PARAMETER SPECIFICATION
Main Body High-strength engineering plastic rack, light, beautiful and corrosion
resistant
Scope of Application Adult, child
Gas Source O2: 0.28--0.6MP
Flow Meter O2: 0.05--1.0L/min; 1.1--10L/min
Flow Rate of Rapid Supply 25--75L/min
Low Oxygen Pressure Alarm It will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer It has the function of automatic compensation based on pressure,
temperature, and flow rate. The regulation range of evaporator
concentration is 0-5vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be
chosen for application as required by the customer.
Respiratory Circuit Working mode: all-close, semi-close, semi-open
APL: ≤12.5kPa.
Respiratory Bellows Bellows for adults, bellows for children, tidal volume range:0-1500ml
VENTILATOR
Display Mode High-definition 5.7” LCD screen display
Ventilation Mode IPPV, SIPPV, SIMV, MANUAL
Ventilation Function PEEP, SIGH, IRV
Ventilation Parameters Tidal volume 50~1500ml
Rate 2~99bpm
SIMV rate 2~20bpm
1
I:E 2:1~1:8
SIGH 1.5 times the inspiratory time/100bpm
Inspiratory trigger pressure -10~20cmH 2O
PEEP 2~20cmH 2 O
Pressure limitation 10~60cmH 2 O
Parameters for Ventilation Tidal volume, Minute ventilation volume, IPPV rate, SIMV rate, total
Monitoring respiratory rate, I/E, peak pressure of airway-pressure-time
waveform, flow rate-time waveform, PEEP, inspiratory trigger
pressure, (average pressure, oxygen concentration, battery capacity).
SECURITY ALARM SYSTEM
Oxygen Concentration Alarm Upper limit setting range (21%~99%)
Low limit setting range (18%~80%)
Airway Pressure Alarm Upper limit setting range 55~60cmH 2O
Low limit setting range 0~50cmH 2O
Minute Ventilation Volume Upper limit setting range 1~20L/min
Alarm Low limit setting range 0~19L/min
Tidal volume Upper limit setting range 100~1500ml
Low limit setting range 0~1400ml
Sustained High-pressure It will give alarm when stress have consistently been higher than
Alarm 25cmH 2O
(Suffocation Alarm, Power alarm, Intubation off alarm, battery voltage alarm)
Power AC 220V, 50Hz, (UPS, Storage battery)
(Mechanical Arms)
(End-tidal CO2 Monitoring)
(Specific interface of waste anesthesia gas)
*Note: Items within brackets are optional
2
AR-322
Anesthesia Machine
Technical Specifications
PARAMETER SPECIFICATION
Main Body High-strength engineering plastic rack, light, beautiful and corrosion
resistant
Scope of Application Adult, child
Gas Source O2: 0.28--0.6Mpa;
N2O: 0.28--0.6MPa.
Flow Meter O2: 0.05--10L/min;
N2O: 0.05--10L/min
O2, N2O Linkage and N2O When using nitrous oxide, oxygen concentration>25%;
Stopper When the oxygen pressure < 0.2MPa, the flow of nitrous oxide would
be cut off.
Flow Rate of Rapid Oxygen 25--75L/min
Supply
Low Oxygen Pressure Alarm There will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer It has the function of automatic compensation based on pressure,
temperature, and flow rate. The regulation range of evaporator
concentration is 0-5vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be
chosen for application as required by the customer. And vaporizers
that are imported with original packaging are also available.
Respiratory Circuit Working mode: all-close, semi-close, semi-open;
APL: ≤12.5kPa.
Respiratory Bellows Bellows for adults, bellows for children, tidal volume range 0-1500ml
3
VENTILATOR
Display Mode High-definition 5.7” LCD screen display
Ventilation Mode IPPV, SIPPV, SIMV, MANUAL
Ventilation Function PEEP, SIGH, IRV
Ventilation Parameters Tidal volume 50~1500ml
Rate 2~99bpm
SIMV rate 2~20bpm
I:E 2:1~1:8
SIGH 1.5 times the inspiratory time/100bpm
Inspiratory trigger pressure -10~20cmH 2O
PEEP 2~60cmH 2O
Pressure limitation 10~60cmH 2O
Parameters for Ventilation Tidal volume, Minute ventilation volume, IPPV rate, SIMV rate, total
Monitoring respiratory rate, I/E, peak pressure of airway, pressure-time waveform,
flow rate–time waveform, PEEP, inspiratory trigger pressure, (average
pressure, oxygen concentration, battery capacity).
