lab mannual of EMI ex completed
lab mannual of EMI ex completed
List of Experiments
Sr.
Name of Experiment
No
1 To study measuring instruments and Classification of Measuring Instruments.
2 To study electrical measuring instruments and their purposes.
Experiment no. 1
The instrument used for measuring the physical and electrical quantities is known as the measuring
instrument. The term measurement means the comparison between the two quantities of the same unit.
The magnitude of one of the quantity is unknown, and it is compared with the predefined value. The
result of the comparison obtained regarding numerical value.
• Electrical Instrument
• Electronic Instrument
• Mechanical Instrument
The mechanical instrument uses for measuring the physical quantities. This instrument is suitable for
measuring the static and stable condition because the instrument is unable to give the response to the
dynamic condition.
The electronic instrument has quick response time. The instrument provides the quick response as
compared to the electrical and mechanical instrument.
The electrical instrument is used for measuring electrical quantities likes current, voltage, power, etc.
The ammeter, voltmeter, wattmeter are the examples of the electrical measuring instrument.
The ammeter measures the current in amps; voltmeter measures voltage and Wattmeter are used for
measuring the power.
The classification of the electric instruments depends on the methods of representing the output reading.
The instrument used for measuring the physical and electrical quantities is known as the measuring
instrument. The term measurement means the comparison between the two quantities of the same unit.
The magnitude of one of the quantity is unknown, and it is compared with the predefined value. The
result of the comparison obtained regarding numerical value.
Types-of-measuring-instrument
Absolute Instrument
The absolute instrument gives the value of measures quantities regarding the physical constant. The
physical constant means the angle of deflection, degree and meter constant. The mathematical calculation
requires for knowing the value of a physical constant.
The tangent galvanometer is the examples of the absolute instruments. In tangent galvanometer, the
magnitude of current passes through the coil determines by the tangent of the angle of deflection of their
coil, the horizontal component of the earth magnetic field, radius and the number of turns of wire used.
The most common applications of this type of instrument are found in laboratories.
Secondary Instrument
3
In the secondary instrument, the deflection shows the magnitude of the measurable quantities. The
calibration of the instruments with the standard instrument is essential for the measurement. The output
of this type of device is directly obtained, and no mathematical calculation requires for knowing their
value.
Digital Instrument
The digital instrument gives the output in the numeric form. The instrument is more accurate as compared
to the analogue instrument because no human error occurs in the reading.
Analog instrument
The instrument whose output varies continuously is known as the analogue instrument. The analogue
instrument has the pointer which shows the magnitude of the measurable quantities. The analogue device
classifies into two types.
1. Indicating Instrument – The instrument which indicates the magnitude of the measured quantity is
known as the indicating instrument. The indicating instrument has the dial which moves on the graduated
dial. The voltmeter, ammeter, power factor meter are the examples of the indicating instrument.
2. Integrating Instrument – The instrument which measures the total energy supplied at a particular
interval of time is known as the integrating instrument. The total energy measured by the instrument is
the product of the time and the measures electrical quantities. The energy meter, watt-hour meter and the
energy meter are the examples of integrating instrument.
3. Recording Instrument – The instrument records the circuit condition at a particular interval of time
is known as the recording instrument. The moving system of the recording instrument carries a pen which
lightly touches on the paper sheet. The movement of the coil is traced on the paper sheet. The curve drawn
on the paper shows the variation in the measurement of the electrical quantities.
5
Experiment no. 2
Signal analyzer Measures both the amplitude and the modulation of a RF signal
CRO(Cathode Ray
Check transistor
Oscilloscope)
Experiment no. 3
Experiment no. 4
Digital multimeter-1no,
Patch codes.
Theory-
A very important device used in the measurement of medium resistances is the Wheat stones
bridge .it is an accurate and reliable instrument .The wheat stone bridge is an instrument based on
the principle of null indication and comparison measurements. The basic circuit of a wheat stone
bridge is shown in fig. it has four resistive arms, consisting of resistances P,Q,R and S together
with a source of emf and a null detector , usually a galvanometer G or other sensitive current meter
is used.
