MPFM Operating Instructions
MPFM Operating Instructions
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Operating instructions
for
Page 1 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
TABLE OF CONTENTS
1. Purpose ......................................................................................................................................... 4
2. Abbreviations / Definitions........................................................................................................... 4
3. Revision history ............................................................................................................................ 6
4. Documentation.............................................................................................................................. 6
4.1 Standard documentation and records ...................................................................................... 6
4.2 Location and Precautions ........................................................................................................ 6
4.3 Storage time ............................................................................................................................ 6
5. System description ....................................................................................................................... 7
6. Software installation..................................................................................................................... 8
7. Service console .......................................................................................................................... 10
7.1 Configuration and communication setup of SCP .................................................................... 10
7.2 Starting the program .............................................................................................................. 10
7.3 Graphical user interface......................................................................................................... 12
7.3.1 Upper status panel ......................................................................................................... 13
7.3.2 Lower status panel.......................................................................................................... 14
7.3.3 Upper graph ................................................................................................................... 14
7.3.4 Lower graph.................................................................................................................... 14
7.3.5 Process and technical log ............................................................................................... 15
7.3.6 Process log ..................................................................................................................... 15
7.3.7 Technical log .................................................................................................................. 15
7.3.8 Standard Conditions or Actual Conditions....................................................................... 16
7.3.9 Accumulated flow volumes ............................................................................................. 16
7.3.10 Graph axes ..................................................................................................................... 17
7.3.11 Mass flow........................................................................................................................ 17
7.4 Data logging .......................................................................................................................... 18
7.4.1 Parameter logging .......................................................................................................... 19
7.4.2 Retrieve MPFM logfiles ................................................................................................... 20
7.4.3 Exporting data ................................................................................................................ 21
7.4.4 Printing the Graphical Display......................................................................................... 22
7.5 Well test ................................................................................................................................ 23
8. Setup and configuration............................................................................................................. 26
8.1 Roxar MPFM 2600 service console program configuration .................................................... 26
8.1.1 Password ........................................................................................................................ 26
8.1.2 Parameter files ............................................................................................................... 26
8.1.3 Units of measure ............................................................................................................ 27
8.1.4 Time constants ............................................................................................................... 28
8.1.5 Panel title ........................................................................................................................ 29
8.1.6 Language ....................................................................................................................... 29
8.2 Roxar MPFM 2600 system configuration ............................................................................... 30
8.2.1 DP, Pressure and Temperature Transmitter set – up ..................................................... 30
8.2.2 Roxar Multiphase Salinity system set up ......................................................................... 31
8.2.3 Communication between Flow Computer and Service Console PC ................................ 32
8.2.4 Sensor and Venturi geometry settings ............................................................................ 33
8.2.5 Sensor calibration ........................................................................................................... 33
8.2.6 Electronics calibration ..................................................................................................... 33
8.2.7 Roxar MPFM 2600 module set – up ............................................................................... 33
Page 2 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
8.2.8 Sensors software upgrade .............................................................................................. 34
8.3 Data acquisition configuration ................................................................................................ 35
8.3.1 COM 6 serial port ........................................................................................................... 35
8.3.2 MODBUS register – fixed values .................................................................................... 36
8.4 Clock synchronization ............................................................................................................ 36
9. Calibration ................................................................................................................................... 37
9.1 Sensor calibration .................................................................................................................. 38
9.1.1 Permittivity calibration ..................................................................................................... 38
9.1.2 Conductivity calibration ................................................................................................... 38
9.1.3 Densitometer unit calibration .......................................................................................... 39
9.2 Fluid calibration...................................................................................................................... 40
9.2.1 Fluid parameter sets ....................................................................................................... 40
9.2.2 Density calculation .......................................................................................................... 41
9.2.3 Permittivity and conductivity ............................................................................................ 43
9.2.4 Conversion to standard conditions .................................................................................. 44
10. Diagnostics ................................................................................................................................. 47
11. Roxar MPFM2600 wetgas mode (if applicable) ......................................................................... 48
11.1 Configuration of Wetgas functionality .................................................................................... 48
11.1.1 Activate Roxar MPFM 2600 wetgas mode. ..................................................................... 50
11.1.2 Configure Wetgas calculation ......................................................................................... 50
11.1.3 Configure Wetgas permittivity calibration ........................................................................ 52
11.2 Fluid data ............................................................................................................................... 53
12. Units ............................................................................................................................................ 54
13. Nomenclature .............................................................................................................................. 55
Page 3 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
1. PURPOSE
This document contains the standard operating instructions submitted with each Roxar Multiphase
meter 2600. Unless you are already familiar with the Meter, we recommend that you review these
instructions before using the Meter.
2. ABBREVIATIONS / DEFINITIONS
Acc Accumulate Perm. Permittivity of the multiphase flow
Avg. Average Press Pressure
dP Diferential preasure across the Venuri Ref. Reference
element of the Roxar Multiphase meter
FC Flow Computer Roxar MPFM2600 Roxar Multiphase meter 2600
GOR Gas- oil ratio Roxar MSS Roxar Multiphase Salinity System
GVF Gas volume fraction Roxar SAM Roxar Sand monitor
Inst Instantaneous Std. Standard conditions
GMF Gas Mass Fraction Tflow Temperature of the multiphase flow
through the multiphase flow meter
LGR Liquid Gas Ratio Temp. Temperature
Pline Pressure of the multiphase flow WLR Water Liquid Ratio
through the multiphase flow meter
PC Personal Computer WVF Water Volume Fraction
PVT Pressure Volume and Temperature Xlm Lockhart-Martinelli parameter
Page 4 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Mix flow rate
The total amount of the two or three phases of a multiphase flow flowing through the cross-section of a
conduit per unit time; should be specified as either mix volume flow rate or mix mass flow rate
Mix velocity
The flow velocity of a multiphase flow; alternatively defined as multiphase volume flow rate divided by
conduit cross-section
Roxar MPFM Flow Computer
The heart of the Roxar Multiphase flow meter; controls sensor electronics, acquires sensor data and
calculates flow measurements
Multiphase flow
Two or more phases flowing simultaneously in a conduit. This document deals in particular with
multiphase flows of oil, gas and water.
