7''Square Hotplate

User Manual

HP550-S LED Digital 7''Square Hotplate

Please read the User Manual carefully before use, and follow
all operating and safety instructions!
Technical specifications and outline are subject to change
without prior notice.
Contents
Preface ................................................................................. 1
Service ................................................................................. 1
Warranty .............................................................................. 1
1. Safety Instructions ....................................................... 2
2. Proper use .................................................................... 3
3. Inspection .................................................................... 3
3.1 Receiving Inspection ............................................. 3
3.2 Listing of Items ..................................................... 3
4. Control......................................................................... 4
4.1 Control elements.................................................... 4
4.2 Display................................................................... 5
5. Trial run ....................................................................... 5
6. Working with external temperature sensor .................. 6
7. Residual heat warning ................................................. 6
8. Faults ........................................................................... 7
9. Maintenance and Cleaning .......................................... 7
10. Associated standards and regulations .......................... 8
11. Specifications .............................................................. 9
12. Ordering information................................................. 10
Preface Please provide the customer care representative with the
following information:
Welcome to the 7''Square Hotplate User Manual. . Users · Serianlumber(on the rear panel)
should read this Manual carefully, follow the instructions · Certification
and procedures, and be aware of all the cautions when using · Description porfoblem (i.e., hardware or software)
this instrument. · Methods anprdocedures adopted to resolve the problems
· Your contact information
Service
When help needed, you can always contact the service
department of manufacturer for technical support in the Warranty
following ways: This instrument is warranted to be free from defects in
materials and workmanship under normal use and service,
for a period of 24 months from the date of invoice. The
ONiLAB LLC. warranty is extended only to the original purchaser. It shall
not apply to any product or parts which have been damaged
25415 Prado De Las Peras, Calabasas, CA 91302,USA
on account of improper installation, improper connections,
Tel: +1-818-318-3771 misuse, accident or abnormal conditions of operation.
For claims under the warranty please contact your local
supplier. You may also send the instrument directly to
manufacturer, enclosing the invoice copy and by giving
reasons for the claim.

1
substances or under water.
1. Safety Instructions • Temperature must always be set to at least 50℃ lower
Warning! than the fire point of the media used.
• Read the operating instructions carefully • Be aware of hazards due to unsafe container.
before use. • Process pathogenic materials only in closed vessels.
• Ensure that only trained staff works with the • Check the instrument and accessories prior to each use.
instrument. Do not use damaged components. Safe operation is only
Risk of burn! guaranteed with the accessories described in the
• Caution when touch the housing parts and “Accessories” chapter. Accessories must be securely
the hotplate which can reach temperature of attached to the device and can not come off by themselves.
550° C. Always disconnect the plug before fitting accessories.
• Pay attention to the residual heat after • When the external temperature sensor needed, the tip of
switching off. the measuring sensor must be at least 5-10mm from vessel
Protective ground contact! bottom and wall.
• Make sure that socket must be grounded • The instrument can only be disconnected from the main
(protective ground contact) before use. power supply by pulling out the main or the connector plug.
• The voltage stated on the label must correspond to the
• When working wear personal safety guards to avoid the main power supply.
risk from: • Ensure that the main power supply cable does not touch
- Release of toxic or combustible gases the hotplate. Do not cover the device.
• Set up the instrument in a spacious are on a stable, clean, • Forbid to put pressure and over heat media on the surface
non-slip, dry and fireproof surface. Do not operate the of glass ceramic, that can be caused surface broken.
instrument in explosive atmospheres, with hazardous • The instrument may only be opened by experts.

2
2. Proper use 3. Inspection
The instrument is designed for mixing and/or heating 3.1 Receiving Inspection
liquids in schools, laboratories or factories.
Unpack the equipment carefully and check for any damages
• Observe the minimum distances between the devices,
which may have arisen during transport. Please contact
between the device and the wall and above the assembly
manufacturer/supplier for technical support.
(min. 100 mm)

Note:
If there is any apparent damage to the system,
>100mm please do not plug it into the power line.

3.2 Listing of Items


>100mm The package includes the following items:
Items Qty
Main unit 1
Power cable 1
User Manual 1
Table 1
Figure 1
This device is not suitable for using in residential areas or
other constraints mentioned in Chapter 1.

3
4. Control Items Descriptions
Temperature The heating function is switched
4.1 Control elements control knob ON or OFF by rotating knob.
If rotate the heating knob, LED
displays the temperature setting
LED Display
value and shift to real value in the
duration of 5 seconds.
When the heating function is
LED Heat
switched ON, the LED Heat flash.
LED Heat
LED Display Power
Switch ON or OFF the instrument.
Temperature Control Knob Switch
Table 2
Figure 3

Power Switch

Figure 4

4
4.2 Display 5. Trial run
• Make sure the required operating voltage and power
Display area supply voltage match.
• Ensure the socket must be properly grounded.
• Plug in the power cable,ensure the power is on and begin
Figure 5
initializing.
Display Descriptions • Place vessel on the work plate.
When the heating function is switched ON and • Set the target temperature and start heating.
rotate the stirring knob, LED displays the • Stop the heating functions.
temperature setting value and shifts to real value
Display If these operations above are normal, the device is ready to
in 5 seconds.
area operate. If these operations are not normal, the device may
When the heating function is switched OFF and
the hotplate temperature is still above 50℃, be damaged during transportation, please contact
LED displays Hot, otherwise LED displays 0. manufacturer/supplier for technical support.

