2.4A__Macroscale_Columns
2.4A__Macroscale_Columns
be separated are very close (if the difference in R is < 0.2), it's best if the middle of the spots has an R of 0.35. An R near 0.35
f f f
is ideal because it is slow enough that stationary-mobile phase equilibration can occur, but fast enough to minimize band widening
from diffusion.
There are a few variables which aren't applicable to TLC, but which affect the separation of components in column
chromatography. These include the column diameter, quantity of adsorbent used, and solvent flow rate. Table 2.5 summarizes
variable recommendations based on the sample size and degree of separation between components. In all scenarios, the columns
are to be prepped between 5-6 inches high.
Table 2.5: Summary of recommended values for column chromatography. 9
For example, a column of one-inch diameter (1 in is 25.4 mm) should be able to purify around 400 mg of material if the separation
is good (ΔR > 0.2, third column in Table 2.5), or around 160 mg if the separation is difficult (ΔR > 0.1). The column should be
f f
able to be prepared and eluted using around 200 mL of solvent, and the fractions can be collected with approximately 10 mL of
solution each. 9
There are multiple variations on how to physically run a column, and your instructor may prefer a certain method. One large
difference in methods is how the column is prepared. In the "dry packing" method, dry silica or alumina is added directly to a
column, and solvent is allowed to trickle through in portions, then with pressure. In the "wet packing" method, the column is filled
with solvent first, then dry silica or alumina is lightly shaken in, then packed with pressure. In the "slurry" method, solvent is added
to the silica or alumina in an Erlenmeyer flask, poured onto the column as a sludgy material, then packed with pressure.
It is important to know that heat is liberated when solvent is added to silica or alumina (they have an exothermic heat of solvation).
The slurry method is presented in this section, with the main reason being that it allows this exothermic step to happen in an
Erlenmeyer flask instead of on the column. If heat is liberated during the packing of the column, it may generate bubbles from the
boiling of solvent. These can interfere with the separation of the column if they are not adequately removed, and can crack the
adsorbent material in the column.
Step-by-Step Procedures
2.4A.1 https://chem.libretexts.org/@go/page/93526
Figure 2.51: a) TLC of ferrocene/acetylferrocene mixture, b) Column chromatography.
The column pictured in this section shows purification of a 0.20 g sample containing a mixture of ferrocene and acetylferrocene
(crude TLC is in Figure 2.51a). Roughly 8 mL fractions were collected into small test tubes, and roughly 400 mL of eluent was
used.
Figure 2.52: a) TLC plate of crude ferrocene/acetylferrocene mixture before elution, b) After elution, c) Column with frit, d)
Column with cotton wedged in the bottom, e) Column lacking the cotton wedge (would need to be inserted before use).
Run a TLC
1. Run a TLC of the sample to be purified (Figures 2.52 a+b) to determine the appropriate solvent for chromatography. The
desired component should have an R around 0.35 and should ideally be separated from all other spots by at least 0.2 R
f f
units.
2. Prepare a batch of eluent that gives the proper R value. The quantity prepared depends on the quantity of sample, the size of
f
the column, and whether or not the solvent composition is planned to be changed midway. (See Table 2.5 for guidelines and the
eluotropic series for trends in "solvent power.")
2.4A.2 https://chem.libretexts.org/@go/page/93526
Figure 2.53: a) Pouring in silica to 5-6" high in the fume hood, b) Pouring the silica into an Erlenmeyer flask, c+d) Making a slurry.
4. Secure your column perfectly vertically to a ring stand or latticework, clamping it with three-fingered clamps in two locations.
In the fume hood, pour silica gel or alumina adsorbent into the column to between 5-6 inches high (Figure 2.53a).
Safety note: Powdered silica and alumina are lung irritants, and should always be handled carefully in a fume hood. Spilled
dust should be disposed of by mopping it with a wet paper towel (if wet, the fine particles are less dispersive).
