〈1603〉 Good Cascade Impactor Practices
〈1603〉 Good Cascade Impactor Practices
Printed by: Titon Saha Official Date: Official as of 01-May-2021 Document Type: GENERAL CHAPTER @2022 USPC
DOI Ref: k9w4a DOI: https://doi.org/10.31003/USPNF_M13055_03_01
1
al
6. IN-USE ASPECTS
6.1 Initial Considerations
6.2 Inspection of Cascade Impactor Components Susceptible to Deterioration
6.3 Assurance of Stage Collection Surface Coating
6.4 Assertion of Correct Assembly
6.5 Mitigation of Air Leakage into the Apparatus
6.6 Setting of Flow Rate
6.7 Cleaning and Storage Between Uses
ci
6.8 Assertion that Individual Cascade Impactor Assemblies of the Same Type (i.e., Andersen Cascade Impactor or
Next Generation Impactor) are Interchangeable
ffi
1. INTRODUCTION
The multi-stage cascade impactor is the mainstay methodology for the determination of aerodynamic particle size
distribution (APSD) of aerosols emitted by Inhalation Aerosols and Sprays, and Inhalation Powders, as described in Inhalation
and Nasal Drug Products: Aerosols, Sprays, and Powders—Performance Quality Tests á601ñ. This equipment is also recommended
for APSD assessment associated with the use of drug products dispersed by nebulizing systems in Products for Nebulization—
Characterization Tests á1601ñ as well as for the assessment of add-on devices for use with inhalation aerosols in Spacers and
Valved Holding Chambers Used With Inhalation Aerosols—Characterization Tests á1602ñ. Both chapters provide instructions on
how to determine APSD, but due to necessity are focused on the measurement process itself, and are therefore limited in the
advice given to assist in obtaining consistent measurements by these apparatuses. The cascade impactor is a complex apparatus
to set-up, use, and maintain correctly. This chapter provides guidance to optimize the accuracy and precision of the methods
described in á601ñ, á1601ñ, and á1602ñ for the determination of APSD.
Regulatory activity in the mid-1990s that led to the 1998 publication of a draft US FDA Guidance for Industry on metered
dose inhaler (MDI) and dry powder inhaler (DPI) drug products chemistry, manufacturing, and controls documentation
obviated the need for an informational chapter covering good cascade impactor practice (GCIP). This draft Guidance (Metered
Dose Inhaler (MDI) and Dry Powder Inhaler (DPI) Products—Quality Considerations) was updated in 2018. Following the release
of this document, the International Pharmaceutical Aerosol Consortium on Regulation and Science (IPAC-RS) and later the
Product Quality Research Institute (PQRI), recognized that there was a need to define a method of addressing a failure to
achieve a mass balance in the quality testing of orally inhaled products (OIPs). After the PQRI working group was concluded,
the Cascade Impactor Working Group of the IPAC-RS continued to address reports from stakeholders that the cascade impactor
method was complex, resulting in measurement failures that were not attributable to product quality. Later, the European
Pharmaceutical Aerosol Group (EPAG) established the acceptability of stage mensuration as a means for a regular check on
aerodynamic size-separating performance of the component stages and exploring methods of collection surface coating,
eliminating leakage into the apparatus and setting the volumetric flow rate correctly.
The purpose of GCIP is therefore:
1. To ensure appropriate size-fractionation of the inhaled product;
2. To establish the APSD profile of the delivered dose reliably, minimizing sources of potential bias and optimizing
measurement precision;
3. To obtain adequate resolution of the fine particle mass component of the delivered dose;
https://online.uspnf.com/uspnf/document/1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US 1/13
Printed on: Mon Mar 21 2022, 08:10:54 AM(EST) Official Status: Currently Official on 21-Mar-2022 DocId: 1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US
Printed by: Titon Saha Official Date: Official as of 01-May-2021 Document Type: GENERAL CHAPTER @2022 USPC
DOI Ref: k9w4a DOI: https://doi.org/10.31003/USPNF_M13055_03_01
2
4. To establish the precision of manufacture of impactor apparatuses and accessories with consideration for allowed
tolerances and material of construction; including information on calibration, verification, and mensuration-related issues
of the apparatus upon receipt and after use.
The rationale for the chapter is to guide both new and existing cascade impactor equipment users as to industry best practices
in connection with set-up, use, maintenance, and storage between uses. The ultimate goal is to establish and maintain the
capability to determine the APSD and quality consistency of drug product batches.
al
cascade impactor and the associated equipment is in service, and lastly advising on checks that should be made to provide
assurance of component interchangeability following appropriate component validation procedure(s) (for example, the
use of a common PS with several different impactors or the use of the same collection plate or cup in several different
impactors of the same type).
