.
• Instructions on how the dangers are to be avoided.
We also use the following signal word box in the text or in the instructions on action to be taken:
WARNING
Moderate or mild injury may result from failing to heed these safety warnings. We give you
the following warnings and corresponding content.
CAUTION
Type of danger / personal injury
Source of danger
– Possible consequences if the dangers are not observed.
• Instructions on how the dangers are to be avoided.
We also use the following signal word box in the text or in the instructions on action to be taken:
CAUTION
In the event of possible property damage we inform you with the word “Instructions” and the
corresponding content.
7
Discharging the oscillator
NOTICE
We also use the following signal word in the text or in the instructions on action to be taken:
NOTICE
In order to carry out work on the oscillator of the CS800/2000, it is necessary to ensure that the
oscillator has no stored energy. This is achieved by short circuiting the various potentials.
Attention:
This activity may only be carried out by a qualified electrician with sound knowledge of the
CS800/2000 devices.
Only the Eltra discharge set (Order Number 11001-2005) may be used for discharging. The
discharge set must always be inspected for visible faults before use.
1000V insulating gloves (order number 8840-0453) must be worn during the discharging
process to prevent an electric shock.
The protective measure is only effective if the discharge set and the insulating gloves are
always worn together.
8
Discharging the oscillator
Implementation:
Disconnect the mains plug.
Use a multimeter to check that the device is
voltage-free
9
Discharging the oscillator
Attention!
The oscillator cover must be removed carefully to
expose the transformer connectors
10
Discharging the oscillator
Attention!
The measuring tip is now held first on the left-hand
and then on the right-hand 4.4 KV connector of the
transformer for approx. 5s
Attention!
All marked components inside the oscillator are short circuited to ground using the measuring tip.
11
Faults
Pos : 3.1 /0025 Ü bersc hriften/ELTR A 1. Ü berschriften/1 F aults Eltra @ 8\mod_1396880810332_9.doc x @ 60058 @ 1 @ 1
3 Faults
Pos : 3.2 /0025 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 C hec k base lines ELTRA @ 6\mod_1368452266990_9.doc x @ 40029 @ 2 @ 1
When this message appears on the PC screen, at least one of the base lines is either over
+9.5Volts or below -9.5 Volts.
In order to read the base line voltages of all IR cells available, click on “base lines”
If any of the “Inputs” or “Averages” is over 9.5V, adjust a voltage of about 6 to 7V.
See chapter 3.2 “Infrared base line adjustment”.
The adjustment is preferably made when having carrier gas flow through the cells.
12
Faults
Notice
The IR cells have normally base lines adjusted to about 8V and the peak comes down during
analysis, respectively during IR absorption by the measuring gas.
In case of golden paths, in the first couple of months after being in operation, the base line may
goes up, which means that the paths become more reflective during this period. Therefore, the
base lines of golden paths should be initially set rather at 5 or 6 volts instead of 8 volts. In the
first couple of months the base lines go up, but the growing slows down and growing
contamination finally causes the base line to start coming down.
The initial base line adjustment at 5 to 6 volts, prevents the maximum base line value from
reaching 9.5V, where the error message “Check base line” appears. Note that this message is
just a warning. The analyzer is not prevented from carrying out further analyses. As long as the
base line remains below 10V, the results are correct.
If a base line is below -9.5V, an adjustment should be done as advised above. If this is not
possible, then most probably there is a fault or a very strong contamination. See chapter 4.1 “IR
paths cleaning”. In case of contamination, all cells available should have very low voltages,
even negative ones.
If all available cells are below -9.5V there is most probably a fault that is common for all cells
like no rotating chopper motor or a faulty infrared source. See chapter 2.35 “The chopper
motor doesn’t rotate” and 3.3 “IR source voltage setting”.
See also Chapter 3.4 “Analysis takes too long / RF-interference on the IR-cell, or on the
electronic units”
Pos : 3.4 /0025 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 Fluc tuati ng gas flow ELTRA @ 6\mod_1368452950593_9.doc x @ 40051 @ 2 @ 1
The flow rate is electronically controlled; therefore there should be no significant variation from
the set value.
13
Faults
In case of a fault however, a control of the flow rate is no longer possible and the gas flow
fluctuates. This can be caused, for example, by an excessive obstruction of the gas flow system
or by a faulty component that needs to be replaced.
See the next pages of this chapter and chapter 3.1 “Gas flow controller adjustment and
jumper settings”
Mostly, the problem lies in the saturation of the dust trap or of the chemicals.
In case the flow becomes lower and the regulating valve or the pump voltage goes higher
during combustion, it is obvious that the flow regulator is working. Therefore the problem must
be a restricted oxygen supply. When the accelerators start burning they take a lot of oxygen, so
that the pressure in the furnace goes down, and as a result for this the flow may also go down
14
Faults
Check by removing the filter with the dust filter housing. See chapter: “High carbon results are
erratic”.
Don’t remove the filter from its housing.
Turn the filter housing upside down and try to blow air through the smaller whole (about 16mm
diameter) of the filter housing.
If you feel a remarkable restriction, replace the filter by a new one. With a new filter the air
should pass practically free without any restriction.
Blocked chemicals:
The grains of magnesium perchlorate as well as of sodium hydroxide have to move when
knocking on their glass tube. If they are cloggy, they are completely depleted and they have to
be replaced. When the glass tube is turned upside down, the grains should fall off. If a screw
driver is needed to stab the grains free, the chemicals had to be replaced much earlier.
Blocked paper filters:
The paper filters are small so that they are frequently forgotten or neglected, although they can
block the flow when they are not replaced for long.
Blocked lance hole over the crucible:
The hole in the ceramic heat shield over the crucible in the induction furnace can be restricted
or blocked.
Clean by using the brass brush as described in the operation manual. If this may not be enough
then clean the hole by using a piece of wire or a pipe cleaner.
Blocked furnace outlet:
For induction furnace, unscrew the cap-nut at the furnace outlet and remove the plastic tube.
Pass through the fitting a pipe cleaner. Open the furnace in order to be sure that the pipe
cleaner has passed thru the fitting reaching the furnace area.
Completely remove the plastic tube by unscrewing the left hand cap-nut on the dust filter
assembly. Clean this fitting the same way like before. If necessary clean the plastic tube as well.
In case of resistance furnace operation clean the furnace outlet as shown below.
15
Faults
Push the tube (1) through the upper glass tube assembly (2), pass it through the tube (3) up to
the inside of the dust trap box (4).
This should be done when the glass tube has to be removed anyway, for replacing the
magnesium perchlorate (anhydrone). See also operation manual, chapter “Dust cleaning”.
Too low oxygen pressure:
Check the oxygen gauge (6). The reading should be 1.5 bar (22.5psi). If the pressure is too low,
it can be adjusted by pulling and turning clockwise the knob of the pressure regulator on the
rear panel of the analyzer. It is placed on the left hand side and the knob is accessible from
inside the analyzer. An adjustment is possible provided that there is sufficient pressure at the
input of this pressure regulator.
Outlet tap of the oxygen bottle is not properly open:
It is important to properly open the small tap at the outlet of the pressure regulator of the oxygen
bottle. If this outlet tap of the regulator is not properly open, you may have enough pressure at
the analyzer when there is no or little oxygen demand. But when the oxygen demand grows, for
example at the beginning of combustion in the
induction furnace when the sample needs a lot of oxygen to burn, the pressure may drop to a
very low level if not enough oxygen can be supplied to the analyzer at the moment of high
oxygen demand. The same problem will appear during furnace purging before or after the
analysis, depending on the actual settings in the software.
Contaminated or corroded solenoid valve:
In case of exceptional materials like ores, solenoid valves can be contaminated up to the extent
of failing. This depends mainly on the amount of sulfur and moisture in the sample. The most
affected valves are the outflow valve V1 and the bypass valve V4 in as well as the valves
switching from Induction to resistance furnace V11 and V12. See chapter “Gas flow system”.
