Combination Imaging System

Technical Manual- Appendix


2017.9.14 Document Ver.1.4

Copyrightⓒ2015 by GENORAY Co.,Ltd. All Rights Reserved


Appendix 1. Alignment & Calibration - Panoramic

Appendix 1. Alignment & Calibration – Panoramic

Table of Contents
1. INTRODUCTION ............................................................................................................................................. 3
1.1 General term.......................................................................................................................................................... 3
1.2 Preparation............................................................................................................................................................ 3
1.2.1 Utility tool ................................................................................................................................................... 3
1.2.2 JIGs and PHANTOMs................................................................................................................................... 4

2. Device connection ......................................................................................................................................... 5


2.1 PAPAYA_REC program and how to connect with the machine. .................................................................................. 5

3. Device connection ......................................................................................................................................... 8


3.1 PAPAYA_REC program and how to connect with the machine ................................................................................... 8
3.2 Preparing .............................................................................................................................................................. 9

4. Alignment & Calibration ........................................................................................................................... 13


4.1 Step1 : Gantry reference positioning.......................................................................................................................13
4.2 Step2 : Adjustment of center of rotation..................................................................................................................14
4.3 Step3 : Collimator reference positioning..................................................................................................................18
4.4 Step4 : Detector reference positioning ....................................................................................................................19
4.5 Step5 : Detailed Offset Angle adjustment ................................................................................................................21
4.6 Step6 : Panoramic Laser alignment ........................................................................................................................22
4.7 Step7 : Panoramic Collimator alignment..................................................................................................................23
4.8 Step8 : Panoramic Detector Calibration ..................................................................................................................24
4.8.1 Detector Model : Extor ................................................................................................................................24
4.8.2 Detector Model : I3 .....................................................................................................................................25
4.9 Step9: Auto Shading Calibration .............................................................................................................................26
4.10 Step10: How to check Panoramic Alignment & Calibration......................................................................................27
4.11 Offset Overall : Live image and manual input of Offset parameter ...........................................................................28

PAPAYA Technical Manual (ver 1.4) 2 / 69


Appendix 1. Alignment & Calibration - Panoramic

1. INTRODUCTION
1.1 General term
This manual includes necessary contents, especially machine alignment and calibration, of when production and
installation of machine. The user must acknowledge the appropriate ways to protect oneself from the potential
danger that could occur at any time, and be aware of the condition of the product that can cause danger.

1.2 Preparation

1.2.1 Utility tool

[(+) Driver] [(-) Driver]

[Wrench Set]

3 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 1. Alignment & Calibration - Panoramic

1.2.2 JIGs and PHANTOMs

[Theta Jig] [SD-Line Jig]

[Ball Phantom] [Laser Jig]

PAPAYA Technical Manual (ver 1.4) 4 / 69


Appendix 1. Alignment & Calibration - Panoramic

2. Device connection
2.1 PAPAYA_REC program and how to connect with the machine.

Before Alignment & Calibration, Check surely whether this equipment keeps horizontal by using a level, and
then do next steps.

[Figure 2.1 PAPAYA REC program]

1) “C:/GENORAY/PAPAYA”. Inside of the folder, find PAPAYA_REC.exe and execute the file.
2) Select About-→ Detector Option

5 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 1. Alignment & Calibration - Panoramic

In case of the using current Extor Detector, Only 'Extor(Laon)' of Use Detector Option is applied.
(Extor Imi is not available)

3) Select machine

4) After execution of the machine alignment program, push [Connect]


button where it is located at middle upper side and connecting program with machine.

PAPAYA Technical Manual (ver 1.4) 6 / 69


Appendix 1. Alignment & Calibration - Panoramic

[Figure 2.2 Vesion Info Confirmation Window ]

5) Click About → Version Info to verify the version of the Detector and each Board.

7 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 1. Alignment & Calibration - Panoramic

3. Device connection
3.1 PAPAYA_REC program and how to connect with the machine

[Figure 3.1 Panoramic tab’s Location]

1) After connecting with machine, refer to Figure 3.1 and choose Panoramic tab.

[3.2 Calibration Wizard’s Location]

2) At Panoramic tab, on right down corner, click [Calibration Wizard] button Figure 3.2.

PAPAYA Technical Manual (ver 1.4) 8 / 69


Appendix 1. Alignment & Calibration - Panoramic

3.2 Preparing

[Figure 3.3 Preparing1 tap -Panoramic]

Before machine alignment and modification, this explains necessary tools.

No. Name Description


1 Theta jig block Using for machine’s alignment of rotating degree to starting point.
Using for machine’s center of rotation alignment and for alignment of
2 SD-line jig
ball phantom’s position.
3 Ball phantom Using with SD-line jig and use for chin-rest’s alignment.
4 Laser alignment jig Using as put above Ball phantom and use for laser alignment.

[Table 3.1 Preparation Tools]

9 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 1. Alignment & Calibration - Panoramic

- How to remove the cover and Generator Driving is as follows:

Remove upper cover of driving.


