D16949_
D16949_
DataStick™ Measurement
System
User Guide
AV38 interchangeable
Local Display communication interchangeable
4-20 mA relays sensor heads
adapters
PLC
Toroidal
PDA MODBUS
Resistivity
PC
USB/RS-232 universal
SCADA Conductivity
sensor module
RS-485
Ozone
DCS
ProfiBus ORP
MODBUS Dissolved
PLC Oxygen
DeviceNet pH
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. U.S. patent 6,793,787.
AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future,
SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ,
AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product
Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum
Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS,
pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK,
SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are
trademarks of Thermo Fisher.
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher.
PerpHecT meters are protected by U.S. patent 6,168,707.
PerpHecT ROSS electrodes are protected by U.S. patent 6,168,707.
ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753.
ionplus electrodes and Optimum Results solutions are protected by U.S. patent 5,830,338.
ROSS Ultra electrodes are protected by U.S. patent 6,793,787.
ORP standard is protected by U.S. patent 6,350,367.
No Cal electrodes are protected by U.S. patent 7,276,142.
© 2010 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice.
This publication supersedes all previous publications on this subject.
Preface
This user guide serves to explain the use of the Thermo Scientific AquaSensors DataStick measurement
system and is written to cover as many applications as possible. Please do not hesitate to contact
Thermo Fisher Scientific or an authorized representative with questions or concerns.
The information presented in this user guide is subject to change without notice as improvements
are made, and does not represent any commitment whatsoever on the part of Thermo Fisher Scientific.
Thermo Fisher Scientific cannot accept any responsibility for damage or malfunction of the sensor due
to improper use.
Contact Information
In Europe, the Middle East and Africa, contact your local authorized dealer.
Before using the sensor, make sure that the sensor cable is connected
as specified. Failure to do so may result in permanent damage to the sensor
or controller.
4 MEASUREMENTS.................................................................................................................................................................................24
4.1 DIFFERENTIAL PH MEASUREMENT ....................................................................................................................................................25
4.1.1 pH Interface Overview ..........................................................................................................................................................26
4.1.2 pH Configuration...................................................................................................................................................................26
4.1.3 pH Calibration.......................................................................................................................................................................30
4.1.4 pH Sensor Diagnosis............................................................................................................................................................38
4.1.5 pH Sensor Maintenance .......................................................................................................................................................39
4.2 DIFFERENTIAL ORP MEASUREMENT .................................................................................................................................................42
4.2.1 ORP Interface Overview.......................................................................................................................................................42
4.2.2 ORP Configuration ...............................................................................................................................................................43
4.2.3 ORP Calibration....................................................................................................................................................................45
4.2.4 ORP Sensor Maintenance....................................................................................................................................................47
4.2.5 ORP Spare Parts..................................................................................................................................................................49
4.3 TWO ELECTRODE CONDUCTIVITY AND RESISTIVITY ............................................................................................................................49
4.3.1 Contacting Conductivity/Resistivity Overview.......................................................................................................................49
4.3.2 Configuration of Contacting Conductivity Sensors ...............................................................................................................50
4.3.3 Contacting Conductivity Calibration......................................................................................................................................62
4.3.4 Contacting Conductivity Maintenance ..................................................................................................................................65
4.4 TOROIDAL CONDUCTIVITY ................................................................................................................................................................66
4.4.1 Toroidal Conductivity Overview ............................................................................................................................................66
4.4.2 Configuration of Toroidal Sensors ........................................................................................................................................67
4.4.3 Toroidal Conductivity Calibration..........................................................................................................................................78
4.4.4 Toroidal Conductivity Maintenance ......................................................................................................................................81
4.5 DISSOLVED OXYGEN (PPM AND PPB) AND OZONE ...........................................................................................................................82
4.5.1 Dissolved Oxygen / Ozone Interface Overview ....................................................................................................................82
4.5.2 Configuration of Dissolved Oxygen and Ozone Sensor Heads............................................................................................83
Table of Figures
Page
Figure Number and Caption
Number
Figure 1: Three Parts of the DataStick 9
Figure 2: The Universal Body
10
Figure 3: Full Array of DataStick Options
10
Figure 4: Configured Analytical Measurements
11
Figure 5: Drinking Water Turbidity System
11
Figure 6: DataStick Interface Options
12
Figure 7: Communications Adapter
12
Figure 8: Mounting Hardware Overview
13
Figure 9: The inner workings of the User-Entered Concentration Table.
61
Figure 10: The inner workings of the User-Entered Concentration Table.
77
Table of Tables
Page
Table Number and Description
Number
Table 1: 10 Point Conductivity Vs. Concentration Table
57
Table 2: Concentration Vs Conductivity after Editing
60
Table 3: 10 Point Conductivity Vs. Concentration Table
73
Table 4: Concentration Vs Conductivity after Editing
76
Table 5: Data Reporting
99
Table 6: Change Measurement Configuration Settings
100
Table 7: View Configuration Settings
101
Table 8: Calibration Commands
102
Table 9: Diagnostics and System Information
105
Thank you for purchasing the Thermo Scientific AquaSensors DataStick Analytical Measurement System. This is a very robust and
versatile product intended for continuous use in industrial applications and complies with safety regulations currently in force. Unique
features include:
• Pre-calibrated sensors for plug and play capability.
• Direct connection to computer control systems without the need for analyzer electronics boxes.
• Sensor heads are easily serviced and replaced in the field minimizing downtime.
• Any sensor type in the DataStick family can be used with any DataStick thus making it possible to convert
a measurement point from pH to toroidal conductivity (for instance), simply by changing sensor heads.
• Performance is not affected by cable length and can be run up to 3,000 feet depending on the selected
communications protocol.
Thermo Fisher Scientific does not accept any liability for damage that may
arise if information in this user guide is not followed. Therefore, the operating
instructions and specifications must be read and understood by all persons
involved in installation and operation of this equipment.
This user guide identifies safety instructions and additional information by means of the following symbols:
It is recommended that this user guide be made accessible to everyone who may need it as a reference.
Please contact Thermo Fisher Scientific or an authorized representative with any questions.
The DataStick Analytical System is designed for continuous industrial applications. Interchangeable measurement and communications
components allow for quick replacement or reconfiguration.
Any other use, or use not mentioned here, that is incompatible with the technical specifications is deemed inappropriate. The operator
is solely responsible for any damage arising from such use.
Other prerequisites for appropriate use include:
• Observing the instructions, notes and requirements set out in this user guide.
• Observing all local safety regulations.
• Observing all warnings and cautions in the documentation regarding all products used in this measurement system,
including the sensor, mounting hardware, and cabling.
• Observing the prescribed environmental and operational conditions.
• Observing chemical compatibility with all wetted materials.
The DataStick Analytical System should be installed and operated only by personnel familiar with
the sensor and qualified for such work.
A defective sensor should be returned to Thermo Fisher Scientific for repair or replacement.
Contact Thermo Fisher Scientific to obtain a Return Material Authorization (RMA) number.
No modifications to the DataStick Analytical System components are allowed. The manufacturer/
supplier accepts no responsibility for damage caused by unauthorized modifications. The risk is borne
entirely by the user.
Input and output wiring must be in accordance with Class I, Div.2 wiring
methods and in accordance with the authority having jurisdiction.
Communications Adapter
Provides power and direct interactive communication
with control systems and computers.
Communications Selections: RS-232 ASCII, RS-232 DF1,
RS-485 MODBUS, PROFIBUS, DEVICENET, ETHERNET,
USB, WIRELESS…
Body Material: 316-Stainless Steel, CPVC, Ryton®, PEEK®
Cable length and termination: 10 feet to 1000 feet
(Consult the factory for other material choices).
Insert communications
adapter this end. Wrench Flats
This family of plug and play communications adapters and sensor heads makes the DataStick the most versatile and accurate analytical
measurement system on the market. Any sensor type can be plugged in to obtain calibrated data. Any computer system can be set up to
communicate directly by configuring the DataStick system with the appropriate communications adapter. If local display and standard 4-20
milliamp data reporting is desired then the DataStick can be connected to Thermo Scientific AquaSensors AV38 Local display as shown in
the upper left corner of Figure 3. The AV38 can work with multiple DataSticks running the Modbus protocol.
Dissolved Oxygen
Differential pH Series ORP Series and Ozone Series
Turbidity
Two Electrode Toroidal 0 to 4000 NTU
Conductivity Series Conductivity Series
DataStick systems can also be configured into more complex systems. Drinking water turbidity measurements are provided using
a DataStick installed in a specially designed sample chamber that has a white light source in the lid. The system provides very accurate
turbidity measurements that can be calibrated and configured locally using the AV38 display or it can be networked directly to a computer
system for measurement, calibration, configuration and diagnostics.
Digital Gateway
PC, PLC, DCS or PDA
Digital Gateway
Modbus
Tee Mount
Use the front threads for
Union Mount insertion into pipe tees.