The resistance furnace operation can only contaminate V11 and V12. The valves can be
cleaned.
Faulty oxygen regulator:
A faulty oxygen regulator can of course cause a similar problem when it does not keep the
pressure level at moments of higher oxygen consumption.
Bent plastic tube of oxygen supply:
When the plastic tube supplying oxygen from the bottle to the analyzer is bent, there will be the
same problem like described in the above two subjects. There will be too low oxygen pressure
in the analyzer, especially at moments of high oxygen consumption, like when burning samples
or purging the furnace.
Oxygen supply is interchanged with the compressed air supply:
In case of reported bad combustion, check for correct connection of oxygen and compressed air
supply. Compressed air in the induction furnace will provide combustion but a bad one, because
of only 20% of oxygen existing in the air. Therefore the peaks are low and long, consisting of
more than one peak, looking like mountains.
CAUTION Oxygen to the compressed air inlet can lead to the explosion of the pneumatic
cylinder.
Blocked dust filter cartridge:
The dust filter cartridge is on the front panel so that contamination is visible for the operator. In
the gas flow circuit, this dust cartridge is connected after the metal filter and its following glass
tube with magnesium perchlorate, so that the carrier gas passing through the cartridge is
already filtered.
16
Faults
Blocked fittings:
This case is very unlikely to happen in case of induction furnace operation, because the
analyzed materials in question are inorganic materials like metals and ceramics, so that they
don’t create fumes, volatiles and moisture. Even cement analysis is not as critical as organics
analysis. In case of analysis of materials causing major contamination, mainly when analyzing
organic materials in the resistance furnace, the fittings of the IR cells can be blocked so that
they have to be cleaned. See chapter “IR paths cleaning and replacing”.
Also fittings elsewhere can be blocked, like the fittings for plastic tubes on the reagent glass-
tube holders.
Blocked flow sensor assembly:
If the flow sensor is blocked, it should normally be replaced.
For those who may decide to clean them - please read the description about cleaning. Chapter
“Flow sensor replacing / cleaning”.
Electronic reasons for flow problems:
The flow controller consists of 3 components.
Induction furnace operation
1. The flow sensor
2. The board HF-42 (The lower half of the board only).
3. The proportional valve V6 (see chapter “Gas flow system”).
4. Interferences from the induction generator. See chapter “Analysis takes too long / RF-
interference on the IR-cell, or on the electronic units”.
The electronic devices involved in the flow regulation are the board HF-42 and the flow sensor.
The only influence of the PC and of the microcontroller on the flow, is the start/stop the flow, for
example to start it when oxygen pressure available, to stop it in case of gas saving mode etc.
For chasing flow problems, there is a possibility to avoid any involvement of the PC and of the
microcontroller, by setting a jumper on the HF-42. To do this, move the jumper of J3 from A1-A2
to B3-B4. See chapter “Gas flow controller and jumper settings”.
DANGER G0002
17
Faults
Fig. 5: HF-42
3.3 No combustion
Pos : 3.7 /0010 ELTR A/0015 s er vice_i nstr uctions /CS- 2000_s er vic e/05F aul ts_C S-2000/0515 M odul No combustion @ 9\mod_1425539352573_9.doc x @ 74117 @ 33333334444444 @ 1
18
Faults
No flow
If this message permanently remains in the “status window”, check the flow on the lower flow
meter (Rotameter). The ball has to be in the range of 180 L/h. If it is in this range, then click on
the “device state” window icon.
19
Faults
1. Measurements on TP9
– a signal of about 0.5 V should turn the generator on,
– a signal of about 12 V should turn the generator off.
If 12 V are measured, despite "Analyzing" appearing on the display, then the microcontroller
board is defective or the connection from the HF 42 circuit board, pin 7 is loose.
20
Faults
frequent problem. Tighten the screws of the plug properly, to ensure a proper connection to the
board.
DANGER
Mortal danger from electric shock
Exposed power contacts - High Voltage
Ensure that the power switch is at position 3.
Check if the filament of the generator tube glows:
This can be easily checked:
Through the air vent holes of the furnace, the reflected light of the glowing filaments (heater of
the cathode) should be visible.
DANGER
Mortal danger from electric shock
Exposed power contacts - High Voltage
Before touching any components or wiring, make sure the capacitors are discharged.
21
Faults
10 VAC
220 VAC
(between contact 2
and 3)
22
Faults
DANGER
Mortal danger from electric shock
Exposed power contacts - High Voltage
When checking for induction power, always keep the furnace closed!
On the primary side of the high voltage transformer, you should be able to measure about 210 V
AC; if not, the triac is possibly defective.
The filament voltage can be also measured as shown below. A direct measurement at the
tube’s socket rules out eventual interruptions on the way of the voltage from the transformer to
the socket.
23
Faults
DANGER
Mortal danger from electric shock
Exposed power contacts - High Voltage
Never try to measure the voltage here!
Due to high voltage in the range of 7000 Volts (amplitude), there should be no measurement on
the rectifier under power. If no spare rectifier available to exchange, then test by using a scope
according to the following drawing.
220 VAC
Insulating
transformer
220 VAC
Adjustable
transformer
Rectifier
about
4000 Ω
24
Faults
10
11
6
1
2
25
Faults
The filament lead (socket) connected to the chassis, is the upper one of the two which are
connected to the multimeter. The grownd connection is done by using a small copper sheet. It is
on one side soldered to the socket and on the other end it is fixed on the chassis with a screw.
26
Faults
One reason could be the resistor (8) having a short circuit. Normally its resistance has the value
of 200K.
One more possibility is a faulty capacitor. Check the capacitors, mainly the one on top (7) of the
oscillator circuit which is connecting the anode of the tube with the oscillator circuit. Check also
the grid capacitor (6) which is placed underneath the oscillator chassis. Try to find out whether
they are broken by applying a reasonable force on them to see whether they are still in one
piece.
Check also the other, bigger capacitors (9), (10) and (11) however if one of them was faulty,
most probably the fuses would blow.
3.3.2 Resistance furnace operation
See Chapter 2.4.2 “IResitance furnace operation”.
Pos : 3.8 /0025 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 Bad combusti on ELTRA @ 6\mod_1368608656440_9.doc x @ 40248 @ 2 @ 1
27
Faults
28
Faults
operation manual chapter “Replacing the heating elements”. If you want to be sure that
the heating elements are broken, then measure the voltage on the wires to the heating
elements.
29
Faults
Pos : 3.10 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 T he furnac e does not cl os e ELTR A @ 6\mod_1368610904372_9.doc x @ 40292 @ 2 @ 1
V7
4
V10
7
V8
12
13
14
15
11190
30
Faults
15
V7
12
V10
V8
31
Faults
Pos : 3.12 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Anal ys edauer zu lang - The anal ysis takes too long ELTR A @ 6\mod_1368613082100_9.doc x @ 40314 @ 2 @ 1
32
Faults
DANGER
Mortal danger from electric shock
Exposed power contacts - High Voltage
2
1
3
1
5
9
2
10
6
11
12
1
7
33
Faults
34
Faults
Pos : 3.14 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 D evelopment of dus t when cl osing the fur nac e ELTR @ 6 \mod_1368620567231_9.doc x @ 40336 @ 2 @ 1
35
Faults
4
5
If the combustion tube (4) was in use for a very long time and many samples with high Sulfur
were analyzed, then sulfuric acid may build up in the rear-side of the combustion tube.
When the heated tube is removed together with the connector (5), a brown fluid can be seen in
the ceramic tube.
During the analysis, some SO2 is trapped in this fluid, causing non
repeatable results. It is recommended to clean the furnace outlet tube.