On circled with red, remove bolts and remove cover by
pulling up.

Remove cover, following the steps. The removal method is as


below contents

After removing red circle and their bolts, and remove front
cover of where X-ray generator.

After removing of red circle and remove bolts and on step


by step removing cover.

PAPAYA Technical Manual (ver 1.4) 10 / 69


Appendix 1. Alignment & Calibration - Panoramic

[Figure 3.4 Preparing2 tap-Panoramic]

Before alignment of machine, on patient’s support part, and need to remove parts, in which are machine’s
cover and handle, and explains how to remove them.

- How to remove the cover Chinrest are as follows:

After removing bolts on red circles and remove the handle.

On red circle, remove indicated bolts and remove cover on that


parts.

11 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 1. Alignment & Calibration - Panoramic

On red circle, remove indicated bolts and remove cover on that


parts.

On red circle, remove indicated bolts and remove below cover.

On red circle, remove indicated bolts and remove upper cover.

PAPAYA Technical Manual (ver 1.4) 12 / 69


Appendix 1. Alignment & Calibration - Panoramic

4. Alignment & Calibration


4.1 Step1 : Gantry reference positioning


[Figure4.1 Y-Axis adjusting of standing point - Panoramic]

1) After preparing all preparation, set up Y-Axis starting point.


2) Machine’s OP and using machine’s [FORWARD], [BACKWARD] button and move machine’s Y-Axis to front and
back to match of indicated driving part’s and aluminium block for Y-axis, refer to Figure 4.1
3) Check kV offset figures written on x-ray tank, input the figures of [Exposure Condition Offset] in ⓑ.
4) Click [Set current position] in ⓒ, and then the position of Y axis is shown in the screen and saved to
equipment.

For checking kV offset figures on the X-ray generator, Take away X-ray generator part's cover first and

Check.

13 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 1. Alignment & Calibration - Panoramic

4.2 Step2 : Adjustment of center of rotation

This tool use for processing this stage

Theta Jig SD-Line Jig Ball Phantom

[Figure4.2 Using Theta jig-Panoramic]

1) Fix Theta jig to driving side and fix with bolts


2) Then, installing ‘Theta jig block’ if because Y-Axis is coming front and rotating side and Theta jig
conflicts, and then using OP’s [BACKWARD] button and move Y-Axis to backward and make installation
3) On rotating part and ‘Theta jig block’ meets and parallel each other, and then push ⓐ [Set current
position] button and save current rotating side’s location by pushing button.

PAPAYA Technical Manual (ver 1.4) 14 / 69


Appendix 1. Alignment & Calibration - Panoramic

[Figure 4.3 Using Ball phantom and SD-Line jig-Panoramic]

4) Ball phantom and SD-line jig is installed on machine


5) At this point, the home -wave excitation and installation carefully the face of the fixed part of the SD-
line jig to exactly match .
6) Once the installation is complete, Ball phantom of SD-line jig position of the jaw with the two screws that
secure the adjustable jaw helps to loosen

See the Figure below to align the SD-Line jig in the hall of the Ball phantom.

15 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 1. Alignment & Calibration - Panoramic

[Figure 4.4 Ball phantom and SD-Line jig’s starting point]

1) Adjusting as repeating below steps to align ‘Ball phantom’ and rotating axis’ location.
2) By pushing ⓐ [0°] button, making perpendicular between rotating part and driving part.
3) SD-line jig’s line and Ball phantom’s reference line need to be matched by turning Ball phantom
stage’s knob and move Ball phantom’s location.
4) ⓑ [180°] button need to be pressed and rotating part move to 180°.
5) If, at 0° location and 180° location, SD-line jig is not on same location, on 0° adjusted Ball phantom’s
reference line and at 180° location. ‘Ball phantom’s reference line and SD-line jig’s line that needs to be
located by turning each knob and locating them to the middle of ‘SD-line jig’s locations of 0° and 180°.
6) By repeating 1-4’s to reduce the error.

PAPAYA Technical Manual (ver 1.4) 16 / 69


Appendix 1. Alignment & Calibration - Panoramic

[Figure 4.5 Fixing Ball Phantom’s location]

7) SD-line jig is located on machine’s rotating axis. SD-line jig line is at Ball phantom matching with appropriate
reference line and adjust chin-rest’s location and tighten chin-rest’s bolt. Then, need to be caution to not
move Ball phantom knob.
8) After chinrest is tightly fixed, again at ‘0°’, ‘90°’, ‘180°’ location, and checking ‘Ball phantom’ and ‘SD-line jig’ are
matching. If location is not matching, again chin-rest’s bolt need to be unscrew and adjusted.