5
6. Working with external
temperature sensor

Figure 6
The external temperature sensor PT1000 is the
Manufacture’s standard accessory. If the sensor is plugged
in and rotate the heating knob, LED displays the
temperature setting value and shifts to real value in 5
seconds. Safe circuit controls hotplate temperature.
Comparing with the temperature control of the hotplate, the
external temperature sensor can control the medium’s
6
temperature more precisely. The heating function will be
stopped automatically under abnormal conditions. Please
operate follow the instructions below:
• Switch OFF the instrument.
• Ensure the external temperature sensors is inserted in
the media heated.
• Switch ON the instrument and run heating function.
If the heating function did not work, please contact
manufacturer/supplier for technical support.

7. Residual heat warning


In order to prevent the risk of burns from a hotplate, digital
hotplate has a residual heat warning function. When the
heating function is switched off and the heating plate
temperature is still above 50°C, “Hot” will flash to warn
that there is a hazard of burns from the hotplate. When the
hotplate temperature drops to below 50° C, the unit will
automatically switch off. If users want to turn off the LCD
or LED immediately, just pull out the plug directly. When
the plug is pulled out, the residual heat warning function
cannot be run.

7
8. Faults Isopropyl alcohol
Foodstuffs Water containing tenside
• Instruments can’t be power ON
Fuels Water containing tenside
- Check whether the power line is unplugged
• Wear the proper protective gloves during cleaning of
- Check whether the fuse is broken or loose
the instrument.
• Fault in power on self test
• Before using other method for cleaning or
- Switch OFF the unit, then switch ON and reset the
decontamination, the user must ascertain with the
instruments to factory default setting.
manufacturer that this method will not harm the
• Unit cannot be powered off when switched off.
instrument.
If these faults are not resolved, please contact
• Send in the case of service the instrument back in the
manufacturer/supplier.
packaging carton. Storage packing is not sufficient for
9. Maintenance and Cleaning the back dispatch. Use additionally a suitable
transportation packing.
• Proper maintenance can keep instruments working
• The enamel makes the hotplate easier to care for and
properly and lengthen its lifetime.
more resistant to acids and bases. Because of it,
• Do not spray cleanser into the instrument when
however, the hotplate is also more susceptible to
cleaning.
extreme fluctuations in temperature and the force of
• Unplug the power line when cleaning.
impact. This can result in cracks forming or the
• Only use recommended cleansers:
coating flaking off.
Dyes Isopropyl alcohol
Warning!
Construction materials Water containing tenside /
Cut off power when maintenance and cleaning.
Isopropyl alcohol
Cosmetics Water containing tenside /

8
10. Associated standards and commercial environment. This equipment generates, uses,
and can radiate radio frequency energy and, if not installed
regulations and used in accordance with the instruction manual, may
Construction in accordance with the following safety cause harmful interference to radio communications.
standards: Operation of this equipment in a residential area is likely to
EN 61010-1 cause harmful interference in which case the user will be
UL 3101-1 required to correct the interference at his own expense.
CAN/CSA C22.2(1010-1)
EN 61010-2-10
Construction in accordance with the following EMC
standards:
EN 61326-1
Associated EU guidelines: EMC-
guidelines: 89/336/EWG Instrument
guidelines: 73/023/EWG
Changes or modifications not expressly approved by the
party responsible for compliance could void the user's
authority to operate the equipment.
NOTE: This equipment has been tested and found to
comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful
interference when the equipment is operated in a
9
11. Specifications
Items Specifications *Temperature sensor in medium PT1000
Voltage [VAC] 100-120/200-240 *Control accuracy of heating
temperature with temperature ±0.5
Frequency [Hz] 50/60
sensor [℃]
Power [W] 1010
*Residual heat warning 50℃
Plate material Glass ceramic
Dimensions (mm) 215×360×112
Dimensions of workplate (mm) 184×184
Weight [kg] 3.8
*Heating power [W] 1000
Permitted ambient temperature [°
C] 5-40
RT-550, increment:
*Temperature range [°
C] Permitted relative humidity 80%
5
*Temperature display [°
C] LED Protection class acc. to DIN 60529 IP21

*Temperature display accuracy [°


C] ±1 Table 3
*Control accuracy of heating ±10
temperature [°
C]
*The safety temperature of the
hotplate [℃] 580

10
12. Ordering information
Accessories
PT1000-A Temperature sensor for digital
18900016
hotplate model, length of 230mm
PT1000-B Glass coated temperature
18900136 sensor for digital hotplate model, length
of 230mm
18900017 Support clamp of PT1000

1
1
ONiLAB LLC.
Add: 25415 Prado De Las Peras, Calabasas,

CA 91302,USA

Tel: +1-818-318-3771

12

You might also like