This chapter is not intended to provide a line-by-line guide to APSD measurement, as some of this information is provided
in á601ñ, á1601ñ, and á1602ñ for the various inhaler product types. Neither does it provide detailed procedures, because in many
ci
cases, such information will be product-specific. Rather, the information presented is meant to summarize current thinking as
to the implementation of best practices in cascade impactor maintenance and use in terms of robust protocols for the laboratory
making these measurements. When developing such protocols, users are also advised to consult with current regulatory
guidance documents that address inhaler APSD measurement, for example, those published from time to time by the FDA.
ffi
1.3 Specific Information Regarding Impactor Stage Mensuration
In 4.1 Stage Mensuration, Measurement Traceability, and Mensuration Interval, information regarding the stage optical
mensuration of nozzles associated with the sizing stages of cascade impactors is given to aid the user in identifying the critical
aspects of the process and how mensuration provides traceability of the measures of size ultimately to the international length
O
standard via intermediate calibrated objects, such as optical reticles or orifices of known diameter, having a circular profile.
However, since specialized optical inspection equipment is required to acquire the mensuration data, it is assumed likely that
the user will employ a specialized service in order to undertake the task, so detailed methodology is not provided.
Aerodynamic diameter (Dae): The aerodynamic diameter of an irregular particle is defined as the diameter of the spherical
particle with a density of water [1000 kg/m3 (ρ0)] and having the same settling velocity under the influence of gravity as
the irregular particle. Dae is related to the microscopy-measured physical diameter (volume equivalent diameter, Dp) in
accordance with:
�� 1/2
DaeCae = � � DρCρ [1]
0
Cae and Cp = slip correction factors corresponding to Dae and Dp, respectively, both of which can be regarded as being close
to unity for the sizes of aerosol particles from inhalers for which APSD is being determined
ρp = particle density
χ = dynamic shape factor1 (1.00 for spherical solid particles or liquid droplets, >1.00 for non-spherical particles)
Effective diameter (Weff): The representative diameter for a collection of nozzles forming a multi-nozzle stage of a
multi-stage cascade impactor. Effective diameter is defined analytically as the product of the area mean diameter (W*)
raised to the power of 2/3 and the area median diameter (Wmedian) raised to the power of 1/3:
1 Hinds W.C. 1999 “Properties, Behavior and Measurement of Airborne Particles.” Wiley Interscience, NY 1999.
https://online.uspnf.com/uspnf/document/1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US 2/13
Printed on: Mon Mar 21 2022, 08:10:54 AM(EST) Official Status: Currently Official on 21-Mar-2022 DocId: 1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US
Printed by: Titon Saha Official Date: Official as of 01-May-2021 Document Type: GENERAL CHAPTER @2022 USPC
DOI Ref: k9w4a DOI: https://doi.org/10.31003/USPNF_M13055_03_01
3
For a single-nozzle stage, Weff defaults to the mensurated diameter of that nozzle.
Stage cut-point size (d50): The size on the aerodynamic diameter scale at which a given impactor stage collects incoming
particles with 50% efficiency. The relationship between collection efficiency and Dae is described further in 3. Cascade
Impactor Operating Principles.
Stokes number (St): The dimensionless term characterizing the behavior of solid or liquid particles suspended in a fluid
flow, such as air. A particle with a low Stokes number follows the air streamlines. In contrast, a particle possessing a large
Stokes number is dominated by its inertia and so that it continues along its initial trajectory when the flow passes an
obstruction, for example, as it diverges upstream of the collection surface of an impactor stage (see Figure 1). St50 is the
Stokes number corresponding to the cut-point size at which the impactor stage is 50% efficient at collecting the incoming
particles.
Stage mensuration: The process of establishing the representative diameter at the exit of each nozzle of a stage forming
part of a multi-stage impactor. Although pin or go/no gauges can be used to mechanically measure nozzles having
diameters in excess of about 2 mm, it is more common currently to make the determination non-invasively by using
automated optical microscopy combined with image analysis.