17 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 1. Alignment & Calibration - Panoramic

4.3 Step3 : Collimator reference positioning

[Figure 4.6 Alignment of Collimator Position-Panoramic]

1) Click ⓐ [Find Collimator Offset] button


2) After clicking button, the machine is automatically finding for Collimator’s position by proceeding necessary
steps and displays ⓑ [Adult Panoramic collimator position]’s value
3) In this condition, click ⓒ [Set Offset & Get Image] button then X-ray is exposed and display an image. Then
distinguishing whether collimator’s is good position by checking with left illustrated picture.
4) Collimator well find that the position of the judge on the basis of ⓓ.

If Collimator’s position is well found, according to picture’s Good position, on image, by Collimator, X-ray’s

exposure’s limited area is almost same with left area and right area.

PAPAYA Technical Manual (ver 1.4) 18 / 69


Appendix 1. Alignment & Calibration - Panoramic

4.4 Step4 : Detector reference positioning

[Figure 4.7 Alignment of detector position-Panoramic]

1) Click ⓐ [Rotate 0 ° & Get Image] button to expose X-ray and to get an X-rayed image.
2) Select line1 of ⓑ [Select Line] Combo box located on bottom of display window and then click the center of
the pin, imaged by X-ray exposure, then blue dashed line is being on the center of the pin.
3) Click ⓐ [Rotate 180 ° & Get Image] button to expose X-ray and to get an X-rayed image.
4) Select line2 of ⓑ [Select Line] Combo box located on bottom of display window and then click the center of
the pin, imaged by X-ray exposure, then yellow dashed line is being on the center of the pin.
5) Red line is being on the middle of two lines between blue and yellow line. Check whether the red line is
located inside of two green line or not (Pass/Fail information will be displayed).
6) If fail, refer to ⓔ then loosen screws marked with red circle and move the detector position left
or right and repeat 1-5 until Pass is displayed.

19 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 1. Alignment & Calibration - Panoramic

[Figure 4.8 Alignment of detector position]

8) In the condition Pass in on, refer to Figure 4.8, join the red circled screws and put the detector c
over on.

PAPAYA Technical Manual (ver 1.4) 20 / 69


Appendix 1. Alignment & Calibration - Panoramic

4.5 Step5 : Detailed Offset Angle adjustment

ⓐ ⓓ

ⓑ ⓒ

[Figure 4.9 Detailed Offset Angle adjustment]

1) Click ⓐ [Set angle and Get image] to expose X-ray and to get an X-rayed image of degree 90°.
2) Select blue line of ⓑ [Select Line] Combo box located on bottom of display window and then click the center
of the pin, imaged by X-ray exposure, then blue dashed line is being on the center of the pin
3) Select yellow line of ⓑ [Select Line] Combo box located on bottom of display window and then click the
center of the metal ball, imaged by X-ray exposure, then yellow dashed line is being on the center of the pin
4) Make a judgment of Pass/Fail. If failed, refer to Figure 4.9, then adjust ⓓ [Gantry rotation angle offset of 90°]
value slightly little. Repeat 1-4 until Pass is displayed

Change offset angle

- On the right : less than 15 pixels

- On the left : increase than 1pixels

21 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 1. Alignment & Calibration - Panoramic

4.6 Step6 : Panoramic Laser alignment

This tool use for processing this stage.

Laser Jig Ball Phantom

[Figure 4.10 Laser alignment]

1) Click ⓐ [Laser alignment position]


2) Ball phantom is installed on Chinrest
3) Laser jig is installed on Ball phantom
4) Install 'Laser Alignment jig' on the Ball phantom and adjust laser position, refer to Figure 4.10

PAPAYA Technical Manual (ver 1.4) 22 / 69


Appendix 1. Alignment & Calibration - Panoramic

4.7 Step7 : Panoramic Collimator alignment

[Figure 4.11 Collimator alignment]

1) Click ⓐ [Find Collimator Offset] button to find collimator position


2) Click ⓑ [Set Offset & get Image] button to expose X-ray and to get an X-rayed image.
3) Make a decision whether collimator position and shape is right or not, refer ⓒ. If the collimator is
rotated or sided to one side then refer to ⓓ image and adjust collimator position correctly, refer to
Figure 4.11
4) Judge whether it is Pass or Fail. If failed, repeat 1-3 until Pass is displayed.

If you find the wrong position Collimator , refer to the Recommended collimator 'move'
position value , Adult panoramic collimator position type .

23 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 1. Alignment & Calibration - Panoramic

4.8 Step8 : Panoramic Detector Calibration

4.8.1 Detector Model : Extor



ⓑ ⓗ

[Figure 4.13 Extor Detector Calibration-Panoramic]

1) Select the item in the Name Preset Option to ⓐ [PANO_PANORAMA].