36
Faults
Pos : 3.18 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Ergebniss e si nd 0,0% oder 00,0ppm ELTR A @ 6 \mod_1368621911416_9.doc x @ 40380 @ 2 @ 1
Pos : 3.20 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 N o oxygen flow @ 7\mod_1380623315133_9.doc x @ 49429 @ 2 @ 1
37
Faults
38
Faults
If in the device state the graphs of IR-temperature °C or %IR are not in the green area, adjust
the temperature. See chapter “IR-cell temperature control”.
Window . The power is 100% after switching on from cold while heating up. It takes about
30 minutes for the temperature to reach the set point. In stable condition, the power required
should be between 30 and 70%.
If the value is less than 30%, adjust a higher temperature set point. If the value is over 70%
without any tendency for coming down to lower values, adjust a lower set point.
Pos : 3.24 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Breaker turns off at the begi nni ng of c ombusti on ELTR A @ 6 \mod_1368688648887_9.doc x @ 40470 @ 2 @ 1
39
Faults
DANGER
Mortal danger from electric shock
Exposed power contacts - High Voltage
Make sure all capacitors are discharged
After the cables (1) have been connected, plug the analyzer to the mains power again.
– Set the power switch to pos. 3
– Enter a weight manually (for example, 500 mg).
– Press START
– The word "Analyzing" should appear in the status window.
If not, but "Waiting for stability" appears instead, the reason is that after restarting the
device and the oxygen flow, the infrared base line is drifting. In order to save time, there
is a possibility to start the cycle without waiting for stability.
– Quit by pressing STOP
– Go to the "power user" level and click on the button "Stable".
– The analysis cycle starts then without waiting for base line stability of the cells.
40
Faults
When "Analyzing" appears on the status window, the light bulb should glow; this does not
happen suddenly, but gradually (soft start).
Within a few seconds, the light bulb will gradually change from dark to fully bright.
– If not, but the lamp becomes immediately bright like switching on with a switch, replace
the board HF-42.
– If the light is not fully bright but dim and flickering, replace the triac (6).
If yes (O.K.), unplug the mains power plug and reconnect the power cables (1) to the
transformer (2).
Unplug the power plug and reconnect the cables (1).
When reconnecting the power cables (1), do not let these wires be close to the connectors for
the cables (4) and (5). Otherwise the high voltage can create sparks.
The cables (4) and (5) should be unscrewed from the transformer and placed away from it.
DANGER
Mortal danger from electric shock
Exposed power contacts - High Voltage
After the generator starts the cable carries a voltage of about 7000 Volts!
– Plug in the mains power plug to power the analyzer.
A new analysis should be started. When "Analyzing" appears and the fuse is not damaged as
well as the mains power current didn't exceed 10 A (RMS), then the transformer is functioning
properly.
Otherwise, replace the transformer. The cables (4) and (5) are then reconnected.
If the problem still remains:
– Unplug the analyzer from the mains power.
– The cable (7) is unscrewed and removed from the capacitor (8)
After the generator starts (Analyzing) the cable (7) carries a voltage of about 7000 Volts! The
cable (7) therefore should be laid so that its end is at least 10 cm away from the nearest object!
If, after pressing START and after "Analyzing" appears, the breaker turns off, then replace the
rectifier (9).
DANGER
Mortal danger from electric shock
Exposed power contacts - High Voltage
Make sure all capacitors are discharged
41
Faults
10
5 9
1
11
12
42
Faults
Pos : 3.26 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Breaker turns off at the end of the anal ysis ELTR A @ 6\mod_1368692228547_9.doc x @ 40492 @ 2 @ 1
Causes:
The triac is malfunctioning.
The so-called Latch-up effect causes the triac to remain conductive in one direction after
triggering. The transformer is then acting as a short circuit because it is fed with a DC voltage!
– Replace the TRIAC.
Pos : 3.28 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 D ustcl oud when openi ng the furnace @ 9\mod_1425546854305_9.doc x @ 74956 @ 2 @ 1
43
Faults
44
Faults
Possible faults:
3
2
45
Faults
After starting an analysis”, the following message appears: "no oxygen pressure".
If this message permanently remains in the “status window”, then check
– Is the power switch (1) set on position 3?
– The pressure on the oxygen gauge (2) may be too low. 1.5bar is normal (22psi). The
pressure remains constant when the furnace is closed, and it sinks during the upward
movement of the furnace floor.
– Check the oxygen pressure in the supply line or tank. 4 to 6 bar (60 to 90psi)
showing on the pressure gauge is normal.
– Check if the pressure gauge for compressed air (5) shows 4 to 6 bar (60 to 90 psi).
– Check if the cock (3) is properly shut, or else the O-ring (4) will not seal Check if the
O-ring (4) is in a good condition.
– Check if the combustion tube is broken or cracked.
– Do leakage check. See chapter: “Leak checking”.
– Check if the chemicals are saturated and have become hard.
– The paper filter could also be the cause. For changing the paper filter, read the
operating manual.
– See chapter “Fluctuating gas flow”
– The contents of the gas cleaning furnace could be worn out and clogged.
46
Faults
Other faults:
– The screw for the upper furnace cushioning is too tight, see chapter “Pneumatics”.
– The pressure switch does not close, see chapter “Pressure switch”.
– The furnace closes too slowly, adjust according to chapter “Pneumatics”.
– The furnace does not close: see chapter “The induction furnace does not close”.
Pos : 3.32 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Waiting for s tability mess age @ 9\mod_1421835931277_9.doc x @ 72389 @ 2 @ 1
When this message appears in the status window after starting analysis, it means that the base
lines are not within the "deviation-limit", set in the "configuration" window. A usual value is 10mV
and the set time is 20 seconds. By setting the voltage and the time, both criteria for base line
stability are considered, the noise and the drift.
Reasons for instability:
1. Too low deviation value
One reason for the above message is that the deviation value in the configuration window is too
low, so that a higher value should be entered. The levels of noise of all cells are not necessarily
identical. Longer cells have normally higher noise than shorter ones and the sensors
themselves have noise level deviations between exemplars. The deviation can be increased in
the Configuration window, Base lines. Usual value is are 10mV / 20 seconds.
47
Faults
Induction furnace:
– The sample weight or the weight of the accelerator is too high.
– Don’t use copper as accelerator.
Resistance furnace:
– The furnace temperature is too high. Usual operation temperature are 1350°C.The higher
the temperature the shorter the service expectancy of the ceramic tube.
– The faster the furnace heats up, the higher the stress for the combustion tube. A
compromise is to lower the current limit of the heating up current.
See chapter “Resistance furnace”. in this manual
48
Faults
Pos : 3.36 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Oxygen pres sur e falls when the fur nac e opens ELTRA @ 6\mod_1368705931127_9.doc x @ 41187 @ 2 @ 1
49
Faults
When the furnace is open, the valve ( V2 ) should have 23 V DC on its magnetic coil.
When the furnace is closed, the voltage should be at zero volts.
– If the voltage on the valve ( V2 ) remains at 23 V DC, no matter whether the furnace is
opened or closed, then there is an electronic problem.
– If the valve voltage always stays at zero, then either the wiring is broken, see chapter
“Wiring diagrams” or the electronic unit is defective.
– If the voltage on the valve ( V2 ) is correct, then the valve itself is defective, see chapter
“Cleaning the solenoid valves”.
Pos : 3.38 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 C ombusti on without starting anal ysis ELTR A @ 6 \mod_1368706920531_9.doc x @ 41209 @ 2 @ 1
50
Faults
51
Faults
When the pulses at TP13 are present without starting analysis and with 12 V DC at TP9, then
the circuit board HF 42 is out of order.
If no scope is available, the pulses can be detected by using a voltmeter.
Normal values are about 0 V AC without pulses, and about 1 V AC with pulses.
This indicated value can be different, depending on the type of the voltmeter used.