Oil
Hydrocarbons in the liquid state at the prevailing temperature and pressure conditions
Oil-continuous multiphase flow
A multiphase flow of oil/gas/water characterized when the water is distributed as water droplets
surrounded by oil. Electrically, the mixture acts as an insulator.
Phase flow rate
The amount of one phase of a multiphase flow flowing through the cross-section of a conduit per unit
time, should be specified as either phase volume flow rate or phase mass flow rate
Phase volume fraction
The phase volume flow rate of one of the phases of a multiphase flow, relative to the multiphase
volume flow rate
Service Console PC
A personal computer which serves as the operator’s user interface to the Roxar MPFM flow computer
Volume flow rate
The volume of a fluid flowing through the cross-section of a conduit per unit time at the pressure and
temperature prevailing in that section
Water-continuous multiphase flow
A multiphase flow of oil/gas/water characterized when the oil is distributed as oil droplets surrounded
by water. Electrically, the mixture acts as a conductor.
Water cut (WC)
The water volume flow rate relative to the total liquid volume flow rate (oil and water), both converted to
volumes at standard pressure and temperature; normally expressed as a percentage
Wet gas
Gas that contains liquids, generally wet gas is defined as gas/liquid systems with a Lockhart-Martinelli
parameter smaller than approximately 0.3. Hydrocarbon gasses that contain heavy components that
will condensate during further processing (but at a particular p and T behaves as a pure gas) are not
considered to be a wet gas from a measurement point of view.
Page 5 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
3. REVISION HISTORY
Revision Issued Reason for issue Author Checker QA Checker Releaser
Sanjay Frode Hugo
A 12.11.2009 Original document Janne Saether
Srivastava Aase
Sara Frode Hugo
B 22.07.2011 New template
Paraschiv Aase
Updates in section: 2, 4.1, 7.3.2.
Iulia Geir Martijn
C 17.12.2013 Added section 8.2.2;
Isman Kristensen Tol
Included chapter 11
4. DOCUMENTATION
4.1 Standard documentation and records
Note: If the client requires notification before records are discarded, this must be stated on the record itself.
Page 6 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
5. SYSTEM DESCRIPTION
The Roxar MPFM 2600 system has an embedded flow computer that acquires and processes data
from the sensor electronics and performs the necessary calculations in order to determine flow rates.
Since this embedded computer has no operator interface, a separate computer running the Roxar
MPFM 2600 service console software is used for calibration and configuration. The Roxar MPFM 2600
Service console software is the operator’s access point to the Roxar Multiphase flow meter. The Roxar
MPFM 2600 service console software has several options. Figure 1 shows a typical system installation
and available options.
The Roxar MPFM 2600 flow computer is the heart of the system. The service console PC is a personal
computer that serves as an operator interface to the flow computer. The data acquisition system is a
customer supplied system which connects multiphase flow measurements to the process management
system at the installation site.
The service console software displays measurement data and instrument status from the meter unit.
The program may also function as a well test program and a data logging program where an external
data acquisition system is not available.
Under normal operations, the program displays flow parameters in real time, both numerically and
graphically. The graph axes may be easily changed to suit the needs of current flow conditions.
Measured data may also be logged in a buffer in the Roxar MPFM 2600 flow computer and/or to a file
on a PC.
Page 7 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
6. SOFTWARE INSTALLATION
The entire system will normally be installed and properly configured by the Roxar engineer who installs
and commissions the flow meter. The Roxar MPFM 2600 service console software is installed by
running the installation program
which creates all of the necessary folders and shortcuts and copies all of the necessary files to their
appropriate locations. Follow instructions on screen (see Figure 2, Figure 3 and Figure 4) to complete
the installation.
Page 8 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Restart computer by selecting “Yes” for the configuration changes made to Roxar MPFM 2600 service
console to take effect, as shown in Figure 5.
Page 9 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
7. SERVICE CONSOLE
Page 10 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Once the Roxar MPFM 2600 flow computer is initialized, start the MPFM 2600 service console
software. First click the Start button in the lower right corner of the screen to open the Start menu.
Next, select Programs to open the Programs submenu and then select Roxar MPFM 2600 to open
the Roxar submenu. Finally, select the Roxar MPFM 2600 Service Console to start the program (see
Figure 7).
Page 11 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Page 12 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Page 13 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
7.3.2 Lower status panel
The lower status panel displays real-time measurement values from the Roxar MPFM 2600 flow
computer.
This panel has four sections:
The phase volume flow rates are in real time and are indicated in different colors; oil in red, water in
blue and gas in green, see Figure 9 The accumulated values are always from the time the register was
last reset. The accumulation registers may be reset by executing the Options|Reset Accumulated
Volumes command. See section 7.3.9
The real time water cut, gas volume fraction, gas oil ratio (at standard conditions) and mix velocity are
displayed.
The real time transmitter values with flow temperature, pressure inside the meter and the differential
pressure across the Venturi are displayed. The real time mix density and the permittivity of the flow
are also displayed.
As an option, the Roxar MPFM 2600 can be installed together with a Roxar Sand monitor. The bottom
section displays the sand flow rate and accumulated sand mass, as well as the status of the Sand
meter. If no Sand meter is installed, this section displays the Roxar logo.