A cascade impactor size-fractionates incoming aerosol particles into several sub-fractions on the basis of their differing
inertia. A typical design, such as the apparatuses described in á601ñ, comprises several consecutive stages, each of which
functions as an inertial size-separator or fractionator of the incoming aerosol in a gas stream moving at constant velocity (flow
rate). The basic component is a single stage that comprises a nozzle plate containing one or more circular orifices located a
al
fixed distance from an impaction plate/cup surface that is usually horizontal (see Figure 1). The following characteristic
dimensions describe the operating property of the stage:
Weff: effective nozzle diameter—most stages will contain n nozzles machined to be as close to identical as possible
T: thickness of the nozzle plate
S:
ci
separation distance between exit plane of nozzle plate and the collection surface
ffi
O
Figure 1. Generic single-stage impactor showing nozzle plate and collection surface located immediately below the plane of
the nozzle exit. Several consecutive stages connected together with nozzle sizes that progressively get smaller constitutes a
cascade impactor.
The stage functions by fractionating incoming particles of various sizes on the basis of their differing inertia, the magnitude
of which reflects the resistance to a change in direction of the laminar flow streamlines passing through each nozzle. As the
incoming flow passes through the nozzle plate, the streamlines diverge on approach to the collection surface, whereas the finite
inertia of the particles causes them to cross the streamlines. Depending upon how far a given particle diverges from the
streamlines, its trajectory may cause it to collide (impact) with the collection surface located at a distance, S, close to the exit
plane of the nozzle set.
The magnitude of T is set by the manufacturer during design of the impactor stage, and in general, does not change
measurably in use. In contrast, the size of S can be reduced by coating of the collection surface to mitigate particle bounce (see
5.4 Mitigation of Particle Bounce and Re-Entrainment), as well as by the accumulation of deposited particles on the collection
surface. The ratio, S/Weff, can vary widely in the range 1–10 for effective size fractionation to occur, and therefore changes in
this ratio are unlikely to influence stage performance, given typical mass loadings associated with inhaler APSD measurements,
even on the stages where most of the particulate deposits.
The dimensionless Stokes number (St), which is the ratio of the stopping distance of a particle to a characteristic dimension,
in this case the nozzle diameter, Weff, describes the size fractionation process by defining a critical particle size that will reach
https://online.uspnf.com/uspnf/document/1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US 3/13
Printed on: Mon Mar 21 2022, 08:10:54 AM(EST) Official Status: Currently Official on 21-Mar-2022 DocId: 1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US
Printed by: Titon Saha Official Date: Official as of 01-May-2021 Document Type: GENERAL CHAPTER @2022 USPC
DOI Ref: k9w4a DOI: https://doi.org/10.31003/USPNF_M13055_03_01
4
the collection surface for a particular stage geometry. St is defined in terms of the stage effective diameter by the following
relationship:
���ae�2
���
St = 9�� [3]
���
In a cascade impactor, several stages are connected together such that u progressively increases through the cascade by
reducing Weff (see Figure 2).
al
ci
ffi
O
Figure 2. Conceptual multi-stage cascade impactor—Typically there are 7 or 8 stages from the entry to the back-up filter or
micro-orifice collector (MOC) in the case of the Next Generation Impactor; the function of the filter/MOC is to capture any
particles that have penetrated beyond the lowermost stage ‘N’.
Equation 4 predicts that the particle collection efficiency of an ideal impactor stage, Estage, expressed as a percentage, will
increase in a step-wise manner between limits of 0%–100%, as dae increases. In practice, for a well-designed stage, Estage is a
monotonic sigmoidal function of dae that increases steeply from E ≈ 0% to >95%, reaching its maximum steepness when E stage
is 50% (see Figure 3). At this location, defined as the stage cut-off diameter (d50), this equation also defines the relationship
between d50, St50, and volumetric flow rate (Q) for a multi-orifice stage comprising n circular nozzles:
9������� 1/2
d50 = 4�0�
��50 [4]
��50 is close to 0.49 at d50 for all stages of each cascade impactor type described in á601ñ.
https://online.uspnf.com/uspnf/document/1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US 4/13
Printed on: Mon Mar 21 2022, 08:10:54 AM(EST) Official Status: Currently Official on 21-Mar-2022 DocId: 1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US
Printed by: Titon Saha Official Date: Official as of 01-May-2021 Document Type: GENERAL CHAPTER @2022 USPC
DOI Ref: k9w4a DOI: https://doi.org/10.31003/USPNF_M13055_03_01
5
al
ci
Figure 3. Generic impactor stage collection efficiency curve—The ideal step function at the size corresponding to the cut-off
diameter is shown together with the sigmoidal curve representing normal behavior.
In a well-designed stage, the area ABC of Figure 3, representing the portion of the incoming aerosol contained in particles
ffi
larger than the cut-point size that should have been collected, but instead bypassed the collection plate/cup, equals the
corresponding area ADE that represents the portion of the aerosol contained in particles smaller than the cut-point size that
should have penetrated the stage, but instead were collected there. The result is that d50 for a given stage can be used to define
its size-fractionating performance as if it was ideal in this respect.