If this AC voltage is practically zero but even though the generator is powered, the next possible
reason: The triac is faulty (short circuit), and has to be replaced.
Pos : 3.40 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Cr ucibl e or boat cr ac ks or melts ELTRA @ 6 \mod_1368708007091_9.doc x @ 41231 @ 2 @ 1
52
Faults
Either the plastic tube supplying oxygen to the analyzer is squashed, or the oxygen regulator is
not properly open.
Pos : 3.44 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Pump voltage too high @ 9\mod_1425548769110_9.doc x @ 74988 @ 2 @ 1
53
Faults
54
55
14 31
Faults
3
2
1
9
32 36
29
10
24
Pressure regulator
Pressure regulator
22 33
35
20
28
5
27
11
25
19 37
23 12
26
3
34
6
7
2
16 38
38 4
1 17
18 13 15
Faults
4 Bypass valve V4
5 Oxygen valve V5
6 Flow regulating valve V6
7 Purge valve V3
8 ----
9 Exhaust valve V9
10 Valve V10
11 Solenoid valve V11
12 Solenoid valve V12
13 Solenoid valve V13
14 Gas outlet
15 Lance valve V2
16 Pressure switch
17 Oxygen pressure gauge 1,5 bar
18 Restrictor
19 Restrictor
20 Carrier gas flow meter
21 IR purge flow meter
22 Moisture trap after the furnace
23 CO2 / H2O – purification trap
24 Catalyst furnace
25 SO3 - trap
26 H2O – trap
27 Dust filter
28 IR-cell
29 Gas pump
30 Attenuator before the pump
31 Attenuator after the pump
32 Resistance furnace
33 Induction firnace
34 Furnace input flow meter
35 Furnace inlet-oxygen
36 Furnace outlet measuring gas
37 Flow sensor
Detailed explanation of the faults indicated in the flow diagram
1. Blockage of the lower connection of the glass tube (22)
The bore hole of the lower connection of the glass tube (22) can get blocked when there
isn’t enough quartz wool at the bottom end of the glass tube (22) enabling fine particles of
the magnesium perchlorate to fall into the bore blocking it. It is therefore absolutely
important to fill the glass tubes exactly according to the Instructions. See Operation
Manual “Filling the reagent tubes”.
Holes of other reagent tube connectors need to be cleaned from inside and outside of the
analyzer, due to the fact that the passage is not straight, but vertical outside and
horizontal inside the analyzer.
56
Faults
CAUTION Switch off the mains power when handling inside the analyzer.
If a tube fitting is blocked, then the tube may be blocked as well.
Therefore the tube needs to be cleaned or replaced, otherwise the contamination could reach
the IR-cell as well.
2. The IR-cell is blocked
This fault is caused mainly by a kink in one of the tubes.
A blockage due to contamination may appear in one of the fittings of the IR paths, where
is a minimum diameter in the gas flow, compared to the inner diameter of the tubes.
3. Leakage
Not only blockages of the flow can cause the pump to run with full power, but also
leakages between the flow sensor (37) and the pump (29) can cause the pump to run at
maximum power.
Occasionally, the cover of the attenuator (30) (see also below) is not properly sealed, or
the gas flow meter (20) has a crack at the fittings for the tube connections. Disconnect
the gas flow gauge and check for cracks.
4. Blockage in the lower connection of the thick glass tube (22)
If there is not enough quartz wool used underneath the magnesium perchlorate, fine
particles fall down to the glass fitting causing blockage in its hole.
– Remove the glass tube (22). See operation manual “Filling the reagent tubes”.
– Remove the tube from the lower fitting.
– Clean the hole by passing a wire or a pipe cleaner.
5. Cotton wool is dirty
Another possibility could be that the cotton wool inside the glass (30) is very dirty.
CAUTION Don’t interchange the tubes of the glasses
57
Faults
4
5
Normal warm-up time: from room temperature to 1300°C, in 15-20 min unless the current limit
on the board is set for lower heating-up current.
Faulty heating elements:
Investigate the problem the following way:
Set the temperature to maximum value and measure the voltage across the heating elements.
58
Faults
This problem can appear in resistance furnace operation, when some material from previous
samples, fall off the combustion boat when shifted into the furnace. Then, when removing the
boat out of the furnace, the hot combustion boat may touch and burn this material, causing an
unwanted small peak on the base line.
If the following analysis is started while the base line is still above zero, then the electronics will
take this level as base line at the beginning of analysis so that the base line drifts negative when
that small peak comes down.
A problem of the IR-cell itself can be ruled out, when running a few blank analyses by just
entering manually a sample weight into the memory and pressing “START”, without inserting
any sample into the furnace. If the base line doesn’t show a negative drift, then the IR-cell is
O.K.
59
Faults
Pos : 3.50 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 R esistance furnace c ompl etl y c ol d or too hot @ 9\mod_1425625509164_9.doc x @ 75116 @ 2 @ 1
Pos : 3.52 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 F umes in the resis tanc e furnac e @ 9 \mod_1425626045991_9.doc x @ 75128 @ 2 @ 1
60
Faults
CAUTION! Never turn on the oxygen flow, when fumes come out of the furnace, or else
there can be an EXPLOSION!
Wait for at least 15 minutes, until the smoke has disappeared from the furnace before turning
on the oxygen supply!
When smoke comes out of the resistance furnace front, mostly the reason is that a sample is
entered into the hot zone of the furnace without having oxygen flow. If at that moment the
oxygen flow is restored, the gases in the furnace will ignite and explode with a loud bang and
with a flame shooting out of the combustion tube.
The explosive gases consist of carbon monoxide from carbon in the sample and (very little)
oxygen from the air which was in the combustion tube and of volatiles driven out of the sample
by the temperature without burning due to missing of oxygen.
O2
6
7 VAC
5 2
220 VAC
4
3 2
1 1-15-3
If not:
– There is no voltage (220 V) on the leads (2) of the coil (4.)
– The valve is blocked. Clean the valve by
a) closing the oxygen supply,
b) setting the mains switch to pos. 0
61
Faults
– The heating elements are used up. See chapter “Bad combustion”.
– The furnace is faulty. See Chapters “No combustion”, “Resistance furnace”,
“Resistance furnace doesn’t reach full temperature”, “The resistance furnace is too
cold or too hot” and “Resistance furnace takes too long to warm up”.
Pos : 3.56 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 U nstable r esistance furnace temperatur @ 9\mod_1425626418438_9.doc x @ 75152 @ 2 @ 1
– The heating elements are used up, and the required temperature can only hardly be
reached so that there are no reserves of power, to properly regulate the temperature at a
stable level. See chapter “Resistance furnace doesn’t reach full temperature”. To
confirm this fault, set the temperature to a much lower value. If the temperature cannot
reach the new higher level, it is obvious that the unstable temperature was at maximum
level of temperature the heaters can reach.
– The temperature control board TH 43/44 is faulty. Replace the TH 43/44 board and adjust
according to chapter “Resistance furnace”.
– The thermocouple is faulty. See chapter “Thermocouple testing”.
62
Faults
Pos : 3.58 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 D ust or s oot from the r esistance fur nace @ 9\mod_1425626621251_9.doc x @ 75164 @ 2 @ 1
63
Faults
Pos : 3.60 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Vibr ations i n i nduc tion furnac e mode @ 9\mod_1425626754537_9.doc x @ 75176 @ 2 @ 1
If the Analyzer vibrates when the mains power switch is turned to position 3, the reason is dust
accumulation on the blades of the oscillator’s blower. The vibrations are caused by dust on the
blades causing eccentrical rotations.