5. In the fume hood, pour the adsorbent measured in the column into an Erlenmeyer flask (Figure 2.53b), then add some eluent
(Figure 2.53c). Make a loose slurry by swirling and stirring with a glass stirring rod (Figure 2.53d) until all of the adsorbent is
completely wet, gas bubbles are released, and the consistency is somewhat thick but pourable.
Figure 2.54: a) Pouring the slurry into the column, b) Rinsing the flask, c) Adsorbent adhered to the sides of the column, d) Rinsing
off the clinging adsorbent.
6. Put a beaker or Erlenmeyer flask beneath the clamped column and open the stopcock. In one quick motion, swirl and pour the
silica or alumina slurry into the column using a large-mouthed funnel (Figure 2.54a). Immediately use more eluent to rinse
residual slurry out of the Erlenmeyer flask (Figure 2.54b) and onto the column.
7. Immediately rinse any silica or alumina off the sides of the column reservoir using eluent and a swirling motion from a Pasteur
pipette (Figures 2.54 c+d). If allowed to dry, the adsorbent will cling to the glass and won't easily be rinsed down.
2.4A.3 https://chem.libretexts.org/@go/page/93526
Figure 2.55: a) Jostling the column to remove air bubbles, b) Applying air pressure, c) Adding sand, d) Rinsing sand from the sides.
8. Jostle the column firmly using a cork ring or your knuckles (Figure 2.55a) to dislodge any air bubbles in the column (which
could cause poor separation or cracking of the adsorbent in the column), and to promote an even deposition of adsorbent.
9. Apply gentle air pressure to the top of the column (Figure 2.55b) to compress it, stopping when the eluent level is 1 cm from
the top of the column. If a T-adapter is used with the air line as in Figure 2.55b, fine control of the airflow can be accomplished
by adjustment of the pinch clamp on the rubber tubing.
Throughout the entire elution process, keep the white column of adsorbent wet, with the eluent level above the top of the
silica or alumina.
Gently break the seal to cease application of pressure, and close the stopcock to prevent liquid from dripping out further.
10. Add a thin layer of sand (Figure 2.55c), approximately 0.5 cm high. Rinse the sides of the column with eluent using a swirling
motion to dislodge the sand off the sides of the glass (Figure 2.55d). Open the stopcock and allow liquid to drip out until the
liquid is just above the sand layer. Apply air pressure if the dripping is too slow.
Figure 2.56: a) Dissolving a solid with a small amount of dichloromethane, b) Applying the sample, c) Rinsing the sample in its
flask, d) Applying pressure to push the sample onto the column just past the sand layer.
11. If the crude sample is a liquid, use it directly (go on to step 13).
2.4A.4 https://chem.libretexts.org/@go/page/93526
12. If the crude sample is a solid, do one of the following things:
a. Ideal situation: dissolve the solid in the minimum amount of the eluent (a few mL at most).
b. If the solid is not particularly soluble, or will not dissolve in a few mL of the eluent, dissolve it in the minimum amount of
dichloromethane (few mL at most, Figure 2.56a).
c. If the solid is insoluble in the eluent, an alternate procedure is also possible. Dissolve the solid in a round-bottomed flask
using a few mL of a low-boiling solvent (e.g. dichloromethane or acetone). Add to the flask approximately 1 g of silica or
alumina, then remove the solvent on the rotary evaporator in order to leave a solid that contains the sample deposited on the
adsorbent. With an inch of eluent resting atop the packed column (skip adding the sand layer if this method is used), pour
the silica-adsorbed sample onto the column using a wide mouthed funnel. If any dust clings to the glass, rinse it down with
more eluent (go on to step 15.)
13. Delicately add the sample to the column via pipette, dripping the liquid or solution directly onto the sand with the pipette tip as
close as you can manage, not down the sides (Figure 2.56b). Take care to not squirt liquid in forcibly such that indentations
would be caused in the sand or silica /alumina column.
14. Rinse the sample container with a little solvent (or dichloromethane if used, Figure 2.56c) and add the rinsing to the column
using the same pipette (in order to rinse the pipette as well).