2) Click the ⓑ [Check All] of X-Ray Option click the ⓒ [Exposure] button to shoot the X- ray .
3) After finishing an exposure, Avg.Level in ⓔ for whole image is shown, the sheets of images for Avg is shown
in the graph in ⓗ. If there is warming message of Saturation after getting Avg. figure, re-take an exposure by
lower condition following the manual.
4) Of X-Ray Option ⓓ [Ref.Level] Compare the value of Level
5) ⓔ [Avg.Level] value Ref. If within 30 Level ± the standard values ⓕ [... ] button to confirm saving path is
correct , click the ⓖ [Create] button.
5) If ⓔ [Avg.Level] value is greater than or less than ± 30 mA and kVp by changing the value of the X-Ray
position Adjust brought into tolerance.
6) ⓓ [Ref. Level] Align the reference value , click the ⓖ [Create] button shown in the Figure below, by the
Panorama Calibration Complete .
7) Once completed, you will be prompted to "Parameters are generated" as ⓗ of Figure 4.13.

If the gap between output value and initial value is lower than 50 → change mA value
If the gap between output value and initial value is higher than 50→ change kVp value

PAPAYA Technical Manual (ver 1.4) 24 / 69


Appendix 1. Alignment & Calibration - Panoramic

4.8.2 Detector Model : I3

[Figure 4.12 I3 Detector Calibration-Panoramic]

1) Ref Make sure that the Dark Ref. Level value is 0, Bright1 is 200, Bright2 is 800 and press 'Auto' button
2) ⓐ [Auto] button to perform calibration automatically
3) Click [OK] button.

25 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 1. Alignment & Calibration - Panoramic

4.9 Step9: Auto Shading Calibration

[Figure 4.14 Collimator shading calibration]

Step9 is where to calibrate attenuation of X-ray due to collimator.


ⓐ [Auto shading calibration] button will proceed automatically.

PAPAYA Technical Manual (ver 1.4) 26 / 69


Appendix 1. Alignment & Calibration - Panoramic

4.10 Step10: How to check Panoramic Alignment & Calibration

This tool use for processing this stage

Ball Phantom

[Figure 4.15 Final test-Panoramic]

1) Fix “Ball Phantom” on the chinrest and click ⓐ [Confirm]


2) After Confirming, align canine laser to the frond reference side of ball Phantom by moving Y-axis
3) Click ⓒ [Ready & Shot] to take an image of Ball Phantom.
4) After getting an image of ball Phantom, the screen shows the image as Figure 4.15
5) Select the third line and draw four lines on the image.
6) After drawing four lines, If the message, marked 4, shows “PASS” three time, the panoramic alignment and
calibration will be finished.
7) If any “FAIL” message occurs, back to STEP2, STEP3, and STEP7 with instruction given.

27 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 1. Alignment & Calibration - Panoramic

4.11 Offset Overall : Live image and manual input of Offset parameter

[Figure 4.16 Offset Overal]


Offset Overall , the offset value by manually entering all the conditions and X- ray equipment to the value
of the sort you want you can shoot a live video.

[Getting Dark Auto Image]


In that getting dark image,
1) Click ⓑ [Dark on],
2) And you can get image with automatically set condition

[Getting X-ray Manual Image]


1) Set figures at ⓒⓓⓔ,
2) Click ⓐ [Get Xray Image] in order. And you can get an image.

PAPAYA Technical Manual (ver 1.4) 28 / 69


Appendix 2. Alignment & Calibration - Cephalo

Appendix 2. Alignment & Calibration – Cephalo

Table of Contents
1. Introduction ................................................................................................................................................... 30
1.1 Regular condition ..................................................................................................................................................30
1.2 Preparation...........................................................................................................................................................30
1.2.1 Required Tools ...........................................................................................................................................30
1.2.2 JIG ............................................................................................................................................................30

2. Device connection ....................................................................................................................................... 31


2.1 PAPAYA_REC program and how to connect with the machine. .................................................................................31

3. Preparation process .................................................................................................................................... 33


3.1 How to run the program ........................................................................................................................................33
3.2 Preparing .............................................................................................................................................................34

4. Alignment & Calibration ........................................................................................................................... 36


4.1 Step1 : Detector reference positioning ....................................................................................................................36
4.2 Step2 : Step3 : Alignment of 1st Collimator positioning ............................................................................................37
4.3 Step3 : Alignment of Ear rod ..................................................................................................................................38
4.4 Step4 : 2nd Collimator reference positioning ...........................................................................................................38
4.5 Step5 : Detector Calibration ...................................................................................................................................39
4.5.1 Detector Model : Extor ................................................................................................................................39
4.5.2 Detector Model : I3 .....................................................................................................................................40
4.6 Step6 : Cephalo Scan FFC calibration ....................................................................................................................41
4.7 Step7 : How to check Cephalo Alignment & Caliration .............................................................................................42
4.8 Offset Overall........................................................................................................................................................46

29 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 2. Alignment & Calibration - Cephalo

1. Introduction
1.1 Regular condition

This manual includes necessary contents, especially machine alignment and calibration, of when production and
installation of machine. The user must acknowledge the appropriate ways to protect oneself from the potential
danger that could occur at any time, and be aware of the condition of the product that can cause danger.

Before reviewing this Appendix 2, read “Appendix 1 Align & Calibration –Panoramic” and familiar with all
calibration steps first.