The current through the galvanometer depends on the potential difference between point’s b and d
.The bridge is said to be balanced when there is on current through the galvanometer or when the
potential difference across the galvanometer is zero. this occurs when the voltage from point ‘b’ to
point ‘a’ equals the voltage from point ‘d’ to point ‘a’ or by referring to other battery terminal,
when the voltage from point ‘d’ to point ‘c’ equals the voltage from point ‘b’ to point ‘c’. For
bridge balance;
I1P=I2 R ......................................................... 1)
I1=I3=E/P+Q ............................................................. (2)
I2=I4=E/R+S ............................................................ (3)
E=emf of battery.
Combining equ (1) and (2)
we get P/P+Q=R/R+S…………………….. (4)
QR=PS
Equ……………….. (5)
Shows the balance condition of wheat stone bridge. If three of the resistances are known then
fourth may be determined by formula…
R=S*P/Q
Where R is the unknown resistance, S is called the standard arm resistor and P and Q are called the
ratio arms.
Procedure: -
1) Connect the patch cord as per the circuit diagram.
2) Note the resistance of P and Q using multimeter.
3) Adjust the resistance of P, Q, R, S
4) Switch on the power supply and adjust the resistance S such that galvanometer
shows the zero deflection.
5) Now calculate R, R=P*S/Q
Circuit Diagram:-
11
Observation Table: -
Result:- Hence we have studied the medium resistance by using whetstone bridge…………
Experiment no. 5
Aim: - Measurement of the low resistance by using Kelvin Double bridge method.
Equation (4) is the usual working equation. For the Kelvin’s Double Bridge .It indicates the
resistance of connecting lead r. It has no effect on measurement provided that the two sets of
ratio arms have equal ratios. Equation (3) is useful however as it shows the error that is
introduced in case the ratios are not exactly equal. It indicates that it is desirable to keep r as
small as possible in order to minimize the error in case there is a diff. between the ratio P/Q and
p/q. R=P/QS
Circuit Diagram:-
Observation Table
Procedure: -
1) The circuit configuration on the panel is studied.
2) Supply is switched on and increased upto 5v.
3) The unknown resistance is connected as shown .
4) The value of P,Q was selected such that a. P/Q=p/q
13
5) S was adjusted for proper balance and balance value of s was balanced.
6) The value of known resistance was calculated.
Precautions:-
Current Transformer:
– Primary winding is connected in series with line carrying the current to be measured
• Therefore, IPri α Load
– Primary winding consist of very few turns
• Therefore, no appreciable voltage drop across it
– Secondary winding has larger number of turns
• Exact number is being determined by the turn ratio
• Relationship in a C.T.: rs and xs = resistance and reactance of secondary winding
• re and xe = resistance and reactance of external burden
• Ep and Es = primary and secondary winding induced voltage
• Np and Ns = number of primary and secondary winding turns
• Vs = voltage at secondary winding terminals
• Ip and Is = primary and secondary winding currents
• Φ = working flux of the transformer
• θ = phase angle of transformer (angle between Is reversed and Ip)
• δ = angle between Es and Is
• Δ = phase angle of secondary winding load circuit
• Io = exciting current
• Im and Ie = magnetizing and loss component of Io
• α = angle between Io and Φ
Errors in C.T.:
• The secondary winding current is not a constant fraction of the primary winding current
– depend upon magnetizing and loss component of exciting current
– this introduces considerable errors into current measurements
error in C.T.:
o Ideally, R=n and θ=0
o But, as a result of physical limitations inherent in electric and magnetic circuit, the
ideality will be lost and errors are induced
o The expression of R and θ indicates that
o Both depend upon the Ie and Im respectively
▪ Thus, they are chosen small.
o Specific design feature will help in minimization of the errors Effect of The resultant
mmf maintains the flux in core and supply the iron losses secondary winding openEffect
of secondary winding open
o C.T. are always used with the secondary winding closed
o Never open the secondary windings circuit of a C.T. while its primary winding is
energized
o Failure to this may lead to serious consequences for both
o In case of P.T., the current flowing in the primary winding is largely the reflection of that
flowing in the secondary circuit.
o whereas, in case of a C.T., the primary winding is connected in series with the line
whose current is being measured
o This current is in no ways controlled or determined by the condition of secondary
winding circuit
o Under normal operating conditions, both primary and secondary windings produces mmf,
which act against each other .since the resultant mmf is small, the flux density is also quite low
hence, a small voltage is induced in the secondary winding
o If the secondary winding is open-circuited when the primary winding is carrying current the
primary mmf remains same while the open secondary mmf reduces to zero therefore, the
resultant mmf is very large (i.e., equal to primary mmf =IpNp) this large mmf produces a large
flux in core till it saturates this large flux linking the turn of secondary winding would induce a
high voltage in the secondary winding which could be dangerous to the transformer insulation
and to the person operating it Also, the eddy current and hysteresis loses would be very high
under this condition thus the transformer may be overheated and completely damaged Even if
this does not happen, the core may become permanently magnetized and give erroneous results.
