XRF
XRF
Introduction
XRF spectroscopy is a very common, simple, and relatively quick spectroscopic
technique used in a wide range of applications since its invention in the 1920s. It’s based on a
very similar principle to EDX and microprobe WDX spectroscopy, in which the interaction of an
electron beam with the sample produces a series of X-rays that are used to identify the
composition of the sample. In XRF spectroscopy, X-ray radiation is used instead of an electron
beam as the primary excitation source, which simplifies the required equipment and lowers the
cost of sample preparation.
How it works
XRF spectroscopy consists in studying the interaction between the sample and X-ray
radiation. Consequently, every XRF spectrometer is equipped with an X-ray generator, typically
with a Rhodium filament and in the range of 20-60 kV, although W, Mo, Cr and others can also
be used, depending on the application. Gamma ray sources can be used too without the need for
an elaborate power supply, allowing an easier use in small portable instruments.
Much like in the other kinds of spectrometers, the beam travels through a series of lenses
which narrow it and focus it on the surface of the sample. X-ray radiation doesn’t require
vacuum like the electron beam used in microprobe spectroscopy, and the beam usually can’t be
displaced along the surface of the sample like in a SEM. This simplifies and lowers the cost of
the equipment.
When excited with the primary X-ray beam, each element emits a series of characteristic
X-rays with different wavelengths from the primary beam, which are used to determine the
composition of the sample. As explained in the SEM-EDX page, when the primary beam reaches
an atom, it can become ionized (i.e. lose an electron). If the electron is lost from one of the inner
shells, another electron from a further outer shell usually occupies its place. When this happens,
energy is released due to the lower binding energy of the inner electron orbital compared with an
outer one, most commonly in the form of an X-ray. As the energy of the emitted X-rays matches
the energy gaps between the orbitals of the atoms of a specific element, these X-rays can be used
to detect the abundances of elements that are present in the sample.
The detection of the emitted X-rays can be made both via wavelength dispersive (WDX) or
energy dispersive (EDX) detectors. In WDX systems, the different wavelengths are selected one
by one with the help of an analytical crystal and passed on to the detector, which registers the
intensity of the radiation at every wavelength to construct a spectrum. The utilization of this
technique grants a better spectral resolution and a higher quantitative power than EDX, which is
why it is a bit more common in X-ray spectroscopy. For a deeper view on WDX spectroscopy
read the dedicated page on our website.
EDX detection, on the other hand, offers a faster response by yielding the emitted radiation of all
wavelengths at the same time to a semiconductor crystal, in which an electron-hole pair is
generated for each X-ray, being the energy of these pairs easily measurable and corresponding to
the energy of the incoming X-rays. A spectrum is formed by measuring the amount of electron-
hole pairs with each energy.
These factors allow its routine usage in a vast number of fields such as geology and
gemmology, the chemical and pharmaceutical industries, art and antique or the food industry.
The simplicity of the XRF systems has lead to the development of handheld XRF analyzers.
These are the cheapest and more widespread within the field of portable chemical analyzers,
where they rival with Raman and FTIR devices.
To overcome some of these problems, especially the one concerning the limits of
detection, Total-Reflection XRF (TXRF) has been developed recently. This technique introduces
some modifications to classical XRF, with the most important one being that the beam of the X-
ray source is directed at the target at very low angles of incidence (≤0.1°), greatly increasing the
reflectivity of the sample. By using this setup with a thin film of sample on a flat support, it was
possible to detect secondary X-rays emitted from the sample eliminating matrix effects, and
reducing the limit of detection by several orders of magnitude. A smaller amount of material is
needed in this technique as well, as long as it can be deposited on a thin layer onto a carrier.
References and links
Beckhoff, B., Kanngießer, B., Langhoff, N., Wedell, R., Wolff, H., Handbook of
Practical X-Ray Fluorescence Analysis, Springer, 2006
Bertin, E. P., Principles and Practice of X-ray Spectrometric Analysis, Kluwer
Academic / Plenum Publishers
Klockenkämper, Reinhold. Total Reflection X-Ray Fluorescence Analysis. 1997, John
Wiley & Sons, Inc.
Buhrke, V. E., Jenkins, R., Smith, D. K., A Practical Guide for the Preparation of
Specimens for XRF and XRD Analysis, Wiley, 1998
http://www.learnxrf.com/Principles_of_XRF.htm
http://serc.carleton.edu/research_education/geochemsheets/techniques/XRF.html
http://www.horiba.com/scientific/products/x-ray-fluorescence-analysis/tutorial/
https://en.wikipedia.org/wiki/X-ray_fluorescence
http://archaeometry.missouri.edu/xrf_overview.html#applications