Submersion
Use the back threads of DataStick for
extension pipes. Integral junction Sanitary
boxes are used for wire termination. For Sanitary applications use
Thermo Scientific
AquaSensors fittings.
Ball Valve
For applications requiring
extraction in a pressurized line
Use a 1-¼ inch wrench to turn the sensor body into pipe fittings that have internal 1-inch NPT threads. Use Teflon tape on all threads
to insure a proper seal. Do not use liquid thread compounds on DataStick threads.
3.2 Mounting
Use a 1-¼ inch wrench to turn the sensor body into pipe fittings that have internal 1-inch NPT threads. Use Teflon tape on all threads
to insure a proper seal. Do not use liquid thread compounds on DataStick threads.
DataStick systems use 1-inch NPT threads for insertion and immersion applications. The threads on the back-body are used for
submersion and Thermo Scientific AquaSensors Union Mount hardware. The front threads are used for insertion into standard plumbing
tee’s. PEEK® sensor heads with sanitary face seals should be used in sanitary applications along with stainless steel sensor bodies that
have special mounting rings between the wrench flats and the NPT threads.
We recommend the Thermo Scientific AquaSensors Union Mount hardware for applications where the sensor is removed more than once
per week for calibration or maintenance.
1. Wiring:
a. Route the cable either through conduit or on a cable tray.
b. Insert the sensor cable into junction box through a watertight cord grip.
c. Connect the sensor wires into the terminal block as specified on the cable label.
2. Mount:
a. Apply Teflon® tape to the sensor body threads.
b. Insert the sensor into the tee and slowly turn clockwise until secure. Tighten the
sensor with a wrench until snug to prevent leaking. Be sure the pipe remains full
when the sensors are installed.
The sensor head is pre-calibrated at the factory and should not require additional calibration
on initial installation unless the DataStick has been in storage for more than two weeks.
1. Wiring:
a. Apply Teflon® tape to the rear sensor body threads and pass the cable through the adapter. Thread the adapter onto the sensor
clockwise until it is secure.
b. Pass the union collar over the cable and onto the adapter with the threads facing the front of the sensor.
c. Route the cable either through conduit or on a cable tray.
d. Insert the sensor cable into the junction box through a watertight cord grip.
e. Connect the sensor wires into the terminal block as specified on the cable label.
2. Mount:
a. Assemble the lower portion of the mounting hardware by threading the pipe nipple into the threaded flange and the tee.
Apply Teflon® tape to the threads.
b. Inspect the O-ring on the union-mounting threaded flange for imperfections or particles of dirt that may prevent the O-ring seal
from seating properly.
c. Carefully insert the sensor into the Thermo Scientific AquaSensors union-mounting tee. Turning the retaining collar clockwise
and hand tighten until snug. Be sure the pipe remains full when the sensor is installed.
2. Mount:
a. Secure the pipe assembly so that the sensor is fully immersed
in the process. Handrail mounting hardware can be ordered from
Thermo Fisher Scientific.
Wiring:
a. Thread the retaining collar over the cable and onto the sensor
with the threads facing the front of the sensor.
b. Route the cable either through conduit or on a cable tray.
c. Insert the sensor cable into the junction box through
a watertight cord grip.
d. Connect the sensor wires into the terminal block as shown
on the cable label.
Mount:
a. Insert the sensor into the adapter making sure the front o-ring
on the sensor is seated.
b. Thread the retaining collar into the adapter and tighten with
a wrench until the face of the sensor head is flush with the
adapter and the o-ring forms a seal.
c. Place a sanitary gasket on the branch flange.
d. Insert the sensor with the adapter into the branch of the tee;
be sure the pipe remains full while installing to minimize trapped air.
e. Install and tighten the sanitary clamp to retain the sensor and
adapter in the tee.
Please refer to Insertion Mounting Hardware User guide for detailed instructions.
Modbus RTU
Color DeviceNet over RS-232 RS-232
or RS-485
USB, Profibus and Ethernet Adapters are provided with standard connectors and discrete wiring for these adapters is not needed.
Host Protocol Reference User guide to Set up Address, Baud Rate and Parity
Modbus RTU over
Modbus Communications Adapter User guide (258481-001)
RS-232 or RS-485
DeviceNet DeviceNet Communications Adapter User guide (258482-001)
RS-232 DataStick User guide (258480-001; this document): 9600 bps, no parity, 8 data bits, 1 stop bit
USB AquaComm USB driver with settings to 9600 bps, no parity, 8 data bits, 1 stop bit (258490-001)
Profibus Modbus/Profibus Gateway User guide (258488-001)
Other Look for addendums on setting up other protocols that may be configured.
DataStick sensors connected to the AV38 communicate via Modbus RTU. As such, the AV38 can
be used to select one of several DataStick sensors on the bus for display, current output reporting
and relay alarms by selecting the desired station address. When there is no DataStick sensor connected
at the selected network address, the measure screen will show dashed lines.
The AV38 can operate in “single sensor mode” with current outputs and relays, or it can be operated in “multi-sensor mode” with
digital communications to a host computer. Host communication options include, Modbus, DeviceNet, RS-485 ASCII, Profibus
and others. The host protocol is independent of the Modbus RTU protocol used by the AV38 to communicate
with the DataSticks.
In the multi-sensor mode the AV38 can be used as a switch that allows a host computer to measure calibrate, configure and diagnose
all DataSticks connected to the AV38. In this mode the AV38 measure screen displays the measure data that the host computer
is currently asking for. If the menu key on the AV38 is pressed, control of the local network of DataSticks is transferred to the AV38.
The host computer regains control after the user navigates back to the measure screen.
There are seven keys for menu navigation. The MENU key is used to toggle between the menu and the measure screen. Pressing the
Menu key provides options for calibration, configuration, communications, outputs and relays. Refer to the AV38 User guide for Details.
The first time the program is launched it will most likely come up with blanks in the screen. Select FILE and the NEW CONNECTION
to connect the program to the DataStick. If there is more than one DataStick connected, set the first one up and then open up another
instance of the program to establish a connection with the next sensor. Once these applications close, the connection configuration
will be saved.
To configure AquaComm™ for use after installing the program, select the FILE menu:
Select NEW CONNECTION or OPEN CONNECTION to set up the port and protocol. When necessary, the station number, baud rate
and parity will also be set. Thermo Scientific AquaSensors communcations adapters are preset for baud rate, parity and station number.
These variables can be changed to match network requirements as described in the programming section of this user guide.
Once the connection has been established, the display fields will be populated with current measurement data from the DataStick.
Pull-down windows will provide calibration, configuration and diagnostic choices specific to the sensor head that is connected to the
DataStick. Data logging options are available under FILE. Calibration, configuration and diagnostic choices are not available when
logging is active.
Open protocol ASCII commands are available with all communications adapters. The ASCII command set for the DataStick is available
for any application and the programming guide is detailed SECTION 7, “Commands and Responses”.
ASCII commands can be typed into any terminal program and with the correct communications setting; data can be obtained very quickly.
Communications setups vary depending on the protocol chosen. Some examples:
• RS-232 ASCII: 9600 Baud, 8 data bits, no parity, 1 stop bit.
• RS-485 MODBUS: Station address, baud rate, and parity.
These ASCII commands can be used directly with network protocols like Modbus or DeviceNet but it is much more effective to use the
standard maps for these protocols.
Pre-calibrated sensor heads include differential pH and ORP, two electrode conductivity, two electrode resistivity, toroidal conductivity,
dissolved oxygen and ozone. Additional sensor types include turbidity, chlorine and others. Just plug them in and tighten the retaining ring.
Sensor heads are available in different materials and with different electrode profiles.
Once a DataStick has been configured for the desired units of measure, data filters and calibration standards, sensor head calibrations
can be performed at any time.
Measurement accuracy is maintained when transferring the sensor head from one DataStick to another. This makes it possible to set up
a DataStick system in the laboratory for calibration and test purposes. Sensor heads can then be carried to the field site and exchanged
with heads that need recalibration or refurbishment. This process reduces down time and actually extends the life of the sensor. Normal
field calibrations can be done through the AquaComm™ interface in local interface boxes.
Encapsulated Quad-Junction
Product Number/ Allows ion exchange
Serial Number to reference chamber.
Removable
Salt Bridge
(Other bridge profiles, materials
and salt formulas available).
Select BUFFER STANDARD from the Configuration pull-down menu to view or change the buffer standard that will be used
as a reference for buffer calibrations.
There are two pre-defined standards currently used with the DataStick system. The Thermo Scientific AquaSensors buffer
standard is the default when the system is reset.
Thermo Scientific AquaSensors Buffer Standards: 4 pH, 7 pH and 10 pH. (DEFAULT).
DIN 19267 Buffer Standards: 1.09 pH, 4.65 pH, 6.79 pH, 9.23 pH and 12.75 pH.
When using a buffer standard that is not included in the pre-defined buffer standard, use the
‘2-point sample’ calibration method.
Select SENSOR FILTER from the Configuration pull-down menu to view or change the sensor filter constant.