1.2 Preparation
1.2.1 Required Tools

[Wrench Set] [(+) Driver]

1.2.2 JIG

[Pin Jig]

PAPAYA Technical Manual (ver 1.4) 30 / 69


Appendix 3. Sesonging

2. Device connection
2.1 PAPAYA_REC program and how to connect with the machine.

[Figure 2.1 PAPAYA REC program]

1) “C:/GENORAY/PAPAYA”. Inside of the folder, find PAPAYA_REC.exe and execute the file.
2) Select About → Detector Option

31 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 2. Alignment & Calibration - Cephalo

3) Select machine

4) Click About → Version Info to verify the version of the Detector and each Board.

PAPAYA Technical Manual (ver 1.4) 32 / 69


Appendix 3. Sesonging

3. Preparation process
3.1 How to run the program

[Figure 3.1 Cephalo tap’s location]

1) After connecting with machine, refer to Figure 3.1 and choose Cephalometric tab.

[Figure 3.2 Calibration Wizard’s location]

2) At Cephalometric tab, on right down corner, click [Calibration Wizard] button Figure 3.2.

33 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 2. Alignment & Calibration - Cephalo

3.2 Preparing

[Figure 3.3 Preparing1 - Cephalo]

1) Preparing the Calibration Wizard appears when running a screen like Figure 3.3.
2) Refer to Figure 3.3 to remove the cover of the Cephalo.

PAPAYA Technical Manual (ver 1.4) 34 / 69


Appendix 3. Sesonging

- The image on bottom side explains how to remove the covers ahead to calibration and alignment.

Remove the top cover.


The position is marked with a red circle and remove the
screws to remove the cover and lift it up.

Remove the cover of the 2nd collimator. Remove the


screws on the red circle location Peel off the cover.

35 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 2. Alignment & Calibration - Cephalo

4. Alignment & Calibration


4.1 Step1 : Detector reference positioning

This tool use for processing this stage

Pin Jig

[Figure 4.1 Setting of refetence position of detector – Cephalo]

1) Match Detector Moving Block’s hole with the reference point of Cephalo Base as Figure 4.1
2) After then, make them to be aligned by the Pin Jig and then remove the Pin Jig. Select [Set current position]
3) When you click [Set current position] Result detector offset value will be appeared.
4) This value is a reference position of the detector.

PAPAYA Technical Manual (ver 1.4) 36 / 69


Appendix 3. Sesonging

4.2 Step2 : Step3 : Alignment of 1st Collimator positioning

ⓐ ⓑ

[Figure 4.2 1st Collimator reference positioning - Cephalo]

1) Click [Find Collimator Offset] button then procedure will proceed automatically.
2) When automatic procedure is finished, then click [Set Offset & Get image] button to check 1st collimator is on
proper position.

37 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 2. Alignment & Calibration - Cephalo

4.3 Step3 : 2nd Collimator reference positioning

In case the location of


collimator is incorrect.

In case the position of collimator In case the location of


is perfectly correct. collimator is incorrect.

[Table 4.1 2nd Collimator’s reference positioning]

ⓐ ⓑ

[Figure 4.6 2nd Collimator alignment - Cephalo]


1) Click ⓐ [Find 2nd Collimator position] to find 2nd Collimator value.
2) Click ⓑ [Set Offset & Get Image] button to get an X-ray exposed image.
3) Figure out that the collimator has a right shape and is on a proper position. If it has not a right shape or be
sided. Then adjust 2nd collimator position to right position, refer to the bottom-left image of ⓒ

PAPAYA Technical Manual (ver 1.4) 38 / 69


Appendix 3. Sesonging

4.5 Step5 : Detector Calibration


4.5.1 Detector Model : Extor


ⓕ ⓔ


[Figure 4.8 Extor Detector Calibration-Cephalo]

1) Select an item in the Name Preset Option to ⓐ [CEPH_CEPHALO].


2) Click the ⓑ [Check All] of the X-Ray Option ⓒ [Exposure] click to shoot X- ray .
3) After finishing an exposure, Avg.Level in ⓓ for whole image is shown, the sheets of images for Avg is shown in
the graph in ⓙ.
If there is warming message of Saturation after getting Avg. figure, re-take an exposure by lower condition
following the manual.
4) X-Ray Option ⓔ [Ref.Level] Compare the value of Level
5) The value ⓓ [Avg.Level] ⓔ [Ref.Level] If within 30 Level ± the standard values ⓕ [....] Click the Save button to
confirm the correct path, click ⓕ [Create] button.
6) If greater than or less than the value ⓓ [Avg.Level] ± 30 mA and kVp by changing the value of the X-Ray
position Adjust brought into tolerance.
7) Ref. Level Align the reference value shown in Figure 4.8 , click the Create button to complete the Cephalo
Calibration.
8) Once completed, you will be prompted to "Parameters are generated" as ⓗ

39 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 2. Alignment & Calibration - Cephalo

If the gap between output value and initial value is lower than 50 → change mA value
If the gap between output value and initial value is higher than 50→ change kVp value

4.5.2 Detector Model : I3

[Figure 4.7 I3 Detector Calibration-Cephalo]

1) Ref Make sure that the Dark Ref. Level value is 0, Bright1 is 200, Bright2 is 800 and press 'Auto' button
2) ⓐ [Auto] button to perform calibration automatically
3) Click ⓑ [OK] button.