15. Open the stopcock and allow liquid to drip out until the sample is just past the sand layer (Figure 2.56d) and into the white area
of the column (apply air pressure if this takes more than 20 seconds).
16. Gently rinse the sides of the column with a swirling motion using 1-2 pipettes-full of eluent to rinse any splashed sample.
Again, allow liquid to drip out (or apply air pressure) until the sample is pushed into the white adsorbent.
Repeat the rinsing step until you feel confident that the entire sample is deposited on the adsorbent. If some of the sample is still
located in the sand layer, it may dissolve in the eluent when more solvent is added, leading to a loss of yield. If the compound is
colored, the rinsing should be completely clear.
Figure 2.57: a+b) Filling the solvent reservoir, c+d) Eluting the column.
The optimal drip rate during elution depends on the size of the column. The ideal flow of eluent is when the solvent in the
cylindrical section of the column above the adsorbent drops at a rate of 2.0 inches per minute. Therefore, the drip rate should
10
2.4A.5 https://chem.libretexts.org/@go/page/93526
The drip rate for a one-inch column should be where individual drops can be barely distinguished. A stream of liquid pouring
out the stopcock with this size of column is slightly too fast.
Collect fractions
19. Immediately start collecting the eluting liquid into test tubes on a rack (Figure 2.58a). See Table 2.5 for recommendations on
volumes to collect in each test tube.
20. When the first test tube fills, or if a certain height of liquid has been collected as recommended by your instructor or Table 2.5,
move the rack over to start collecting into a different tube (Figures 2.58 b+c). Fill and keep the tubes in order on the rack.
These different tubes are called "fractions". The goal of a column is to collect small enough fractions that most (or some)
fractions contain pure material. If the separation of the mixture is difficult (if the ΔR of the components is low), it may be best
f
Figure 2.59: a-c) Rinsing material splashed onto the tip of the column, d) Students running a column.
21. As liquid drains off the column, it often splashes onto the outsides of the tip of the column, and when the solvent evaporates
you may see a ring of material on the tip (you will see a ring of solid if the component is a solid as in Figure 2.59b, or oily
droplets if the component is a liquid). If the components are colored, the column tip should be rinsed (Figure 2.59c) when it
appears as if one component has completely eluted and before the other component approaches.
22. Periodically keep an eye on the eluent level, and refill before it drops below the sand layer.
2.4A.6 https://chem.libretexts.org/@go/page/93526
Figure 2.60: a) Elution, b) Addition of ethyl acetate to increase the solvent polarity, c) Solvent level nears the sand layer, d)
Refilling.
However, if the components have very different R values, the solvent polarity may be increased after one component has
f
Increasing the solvent polarity will make components travel "faster." There are several reasons a faster elution is desired. First,
if one component has already exited the column, the column has already done its job with separation, so speeding up the
process will not affect the purity of the collected fractions. Second, the longer it takes to run a column, the wider will be the
component bands (due to diffusion), and collecting a broad band of material will use (and waste) a lot of solvent.
Table 2.6 contains a partial list of the eluotropic series, a list of common solvents ranked according to their "solvent power" in
normal phase chromatography. The more polar solvent causes the most dramatic increase in R . f
24. To increase the solvent polarity, the polar solvent can be dripped directly into the eluent on the column reservoir (Figure 2.60b).
For example, if using a hexanes:ethyl acetate mixture, addition of pure ethyl acetate to the eluent currently in the reservoir
would increase its polarity. If the eluent level is running low, a solution could be prepared that contains a higher percentage of
the more polar component. For example, if the column first used a 4:1 hexanes:ethyl acetate mixture, using a 1:1 mixture would
be a more polar solvent.
25. Elute the column with the more polar solvent as before, and always remember to watch the eluent level, and refill (Figure
2.60d) before it drops below the sand layer.
2.4A.7 https://chem.libretexts.org/@go/page/93526
Figure 2.61: a) Original TLC plate, b) Fractions collected by the column, c) Spotting fractions onto a TLC plate, d) A visualized
TLC plate of samples of each fraction.