The filter setting can be changed by selecting the edit box and typing in the desired constant. To accept the change, click OK.
To abort the change, click CANCEL.
If the sensor filter is set to 5 seconds, it will take 5 seconds to reach 90% of the final value when the reading changes.
Measurements are averaged to reduce periodic variations.
The default filter setting is 1 second and the maximum filter is 100 seconds. Use zero seconds for a completely unfiltered reading.
Select TEMPERATURE FILTER from the Configuration pull-down menu to view or change the temperature filter constant.
The filter setting can be changed by selecting the edit box and typing in the desired constant. To accept the change, click OK.
To abort the change, click CANCEL.
If the temperature filter is set to 5 seconds, it will take 5 seconds to reach 90% of the final value when the reading changes.
Measurements are averaged to reduce periodic variations.
The default filter setting is 1 second and the maximum filter is 100 seconds. Use zero seconds for a completely unfiltered reading.
Although temperature and sensor filter constants are independently controlled the best results are
generally achieved when both filters are set to the same constant.
Select SENSOR UNITS from the Configuration pull-down menu to view or change units of measure.
Units shown in the sensor units window are active. To change units or to review options, scroll the selection list. Click on the
desired units of measure to place the new choice in the sensor units window. To accept the change, click OK. To abort the
change, click CANCEL.
For pH sensors, the units of measure can be set to pH or millivolts. The default units are pH.
Select TEMPERATURE UNITS from the Configuration pull-down menu to view or change units of measure.
Units shown in the temperature units window are active. To change units or to review options, scroll the selection list. Click on the
desired units of measure to place the new choice in the sensor units window. To accept the change, click OK. To abort the change,
click CANCEL.
Temperature can be displayed in °C or °F. Default units are °C.
Select 1-POINT BUFFER from the calibration menu to begin a calibration in a known buffer standard.
In a buffer calibration, the DataStick assumes that the sensor has been placed in a pre-defined buffer standard and automatically
equates the current reading to the closest standard value. For a 1-point pH buffer calibration, instructions are shown in the status
box on the left side of the calibration screen. The sensor and temperature values are shown on the right side of the screen.
Place the pH DataStick sensor into the buffer solution. Wait for the sensor and temperature readings to stabilize. Press the
CALIBRATE button and the sensor will begin the process of changing the current pH reading to the closest buffer standard
reading as defined by the buffer standard selected.
Once the calibration has been initiated it will take between 30 seconds and 2 minutes depending on sensor and temperature
measurement stability. During this period, the abort sequence can be initiated as shown with the active Abort button.
When the point has been successfully calibrated, measured values will be based on the new calibration. The rate of
change in reading from the old measured value to the new will be dependent on the sensor filter value that is entered
in the configuration menu.
Exiting the calibration menu will return the user to the measure screen.
The status of the calibration can be reviewed with the CALSTATUS command. When the last number returned is a 1,
the calibration is in process. When the last number returned is a 2, the calibration has been successful. Any other number
indicates a specific failure mode (see appendix).
Select 2-POINT BUFFER from the calibration menu to begin a pH calibration sequence.
A 2-point buffer calibration is the recommended calibration method for most accurate results and requires 2 buffer standards
at least 3 pH units apart. Every pH sensor is calibrated this way when it leaves the factory and will maintain accuracy for a period
of time when the sensor head is kept wet.
Either Thermo Scientific AquaSensors pH standards or DIN pH standards can be selected from the configuration menu for this
calibration method. This process allows calibration to be done without entering any numbers but requires that the buffer standards
used for the calibration match the standard family that is selected in the configuration menu. The default standards are 4 pH, 7 pH
and 10 pH.
After selecting the 2-point pH buffer calibration method, the sensor should be placed in the first standard and when sensor
and temperature readings are stable, click the CALIBRATE button to initiate the calibration or click EXIT to return to the main
measure screen.
Once the calibration has been initiated it will take between 15 seconds and 2 minutes depending on sensor and temperature
measurement stability. During this period, the abort sequence can be initiated as shown with the active Abort button.
Once the 2-point buffer calibration has successfully completed storage of the first point, the status box will prompt the user to place
the sensor in the second solution before initiating calibration of the second point. Please note that the reading will not be adjusted
after the first calibration point. Calibration changes will only take effect after both calibration points have been accepted.
While the calibration is in progress the operation can be aborted at any time. The abort will take the system back to normal
measurements using the previous calibration values.
Once the calibration for both points is successful, the displayed sensor value will be based on the new calibration.
Exiting the calibration function will return the display to the measure screen. In this case the measured value of 4.01 pH confirms
that the sensor calibration in 4.01 pH buffer solution was successful.
The status of the calibration can be reviewed with the CALSTATUS command. When the last number returned is a 1,
the calibration is in process. When the last number returned is a 2, the calibration has been successful. Any other number
indicates a specific failure mode (see appendix).
Select 2-POINT SAMPLE from the calibration menu to begin the calibration sequence.
If buffer standards are not available or if it is desired to calibrate to process values, the 2-point sample calibration method may
be appropriate. This calibration method requires the input of desired values for a given measurement. The points must be 3 pH
units apart and the readings must be stable before calibration can be successful.
When the 2-point sample method is chosen, the status box will prompt the user to enter the desired calibration value. Stable
temperature and sensor readings will help to establish an accurate calibration.
Once the desired value for calibration is entered, the CALIBRATE button becomes active. Once the displayed sensor and
temperature values are stable and near expected values, click the CALIBRATE button to initiate the calibration of the first sample
point or click EXIT to return to the main measure screen.
When the first calibration point is accepted, the status box will prompt the user to place the sensor in the second solution
of known value.
Wait until the sensor and temperature values are stable in the second solution.
Enter the desired value and initiate calibration by clicking the CALIBRATE button.
When the second point has been successfully calibrated, measured values will be based on the new calibration. The rate of
change in reading from the old measured value to the new will be dependent on the sensor filter value that is entered in the
configuration menu.
Exiting the calibration menu will return the display to the measure screen.
The status of the calibration can be reviewed with the CALSTATUS command. When the last number returned is a 1,
the calibration is in process. When the last number returned is a 2, the calibration has been successful. Any other number
indicates a specific failure mode (see appendix).
Stable temperature and sensor readings will also help to establish an accurate calibration.
Once the desired value for calibration is entered, the CALIBRATE button becomes active. Once the displayed sensor and
temperature values are stable and near expected values, click the CALIBRATE button to initiate the calibration or click EXIT
to return to the main measure screen.
Once the calibration has been initiated it will take somewhere between 30 seconds and 2 minutes depending on sensor
and temperature measurement stability. During this period, the abort sequence can be initiated as shown with the active
ABORT button.
Exiting the calibration menu will return the display to the measure screen.
The status of the calibration can be reviewed with the CALSTATUS command. When the last number returned is a 1,
the calibration is in process. When the last number returned is a 2, the calibration has been successful. Any other number
indicates a specific failure mode (see appendix).
2. Soak the sensor in a container of hot detergent water for 30 minutes. Do not use detergents that contain oily skin softeners
like aloe or lanolin that can coat the glass electrode. Powdered Alconox™ and Dawn™ dishwashing liquid work well.
3. Use a soft-bristled brush, such as a soft toothbrush, and hot detergent water to scrub the entire electrode end of the sensor,
being careful not to scratch or break the glass electrode.
5. If deposits are still present on glass electrode repeat steps 2 and 3. In the case of lime or other mineral deposits a weak
solution (about 0.1 M) of hydrochloric acid may be used. In some cases, a dilute solution (about 10:1) of water and chlorine
bleach or a solution of water and EDTA may also work. Stubborn oil or grease deposits may require cleaning with a solvent
such as acetone or alcohol. Assure that the sensor body in compatible with the solvent. Protein deposits may be cleaned with
a pepsin-based cleaning solution. Bacterial or mold growths may be removed with dilute chlorine bleach.
Note: Cleaning can be process-specific. Some users will have to develop a procedure that works
best for them.
Caution: Ensure that a dangerous chemical reaction will not occur between process deposits and
the cleaning solution, such as compounds of cyanide, and the hydrochloric acid. Cyanide is often
used in electroplating and in gold refining. If in doubt check with a chemist.
Caution: When using dangerous chemicals such as hydrochloric acid or chlorine bleach, use all
appropriate safety precautions, especially eye protection.
Before returning the sensor to service, allow it to soak in water or buffer at ambient temperature for about an hour to stabilize.
After cleaning the sensor, always calibrate the sensor head before installing in the process.
If the sensor head has been cleaned and calibration cannot be achieved, replace the salt bridge and reference
solution.
Hold the sensor firmly with the electrode tip facing upwards. Remove the existing salt bridge by using
a 15/16 wrench (24mm) and turning it counterclockwise. Dispose of the salt bridge using an approved method.
Pour out the old reference filling solution. Rinse the reservoir with distilled or de-ionized water.