PAPAYA Technical Manual (ver 1.4) 40 / 69


Appendix 3. Sesonging

4.6 Step5 : Cephalo Scan FFC calibration

[ Figure 4.9 Cephlao Scan FFC Clibration]

Step 6 performs FFC(Flat Field Correction) and Scan calibration to calibration 2nd collimator position of scanning
and to equalize brightness of cephalometric image.

set ear rod and nasal positioner to AP direction and click ⓐ [Auto scan/FFC calibration] button, then procedure
will proceed automatically.

41 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 2. Alignment & Calibration - Cephalo

4.7 Step6 : Cephalo Alignment & Calibration


4.7.1 Ear rod and Cephalo Calibration reference positioing

1) Adjust the height of two ear rods so that they have same length

2) Adjust the position of the Laser.

PAPAYA Technical Manual (ver 1.4) 42 / 69


Appendix 3. Sesonging

ⓒ ⓐ

[Figure 4.4 Alignment of Ear rod]

3) Ear rod and the Lateral nasal positioner to position the shooting location .
4) Click ⓐ [Confirm] button
5) Click ⓑ [Ready & Shot] button when it is ready.
6) Select blue cross of ⓒ [Select Cross] combo box which is located under the image window and click
center of one of two ear posts.
7) Select yellow cross of ⓒ [Select Cross] combo box which is located under the image window and click
center of another one.

43 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 2. Alignment & Calibration - Cephalo

8) If fail message is shown, Follow the second fig. by [Help] button and remove Nasal postinoner.
9) Loosen the bolt a bit and move left&right to adjust range of Ear rod like a third fig.
10) Re-take an exposure after taking Ear rod.

PAPAYA Technical Manual (ver 1.4) 44 / 69


Appendix 3. Sesonging

[Figure 4.10 Final image test – Cephalo]

11) If the position of the Ear rod is confirmed to PASS, adjust the brightness of the displayed image with the
right mouse button to verify that the quality is equal to the image shown below.
If the images such as ⓒ is released, Cephalo Alignment & Calibration were completed without defect.
12) If the images are not normal, back to step5 and do step again repeatedly.
13) If the images is still not normal after step5, back to step1 and do step again.

45 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 2. Alignment & Calibration - Cephalo

4.8 Offset Overall

[Figure 4.16 Offset Overal]


Offset Overall , the offset value by manually entering all the conditions and X- ray equipment to the value
of the sort you want you can shoot a live video.

[Getting Dark Auto Image]


In that getting dark image,
1) Click ⓑ [Dark on],
2) And you can get image with automatically set condition

[Getting X-ray Manual Image]


1) Set figures at ⓒⓓⓔ,
2) Click ⓐ [Get Xray Image] in order. And you can get an image.

PAPAYA Technical Manual (ver 1.4) 46 / 69


Appendix 3. Sesonging

Appendix 3. Seasonging

Table of contents

1. Ready for seasoning ................................................................................................................................... 48


2. Running Seasoning ..................................................................................................................................... 49
3. Completion of Seasoning .......................................................................................................................... 50
4. Notes ............................................................................................................................................................... 50

47 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 3. Sesonging

This manual contains general usage descriptions and other information about Seasoning.
You have to finish the process (Align & Calibration) first and keep power off for a long time to install.

1. Ready for seasoning

1) After turning on the appliance, run the PAPAYA_REC.exe file in C:\GENORAY\PAPAYA folder.

2) When the program runs as shown below, click the [Connect] button to connect to the equipment.

3) When connected to the equipment, the [Connect] button changes to [Disconnect]

4) Run Seasoning tab.

PAPAYA Technical Manual (ver 1.4) 48 / 69


Appendix 3. Sesonging

2. Running Seasoning

1) Select [START] button in the figure.

2) It proceed like the below.


- 1 : Displaying the number and steps for exposure.
- 2 : Displaying the progress of exposure.
- 3 : Displaying status of waiting time.

49 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 3. Sesonging

3. Completion of Seasoning

After finishing Seasoning, Pop-up is shown. Click ‘OK’.

4. Notes

The below table shows Seasoning steps, please be informed.

The number of Interval of


kV mA Exposure time
exposure exposure
5 10 60 4 0.5
2min for waiting
5 10 75 8 1
2min for waiting
5 10 90 12 1

PAPAYA Technical Manual (ver 1.4) 50 / 69


Appendix 4. Auto Save Raw Manage(DB)

Appendix 4. Auto Save Raw Manage(DB)

Table of contents

1. The screen configuration ................................................................................................................52

51 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 4. Auto Save Raw Manage(DB)

This manual contains information about the confirming and management of RAW files after exposure.
Ensure that this manual is interlocked to the database, please keep fully in mind with the contents of this manual.

