1. Software installation and configuration
.
The SmartExplorer software allows you to configure a smart sensor and to collect real-time data from a sensor.
• Configuration makes it possible to configure a smart sensor ‘out of the box’ and test the smart sensor before the sensor is used
in an operational network.
• Collecting data makes it possible to store data from the smart sensor in a comma separated file. The comma separated file is
created at the beginning of every new day or at the beginning of the first day of the week.
• The SmartExplorer software can also be used to monitor and/or log up to 10 instruments simultaneously and works with all
Smart Radiometers (SMP, SHP, SGR, SUV)
• Install the SmartExplorer program from the supplied DVD or download the latest version from the relevant product page
under the tab download/software from our website
• Run the setup program from your local computer with administrator rights.
The installation program should download the .NET 4.5 frame work from the Microsoft Server, but if the .NET 4.5 framework is
not downloaded under Windows VISTA after running the setup program, please download the .NET 4.5 framework manually from
the Microsoft website.
The installation program will create a work directory in the program data directory:
C:\ProgramData\KippZonen\SmartExplorer
The program data directory on most computers is located on the C drive (please contact your network administrator for details).
The program data directory is a hidden directory. Open the file browser and select the menu: ‘organize map and search options’
to make this directory visible.
5
1.2 Running the SmartExplorer software
Start the application SmartExplorer.exe for Windows 7, Windows 8 and Windows VISTA.
The workspace file is located in the work directory C:\ProgramData\KippZonen\SmartExplorer. The application will look for this
workspace file.
The workspace file defines the default options, login-password, a list of the default devices and the default communication
parameters. Changes made by the user can be saved in the workspace file. The next time the application is started it will start
up with the last saved changes (see also Menu: File → Save Workspace).
After the first installation of the software and starting up the following menu is shown.
6
1.2.1 Program Start
• Select the option Service mode if you want to configure smart-sensors. Select the option View only if you want to use the
program as simple data-logger.
• Select the option Enable Auto Connect if the application should make a connection with sensors every time when the application
is started and select the option ‘Enable logging’ to start logging every time when the application is started. Note: the option
‘Enable logging’ is shown when the option ‘Enable Auto Connect’ is set.
• Select the option ‘Enable single instrument use’ if you want to configure sensors, change the Modbus® address and change the
communication parameters of the sensor.
Select the following options after the very first installation: Service Mode, Enable Auto Connect and Enable ‘single instrument use’.
Press ‘Update’. The changes are not applied until after a full restart. Press ‘Ok’ to make a full restart.
7
1.3 Selecting the communication parameters
Selecting the communication parameters is not possible when the Modbus® TCP/IP mode is enabled (see the description of the
workspace file for more details).
Press the button ‘Setup Connection’ and select the Modbus® Protocol (for example Serial RTU protocol).
Enter the Modbus® ID of the device and select the option ‘Enable’ to enable the communication with the sensor or press the
button ‘Set Defaults’ to enable the default sensors (the default sensors are defined in the workspace file). Press ‘Ok’ to continue.
After adding all sensors it is recommended to save the configuration in the workspace file.
8
The following menu is shown:
Press ‘Yes’ to make a full restart and make the changes effective, or ‘No’ to save the changes only.
Select the tab ‘Status’ to show the status of the sensor. Sensor errors are shown here.
Note After power up, the sensor will show a power fail error. This error should disappear after resetting.
9
Select tab page ‘Statistics’ to show an overview of all errors, the button ‘Update’ will refresh the information.
1.6 Chart
Select the tab ‘Chart’ to view the graph from the selected sensor.
Press the keys Control + → or Control + ← to increase or reduce the vertical scale.
10
1.7 Log file options
Select the tab ‘File Output’.
In the left column select the devices to enable the devices for logging.
11
Select the tab ‘Connections’.
12
Change Communication Parameters
Note In an operation network it is allowed to change the Modbus® address but it is not allowed to change the baud rate or
parity because this will have an impact on the communication to all other sensors.
Note It is not recommended to change the baud rate if the sensor is connected to a wireless network (wireless Modbus®,
Bluetooth, Bluegiga, Zigbee etc) because changing the baud rate of the sensor does not change the baud rate of the
wireless network.
