V8

Unifying intelligence and performance

Quick Manual (Based on V 1.03)

Produced by Global Clinical Group / SAMSUNG MEDISON CO.,LTD

※ This Quick Manual does not include all of the details of instruction, For more detail, please refer to V8 User Manual.
※ Do not distribute this document to customers unless relevant regulatory and legal affairs officers approve such distribution. SAMSUNG MEDISON CO.,LTD.
V8 Quick Manual

Power Mode
Turn on/off the system Enter 2D, C, PD, PW, M mode
and adjust Gain

Angle
Depth / Focus
Adjust the angle of SV on
Adjust depth and focus on
PW/CW mode, the probe
scanning mode
cursor of Body Marker and the
angle of the Indicator
Zoom
Rotate: Magnify whole image
Text (+Indicator) area
Insert Annotation or Push: Magnify an ROI image
Indicator(Setup)

Calculator
Clear
Enter the measurement of each
Delete Indicator, Measure, application
Text, Body Marker.
*Long press is all clear Q Scan
Adjust image optimization
Body Marker automatically
Body Marker on/off
Caliper
Pointer
Measure Distance, Area, Volume
Activate an arrow for image and Velocity
review
L & R (Dual)
Menu Split the screen (Left & Right)
Preset/Probe change or
Enter the EZ-Exam+ P1, P2, P3
User defined function keys
U1, U2
(e.g. Save+DICOM, Print+Save,
User defined function keys
(i.e. CW, TDI etc.) Clip Store etc.)
Change key
Change the trackball function that depends
Set or Context Freeze
on current mode (i.e. Color box, SV size etc.)
Select/Deselect items or * When the Text is activated, it has indicator Pause/Resume scan
Provide shortcuts around
function.
trackball (Position is possible
to be changed)
Produced by Clinical Training Center/ SAMSUNG MEDISON CO.,LTD. I 02
V8 Quick Manual

Worklist Search Registration of Patient


■ Tap the [Patient] button on the touch screen ■ Tap the [Patient] button on the touch screen
* If you have already linked worklist server then it will be moved [Worklist] automatically.

Click the [Search] button: Patient list will be updated ❶ Insert ID Input the Patient ID, Name, Birth, Age, etc.
❶ Search (Available to search Patient ID, Name, Procedure ID, Exam Date,
If you need to add the patient’s further information, e.g. Input
Accession number)
❷ Study the OB Information.
❷ ID Select Select patient ID which you are going to start the study. Information ① Enter the OB Application tab.
② Enter the GA, LMP, EDD, Number of fetuses etc.
❸ Start Exam Click [Start Exam] to enter the scan mode.
Select the application and Exam type then start, chosen
❸ Start EzExam+
EzExam+™ protocol will be started.

Click the Patient ID → Click the Start Exam to enter the scan mode. Input the operator information.
❹ Operator
Double click the Patient ID → Page will be changed into Patient tab and check or (Diagnosis / Reference Physician / Sonographer)

input patient/study information. Then click the Start Exam to start scan. ❺ Start Exam Click [Start] or [Freeze] button to start examination.

Produced by Clinical Training Center / SAMSUNG MEDISON CO.,LTD. I 03


V8 Quick Manual

Probe Selection Quick Preset Setup

❶ Probe Tap the [Probe] button on the touch screen. Available to save the preset up to 6 which is frequently used.
It will be displayed on the right side of the touch screen.
❷ Probe List Display probe lists which are connected by system.
❻ Quick Preset
Tap the [Quick Preset Setup] button to customize
❸ Default Preset Select if you want to use the factory (default) presets. Set up
Tap the preset want to use in Default or User tab.
Tap the location where you want to save onto the right.
❹ User Preset Select if you want to use user preset.
To finish the set up, tap the Quick Preset Setup button once
Available to revise the order of application. again.
❺ Edit App. Order
(It is available into the Default.)

Select the application that you want to move.


Adjust the order by clicking the up/down arrow.
Click [OK] button to save.
→ The changed application order will be
displayed.

Produced by Clinical Training Center / SAMSUNG MEDISON CO.,LTD. I 04


V8 Quick Manual

Touch Screen - Scan Mode 1

■ Digital TGC

Change the TGC line by drawing with your finger.