Slowly fill the reservoir with Thermo Scientific AquaSensors Reference Filling Solution (P/N RCS02)
so the solution just covers the reference O-ring.
Slowly screw the new salt bridge clockwise onto the sensor head until secure. Tighten the salt bridge with a 15/16
wrench (24mm) until snug.
After replacing the salt bridge and reference solution, calibrate the sensor.
If after cleaning and refurbishment sensors still have slow responses, low measurement slopes or are
seriously abraded or scratched, they must be replaced. Contact your local distributor or call Thermo Fisher
Scientific Customer Service.
There are several salt bridge replacement options. Protected and unprotected electrodes (with face seals for sanitary
applications). CPVC and PEEK® material. Custom salt bridge chemistry can be provided on request. The salt bridge
options can also be used on ORP sensor heads.
O-Ring
Water tight seal Retaining Ring
Connector Key Locks the sensor head
into the DataStick body
Gold Pins
Platinum electrode inside
concentric stainless steel
ground rod
Encapsulated
Product Number/
Serial Number
Quad-Junction
Allows ion exchange
to reference chamber
Removable
Salt Bridge
(Other bridge profiles, materials
and salt formulas available).
Select SENSOR FILTER from the Configuration pull-down menu to view or change the sensor filter constant.
The filter setting can be changed by selecting the edit box and typing in the desired constant. To accept the change, click OK.
To abort the change, click CANCEL.
If the sensor filter is set to 5 seconds, it will take 5 seconds to reach 90% of the final value when the reading changes.
Measurements are averaged to reduce periodic variations.
The default filter setting is 1 second and the maximum filter is 100 seconds. Use zero seconds for a completely
unfiltered reading.
Select TEMPERATURE FILTER from the Configuration pull-down menu to view or change the temperature filter constant.
The filter setting can be changed by selecting the edit box and typing in the desired constant. To accept the change, click OK.
To abort the change, click CANCEL.
If the temperature filter is set to 5 seconds, it will take 5 seconds to reach 90% of the final value when the reading changes.
Measurements are averaged to reduce periodic variations.
The default filter setting is 1 second and the maximum filter is 100 seconds. Use zero seconds for a completely unfiltered reading.
ORP and temperature measurements are completely independent of one another and filter constants
can also be independent.
The SENSOR UNITS selection is not available for ORP measurements. Units of measure are in millivolts (mV).
Select TEMPERATURE UNITS from the Configuration pull-down menu to view or change units of measure.
Sensors are calibrated at the factory. When field calibrations are necessary a 1-POINT SAMPLE calibration can be done when the sensor
is in process liquid. The ORP measurement is not compensated by temperature. Temperature is offered as an independent measurement.
Select 1-POINT SAMPLE from the calibration menu to begin an ORP calibration.
When the 1-point sample method is chosen, the status box will prompt the user to enter the desired calibration value. Use
calibration solutions with millivolt potentials that are similar to those used in the process measurement. Solutions of opposite
polarity may react on the sensor electrode and cause it to foul.
Once the desired value for calibration is entered, the CALIBRATE button becomes active. Once the displayed sensor value
is stable and near the expected value, click the CALIBRATE button to initiate the calibration or click EXIT to return to the main
measure screen.
Once the calibration has been initiated it will take somewhere between 30 seconds and 2 minutes depending on sensor
measurement stability. During this period, the abort sequence can be initiated as shown with the active ABORT button.
Exiting the calibration menu will return the user to the measure screen.
Before returning the sensor to service, allow it to soak in water or buffer at ambient temperature for about an hour to stabilize.
After cleaning the sensor, always calibrate the sensor head before installing in the process.
If the sensor head has been cleaned and calibration cannot be achieved, replace the salt bridge and
reference solution.
Hold the sensor firmly with the electrode tip facing upwards. Remove the existing salt bridge by using
a 15/16 wrench (24mm) and turning it counterclockwise. Dispose of the salt bridge using an approved method.
Pour out the old reference filling solution. Rinse the reservoir with distilled or de-ionized water.
Slowly fill the reservoir with Thermo Scientific AquaSensors Reference Filling Solution (P/N RCS02) so the solution
just covers the reference O-ring.
Slowly screw the new salt bridge clockwise onto the sensor head until secure. Tighten the salt bridge with
a 15/16 wrench (24mm) until snug.
After replacing the salt bridge and reference solution, calibrate the sensor.
If after cleaning and refurbishment sensors still have slow responses, low measurement slopes or are
seriously abraded or scratched, they must be replaced. Contact your local distributor or call Thermo Fisher
Scientific Customer Service.
There are several salt bridge replacement options. Protected and unprotected electrodes (with face seals for sanitary
applications). CPVC and PEEK® material. Custom salt bridge chemistry can be provided on request. The salt bridge
options can also be used with pH sensor heads.
Encapsulated
Product Number/
Serial Number Resistivity (Cell K = 0.1) Conductivity (Cell K = 1.0)
The AquaComm™ CONFIGURE functions can be accessed from a pull-down menu that automatically changes based on the sensor type
connected to the DataStick. When a conductivity sensor is connected, settings for data filters, units of measure, reference temperature,
temperature compensation and concentration tables can be viewed or changed. It is also possible to manually adjust the cell constant of
the connected sensor if a user guide adjustment of the measure value is desired. It is best to view the adjusted cell constant after
performing a 1-point sample calibration.
Select SENSOR FILTER from the Configuration pull-down menu to view or change the sensor filter constant.
The filter setting can be changed by selecting the edit box and typing in the desired constant. To accept the change, click OK.
To abort the change, click CANCEL.
If the sensor filter is set to 5 seconds, it will take 5 seconds to reach 90% of the final value when the reading changes.
Measurements are averaged to reduce periodic variations.
The default filter setting is 1 second and the maximum filter is 100 seconds. Use zero seconds for a completely
unfiltered reading.
Select TEMPERATURE FILTER from the Configuration pull-down menu to view or change the temperature filter constant.
The filter setting can be changed by selecting the edit box and typing in the desired constant. To accept the change, click OK.
To abort the change, click CANCEL.
If the temperature filter is set to 5 seconds, it will take 5 seconds to reach 90% of the final value when the reading changes.
Measurements are averaged to reduce periodic variations.
The default filter setting is 1 second and the maximum filter is 100 seconds. Use zero seconds for a completely
unfiltered reading.
Although temperature and sensor filter constants are independently controlled the best results
are generally achieved when both filters are set to the same constant.
Select SENSOR UNITS from the Configuration pull-down menu to view or change units of measure.
Units shown in the sensor units window are active. To change units or to review options, scroll the selection list. Click on the
desired units of measure to place the new choice in the sensor units window. To accept the change, click OK. To abort the
change, click CANCEL.
For contacting conductivity sensors, conductivity measurements can be set to conductivity in µS/cm, concentration in %,
total dissolved solids in ppm and resistivity in MΩ•cm.
Default units of measure are µS/cm.
Concentration measurements are converted from the data points entered in a 10-point concentration table that can be viewed
and edited from the Configuration pull-down menu.
Select TEMPERATURE UNITS from the Configuration pull-down menu to view or change units of measure.
Units shown in the temperature unit’s window are active. To change units or to review options, scroll the selection list. Click on
the desired units of measure to place the new choice in the sensor units window. To accept the change, click OK. To abort the
change, click CANCEL.
Temperature can be displayed in °C or °F. Default units are °C.
Select CELL CONSTANT from the Configuration pull-down menu to view or change the conductivity reference temperature.
The setting can be changed by selecting the edit box and typing the desired number. To accept the change, click OK. To abort
the change, click CANCEL.
The value shown in the edit box is used as a calibration value. Changing it will change the measured value. If the sensor type
is a nominal 0.1 Cell for RESISTIVITY measurements, values can be changed from 0.05 to 0.15. If the sensor type is a nominal
1.0 Cell for low range conductivity measurements, values can be changed from 0.5 to 1.5. It is important to clean the sensor
electrodes in detergent that does not contain Lanolin or Aloe prior to calibration.
The commands in this section are available in DataStick firmware D3.32 or later. The DataStick firmware version can be obtained using
the GCVSN command.
ASCII Command to View Conductivity Temperature Compensation Type
Function Command Response Explanation
The type of temperature compensation used
0
is Linear (default)
View conductivity
temperature GCTCTYPE The type of temperature compensation used
1
compensation type is Natural Water
The type of temperature compensation used
2
is None
Select REFERENCE TEMPERATURE from the Configuration pull-down menu to view or change the conductivity
reference temperature.
The setting can be changed by selecting the edit box and typing the desired number. To accept the change, click OK.
To abort the change, click CANCEL.
The default reference temperature for conductivity is 25°C. The system will accept any number between 0 and 100
with up to one decimal place.
Select COMPENSATION SLOPE from the configuration menu to view or change the compensation slope for conductivity
measurements.
The default compensation slope is 2.00%/°C. The system will accept any number from 0 to 9.9 with up to two decimal places.