SECURITY ALARM SYSTEM
Oxygen Concentration Upper limit setting range (21%--99%)
alarm Low limit setting range (18%--80%)
Airway pressure alarm Upper limit setting range 5--60cmH 2 O;
Low limit setting range 0--50cmH 2 O.
Minute ventilation volume Upper limit setting range 1--20L/min;
alarm Low limit setting range 0--19L/min.
Tidal Volume Upper limit setting range 100--1500ml
Low limit setting range 0--1400ml
Sustained high-pressure It will give alarm when stress have consistently been higher than
alarm 25 cmH 2 O
Suffocation alarm It will give sound and light alarm if there is no tidal volume input for 15
seconds
(Suffocation Alarm, Power alarm, Intubation off alarm, battery voltage alarm)
Power AC 220V, 50Hz, (Storage battery, UPS)
Mechanical Arms
(End-expiratory CO2 monitoring)
(Specific interface of waste anesthesia gas)
*Note: Items within brackets are optional.
4
AR-323
Anesthesia Machine
Technical Specifications
PARAMETER SPECIFICATION
Main Body High-strength engineering plastic rack, light, beautiful and corrosion
resistant
Scope of Application Adult, child
Gas Source O2: 0.28--0.6MPa;
N2O: 0.28--0.6MPa
Flow Meter O2: 0.05--1.0L/min; 1.1--10L/min
N20: 0.05--1.0L/min; 1.1--10L/min
O2, N2O Linkage and N2O When using nitrous oxide, oxygen concentration>25%;
Stopper When the oxygen pressure < 0.2 MPa, the flow of nitrous oxide would
be cut off.
Rapid Oxygen Flow 25--75L/min
Low Oxygen Pressure Alarm There will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer It has the function of automatic compensation based on pressure,
temperature, and flow rate. The regulation range of evaporator
concentration is 0-5vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be
chosen for application as required by the customer. And vaporizers
that are imported with original packaging are also available.
Respiratory Circuit Working mode: all-close, semi-close, semi-open;
APL: ≤12.5kPa.
Respiratory Bellows Bellows for adults, tidal volume range: 0--1500ml;
bellows for children, tidal volume range: 0--300ml.
5
Anyone can be chosen for application as required by the customer
VENTILATOR
Display Mode High-definition 10.4” TFT color LCD Screen Display
Ventilation Mode VCV, PCV, SIMV, MANUAL
Ventilation Function PEEP, SIGH, IP, IRV, Standby
Ventilation Parameters Tidal volume 20~1600ml
Rate 2~99bpm
SIMV rate 2~20bpm
I:E 4:1--1:8
Inspiratory trigger pressure -10~20cmH 2O
PEEP 2~20cmH 2O
Pressure limitation 5~60cmH 2 O
SIGH 1.5 times the inspiratory time, 60-120 adjustable
Inspiratory platform 0~50%
Pressure control 3~50cmH 2 O
Pressure slope 1~10 gear
Parameters for Ventilation Expired Tidal volume, Inhale tidal volume, Minute ventilation volume,
Monitoring IPPV rate, SIMV rate, total respiratory rate, autonomous breathing
rate, I/E, peak pressure of airway, average pressure, PEEP,
inspiratory trigger pressure, Inspiratory platform, oxygen
concentration, battery capacity, pressure-time waveform, flow-rate–
time waveform, flow-volume loop, pressure-volume loop
SECURITY ALARM SYSTEM
Oxygen Concentration alarm Upper limit setting range 21%--99%;
Low limit setting range 18%--80%.
Airway Pressure Alarm Upper limit setting range 5--60cmH 2 O;
Low limit setting range 0--50cmH 2 O.
Minute Ventilation Volume Upper limit setting range 1--20L/min;
Alarm Low limit setting range 0--19L/min.
Respiratory Rate Upper limit setting range 10--99bPm
Low limit setting range 0--50bPm
Tidal Volume Upper limit setting range 10--1500ml
Low limit setting range 0--1400ml
Sustained High-pressure It will give alarm when stress have consistently been higher than
Alarm 25 cmH 2 O
Suffocation Alarm Power alarm, Battery voltage alarm, Intubation off alarm
Power AC 220V, 50Hz, Storage battery, UPS
Manipulators
(End-expiratory CO2 monitoring)
(Specific interface of waste anesthesia gas)
*Note: Items within brackets are optional.