Initiate the TGC line. It will be all aligned at center.
TGC more menu
* How to save the TGC Line)
Change the TGC line → Save → Choose the save location out of
the five at the bottom.

■ Lateral TGC
❶ Patient Select a patient ID from worklist or enter a new patient Information manually.

❷ Probe Display the Probe Selection screen to select or change the probe and application.

❸ Report Show the measurement results of the current application and other information.

❹ SonoView Image management program. (Review, Storage, Delete, Export and Backup)

❺ End Exam Finish the exam of the currently selected patient and resets the related data.

Go to Utility page to activate as fallowing as a Demo Play, Setup, Touch Customize


❻ Utility
and etc.
Swipe a lateral TGC line by drawing your finger.
Save Changed lateral TGC line out of the 5 empty panel on the right.
❼ Quick Preset Available to change the probe and preset at once.
( Change line → Save as preset → choose the location )
Initiate the lateral TGC by clicking the blue box on the top left corner.
❽ Next page To go next page of flexible menu, swipe bottom of the touch screen.

Produced by Clinical Training Center / SAMSUNG MEDISON CO.,LTD. I 05


V8 Quick Manual

Touch Screen - Scan Mode 2 Touch Screen - Touch Customize

❶ Harmonic Adjust contrast resolution of image. Select mode between 2D, C, PW, PD, M, and More.
❶ Mode Selection
Each [General] and [Flexible] menus are customized.
Provide clearer image from near to far zone using wide-band
❷ S-Harmonic Touch Customize menu can be separately arranged in [Live] and
frequency. ❷ Live / Frozen
[Frozen]image mode.
Flip up or down & left or right the image.
❸ Flip / Rotate
Rotate clockwise or counterclockwise. ❸ General [General] menu includes main functional image items.

❹ Dual Live Display 2D image and Color Doppler image simultaneously. [Flexible] menu represents soft menu items that are available in
❹ Flexible
the current input mode as shown on the bottom of screen.
Utilize high and low-frequency of the ultrasound to optimize
❺ ShadowHDR
attenuated shadow areas. ❺ Save Tap to save after customization.

Enhance the spatial resolution which improves sharpness of


❻ HQ-Vision
structure.

❼ MultiVision The spatial compound image by using multi-beam.


How to customize touch screen menu
Remove noise and enhance outlines to make the image more Tap the Touch Edit [ ] on the left side of touch screen.
❽ ClearVision
clearly.

Produced by Clinical Training Center / SAMSUNG MEDISON CO.,LTD. I 06


V8 Quick Manual

Color/PD/S-Flow PW/CW Doppler

High sensitivity bi-directional Power Doppler mode for tiny View real-time 2D and Spectral Doppler image at the same
❶ S-Flow ❶ Simultaneous
vessels. time.

❷ Auto Gain Adjust proper Color Doppler gain automatically frame by frame. ❷ Doppler Invert The spectrum inverts position of + and -.

❸ Low/ Mid/ High Automatically adjust the appropriate Color Doppler scale. ❸ Dual Layout Select display format as Top-bottom or Side by side.

❹ Steer Invert Invert the direction of Color Doppler box. Automatically calculate the values of Doppler.
❹ Auto Calc.

❺ Scale Adjust the Color PRF. (Pulse Repetition Frequency) ❺ Steer Invert Invert the angle of the sample volume.

❻ Steer Adjust the angle of Color Doppler box. (-30 ̊~ 30 ̊) Select a SV size. (0.5 ~ 25mm)
❻ SV Size

Adjust the PW/CW PRF.


Smart Auto Doppler ❼ Scale
(Adjust the velocity depending on vascular velocity.)
Press [Q Scan] on the control panel in the Doppler mode.
- Control the location of ROI and sample volume, direction of steer angle by one touch.

- Improve the complex UI workflow of Doppler mode.

* Only available in vascular application of linear probe


Produced by Clinical Training Center / SAMSUNG MEDISON CO.,LTD. I 07
V8 Quick Manual

Measurement - Calculator Measurement - Caliper


■ Calculator: Data measurement per each application. ■ Caliper: Measure (by using a trackball and [Set]) selected items.
(e.g., OB, Cardiac, Vascular, Prostate, etc.) (e.g., Distance, Area, Volume, Doppler measure, etc.)

