GEMDose Imaging Detectors for Radiotherapy Based on Gas Electron Multipliers
GEMDose Imaging Detectors for Radiotherapy Based on Gas Electron Multipliers
Klyachko
Nuclear et al. / Nuclear
Instruments Instruments
and Methods andinMethods
Physicsin Physics
ResearchResearch
A 628 A 628 (2011)434–439
(2011) 434–439 434
Available online 8 July 2010 New techniques in charged particle therapy and widespread use of modern dynamic beam delivery systems demand new beam
monitoring devices as well as accurate two-dimensional dosimetry systems to verify the delivered dose distribution. We are
Keywords:
GEM detector
developing dose imaging detectors based on gas electron multipliers (GEM) with the goal of improving dose measurement
Hadron therapy linearity, position and timing resolution, and to ultimately allow pre-treatment verification of dose distributions and dose
Dose imaging delivery monitoring employing scanning beam technology. A prototype 10 10 cm 2 double-GEM detector has been tested in the
205 MeV proton beam using electronic and optical readout modes. Preliminary results with electronic cross-strip readout
demonstrate fast response and single-pixel (4 mm) position resolution. In optical readout mode, the line spread function of the
detector was found to have s¼ 0:7 mm. In both readout modes, the detector response was linear up to dose rates of 50 Gy/min,
with adequate representation of the Bragg peak in depth-dose profile measurements.
& 2010 Elsevier B.V. All rights reserved.
articleinfo abstract
$
Funded by: NIH.
Corresponding author.
E-mail address: [email protected] (A.V. Klyachko).
0168-9002/$ - see front matter & 2010 Elsevier B.V. All rights reserved.
doi:10.1016/j.nima.2010.07.019
2. Materials and methods narrow strips were connected together to form 4 mm wide anodes
providing a 4 4 mm2 readout pixel size. A 12 13 array of strips was
2.1. Detector setup, electronic readout read out using gated integrator cards, developed at the Indiana
University Cyclotron Facility (IUCF), and a VME-based data
The detector, schematically shown in Fig. 1, consists of a double- acquisition system. During the beam measurements, the GEM 1 and
GEM amplification structure with a copper-clad Kapton cathode and a GEM2 were operated at 350 and 340 V, respectively, with the drift,
transfer and induction fields set, respectively, at 1.5, 1.7 and 1.7
recycling integrator card. The detector volume was continuously 2.3. Experimental setup
flushed with an Ar/CF4 (95/5%) gas mixture. Both GEMs were
operated at 300 V, with the same drift and transfer fields as described in
Section 2.1. The light produced by the electron avalanches was
detected by an SBIG ST-6 astronomical camera with thermoelectric
cooling from the Santa Barbara Instrument Group [10] and a Tamron
CCTV CS zoom lens. The camera was positioned away from the beam
and was shielded with lead and concrete blocks to reduce radiation
damage to the CCD sensor. The light path was enclosed in light-tight
shielding made from a black aluminum foil. The Texas Instruments
TC-241 CCD sensor, with a quantum efficiency of 62% at 650 nm, is
well matched to the emission spectrum of the Ar/CF 4 gas mixture [11].
The sensor has 375 241 pixels with dimensions of 23 27 mm2, which
translates to 0.36 0.42 mm 2 at the GEM2 location. During the
measurements, the camera was cooled to 30 1C. The images were read
out by the camera’s native SBIG software and analyzed using SBIG
[10] and ImageJ [12] software packages.
The detector was tested in the IUCF Proton Dosimetry Test Facility
with a 205MeV proton beam. The experimental setup is shown in Fig. 3.
The Ionization Chamber Beam Monitor (ICBM) was used to measure
beam intensity and beam profile, and to control the beam delivery,
terminating the beam upon the delivery of a pre-set dose. The dose rate
was varied by changing the beam current in the cyclotron. A 2.4mm thick
436 A.V. Klyachko et al. / Nuclear Instruments and Methods in Physics Research A 628 (2011) 434–439
Fig. 3. Schematics of the experimental setup. 1—ICBM; 2—scatter foil; 3—collimator; 4—acrylic phantom; 5—GEM detector; 6—Markus chamber; 7—therapy dosimeter/
copper scatter foil spreaded the beam to about 6cm diameter FWHM at the determine the light yield by integrating over an area of interest identical
detector location, also degrading the beam energy to 198MeV. For for all measurements.
depthdose measurements (see Section 3), the thickness of an acrylic picoampermeter; 8—VME DAQ; 9—mirror; 10—CCD camera.
phantom placed in front of the detector was varied by adding/ removing
acrylic sheets, with 1–10mm steps. The dose linearity tests were carried
out without any phantom material. Brass collimators were used to shape
the beam impinging on the detector. A commercial Markus ionization
chamber with NISTtraceable calibration in absorbed dose to water (PTW,
Model TN23343) with a 5.3mm diameter active area, positioned behind
the detector, was used to estimate the dose rate.