As another option, the Roxar MPFM 2600 can be installed together with a Roxar Multiphase Salinity
System (Roxar MSS). The bottom section displays the measured water conductivity (CondP) and
calculated water salinity (SalP%) that is submitted by the Roxar Multiphase Salinity System to the
MPFM 2600 flow computer if the Roxar MPFM 2600 is installed with the Roxar Multiphase Salinity
System connected.
The graphs always display the most recent measurements. Once a graph is filled, new measurements
are added at the right edge and previous measurements are moved to the left. The horizontal and
vertical scales of this graph may be changed by executing the View|Graph Axis Span command. See
section 7.3.10.
Page 14 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Figure 10 : Choose the parameter to display in the lower graph via the graph select button
Figure 11 : The Process and Technical Alarm panels are located at the bottom of the main SCP window.
Page 15 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Figure 12: Process Status Log Figure 13: Technical Status Log
Page 16 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
7.3.10 Graph axes
The scale of the upper and lower graphs may be changed by executing the View|Graph Axis Span
command. This command opens the dialog box shown in Figure 14.
In the Upper Graph Settings panel, enter the desired minimum and maximum values for the Y-axis of
the upper graph. The Flow Display Span is the Y-axis span for oil and water flow rates regardless of
whether actual or standard conditions are used. The Actual Gas rate Span is the Y-axis span for gas
flow rates displayed according to actual conditions. The Std. Gas rate Span is the Y-axis span for gas
flow rates displayed according to standard conditions.
Figure 14: Graph Axes Span dialog box used to change the scale of the upper and lower graphs displayed
in the Multiphase Flow Meter dialog box.
In the Lower Graph Settings panel, enter the desired minimum and maximum values for the Y
(vertical) axis of the lower graph. The values to be entered here will be dependent on which parameter
is selected to be displayed in the lower graph. In the Common Graph Settings panel, enter the time
span to be displayed across the X-axis of both graphs.
Once the desired settings have been entered, click on the OK button to close the dialog box. Note that
when changing graph scales, the graph is cleared before data is presented with the new scale(s).
The units in which the measured values are displayed may be changed by executing the
Options|Units command (described in section 8.1.3).
Mass flow data is shown numerically. The Service Console program does not display mass flow
graphically in this display.
As an alternative, mass flow can be displayed graphically and numerically on the normal display (see
Figure 8) by selecting Mass Flow rate instead of volumetric flow rates (described in section 8.1.3).
Page 17 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Since the data acquisition system is a customer supplied system, this document can only mention this
possibility for data logging. The other data logging methods are described below.
Page 18 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
7.4.1 Parameter logging
The parameter logging is enabled by executing the File|Parameter logging command, which opens
the log configuration dialog box shown in Figure 16.
In the upper part of the menu select the desired parameter to log from a scroll down menu.
In the log setup in the lower left corner logging data to the PC will be enabling and disabling
accordingly. It is not possible to disable internal logging within the flow computer.
In the text box “Log interval”, enter the desired number of minutes between each entry in the log files.
Click “OK” to save any changes.
Data is logged in the same units as those displayed by the program. If the units are changed while
data logging is enabled, new column headers with the new units are added to the log file.
The local log file may be easily imported into a spreadsheet, word processor or database. It is a text
file containing the selected parameters in addition to time and date.
The local log file is stored in the LogFiles subfolder of the program folder (typically C:\Program
Files\Topside\) with a filename in the form YY_MM_DD.LOG where YY, MM and DD are the year,
month and day of the current date1.
1
NOTE: The offline version of the Service Console program creates all logfiles in the folder named
Offline located under the Topside folder.
Page 19 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
7.4.2 Retrieve MPFM logfiles
Because the log buffer is stored in the flow computer, its contents must be retrieved in order to be
accessible in the service console PC (or other PCs). This is done by executing the File/Retrieve
MPFM Logfile command, which opens the dialog box shown in Figure 17.
Figure 17: Log buffer must be retrieved from the flow computer to be used in another program.
The retrieved log file will be stored in the MPFM 2600 files subfolder of the program folder (typically
C:\Program Files\Topside\). By default, the filename will have the form YY_MM_DD_A240 or
A241.BUF where YY, MM and DD are the year, month and day of the current date. The file name may
be changed by clicking on the Change Name button and entering the desired file name in the dialog
box which opens. It is not possible to change the folder to which the file will be saved.
The retrieved log file may be easily imported into a spreadsheet, word processor or database. It is a
tab-delimited text file containing the parameters specified in the log configuration dialog box in Figure
16.
Page 20 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
7.4.3 Exporting data
Data acquired by the Service Console program may be exported to Microsoft Excel using a DDE
2
(Dynamic Data Exchange) link . This data has a format similar to the local log file format, but includes
all data points rather than just those included in the log file.
Users which have access to the Microsoft Excel spreadsheet program, may export data to an Excel
spreadsheet directly by executing the File |Excel DDE command. This command opens the dialog
box shown in Figure 18.
The full path and file name of the Excel program file must be entered in the Excel Path text box. Click
on the Launch Excel button to run the program specified in the text box. The full path and file name of
the Excel file to be opened must be entered in the Excel File text box3. Click on the Open Excel Sheet
button to open the specified Excel file and transfer the specified number of lines of data to the file.
Status and error messages for the transfer are shown in the Conversation Status panel.
The Topside program has no limitation on the number of lines that may be transferred. However, the
maximum number of rows allowed in an Excel spreadsheet should not be exceeded. (Remember to
allow for 2 column header rows, plus an additional 2 rows for each time the display measurement units
are modified.)
