Certificate of Validation_09May
Certificate of Validation_09May
Frigga certifies that the products mentioned on the following page(s) have been thoroughly tested and validated under
the observation of ISO9001:2008 certified quality assurance system and meet performance accuracy specifications
over the stated ranges.
Reference Instrument
Device Model Calibration Date Accuracy Serial Number
Thermometer FLUKE 51II Dec 21th,2021 +/-0.05°C 56110274WS
Note:All reference instruments used in validation are periodically calibrated by CNAS China National Accreditation
Service accredited laboratories, CNAS is full member of the International Laboratory Accreditation Cooperation,
so this validation certificate is also accepted by all signatories to the ILAC MRA.
Environment Conditions
Ambient Temperature: 20~25°C; Ambient Humidity:50~60%RH;
Validation Result
Validation points Result Tolerance Status
+60°C -0.1°C +/-1°C PASS
+40°C -0.4°C +/-0.5°C PASS
+25°C -0.4°C +/-0.5°C PASS
0°C -0.3°C +/-0.5°C PASS
-5°C -0.3°C +/-0.5°C PASS
-20°C -0.2°C +/-0.5°C PASS
Product Information
Product Name Serial Number Production Expiry
U2 UF52000128 02/2023 02/2025
This certificate was automatically generated and valid for 2(two) years from date of production
1/2
Texas Instruments Incorporated
12500 TI Boulevard
Dallas, Texas 75243
October 10th, 2019
The accuracy of temperature testing for the above mentioned products is verified with equipment that is
calibrated by an accredited lab that complies with ISO/IEC 17025 policies and procedures. Each device is
tested and trimmed to conform to its respective datasheet specification limits.
Per ISO 9001:2008 procedures and TI's quality assurance program, there is a scheduled and documented
calibration procedure for the equipment. These procedures provide for traceability of the measurement
results to a national standards laboratory such as the National Institute of Standards and Testing (NIST) or an
equivalent regional standards laboratory.
Sincerely,
Ivon Ayala
Customer Quality Engineering
2/2