An important outcome arising from Equation 4 is that the cut-off diameters of all the stages in a given cascade impactor are
constant only when Q is also constant. It follows that for the testing of dry powder inhalers in accordance with the methodology
O
in á601ñ, the stage cut-points will not be at their final steady state values until the volumetric flow rate is constant after initiation
of sampling from no flow by opening the two-way solenoid valve located between the exit from the impactor and the vacuum
source. This process can take up to 400 ms for impactors having large internal volume, in particular, the Next Generation
Impactor (NGI) with its PS. In practice, this finite start-up time is ignored in such assessments on the basis that it represents a
small fraction of the total time during which the 4.0-L volume is sampled for inhalation powders. More importantly, however,
any leakage of ambient air into the impactor at locations other than via the induction port, will increase the local value of Q
after the leakage point, and in consequence result in an uncontrolled decrease in d50 for stages downstream. Bias arising from
leakage into the impactor is therefore discussed further in 6.5 Mitigation of Air Leakage into the Apparatus.
4. APPARATUS MAINTENANCE
https://online.uspnf.com/uspnf/document/1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US 5/13
Printed on: Mon Mar 21 2022, 08:10:54 AM(EST) Official Status: Currently Official on 21-Mar-2022 DocId: 1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US
Printed by: Titon Saha Official Date: Official as of 01-May-2021 Document Type: GENERAL CHAPTER @2022 USPC
DOI Ref: k9w4a DOI: https://doi.org/10.31003/USPNF_M13055_03_01
6
traceable ultimately to the international length standard. It is anticipated that the cascade impactor user will have the procedure
undertaken by an external provider, who will typically be the supplier of the impactor, so the focus here is on establishing a
regimen for maintaining assurance that any given stage of any cascade impactor remains within the manufacturer’s specification
as defined in á601ñ.
The scheme shown in Figure 4 illustrates the components of a mensuration protocol for initial receipt and subsequent
assessment of any of the impactors described in á601ñ. The following checks are recommended:
1. Installation qualification (IQ): At receipt of a new impactor, ensure, as part of IQ, that each stage is compliant with the
manufacturer’s dimensional specifications. These initial measurements are intended to establish baseline data for each
stage of each individual cascade impactor. Undertake mensuration for each and every nozzle of each stage and
determine its Weff. Compare this measurement against the critical dimensions specified for that stage/cascade impactor
in á601ñ. Reject any stage whose Weff is outside the specification. For the ACI, it may be appropriate to substitute the
equivalent rejected stage(s) with alternative stages of the same number(s), whose Weff value(s) are within specification.
However, in the case of the NGI, the entire unit will need to be returned to the supplier for remedial action to provide
stage(s) whose Weff values are within specification. Once values of Weff for all stages are ascertained to be within
specification, the entire cascade impactor can be accepted for use.
al
ci
ffi
O
Figure 4. Scheme for the application of stage mensuration to confirm suitability for use of a cascade impactor at initial
receipt and at subsequent performance assessment.
[NOTE—The frequency of the “In-use mensuration” check is determined by the amount of impactor use, the material
from which the impactor is constructed, the product ingredients, and recovery solvent characteristics.]
2. For stages that are deemed acceptable in terms of the specifications in á601ñ, establish the in-use stage mensuration
regimen, whose frequency is determined by the amount of impactor use and the product/active pharmaceutical
ingredient solvent chemical characteristics. Each in-use mensuration will always include measuring the individual stage
nozzle diameters, determining Weff as the performance metric. The specifications established at initial acceptance will
normally be used as benchmarks to judge whether the stage(s) may continue in service or require remediation or
replacement. Note that remediation could be as simple as careful cleaning to remove deposits that may be plugging
nozzles. Acquisition of data from several subsequent mensuration exercises after exposure to the product(s) being
evaluated in the typical testing environment may also provide the opportunity to develop meaningful in-service “at
https://online.uspnf.com/uspnf/document/1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US 6/13
Printed on: Mon Mar 21 2022, 08:10:54 AM(EST) Official Status: Currently Official on 21-Mar-2022 DocId: 1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US
Printed by: Titon Saha Official Date: Official as of 01-May-2021 Document Type: GENERAL CHAPTER @2022 USPC
DOI Ref: k9w4a DOI: https://doi.org/10.31003/USPNF_M13055_03_01
7
mensuration” limits for Weff for a given stage that will likely be chosen to be tighter than those used to begin with. These
internal limits can subsequently be used to indicate the need for remediation before the stage Weff breaches either the
upper or lower boundary of the tolerance range provided in á601ñ.