The secondary mmf is slightly less than the primary mmf Thus, the resultant mmf is small .The
resultant mmf maintains the flux in core and supply the iron losses.
Experiment no. 7
Aim: - Insulation and earth resistance measurement through Megger.
Theory:-
History of Megger
The device is being used since 1889, popularity raised during 1920s since the long back device is same
in its uses and purpose of testing, few real improvements appeared in recent years with its design and
quality of tester. Now high-quality options are available which are easy to use and quite safe.
What is Megger?
Insulation resistance IR quality of an electrical system degrades with time, environment condition, i.e.,
temperature, humidity, moisture and dust particles. It also gets impacted negatively due to the presence
of electrical and mechanical stress, so it’s become very necessary to check the IR (Insulation resistance)
of equipment at a constant regular interval to avoid any measure fatal or electrical shock.
Constructional diagram
Circuit diagram
Types of Megger
This can be separated into mainly two categories:-
1. Still keeps important in such high-tech world as it’s an oldest method for IR value
determination.
2. No external source required to operate.
3. Cheaper available in market.
Disadvantages of Hand Operated Megger
1. At least 2 person required to operate i.e. one for rotation of crank other to connect megger
with electrical system to be tested.
2. Accuracy is not up to the level as it’s varies with rotation of crank.
3. Require very stable placement for operation which is a little hard to find at working sites.
4. Unstable placement of tester may impact the result of tester.
5. Provides an analog display result.
6. Require very high care and safety during use of the same.
Construction of Megger
Circuit Construction features :-
Work philosophy based on ohm-meter or ratio-meter. The deflection torque is produced with
megger tester due to the magnetic field produced by voltage and current, similarly like ‘Ohm’s
Law’.
The torque of the megger varies in a ration with V/I, (Ohm’s Law:- V = IR or R = V/I). Electrical
resistance to be measured is connected across the generator and in series with deflecting coil.
Experiment no. 8
Aim: - Measurement of power in a 3 phase ac circuit by 2 watt meter method .
Required Equipment:-
Theory:-
Three-phase balanced networks are used in the power industry for reasons of economy and
performance. Three-phase generators and motors run smoothly, with no torque pulsations,
unlike single phase machines. In addition balanced three phase systems may be operated as
three wire or four wire systems, with much less copper needed for the power delivered as
compared with three single phase systems.
At a power generating plant, the windings of a three phase machine are arranged to provide
three voltages, each 120° apart in time and, in the common balanced system, usually all of the
same magnitude. These three voltage sources may be connected in a wye (Y) or a delta (∆)
configuration. Three phase loads may also be connected in wye or delta connections. The wye
connection has a central node to which a neutral wire may be joined, but the delta connection
is a three wire system without a node for a neutral or ground) connection.
To measure power in a 3-phase system, it would seem necessary to use three wattmeters, each
connected to neutral for a common terminal, and each responding to a line-to-neutral voltage
and a line current. One would then add up the powers indicated on each wattmeter. Analysis of
such a circuit shows that one wattmeter is redundant, hence the two-wattmeter method of
measuring 3-phase power was developed for three wire systems. This method is satisfactory
even if the loads are unbalanced. It is necessary to connect the wattmeters taking into account
the polarity of their coils. When the current enters the marked terminal of the current coil and
the voltage positive is connected to the marked terminal of the voltage coil, the reading
represents power absorbed. In that case the algebraic sum of the wattmeters determines the total
load power. In reactive circuits it may be necessary to reverse the current coil of one wattmeter
in order to get an upscale deflection. This reading is taken as negative when the total power is
determined algebraically.