Measure the
Volume of
an object

Provide calculate package of selected probes/ applications. Distance Measure the straight distance between 2 points.
❶ Change App.
Can be changed when measuring other application items. ❶ Distance Open Spline Measure traced curved-line.
Trace Length Use the trackball to draw the desired curve.
Select Right , Left , or N/A as the location of your measurement
❷ Laterality target. This is displayed on the touchscreen only for specific
Ellipse Measure the circumference area.
packages.
❷ Area Trace & Measure the area by tracing and drawing.
❸ Location Specify the location of the measurement target. Closed Spline Measure the area with a connected point.

❹ Trace Direction Select the trace direction of the Doppler spectrum.


❸ %Stenosis(D) Diameter & Area of a vessel are measured and
❺ Sensitivity Adjusting the Sensitivity facilitates Doppler spectrum analysis. %Stenosis(A) the stenosis ratio is calculated.

3 Points Measure the angle using 3 points.


❹ Angle
2 Lines Measure the angle between 2 lines.

Produced by Clinical Training Center / SAMSUNG MEDISON CO.,LTD. I 08


V8 Quick Manual

Image Save Cine Save - Freeze Mode


■ Save Still Image, Volume data and Cine. ■ Save a video interval of Unfreeze and Freeze mode up to maximum frame.

Various functions can be saved such as Save, Cine, B/W Print, Tap the [Cine Save] to save the specified image frames by tapping it.
❶ Cine Save
P1, P2, P3, U1, U2 Color Print, Send to DICOM, Biopsy, TDI and etc. (Only available on Freeze mode)
[Peripheral Key] * Possible to set the functions: Utility → Setup → Customize →
User Key ❷ Trim Start Set the first frame of the cine images to be saved.
Trim End Set the last frame of the cine images to be saved.
Still Image, Volume data and Store Clip can be saved on to using
Still or
[Peripheral Key]. Set Cine play speed.
Cine Save, ❸ Cine Speed
* To check the assigned key, refer to key guide on the bottom (It can be set from 6% ~ 300%)
Volume Image
of the monitor.

Produced by Clinical Training Center / SAMSUNG MEDISON CO.,LTD. I 09


V8 Quick Manual

SonoView - Image Review, Export & Backup


■ Tap the [SonoView] on the touch screen. ■ The method to manually transfer images

❹ Export

Select Drive. (USB - F:\)


Directories: Select the location in which the exam will be saved.
Enter the file name
Select file and video format. (BMP, JPEG, TIFF, DICOM, AVI, MP4)
∙ Create Patient Folder(s): Categorized files per date and patient ID Click
∙ Hide Patient Information: To remove patient ID and name.
Select the [OK].

❶ Select Exam Check the combo box ( ) in front of the patient ID in the [Exam List].

·Review Exam: The exams that were performed prior to the last 24
❷ Review or hours.
Continue Exam ·Continue Exam: The exams that were performed within the last 24
hours.

Transfer the saved images to PACS server.


❸ Send (Press [Send] → Select [DICOM Destination] → Press [Transfer]
* Transfers all images on selected exams.

Convert image to PC compatible format such as BMP, JPEG, DICOM, TIFF,


❹ Export
AVI and MP4.

Only the back up data can be reviewed on the system.


❺ Backup
Press [Backup] → Select [Drive] → USB(F: \)

❻ Delete Delete the saved exams.


Produced by Clinical Training Center / SAMSUNG MEDISON CO.,LTD. I 10
V8 Quick Manual

SonoView - Image Management


■ Tap the [Patient] on the touch screen.

Select the exam to change and enter the [Edit Patient]


❸ Edit Patient
* Only available patient who has been tested within 24 hours.

Check the patient information to edit. Then Create patient information manually.
Select the Local(HDD) or Worklist
∙ Local Storage: Change patient information to saved exam in the HDD.
∙ Worklist: Change patient information to information in the Worklist.
Hide patient information on the image after change patient information.
DICOM send option can be selected after change patient information.
Press [Change].

Search saved exams by Patient ID, Name, Date of Birth, Exam Date
❶ Search
(Today, Last week, Last Month, Manual)

Select exam that you want the Review, EzCompare, Export, Send or
❷ Exam List
Backup

❸ Edit Patient Edit patient information.

Produced by Clinical Training Center / SAMSUNG MEDISON CO.,LTD. I 11