Once the transfer is initiated, the dialog box may be closed. Once all transfers are completed, click on
the Close DDE connection button (execute the File|Export|Excel DDE command again, if
necessary) to close the connection to the Excel file, and then close the dialog box.
The following parameters may be transferred using the DDE functionality:
2
DDE: A form of interprocess communication (IPC) implemented in the Microsoft Windows family of operating
systems. Two or more programs that support dynamic data exchange (DDE) can exchange information and
commands.
3
NOTE: The DDE interface can only open existing Excel-files, not create new files.
Page 21 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Time and Reset Time (current time and time registers were last reset)
Standard Volumes (Oil, Water and Gas)
Standard Flow Rates (Oil, Water and Gas)
Actual Volumes (Oil, Water and Gas)
Actual Flow Rates (Oil, Water and Gas)
Temperature
Pressure
Differential Pressure
Alarms
Some of these parameters may also automatically be plotted in graphs:
Standard Flow Rates (Oil, Water and Gas)
Actual Flow Rates (Oil, Water and Gas)
Figure 19 : The Print to File dialog box creates a bitmap file of the program window
Page 22 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
7.5 Well test
The well test panel is a data acquisition application that can run for one or more hours where data is
acquired and saved in a standard reporting format.
The well test window (Figure 20) is opened by executing the File|Well Test command. The well test
window is similar to the main window displaying flow rates both graphically and numerically.
Instantaneous and average flow rates as well as accumulated flow volumes are displayed at both
actual conditions and standard conditions. All values are shown in the same units which are currently
selected for the main window.
Figure 20 : The Well Test Program window presents results of the well test both graphically and
numerically.
A text box located below the lower graph may be used by the operator to enter any comments. This
text will be included in reports and printouts.
The well test window has a number of menu items. Unlike the main window, clicking on a menu in the
well test window executes the selected command immediately rather than opening a menu.
After first opening the well test window, the dialog box shown in Figure 21 is displayed. This dialog box
may also be opened by executing the Setup|Well Test Setup command. The well test duration and
the well identification number may be entered.
Page 23 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Figure 21 The Well test Setup window is used to enter well names, assign parameter sets and choose
which well is to be tested.
To start a well test, click on the Start Well test button. When asked to confirm starting the well test,
click on the Yes button to start the well test. To stop a well test, click on the Stop Well test button and
confirm accordingly.
At the completion of a well test, the dialog box shown in Figure 22 opens. This dialog box may also be
opened by executing the Setup|Report Setup command. In this dialog box, the filename of the well
test report may be specified. By default, the filename will have the form YYMMDD_N.REP where YY,
MM and DD are the year, month and day of the current date and N is an index number for the well
test(s) carried out on this date. In the selected file dialog box, the file name may be changed by
clicking on the Change Name button and entering the desired file name in the filename text box. It is
not possible to change the folder to which the file will be saved. The report file will be stored in the Well
Tests subfolder of the program folder (typically C:\Program Files\Topside\).
Page 24 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Figure 22 : The Well Test Reports dialog box, for specifying output of the well tests.
By executing the Setup View Completed Well test command, a table of data from the previous 50
well tests as shown in Figure 23 is displayed. This command is disabled during a well test.
Figure 23 : The Well test overview dialog box is for viewing well test results.
By executing the Setup|Close panel command, the well test window is closed and the main program
window returns control. This command is disabled during a well test. It is not possible to start a new
well test without closing and reopening the well test window.
Page 25 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
8. SETUP AND CONFIGURATION
The Roxar MPFM 2600 service console program allows data to be displayed in different units and
languages. In addition, the program may be used to configure communication with the Roxar MPFM
2600 flow computer and between the Roxar MPFM 2600 flow computer and a customer supplied data
acquisition system.
Access to all parameters which influences the set-up and configuration is password protected. It
should not be necessary to modify configuration parameters under normal operation, since the meter
and software will be properly installed and configured by Roxar personnel when commissioned.
The first level default user password is set by the manufacturer and is available to operators who have
completed the basic user training course. The configuration menu is protected with a second level
password. This menu contains settings that may affect the measurements and should only be used by
operators with detailed knowledge of the system, normally Roxar service engineers or operators with
MPFM advanced training.
The current user password may be changed by executing the Options|Change User Password
command. First enter the current password. Then, enter the new password twice. If the current
password is entered incorrectly, or there is any discrepancy between the two entries of the new
password, the user password will not be changed (see Figure 24).
Page 26 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Figure 25 : The Parameter File in use by the flow computer is shown in the main SCP window.
To download a new set of parameters from the PC to the flow computer, execute the File|Change
Parameter File command and select an existing parameter file to download. To save the current
parameters in a new parameter file, execute the File|Save Parameter file command.
The units, in which measurement data are displayed, may be changed by executing the
Options|Units command which opens the dialog box shown in Figure 26.
Figure 26 : The Parameter Unit Setup dialog box is used to change the units of the measurement data.
Page 27 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
8.1.4 Time constants
The average time for the data may be changed by executing the Configuration|Time Constants
command which opens the dialog box shown in Figure 27.
There are two time constants that may be set. The Measurement average period specifies the
number of seconds over which the raw data is averaged before being applied in the multiphase
calculations.
Caution: If this value is specified incorrectly, the accuracy of flow measurements will
be compromised.
The Flow rate average period (if different from the Measurement average period) causes the
graphically displayed flow rates curves to be averaged over a different period than other
measurements. This averaging only affects the visual presentation of the curves and does not alter
the multiphase calculations.
