DEA2 Software Upgrading
DEA2 Software Upgrading
There are two parts upgrading, software in the mainboard and in the touch screen.
2. Open “Property” menu of “My computer”, then open “Device manager” menu.
3. Find “Port (COM and LPT)”, click right button on “Prolific USB-to-Serial Bridge” to update driver. After
you finish the driver installation, you will find the COM Port number as below. Remember the port number.
Note: Please make sure the driver is installed well; otherwise you cannot upgrade software.
1. Turn on analyzer.
4. Select COM Port which you find in Part 1, Step 3. Then select and input related setting as the showed above.
5. Click “Browse…” button and then find the software (such as ......hex)
6. Click “Start” button to start software upgrading.
7. If you get error report as above, you need to restart analyzer and then click “Start” button again.
8. During upgrading, you will hear sound. After software upgrading finished, the sound will disappear. Adjust the switch
to 1, 2 side. And please turn off the device.
Part 1. Preparing
Make sure the screen software in the folder of XRD_TFT at the root directory of the TF card.
When upgrading finish, the machine will show self-checking and stop at the page of asking for calibration.
4. Switch off the machine, and take away the TF card, tighten the screws and close the shell.