MI MODULE -1
MI MODULE -1
E-mail: [email protected]
URL: http://www.vssut.ac.in
Google Scholar: Gyan Biswal
Home Page: http://in.linkedin.com/pub/gyan-biswal/14/458/a8
ORCID id: https://orcid.org/0000-0001-7730-1985
Reference Books:
* Larry Jones & A Foster Chin, “Electronic Measurement & Instrumentation Systems”,
John Wiley & Son Publishers.
* Golding & Waddis, “Electrical Measurement and Measuring Instruments”, Reem
Publishers.
* David A. Bell, “Electronic Instrumentation and Measurements”, 3rd ed., Oxford
University Press.
* P. Purkait, B. Biswas, S. Das and C. Koley, “Electrical and Electronics Measurement
and Measuring Instruments”, McGraw Hills.
Measurement
ADC
system /
sensors
Readable form
• Visual
• Motion of pointer on a scale
• Series of light (digital)
Danish physicist who discovered the French physiologist who discovered the
relationship between current and moving-coil galvanometer – from muscle
magnetism – from the deflection of contractions in frogs using a telephone,
a compass needle which operates on an extremely feeble
currents similar to animal electricity
Deprez-d'Arsonval Galvanometer
Sinusoidal wave
Square wave
Triangle wave
Vavg = 0
Vrms = 0.707Vp
Sine Wave
Vavg = 0.636Vp
Vrms = 0.707Vp
Full Wave
Vavg = 0.318Vp
Vrms = 0.5Vp
Half Wave
1. Electrodynamometer
2. Iron Vane
3. Electrostatic
4. Thermocouple
5. D’Arsonval with rectifier
D’Arsonval YES YES with Most widely used meter movement for
rectifier measuring direct current or voltage and
resistance
Current from the circuit in which measurements are being made with the meter passes
through the windings of the moving coil. Current through the coil causes it to behave as
an electromagnet with its own north and south poles. The poles of the electromagnet
interact with the poles of the permanent magnet, causing the coil to rotate. The pointer
deflects up scale whenever current flows in the proper direction in the coil. For this
reason, all dc meter movements show polarity markings.
T c = ks x θ = T d = k1 x I
As, θ α I
E dc 0.45E rms
Rs = − Rm = − Rm
I dc I dc
Sac = 0.45Sdc
Benefits
Limitations
▪ this type of instrument suits for DC only. In AC, the instrument does
not operate because of equal amount force experienced by the pointer
but opposite direction. Due to the inertia of the pointer, it (pointer)
remains at zero position.
▪ the moving system is highly delicate so needs careful handling.
▪ the coil being very fine, can not withstand prolonged overloading.
▪ Costlier
▪ Error occur at the later stage due to Ageing effect (permanent magnet
and control spring).
Where,
Vm = I m Rm
• The shunt resistor is parallel with the meter movement, thus the voltage
drop for both is equal
Vsh = Vm
• Then the current through the shunt is,
I sh = I − I m
Vsh I m Rm I m Im
Rsh = = = Rm = Rm..............1.0
(I − I m )
Ohm
I sh I sh I sh
Example 1-1
Calculate the value of the shunt resistance required to convert a 1-mA
meter movement, with a 100-ohm internal resistance, into a 0- to 10-mA
ammeter?
Solution:
Vsh = Vm = 0.1V
I sh = I − I m = 10 mA − 1mA = 9mA
Vsh 0.1V
Rsh = = = 11.11
I sh 9mA
Rm
Rsh =
n −1
Examples 1-2
• A 100 µA meter movement with an internal resistance of 800 Ω is used in
a 0- to 100 mA ammeter. Find the value of the required shunt resistance.
Answ: ~ 0.80 ohm
• On this range the shunt resistance is equal to Rsh and can be computed
by Equation
Rm
Rsh =
n −1
I m ( Rsh + Rm )
Rb + Rc =
I2
I m ( Rsh + Rm )
Rc =
I3
Ra = Rsh − ( Rb + Rc )
Rb = ( Rb + Rc ) − Rc
• To find the value of the multiplier resistor, first determine the sensitivity,
S, of the meter movement.
1
Sensitivit y = (/V)
I fs
Example 1-2
1 1
= 2k
Sensitivity,
S= =
I fs 500 V
Example 1-3
Calculate:
(a) Voltage across RB without any meter
connected across it. Fig. 2.2
(b) Voltage across RB when meter A is used.
(c) Voltage across RB when meter B is used
(d) Error in voltmeter readings.