1. The screen configuration

The Auto Save Raw Manage tab is divided by Panorama / Cephalo / CT like below.
Set the number of exposure list for the management in [Auto Save management] of PAPAYA OP Setting.
The button instruction is like a below.
1) After connecting DB, renew the list.
2) Open the folder saved for Raw file.
3) After connecting DB, remove the item.

1 2 3

PAPAYA Technical Manual (ver 1.4) 52 / 69


Appendix 5. Remote Update

Appendix 5. Remote Update

Contents

1. Remote update ............................................................................................................................................. 54


2. Run LisaTools................................................................................................................................................. 54
3. Process of Remote Update ....................................................................................................................... 55
3.1 Indivisual Remote update........................................................................................................................................................................................................................... 58
3.2 Selecting Remote update ........................................................................................................................................................................................................................... 63

4. List of reference for firmware .................................................................................................................. 67

53 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 5. Remote Update

This manual includes information and method to use remote update of PAPAYA Series(unify word "PAPAYA").
Keep informed this manual to use equipment effectively.

1. Remote update

1) Find update file.


- Find binary file of remote update.
2) Check network connection.
3) Run remote control program after checking pc network connection for PAPAYA.

2. Run LisaTools

1) Move ‘LisaTools.exe’ to pc wallpaper of equipment’s pc.


2) Run ‘LisaTools’.

[LisaTools ‘exe file’ icon]

PAPAYA Technical Manual (ver 1.4) 54 / 69


Appendix 5. Remote Update

3. Process of Remote Update

1) Press Step1 [Get system Info].

[Execution screen of LisaTools]





On the initial screen, the equipment name is PAPAYA slim

55 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 5 Remote Update

2) As shown in the below, the equipment picture is shown in the center, and the equipment name is displayed in
the upper right corner.
- Check picture and name whether it is same with the equipment you want to update.

[Check equipment name]

In the case of PAPAYA PLUS or SLIM, Check equipment name surely because it is equal to equipment
picture.

3) Like below, current firmware version is shown per B/D.

PAPAYA Technical Manual (ver 1.4) 56 / 69


Appendix 5. Remote Update

- When the light beside each B/D is green, The B/D is possible for remote update.
When the light beside main B/D(FPGA) is grey, it is impossible for remote update, and check if main B/D
update harness(JP02A030) is connected.

[Checking possible B/D for remote update – Impossible for main FPGA update]

[CheckB/D for remote update – It is possible for main FPGA update]

- Once you complete the process, you can run indivisual/selective update.

57 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 5. Remote Update

3.1 IndivisualRemote update

1) Press Step2 [Select File to upload]

[Step2 Run selecting screen for binary file]

2) Select binary file of B/D for update.


ex) Chinrest B/D Update

[Select binary]

PAPAYA Technical Manual (ver 1.4) 58 / 69


Appendix 5. Remote Update

3) Once you select the file, the current version and the version for update are shown.

[Check current version and update version.]

- Compatible : Compatible: firmware compatibility.


If the number of each binary is different, it is impossible for update.
- Major : Version of function.
- Minor : Version of bug revision.

* Refer to "(P.15) Firmware management system"to find binary file name

[Check firmware compatibility]

59 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 5. Remote Update

4) Once you press Step3 [Upload files on boards], the update is run.

[Step3 Upload files on boards - Update]

PAPAYA Technical Manual (ver 1.4) 60 / 69


Appendix 5. Remote Update

5) The checking marks beside each B/D, like below, is shown when finishing update.

[Check finish of Chinrest B/D update]

6) If you want to do another update, please do step2 [Select Files to uploard] (P.69) first.

61 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 5. Remote Update

7) Press step 4 [Reset system] if you no longer proceed with update.


Check if the equipment set initial value after reset.

[Step4Reset(Reboot) system after finishing update.]

8) After finishing Inirializing, turn on and off the power of equipment.

PAPAYA Technical Manual (ver 1.4) 62 / 69


Appendix 5. Remote Update

3.2 Selecting Remote update

1) Press step2 [Select File to updaoad]

[Step2 Select File to upload - Binary File open]

2) Select binary files of B/D for update

[Select twice]

63 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 5. Remote Update

3) Once you finish selecting file, current version and the version for update are shown.

[Versions for select binary B/D]

- Like below, all versions for B/D and update are shown if you select all binary of B/D.

[Select all B/D update binary.]

PAPAYA Technical Manual (ver 1.3) 64 / 69


Appendix 5. Remote Update

4) Once you press step3 [Upload files on boards], the update is started.

[Step3 Upload files on boards - Run update]

- The update process screen is shown from no.1 B/D in order.

[FromMain(MCU) to Main(FPGA), the update process screen is shown.]

65 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 5. Remote Update

5) After finishing update, checking marks of all status lights are shown.

[After updating B/D, check status light.]

