Instrumentation, Principles of Instruments, And Spectrophotometry (1)
Instrumentation, Principles of Instruments, And Spectrophotometry (1)
Wavelength Regions
● Visible spectrum: 400 - 700 nm
● Ultraviolet region (UV): <400 nm
● Infrared region (IR): >700 nm
Planck's formula: E = hv
where:
E - is the energy of a photon in Joules
h - constant (6.626 x 10-27 erg sec) → Bishop
v - frequency
Frequency
● Frequency is the number of vibrations of wave motion per second.
● The lower the wave frequency, the longer the wavelength.
● The wavelength is inversely related to frequency and energy; the shorter the
wavelength, the higher the frequency and energy, and vice versa.
Single-beam Spectrophotometer
Major Analytical Methods in the Clinical Chemistry ● It is the simplest type of an absorption spectrophotometer.
Laboratory ● It is designed to make one measurement at a time at one specified
wavelength.
I. COLORIMETRY ● The maximum absorption of the analyte must be known in
● Spectrophotometric measurement - is the measurement of light intensity in advance when a single-beam instrument is used.
a narrower wavelength.
● Photometric measurement - is the measurement of light intensity using a
specific wavelength
● The primary analytical utility of spectrophotometry or filter photometry is
the isolation of discrete portions of the spectrum for purposes of
measurement.
Spectrophotometry
● It involves measurement of the light transmitted by a solution to
determine the concentration of the light-absorbing substances in
the solution.
Double-beam Spectrophotometer
● A spectrometer is a device that measures the wavelengths of light
● It is an instrument that splits the monochromatic light into two
or the intensity of radiation.
components: one beam passes through the sample, and the other
through a reference solution or blank.
● The additional beam corrects for variation in light source intensity.
● The absorbance of the sample can be recorded directly as the
electrical output of the sample beam.
Phototube
Kinds of Monochromators
● It contains cathode and anode enclosed in a glass case.
Prisms
● It has a photosensitive material that gives off electrons when
● These are wedge-shaped pieces of glass, quartz, or sodium chloride.
light energy strikes it.
● These can be rotated, allowing only the desired wavelength to pass
● It requires an external voltage for operation.
through an exit slit.
● A narrow light focused on a prism is refracted as it enters the more
dense glass.
Photomultiplier Tube (PMT)
● It is the most sensitive photodetector. Absorbance (A)
● The response of the PMT begins when incoming photons strike a ● It is the amount of light absorbed.
photocathode. ● It is proportional to the inverse log of transmittance.
● It is limited to measuring low power radiation because intense light causes ● It is mathematically derived from % T.
irreversible damage to the photoelectric surface.
● It should never be exposed to room light because it will be damaged. A = abc = 2 - log%T
where:
A = Absorbance
a = Molar absorptivity; absorptivity of the compound under standard
conditions
b = Length of light through the solution
c = Concentration of absorbing molecules/solution
Photodiode
● It is not as sensitive as PMT (Photomultiplier Tube) but with excellent
linearity.
● It measures light at a multitude of wavelengths - detects less amount of
light.
● It is most useful as a simultaneous multichannel detector.
Where:
7. Meter or Readout Device Au → Absorbance of the Unknown Solution
● It displays the output of the detection system. As → Absorbance of the Standard Solution
● Example: Galvanometer / ammeter / light-emitting diode (LED) display Cs → Concentration of the Standard Solution
Beer’s Law
● It states that the concentration of the unknown substance is
directly proportional to the absorbed light (absorbance or optical
density) and inversely proportional to the amount of transmitted
light (% Transmittance).
● It mathematically establishes the relationship between
concentration and absorbance.
II. VOLUMETRIC (Titrimetric)
● Principle: The unknown sample is made to react with a known solution in
the presence of an indicator.
○ Examples:
■ Schales and Schales method (Chloride test)
■ EDTA Titration method (Calcium test)
III. TURBIDIMETRY
● Principle: It determines the amount of light blocked (reduction of light)
by a particulate matter in a turbid solution.
● It depends on specimen concentration and particle size.
● The measurement of the reduction of light is due to particle formation.
● For measuring abundant large particles (proteins) and bacterial suspensions.
● Solutions requiring quantitation by turbidimetry are measured using visible
photometers or visible spectrophotometers.
● Uses: Protein measurements (CS and urine); to detect bacterial growth in
broth cultures, antimicrobial test (broth method); and to detect clot formation
IV. NEPHELOMETRY
Blanking Technique
● Principle: It determines the amount of scattered light by a
● It means the blank contains serum but without the reagent to complete the
particulate matter suspended in a turbid solution.
assay.