The setting can be changed by selecting the edit box and typing the desired number. To accept the change, click OK.
To abort the change, click CANCEL.
The reference can be changed to any temperature between 0°C to 100°C with one-tenth degree resolution. Edit the reference
temperature by typing a new value in the edit box and clicking OK.
Select CONCENTRATION TABLE from the configuration menu to view or change any or all of the points in the table.
When selecting the ten-point concentration table, there will be an automatic update of each of the 10 points before any editing
or scrolling of the table can occur. This scan takes about 10 seconds.
The concentration table coordinates must be monotonic. Conductivity must increase with each point. Concentration can either
increase or decrease with each point. The table must start at POINT 1 and continue incrementally from there. Unused points
(except POINT 1) have zero, zero coordinates.
Current entries can be viewed by scrolling up and down on the “Number” window.
Table entries for specific chemicals are available from Thermo Fisher Scientific Customer Service on request.
Conductivity Concentration
Point
(µS/cm) %
1 0 0
2 20,000 99.99
3 0 0
4 0 0
5 0 0
6 0 0
7 0 0
8 0 0
9 0 0
10 0 0
The following example shows how to change and add data points to the table. In this case, POINT 2 is changed and
POINTS 3 and 4 are added.
To change point 2 from 20,000 μS/cm at 99.99%, to 20,000µS/cm at 50%, scroll to point 2 and place the cursor in the
concentration coordinate window.
Scroll to point 3 and replace the zero coordinates with 50,000µS/cm at 70% concentration. Then click on the STORE button
to save point 3 to temporary memory.
Data for points 5 through 10 should be zero. Data stored in temporary memory can now be permanently stored using
the ACTIVATE STORED TABLE button.
If any error messages are shown, review the data points to be sure they are monotonic, that conductivity increases from
point 1 and that concentration increases or decreases from point 1.
Once the table has been activated, use the EXIT button and return to the main measure menu.
Table entries for specific chemicals are available from Thermo Fisher Scientific Customer Service on request.
The table entries made in this example are shown in Table 2.
Conductivity Concentration
Point
(µS/cm) %
1 0 0
2 20,000 50
3 50,000 70
4 100,000 97
5 0 0
6 0 0
7 0 0
8 0 0
9 0 0
10 0 0
When the units of measure are set to “%” the conversion from conductivity to concentration will make use of the entered table
and will present the readings as shown.
SUPNT GUPNT
GSPNT
User
For the scratch table to be considered valid, the following must be true:
1. All successive conductivity point values must be monotonic and increasing and be within the bounds of 0 to 2,000,000.0
µS/cm.
2. The concentration values must be monotonic and either increasing or decreasing and be within the bounds of 0 to 99.99%.
3. If fewer than ten points are entered, then the scratch table must be terminated with a last point containing zero values.
Provided the above criteria are met, the points in the scratch table are transferred to the working table when the SUTBL command is used;
otherwise the current working table is retained.
Select ZERO from the calibration menu to begin the contacting conductivity zero calibration.
A zero calibration is applied to a sensor placed in dry air. The calibration establishes the zero for the measurement system
and will require more than 1 minute to complete. Please be patient.
Suspend the sensor in dry air, away from surfaces. Be sure that the sensor body is dry by wiping it clean with a dry rag.
For best results, wait for the sensor reading to reach a steady state.
Initiate the zero calibration by clicking the CALIBRATE button.
When the point has been successfully calibrated, measured values will be based on the new zero. The rate of change in reading
from the old measured value to the new will be dependent on the sensor filter value that is entered in the configuration menu.
Exiting the calibration menu will return the user to the measure screen.
Select 1-POINT SAMPLE from the calibration menu to begin a contacting conductivity span calibration.
When the 1-point sample method is chosen, the status box will prompt the user to enter the desired calibration value.
This calibration should be initiated when the sensor is in a liquid that is at the high end of the expected measuring range.
Stable temperature and sensor readings will also help to establish an accurate calibration.
Once the desired value for calibration is entered, the CALIBRATE button becomes active. Once the displayed sensor and
temperature values are stable and near expected values, click the CALIBRATE button to initiate the calibration or click EXIT
to return to the main measure screen.
Once the calibration has been initiated it will take between 30 seconds and 2 minutes depending on sensor and temperature
measurement stability. During this period, the abort sequence can be initiated as shown with the active ABORT button.
When the point has been successfully calibrated, measured values will be based on the new calibration. The rate of change
in reading from the old measured value to the new will be dependent on the sensor filter value that is entered in the configuration
menu.
Exiting the calibration menu will return the user to the measure screen.
Many types of deposits can be removed by wiping the electrodes with a cloth, brush, or paper towel and then rinsing with clean water.
Additional deposits may be removed by soaking in water and detergent. In general, never use a cleaner that contains a skin softener
such as aloe or lanolin. These can coat the electrode. Powdered Alconox™ and Dawn™ are known to work well.
Stubborn mineral deposits may be removed by soaking in dilute hydrochloric acid (about 0.1 M to 1.0 M). HCl will in time, attack both
316 stainless steel and titanium, so the cell should be soaked no longer than required. Rinse the cell to remove the acid.
Oily deposits may be removed with a solvent such as alcohol or acetone. Assure that the sensor body in compatible with the solvent.
A mild alkaline cleaner such as Alconox Solujet™ may also work.
Rinse or soak the cell in clean water before returning to service.
Caution: Ensure that a dangerous chemical reaction will not occur between process deposits and the
cleaning solution, such as compounds of cyanide, and the hydrochloric acid. Cyanide is often used in
electroplating and in gold refining. If in doubt check with a chemist.
Caution: When using dangerous chemicals such as hydrochloric acid or chlorine bleach, use all
appropriate safety precautions, especially eye protection.
Encapsulated
Product Number/ O-Ring
Serial Number Water tight seal
Select SENSOR FILTER from the Configuration pull-down menu to view or change the sensor filter constant.
If the sensor filter is set to 5 seconds, it will take 5 seconds to reach 90% of the final value when the reading changes.
Measurements are averaged to reduce periodic variations.
The default filter setting is 1 second and the maximum filter is 100 seconds. Use zero seconds for a completely
unfiltered reading.
4.4.2.2 TEMPERATURE FILTER
Select TEMPERATURE FILTER from the Configuration pull-down menu to view or change the temperature filter constant.
The filter setting can be changed by selecting the edit box and typing in the desired constant. To accept the change, click OK.
To abort the change, click CANCEL.
If the temperature filter is set to 5 seconds, it will take 5 seconds to reach 90% of the final value when the reading changes.
Measurements are averaged to reduce periodic variations.
The default filter setting is 1 second and the maximum filter is 100 seconds. Use zero seconds for a completely unfiltered
reading.
Although temperature and sensor filter constants are independently controlled the best results are
generally achieved when both filters are set to the same constant.
The toroidal sensor head has significant mass and although the temperature element is located as close to the process as
possible, temperature readings can take up to 40 minutes to reach steady state when exposed to extreme temperature changes.
This is important to note when calibrating the sensor.
Select SENSOR UNITS from the Configuration pull-down menu to view or change units of measure.
Units shown in the sensor units window are active. To change units or to review options, scroll the selection list. Click on
the desired units of measure to place the new choice in the sensor units window. To accept the change, click OK. To abort
the change, click CANCEL.
For toroidal conductivity sensors, conductivity measurements can be set to conductivity in μS/cm, concentration in %, total
dissolved solids in ppm and resistivity in MΩ·cm.
Default units of measure are µS/cm.
Concentration measurements are converted from the data points entered in a 10-point concentration table that can be viewed
and edited from the Configuration pull-down menu.
Select TEMPERATURE UNITS from the Configuration pull-down menu to view or change units of measure.
Units shown in the temperature units window are active. To change units or to review options, scroll the selection list. Click on
the desired units of measure to place the new choice in the sensor units window. To accept the change, click OK. To abort the
change, click CANCEL.
Temperature can be displayed in °C or °F. Default units are °C.
The commands in this section are available in DataStick firmware D3.32 or later. The DataStick firmware version can be obtained using
the GCVSN command.
ASCII Command to View Conductivity Temperature Compensation Type
Function Command Response Explanation
The type of temperature compensation used
0
is Linear (default)
View conductivity
temperature GCTCTYPE The type of temperature compensation used
1
compensation type is Natural Water
The type of temperature compensation used
2
is None
Select REFERENCE TEMPERATURE from the Configuration pull-down menu to view or change the conductivity
reference temperature.
The setting can be changed by selecting the edit box and typing the desired number. To accept the change, click OK.
To abort the change, click CANCEL.
The default reference temperature for conductivity is 25°C. The system will accept any number between 0 and 100
with up to one decimal place.
Edit the reference temperature by typing a new value in the edit box and clicking OK.
Select COMPENSATION SLOPE from the configuration menu to view or change the compensation slope for conductivity
measurements.
The default compensation slope is 2.00%/°C. The system will accept any number from 0 to 9.9 with up to two decimal places.