6) Press step4 [Reset (Reboot) system].


Check if equipment is initialized after reset

[After finishing update, do step4Reset(Reboot) system]

8) After finishing Initializing, turn on and off the power of equipment

PAPAYA Technical Manual (ver 1.3) 66 / 69


Appendix 5. Remote Update

4. List of reference for firmware

Below is the chart for management system. Please refer to this chart.

File name filename_B/D name_CANID_function_version_system.bin

PPY : PAPAYA
Product name
P3D : PAPAYA 3D
[3 letters with english]
SLM : PAPAYA SLIM

MAIN : MAIN B/D

COLL : COLLIMATOR B/D

B/D name CHIN : CHINEREST B/D

[3 letters with english] LIFT : LIFTER B/D

LANS : LAN SWITCH B/D

FPGA : MAIN B/D FPGA

MAIN : 01

CHIN : 02

COLL : 03
CAN_ID
LIFT : 04

LANS : 05

FPGA :06

BOOT : Bootloader

function APPL : Application Firmware

[4letters with english] FACT : FPGA Factory Image

UFDA : FPGA Application

Version system A_B_C_D

1 : PAPAYA
Number for A-1
2: PAPAYA 3D
[Product classification]
3 : PAPAYA SLIM

Number for B -1 1 : CURRENT VERSION [CAN]

[F/WCompatibility] 2 : REMOTE UPGRADE

Number for C -2
0~99
[Function modification]

In a quality test : B00~B99


Number for D -3
Mass production distribution : 000~999
[Bug modification]
B : For beta version test

67 / 69 PAPAYA Technical Manual (ver 1.4)


Appendix 5 Remote Update

Example to Firmware file name. Refer to chart for management system.

1) In case of Papaya3D Main B/D Application Firmwar (Version : 2_2_5_001)


→Filename : P3D_MAIN_01_APPL_2_2_5_001.bin
- Partial marking ofLisaToolsUpdate(Refer to P.7)
01. Main(MCU)
Compatible : 2
Major : 5
Minor : 001

2) In case of Papaya3D Main B/D FPGA Application Image (Version : 2_2_9_010)


→Filename : P3D_ FPGA_06_UPDA_2_2_9_010.bin
- Partial marking ofLisaToolsUpdate(Refer to P.7)
06. Main(FPGA)
Compatible : 2
Major :9
Minor : 010

3) In case of Papaya Slim Lifter B/D Application Firmware (Version : 3_2_7_003)


→Filename :SLM_LIFT_04_APPL_3_2_7_003.bin
- Partial marking ofLisaToolsUpdate(Refer to P.7)
04. Lifter
Compatible : 2
Major : 7
Minor : 003

PAPAYA Technical Manual (ver 1.4) 68 / 69


Appendix 5. Remote Update

The below chart shows the firmware versioncompatibilityof equipment for remote update.

F/W version for


PAPAYA PLUS PAPAYA SLIM PAPAYA 3D(PLUS)
remote update

MAIN B/D(FPGA) 1_2_9_001 or higher 3_2_9_001 or higher 2_2_9_001 or higher

MAIN B/D(STM) 1_2_0_001 or higher 3_2_0_001 or higher 2_2_5_001 or higher

1_2_2_001(new model B/D) 3_2_2_001(new model B/D) 2_2_2_001(new model B/D)

or higher or higher or higher


LIFT B/D
1_2_1_001(old model B/D) 3_2_1_001(old model B/D) 2_2_1_001(old model B/D)

or higher or higher or higher

CHINEREST B/D - - 2_2_5_001 or higher

COLLIMATOR - - 2_2_2_001 or higher

LAN SW B/D
- - 2_2_5_001 or higher
(LING LED MAIN)

LAN SW B/D
- - 2_2_5_001 or higher
(LIN LED LA)

69 / 69 PAPAYA Technical Manual (ver 1.4)


Genoray Co., Ltd. develops equipments continually. Manual and specifications subject to change
without notice

Genoray Co., Ltd.


512, 560, Dunchon-daero, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea
Tel: +82-31-740-4100 Fax: +82-31-737-8018 [email protected] www.genoray.com

Genoray America Inc.


147 E. Bristol Lane, Orange, CA 92865 USA
Tel: +1-855-436-6729 FAX: +1-714-786-8919 [email protected] www.genorayamerica.com

Genoray EU GmbH
Westhafenstr. 1 13353 Berlin, Germany
Tel: +49-30-509-694-98 Fax: +49-30-530-198-08 [email protected]

Genoray Japan
2F Ishibashi-Bldg, 1-4-15 Shinyokohama,Kouhoku-ku, Yokohama-city, kanagawa, 222-0033 Japan
Tel: + 81-45-620-4971 Fax: +81-45-620-4972 [email protected] www.genorayjapan.co.jp

Boulevard Général Wahis 53, 1030 Brussels, BELGIUM


Tel: +32-2-732-59-54 Fax: +32-2-732-60-03 [email protected]

You might also like