If a 3-phase system has four wires, it is necessary to use three wattmeters, unless it is known
that the system is balanced and therefore no current is flowing in the neutral wire. For any
balanced N wire system it is necessary to use N − 1 wattmeters to measure the total power.
1. Assume a phase voltage of 120 V (a line voltage of 208V) in Figure 1.1 and that the three
resistors have values of 800 Ω. Calculate the expected values of I1 = I2 = I3 for the completely
balanced circuit.
2. Review the two power meter method for measuring three phase power. Determine how to
connect the meters into the circuits of figures 1.1 and 1.2 to measure power supplied by the
variac. The power meters we use will also read out the voltage and currents they are measuring
but you will need to connect DMVs to measure the remaining phase voltage and line current as
well as to measure the additional voltages and currents asked for in the lab description. (Line
voltages: VAB, VBC, VCA. Phase Voltages: VAN, VBN, VCN. Powers: W1, W2. Line currents: I1, I2, I3.
Phase Currents: IP1, IP2, IP3. Neutral current: IN) Print out these circuits and indicate on them
where your watt meters and DVMs will connect.
3. Under what conditions will one of the watt meters in a two-wattmeter measurement read
negative powers with a balanced source feeding a balanced load?
1. Set Load Rack switches so that all 3 resistances are nominally identical. Measure the resistor
values before the experiment; their values should be closely matched.
2. Connect the three-phase wye circuit as shown in figure 1.1. Connect in the power meters and the
DVM's to allow measurement of the power flowing into the load, the line voltages (VAB, VBC,
and VCA), the phase voltages across the resistors (VAN, etc.), neutral current (IN) and the line
currents.
Note: It is important to monitor the current through power meters to ensure that it doesn't exceed
the rated current. Low power can be observed when there are large voltages and large currents if
there is a low power factor. Note that all measurements in this experiment are AC. Estimate all
instrument readings for a source phase voltage of 120V (line voltage of 208V between phases).
Select your meter scales accordingly.
3. A voltage distribution panel is located on the side of the bench. Use a voltmeter to verify that the
voltage is 208 volts between lines. Connect the three-phase variac to voltage distribution panel.
4. Carefully adjust the variac output voltage to a phase voltage of 120V (208V line voltage).
5. Without connecting the neutral switch at open position, measure and record all currents, voltages
(line and phase) , and power it with different balanced loads of the resistor load cart. Record the
results in table 1.1. Turn down the variac and shut the power off.
Note: Measuring power requires the measurement of Voltage, Current, and the phase between
them. The Fluke meter has a current clamp which is an inductive pick up that converts current to
voltage for measurement by the instrument.
The clamp has two scale settings and it is important to check that the meter is set to the same scale as
the current clamp. The small black box meters should have their current “coil” connections in series
with the circuit. For most measurements (all in this lab) you will want to short the input current
connection to one of the voltage “coil” connections.
These meters turn on when the voltage exceeds about 65 Volts. They do not read negative power
(power flow from the load towards the source). If the meter shows current and voltage but no power
then the direction of current through the device must be reversed. The watt meter reading should then
be considered negative.
Both Current and Voltage can be very high while there is still almost no power dissipated in the circuit
when they are out of phase (low power factor). Hence it is important to always monitor the voltage,
current, and power to ensure that none of them exceeds the ratings of the power meters.
6. Move switch to closed position to connect an ammeter from the neutral of the resistor circuit to
the neutral of the three-phase variac and observe the current flow. The current should be read on
the 300 mA (or lower) scale.
7. Measure all currents, voltages and power readings at same load settings of resistor load cart from
step 5. Record all measurements in table 1.1. Turn down the variac and shut the power off.
8. Connect the 3-phase circuit as shown in figure 1.2. Raise the line voltage to 120 Volts (phase
voltage of 69.3V). Measure and record all currents, voltages and power readings at the same
balanced load settings of resistor load cart from step 5.
Note: There will not be enough ammeters for measuring all the phase currents and phase
voltages at the same time. Measure the phase currents first, then reconnect to measure the phase
voltages.
Report:-
1. Why do we use 208 V for the line voltage on the "wye" circuit but only 120 V for the line
voltage on the "delta" load?
2. Calculate the total load power in a wye(ү) and delta(Δ) configuration at each balance load from
experiment, using the current and voltage data, by two different methods.