In column chromatography, the sample is deposited on the top of the column and eluted down, while in thin layer
chromatography the sample is spotted on the bottom of the plate and eluted up. Therefore, a column can be thought of like an
upside-down TLC plate. A compound with a higher R runs "faster," meaning it will end up higher on a TLC plate, and will be
f
In the column pictured in this section, the component with the lower R (orange on the TLC plate in Figure 2.61a), is the
f
2.4A.8 https://chem.libretexts.org/@go/page/93526
Figure 2.63: a) Eluted TLC plates of fractions containing possible compound, b) Combining fractions, c) Rinsing a fraction tube.
28. Run a TLC of all fractions that contain compound, spotting up to five samples per 1-inch wide TLC plate. Wider TLC plates
may be used for this purpose if available.
29. Identify the compound with the desired R by comparison with the original crude TLC plate. Choose to retain the fractions that
f
have the desired compound in pure form, as evidenced by the eluted TLC plate. For example, if the compound with the higher
R is desired in Figure 2.63a, fractions 6-10 should be kept.
f
30. Combine the pure fractions into an appropriately sized round-bottomed flask (no more than half-full, Figure 2.63b). Rinse each
test tube with a small amount of eluent (or other solvent if solubility is an issue), and add the rinsing to the round-bottomed
flask (Figure 2.63c).
31. Evaporate the solvent on the rotary evaporator to leave the purified compound in the flask.
Figure 2.64: a) Using air pressure to dry the column, b) Allowing the column to dry upside-down, c) Collecting the waste silica.
2.4A.9 https://chem.libretexts.org/@go/page/93526
Safety note: Powdered adsorbents are a lung irritant, and their hazard is exacerbated if the column contains residual compounds
that can now make their way into your lungs. Pouring silica or alumina powders should always be done in the fume hood.
34. When the majority of the adsorbent has been collected in a waste container, use water to rinse any residual solid into the sink,
and then rinse the column with acetone into a waste beaker. Further clean the column with soap and water and dry with the
stopcock parts separated.
Troubleshooting
2.4A.10 https://chem.libretexts.org/@go/page/93526
The Pipette Broke in the Column
It is quite common to break the tip of a Pasteur pipette while rinsing the column, which often falls and wedges itself into the
delicate column. Unfortunately, a broken pipette in a column can cause problems with the separation of components.
For example, Figure 2.65 shows the effect of a broken pipette on the separation of two components. A broken pipette is embedded
in the column and is the nearly vertical line of orange seen between the two bands on the column. Since stationary-mobile phase
equilibration does not occur on the glass surface of the pipette, compounds stream down quickly and move faster than they should
based on their R . The orange vertical line is the top component draining into the band of the bottom component, contaminating it.
f
2.4A.11 https://chem.libretexts.org/@go/page/93526
Figure 2.66: Effect of an air bubble on the separation in a column.
If air bubbles are seen in the column and the sand or sample has not yet been applied, give the column a good jostling during
packing to remove all air bubbles. See your instructor if the bubbles are not budging as you may be approaching the task too
delicately. If the sand or sample has already been applied, it's best to leave the column as is and hope the air bubbles do not affect
the separation.
bands may have no effect on the separation. In the future, be sure to check that the column is perfectly vertical in both the side-to-
side and front-back directions.
8
W.C. Still, M. Kahn, A. Mitra, J. Org. Chem., Vol. 43, No. 14, 1978.
9
A typical small test tube (13x100 mm) has a capacity of 9 mL, and a typical medium test tube (18x150 mm) has a capacity of
27 mL .
10
W.C. Still, M. Kahn, A. Mitra, J. Org. Chem., Vol. 43, No. 14, 1978.
This page titled 2.4A: Macroscale Columns is shared under a CC BY-NC-ND 4.0 license and was authored, remixed, and/or curated by Lisa
Nichols via source content that was edited to the style and standards of the LibreTexts platform.
2.4A.12 https://chem.libretexts.org/@go/page/93526