6
AR-324
Anesthesia Machine
Technical Specifications
PARAMETER SPECIFICATION
Main Body High-strength engineering plastic rack, light, beautiful and corrosion
resistant
Scope of Application Adult, child
Gas Source O2: 0.28--0.6Mpa;
N2O: 0.28--0.6MPa.
Rapid Oxygen Flow O2: 0.05--1.0L/min; 1.1--10L/min
N20: 0.05--1.0L/min; 1.1--10L/min
O2, N2O Linkage and N2O When using nitrous oxide, oxygen concentration>25%;
Stopper When the oxygen pressure < 0.2KPa, the flow of nitrous oxide would
be cut off.
Flow Rate of Rapid Oxygen 25--75L/min
Supply
Low Oxygen Pressure Alarm There will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer It has the function of automatic compensation based on pressure,
temperature, and flow rate. The regulation range of evaporator
concentration is 0-5vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, two can be
chosen for application as required by the customer. And vaporizers
that are imported with original packaging are also available.
Respiratory Circuit Working mode: all-close, semi-close, semi-open;
7
APL: ≤12.5kPa.
Respiratory Bellows Bellows for adults, bellows for children, tidal volume range: 0-1500 ml.
VENTILATOR
Display Mode High-definition 5.7” LCD screen display
Ventilation Mode IPPV, SIPPV, SIMV, MANUAL
Ventilation Function PEEP, SIGH, IRV
Ventilation Parameters Tidal volume 50~1500ml
Rate 2~99bpm
SIMV rate 2~20bpm
I:E 2:1~1:8
Inspiratory trigger pressure -10~20cmH 2O
PEEP 2~20cmH 2O
Pressure range 10~60cmH 2 O
SIGH 1.5 times the inspiratory time/100bpm
Parameters for Ventilation Tidal volume, Minute ventilation volume, IPPV rate, SIMV rate, total
Monitoring respiratory rate, I/E, peak pressure of airway, pressure-time
waveform, flow rate–time waveform, PEEP, inspiratory trigger
pressure, average pressure, oxygen concentration, battery capacity.
SECURITY ALARM SYSTEM
Oxygen Concentration Alarm Upper limit setting range (21%--99%);
Low limit setting range (18%--80%)
Airway Pressure Alarm Upper limit setting range 5--60cmH 2 O;
Low limit setting range 0--50cmH 2 O.
Minute Ventilation Volume Upper limit setting range 1--20L/min;
Alarm Low limit setting range 0--19L/min.
Tidal Volume Upper limit setting range 100--1500ml
Low limit setting range 0--1400ml
Sustained High-pressure It will give alarm when stress have consistently been higher than
Alarm 25 cmH 2 O
Suffocation Alarm Power alarm, Intubation off alarm, Battery voltage alarm
Power AC 220V, 50Hz, Storage battery, UPS
Mechanical Arms
(End-expiratory CO2 monitoring)
(Specific interface of waste anesthesia gas)
*Note: Items within brackets are optional.
8
AR-325
Anesthesia Machine
Technical Specifications
PARAMETER SPECIFICATION
Main Body High-strength engineering plastic rack, light, beautiful and corrosion
resistant
Scope of Application Adult, child
Gas Source O2, N2O, Air: 0.28--0.6MPa.
Flow Meter O2: 0.05--1.0L/min; 1.1--10L/min
N20: 0.05--1.0L/min; 1.1--10L/min
Air: 0.05--10L/min
O2, N2O Linkage and N2O When using nitrous oxide, oxygen concentration>25%;
Stopper When the oxygen pressure < 0.2MPa, the flow of nitrous oxide would be
cut off.
Rapid Oxygen Flow 25--75L/min
Low Oxygen Pressure There will be sound alarm when the oxygen pressure < 0.2MPa
Alarm
Vaporizer It has the function of automatic compensation based on pressure,
temperature, and flow rate. The regulation range of evaporator
concentration is 0-5vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, two can be
chosen for application as required by the customer. And vaporizers that
are imported with original packaging are also available.
Respiratory Circuit Working mode: all-close, semi-close, semi-open;
APL: ≤12.5kPa
Respiratory Bellows Bellows for adults, tidal volume range:0--1500ml;
9
Bellows for children, tidal volume range:0-300ml.