DANGER
Mortal danger from electric shock
Clean it carefully with a vacuum cleaner, if necessary:
Vacuum cleaning
Vacuum cleaning
The minimum analysis time is set too short in the configuration window, so that the analysis time
expires before the peak starts coming up. Normally the peak starts about 6 to 8 seconds after
the start of combustion. If the minimum analysis time is set for example to 4 seconds, the
analysis may be terminated before the peak starts. If the minimum analysis time is set far over 8
seconds, the peak starts before the minimum time expires. Due to the IR signal being then
higher than the comparator level, the analysis continues until the IR signal comes down to the
comparator level, unless the maximum analysis time expires earlier. If so, it is advisable to
increase the maximum analysis time in order to integrate the whole peak.
64
Faults
Especially in case of resistance furnace operation, the peak of hard-to-burn samples starts
coming up quite late. In such cases, the minimum analysis time should be set to about 60
seconds.
If after prolongation of the minimum analysis time to a reasonable value, a peak still doesn't
appear, possible reasons for the fault are:
– No combustion. See chapter “No combustion”.
– There is no gas flow. See chapter “No flow”.
– There is gas flow, but bypassing the furnace. See chapter “Sample burns but no peak
on the screen“.
– The IR cell is faulty. See chapter “Check base lines“.
Pos : 3.64 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 C hopper motor does n‘t r otate ELTR A @ 6\mod_1368709089073_9.doc x @ 41319 @ 2 @ 1
The rotation of the chopper motor needs to be checked in case of problems only, like when
there are no combustion peaks on the screen or the IR signals are minus -10V (take care,
minus) displayed on the PC screen in the window “inputs”.
In order to check whether the chopper motor is rotating, the easiest way is to remove an IR
path. See chapter “IR-paths, cleaning and replacing”. Mostly, the easiest path to remove is
the shortest one. Then, the chopper blade can be seen thru the hole of the chopper housing,
which becomes free (open) after removing the path.
The reason of failure can be a faulty motor but also the controlling electronics of the motor may
have become defective. If so, the motor or the IRC board has to be replaced. Before replacing,
check at least the voltages on the board, mainly the supply voltage of 24VDC. See chapter
“Infrared cell adjustments”.
For replacing the chopper see chapter “Chopper”.
Pos : 3.66 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Sample burns but no peak on the s creen @ 7\mod_1381400077047_9.doc x @ 49828 @ 2 @ 1
When the analysis cycle is carried out normally and the sample is burned but no peak on the
screen, possible reasons are the following:
- The IR cell could be faulty. Check the base lines. See chapters “Check base lines” and
“Infrared cell adjustments”. If the base lines look normal, the IR cell can be regarded as
ruled out. If the base lines are practically at zero volts, the cell is faulty or without power.
(24VDC). See chapter “Bad combustion”.
- The plunger of bypass valve V4 jams. See chapter “Cleaning solenoid valves”.
This may happen in case of induction furnace operation. There is gas flow, but bypassing the
furnace, so that there is gas flow shown on the flow meter and by the flow sensor on the screen,
but no gas flow thru the induction furnace. In this case the bypass valve V4 jams, causing the
oxygen flow to bypass the furnace like in case of open furnace. See chapter “Gas flow
system”.
The furnace has though oxygen pressure but without having oxygen flow. Due to oxygen being
present, the metal sample and accelerator(s) burn, because there is availability of oxygen in the
furnace, coming thru the chamber valve V3 and/or thru the lance valve V2. In case of metal
65
Faults
analysis, the burned sample in the crucible looks like normally combusted sample. If so, check
the bypass valve V4. See chapter “Cleaning Solenoid valves”.
Pos : 3.68 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 IR signal does not c ome down to the base line at the end of anal ysis @ 7\mod_1381404509510_9.doc x @ 49850 @ 2 @ 1
3.34 IR signal does not come down to the base line at the end of analysis
Pos : 3.69 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-2000_s er vice/05Faults _C S-2000/0685 M odul IR signal does not c ome down to the base line @ 9\mod_1425541763229_9.doc x @ 74537 @ 3333 @ 1
66
Faults
67
Faults
68
Faults
69
Adjustments
Pos : 5.1 /0025 Ü bersc hriften/ELTR A 1. Ü berschriften/1 Adj ustments eltr a @ 8 \mod_1396882697255_9.doc x @ 60262 @ 1 @ 1
4 Adjustments
Pos : 5.2 /0025 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 Gas fl ow c ontroll er adj ustment and j umper s etti ngs ELTR A @ 6\mod_1368776355883_9.doc x @ 41371 @ 2 @ 1
70
Adjustments
Jumper settings:
71
Adjustments
72
Adjustments
73
Adjustments
In normal operation the pump voltage is displayed in the “device state window” lies
between 5V to 8V.
TP6: Flow sensor output voltage. This voltage is displayed on the PC screen in the “device
R13: Pump frequency adjustment. The pumping frequency of the pump can be adjusted by
turning this trimmer, clockwise for higher and counter clockwise for lower pump frequency. The
pumps are basically designed for 50 Hz operation. The pump frequency has a factory
adjustment so that it is not advisable to change it.
X1: Flow sensor connector FD9.
Pin 2: Sensor output
Pin 3: GND
Pin 5: +12V
X2: Connector to UNI board.
Pin 1: Sensor output / TP6
Pin 2: Input for flow start command/TP4. 12V means no flow. 0.5V for flow.
X3: 24V power supply.
Pin (middle) +24V. Outer contact GND.
X4: Pump connector.
Alternating 24V square voltage voltage/TP1 – TP2. Frequency adjustable by R13.
The duty cycle of this square voltage defines the power of the pump.
JP1: The pins 1 and GND must be connected with a jumper for late analyzers.
Older analyzers work also with a jumper, however the pump and flow voltages in the device
state window are more stable without jumper.
For details about wiring see chapter “Wiring diagrams”.
74
Adjustments
Remark:
Due to a few failures of the flow sensors, almost exclusively with users who analyse samples
with high sulphur and high moisture or regularly treat their samples with acid, we concluded that
the reason of the failures should be of chemical nature i.e. sulphuric acid and other acids
damage the sensor.
The solution is to prevent the analysis gases to get in contact with the sensors. Therefore the
sensor ports are connected with long tubes to the gas flow system. There is no gas flow through
the sensor tubes but only the pressure is sensed by the sensor.
In case of failure of a previous version sensor, it is advisable to replace it by the new version. In
this case the part numbers of the complete assemblies should be used for ordering. In case the
analyzer has already the new version sensor, the order should include the spare part number.
See drawings below and chapter “Ordering numbers”.
Pos : 5.4 /0025 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 Infrar ed bas e line adj ustments ELTRA @ 6\mod_1369121936836_9.doc x @ 41542 @ 2 @ 1
75
Adjustments
The outputs of the IR cells (IR base lines) are factory set at 8V when the paths are new and
clean. The peaks come down starting from 8V coming down to a minimum voltage (at the peak
maximum) and they go back to the 8V baseline after completion of combustion. The 8V IR-base
line is obtained through preamplifier adjustment only.
In real operation, with growing contamination resulting in gradually weakened radiation reaching
the detector, the IR cell output base lines will move down to levels below 8V depending on the
grade of contamination. The (negative output) peaks of the cell will also change to smaller
amplitude.
The base line is not regulated any more.
Due to a recalculation formula, the base lines on the screen are always zero volts, the peaks
are positive and they are always of equal amplitude and surface (for same sample and same
weight) regardless the path contamination, regardless the cell base line level, and regardless
the cell output peak amplitude. Consequently the integrated value (of the peak on the screen)
remains the same and of course also the analyses results remain constant at any grade of
contamination.
– When loading the UNI software select from now on "New cell" mode.