3. Following mensuration failure of an impactor, undertake the following:
A. In the case of the ACI, remove those stages from service that are not acceptable in terms of the manufacturer’s
specifications, and either recondition them or replace the affected stages altogether.
B. In the case of an NGI, the entire unit will need to be returned to the supplier for remedial action, if Weff for one or
more stages is/are found to be out of specification.
4. Performance qualification (PQ): Establish the frequency between mensuration events. Although annual evaluation will
be sufficient for most users, the duration between mensuration exercises should ideally be governed by the frequency
of cascade impactor use, the type of compounds being collected, and the solvents used to recover the collected active
pharmaceutical ingredient(s) particulate matter as well as the end user established cleaning procedures.
5. Ensure that other non-nozzle dimensional data for the cascade impactor apparatus are checked and confirmed as being
acceptable through a mensuration or appropriate visual inspection process. Guidance on the important dimensions for
ancillary equipment, such as the induction port and PS (if used), has been provided by Nichols, et al.2 In the unlikely
event that any of the critical dimensions (e.g., entry and exit diameters for the induction port, or nozzle diameters for
the PS) are deemed to be unacceptable, remedial action may be as simple as cleaning the affected component. Under
such circumstances, discussion with the manufacturer to establish the most appropriate procedure beforehand is strongly
recommended. However, if remedial action does not resolve the problem, then the individual component should be
replaced before the ancillary apparatus is placed back into service.
6. If any impactor stage or item of ancillary equipment fails specification, establish an appropriate remediation strategy,
al
most likely involving the supplier/manufacturer of the affected equipment.
In the initial development of a method for a new drug product or when a new design of cascade impactor is placed in service,
it is recommended that internal surfaces of the induction port, PS (if used) and cascade impactor (entrance cone and nozzle
plates for the ACI or inter-stage passageways for the NGI) be washed from both each other and from the collection surfaces
separately with the established minimum necessary volume of recovery solvent to determine the magnitude of the associated
internal losses as well as their location. In the product development stage, if the overall losses in APSD determination exceed
5% of the mass balance, and steps to mitigate particle bounce in the cascade impactor (see 5.4 Mitigation of Particle Bounce
and Re-Entrainment) and electrostatic charge (see 5.5 Mitigation of Electrostatic Charge Accumulation) prove ineffective at
reducing such losses, the user may wish to consider an alternative cascade impactor (e.g., replace the ACI with the NGI) to
investigate if losses can be reduced below 5%. If this outcome is not possible, the inhaler aerosol APSD measurement procedure
can be performed including the total internal wall losses in the assessment of mass balance to confirm that the mass recovery
is within the recommended range of label claim for mass of active pharmaceutical ingredient delivered per actuation. In defining
the ASPD profile, it is important to note that it is not possible to apportion wall losses from stage to stage within the cascade
impactor in an attempt to assign specific size ranges to these losses. The wall loss deposition processes lack the requisite size
selectivity associated with each impaction stage.
2 NicholsSC, Mitchell JP, Shelton CM, Roberts DL. Good cascade impactor practice (GCIP) and considerations for “in-use” specifications. AAPS
PharmSciTech. 2013;14(1):375–390.
https://online.uspnf.com/uspnf/document/1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US 7/13
Printed on: Mon Mar 21 2022, 08:10:54 AM(EST) Official Status: Currently Official on 21-Mar-2022 DocId: 1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US
Printed by: Titon Saha Official Date: Official as of 01-May-2021 Document Type: GENERAL CHAPTER @2022 USPC
DOI Ref: k9w4a DOI: https://doi.org/10.31003/USPNF_M13055_03_01
8
al
subsequent assay for the active pharmaceutical ingredient are often limiting factors.
further into the impactor. However, for inhalation powders, coating of collection surfaces other than the filter/MOC will almost
certainly be necessary. Such pre-coating is normally used for inhalation powders where the incoming drug particles are
accompanied by larger carrier particles.
Particle bounce and associated re-entrainment is easily diagnosed by the presence of a larger than expected mass of drug
recovered from the back-up filter/MOC, because once re-entrained, the particles have sufficient velocity to bounce off the
collection surfaces of subsequent stages. In method development, it is therefore recommended that evaluation of such behavior
be undertaken in a validation study with and without collection surface coating present.