The setting can be changed by selecting the edit box and typing the desired number. To accept the change, click OK. To abort
the change, click CANCEL.
The reference can be changed to any temperature between 0°C to 100°C with one-tenth degree resolution. Edit the reference
temperature by typing a new value in the edit box and clicking OK.
Select CONCENTRATION TABLE from the configuration menu to view or change any or all of the points in the table.
When selecting the ten-point concentration table, there will be an automatic update of each of the 10 points before any editing
or scrolling of the table can occur. This scan takes about 10 seconds.
The concentration table coordinates must be monotonic. Conductivity must increase with each point. Concentration can either
increase or decrease with each point. The table must start at POINT 1 and continue incrementally from there. Unused points
(except POINT 1) have zero, zero coordinates.
Current entries can be viewed by scrolling up and down on the “Number” window.
Table entries for specific chemicals are available from Thermo Fisher Scientific Customer Service on request.
The default concentration table is set up as shown in Table 1. The numbers can be viewed by scrolling the up/down arrows.
The following example shows how to change and add data points to the table. In this case, POINT 2 is changed
and POINTS 3 and 4 are added.
Delete the 99.9 in the concentration coordinate and replace it with 50. Note that the STORE and REVERT buttons
are now active. If the point is what is intended, click the STORE button to save the point to temporary memory.
Clicking REVERT will restore the point to what it was before editing was initiated.
Scroll to point 4 and replace the zero coordinates with 100,000 μS/cm at 97% concentration. Then click on the STORE button
to save point 4 to temporary memory.
If any error messages are shown, review the data points to be sure they are monotonic, that conductivity increases
from point 1 and that concentration increases or decreases from point 1.
Once the table has been activated, use the EXIT button and return to the main measure menu.
Table entries for specific chemicals are available from Thermo Fisher Scientific Customer Service on request.
The table entries made in this example are shown in Table 2.
Table 4: Concentration Vs Conductivity after Editing
Point Conductivity (µS/cm) Concentration %
1 0 0
2 20,000 50
3 50,000 70
4 100,000 97
5 0 0
6 0 0
7 0 0
8 0 0
9 0 0
10 0 0
When the units of measure are set to “%” the conversion from conductivity to concentration will make use of the entered
table and will present the readings as shown.
SUPNT GUPNT
GSPNT
User
For the scratch table to be considered valid, the following must be true:
1. All successive conductivity point values must be monotonic and increasing and be within the bounds of 0 to 2,000,000.0
µS/cm.
2. The concentration values must be monotonic and either increasing or decreasing and be within the bounds of 0 to 99.99%.
3. If fewer than ten points are entered, then the scratch table must be terminated with a last point containing zero values.
Provided the above criteria are met, the points in the scratch table are transferred to the working table when the SUTBL command is used;
otherwise the current working table is retained.
Select ZERO from the calibration menu to begin the toroidal conductivity zero calibration.
A zero calibration is applied to a sensor placed in dry air. The calibration establishes the zero for the measurement system
and will require more than 1 minute to complete. Please be patient.
Suspend the sensor in dry air, away from surfaces. Be sure that the sensor body is dry by wiping it clean with a dry rag.
For best results, wait for the sensor reading to reach a steady state. If the sensor is being calibrated for the first time,
the reading may be as high as 100 μS/cm. In most situations however, the sensor will already have a zero calibration
stored in it, which may require some periodic adjustment.
Once the calibration has been initiated it will take approximately 1 to 2 minutes. During this period, the abort sequence
can be initiated as shown with the active Abort button.
When the point has been successfully calibrated, measured values will be based on the new zero. The rate of change in reading
from the old measured value to the new will be dependent on the sensor filter value that is entered in the configuration menu.
Exiting the calibration menu will return the user to the measure screen.
Select 1-POINT SAMPLE from the calibration menu to begin a toroidal conductivity span calibration.
When the 1-point sample method is chosen, the status box will prompt the user to enter the desired calibration value.
This calibration should be initiated when the sensor is in a liquid that is at the high end of the expected measuring range.
Stable temperature and sensor readings will also help to establish an accurate calibration.
Once the desired value for calibration is entered, the CALIBRATE button becomes active. Once the displayed sensor
and temperature values are stable and near expected values, click the CALIBRATE button to initiate the calibration
or click EXIT to return to the main measure screen.
When the point has been successfully calibrated, measured values will be based on the new calibration. The rate
of change in reading from the old measured value to the new will be dependent on the sensor filter value that is entered
in the configuration menu.
Exiting the calibration menu will return the user to the measure screen.
Encapsulated
Product Number/
Serial Number
Retaining Ring
Locks the sensor head
into the DataStick body
Replaceable Membrane
Cap
Factory default configurations for a dissolved oxygen or ozone sensor head are:
• Units of measure: ppm ) (options include % and ppb)
• Units for temperature: °C (options include °F)
• Sensor Filter: 1 Second (options include 0 to 100 seconds in 1 second increments)
• Temperature Filter: 1 Second (options include 0 to 100 seconds in 1 second increments)
• Salinity: 0 mS/cm (options from 0 to 999.9 mS/cm)
• Atmospheric Pressure: 760 mmHg (options from 539.2 to 792.4 mmHg)
Select SENSOR FILTER from the Configuration pull-down menu to view or change the sensor filter constant.
The filter setting can be changed by selecting the edit box and typing in the desired constant. To accept the change, click OK.
To abort the change, click CANCEL.
If the sensor filter is set to 5 seconds, it will take 5 seconds to reach 90% of the final value when the reading changes.
Measurements are averaged to reduce periodic variations.
The default filter setting is 1 second and the maximum filter is 100 seconds. Use zero seconds for a completely
unfiltered reading.
Select TEMPERATURE FILTER from the Configuration pull-down menu to view or change the temperature filter constant.
If the temperature filter is set to 5 seconds, it will take 5 seconds to reach 90% of the final value when the reading changes.
Measurements are averaged to reduce periodic variations.
The default filter setting is 1 second and the maximum filter is 100 seconds. Use zero seconds for a completely unfiltered
reading.
Although temperature and sensor filter constants are independently controlled the best results
are generally achieved when both filters are set to the same constant.
Select SENSOR UNITS from the Configuration pull-down menu to view or change units of measure.
Units shown in the sensor units window are active. To change units or to review options, scroll the selection list.
Click on the desired units of measure to place the new choice in the sensor units window. To accept the change, click OK.
To abort the change, click CANCEL.
For dissolved oxygen sensors, units of measure can be set to ppm or %. For special DO sensors, ppb units can also be selected.
Consult the factory for details.
Select TEMPERATURE UNITS from the Configuration pull-down menu to view or change units of measure.
Units shown in the temperature units window are active. To change units or to review options, scroll the selection list.
Click on the desired units of measure to place the new choice in the sensor units window. To accept the change, click OK.
To abort the change, click CANCEL.
Temperature can be displayed in °C or °F. Default units are °C.
Select PRESSURE from the Configuration pull-down menu to view or change the setting of atmospheric pressure used
during an air calibration.
Edit the pressure setting by typing a new value in the edit box. To accept the change, click OK. To abort the change,
click CANCEL.
The system will accept any number between 539.2 and 792.4 with up to one decimal place.
The default PRESSURE setting for dissolved oxygen and ozone measurements is 760mm•Hg.
Select SALINITY from the configuration menu to view or change the salinity compensation for dissolved oxygen measurements.
The default compensation slope is 0 mS/cm. The system will accept any number from 0 to 999.9 with up to one decimal place.
Sensors are calibrated at the factory. When field calibrations are necessary, Air and 1-point sample calibrations are available.
For DO sensors that have not been active for several hours, it may take up to 1 hour before the reading becomes stable and accurate.
Select 1-POINT SAMPLE from the calibration menu to begin a dissolved oxygen span calibration.
When the 1-point sample method is chosen, the status box will prompt the user to enter the desired calibration value.
This calibration should be initiated when the sensor is in a liquid that is at the high end of the expected measuring range.
Stable temperature and sensor readings will also help to establish an accurate calibration.
Once the desired value for calibration is entered, the CALIBRATE button becomes active. Once the displayed sensor and
temperature values are stable and near expected values, click the CALIBRATE button to initiate the calibration or click EXIT
to return to the main measure screen.
Once the calibration has been initiated it will take between 30 seconds and 2 minutes depending on sensor and temperature
measurement stability. During this period, the abort sequence can be initiated as shown with the active ABORT button.
Exiting the calibration menu will return the user to the measure screen.
Select AIR from the calibration menu to begin a dissolved oxygen air calibration.
Hold the sensor in the air and wait for temperature and sensor readings to stabilize. Do not allow the sensor membrane
to dry out.
Once the calibration has been initiated it will take between 30 seconds and 2 minutes depending on sensor and temperature
measurement stability. During this period, the abort sequence can be initiated as shown with the active ABORT button.