Anyone can be chosen for application as required by the customer
VENTILATOR
Display Mode High-definition 10.4” TFT color LCD screen display
Ventilation Mode VCV, PCV, SIMV, MANUAL
Ventilation Function PEEP, SIGH, IP, IRV, standby
Ventilation Parameters Tidal volume 20~1600ml
Rate 2~99bpm
SIMV rate 2~20bpm
I:E 4:1~1:8
Inspiratory trigger pressure -10~20cmH 2 O
PEEP 2~20cmH 2O
Pressure limitation 5~60cmH 2 O
SIGH 1.5 times the inspiratory time, 60~120 adjustable
Inspiratory platform 0~50%
Pressure control 3~50cmH 2 O
Pressure slope 1~10 gear
Parameters for Ventilation Expired tidal volume, Inhale tidal volume, Minute ventilation volume,
Monitoring IPPV rate, SIMV rate, total respiratory rate, autonomous breathing rate,
I/E, peak pressure of airway, average pressure, PEEP, inspiratory trigger
pressure, Inspiratory platform, oxygen concentration, battery capacity,
pressure-time waveform, flow-rate-time waveform, flow-volume loop,
pressure-volume loop
SECURITY ALARM SYSTEM
Oxygen Concentration Upper limit setting range 21%--99%;
Alarm Low limit setting range 18%--80%
Airway Pressure Alarm Upper limit setting range 5--60cmH 2 O;
Low limit setting range 0--50cmH 2 O.
Minute Ventilation Volume Upper limit setting range 1--20L/min;
Alarm Low limit setting range 0--19L/min.
Respiratory Rate Upper limit setting range 10--99bPm
Low limit setting range 0--50bPm
Tidal Voulme Upper limit setting range 10--1500ml
Low limit setting range 0--1400ml
Sustained High-pressure It will give alarm when stress have consistently been higher than
Alarm 25 cmH 2 O
Suffocation Alarm Power alarm, Battery voltage alarm, Intubation off alarm
Power AC 220V, 50Hz, UPS, Storage battery
Manipulators
(End-expiratory CO2 monitoring)
(Specific interface of waste anesthesia gas)
*Note: Items within brackets are optional.
10
AR-325B
Anesthesia Machine
(Including M-9000E Patient Monitor)
Technical Specifications
PARAMETER SPECIFICATION
Main Body High-strength engineering plastic rack, light, beautiful and corrosion
resistant
Scope of Application Adult, child
Gas Source O2, N2O, Air: 0.28--0.6MPa.
Flow Meter O2: 0.05--1.0L/min; 1.1--10L/min
N20: 0.05--1.0L/min; 1.1--10L/min
Air: 0.05--10L/min
O2, N2O Linkage and N2O When using nitrous oxide, oxygen concentration>25%;
Stopper When the oxygen pressure < 0.2MPa, the flow of nitrous oxide would be
cut off.
Rapid Oxygen Flow 25--75L/min
Low Oxygen Pressure There will be sound alarm when the oxygen pressure < 0.2MPa
Alarm
Vaporizer It has the function of automatic compensation based on pressure,
temperature, and flow rate. The regulation range of evaporator
concentration is 0-5vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, two can be
chosen for application as required by the customer. And vaporizers that
are imported with original packaging are also available.
Respiratory Circuit Working mode: all-close, semi-close, semi-open;
APL: ≤12.5kPa
Respiratory Bellows Bellows for adults, tidal volume range:0--1500ml;
Bellows for children, tidal volume range:0-300ml.
Anyone can be chosen for application as required by the customer
VENTILATOR
11
Display Mode High-definition 10.4” TFT color LCD screen display
Ventilation Mode VCV, PCV, SIMV, MANUAL
Ventilation Function PEEP, SIGH, IP, IRV, standby
Ventilation Parameters Tidal volume 20~1600ml
Rate 2~99bpm
SIMV rate 2~20bpm
I:E 4:1~1:8
Inspiratory trigger pressure -10~20cmH 2 O
PEEP 2~20cmH 2O
Pressure limitation 5~60cmH 2 O
SIGH 1.5 times the inspiratory time, 60~120 adjustable
Inspiratory platform 0~50%
Pressure control 3~50cmH 2 O
Pressure slope 1~10 gear
Parameters for Ventilation Expired tidal volume, Inhale tidal volume, Minute ventilation volume,
Monitoring IPPV rate, SIMV rate, total respiratory rate, autonomous breathing rate,
I/E, peak pressure of airway, average pressure, PEEP, inspiratory trigger
pressure, Inspiratory platform, oxygen concentration, battery capacity,
pressure-time waveform, flow-rate-time waveform, flow-volume loop,
pressure-volume loop
SECURITY ALARM SYSTEM
Oxygen Concentration Upper limit setting range 21%--99%;
Alarm Low limit setting range 18%--80%
Airway Pressure Alarm Upper limit setting range 5--60cmH 2 O;
Low limit setting range 0--50cmH 2 O.