76
Adjustments
77
Adjustments
78
Adjustments
79
Adjustments
Example
A three cell module with a low carbon, a high carbon and a sulfur cell has
– the low carbon signal on section [1], the LoC preamp is on X1
– the high carbon signal on section [2], the HiC preamp is on X2
– the Sulfur cell signal on section [3], the S preamp is on X3
The base line adjustments are done by adjusting the gain of the preamplifier of each IR cell in
the IR module (rack). The gain adjustment is done by turning the trimmer which is inside of each
preamplifier. The base line voltages can be measured on the following test points:
80
Adjustments
81
Adjustments
13. Adjust the outputs of other cells (not golden paths) to have between 7 to 8V by adjusting
the gains of their preamplifiers.
14. Set the jumper of the JP1 from position 2 (Test °C) to Position 3 (Analysis).
15. Install the IR module into the analyzer.
16. Allow sufficient time for the base lines to stabilize.
NOTICE It is necessary to turn slowly in order for the IR sources to stabilize their temperature
with increasing IR source voltage.
– Adjust the other cells of the module (if any) at 8V baselines or 5V to 6V in case of clean
golden paths.
– Install the IR module into the analyzer. Allow sufficient time for the base lines to stabilize.
– Start gas flow and check the base lines again. If any of them is significantly over 8V then
readjust.
Drift too large
– Check the temperature regulation.
– Unstable power supply.
– Defective IR-source.
NOTICE If it is shown on the display that the infrared cell's output voltage is over 10 volts then
there is a serious defect.
82
Adjustments
Pos : 5.6 /0025 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 IR s ource voltage setti ng Eltr a @ 8\mod_1396940082361_9.doc x @ 60276 @ 2 @ 1
83
Adjustments
Pos : 5.8 /0025 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 Infrar ed cell temper ature regul ati on ELTRA @ 6\mod_1369641998919_9.doc x @ 42289 @ 2 @ 1
84
Adjustments
order to have the temperature adjustable in the above range, a jumper has to be set on position
5-Hi Temp. of JP1.
The temperature is set by turning the pot and measuring the set value on
The actual value of the temperature is measured on TP26.
Caution: After changing the temperature set point, it takes some time until the actual
temperature value is stabilized at the new set level.
85
Adjustments
Pos : 5.10 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Pneumatic s ELTR A @ 6\mod_1369229118822_9.doc x @ 41844 @ 2 @ 1
4.5 Pneumatics
Pos : 5.11 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-2000_s er vice/10Adjustments_C S-2000/1025 Modul Pneumatics @ 9\mod_1425542494387_9.doc x @ 74610 @ 3 @ 1
86
Adjustments
5
3
6 10
7 11
12
14
13
15
87
Adjustments
5 Adjusting screw
6 Pneumatic cylinder
7 Lower cushioning
8 Pressure regulator
9 Piston valve
10 Cushioning valve
11 Coil cooling valve
12 Coil cooling adjustment
13 Pressure gauge
14 Combustion coil
15 Muffler
Pneumatics adjusting for correct furnace movement:
When closing the furnace, the piston moves.
Open the right side panel of the analyzer to find the pneumatic valves (9 / V7), (10 / V10) and
(11 / V8) as well as the pressure regulator (8).
- Turn the restrictor (12) clockwise up to the end for complete restriction.
- Check the pressure on gauge (13) to be between 4-6 bar (60-85psi).
- Press F2 to open the furnace.
- Pull to unlock and then rotate the knob of the pressure regulator (8) counter clockwise for
minimum pressure (zero pressure).
- Turn the screw (5) counter clockwise until the magnetic switch support (4) is free to move
up and down on the cylinder (6).
- Shift the magnetic switch upwards up to the end of its adjusting range.
- Press F2 for closing the furnace. The furnace closing cone (1) won’t yet move or move
very slowly.
- Turn clockwise the knob of the pressure regulator (8) until the closing cone (1) can rise
smoothly (without shaking) up to the upper end of its stroke. To do this, open and close
the furnace (F2) as often as needed. Take software delays into account.
- The pressure should be only as high as necessary and low enough for a person to be
able to stop by hand the cone from rising.
- Push the knob to lock the pressure regulator (8).
- When the furnace is closed, slowly shift the magnetic switch (3) with its support (4)
downwards until it just switches on. This can be recognized due to the clicking of a
pneumatic valve (10/V10) and the piston moving then slightly higher.
- Make a marking on the cylinder (6) at the upper end of the magnetic switch holder (4).
- Shift the switch with holder downwards until the magnetic switch is off (A valve clicks and
the piston slightly moves downwards).
- Make a second marking on the cylinder (6) at the upper end of the switch holder (4).
- Adjust the switch holder (4) to have the upper end in the middle of the two markings.
- Open the furnace and then close it again (F2) to verify the correctness of the adjustment.
- Adjust the lower cushioning (7), so that the furnace opens without the piston banging at
the bottom.
88
Adjustments
- Click on “Communication test”-“Digital outputs”, on the 9th button counting from the left,
to energize the coil cooling valve (11/V8), or start analysis cycle.
- Turn the restrictor (12) counter clockwise until the pressure on the gauge (13) is reduced
by about 0.2 to 0.3 bar (3 to 4 psi) when the coil cooling valve (11/V8) is open.
- Fix the locknut of the restrictor (12).
REMARK:
In case of using valuable gas for operating the pneumatics, the restrictor (12) should be
adjusted for lower flow by turning it clockwise. It can even be completely closed by turning it
clockwise up to the end of its range.
The flow adjusted with this restrictor is for cooling the combustion coil to ensure very long
service life. According to our experience, the combustion coils of our analyzers don’t need
replacement, even after years of operation.
89
Adjustments
Pos : 5.12 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Press ure regul ati on ELTR A @ 6 \mod_1369298860002_9.doc x @ 41930 @ 2 @ 1
Inlet pressure
90
Adjustments
– Pull and rotate the knob of the regulator, until the pressure gauge (6) shows the correct
value.
– Ensure that the flow shown on the flow meter (5) shows between 180 and 200l/h; if
not, adjust according to the related chapter.
91
Adjustments
Purge pressure
1 1
1.5
0.5 2
bar
0 2.5
PR2
– Ensure that the shaft (1) is securely fastened to the (1). (engineers wrench size 17).
– In order to adjust the cone, the screws (4) should be loosened so that the cone (3) moves
horizontally on the bar (2).
– Close the furnace ( with the lift upwards ).
– Tighten the screws (4).
– The cone is now positioned correctly.
– The screws (5), (6), (7), must now be well tightened in this order.
– The screws (8) should be properly fastened (with a wrench).
– The screws (9) should be tightened by hand only.
92
Adjustments
Pos : 5.16 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Linearity c orrecti on @ 9\mod_1425629238672_9.doc x @ 75212 @ 2 @ 1
93
Adjustments
Pos : 5.18 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Induction g enerator c ontrol ELTRA @ 6\mod_1369387490836_9.doc x @ 42096 @ 2 @ 1
TP13 TP14
TP10
TP11
J2
TP6
TP2 TP7 TP8
J3
TP4
TP5 TP12
TP3
TP9
Fig. 74: HF 42
94
Adjustments
The following diagrams show the relationship between the set value (adjustment with P2 on the
HF 42 board) and the generator parameters.
95
Adjustments
Pos : 5.20 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 R esistance furnace @ 9\mod_1425629579658_9.doc x @ 75224 @ 2 @ 1
96
Adjustments
NTC TP6
J5/9
Mains L
Temp.compensation Current
limitation
+15V
J5/8
+15V
TP2 +5V
TP4
max.
J2/5 setpoint
TP7
2-6-3
97
Adjustments
Pos : 5.22 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 T her moc ouple testi ng @ 9\mod_1425629857432_9.doc x @ 75276 @ 2 @ 1
TH43
36907
TP2
J1 1
TP3
TP5
TP4
AB
8
7
1 2
6 TP7 A
J2 5
B
4 1 2
3
2
1
J3 TP11
TP12
TP6
TP1
J5
P1 P2 P3 P4 2-15
OHM
98
Service
Pos : 7.1 /0025 Ü bersc hriften/ELTR A 1. Ü berschriften/1 Ser vic e @ 6\mod_1369640568215_9.doc x @ 42168 @ 1 @ 1
5 Service
Pos : 7.2 /0025 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 IR-paths, cleaning and r epl acing ELTRA @ 6\mod_1369640666897_9.doc x @ 42179 @ 2 @ 1
– Remove the plastic tubes from the IR-paths (gas inlet / outlet).