There are numerous methods for collection surface coating, all of which involve the deposition of a thin layer of a viscous
substance, such as glycerol, silicone oil, or other suitable substance, such as the surfactant, polyoxyethylene lauryl ether.3 The
coating substance is typically prepared as a solution in a volatile solvent which subsequently evaporates, leaving behind a
uniform deposit. The choice of coating material will depend upon the physicochemical properties of the drug product being
sampled as well as the subsequent recovery and analytical procedure. Whatever method is chosen for applying the coating
substance, each collection surface should be visually inspected for non-uniformities in the coating, and the procedure repeated
after cleaning the collection surface, if imperfections are found.
The interior surfaces of the induction port are not normally coated in the procedures described in á601ñ. However, an
appropriate pre-defined volume of an appropriate solvent may be dispensed when the PS is used (up to 10 mL to coat the PS
interior used with the ACI or up to 15 mL to the central cup of the PS used with the NGI). Alternatively, it may be appropriate
to coat the particle collection surfaces with a viscous substance as described for the stage collection surfaces in order to prevent
re-entrainment of impacted particles therein.
3 Polyoxyethylene-23-lauryl ether.
https://online.uspnf.com/uspnf/document/1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US 8/13
Printed on: Mon Mar 21 2022, 08:10:54 AM(EST) Official Status: Currently Official on 21-Mar-2022 DocId: 1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US
Printed by: Titon Saha Official Date: Official as of 01-May-2021 Document Type: GENERAL CHAPTER @2022 USPC
DOI Ref: k9w4a DOI: https://doi.org/10.31003/USPNF_M13055_03_01
9
Appropriate setup and handling of the inhaler as well as the entire measurement apparatus can significantly reduce variability
in test results caused by electrostatics, and practical steps include grounding the measurement equipment as well as the operator
(typically via use of a wrist strap and an antistatic laboratory coat). Controlling the local environmental conditions such that
testing is undertaken at values of relative humidity in the range 40%–60% is also recommended to avoid variability from
electrostatic effects.
Change to read:
6. IN-USE ASPECTS
al
Regardless of incidents involving accidental damage, a regimen should be established for periodic visual inspection of all
cascade impactor apparatus components, including the induction port and PS (if used), connection tubing, and related flow
control equipment. This physical examination is in addition to the stage mensuration of the nozzles of individual stages
described in 4.1 Stage Mensuration, Measurement Traceability, and Mensuration Interval. The frequency of such inspections will
depend upon the amount of use the components receive, as well as experience gained with time in service; however, as a start
an annual inspection is recommended. Inspected components that are deemed damaged or marginally functional should be
ci
identified as such and removed from service. Visual inspection of the seal body and inter-stage passageways of the NGI can be
difficult, and may therefore best be undertaken by the supplier/manufacturer. The component should be replaced or repaired
to the manufacturer’s specification, if any damage/deformation is apparent.
The external ‘O’-ring seal associated with each stage of the ACI is particularly susceptible to crack formation in normal use.
Each seal should therefore be visually inspected for cracking, scratch marks, and gross deformation on each occasion of use and
ffi
replaced if damaged. The corresponding elastomer rings located in the seal body of the NGI should likewise be visually inspected
before use and replaced if damaged.
Coating of the stage collection plates (ACI) or cups (NGI) may be necessary to mitigate bias in the APSD measurement caused
by particle bounce and re-entrainment (see 5.4 Mitigation of Particle Bounce and Re-Entrainment). Since the coating procedure
can be highly operator dependent, even if semi-automated, once a suitable coating method has been established and validated,
each coated collection plate or cup should be visually inspected before assembly into the impactor to be sure that the applied
coating is uniform and that no locations are apparent that remain uncoated.
https://online.uspnf.com/uspnf/document/1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US 9/13
Printed on: Mon Mar 21 2022, 08:10:54 AM(EST) Official Status: Currently Official on 21-Mar-2022 DocId: 1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US
Printed by: Titon Saha Official Date: Official as of 01-May-2021 Document Type: GENERAL CHAPTER @2022 USPC
DOI Ref: k9w4a DOI: https://doi.org/10.31003/USPNF_M13055_03_01
10
the corresponding nozzle plate. Attach the inlet cone to the uppermost nozzle plate if an inhalation aerosol or spray is being
tested, or the PS, assembled in accordance with the instructions in á601ñ, if the impactor is being used to assess an inhalation
powder, unless justified by data that PS is not necessary. Finally, secure the induction port to the entry of the inlet cone
(inhalation aerosol or spray assessment) or PS (inhalation powder assessment) to complete the assembly of the impactor stack.