Press ‘Next’.
Press ‘Next’.
13
The recommended options at power on are ‘Round’ and ‘Auto Range’ not enabled and ‘Tracking Filter’ and ‘Fast Response’
enabled. In some applications auto range and resolution can be set by the user in real time mode. To enable this feature enable
the remote controllable options ‘Auto Range’ and ‘Resolution’. However SmartExplorer software does not allow you to change
this in real time mode.
Press ‘Next’.
Press ‘Update’.
The selected sensor will restart with the new configuration. At the same time the baud rate of the computer (PC) is changed to
the new baud rate.
After restarting the selected device the ‘Communication Status’ field in the main menu must show ‘Status √’. If a boot error is shown,
disconnect the power for at least 10 seconds and reconnect the power again. Press the ‘Send Broadcast’ button to rediscover
the device.
Select in ‘Setup’ → ‘Program Start’ → ‘Service Mode’ → ‘Single Instrument Use’. Then press ‘Update’ and disconnect all sensors
except the faulty sensor.
Note Do not press this button in an operational network with 2 or more sensors.
If the sensor has the same parameter as the computer, the status line will show: ‘Communication parameters for device <Modbus® ID>
changed <baud rate, parity etc.>’
14
1.11 The sensor does not respond after a send broadcast
If the communication parameters (baud rate, parity etc.) of the computer and device are not the same, or the communication
parameters are not known, the sensor will not respond on a sent broadcast command. There is an escape sequence to start the
sensor with default communication parameters (19200 baud, 8 bits - even - 1 stopbits) and default Modbus® ID (address 1).
Transmitting a broadcast over a Modbus® TCP network is not possible, because the Modbus® gateway will block the special
broadcast address (0xFF).
Press ‘Setup Connection’ and select the default baud rate (19200), data bits, parity and stop bits (8, even, 1).
After this escape sequence the sensor will start with a temporary Modbus® ID 1, default baud rate (19200) default parity (even)
default data bits (8) and default stop bits (1). The sensor operates now in a special mode. The communication parameters in the
EEPROM of the sensor are not changed.
To return to the original configuration and leave the special mode after the ‘Start From Boot’ you must disconnect the power for
at least 10 seconds in order to restart the sensor with the original parameters stored in EEPROM.
1. Select the tab ‘Connections’ and select option ‘Single Instrument Use’
2. Press ‘Start From Boot’ and disconnect the power immediately after pressing the key
3. Disconnect the power of the sensor and wait a few seconds
4. Click ‘Ok’
5. Reconnect the power as soon as possible (at least within 10 seconds)
6. Wait until the sensor is ready, if not, repeat step 2 and 3
7. On the tab ‘Configuration’, press ‘Configure Device’ or press function key F3
8. Change the communication parameters and Modbus® ID
9. After a few seconds a reboot error is shown in the ‘Communication Status’ field
10. Disconnect the power of the sensor and wait at least 10 seconds, reconnect the power and press ‘Send Broadcast’
Always disconnect the power at least 10 seconds after you change the configuration and start from boot.
15
Our customer support remains at your disposal for any maintenance or repair, calibration,
supplies and spares.
Für Servicearbeiten und Kalibrierung, Verbrauchsmaterial und Ersatzteile steht Ihnen unsere
Customer Support Abteilung zur Verfügung.
Notre service 'Support Clientèle' reste à votre entière disposition pour tout problème de
maintenance, réparation ou d'étalonnage ainsi que pour les accessoires et pièces de rechange.
HEAD OFFICE
Kipp & Zonen B.V.
Delftechpark 36, 2628 XH Delft +31 15 2755 210
P.O. Box 507, 2600 AM Delft [email protected]
The Netherlands www.kippzonen.com
SALES OFFICES
Kipp & Zonen France S.A.R.L.
88 Avenue de l’Europe T: +33 (0) 1 64 02 50 28
77184 Emerainville F: +33 (0) 1 64 02 50 29
France [email protected]
Go to www.kippzonen.com for your local distributor or contact your local sales office