Solution:
RB
VRB = E
(
A R + R )
B
5kΩ
= 30V
25k + 5k
= 5V
Re1
VRB = E
e1
R + R A
Therefore, the voltage reading
obtained with meter A, determined 3.33kΩ
= 30V
by the voltage divider equation, is 3.33kΩ + 25kΩ
= 3.53V
Cont…
(d)
(Expected value - Measured value)
Voltmeter A error = 100%
Expected value
E
Ie =
R1
E
Im =
R1 + Rm
hence;
Im R1
=
I e R1 + Rm
Therefore
Im
Insertion error = 1 − 100%
Ie
Example 1-4
A current meter that has an internal resistance of 78 ohms is used to
measure the current through resistor Rc in Fig. 2.5. Determine the
percentage of error of the reading due to ammeter insertion.
Fig. 2.5
Solution:
Fig. 2-6
The current meter will be connected into the circuit between points X and Y
in the schematic in Fig. 2.6. When we look back into the circuit from
terminals X and Y, we can express Thevenin’s equivalent resistance as
RTH = Rc +
(Ra Rb )
(Ra + Rb )
RTH = 1 k + 0.5 k = 1.5 k
Im R1
=
I e (R1 + rm )
E
I fs =
R Z + Rm
Fig. 2-7 Basic ohmmeter circuit
Rz = variable resistor
Summary
How it works?
▪ One iron vane is held fixed to the coil frame and other is free to
rotate, carrying with it the pointer shaft. Two irons lie in the
magnetic field produced by the coil that consists of only few turns
if the instrument is an ammeter or of many turns if the
instrument is a voltmeter.
▪ Current in the coil induces both vanes to become magnetized and
repulsion between the similarly magnetized vanes produces a
proportional rotation. The deflecting torque is proportional to
the square of the current in the coil, making the instrument
reading is a true ‘RMS’ quantity. Rotation is opposed by a
hairspring that produces the restoring torque. Only the fixed coil
carries load current, and it is constructed so as to withstand high
transient current.
▪ Limitation: Moving iron instruments having scales that are
nonlinear and somewhat crowded in the lower range of
calibration.
Ammeter
▪ Instrument used to measure current in the circuit.
▪ Always connected in series with the circuit and carries the current
to be measured.
▪ This current flowing through the coil produces the desired
deflecting torque.
▪ It should have low resistance as it is to be connected in series.
Voltmeter
• Instrument used to measure voltage between two points in a
circuit.
• Always connected in parallel.
• Current flowing through the operating coil of the meter produces
deflecting torque.
• It should have high resistance. Thus, a high resistance of order of
kilo ohms is connected in series with the coil of the instrument.
✓Let, ‘I’ be the initial current; ‘L’ be the instrument inductance; and ‘θ’ is
the deflection. (change in ‘dI’ causes change in ‘dθ’ and ‘dL’.
1 2
I dL + IL dI = ILdI + I dL + Td d
2
2
1 2
I dL = Td d
2
1 2 dL
Td = I ( N − m)
2 d
Tc = k s d
Benefits
Limitations
Source
Operating Principle
• It would have a torque in one direction during one half of the cycle and an
equal effect in the opposite direction during the other half of the cycle.
• If the frequency were very low, the pointer would swing back and forth
around the zero point.
• If, however, we were to reverse the direction of the flux each time the
current through the movable coil reverses, a unidirectional torque would be
produced for both positive and negative halves of the cycle.
Construction
Fixed Coils
• The field is produced by a fixed coil.
• This coil is divided into two sections to give a more uniform field near the centre and
to allow passage of the instrument shaft.
Moving Coil
• A single element instrument has one moving coil.
• The moving coil is wound either as a self-sustaining coil or else on a non-metallic
former.
• A metallic former cannot be used as eddy current would be induced in it by the
alternating field.
• Light but rigid construction is used for the moving coil.
• It should be noted that both fixed and moving coils are air cored.
Control
• The controlling torque is provided by two control springs.
• These springs act as leads to the moving coil.
Moving System
Damping
Shielding
• The field produced by the fixed coils is somewhat weaker than in other
types of instruments
• It is nearly 0.005 to 0.006 Wb/m
• In D.C. Measurements even the earth magnetic field may affect the
readings.
• Thus it is necessary to shield an electrodynamometer type instrument from
the effect of stray magnetic fields.
• Air cored electrodynamometer type instruments are protected against
external magnetic fields by enclosing them in a casing of high permeability
alloy.
• This shunts external magnetic fields around the instrument mechanism and
minimizes their effects on the indication.
Torque Equation
i) Frequency error
Advantages
i) These instruments can be used on both a.c & d.c
Disadvantages
(i) They have a low torque/weight ratio and hence have a low sensitivity.
(iii) They are more expensive than either the PMMC or the moving iron type
instruments.
(v) The operating current of these instruments is large owing to the fact that
they have weak magnetic field. The flux density is about 0.006 Wb/m as
against 0.1 to 0.5 Wb/m in PMCC instruments