Table 1. The Three Configurations of the Andersen Cascade Impactor—Configuration A is Used at 28.3 L/min for the
Evaluation of Inhalation Aerosols and Sprays; Configurations A, B, or C are Used for Evaluation of Inhalation Powders
Depending Upon the Target Flow Rate (28.3, 60, or 90 L/min, Respectively)
Stage Configuration
A B C
28.3 L/min 60 L/min 90 L/min
0 –1 –2
1 –0 –1
2 1 –0
3 2 1
4 3 2
5 4 3
6 5 4
al
7 6 5
[NOTE—Stage ‘0’ is modified to accept stage ‘−1’ for operation at 60 L/min or stages ‘−1’ and ‘−2’ together for operation at
ci
90 L/min. In either case, the inlet cone is replaced by the modified stage ‘0’ and used above either stage ‘−1’ or stage ‘−2’
depending upon the chosen flow rate.]
determination, since their order with respect to the stage number is easily preserved after each measurement by removing the
tray and cups as a single unit from the bottom frame for transfer to the station where drug recovery takes place. Alternatively,
multiple cup-sets may be used with a single cup tray or each cup set can be assigned to a particular cup tray, depending on
the number of APSD determinations that need to be made per day, as well as the operating regimen in force within the
laboratory. For instance, assigning cup sets to a particular tray is useful if separate operators are responsible for the APSD
determination and subsequent drug recovery/assay processes. Nevertheless, it is advisable to mark the base of each cup so they
are identified as a single set.
In typical use, following completion of the collection cup coating procedure (see 5.4 Mitigation of Particle Bounce and
Re-Entrainment) and inspection of the seals in the seal body, each cup is located in the appropriate order in the cup tray. The
MOC is normally present to capture any particles that penetrate beyond stage 7. However, for some inhaled products (i.e.,
inhalation aerosols dispersing a plume in which a significant portion of the particles have sizes finer than about 1.0 μm in
aerodynamic diameter), the MOC is replaced by either an internal filter and/or a separate external filter assembly which can be
inserted between the impactor body and the vacuum source. Next, insert the cup tray into the impactor, such that the cup
bases fit into the corresponding spaces of the bottom frame. Close and secure the seal body containing the stages nozzle plates
over the cup tray using the external lever located at the front of the impactor. Insert the tapered fitting of the induction port
to the entry port located on the uppermost surface of the seal body if an inhalation aerosol or spray is being tested. Alternatively,
insert the tapered fitting of the PS, assembled in accordance with the instructions in á601ñ, if the impactor is being used to
assess an inhalation powder, unless justified by data that PS is not necessary, and secure the induction port to the entry of the
PS to complete the assembly of the impactor.
https://online.uspnf.com/uspnf/document/1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US 10/13
Printed on: Mon Mar 21 2022, 08:10:54 AM(EST) Official Status: Currently Official on 21-Mar-2022 DocId: 1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US
Printed by: Titon Saha Official Date: Official as of 01-May-2021 Document Type: GENERAL CHAPTER @2022 USPC
DOI Ref: k9w4a DOI: https://doi.org/10.31003/USPNF_M13055_03_01
11
Leakage testing is most conveniently undertaken immediately following assembly (see 6.4 Assertion of Correct Assembly) by
connecting the set-up, including the induction port, PS (if used), and external filter (if used with the NGI), to the vacuum source.
This procedure also serves to check the leakage integrity of the entire set-up including tubing and flow control equipment
before use. The entry of the induction port is temporarily plugged with an elastomeric bung for the duration of the test. Figure
5 illustrates the configuration for the ACI without PS, but the external components, if equipped with its PS or for the NGI with/
without PS, are similar. A side-connection via a gate valve (G2) is made to a sensitive differential pressure transducer (0–50 kPa
and capable of determining a change of ±0.5 kPa at 20.0 kPa) that is located immediately after the exit port from the cascade
impactor or the external filter assembly if the NGI is used with this additional equipment. Another gate valve (G1) is located
between the side-connection and the variable flow control valve that, in turn, is connected to the vacuum source/pump.
al
ci
ffi
Figure 5. Set-up for leak testing impactor and associated equipment before use—the Andersen Cascade Impactor is shown
O
without its PS, but the set-up is similar if the PS is present or the Next Generation Impactor is substituted.
[NOTE—The system is shown for inhalation aerosols, but the same procedure will work by substituting the appropriate flow
control equipment for testing inhalation powders in place of the flow control valve.]