In the electronic readout mode, the main sources of background are the
noise pickup in the detector and in the cables. The background was
measured before each beam measurement and subtracted from the data. In
the optical readout mode, the main background sources are: (a) camera
offset; (b) dark current (noise); (c) ambient light. To take into account the
backgrounds (a–c), an ambient light image was taken without beam and
with the same exposure time as in beam measurements, and then
subtracted from the images taken with the beam. Corrections for additional
possible backgrounds (such as interactions of scattered beam and
secondary particles with the camera’s sensor, and scintillations in the
detector gas and exit window caused by beam particles) were estimated at
less than 1% total [5] and therefore were not applied. The images have
been processed offline to correct for background and to remove extremely Fig. 4. Dose-rate response of the detector in electronic readout mode. The lines are
linear fits to the data below 40 Gy/min.
hot and cold isolated pixels, using the routines provided with the SBIG
camera. The images then have been analyzed with ImageJ software to
Fig. 5. Depth-dose response of the detector in electronic readout mode. Insert:
expanded view of the Bragg peak.
3. Results and discussion Bragg peaks from detector and Markus chamber. In future, we intend to
increase the readout channel count and to use a wide aperture parallel
In the electronic readout mode, to estimate the position resolution plate ionization chamber instead of the Markus chamber for
of the detector, we performed series of measurements with different comparative measurements.
diameter collimators. The results obtained in the proton beam and with In the optical readout mode, collimators of different shapes were used
radioactive sources suggest that the spatial resolution is close to the to estimate the spatial resolution of the detector. An image of a 1.4 20
single-pixel size, i.e. 4 mm. The dose-rate response has been measured mm2 collimator obtained with 5 nA beam and 3 s exposure is shown in Fig.
with a 20 mm diameter collimator. The summed response of three X- 6 (insert) together with the light
and three Y-strips corresponding to the central part of the beam is Fig. 6. The 1.4 20 mm2 collimator image (insert) and the light intensity profile along the
one-pixel wide horizontal line. The data are fitted with a superposition of 3 Gaussians.
shown in Fig. 4.
To evaluate the energy dependence of the detector’s output, we
measured the depth-dose response of the detector, with the beam
current of about 6 nA corresponding to dose rate of 12Gy=min without
any material in the phantom, and up to 40Gy=min at the Bragg peak.
The depth-dose curves (summed response of six X- and six Y-strips)
are shown in Fig. 5. The curves are normalized to 1 at zero phantom
thickness. The acrylic thickness is converted to water-equivalent depth.
Note that the detector and the Markus chamber have sensitive areas of
different size and both detect only a small (and different) portion of all
particles coming out of the phantom, so one should not expect identical
438 A.V. Klyachko et al. / Nuclear Instruments and Methods in Physics Research A 628 (2011) 434–439
A.V. Klyachko et al. / Nuclear Instruments and Methods in Physics Research A 628 (2011) 434–439 440
Fig. 8. Dose rate response of the detector in optical readout mode. The line is a linear fit to
the data below 55 Gy/min.
intensity profile along a single-pixel line. The main peak has been fitted were taken with 3 s exposures, shorter than the duration of the beam
with a Gaussian curve with s¼ 0:7mm. An image of a 20 mm diameter delivery. The depth-dose measurements in the optical readout mode were
collimator is shown in Fig. 7 as well as a light intensity profile along a carried out with a 50 mm diameter collimator and a beam current of 5 nA.
one-pixel wide line (insert). We attribute the 3% variations of intensity on Depth-dose curves, shown in Fig. 9, have been normalized to 1 at zero
the flattop to GEM gain nonuniformities caused by hole size variations. depth. At the peak, the charge signal exceeds the light signal by about
The detector dose-rate response in optical readout mode, measured with 3.5%. One of the reasons for this might be the difference of the integration
the same 20 mm diameter collimator, is shown in Fig. 8. Beam images areas for charge and light signals.
4. Conclusions
Acknowledgements
References