– Unscrew the rings with windows on both ends of the paths and remove it for cleaning
purposes (screws M 2.5 in the following schematics).
– After disassembly check the two windows for damage.
– For cleaning the path use the cleaning brush for golden IR-paths, see picture below.
99
Service
100
Service
101
Service
NOTICE
The position of the IR sources in their housing has to be as shown in the graphics A and B.
Correct: Wrong:
5 5 5
5
B C D
A 6 5 5
5
E F
5
33--22--11
5.3 Chopper
Pos : 7.7 /0010 ELTR A/0015 s er vice_i nstr uctions /CS- 2000_s er vic e/15 Ser vice_CS- 2000/1520 Modul C hopper @ 9\mod_1425543409739_9.doc x @ 74742 @ @ 1
Motor replacement:
The replacement motors are supplied with their wires already soldered on a 9-pin plug, so that
in case of replacing a chopper motor, only the two wires of the IR sources have to be soldered
on the connector.
102
Service
The chopper blade (1) is fastened to the motor (2). The chopper blade must be removed first in
order to dismount the motor (2) from the mounting plate (3)
Disassemble in the following order:
1. safety ring (4)
2. the washer(s) (5) ( there could be two )
3. the chopper blade (1)
4. the cylinder (6)
5. the spring (7)
103
Service
104
Service
– After pressing the button, the pressure drops on gauge (3) for about 5 seconds and then
it remains constant. Watch it for about a minute. If there is no pressure drop any more,
the complete gas system is seal so that the leakage test is successfully completed.
– Keep holding the button (2) for about one minute. Should there be a continuous pressure
drop, then release the button (2), open the furnace, press and hold the button (2) again. If
the pressure with open furnace remains stable then the leakage is in the furnace area. To
the furnace area belong also the metal dust trap (6), the tube (5) and the moisture trap (7)
underneath the metal filter, the dust filter cartridge on the front panel and the outflow
valve V1.
1. Leakage in the furnace area:
According to experience over years, leakages are almost exclusively in the furnace area. Most
common reasons in the order of probability are: (see graphic below).
- The combustion tube (12) has a crack or a hole due major accumulation of splashes from
the analyzed samples. See graphic below.
- The two small O-rings (9) sealing the top of the furnace are damaged, deformed or dusty.
- The palm grip (8) is not in the right position so that the O-ring, sealing the upper end of
the filter housing, is not properly pressed against the filter housing. Rotate to the right
position (maximum force on the O-ring).
- The O-rings (10) sealing the combustion tube are very unlikely to leak, provided that the
winged nuts fixing the furnace closure are driven normally by fingers. The use of
excessive force is not needed and not recommended.
- The O-ring (11) sealing the furnace with the closing cone is very reliable as well. Only if it
is burned by a hot crucible when taking the crucible from the pedestal, there may be a
sealing problem.
- The tube (5) can be damaged due to aging and pressure.
- The moisture trap underneath the metal filter can be not seal when the glass tube may
crack when fitting it.
- The plunger of the outflow valve V1 jams due to contamination. See chapter “Cleaning
solenoid valves” and “Wiring and gas flow diagrams”.
- The dust filter cartridge is leaking. Either one of the small O-rings sealing the cartridge
leads is not seal or one of the hexagon socket screws on the front of the supporting
fittings is not properly fixed.
9 10
8
10
12
5 11 b7-10-1
105
Service
106
Service
107
Service
ATTENTION do not cause a short circuit while measuring, otherwise the drivers will be
destroyed. See chapter “Electronic drivers malfunctioning”.
– If the voltages on V2, V3, or V4 are wrong, then replace the electronic drivers.
Pos : 7.10 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Sol enoi d val ves cleaning ELTR A @ 6\mod_1369643206541_9.doc x @ 42465 @ 2 @ 1
1
7 2
7 2
3
3
4
4
5
5
6 6
3-13-1
108
Service
1
2
7 2
3
3
4
4
5
5
6 6
3-13-1
109
Service
Pos : 7.12 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Fl ow s ensor r epl ac ement Eltr a @ 8 \mod_1396941567775_9.doc x @ 60384 @ 2 @ 1
11
9
10
1 7
6
Fig. 94: Flow sensor(3-14)
110
Service
Pos : 7.14 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Gas pump replacement Eltra @ 8\mod_1396941649931_9.doc x @ 60396 @ 2 @ 1
111
Service
Pos : 7.16 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 T her moc ouple replac ement @ 9 \mod_1425632233953_9.doc x @ 75300 @ 2 @ 1
77450
36720
77507
36242
36288
36288 77525
75140 70380
90155 36101
36620
70380
36465
90162
70410
36242
36445
36665 36283
36914
Viton 4x1 Länge:65cm
36975 36908
zum unteren Glasanschluß
36217
36216 36350
– Disconnect the chinch connector of the thermocouple cable from the temperature control
board.
112
Service
113
Service
41 board:
Set the jumpers
This option is functional jumper
with EPROMS. to conect A1 with A2 on J2
If theIf the
analyser
analyzer has thehas the Hf 41 board:
HF41 board:
Set the
– jumpers
Set the jumper to jumper
connect A1 withto conect
A2 on J2 A1 with A2 on J2 TP10
TP13 TP14
TP11
TP13 TP14
TP10
TP2 TP7 TP8
TP6 TP11
HF41: J2
I C 1
A B
TP2 TP7 TP8 External A1 B1 On board
TP3 TP4 TP6 J2
TP5 HF41:
flow enable A2 J2 B2 flow enable
I C 1
On board A3 A B B3 Continuous
TP12 power adjust.
External A
A14 B4
B 1 flowboard
On
TP3 TP4
J2
flow enable A2 B2 flow enable
TP5
TP9
On board A3 B3 Continuous
TP12 power adjust. A4 B4 flow
TP1 TP9
TP15 TP16 TP17 TP18 TP19 TP20
HF41 12008
TP1
HF41 12008
114
Service
Pos : 7.20 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Gas s aving also for resis tanc e furnac e @ 9\mod_1425632688373_9.doc x @ 75312 @ 2 @ 1
If this feature is required, the CS-2000 can be upgraded as follows. But the customer must take
care to have oxygen flow first before entering the sample into the furnace. The flow is restored
as soon as any button on the PC-keyboard is pressed. For example taking a sample weight.
If an organic sample is entered into the furnace without oxygen flow and the flow is
subsequently restored, it may lead to a very intensive ignition. Therefore, the gas saving feature
for the resistance furnace operation should only be installed after the customer’s consent to
take full responsibility for entering a sample into the furnace only after the flow is restored.
The modifications are shown in the following schematics. For full
Details, see chapter „Wiring diagrams“.
1. Install a solenoid valve No. 01160 in the analyzer’s cabinet.
Pressure
regulator
Schematic 1
PR1 0.5
1
1 .5
2
1 , 5 b a r / 2 2 p si
200l/h
bar
0 2.5
Oxygen inlet
2-4 bar
30-60 psi
230VAC
Gas saving for
resistance furnace V13
operation Pressure switch
01160 3-21-1
115
Service
By adding a solenoid valve (8) and connecting the coil directly to 220V AC, the piston won’t
change position in case of a power failure. If the piston was down, it will remain down; if it was
up, it will remain up.