The following procedure is then followed:
1. With the flow control valve closed, the gate valves, G1 and G2, are opened, and the vacuum source is activated;
2. A partial vacuum is applied to the test system by adjusting the flow control valve to set P0 to 20.0 ± 0.5 kPa observing
the reading of the output from the differential pressure meter. The differential pressure is measured by connecting the
differential pressure transducer at its negative pressure port to the side-connection from the test system;
3. After the required vacuum is displayed on the meter of the pressure gauge, the gate valve, G1, is fully closed to isolate
the test apparatus from the vacuum source, and a timer is started;
4. The initial differential pressure (P0) is documented;
5. Allow at least 1 min to elapse, then observe and record the displayed pressure (P1);
6. Carefully remove the rubber bung from the inlet, ensuring not to disturb the apparatus, close the gate valve, G2, and
open the gate valve, G1, to isolate the differential pressure meter and to bring the entire test system back to atmospheric
pressure conditions, ready for next use;
7. The vacuum source can be deactivated at this stage;
8. The leak rate (L) is calculated in accordance with Equation 5:
�0 − �1
L= �
[5]
Use the leak test specification from the manufacturer. If this is unavailable, then L should be ≤10 kPa/min.
https://online.uspnf.com/uspnf/document/1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US 11/13
Printed on: Mon Mar 21 2022, 08:10:54 AM(EST) Official Status: Currently Official on 21-Mar-2022 DocId: 1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US
Printed by: Titon Saha Official Date: Official as of 01-May-2021 Document Type: GENERAL CHAPTER @2022 USPC
DOI Ref: k9w4a DOI: https://doi.org/10.31003/USPNF_M13055_03_01
12
If L exceeds this limit, examine the components of the assembly for leakage(s). Begin by inspecting the seals associated with
the stages, then the connections from the induction port to PS (if used) or to cascade impactor, and finally from the impactor
to the vacuum source. Replace defective components as necessary and reassemble. Repeat the leak test above until L is lower
than the specified limit.
al
ci
ffi
O
Figure 6. Flow rate setting for (A) Andersen Cascade Impactor with PS, (B) Next Generation Impactor with PS; the flow sensor
location at the entry to the induction port remains the same if the PS is not used.
Set the volumetric flow rate to the target value for the inhaler being tested immediately before making the APSD
determination. Note that a low flow resistance mass flowmeter may be used to make this measurement, as the air pressure at
the point of flow measurement is close to room ambient.
6.8 Assertion that Individual Cascade Impactor Assemblies of the Same Type (i.e., Andersen
Cascade Impactor or Next Generation Impactor) are Interchangeable
In many laboratories, there are likely to be many cascade impactor assembles of the same type in service. There may therefore
be a desire to interchange different impactors when making APSD performance measurements on the same inhaler product in
order to speed up the process. Provided that stage mensuration (see 4.1 Stage Mensuration, Measurement Traceability, and
Mensuration Interval) has demonstrated that Weff, determined as the representative measure of the diameter of the nozzles of
each stage is within the manufacturer’s specification, as provided in á601ñ, Table 4 (ACI) and á601ñ, ▲Table 6▲ (CN 1-May-2021)
(NGI), and checks have been made to ensure that seals are intact (see 6.2 Inspection of Cascade Impactor Components Susceptible
to Deterioration), and other considerations as discussed in 4. Apparatus Maintenance. The impactors may be interchangeable.
https://online.uspnf.com/uspnf/document/1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US 12/13
Printed on: Mon Mar 21 2022, 08:10:54 AM(EST) Official Status: Currently Official on 21-Mar-2022 DocId: 1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US
Printed by: Titon Saha Official Date: Official as of 01-May-2021 Document Type: GENERAL CHAPTER @2022 USPC
DOI Ref: k9w4a DOI: https://doi.org/10.31003/USPNF_M13055_03_01
13
However, it is advisable to perform comparative testing to validate equivalence. Nevertheless, it is recommended that the stages
together with their collection surfaces are uniquely identified to a particular cascade impactor assembly for the purpose of
traceability as to the source for each APSD determination.
A similar consideration applies to the PS used with the NGI, because there is a specification for the nozzle diameters in
á601ñ, Table 6. The situation is less clear for the PS for the ACI, although a manufacturer-supplied specification for the diameter
of the three exit tubes, where available, would fulfill the same purpose as the PS specification for the NGI PS.
The specifications for the critical dimensions (entry and exit tube internal diameters) of the induction port are provided in
á601ñ, ▲Figure 6b▲ (CN 1-May-2021) and should be used to check for the interchangeability of this component, where multiple
induction ports are in service.
al
ci
ffi
O
https://online.uspnf.com/uspnf/document/1_GUID-7054DEA7-8615-44AE-B612-960462CB4BCC_3_en-US 13/13