Page 28 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
8.1.5 Panel title
In cases where more than one flow meter is installed, it will be desirable to indicate which MPFM 2600
service console program applies to which meter. Execute the Options|Panel Title command which
opens the dialog box shown in Figure 28, and add an additional text string to the window title. The text
string entered in this dialog box will be displayed to the right of the RFM MPFM 2600 Multiphase Flow
Meter text in the title bar of the main window. It is obligatory to save parameters and restart the
program before the text string is added to the title bar.
Figure 28 : The MPFM Panel Title dialog box is used to change the SCP window title.
8.1.6 Language
The Roxar MPFM 2600 service console program is normally supplied in English. If necessary, the
program can be supplied with an alternate language. In such cases, the Options|Language
command, which opens the dialog box shown in Figure 29, may be used to select a language. The
program must be restarted before the selected language will be applied. If an alternate language that
has not been installed is chosen, this command takes no action.
Figure 29 : The SCP language can be changed via the User Interface Language dialog box.
Page 29 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
8.2 Roxar MPFM 2600 system configuration
8.2.1 DP, Pressure and Temperature Transmitter set – up
The Options|Transmitter Inputs command opens the dialog box shown in Figure 30.
Figure 30 : Transmitter Inputs dialog box and RMSS Inputs dialog box
The service console program can raise an alarm if a pressure or temperature exceeds normal
operating limits. When a pressure or temperature is outside its specified alarm limits, the
corresponding text box in the main window will have a red background rather than a white, and a
description of the alarm condition is displayed in the message panel below the lower graph. To set the
high and low alarm limits, access the Transmitter Inputs dialog box via the Options/Transmitter
Inputs menu (see Figure 30, top of the screen).
Enter appropriate low and high alarm limits for each sensor. Click on the up or down arrow to the left
of a value to increase or decrease the currently specified value. Alternatively, the value may be
changed by selecting and entering a new value.
The Transmitter units (i.e. software settings, ref Figure 30) must be set to the same unit as the
transmitter unit itself (i.e. the unit in which the transmitter has been set using a HART communicator).
Page 30 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
The Transmitter units are specified by Roxar during meter installation. Therefore, it
is not required to make changes under normal operations.
Caution: If these parameters are specified incorrectly, the Service Console program
will not be able to obtain correct data from the transmitters.
The “conductivity input” “high limit” and “low limit” on the right hand side of the screen gives the
opportunity to define cut off values for valid conductivity reading from the Roxar MSS.
The conductivity average field indicates the value used in the flow calculations.
There are various alarms and statuses that can be read from the Roxar Multiphase Salinity system in
the MFPM service console. The most important ones are:
“Comtat”: The field turns yellow in case the communication between the Roxar MSS and the Roxar
MPFM flow computer is not up and running.
Alarms: “Process Pressure”, “Process temperature”, “Mix density”, “Venturi dP”, “microwave system”
indicate that one or more of these parameters that are exchanged between the Roxar MPFM flow
computer and the Roxar MSS are out of range or missing and the Multiphase Salinity System cannot
perform optimal.
Warnings: Indication a possible reason why the Roxar MSS is not able to give reliable water conductivity
readings. Interpretation of these alarms is to be done by a Roxar service engineer. These alarms do not
necessarily mean that the system is not working as designed, it just indicates why under certain
conditions there is no reliable output from the system, for example in oil continuous flow. There are also
temperature alarms (“interface card temp” and “microwave temp”) that may be used to initiate preventive
action in case of overheating of the electronics.
Page 31 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
8.2.3 Communication between Flow Computer and Service Console PC
The communications link between the Roxar MPFM 2600 flow computer and the service console PC is
specified by executing the File|PC Communication Setup command which opens the dialog box
shown in Figure 31.
Figure 31 : Communication link between the flow computer and the Service Console PC is set up by using
the PC Communication Setup dialog box
The ComPort is the serial port of the service console PC which is connected to the Roxat MPFM 2600
flow computer.
The Protocol used for communication will normally be Modbus ASCII for RS-232 communication and
Modbus TCP for Ethernet communication.
The default Modbus Slave Address for the flow computer is 240. When the same service console
program accesses more than one flow computer, different slave addresses may be applied.
The connection between the service console PC and the Roxar MPFM 2600 flow computer may be
established in one of the following manners:
Page 32 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
RS232 to Ethernet
The specified serial port of the service console PC is connected to an external Ethernet adapter which
is connected to the local area network. The Roxar MPFM 2600 flow computer is also connected to the
local area network.
RS232
The specified serial port of the service console PC is connected directly to the COM1 serial port of the
Roxar MPFM 2600 flow computer.
Ethernet
The service console PC is directly connected to the local area network. The MPFM 2600 flow
computer is also connected to the local area network.
If an Ethernet or RS232 to Ethernet connection is applied, the IP address of the MPFM 2600 flow
computer must be specified in the lower part of the dialog box.
An Ethernet connection may be tested by clicking on the Connect to MPFM button. An attempt will be
made to communicate with the MPFM 2600 flow computer and the results of the test are displayed in
the Connection Status panel.
The Sensor geometry settings are located in the Configuration|Sensor Geometry/Venturi menu.
This menu contains information about sensor inner diameter; Venturi beta and discharge factor in
addition to some other information regarding distance between measurement electrodes. These
settings should never be changed.
The Sensor calibration settings are located in the Configuration|Sensor Calibration menu. These
settings should never be changed.
The Electronics calibration settings are located in the Configuration|Electronics Calibration menu.
These settings should only be changed if the field electronics units are changed. This task can only
be performed by Roxar service personnel.
The Roxar MPFM 2600 module set-up is located in the Configuration MPFM module menu. These
settings should only be changed if the Roxar MPFM 2600 system is changed by removing or adding
hardware modules. This task should only be performed by Roxar service personnel.