3. Tabulate the total load power from the calculations from previous question and from the two-
wattmeter measurement method. Discuss any differences.
4. Verify phase and line voltage/current relationship of wye(ү) and delta(Δ) configuration circuits.
Table 1.1: Data sheet for Y and Δ connected load.
Y Y Δ
w/o w/ connection
neutral neutral
Line Voltage Vab
in volts Vbc
Vca
Phase Voltage VAN —
in volts VBN —
VCN —
Powers W1
in watts W2
Line/Phase I1 /Ip1
Currents I2 /Ip2
in
amps I3 /Ip3
— —
IN — —
Resistor RA
in ohms RB
RC
Discussion Questions:-
1. Discuss any differences or similarities for the data obtained for the Y connection with or without
neutral connection.
2. Would the results be affected if wattmeter 2 were placed to measure the line current B-B’ and
both wattmeter potential coils were brought to line C, instead of line B.
3. Show a diagram for using only one wattmeter to measure the power in one phase of a balanced
three-phase load.
Experiment no. 9
Apparatus:-
1. Voltmeter – (0-300V) MI
2. Ammeter – (0-20A) MI
3. 1-Ф variac – (230V / 0-270V)
4. Rheostats – 350Ω / 1.2A, 29Ω / 4.1A, 10Ω / 1.2A
5. Stop watch
6. Connecting wires
Circuit Diagram:-
Theory:-
Energy meter mainly consists of 4 parts of the operating mechanism. They are
1. Driving system
2. Moving system
3. Breaking system
4. Registering mechanism
The driving mechanism of the meter consists of two electromagnets. The core of these
electromagnets is made up of steel laminations. The coil of one of the electromagnets is excited
by the load current. This coil is called the current coil. The coil of second electromagnet is
connected across the supply voltage and therefore carries a current proportional to the supply
voltage. This coil is called th pressure coil. Consequently the two electromagnets are known as
series and shunt magnets respectively. Copper shading bands are provided on the central
limb.The function of these bands is to bring the flux produced by the shunt magnet exactly in
quadrature with the applied voltage.
Moving System:
This consists of an aluminum disk mounted on a light alloy shaft. This disc is positioned in the
air gap between series and shunt magnets. The ‘Al’ disc moves in the field of this magnet and
thus provides a breaking torque.
Braking System:
A permanent magnet positioned near the edge of the aluminum disc forms the braking system.
The aluminum disc moves in the field of the magnet and thus provides a braking torque. The
position of the permanent magnet is adjustable and therefore braking can be adjusted by shifting
the permanent magnet to different radial positions.
Registering Mechanism:
Procedure:-
Tabular Column:-
S.No Voltage Curren Time taken No. of Actu Energy Error %Error
t for revolutio al register WR -
revolutions ns Energ ed by Wa(W)
(S) y Wa energy
(W) meter
WR (W)
Theoretical Calculations:-
Where V is voltage
I= Current (A)
t = Time (S)
Energy registered by energy meter WR = n / meter constant
Error = WR - Wa
% Error = [(WR - Wa) / Wa] * 100
Result:-The error and percentage errors are determined for single phase energy meter at
5%, 25% and 125% of full load current.
Experiment no. 10
Aim:- To measure the inductance and power factor of the choke coil using 3Ammeter and
3Voltmeter method.
Apparatus:-
Theory:-
3 — Ammeter method:
12 = I R2 + I L2 + 2 I L I R
Cos G L Cos 4› L = 12 — I R2
—I L2 /2ILIR
Power drawn the load = VI L Cos & L
= I R R I L Cos 4› L
Since power = I R I L R (I2 — I R2 — I L2 / 2 I
L I R)
= (12 — I R 2 — I L 2) R/2.
From the power calculated the inductance of the choke can be calculated.
3 — Voltmeter method:
Circuit Diagram:-
3 - Voltmeter Method
3 - Ammeter Method
3-Ammeter Method:
Voltag I IL IR COSG PL
e
25
50
75
100
125
150
175
COS L — I° — I R 2 — I L 2 / 2 I L I R
— (I 2 — I R° — I L 2) R/2.
3-Voltmeter Method :
Voltage I V VL COS G PL
R
25
50
75
100
125
150
Result:-