When the point has been successfully calibrated, measured values will be based on the new calibration. The reading will be
in the range of 8 ppm to 10 ppm depending on the atmospheric pressure setting. Consult dissolved oxygen tables for more
specific information on expected readings after an AIR calibration.
Exiting the calibration menu will return the user to the measure screen.
• Clean off left over material by wiping with a soft cloth or paper towel and clean water. A bristle brush is not recommended.
• If the membrane is worn or punctured, replace the membrane cartridge as shown in the following figure.
4.6 Turbidity
When a Turbidity system is installed, the DataStick automatically sends measurement information in NTU. Temperature is also provided.
All ASCII commands for measurement, calibration, configuration and diagnostics shown in Appendix 7 apply. For Drinking Water Turbidity
Systems, an AV38 local interface is provided. Refer to the AquaClear™ Low-Range Turbidimeter User Guide (258483-001) for details.
DataStick Communications
Any DataStick communications adapter can
be connected directly to the wide-range turbidity
system for direct computer communications.
Optical Head
The Wide-Range Turbidity system can
measure to 4000 NTU with a sophisticated
series of light sources and detectors.
Select TEMPERATURE from the calibration menu to begin a temperature calibration. The status box will prompt the user to enter
the desired calibration value.
Once the desired value for calibration is entered, the CALIBRATE button becomes active. Once the displayed temperature value is stable
and near expected values, click the CALIBRATE button to initiate the calibration or click EXIT to return to the main measure screen.
Once the calibration has been initiated it will take about 30 seconds depending on temperature measurement stability. During this period,
the abort sequence can be initiated as shown with the active Abort button.
Exiting the calibration menu will return the user to the measure screen.
All Calibration Commands will respond with “OK” if accepted and “ERROR” if not accepted
The status of the calibration can be reviewed with the CALSTATUS command. When the last number returned is a 1, the calibration
is in process. When the last number returned is a 2, the calibration has been successful. Any other number indicates a specific failure
mode (see appendix).
Select STATUS and all information will be detailed in the following report.
In this example specific information can link the specific DataStick to calibration and test information at the factory. Whenever there
are issues in the field, this information can make it much easier for Thermo Fisher Scientific to assist in troubleshooting.
In this example we know that the sensor type has been properly detected and that configuration and calibration memory are valid.
Most importantly, the run status of the DataStick is okay.
Refer to the Programming Guide for specific diagnostic commands and responses.
Start
1. Cycle power.
No 2. Reset DataStick memory.
Detected Memory
OK
Fault not found: If failure is Sensor Memory, replace Sensor Head.
If failure is in Configuration or Calibration Memory, replace DataStick.
Yes
No
Measurement OK
1. Check sensor configuration to be sure appropriate parameters
are properly set. These include: buffer type, pressure, salinity,
temperature, reference temperature, and reference slope.
Yes 2. Check that the temperature reading is correct.
If it reads full-scale then the temperature element is open circuit.
3. Check that the calibration solutions are correct.
End
Fault not found: Refurbish or replace the sensor head.
Note: Any new sensor that has been in storage for more than a few days should be soaked in tap water
or a solution of known pH for at least 30 minutes before reviewing the span and offset performance.
Note: The electrodes of any pH sensor in storage must be kept moist at all times. Protective caps should
be filled with Thermo Scientific AquaSensors storage solution for best performance. The protective cap
should be placed securely over the sensor head.
If the sensor passes the visual inspection, there are three basic measurement tests that can verify normal operation. The temperature
measurement can be verified with an independent thermometer. Temperature comparisons should be done in liquid as it takes the sensor
much longer to reach steady state in air. The pH offset can be verified in 7 pH buffer. If the reading is more that 0.5 pH units off in 7 pH
buffer, then the sensor needs to be refurbished. The sensor should be stirred if the test is done in a beaker or at least the sensor head
should be swirled around for 10 seconds to make sure the solution gets through the reference junction. The pH span can be verified in a
sample solution that is more that 3 pH units away from 7 pH. Generally 4 pH buffer and 10 pH buffer work well. If the reading is more than
1 pH unit off then the sensor needs to be refurbished or replaced.
Make sure the correct configuration settings have been made:
• Correct Buffer Standards selected for buffer calibrations
• Temperature should read within 2 degrees centigrade of the expected value
Note: Any new sensor that has been in storage for more than a few days should be soaked in tap water or
Thermo Scientific AquaSensors ORP Storage Solution for at least 30 minutes before reviewing the offset
performance.
Note: The salt bridge of any ORP sensor in storage must be kept moist at all times. Protective caps should
be filled with Thermo Scientific AquaSensors Storage Solution for best performance. The protective cap
should be placed securely over the sensor head.
If the sensor passes the general inspection, there are two basic tests that can verify normal operation. If the sensor passes the visual
inspection, there are two basic measurement tests that can verify normal operation. The temperature measurement can be verified with
an independent thermometer. Temperature comparisons should be done in liquid as it takes the sensor much longer to reach steady state
in air. The ORP measurement should be tested with a solution that is at least +200 mV. If the reading is more than 100 mV off then the
sensor needs to be refurbished or replaced.
If the sensor has been given sufficient time after power is applied to reach the proper reading, several things can happen:
• If the sensor reads zero all the time: Replace the electrolyte and membrane cartridge.
• If the sensor reads near full scale and does not come down, give it additional time. If the electrolyte was shaken
vigorously prior to filling, additional time may be needed.
When the sensor head is placed in storage, keep the membrane moist with a wetted sensor cap.
When the sensor head is placed in storage, cap the metal electrodes to keep them clean and dry.
Please note that commands are common for each sensor type but that response codes may vary depending on the sensor connected.
A 1-point sample calibration command will be the same for dissolved oxygen as it is for toroidal conductivity. This simplifies ladder logic
coding for applications that presume the possibility of changes in sensor type.
Table 5: Data Reporting
Salinity is set from 0 to 999.9 mS/cm. Set up the average salinity for
Salinity for Dissolved
SSALT X compensation of dissolved oxygen
Oxygen Compensation X = 0 to X = 999.9 mS/cm measurement.
Pressure for Dissolved Pressure is set from 539.2 to 792.4 mmHg Set up the average pressure for
SPRESS X
Oxygen Compensation X = 539.2 mmHg to X = 792.4 mmHg compensation of measurement.
Conductivity Reference Conductivity reference temperature is set
SCRTEMP X Default value is 25˚C.
Temperature from 0˚C to 100˚C. X = 0˚C to X = 100˚C
Conductivity Conductivity slope is set from:
SCCSLOPE X Default is 2.0%/˚C
Compensation Slope X = 0 to X = 9.9% / ˚C
Points in the table where:
X is the table entry (X = 0 through 9) This table is designed to convert
Customized conductivity measurements to
Y is the Conductivity value in µS/cm.
Concentration Table SUPNT X Y Z concentration. Currently uses a built-in
Z is the Concentration value in %.
Natural Water table for temperature
Conductivity Range: 0 to 2,000,000.0 compensation. Up to 10 points.
Concentration Range: 0 to 99.99%
Customized When edits to all points are complete use Entries must be monotonic
SUTBL
Concentration Table SUTBL to activate the new table with increasing conductivities.
View Configuration Using the Where the DataStick replies with a number
Comments
Settings for… Command Text… that means…
PH:
0 = pH units
1 = mV units
ORP:
Sensor 0 = mV As different sensor heads are added to
Conductivity: the family of DataStick measurements
Units of Measure GSUNITS 0 = µS/cm this list will be expanded.
1=%
2 = TDS
3 = MΩ•cm
DO/Ozone:
0 = ppm
1=%
Temperature Units 0 = ˚C
GTUNITS
of Measure 1 = ˚F
Filter time constant from 0 to 100 seconds. The sensor process variable will reach
Sensor Data Filter GSFIL
Integers only. 90% of the final value in “X” seconds.
The temperature variable will reach
Filter time constant from 0 to 100 seconds. 90% of the final value in “X” seconds.
Temperature Data Integers only.
GTFIL Caution: It is advised to keep the
Filter
temperature and process filter
constants equal.
pH buffer calibration: the current
Buffer type set for auto recognition during reading will be changed to the closest
Buffer Set for pH buffer calibration. value in the buffer set selected.
GPHBUF
Calibration
0 = 4,7, 10 pH Buffer DIN Buffer Standards are:
1 = DIN Buffer Type 1.09, 4.65, 6.79, 9.23 and 12.75 pH
Salinity for Salinity setting between: Average salinity for compensation
Dissolved Oxygen GSALT
0 and 999.9 mS/cm of dissolved oxygen measurement.
Compensation
Pressure for Pressure setting between: Average pressure for compensation
Dissolved Oxygen GPRESS
539.2 to 792.4 mmHg of dissolved oxygen measurement.
Compensation
Calibration Procedure
The DataStick system can perform many different calibration types, e.g., air, 1-point buffer, 2-point sample, etc.
Aside from the calibration command used, the procedure is the same for all of them. Below is a description of that procedure.