● Photodetector: Photomultiplier tube
● Reagent blank corrects absorbance caused by the color of the reagents - the
● Use: For measuring the amount of antigen-antibody complexes (proteins)
absorbance of reagents is automatically subtracted from each of unknown
● Light scattering depends on wavelength and particle size.
reading.
● Light scattered by particles is measured at an angle, typically 15 to 90
● Sample blank measures absorbance of the sample and reagent in the
degrees to the beam incident on the cuvette.
absence of the end product, and corrects the measurement for optical
● Most antigen-antibody complexes have a diameter of 250 nm to 1500 mm,
interference (like hemoglobin) absorbing the wavelength of measurement.
and the wavelengths used are 320 mm to 550 nm, thus light is scattered
● A blanking process may not be effective in some cases of turbidity, and
forward (Rayleigh-Debye type).
ultracentrifugation may be necessary to clear the serum or plasma of
● Basic Components: Light source (mercury-arc lamp, a tungsten-filament
chylomicrons.
lamp, light emitting diode, and a laser), collimator, monochromator, sample
● Lipids interfere mainly by increasing light scatter (turbidity).
cuvette, stray light trap, and photodetector.
● To correct for artifactual absorbance readings, "blanking"
procedures or dual-wavelength methods may be used.
VI. CHEMILUMINESCENCE
● Principle: The chemical reaction yields an electronically excited compound
that emits light as it returns to its ground state or transfers its energy to
another compound which then produces an emission.
● It quantifies sample analytes in a wide range of wavelengths not observed in
the visible spectrum.
● This method is best when differentiating two compounds having excitation Cleaning of Glassware
reaction at the same wavelength but emit at different wavelengths. ● Pre Soaking glassware in soapy water.
● It is widely used nowadays due to its high sensitivity while even, more ● Multiple rinses must be done with appropriate grade water.
sensitive than fluorometry. ● Detergent-contaminated water have more alkaline ph as compared with the
● Methods: Chemiluminescence Immunoassay (CLIA) ph of the appropriate grade water.
● Uses: Autoantibody testing; measurement of hormones, drugs, vitamins, ● Acid dichromate and nitric acid is the cleaning solution for glassware.
tumor markers; forensic analysis; microbial and infectious disease marker ● Ultrasonic cleaners help remove debris coating the surfaces of glass or
studies; and toxicology Photodetector: Photomultiplier tube (luminometer). plasticware.
I. CALIBRATION MARKS/DESIGN
To Contain (TC) Pipette
● It holds a particular volume but does not dispense the exact volume.
● It is also known as the rinse-out pipette.
Boron-free Glassware/Soft Glass
● It does not meet Class A certification criteria.
● It has high resistance to alkali.
● Examples: Sahli hemoglobin and Long-Levy pipettes
● Its thermal resistance is less as compared to borosilicate glass.
II. DRAINAGE CHARACTERISTICS
Blow-out Pipette
● It has continuous etched rings on top or near the mouthpiece of the pipette
and exact volume is obtained when the last drop is blown out.
● This type of pipette is not rinsed out.
● Examples: Ostwald-Folin and serological pipette
● Self-draining Pipette
● It allows the liquid to drain by gravity.
● It does not have etched or frosted rings.
Corex
● It is a special alumina-silicate glass that has been strengthened chemically III. TYPES OF PIPETTES
than thermally; six times stronger than borosilicate. Transfer Pipette
● Volumetric Pipette - for nonviscous fluid; self-draining; small amount left in
the tip should not be blown out
● Ostwald Folin - for viscous fluid; with etched ring
● Pasteur Pipette - transfers fluids without consideration of a specific volume
● Automatic macro or micropipettes
MECHANICAL/AUTOMATIC PIPETTES
Air Displacement Pipette
● It relies on piston for suction creation to draw the sample into a disposable
tip.
● The piston does not come in contact with the liquid.
● A disposable, one-time use polypropylene tip is attached to the pipette
barrel.
Thermometer
● Temperature-monitoring devices should be verified for accuracy at 6or
12-month intervals.
● Types of thermometers: Total immersion (freezers and refrigerators)
and Partial immersion (water baths and heating blocks)
Reflectance Photometry
● It is the measurement of light reflected from solid surfaces.
● The intensity of the reflected light from the reagent carrier is
compared with the intensity of light reflected from a
reference surface.
● A reflectometer is used to measure analytes by measuring
the quantity of light reflected by a liquid sample that has
been dispensed onto a grainy or fibrous solid support.