Page 33 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
8.2.8 Sensors software upgrade
The Roxar MPFM 2600 sensor software can be upgraded by selecting the Sensor V3.xx.exe file in the
Configuration|Update Flow computer SW menu. Downloading will start automatically. See Figure
32.
A progress bar will indicate the downloading process. After downloading has been completed, the
system will automatically enter the parameter file menu where the correct parameter file is
downloaded. Also, it will automatically enter the Sensor Computer clock menu where the clock is
synchronized.
This procedure should only be carried out by trained MPFM specialist or by Roxar
service personnel.
Caution: If the incorrect code is downloaded, malfunction of the system may occur or
cause incompatibility between service console software and flow computer code.
Page 34 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Specify the Modbus address of the Roxar MPFM 2600 flow computer (slave address), Modbus protocol
type, parity, number format, request quantity and baud rate for the communication link.
In order for changes in the COM 6 communication setup to take effect, the Roxar MPFM 2600 flow
computer must be powered off and restarted.
Page 35 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
8.3.2 MODBUS register – fixed values
In order to ease the verification of the serial link, most Modbus registers can be set to a predetermined
fixed value. In this manner the operator can easily verify that the numbers are read correctly. This
command is activated by executing the Configuration|Set fixed values command (See Figure 34).
Check the Activated box in order to activate this function. Note that if the service console program or
flow computer are switched off and on, this function will be cancelled.
The current date and time of the service console PC is shown in the yellow text boxes in the upper part
of the dialog box. Clicking on the Transfer PC time to MPFM Flow Computer copies the current date
and time of the service console PC into the white text boxes in the lower part of the dialog box.
Clicking on the OK button resets the Roxar MPFM 2600 flow computer clock to the time shown in the
white text boxes, and closes the dialog box.
Page 36 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Figure 35 : Use this dialog box to synchronize Flow computer & PC Clock
9. CALIBRATION
There are a number of parameters that affect the calibration of the multiphase flow meter. Calibration
of the field impedance sensors and densitometer are related to the meter hardware. Properties of the
fluids, such as permittivity, conductivity and density are related to the flow at the installation site. It is
important that these parameters are properly specified in order to obtain the most accurate and reliable
measurement results.
Access to calibration parameters are password protected. It should not be necessary to modify
calibration parameters under normal operation, since the meter and software will be properly installed
and configured by Roxar personnel when commissioned. Roxar makes the default user password
available to operators during training courses.
All of the configuration and calibration parameters are stored in a parameter file. It is possible to
create different parameter files for different measurement conditions. By default, the most recently
used parameter file is used when the program is restarted. To load the parameters from an existing
parameter file, execute the File|Change Parameter File command. To save the current parameters in
a new parameter file, execute the File|Save Parameter File command.
Page 37 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
9.1 Sensor calibration
9.1.1 Permittivity calibration
The Option|Capacitance Zero Calibration command opens the dialog box shown in Figure 36. The zero
calibration procedure for the impedance sensors is described in the Maintenance Instructions [1]. The
Update MPFM button stores the calibration results in the Roxar MPFM 2600 Flow Computer.
Page 38 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
9.1.3 Densitometer unit calibration
The densitometer calibration is required only when the hardware is installed. The Gamma source &
detector hardware is not installed in the Non-Gamma version of the Roxar MPFM 2600. The Non-
Gamma method is based on improved algorithms for time series analysis, and analyse the variation in
permittivity measurements combined with a characterisation of the signals over time.
When the gamma densitometer hardware is installed, the calibration procedure described in the
Maintenance Instructions [1] should be followed. The Options|Gammameter Calibration command
opens the dialog box shown in Figure 37. The OK button stores the calibration results in the Roxar
MPFM 2600 Flow Computer.
This procedure should only be carried out by trained personnel. Caution: If these
values are specified incorrectly, the accuracy of flow measurements will be
compromised.
Page 39 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
9.2 Fluid calibration
9.2.1 Fluid parameter sets
There are a number of parameters that affect the calibration of the Roxar Multiphase meter.
Calibrations of the gamma densitometer and capacitance units are related to the meter hardware.
Properties of the fluids, such as permittivity, conductivity and density are related to the flow at the
installation site.
The meter and software, including fluid properties, will be properly installed and configured by a Roxar
service engineer during the commissioning and start-up phase. Operators may notice after some time
what appears to be a change in the performance of the meter, this is not necessarily an indication of a
problem with the meter; it could simply mean that the properties of the well(s) have changed. A
possible solution is to update the fluid properties and the PVT tables.
Caution: If these values are specified incorrectly, the accuracy of flow measurements
will be compromised.
There are several types of fluid parameters that affect flow calculations: density, permittivity and
conductivity of the phases, as well as the relationship between standard and actual pressure and
temperature. Fluid parameter sets may be specified by executing the Fluid/Fluid Calibration Set
command, which opens the dialog box shown in Figure 38, Clicking on an Edit button opens the
appropriate dialog box for parameter entry. In the case of both density calculations and conversion to
standard conditions, the user will first be asked to specify which method of parameter entry (lookup
table or default equations) is to be used.
Figure 38 : Fluid Calibration dialog Box, which gives access to fluid property Data
Up to 5 different fluid parameter sets can be specified (called Well 1 – 5). Parameters for conversion to
standard conditions and for density may be entered as default equations or as lookup tables. Default
equations assume linear temperature and pressure coefficients. Lookup tables allow parameters to be
specified at 5 different pressures for each of 5 different temperatures.
Page 40 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
In general, there are three common situations where an operator may want to specify different fluid
sets:
Different wells have the same downhole pressure but significantly different GOR (gas-oil ratios).
Wells may be in the same reservoir but located at significantly different reservoir depths.
There are significant differences in the density of the crude being pumped up at different locations.