1. Observe the calibration status by monitoring the Z field in the response from the CALSTATUS command (see above).
Prior to using the calibration command, the value will be 0 (no sensor calibration).
2. Prepare the DataStick for calibration and allow the measured readings to stabilize.
3. Issue the appropriate calibration command, e.g., CALS1PS to do a 1-point sample calibration.
4. Monitor the Z field and observe the change from 0 to 1 indicating that the calibration is in progress.
5. Continue monitoring the Z field. When it changes from 1 to 2 (Cal OK), the calibration procedure is complete and has been
successful. If it changes from 1 to 3 or greater, then the procedure is complete but has failed. For example, if it changes
from 1 to 7 (Fail – Value too low), it indicates that the measured value is too close to zero for a calibration to take place.
Take corrective action and repeat the procedure.
0 = No Sensor Connected
1 = pH
2 = ORP Confirmation that the correct
Check Sensor 3 = DO
GSTYPE sensor type is connected
Type 4 = Contacting Conductivity or configured properly.
5 = Toroidal Conductivity
6 = Ozone
7 = Turbidity
PH:
0 = Differential
1 = Combination
ORP:
0 = Differential Confirmation that the correct
Check Sensor
GSCAT 1 = Combination sensor category is connected
Sub-Type
Conductivity: or configured properly.
0 = 0.1 Cell (Nominal)
1 = 1.0 Cell (Nominal)
2 = 10.0 Cell (Nominal)
pH = − log10 H + [ ] (2)
Or
[H ] = 10
+ − pH
(3)
According to the Handbook of Chlorination and Alternative Disinfectants, Fourth Edition, 1998, equation (4) relates the free chlorine
concentration in ppm to the hypochlorous acid concentration in ppm, the pH, and the temperature in °C.
H+
⎬
[ ] (4)
⎩ ⎭
[ ]
The Chlorine DataStick measures the hypochlorous acid concentration HOCl and the temperature T of the solution. The pH
DataStick measures the pH of the solution. With these measurements equation (4) can be used to calculate the free chlorine
concentration of the solution.
For example, if the hypochlorous acid concentration is 4.99 ppm and the temperature is 27 °C as measured by the Chlorine
DataStick, and the pH is 7 as measured by the pH DataStick, then the free chlorine concentration in ppm will be
⎧ × 27 + 1.476 × 10−8 ⎫
[HOCl ] + [OCl − ] = 4.99 × ⎨1 + 0.056714 × 10
−8
⎬ = 6.49
⎩ 10− 7 ⎭
7.2 Calibration
[ ]
The Chlorine DataStick requires a hypochlorous acid concentration HOCl value when it is calibrated. If the user has a free
chlorine concentration value obtained from a colorimeter, for example, then the calculation done in equation (4) must be reversed.
⎧ −8
× (F 8 : 2 ) + 1.476 × 10−8 ⎫
(F 8 : 4) = (F 8 :1) × ⎨1 + 0.056714 × 10 ⎬ (6)
⎩ 10− ( F 8:3 ) ⎭
The free chlorine concentration value in F8:4 can then be displayed to the user.
7.3.2 Calibration
[
At calibration time, the user enters a free chlorine value that the PLC stores in F8:5. The PLC then calculates HOCl using ]
equation (5) and stores the value in F8:6:
(F 8 : 6 ) = (F 8 : 5) (7)
0.056714 × 10 × (F 8 : 2 ) + 1.476 × 10−8
−8
1+
10− ( F 8:3 )
The hypochlorous acid concentration value in F8:6 can then be written to the Chlorine DataStick to calibrate it.
Declaration of Conformity
Thermo Fisher Scientific certifies this product was tested thoroughly, inspected, and found to meet its published specifications
when it was shipped from the factory.
The DataStick is a DC powered measurement system with many sensing and communications options. Representative
combinations of the product have been tested and are certified as indicated to the following instrumentation standards:
Immunity: The equipment was tested to the Industrial level of performance in accordance with EN 61326, which includes
test criteria for electromagnetic compatibility for electrical equipment for measurement control and laboratory use per
89/336/EEC EMC. Supporting test records were written and certified by Thermo Fisher Scientific LLC.
Emissions: The equipment was tested to for radio frequency emissions in accordance with EN 61326:1998, which
includes test criteria for electromagnetic emissions limits for electrical equipment for measurement, control and laboratory
use. Supporting test records were written and certified by Thermo Fisher Scientific LLC.
Test Standards include:
EN55011 (CISPER 11) Conducted emissions from 150 kHz to 30 MHz Class B.
EN55011 (CISPER 11) Radiated emissions from 30 MHz to 1 GHz Class B.
Thermo Fisher Scientific warrants its DataStick against material and workmanship defect for a period of one year from the
date of shipment.
In the event that a defect is discovered during the warranty period, Thermo Fisher Scientific agrees, at its option, to repair or
replace the defective product. Any product repaired or replaced under this warranty will be warranted only for the remainder
of the original product warranty period.
This warranty does not apply to consumable products associated with this product including, but not limited to, chemical
reagents and salt bridges.
Products may not be returned without authorization from Thermo Fisher Scientific. To obtain authorization, please call
Thermo Fisher Scientific for a return material authorization number.
Limitations:
This warranty does not cover:
1. Damage caused by misuse, neglect (lack of appropriate maintenance), alteration, accident or improper application
or installation.
2. Damage caused by any repair or attempted repair not authorized by Thermo Fisher Scientific.
3. Any product not used in accordance with the instructions furnished by Thermo Fisher Scientific.
4. Damage caused by acts of God, natural disaster, acts of war (declared or undeclared), acts of terrorism,
work actions, or acts of any governmental jurisdiction.
5. Freight charges to return merchandise to Thermo Fisher Scientific.
6. Travel fees associated with on-site warranty repair.
This warranty is the sole expressed warranty made by Thermo Fisher Scientific in connection with its products. All other
warranties, whether expressed or implied, including without limitation, the warranties of merchantability and fitness for
a particular purpose, are expressly disclaimed.
The liability of Thermo Fisher Scientific shall be limited to the cost of the item giving rise to the claim. In no event shall
Thermo Fisher Scientific be liable for incidental or consequential damages.
This warranty is the sole and complete warranty for Thermo Fisher Scientific. No person is authorized to make any warranties
or representations on behalf of Thermo Fisher Scientific.
Thermo Fisher Scientific reserves the right to change or modify this warranty at any time.
The following terms and conditions will be presumed acceptable unless changes are made in writing and accepted
by both parties in a reasonable amount of time.
Any standard or boilerplate terms and conditions supplied with a written purchase order will not be applicable unless
accepted in writing by both parties.
Quotations: All quotations shall be in writing. Written quotations shall be valid for 30 days from the date issued.
Verbal quotations or price lists are not valid.
Pricing: All pricing is in US Dollars. Thermo Fisher Scientific reserves the right to change pricing without notice.
Terms: Payment terms are net 30 days from the date of invoice with approved credit. Thermo Fisher Scientific reserves
the right to deny credit or revoke previously extended credit. Past due accounts are subject to interest charges. Other
acceptable payment terms are cash, certified check, money order, credit card or letter of credit confirmed by any United
States of America bank. Other payment terms are not valid unless accepted in writing.
Sales taxes shall be included on the invoice unless a valid tax exemption certificate is supplied.
Return Material Authorization: Contact Thermo Fisher Scientific Customer Service for a Return Material Authorization
(RMA) number. Items returned without an RMA number will be rejected.
All returned merchandise must be in unused, resalable condition, and must not be contaminated with hazardous materials.
Cancelled orders must be returned within 30 days of the date on the invoice and shall be subject to expenses incurred
that may include, but are not limited to, inspection and restocking fees. Items returned within 60 days shall be subject
to a restocking charge that is equal to 15% of the purchase price. Items returned after more than 60 days shall be subject
to a restocking charge equal to 25% of the purchase price. Thermo Fisher Scientific reserves the right to reject any return
that is not under warranty after 60 days. Non-stock items are normally not returnable.
Transportation: Orders are shipped FOB Thermo Fisher Scientific, or factory, by the most efficient means available.
Appropriate charges, such as freight and insurance will be added to invoices. All shipments will be insured. Goods damaged
in shipment must be reported by the recipient to the freight carrier for claims.
Thermo Scientific
Environmental Instruments
Process Water Instruments
North America
166 Cummings Center
Beverly, MA 01915 USA
Toll Free: 1-800-225-1480
Tel: 1-978-232-6000
[email protected]
Netherlands
Tel: (31) 033-2463887
[email protected]
India
Tel: (91) 22-4175-8800
[email protected]
Japan
Tel: (81) 045-453-9175
[email protected]
China
Tel: (86) 21-68654588
[email protected]
Singapore
Tel: (65) 6778-6876
[email protected]
www.thermoscientific.com/water
© 2010 Thermo Fisher Scientific Inc. All rights reserved. All trademarks
are the property of Thermo Fisher Scientific Inc. and its subsidiaries.