MassWorksQuickGuide
MassWorksQuickGuide
Apr. 2016
Read the instruction manual thoroughly before you use the product.
Keep this instruction manual for future reference.
This page is intentionally left blank.
Introduction
Notices
・ All rights are reserved, including those to reproduce this publication or parts thereof in any form
without permission in writing from Shimadzu Corporation.
・ Information in this publication is subject to change without notice and does not represent a
commitment on the part of the vendor.
・ Any errors or omissions which may have occurred in this publication despite the utmost care
taken in its production will be corrected as soon as possible, but not necessarily immediately
upon detection.
・ MassWorksTM are registered trademarks of cerno BIOSCIENCE in the United States and other
countries. Other company names and product names mentioned in this manual are trademarks
or registered trademarks of their respective companies.
・ Shimadzu Corporation and cerno BIOSCIENCE are not responsible for errors or injuries
resulting from following the instructions in this document.
・ MassWorks is provided on the "as is", and cerno BIOSCIENCE and Shimadzu Corporation
makes no warranty of any kind with respect to this product or information contained therein.
cerno BIOSCIENCE and Shimadzu Corporation does not warrant the accuracy or fitness for a
particular purpose.
・ In no event will cerno BIOSCIENCE and Shimadzu Corporation be liable for any damages
whether direct, indirect, special, incidental, consequential or punitive (including but not limited
to lose profits) arising out of or in connection with the use of this product. You have sole
responsibility for any consequences from using this product.
Scope of agreement
· The user can use this software only for a specified single computer.
· When the user intends to transfer all rights of this software, the user cannot reserve any
reproductions of this software. The user must transfer the entire software (including all
components, media and documents such as operation manual), and the transferor must agree
on this scope of agreement.
· If the software version has been upgraded, the transferred articles shall include the prior
versions of this software.
Copyright
・ The property right, copyright, and all intellectual property rights on this software are reserved
by Shimadzu Corporation or by a third party to be specified separately.
Indications Used in This Instruction Manual
Throughout the text in this manual, warnings and other information essential when using this unit,
such as cautionary or prohibited items, appear classified as per the following:
Mark Description
1 System Configuration
1.1 Overview
MassWorks enables highly-precise calculation of accurate mass and isotope spectra with a
unique mass calibration technology based on cerno BIOSCIENCE's patented MSIntegrity
technology to expand the capacity of unit mass resolution-base mass spectrometers. This
feature enables the elemental composition estimation that can be usually performed only on
expensive systems such as time-of-flight (TOF) mass spectrometers. This Quick Guide
describes operation procedures for using the software with a GCMS-QP Series or GCMS-TQ
Series system.
This Quick Guide assumes that the reader already has a basic understanding with
respect to the handling of GC-MS(/MS) and GCMSsolution. Names and terms specific
to GCMSsolution are used in many places in this guide and so, if you are a first-time
user or if there are any points that are not clear, refer to the GC-MS(/MS) instruction
manual, the operation guide and the GCMSsolution help as necessary.
If the GCMSsolution version is earlier than Ver. 4.42, upgrade the software.
1
1. System Configuration
2
2 Setting of GCMSsolution
2 Setting of GCMSsolution
Configure settings in GCMSsolution to output mass profile data to be used on the MassWorks.
1) Execute the [GCMS Real Time Analysis] program.
2) Select "Option" from the [Tools] menubar.
3) Click the [Profile Export] tab, Set "CRN format" to the [profile raw data is additionally stored].
3
3 GC/MS Data Acquisition
3.1 Auto-Tuning
1) Perform auto-tuning. Open [Tuning] window from the [GCMS Real Time Analysis]
program and click [New] icon.
3) In case of QP series, click [Auto Tuning Condition] icon and enter "0.5" in the "FWHM of
Peak Profile".
In case of TQ series, click [Auto Tuning Condition] icon and set "-0.1" in the "Relative
Value from initial FWHM".
4
3. GC/MS Data Acquisition
Perform tuning before acquiring calibration data. If tuning is not performed for a long time,
the accurate mass calculation result may not be accurate.
5
3. GC/MS Data Acquisition
Load the method file in the [GCMS Real Time Analysis] - [Acquisition] window. Here,
click [OK] if the following confirmation message is displayed: "[1317] The hardware
configuration for this method is different from the current configuration. The
measurement condition in the method file is modified according to the current
instrument configuration.
6
3. GC/MS Data Acquisition
The detector voltage varies according to the instrument status. If the intensity of each
ion of PFTBA used for calibration has been saturated, lower the detector voltage. If
PFTBA ions are saturated, accurate mass calculation cannot be made correctly.