All five calibration parameter sets are stored in the Roxar MPFM 2600 flow computer as well as in the
current parameter file. The check mark under Well Selection indicates what set of calibration
parameters that are to be applied. Any parameter set may be edited, regardless of whether or not it is
currently selected for use.
Caution: If these values are specified incorrectly, the accuracy of flow measurements
will be compromised.
If the lookup table method is selected, the dialog box shown in Figure 40 will open. This dialog box
has five panels: One for lookup table headers, one lookup table for each phase and one for densities
at standard conditions for mass flow calculations.
Normally, this lookup table is generated by an external program and stored as a text file which may be
imported by clicking on the Update table from file button and selecting the appropriate text file or
entering the numbers manually into the table.
A special version of the PVT software can be supplied with the Roxar MPFM 2600. The PVT software
will, based on composition data, generate a density table which can be imported into the service
console program. This manual does not describe how to operate the PVT software.
Page 41 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Figure 40 : Lookup Table dialog box allows density parameters to be entered manually or from a file
Each lookup table consists of an array of 25 densities at each of 5 different temperatures and
pressures. The lookup table entries are specified in the Oil, Gas and Water density panels.
In the lower panel for mass flow calculations, enter the density of each phase at standard temperature
(15.0 °C) and pressure (1.01 bar (a)). Note that these three values are NOT updated when density
data are imported from file. These data must be manually updated when necessary. Clicking on the
Update MPFM button copies the lookup table data to the Roxar MPFM 2600 flow computer and closes
the dialog box. The Exit without update button closes the dialog box without making any changes to
the parameters in use.
Caution: If these values are specified incorrectly, the accuracy of flow measurements
will be compromised.
If the default equation method is selected, the dialog box shown in Figure 41 opens. This dialog box
has four panels: one for each phase and one for densities at standard conditions to be used for mass
flow calculations. The default equation is:
ρ = ρR + KP(p - pR) + KT(T - TR)
where:
Oil density ρR density of the phase in the process flow
Density calibration temperature TR temperature at which the density was measured
Density temperature coeff. KT linear temperature compensation coefficient
Density calibration pressure pR pressure at which the density was measured
Density pressure coeff. Kp linear pressure compensation coefficient
Page 42 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
In the panel for mass flow calculations, enter the density of each phase at standard temperature
(15.6 °C) and pressure (1 bar(a)).
Clicking on the OK button copies the data to the Roxar MPFM 2600 flow computer and closes the
dialog box. The Cancel button closes the dialog box without changing any of the parameters in use.
Page 43 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Caution: If these values are specified incorrectly, the accuracy of flow measurements
will be compromised.
In the upper panel, the permittivity of the oil is specified. The equation used for calculating permittivity
is: = R + KP(p - pR) + KT(T - TR) where:
Oil Permittivity R permittivity of the oil in the process flow
Oil calibration temperature TR temperature at which the permittivity was measured
Oil cal. temperature coeff. KT linear temperature compensation coefficient
Oil calibration pressure pR pressure at which the permittivity was measured
Oil cal. pressure coeff. Kp linear pressure compensation coefficient
Page 44 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
9.2.4.1 Lookup table method
Caution: If these values are specified incorrectly, the accuracy of flow measurements
will be compromised.
If the lookup table method is selected, the dialog box shown in Figure 44 opens. This dialog box has
six panels: one for lookup table headers and one lookup table for each of the coefficients used in the
conversion to standard conditions.
Normally, these lookup tables will be generated by an external program and stored in a text file which
may be imported by clicking on the Update table from file button and selecting the appropriate text
file or entering the numbers manually into the table.
A special version of the PVT software can be supplied with the Roxar MPFM 2600. Based on
composition data the PVT software will generate a black oil table which can be imported into the
service console program. This manual does not describe how to operate the PVT software.
Page 45 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
9.2.4.2 Lookup table for conversion from actual to standard conditions
Each lookup table consists of a 5x5 array of 25 coefficient values at 5 temperatures and pressures
each. The lookup table entries are specified in the Bo, Rs, Bg, Rv and Bw panels. Clicking on the
Update MPFM button copies the lookup table data to the Roxar MPFM flow computer and closes the
dialog box. The Exit without update button closes the dialog box without making any changes to the
parameters in use.
Caution: If these values are specified incorrectly, the accuracy of flow measurements
will be compromised.
If the default equations method is selected, the dialog box shown in Figure 45 opens. There are 14
coefficients to be specified which are used in the following five default equations:
Figure 45 : Standard Conditions dialog box used to specify the parameters needed to convert actual data
to standard conditions.
Clicking on the OK button copies the coefficients to the Roxar MPFM 2600 flow computer and closes
the dialog box. The Cancel button closes the dialog box without changing any of the parameters in
use.
Page 46 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
10. DIAGNOSTICS
Roxar Multiphase meters are extremely robust and contain redundant systems to protect against
failure. If a meter malfunctions or if the operator experiences a problem which he is not able to correct
himself, then a diagnostics test should be performed. For example, the measurement data from the
meter may be inaccurate, the measurement data may be completely wrong, or there may be no data at
all (all zero). Normally, a malfunctioning meter would be checked and diagnosed by Roxar service
personnel, but some of the situations above might be caused by incorrect set-up or calibration of the
meter. To check these possibilities, the user can create diagnostics files that can be sent to the Roxar
service department for diagnostics.
The diagnostics is enabled by executing the Options|Create Dignostics Files command. This
function will perform a diagnostic of the Roxar MPFM 2600 system.
Before starting the diagnostic test, you will be prompted to name the file Figure 46.