Pos : 7.24 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 H alog en trap - i nstalling ELTR A @ 6\mod_1369643531895_9.doc x @ 42553 @ 2 @ 1
116
Service
Eltra analyzers don’t need any halogen trap when they are equipped with golden paths. However, in case a
user may insist in a halogen trap, the installation can be done as follows:
117
Miscellaneous
Pos : 9.1 /0025 Ü bersc hriften/ELTR A 1. Ü berschriften/1 Misc ellaneous ELTRA @ 6\mod_1370352969876_9.doc x @ 42719 @ 1 @ 1
6 Miscellaneous
Pos : 9.2 /0025 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 Ordering numbers ELTR A @ 6\mod_1373449838307_9.doc x @ 43582 @ 2 @ 1
118
Miscellaneous
119
Miscellaneous
120
Miscellaneous
121
Miscellaneous
122
Miscellaneous
123
Miscellaneous
124
Miscellaneous
70390 O-ring
71010 Cleaning brush for pedestal
125
Miscellaneous
126
Miscellaneous
127
Miscellaneous
77335 Capacitor
77333 Filter
77345 (13261; 13262; 77340; 77341; 77342) Capacitor
128
Miscellaneous
129
Miscellaneous
05278 IR pre-amplifier V4
05279 IR pre-amplifier
05350 IR front panel
05465 IR path ring with window (source side)
05469 IR path cone with window (detector side)
05479 Chopperassembly, 1 channel
05488 Chopperassembly, 4 channels
130
Miscellaneous
6.1.8 Chopper
131
Miscellaneous
70280 O-ring
70330 O-ring
75120 Spring
75190 Washer
132
Miscellaneous
133
Miscellaneous
134
Miscellaneous
135
Miscellaneous
136
Miscellaneous
Pos : 9.4 /0025 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 Wiring di agrams Eltra @ 8\mod_1399966125802_9.doc x @ 60786 @ 2 @ 1
137
Miscellaneous
138
Miscellaneous
139
Miscellaneous
140
Miscellaneous
bl blue
br brown
ge yellow
gr grey
gn green
3+16
2+15
or orange
22
21
11
17
25
rs pink
rt red
sw black
sw
sw
bl
br
bl
gr
vi
vi violett
ws white
Je nach Anzahl der IR-Zellen
According to the
number of IR cells
Heaters
200VDC as long
as the LED is on.
No voltage as long
as the LED is OFF
8 7 6 5 4 3 2 1
06255
PRI SEC
06640
10pin plug
1 2 3 4 5 6 7 8
sw sw
br bl
1 2 3 4 5 6 7 8 9 10
rt rt
rt br gn
gr
vi
gr
sw sw
rt
sw sw
Temperature
ge/gn sensor
rt
D-15 plug pin 9
ge/gn
1-14-4-1
141
Miscellaneous
142
143
7kV
Miscellaneous
500pF 100pF
600pF
22/10W 500pF
500pF
600pF
100k/50W
2k7/100W
100nF
100nF
F11.126/36
Amplitude circle
1µF 1µF
12085
Miscellaneous
Pos : 9.6 /0025 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 Gas fl ow diagrams El tra @ 8\mod_1396942447930_9.doc x @ 60504 @ 2 @ 1
144
Miscellaneous
145
Miscellaneous
146
Miscellaneous
147
Miscellaneous
Pos : 9.8 /0025 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 Spar e parts kits @ 7\mod_1382427709237_9.doc x @ 50008 @ 2 @ 1
quantity
Number
description
Part
148
Miscellaneous
149
Miscellaneous
75190 Washer 1
77140 HF-filter 250 V 1
78010 Mains power switch 1
90290 Copper oxide 100g
92610 Tube of grease 1
6.4.2 Spare Parts kit for CS2000 (including spare parts kits CS800 and HTF540)
quantity
Number
description
Part
05390 Heatsink 1
09090 Reagent tube 1
11105 Metal dust filter 1
11115 Plastic tube 75mm
11120 Paper filter holder 1
11180 Dust filter cartridge 1
11185 Paper filters 10
11370- Gasflow regulating valve 1
2001
11400 Pressure outlet solenoid valve 1
11410 Oxygen stop solenoid valve 1
11430 Purge solenoid valve 1
11440 Bypass solenoid valve 1
11470 One way valve 1
11490 Pressure regulator 1
12005 Gasflow and furnace control board HF3 1
12016 Gasflow and furnace control board HF42 1
12080 Rectifier 1
13067 Combustion coil 1
13175 Insulator 1
13260 High voltage filter 1
13262 Capacitor 700 pF/40KV 1
13270 Resistor 1
14045 Cleaning brush for combustion tube 1
14072 Ceramic heat shield for brush 1
150
Miscellaneous
151
Miscellaneous
Pos : 9.10 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften11 Mai ntenanc e (2015- 12- 03 07:57:57) @ 7\mod_1383121708299_9.doc x @ 50516 @ 2 @ 1
6.5 Maintenance
Pos : 9.11 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-2000_s er vice/20 Misc ellaneous_CS-2000/2025 Modul Mai ntenanc e C S-2000 @ 9\mod_1425544021010_9.doc x @ 74910 @ @ 1
– Is the furnace closure cone still properly fixed? See chapter “Closing cone”.
– Check whether the imprint of all the O-rings is at least of 2 mm thickness.
– Is the inlet pressure regulator set on 1,5 bar ( 22.5 psi )? See chapter “Pressure
regulators”.
– Is the other two pressure regulator set on 0,35 bar ( 5 psi )?
– Check for leaks. See chapter “Leak checking”.
– Is the gas flow rate correct? See chapter “Gas flow controller”.
– Pressure switch: change over at about 1 bar? Measure across the leads. See chapter
“No or low oxygen pressure”.
– Does the thermostatic control work correctly? See chapter “IR thermostatic control”.
– Check the infrared baseline. See chapter “Infrared base line”.
– Check whether all the screws of the generator housing are tight.
– Check whether all four „high voltage“ stickers are still in place.
– Check whether all plugs in the analyzer are properly plugged in.
– Check whether all the warning labels are still attached to the side-panels.
– Calibration of all measuring ranges.
152
Approved methodologies to which Eltra instruments conform
Pos : 11.1 /0025 Übersc hriften/ELTRA 1. Ü bersc hriften/1 Approved methodol ogies to whic h Eltra ins trumen ts c onfor m ELTRA @ 7\mod_1388672840004_9.doc x @ 53931 @ 1 @ 1
153
Approved methodologies to which Eltra instruments conform
CS 800
DIN EN ISO 15349-2 C Steel
CS 2000
CS 800
ISO 13902 S Steel/Iron
CS 2000
CS 800
ISO 4689-3 S Iron ore
CS 2000
CS 800
ISO 7524 C Nickel
CS 2000
CS 800
DIN EN 27526 S Nickel
CS 2000
CS 800
DIN EN ISO 15350 C,S Steel / Iron
CS 2000
DIN EN ISO 3690 H Steel H 500
ON 900
DIN EN ISO 10720 N Steel
ONH 2000
CS 800
ISO 10719 C Steel
CS 2000
Pos : 11.4 /0025 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Organic materials (Oil, C oal, foods tuffs) @ 7\mod_1388733114197_9.doc x @ 53990 @ 2 @ 1
154
Disposal
Pos : 13.1 /0025 Übersc hriften/1. Übersc hriften/1 Ents orgung @ 0\mod_1234258746831_9.doc x @ 6173 @ 1 @ 1
8 Disposal
Pos : 13.2 /0005 RET SCH/0000 Ger äteneutral/M odul eEnts orgung General (2015-10-21 11:05:55) @ 0\mod_1234269404935_9.doc x @ 6180 @ @ 1
155
Index
9 Index
156
Index
157
Index
158
Index
159
Copyright
© Copyright by
Eltra GmbH
Retsch-Allee 1-5
D-42781 Haan
Germany