It’s recommended to check the detector voltage that the intensity of m/z 69 for
PFTBA is approximately 3,000,000 ~ 4,000,000 introducing PFTBA from the [Peak
Monitor] of [Tuning] icon, before GC/MS measurement.
5) Click the (Save) icon on the toolbar to overwrite the method file.
6) Create a batch file to acquire mass calibration data and measure samples. It is
recommended to acquire mass calibration data in each batch.
7) Configure batch settings for mass calibration and sample measurement.
8) Set "0" for the vial # on a row on which a mass calibration method has been set.
7
3. GC/MS Data Acquisition
If the column with thicker film such as over 0.5 μm is used, set the cooling
temperature rate lower.
8
3. GC/MS Data Acquisition
4) Click the [MS] tab and select the [Profile Export] checkbox.
5) Change the MS parameters below.
[End Time (min)]: Enter the GC program time.
[Start m/z]: 40 - 60
[End m/z]: 600 (to acquire the mass spectrum of PFTBA)
[Event Time (sec)]: Set this so that the scan speed becomes approximately
2000 u/sec.
- When the [ON] checkbox of profile export is selected, the mass profile to be used on
MassWorks is output. Due to the large size of mass profile files, it is recommended to
clear this checkbox when you do not plan to use MassWorks.
- If the data is acquired in only SIM or MRM mode, the mass profile can not be loaded
to MassWorks.
9
3. GC/MS Data Acquisition
10
3. GC/MS Data Acquisition
The detector voltage varies according to the instrument status. If the intensity of each
ion of PFTBA used for calibration has been saturated, lower the detector voltage. If
PFTBA ions are saturated, accurate mass calculation cannot be made correctly.
It’s recommended to check the detector voltage that the intensity of m/z 69 for
PFTBA is approximately 3,000,000 ~ 4,000,000 introducing PFTBA from the [Peak
Monitor] of [Tuning] icon, before GC/MS measurement.
11
4 Data Processing Using MassWorks
Toolbar
Chromatogram Graph
Navigation Pane
Graph Window
Spectrum Graph
Utility Pane
Name Contents
The Navigation Pane allows selection between different functional
panes which include "Tasks" "Recent" and "Explore".
Recent provides a listing of the most recently opened data files or
Navigation Pane calibrations
Tasks displays the key modes in MassWorks, Calibration Mode and
Analysis Mode. Select the mode from the explorer panel by clicking
on the icons.
The Utility Pane shows additional information relating to information in
Utility Pane the Graph Window including retention time or mass range windows,
tables, shortcuts, and other information.
The graph window can include spectra, chromatograms, tables (such
as standard ion tables), dialog boxes and buttons. It may be a tab
Graph Window
style, with a different operation on each tab as in the Calibration
mode, or it may contain multiple spectral windows (MDI).
12
4. Data Processing Using MassWorks
The toolbar buttons on the MassWorks toolbar provide the following functionality:
Icon Functionality Contents
New Calibration Close out the current calibration in the calibration task
Open MS Data File Open an MS data file in either calibration mode or analysis mode
Scales the X and Y axis to the full range. The select graph
Set Full Scale
buttons determine which graph to apply.
Scales the Y axis to the full range. The select graph buttons
Set Full Y Scale
determine which graph to apply.
13
4. Data Processing Using MassWorks
14
4. Data Processing Using MassWorks
4) Right click the mouse on the TIC graph and from the drop down menu, select "Mark an
RT window."
5) Specify a range for extracting the spectrum in an area where PFTBA is detected by
clicking the mouse on the start and end points. Then, click the [Standard Ions] tab.
6) Right click the mouse on the table and choose "Load Ions…" from the pop up menu.
15
4. Data Processing Using MassWorks
7) Select the "Shimadzu_Default_CAL.mwc" file that contains the PFTBA ion information.
8) The selected information is input to the lower table and corresponding peaks are shaded.
Confirm that each peak is within the shaded area.
16
4. Data Processing Using MassWorks
10) When data processing is completed, move to the [Review] tab and check the results.
Confirm that each m/z peak is recognized. If the peaks are not recognized, go back to
the [Standard Ions] tab, set the ion and calibration range again, and perform calibrate
again. If the setting is changed and the peaks isn’t recognized, acquire the data again
after increasing the detector voltage.
11) Click [File] - [Save Calibration File As] to save the created calibration file under a new
name. Click [File] - [Print Report] if you want to output the results as a report.
17
4. Data Processing Using MassWorks
1) Click the [Analysis] icon in the navigation pane. Then click [File] - [Open MS Data] to
open a crm format file.
2) Click the [MassWorks Calibration] tab in the utility pane and select the created
calibration file.
3) Right-click the mouse in the chromatogram subwindow. In the popup menu, click [Mark
an RT window] and specify the peak analysis range by clicking the mouse on the start
and end points.