A number of relevant files will be collected. Among these files are technical alarm logs, process alarm
logs, parameter settings, reset logged data files and several other files. All files will be stored in a
folder which will be generated automatically (normally C:\Topside\Diagnostics\). After completing the
diagnostics, you will be asked if the files are to be stored onto a floppy disk. Press YES to copy data
to the floppy disk (See Figure 47).
Page 47 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Page 48 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
The service console report actual mass flow rates and visualize these in graphical trend lines. The
volumetric flow rates at standard condition are also reported.
The steps needed to configure wetgas operation will be described in the following chapters. Operation
of the meter in the multiphase region is covered in the operating instructions for the Roxar MPFM 2600
[2]
Note that to access the screens to configure the wetgas functionality both user and configuration
password is required.
Page 49 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
11.1.1 Activate Roxar MPFM 2600 wetgas mode.
To enable the wetgas functionality the first step is to enter the configuration menu and select sensor
calibration. Then the window tab CapSens conf found under the configuration menu must be selected,
shown in Figure 50.
To activate wetgas calculation the Fraction Model selector must be set to 5 (Wetgas) and then press
the Update MPFM button to update the FC with the new setting. Finally press the Exit button to close
the menu and return to the service console. After a FC update (approximate 10 seconds) the service
console will change to show wetgas readings as shown in Figure 49.
If the meter is equipped with a density system this must be ticked on. Currently the measured density
is only used as an inspection value, the density used in calculations is from PVT-tables. The CapCal
selection should not be ticked on, since this is only used during factory calibration of the meter.
Page 50 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
If the meter is operated with manual WLR input the selector must be set to WLR from manual input
and the current WLR must be entered.
If the meter is to be operated with use of the capacitive sensor to measure the WLR the selector must
be set to WLR measurement mode. To enable the meter to calculate correct WLR it depends on a
separate set of wetgas calibration coefficients normally in place from factory. If the gas permittivity is
possible to measure with shut in of dry gas at operating pressure this will provide a good reference
point otherwise use the default value 1 and pressure factor. The oil/condensate permittivity is also
preferred to be found by the meter, but it is normally difficult to achieve this and an inline calibration
during flow is necessary. This is done by adjusting the permittivity for oil until waterfraction match a
fluid sample taken at the same time.
Flow calibration equations for gas and liquid mass flow rates are also available if needed. This is
normally not needed, but if a good reference system enables improvement of flow rates, a third order
equation is available. In Table 2 an explanation of the setup parameters is listed.
Page 51 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
Table 2 Wetgas Calculation Setup parameters
Note! The use of mass correction equations demands use of calculation tools and additional training.
Page 52 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
The capcal equation is made at factory and only needs to be verified against certificate. The wetgas
calibration is completed at factory if the meter is bought with wetgas mode. If this is a meter where the
wetgas functionality is added post-factory the coefficients for this meter has to be calculated and
entered at site. Document [4] contains calibration equation for different sized meters and the
corresponding has to be entered. Procedure [3] explains how to adjust this equation to compensate for
any offset between empty (gas filled) permittivity. Please note that in field the gas point for empty
sensor is set equal to air in the adjustment document [4]. So the procedure is not fully completed as
only the “air” point is recorded and used to adjust the default equation. It is important that the meter is
bypassed and contains steady gas at a given pressure and temperature. The gas reference permittivity
will also have to be updated according to the measured permittivity @P&T after the adjusted equation
is implemented into the meter. The Roxar MPFM 2600 should then measure and show 100% GVF,
even if no flow.
Page 53 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
12. UNITS
Bar bar
BBL barrels
BPD barrels per day
CFD cubic feet per day
CFM cubic feet per minute
CFT cubic feet
deg C degrees Celsius
deg F degrees Fahrenheit
g/ml grams per milliliter
Gal (US) U.S. gallons
GPM (US) U.S. gallons per minute
Inches H2O inches of water
kg kilograms
kg/d kilograms per day
kg/hr kilograms per hour
kg/m3 kilograms per cubic meter
kl kiloliters
kl/d kiloliters per day
km3 thousand cubic meters
km3/d thousand cubic meters per day
kPa kiloPascals
kT metric kilotons
kT/d metric kilotons per day
m3 cubic meters
3
m /d cubic meters per day
m3/hr cubic meters per hour
mbar millibar
MCFD thousand CFD
MCFT thousand CFT
MMCFD million CFD
MMCFT million CFT
psi pounds per square inch
Sunit unit at standard conditions
(a) absolute pressure (including atmospheric pressure)
(g) gauge pressure
Page 54 of 55
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA
Operating instructions
Document no. / rev.: ROX000091983 / AC
Roxar Multiphase meter 2600
13. NOMENCLATURE
VGAC Gas volume at AC (actual/meter conditions)
V AC
G Measured gas volumetric flow rate at AC (actual/meter conditions)
VOAC Oil volume at AC (actual/meter conditions)
V AC
O Measured oil volumetric flow rate at AC (actual/meter conditions)
V W
AC
Water volume at AC (actual/meter conditions)
V AC
W Measured water volumetric flow rate at AC (actual/meter conditions)
V G
SC
Gas volume from AC gas phase at SC (standard conditions)
V SC
G in O Gas volume from AC oil phase at SC (standard conditions)
V SC
G Measured gas volumetric flow rate at SC (standard conditions)
V O
SC
Oil volume from AC oil phase at SC (standard conditions)
VOSCin G Oil volume from AC gas phase at SC (standard conditions)
V SC
O Measured oil volumetric flow rate at SC (standard conditions)
V W
SC
Water volume at SC (standard conditions)
V SC
W Measured water volumetric flow rate at SC (standard conditions)
Page 55 of 55
Roxar Template doc. no. / rev.: 034708/D