Minute Ventilation Volume Upper limit setting range 1--20L/min;
Alarm Low limit setting range 0--19L/min.
Respiratory Rate Upper limit setting range 10--99bPm
Low limit setting range 0--50bPm
Tidal Voulme Upper limit setting range 10--1500ml
Low limit setting range 0--1400ml
Sustained High-pressure It will give alarm when stress have consistently been higher than
Alarm 25 cmH 2 O
Suffocation Alarm Power alarm, Battery voltage alarm, Intubation off alarm
Power AC 220V, 50Hz, UPS, Storage battery
Manipulators
(End-expiratory CO2 monitoring)
(Specific interface of waste anesthesia gas)
*Note: Items within brackets are optional.
12
AR-N3
Anesthesia Machine
(Adult, Pediatric, Neonate)
Features
-8” color TFT screen, optional touchscreen
-User friendly design interface
-Tidal volume 5~1500 ml
-Up to 6 flow tubes for 3 gases
-PPSU/Metallic material of breathing circuit
-By-pass Tech for soda lime canister
-All ventilation modes to satisfy neonate, pediatric and adult patients
-Integrated breathing circuit with built-in heater
-Highly visible ACGO
Technical Specifications
PARAMETER SPECIFICATION
Screen 8-inch TFT LCD screen, 800x480 pixels
Gas supply O2, N2O, AIR
Selectatec bar Dual positions
Mechanical Flowmeter 6 flow tubes for 3 gases
13
Waveforms Up to 3 waveforms, user configurable
Ventilator Software V-CMV, P-CMV, PSV, V-SIMV, P-SIMV, PRVC, Standby,
Manual/Spont
Spirometry loop P-V, P-F, F-V, reference Loop
Auxiliary power outlets 3
Wheels Four wheels(125mm) with 2 separate brakes
Drawers 2
Reading lamp LED lighting Included
Vaporizer Standard one Chinese brand vaporizer;
Optional extra vaporizers
O2 cell Standard
Built-in Heater Standard
By-pass Standard
ACGO Standard
Li-ion Battery Standard 1 battery, 4800mAh;
Optional 2 batteries, 9600mAh
Spare Cylinder yoke Optional (O2, N2O)
AGSS Optional
Gas monitor module Optional (CO2, AG)
Patient monitor Optional
Suction Device Optional
Standard: Main unit, one Chinese brand vaporizer, O2 cell, Built-in Li-on battery, Disposable adult
anesthesia breathing circuit, O2/N2O/Air gas pipelines, Hospital O2/N2O/Air joints
Options: Extra vaporizer (Chinese/Drager/Penlon/Northern), Touch screen, Patient monitor, ETCO2,
Suction module, Pediatric/Neonate anesthesia breathing circuit, Spare cylinder yoke (O2, N2O),
AGSS, Gas monitor module (CO2, AG), etc.
14
AR-902P Portable
Anesthesia Machine
Introduction
This is manual, light-weight, compact, portable Anesthesia Machine specifically designed
for surgical locations where space or access is limited, such as Office Based Anesthesia
(OBA), Dental anesthesia, Missionary off-shore anesthesia, Remote locations in hospitals
outside O.R., Disaster scenes and mass casualty incidents, Front-line military hospitals,
Shipboard, aircraft, etc., where size, mobility and portability are important.
Technical Parameter
1. Working mode: Manual
2. Patient scope: from Pediatric to Adult (3kgs over)
3. Tidal volume: 20-1500ml
4. Working pressure: O2:0.3—0.4Mpa, N2O:0.3—0.4Mpa
5. Flowmeters: 2 tube 2 gas flowmeter; O2 range: 0.1-10L/min, N2O range: 0.1-10L/min.
6. O2 and N2O linkage device (ORC system), FiO2 ≥25%
7. APL valve adjustment range: 0.5-6.0Kpa
8. O2 Flush: 30-60L/min
Standard Configuration
- Main Unit
- One Vaporizer; Agent type: Halothane, Isoflurane, Enflurane and Sevoflurane.
- Standard Accessories of breathing tubes for connecting and Mask etc.
- In-built Co2 absorber Canister with capacity of 2kg.
15
Two Optional Anesthesia Ventilators
16