18
4. Data Processing Using MassWorks
4) Enlarge the molecular ion peak in the spectrum subwindow and right-click the mouse on
the m/z value displayed at the peak top, and select [CLIPS Search].
5) Set the parameters in the [CLIPS Search] window and click [Search].
19
4. Data Processing Using MassWorks
Functionality Contents
The accurate mass derived from clicking the monoisotope peak of the ion
Accurate mass to search
the search.
Charge The charge state of the ion.
Show All Results Displays all elemental compositions searched.
Displays a specified number of components from the top in the search
Show Top xxx Results
result.
The mass window in milli-Daltons or PPM about the monoisotope peak to
Mass Tolerance
search for formulas.
The Electron State specifies the preferred types of molecule to be included
or excluded from the results.
Electron State Odd: Includes odd electron types only
Even: Includes even electron types only.
Both: Includes both odd and even electron types.
The Double Bond Equivalent (DBE) specifies the maximum and minimum
Double Bond Equivalent Range number of double bonds per molecule considered to be acceptable in
producing a valid formula.
The Element table defines what elements and what number of each,
Element table minimum to maximum, to allow for the search. It is best to restrict this as
best you can to limit the search candidates.
Check box to indicate whether to use empirical elemental rules to eliminate
formulas that are unfeasible or unlikely, according to typically recognized
chemical rules and/or existing known compound libraries. By using these
empirical rules, you may be able to work with only a limited number of
possible formula candidates and quickly arrive at the correct identification.
On the other hand, some new and unique chemical compounds may fall
Empirical Rules
outside of the statistical boundaries and be eliminated from consideration,
leading to incorrect identifications. Another case where these rules may not
apply is in MS/MS or EI experiments where ion fragments instead of
molecular ions are the unknowns to be determined. It is recommended to
always confirm your formula identification results by also considering high
spectral accuracy hits with these empirical rules unchecked.
Sets the mass range for calculating the spectral accuracy taking isotope
Profile Mass Range
ions into account.
Sets a threshold to minimize interferences between the isotope peaks.
Interference Rejection
Set from 0 - 1 where 1 is the maximum peak height.
Ion Series (formerly Mixture Analysis) Allows the inclusion of up to two
Ion Series additional ion series which are interfering with the ion of interest.
The input can be a formula, or a modification of the search formula.
Reset to Factory Defaults Resets the parameters to factory defaults and clears the element table.
Reset to My Defaults Resets the parameters and element table to user saved values.
Saves the current parameters and element table to for future application to
Save As My Defaults
new searches or for recalling from the "Reset to My Default" button.
Shows the CLIPS parameters dialog every time a search is executed. If you
Show this dialog before each do not want this to show, uncheck this box. This will force CLIPS to
search immediately search without presenting the dialog using the existing
parameters.
20
4. Data Processing Using MassWorks
The more the number of elements registered for composition estimation using calculated
accurate mass, the more the candidate composition formulas. The table below shows the
natural isotopic ratios of elements detected by GC/MS. Some elements such as silicon
(Si), sulfur (S), chlorine (Cl), and bromine (Br) have a higher natural isotopic ratio than
other elements. Check the molecular ions (M+) of the acquired peak and investigate the
relative intensity ratio of M+2. Then set elements as search parameters if that ratio is
relatively high (for example 10 % or more). Doing so will reduce the number of search
candidates and make the search process easier.
Isotopic
Element Isotope
Composition (%)
1
H 99.9885
H 2
H 0.0115
12
C 98.93
C 13
C 1.07
14
N 99.636
N 15
N 0.364
16
O 99.757
17
O O 0.038
18
O 0.205
19
F F 100.000
28
Si 92.223
29
Si Si 4.685
30
Si 3.092
31
P P 100.000
32
S 95.99
33
S 0.75
S 34
S 4.25
36
S 0.01
35
Cl 75.76
Cl 37
Cl 24.24
79
Br 50.69
Br 81
Br 49.31
127
I I 100.000
Reference: NIST Atomic Weights and Isotopic Compositions for All Elements
21
4. Data Processing Using MassWorks
6) Move to the [CLIPS Results] tab in the utility pane and check the search results.
Click the list to display the calculated mass profile for the selected composition on the
spectrum in an overlay manner.
7) To print the graph subwindow (chromatogram and spectrum), click [Print Report] in the
[File] menu. To print out search results, right-click the mouse on the table in the utility
pane and click [Print Report]. Or click [Report to Clipboard] to copy the search results in
print image. If clicking [Export to Clipboard], the search results will be copied as ASCII
value.
22
Appendix 1 Installing MassWorks Software
23
Appendix 1. Installing MassWorks Software
7) Insert the USB security key into a USB port on your computer.
24